 Class 2 Device Recall CAPNOSTREAM 20P (INTL) N ROHSsee related information
Date Initiated by FirmJuly 31, 2018
Create DateOctober 03, 2018
Recall Status1 Terminated 3 on September 20, 2022
Recall NumberZ-0038-2019
Recall Event ID 80895
Product Classification Oximeter - Product Code DQA
ProductCAPNOSTREAM 20P (INTL) N - ROHS, Part Number CS08652-02
Code Information B836141168 B836141196 B836141198 B836141206 B836141207 B836141209 B836141210 B836141212 B836141214 B836141217 B836141218 B836141221 B836141222 B836141223 B836141225 B836141226 B836141229 B836141230 B837141231 B837141232 B839141793 B840141875 B839141678 B839141682 B839141712 B841142027 B841142029 B841142033 B843142241 B844142389 B844142406 B844142411 B844142423 B844142426 B844142443 B844142449 B847142891 B847142906 B847142923 B847142938 B847142945 B847142947 B848142952 B848142970 B848142994 B848143001 B848143002 B801153800 B803153963 B803153964 B803153965 B803153968 B803153970 B803153971 B803153974 B803153975 B803153979 B803153981 B803153982 B803153983 B803153984 B803153985 B803153990 B841142053 B841142067 B851143673 B852143682 B850143423 B850143424 B850143425 B850143449 B850143450 B852143708 B852143715 B852143719 B852143736 B852143738 B852143743 B852143747 B852143750 B852143751 B852143755 B852143756 B852143757 B852143758 B852143760 B852143761 B852143762 B852143763 B852143766 B852143767 B852143768 B852143770 B852143772 B852143773 B852143776 B852143777 B852143778 B852143780 B852143782 B852143783 B852143785 B852143786 B852143787 B852143788 B852143792 B852143793 B852143794 B808154912 B808154964 B808155000 B809155003 B810155087 B810155099 B810155108 B810155117 B810155120 B810155144 B812155509 B812155546 B812155549 B812155551 B812155556 B812155559 B812155560 B812155566 B812155569 B812155586 B812155587 B812155592 B812155607 B812155609 B812155610 B809155067 B810155097 B810155195 B810155245 B810155246 B810155256 B810155261 B810155263 B810155264 B816156024 B816156043 B816156045 B816156047 B816156049 B816156051 B816156052 B816156061 B816156063 B816156064 B816156065 B816156069 B816156072 B816156075 B816156083 B803153993 B803154001 B803154036 B806154401 B806154424 B806154455 B807154583 B807154653 B807154718 B807154790 B808154968 B808154981 B809155031 B809155056 B810155076 B810155110 B810155125 B810155143 B810155171 B810155202 B810155220 B810155226 B810155228 B810155241 B810155259 B810155262 B811155417 B811155421 B811155459 B811155464 B811155472 B812155519 B812155605 B812155616 B812155735 B816156032 B816156070 B837141239 B837141272 B837141352 B837141371 B837141387 B842142183 B843142258 B843142297 B846142665 B846142666 B849143256 P835150025 P835150026 P835150027 P835150028 P835150029 P835150030 P835150031 P835150032 P835150033 P835150034 P835150035 P835150036 P835150037 P835150038 P835150039 P835150040 P835150041 P835150042 P835150043 P835150044 P835150045 P835150046 P835150047 P835150048 P844150433 P844150434 P844150435 P844150436 P844150437 P844150438 P844150439 P844150440 P844150441 P844150442 P844150443 P844150444 P844150445 P844150446 P844150447 P844150448 P844150449 P844150450 P844150451 P844150452 P844150453 P844150454 P844150455 P844150456 B805154325 B810155134 B848142973 B848143010 B851143491 P839163298 P839163316 P840163332 P840163381 P840163383 P840163384 P840163385 P840163386 P840163387 P840163388 P840163389 P840163390 P840163391 P840163392 P840163394 P840163395 P840163396 P840163397 P840163399 P840163400 P840163401 P840163415 P840163416 P840163422 P845164133 P845164151 P845164152 P845164153 P845164154 P845164156 P845164157 P845164158 P845164159 P845164160 P845164161 P845164162 P845164163 P845164165 P845164166 P845164167 P845164169 P845164170 P845164172 P845164173 P845164175 P845164177 P845164178 P845164179 P845164180 P845164182 P845164183 P845164185 P845164186 P845164187 P845164188 P845164189 P845164190 P845164191 P845164192 P845164193 P845164194 P845164195 P845164199 P845164201 P845164202 P845164203 