| | Class 2 Device Recall VITEK(R) 2 ASTP640 |  |
| Date Initiated by Firm | August 13, 2018 |
| Create Date | September 15, 2018 |
| Recall Status1 |
Terminated 3 on November 02, 2020 |
| Recall Number | Z-3179-2018 |
| Recall Event ID |
80860 |
| Product Classification |
Enzyme immunoassay, gentamicin - Product Code LCD
|
| Product | VITEK(R) 2 AST-P640, For susceptibility testing of Staphylococcus spp. Enterococcus spp., ad S agalactiae against specified antimicrobials, REF 418579
Product Usage:
VITEK 2 is an automated system consisting of instruments, software and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast. The VITEK 2 utilizes growth-based biochemical patterns to determine identification. The VITEK 2 provides minimal inhibitory concentration (MIC) results for most organism/drug combinations as well as a category interpretation (S, I, or R) consistent with accepted criteria. |
| Code Information |
Lot Numbers: 7400410103, 7400697403 |
Recalling Firm/ Manufacturer |
bioMerieux, Inc. 100 Rodolphe St Durham NC 27712-9402
|
| For Additional Information Contact | 919-620-2000 |
Manufacturer Reason for Recall | Routine internal review of Device History Records identified six VITEK(R) 2 AST card lots were manufactured with less drug (gentamicin) than required for the Gentamicin High-Level Resistance (ghlr01n) screen test due to an error when making 1500mL and 2000mL media well preparations (solutions). |
FDA Determined Cause 2 | Process control |
| Action | The firm initiated the recall by email on 08/13/2018. The firm requested the consignee discard the impacted product and contact the local bioMerieux representative for credit. |
| Quantity in Commerce | 2538 cartons |
| Distribution | Worldwide Distribution US Nationwide and the countries of: Turkey, Austria, Switzerland, Czech Republic, Germany, Netherlands, |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|