| Date Initiated by Firm | November 20, 2018 |
| Date Posted | May 15, 2019 |
| Recall Status1 |
Terminated 3 on July 09, 2020 |
| Recall Number | Z-1350-2019 |
| Recall Event ID |
81610 |
| 510(K)Number | K113598 |
| Product Classification |
Stretcher, wheeled - Product Code FPO
|
| Product | Power-LOAD Cot Fastening System, Model #6390, Catalog #6390-000-000 |
| Code Information |
170842168 170842169 170842170 170140459 170340545 170340546 170441816 170639347 170741449 170841501 170740752 170340455 170440451 170340235 170639548 170839227 170539929 170539931 170339830 170541854 170240183 170240184 170639519 170639520 170841536 170841537 170641385 170440640 170740758 170140879 170839507 170541048 170540760 170641358 170140234 170642299 170541867 170340201 170340202 170439365 170441803 170441804 170441805 170441806 170441807 170441808 170441809 170441810 170441811 170441812 170642291 170642292 170642293 170642294 170641050 170641051 170641052 170641053 170641396 170641397 170641398 170641399 170640894 170640895 170640896 170640897 170741400 170741401 170741402 170741403 170741426 170741427 170741428 170741429 170739492 170739493 170739494 170739495 170839283 170839285 170839286 170841884 170841885 170841886 170841887 170841543 170841544 170841545 170841546 170841538 170841539 170841540 170841541 170839223 170839224 170839225 170839226 170839284 170140880 170340850 170440458 170842145 170240232 170542128 170139635 170241755 170241756 170241773 170240234 170540658 170639587 170340190 170139344 170139345 170339717 170641742 170540642 170440667 170440680 170440681 170839233 170839234 170839235 170839236 170839237 170839238 170839239 170241596 170439382 170439381 170140696 170439691 170439692 170439357 170439358 170439372 170440008 170440453 170540693 170542013 170639583 170641071 170741430 170739550 170841547 170640800 170640899 170741437 170839287 170839288 170340457 170340748 170139637 170139638 170139639 170740838 170139699 170339002 170839271 170340197 170139343 170841511 170140882 170140898 170839279 170440497 170739573 170542126 170140237 170139622 170241848 170839534 170339718 170240456 170839533 170240235 170440498 170739485 170739486 170739487 170439650 170439651 170740150 170540799 170540659 170839251 170440099 170841523 170839511 170539925 170539926 170539927 170541750 170541751 170140233 170839299 170839293 170839294 170839295 170839296 170839297 170839298 170539720 170139705 170741435 170741436 170141221 170141222 170141223 170841526 170841527 170240180 170240181 170240182 170241747 170339837 170439698 170539994 170642250 170641743 170741410 170839266 170839200 170839201 170839206 170839202 170839203 170839204 170839205 170741500 170440058 170440607 170440609 170841528 170841529 170140793 170339832 170541887 170440687 170841509 170841850 170841878 170340222 170641376 170641377 170540694 170540695 170540696 170440608 170539933 170340839 170440004 170440005 170539940 170440414 170440668 170440669 170440670 170141249 170141292 170241780 170240222 170441817 170441822 170639547 170841860 170841861 170841864 170541742 170741417 170340853 170641382 170842163 170842164 170842165 170842166 170340226 170439378 170642148 170642149 170540656 170439193 170440092 170542141 170542142 170542143 170542144 170542145 170542146 170340742 170841877 170539479 170439694 170839281 170839282 170439385 170439386 170439387 170439388 170439389 170439390 170839548 170439362 170139636 170240188 170240189 170240190 170240191 170439695 170340854 170340837 170339828 170340223 170440606 170739554 170541871 170541872 170541873 170540660 170541882 170541883 170541884 170541885 170140243 170140471 170441925 170441926 170140884 170141248 170541859 170139712 170240226 170741404 170741405 170741406 170841522 170441813 170441814 170441815 170739553 170541733 170639543 170640803 170240794 170839273 170741163 170340182 170639584 170639585 170741432 170741433 170741434 170439145 170339834 170140645 170140699 170340221 170340225 170839517 170440057 170440445 170539947 170241354 170241355 170241356 170240230 170339391 170339392 170339835 170339836 170339839 170339840 170340052 170139346 170340192 170841891 170441818 170441819 170441820 170441821 170441823 170440439 170439363 170440459 170440460 170539476 