P845164204 P845164207 P845164210 P845164211 P845164212 P845164213 P845164144 P845164155 P845164168 P845164176 P845164184 P845164198 P845164205 P845164208 P845164209 P845164214 P845164215 P845164217 P845164218 P845164219 P845164220 P845164221 P845164224 P845164225 P845164226 P846164402 P846164405 P846164406 P846164409 P846164413 P802175941 P802175943 P802175944 P802175947 P802175948 P802175949 P802175950 P802175951 P802175952 P802175953 P802175954 P802175956 P802175957 P802175959 P802175960 P802175961 P802175963 P802175964 P802175966 P802175967 P802175968 P802175969 P802175970 P802175981 P802175918 P804176135 P804176186 P804176187 P804176193 P804176204 P804176212 P804176213 P804176215 P804176216 P804176219 P804176220 P804176221 P804176224 P804176225 P804176226 P804176227 P804176228 P804176229 P804176230 P804176231 P804176233 P804176234 P804176236 P804176239 P804176240 P804176241 P804176242 P804176243 P804176244 P804176246 P804176247 P804176248 P804176249 P804176250 P805176251 P805176252 P805176253 P805176255 P805176256 P805176257 P805176258 P805176259 P805176260 P805176261 P805176262 P805176263 P805176264 P805176297 P806176343 P806176399 P806176401 P806176402 P806176403 P806176404 P806176405 P806176406 P806176407 P806176408 P806176409 P806176410 P806176411 P806176412 P806176413 P806176414 P806176416 P806176417 P806176418 P806176419 P806176420 P806176421 P806176422 P806176423 P806176425 P806176426 P806176427 P806176428 P806176429 P806176430 P806176431 P806176432 P806176433 P806176434 P806176435 P806176436 P806176437 P806176438 P806176440 P806176441 P806176442 P806176443 P806176444 P806176446 P806176447 P806176448 P806176478 
Recalling Firm/
Manufacturer
Oridion Medical 1987 Ltd.
Har Hotzvim Industrial Park Box 45025
Har Hotzvim Industrial Park
Jerusalem Israel
Manufacturer Reason
for Recall
The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory default settings when the monitor is powered off.
FDA Determined
Cause 2
Nonconforming Material/Component
ActionUrgent Medical Device Correction notification letters dated 7/31/18 were distributed to customers. The letters included the following instructions for customers: Share this notification with all care environments where the Capnostream" 20 and Capnostream" 20p bedside patient monitors are used, particularly in the areas where nurse call may be enabled (general care floors). If your facility has distributed these bedside patient monitors to other persons or facilities, please promptly forward a copy of this letter to those recipients. Medtronic is recommending continued use of Capnostream" devices. Users should confirm that the date/time is accurately displayed. If the date/time requires reset, all user defined settings, such as alarms and nurse call, will also require reset until the software is updated. Please follow the instruction in the Operators Manual to set user-defined default settings. The Operators Manual is available at http://www.medtronic.com/covidien/enus/support/product-manuals.html, refer to the Institutional Settings section. Return the completed Acknowledgement and Receipt Form by fax or email even if you have no inventory. In October 2018, download the software update from the Medtronic website via the following link which includes directions on how to download and install the software: http://www.medtronic.com/covidien/en-us/support/patient-monitoring-equipmentsoftware-upgrades.html.
Quantity in Commerce474
DistributionThe products were distributed US nationwide and Puerto Rico. The products were distributed to the following foreign countries: Argentina, Australia, Austria, Bahrain, Belgium, Canada, Columbia, Costa Rica, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, Ireland, Israel, Italy, Japan, Kazakhstan, Kuwait, Mexico, Netherlands, Norway, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, UAE, UK, and Vietnam.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.