• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall CHS Custom Convenience Kit

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall CHS Custom Convenience Kitsee related information
Date Initiated by FirmAugust 13, 2018
Create DateOctober 25, 2018
Recall Status1 Terminated 3 on August 01, 2022
Recall NumberZ-0256-2019
Recall Event ID 81179
Product Classification General surgery tray - Product Code LRO
ProductCHS Custom Convenience Kit-ROCHESTER PEAN, CURVED (8") Product Number:D36-16669
Code Information Lot number: 52418  Exp. Date: 5/1/2021
FEI Number 1121091
Recalling Firm/
Manufacturer
Custom Healthcare Systems, Inc.
4205 Eubank Rd
Richmond VA 23231-4328
Manufacturer Reason
for Recall
Convenience kits were sealed in immediate Tyvek pouches using incorrect specifications compromising product sterility
FDA Determined
Cause 2
Packaging
ActionCustom Healthcare Systems (CHS) initiated recall by phone, email and by certified mail on 8/13/18 and carried out to the end user. Custom Healthcare Systems Inc. notified the customers, they were informed to contact their account to receive the product. Product to be returned and credited.
Quantity in Commerce50 units
DistributionConnecticut, Georgia, Kentucky, North Carolina, Virginia, Maryland, Oklahoma, Illinois, South Carolina,
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
-
-