170542028 170839276 170839277 170839278 170440635 170642144 170642145 170642143 170440407 170440408 170839280 170839531 170440682 170839547 170240233 170140883 170241353 170340859 170641745 170641746 170541862 170539978 170639546 170740760 170340491 170841549 170839514 170139600 170340834 170340835 170540697 170540698 170141250 170140895 170140239 170239951 170841534 170541874 170541875 170541876 170741423 170740840 170140874 170240223 170440440 170440441 170440442 170741546 170741547 170741548 170741549 170440050 170440051 170440052 1705042027 170240875 170240876 170340456 170240199 170440636 170141211 170141212 170240877 170839500 170839501 170839502 170340453 170340454 170540796 170540797 170541738 170541739 170639517 170639518 170842171 170842172 170842173 170842174 170239950 170639337 170241393 170741543 170639515 170639516 170439148 170439149 170539120 170539121 170539122 170439141 170540649 170141218 170139633 170139634 170539977 170440637 170141260 170841882 170340494 170440678 170440679 170541863 170139708 170541886 170439687 170439688 170842100 170540643 170841524 170741522 170340784 170340785 170340786 170340237 170340238 170340450 170241772 170241358 170340200 170741544 170541857 170541865 170439375 170841862 170839255 170839256 170440097 170440098 170739576 170741531 170140465 170140466 170140467 170140468 170139704 170539475 171741431 170839231 170839232 170741419 170541868 170539983 170539984 170140899 170539993 170340229 170541726 170340782 170340783 170239357 170139603 170139604 170340856 170340858 170339000 170339001 170139609 170140238 170139640 170541869 170639321 170639322 170639323 170639324 170639325 170639326 170240797 170440044 170741443 170741444 170741445 170540661 170539719 170339716 170842161 170440009 170440010 170342144 170439146 170539992 170241350 170841530 170440604 170839513 170339719 170241754 170440655 170440656 170440657 170440658 170440659 170542137 170542127 170240176 170440492 170440493 170440494 170440495 170440496 170240451 170340199 170440405 170440406 170839207 170139605 170140795 170440413 170241395 170841883 170639549 170740186 170240406 170839268 170541047 170739478 170541049 170541877 170541878 170839258 170841880 170140232 170539924 170241775 170539729 170541753 170541754 170139706 170240220 170639344 170639345 170439693 170841871 170340826 170841505 170541752 170641069 170741424 170440403 170440404 170340753 170340547 170339715 170140870 170139641 170440400 170241592 170839099 170241752 170140476 170340183 170340184 170340185 170240881 170240882 170440090 170440091 170639333 170639334 170639545 170339833 170440692 170440409 170340836 170241746 170339709 170440047 170539949 170441919 170441920 170441921 170441922 170441923 170139629 170641390 170439360 170240224 170240225 170740750 170340195 170741412 170741413 170741414 170741415 170741416 170141246 170141247 170841881 170639500 170841858 170841859 170639588 170639589 170639591 170639592 170639593 170639594 170639595 170639596 170639597 170639598 170139349 170540650 170240231 170539990 170539991 170741530 170339822 170839267 170240221 170140871 170639544 170642050 170740182 170740183 170740184 170339394 170141213 170440660 170539982 170841866 170841869 170439689 170839289 170339720 170241765 170740847 170739450 170739463 170140298 170842142 170841521 170240177 170440046 170539745 170539746 170539747 170539744 170241847 170440638 170842136 170740846 170340827 170240192 170541736 170140231 170240408 170240410 170841500 170841872 170841889 170841890 170839530 170440449 170842178 170641356 170339004 170140700 170140701 170140702 170140703 170140704 170140705 170140706 170140707 170140708 170140709 170140710 170140711 170140712 170140713 170140714 170140715 170140716 170140717 170140718 170140719 170140720 170140721 170140722 170140723 170140724 170140725 170140726 170140727 170140728 170140729 170140730 170140731 170140732 170140733 170140734 170140735 170140736 170140737 170140738 170140739 170140740 170140741 170140742 170140743 170140744 170140745 170140746 170140747 170140748 170140749 170140750 170140751 170140752 170539718 170539996 170539997 170539998 170539999 170540750 170540751 170540752 170540753 170540754 170140240 170140241 170140242 170439366 170439367 170439368 170140300 170140893 170140472 170839506 170339348 170739560 170739561 170739562 170739563 170739564 170639300 170339848 170839259 170439138 170839274 170839275 170842177 170841851 170841852 170841853 170841854 170841855 170841856 170842151 170842152 170539119 170842146 170842147 170139631 170440045 170340198 170741161 170239360 170641074 170641075 170641076 170140296 170339710 170339711 170339712 170339713 170140470 170841506 170841507 170841508 170839542 170839543 170241782 170642110 170539937 170440455 170741534 170741535 170739568 170741545 170139617 170340054 170340241 170340242 170340243 170340244 170340245 170340246 170340247 170340248 170340249 170340750 170139608 170140872 170339846 170339847 170240428 170439652 170340752 170740841 170140469 170339723 170340231 170539995 170540759 170241778 170439696 170439697 170542148 170741447 170241777 170240450 170141294 170339843 170440688 170440689 170440690 170440691 170241357 170641389 170741411 170440448 170839535 170839536 170839537 170839538 170241390 170241391 170241392 170841531 170739571 170640802 170440011 170440012 170440013 170440014 170440015 170440016 170440017 170440018 170440019 170440020 170440021 170440022 170440023 170839545 170439196 170439197 170439198 170439199 170141264 170141265 170141266 170842141 170841874 170740753 170440600 170440059 170839230 170240429 170340852 170541740 170839252 170241769 170841899 170841868 170640887 170640888 170640889 170640890 170640891 170640892 170439194 170741407 170741408 170741409 170739451 170739452 170739453 170739454 170739455 170739456 170739457 170739458 170739459 170739460 170741540 170842159 170541737 170440676 170342030 170342031 170340228 170639695 170639696 170639697 170639698 170841502 170841504 170641747 170241396 170139632 170140228 170240212 170441835 170540758 170241779 170440048 170639348 170340051 170539749 170841519 170242149 170740839 170839208 170841894 170841895 170841896 170841897 170841863 170841865 170441824 170441825 170441828 170441829 170441830 170441831 170139614 170139615 170139616 170842179 170339714 170539936 170539735 170839528 170440060 170440061 170440062 170440063 170440064 170440065 170440066 170340193 170642297 170141217 170440096 170839512 170140473 170139628 170139630 170241590 170539988 170641065 170240407 170440456 170741425 170340236 170739552 170239350 170239351 170239353 170239354 170839262 170839263 170239356 170239355 170240178 170240179 170339396 170339397 170140888 170140889 170140890 170140891 170140892 170642146 170139347 170139348 170139700 170139701 170140642 170140643 170140644 170240409 170439369 170439370 170439371 170440693 170440694 170539741 170539742 170539743 170740198 170739464 170739465 170542100 170839529 170241047 170241048 170542136 170241781 170641384 170340754 170340755 170739466 170440053 170839272 170640801 170140249 170439142 170440454 170541864 170841548 170741541 170440412 170841533 170139709 170439391 170740842 170741418 170241757 170241758 170439140 170842175 170842176 170741536 170340493 170740193 170740194 170740195 170839508 170839509 170839510 170340492 170141210 170140474 170140475 170241776 170240219 170439136 170839210 170440003 170639691 170639692 170639693 170639694 170340053 170641073 170741152 170741153 170741154 170741155 170741156 170741157 170741158 170741159 170741160 170439235 170439236 170439237 170439238 170439239 170439240 170439241 170439242 170439243 170439244 170439245 170439246 170439247 170439248 170542000 170339749 170542149 170841535 170540662 170539748 170842154 170842155 170842156 170842157 170842158 170641072 170541734 170140877 170340788 170639335 170539980 170340234 170340825 170340495 170340496 170340497 170440447 170240227 170240228 170340840 170340841 170340191 170839269 170839270 170540654 170739569 170739570 170440601 170541855 170439147 170139649 170740185 170241751 170241846 170741448 170140697 170140698 170639699 170539981 170842133 170842134 170140246 170540655 170141244 170739551 170340855 170739577 170439374 170240236 170739461 170739462 170241599 170139707 170441928 170739567 170541879 170839260 170140640 170842167 170141258 170141259 170739556 170539728 170440450 170842143 170539979 170240237 170240238 170240239 170240240 170240241 170340746 170542030 170542031 170542032 170542033 170542034 170542035 170542036 170542037 170542038 170542039 170542040 170542041 170542042 170542043 170542044 170542045 170542046 170542047 170542048 170542049 170540646 170140460 170140461 170140462 170140463 170140464 170241049 170642141 170140247 170539948 170239352 170839264 170441927 170839544 170240229 170639329 170639330 170639331 170639332 170440639 170740199 170240198 170139607 170841542 170140878 170541861 170339325 170340499 170641387 170839546 170140299 170440457 170642295 170642296 170139642 170139643 170139644 170139645 170139646 170139647 170139648 170440683 170440684 170440685 170342028 170339821 170841873 170740751 170641741 170842149 170140885 170839265 170540647 170540648 170641057 170641058 170641059 170641060 170641061 170641062 170641063 170641064 170340749 170439137 170439359 170841876 170140293 170140294 170541741 170440446 170241766 170841525 170141251 170240454 170240452 170839228 170339824 170339825 170339826 170339827 170539934 170539938 170740757 170241750 170539939 170439134 170539477 170539478 170539985 170539986 170339726 170340224 170439380 170440056 170440632 170440641 170440651 170440652 170440653 170539946 170539950 170539976 170642142 170740754 170740755 170740756 170542147 170842101 170842102 170842103 170842104 170842105 170842106 170842107 170842108 170842109 170842110 170842111 170842112 170842113 170842114 170842115 170842116 170842117 170842118 170842119 170842120 170842121 170842122 170842123 170842124 170842125 170842126 170842127 170842128 170842129 170839532 170839516 170841893 170139702 170139703 170340745 170841875 170640893 170241759 170241760 170641388 170541880 170739572 170639586 170542029 170140230 170241771 170339838 170740849 170540792 170540793 170540794 170540795 170339845 170839257 170241849 170641357 170739583 170641391 170641392 170641393 170641394 170641395 170741450 170641354 170441832 170441833 170441834 170441836 170439383 170439384 170542012 170141262 170439139 170140897 170842130 170539987 170741533 170539989 170339399 170140235 170140236 170241397 170542138 170740848 170841888 170741524 170741537 170539932 170339844 170340196 170340756 170439699 170440000 170440001 170440002 170340787 170842144 170741539 170740196 170340194 170841898 170540644 170540645 170541860 170139610 170139611 170740759 170340050 170339724 170339725 170440605 170439690 170340833 170241774 170240455 170739559 170541858 170141263 170839250 170140297 170839541 170140875 170140881 170340828 170541866 170641383 170842139 170340240 170839229 170539485 170440452 170340451 170340452 170842180 170839515 170841892 170140290 170140291 170140292 170839261 170241598 170641055 170641056 170440461 170639327 170639328 170541881 170241352 170241591 170741421 170741422 170339722 170440094 170440095 170739578 170140638 170340747 170839254 170839253 170340186 170340187 170340188 170340189 170641378 170641379 170641380 170140876 170240458 170240459 170240460 170240461 170240462 170240463 170241753 170741529 170439144 170542011 170439376 170439377 170842153 170439133 170639541 170639542 170240750 170440602 170440603 170541735 170140639 170441837 170539921 170240753 170439195 170240883 170140245 170140646 170140647 170339390 170339393 170241770 170539941 170539942 170539943 170539944 170539945 170841867 170339831 170641350 170641351 170641352 170641353 170441827 170642109 170542129 170439135 170741521 170441826 170639650 170239358 170239359 170540755 170540756 170841520 170139711 170739555 170140873 170439249 170739581 170739582 170340448 170340449 170839503 170839504 170839505 170340549 170539928 170841510 170839209 170841532 170741420 170140886 170540651 170741523 170239952 170739575 170141215 170141216 170641355 170140248 170240457 170140896 170141293 170440054 170440055 170739491 170140794 170739477 170340227 170439678 170439679 170439680 170439681 170439682 170439683 170439684 170439685 170439686 170339829 170340857 170140894 170240411 170639346 170542014 170542015 170542016 170439373 170639336 170842162 170541850 170541851 170740189 170740190 170740191 170541870 170339708 170340230 170841879 170139606 170539923 170139618 170139619 170139620 170139621 170141219 170241394 170141220 170641077 170641078 170641079 170140458 170842160 170641054 170141261 170739558 170440677 170441924 170739488 170739489 170739490 170140229 170339349 170439180 170539483 170539484 170340851 170140295 170241597 170641744 170340751 170741528 170640804 170141245 170340741 170140086 170140087 170140088 170140089 170139612 170139613 170440686 170841870 170339395 170339398 170339721 170439379 170739579 170842148 170439364 170540798 170739557 170139710 170739565 170739566 170642298 170842150 170340829 170841512 170841513 170841514 170841515 170841516 170841517 170841518 170139601 170141214 170641700 170640898 170241388 170241389 170842131 170540652 170540653 170641386 170440087 170440093 170741542 170540757 170839539 170839540 170241767 170541727 170541728 170440088 170440089 170439143 170440006 170440007 170741150 170440443 170440444 170140887 170339823 170339849 170440438 170539922 170539935 170740844 170839291 170839292 170440499 170340744 170839290 170241748 170140648 170140649 170740843 170240751 170240752 170440633 170440634 170841503 170741538 170741438 170241768 170340232 170741532 170240796 170741162 170241362 170240213 170642147 170641381 170340838 170139602 170440049 170642108 170240453 170340548 170340498 170540657 170539721 170539722 170539723 170539724 170539725 170539726 170539727 170539730 170539731 170539732 170539733 170539734 170539736 170539737 170539738 170539739 170539740 170639590 170842132 170842135 170842137 170842138 170842140 170539930 170440654 170140641 170340743 170340870 170541729 170541730 170541731 170541732 170741439 170741440 170340233 161241400 161241401 161241402 161241403 161241404 161241405 161241406 161241407 161241408 161241409 161241410 161241411 161241412 161241413 161241414 161241415 161241416 161241417 161241418 161241419 161241420 161241421 161241422 161241423 161241424 161241425 161241426 161241427 161241428 161241429 161241430 161241431 161241432 161241433 161241434 161241435 161241436 161241437 161241438 161241439 161241440 161241441 161241442 161241443 161241444 161241445 161241446 161241447 161241448 161241449 161241450 161241451 161241452 170141750 170141751 170141752 170141753 170141754 170141755 170141756 170141757 170141758 170141759 170141619 170141620 170141621 170141622 170141623 170141760 170141761 170141762 170141763 170141764 170141765 170141766 170141767 170141768 170141769 170239361 170239362 170239363 170239364 170239365 170239366 170239367 170239368 170239369 170239370 170239371 170239372 170239373 170239374 170239375 170239376 170239377 170239378 170239379 170239380 170242172 170242173 170242174 170242175 170242176 170242177 170242178 170242179 170242180 170242181 170242182 170242183 170242184 170242185 170242186 170242187 170242188 170242189 170242190 170242191 170241550 170241551 170241552 170241553 170241554 170241555 170241556 170241557 170241558 170241559 170241560 170241561 170241562 170241563 170241564 170241565 170241566 170241567 170241568 170241569 170241200 170241201 170241202 170241203 170241204 170241205 170241206 170241207 170241208 170241209 170241210 170241211 170241212 170241213 170241214 170241215 170241216 170241217 170241218 170241219 170240754 170240755 170240756 170240757 170240758 170240759 170240760 170240761 170240762 170240763 170240764 170240765 170240766 170240767 170240768 170240769 170240770 170240771 170240772 170240773 170240774 170240775 170240776 170240777 170240778 170240779 170240780 170240781 170240782 170240783 170240784 170240785 170240786 170240787 170240788 170240789 170240790 170240791 170240792 170240793 170240560 170240561 170240562 170240563 170240564 170240565 170240566 170240567 170240568 170240569 170240570 170240571 170240572 170240573 170240574 170240575 170240576 170240577 170240578 170240579 170240580 170240581 170240582 170240583 170240584 170240585 170240586 170240587 170240588 170240589 170240590 170240591 170240592 170240593 170240594 170240595 170240596 170240597 170240598 170240599 170240166 170240167 170240168 170240169 170240170 170240171 170240172 170240173 170240174 170240175 170339150 170339151 170339152 170339153 170339154 170339155 170339156 170339157 170339158 170339159 170339160 170339161 170339162 170339163 170339164 170339165 170339166 170339167 170339168 170339169 170339170 170339171 170339172 170339173 170339174 170339175 170339176 170339177 170339178 170339179 170339180 170339181 170339182 170339183 170339184 170339185 170339186 170339187 170339188 170339189 170342100 170342101 170342102 170342103 170342104 170342105 170342106 170342107 170342108 170342109 170342110 170342111 170342112 170342113 170342114 170342115 170342116 170342117 170342118 170342119 170342120 170342121 170342122 170342123 170342124 170342125 170342126 170342127 170342128 170342129 170342130 170342131 170342132 170342133 170342134 170342135 170342136 170342137 170342138 170342139 170339500 170339501 170339502 170339503 170339504 170339505 170339506 170339507 170339508 170339509 170339510 170339511 170339512 170339513 170339514 170339515 170339516 170339517 170339518 170339519 170339520 170339521 170339522 170339523 170339524 170339525 170339526 170339527 170339528 170339529 170339530 170339531 170339532 170339533 170339534 170339535 170339536 170339537 170339538 170339539 170340058 170340059 170340060 170340061 170340062 170340063 170340064 170340065 170340066 170340067 170340068 170340069 170340070 170340071 170340072 170340073 170340074 170340075 170340076 170340077 170340078 170340079 170340080 170340081 170340082 170439050 170439051 170439052 170439053 170439054 170439055 170439056 170439057 170439058 170439059 170439060 170439061 170439062 170439063 170439064 170439065 170439066 170439067 170439068 170439069 170439653 170439654 170439655 170439656 170439657 170439658 170439659 170441350 170441351 170441352 170441353 170441354 170441355 170441356 170441357 170441358 170441359 170441360 170441361 170441362 170441363 170441364 170441365 170441366 170441367 170441368 170441369 170441370 170441371 170441372 170441373 170441374 170441375 170441376 170441377 170441378 170441379 170441380 170441381 170441382 170441383 170441384 170441385 170441386 170441387 170441388 170441389 170441400 170441401 170441402 170441403 170441404 170441405 170441406 170441407 170441408 170441409 170441410 170441411 170441412 170441413 170441414 170441415 170441416 170441417 170441418 170441419 170539951 170539952 170539953 170539954 170539955 170539956 170539957 170539958 170539959 170539960 170539961 170539962 170539963 170539964 170539965 170539966 170539967 170539968 170539969 170539970 170539971 170539972 170539973 170539974 170539975 170542101 170542102 170542103 170542104 170542105 170542106 170542107 170542108 170542109 170542110 170542111 170542112 170542113 170542114 170542115 170542116 170542117 170542118 170542119 170542120 170542121 170542122 170542123 170542124 170542125 170542130 170542131 170542132 170542133 170542134 170542135 170541856 170639521 170639522 170639523 170639524 170639525 170639526 170639527 170639528 170639529 170639530 170639531 170639532 170639533 170639534 170639535 170639536 170639537 170639538 170639539 170639540 170639651 170639652 170639653 170639654 170639655 170639656 170639657 170639658 170639659 170639660 170639661 170639662 170639663 170639664 170639665 170639666 170639667 170639668 170639669 170639670 170639301 170639302 170639303 170639304 170639305 170639306 170639307 170639308 170639309 170639310 170639311 170639312 170639313 170639314 170639315 170639316 170639317 170639318 170639319 170639320 170639671 170639672 170639673 170639674 170639675 170639676 170639677 170639678 170639679 170639680 170642251 170642252 170642253 170642254 170642255 170642256 170642257 170642258 170642259 170642260 170642261 170642262 170642263 170642264 170642265 170642266 170642267 170642268 170642269 170642270 170639681 170639682 170639683 170639684 170639685 170639686 170639687 170639688 170639689 170639690 170642271 170642272 170642273 170642274 170642275 170642276 170642277 170642278 170642279 170642280 170642281 170642282 170642283 170642284 170642285 170642286 170642287 170642288 170642289 170642290 170642111 170642112 170642113 170642114 170642115 170642116 170642117 170642118 170642119 170642120 170642121 170642122 170642123 170642124 170642125 170642126 170642127 170642128 170642129 170642130 170641701 170641702 170641703 170641704 170641705 170641706 170641707 170641708 170641709 170641710 170641711 170641712 170641713 170641714 170641715 170641716 170641717 170641718 170641719 170641720 170641721 170641722 170641723 170641724 170641725 170641726 170641727 170641728 170641729 170641730 170641731 170641732 170641733 170641734 170641735 170641736 170641737 170641738 170641739 170641740 170642131 170642132 170642133 170642134 170642135 170642136 170642137 170642138 170642139 170642140 170640901 170640902 170640903 170640904 170640905 170640906 170640907 170640908 170640909 170640910 170640911 170640912 170640913 170640914 170640915 170640916 170640917 170640918 170640919 170640920 170640921 170640922 170640923 170640924 170640925 170640926 170640927 170640928 170640929 170640930 170640931 170640932 170640933 170640934 170640935 170640936 170640937 170640938 170640939 170640940 170740761 170740762 170740763 170740764 170740765 170740766 170740767 170740768 170740769 170740770 170740771 170740772 170740773 170740774 170740775 170740776 170740777 170740778 170740779 170740780 170740781 170740782 170740783 170740784 170740785 170740786 170740787 170740788 170740789 170740790 170740791 170740792 170740793 170740794 170740795 170740796 170740797 170740798 170740799 170740800 170740801 170740802 170740803 170740804 170740805 170740806 170740807 170740808 170740809 170740810 170740811 170740812 170740813 170740814 170740815 170740816 170740817 170740818 170740819 170740820 170740821 170740822 170740823 170740824 170740825 170740826 170740827 170740828 170740829 170740830 170740831 170740832 170740833 170740834 170740835 170740836 170740837 170740900 170740901 170740902 170740903 170740904 170740905 170740906 170740907 170740908 170740909 170740910 170740911 170740912 170740913 170740914 170740915 170740916 170740917 170740918 170740919 170740920 170740921 170740922 170740923 170740924 170740925 170740926 170740927 170740928 170740929 170740930 170740931 170740932 170740933 170740934 170740935 170740936 170740937 170740938 170740939 170740940 170740941 170740942 170740943 170740944 170740945 170740946 170740947 170740948 170740949 170739475 170739476 170739467 170739468 170739469 170739470 170739471 170739472 170739473 170739474 170741001 170741002 170741003 170741004 170741005 170741006 170741007 170741008 170741009 170741010 170741011 170741012 170741013 170741014 170741015 170741016 170741017 170741018 170741019 170741020 170139950 170139951 170139952 170139953 170139954 170139955 170139956 170139957 170139958 170139959 170139960 170139961 170139962 170139963 170139964 170139965 170139966 170139967 170139968 170139969 170139970 170139971 170139972 170139973 170139974 170139975 170139976 170139977 170139978 170139979 170139980 170139981 170139982 170139983 170139984 170139985 170139986 170139987 170139988 170139989 170139915 170139916 170139917 170139918 170139919 170139920 170139921 170139922 170139923 170139924 170139925 170139926 170139927 170139928 170139929 170139930 170139931 170139932 170139933 170139934 170139935 170139936 170140244 170139623 170139624 170139625 170139626 170139627 170140477 170140478 170140479 170140480 170140481 170140482 170140483 170140484 170140485 170140486 170140632 170140633 170140634 170140635 170140636 170140637 170141252 170141253 170141254 170141255 170141256 170141257 170239394 170239395 170239396 170239397 170239381 170239382 170239383 170239384 170239385 170239386 170239387 170239388 170239389 170239390 170239391 170239392 170239393 170242141 170242142 170242143 170242144 170242145 170242146 170242147 170242148 170242170 170242171 170239398 170239399 170241761 170241762 170241763 170241764 170241351 170241593 170241594 170241595 170240878 170240879 170240880 170240464 170240465 170240466 170240467 170240468 170240469 170240470 170240471 170240412 170240413 170240414 170240415 170240416 170240417 170240418 170240419 170240420 170240421 170240422 170240423 170240424 170240425 170240426 170240427 170240200 170240201 170240202 170240203 170240204 170240205 170240206 170240207 170240208 170240209 170240210 170240211 170240185 170240186 170240187 170240193 170240194 170240195 170240196 170240197 170340445 170340446 170340447 170340830 170340831 170340832 170439350 170439351 170439352 170439353 170439354 170439355 170439356 170439361 170539480 170539481 170539482 170540761 170540762 170540763 170540764 170540765 170540766 170540767 170540768 170540769 170540770 170540771 170540772 170540773 170540774 170540775 170540776 170540777 170540778 170540779 170540780 170540781 170540782 170540783 170540784 170540785 170540786 170540787 170540788 170540789 170540790 170540791 170542001
170542002 170542003 170542004 170542005 170542006 170542007 170542008 170542009 170542010 170539150 170539151 170539152 170539153 170539154 170539155 170539156 170539157 170639501 170639502 170639503 170639504 170639505 170639506 170639507 170639508 170639509 170639510 170639511 170639512 170639513 170639514 170639338 170639339 170639340 170639341 170639342 170639343 170641066 170641067 170641068 170641070 170740845 170741525 170741526 170741527 170741441 170741442 170740187 170740188 170739580 170740151 170740152 170740153 170740154 170740155 170740156 170740157 170740158 170740159 170740160 170740161 170740162 170740163 170740164 170740165 170740166 170740167 170740168 170740169 170740170 170740171 170740172 170740173 170740174 170740175 170740176 170740177 170740178 170740179 170740180 170740181 170839211 170839212 170839213 170839214 170839215 170839216 170839217 170839218 170839219 170839220 170839221 170839222 170840000 170840001 170840002 170840003 170840004 170840005 170840006 170840007 170840008 170840009 170840010 170840011 170840012 170840013 170840014 170840015 170840016 170840017 170840018 170840019 170840020 170840021 170840022 170840023 170840024 170840025 170840026 170840027 170840028 170840029 170840030 170840031 170840032 170840033 170840034 170840035 170840036 170840037 170840038 170840039 170840040 170840041 170840042 170840043 170840044 170840045 170840046 170840047 170840048 170840049 170839518 170839519 170839520 170839521 170839522 170839523 170839524 170839525 170839526 170839527 170839549 170240214 170240215 170240216 170240217 170240218 170340700 170340701 170340702 170340703 170340704 170340705 170340706 170340707 170340708 170340709 170340710 170340711 170340712 170340713 170340714 170340715 170340716 170340717 170340718 170340719 170340720 170340721 170340722 170340723 170340724 170340725 170340726 170340727 170340728 170340729 170439150 170439151 170439152 170439153 170439154 170439155 170439156 170439157 170439158 170439159 170439160 170439161 170439162 170439163 170439164 170439165 170439166 170439167 170439168 170439169 170439170 170439171 170439172 170439173 170439174 170439175 170439176 170439177 170439178 170439179 170440850 170440851 170440852 170440853 170440854 170440855 170440856 170440857 170440858 170440859 170440860 170440861 170440862 170440863 170440864 170440865 170440866 170440867 170440868 170440869 170440870 170440871 170440872 170440873 170440874 170440875 170440876 170440877 170440878 170440879 170541755 170541756 170541757 170541758 170541759 170541760 170541761 170541762 170541763 170541764 170541765 170541766 170541767 170541768 170541769 170541770 170541771 170541772 170541773 170541774 170541775 170541776 170541777 170541778 170541779 170541780 170541781 170541782 170541783 170541784 170641361 170641362 170641363 170641364 170641365 170641366 170641367 170641368 170641369 170641370 170641371 170641372 170641373 170641374 170641375 170741501 170741502 170741503 170741504 170741505 170741506 170741507 170741508 170741509 170741510 170741511 170741512 170741513 170741514 170741515 170741516 170741517 170741518 170741519 170741520 170739574 170440410 170440411 170741151 170740192 170340239 Updated as of 6/12/19: 170541743 170541744 |
Recalling Firm/ Manufacturer |
Stryker Medical Division of Stryker Corporation 3800 E Centre Ave Portage MI 49002-5826
|
| For Additional Information Contact | Victoria Haney 269-389-8306 |
Manufacturer Reason for Recall | The power load floor plates which hold the ambulance cot anchors in place during patient transport, may be
broken. |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | Notification letters were distributed to customers on 11/20/18. |
| Quantity in Commerce | 3285 |
| Distribution | US Nationwide and the following foreign countries: Australia, Canada, Netherlands, South Africa, and Switzerland. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = FPO
|