| Date Initiated by Firm | February 14, 2019 |
| Create Date | March 21, 2019 |
| Recall Status1 |
Terminated 3 on May 07, 2020 |
| Recall Number | Z-1033-2019 |
| Recall Event ID |
82228 |
| 510(K)Number | K170763 |
| Product Classification |
Mesh, surgical - Product Code FTM
|
| Product | Gentrix Surgical Matrix Thick, 20 x 30cm, Catalog Number PSMT2030, Lot 013948 - Product Usage: Gentrix Surgical Matrix Thick is intended for implantation to reinforce soft tissue where weakness exists in patients requiring gastroenterological or plastic and reconstructive surgery. Reinforcement of soft tissue within gastroenterological and plastic & reconstructive surgery includes, but is not limited to the following procedures: hernia and body wall repair, colon and rectal prolapse repair, tissue repair, and esophageal repair. |
| Code Information |
SN KJ000423 |
Recalling Firm/ Manufacturer |
ACell, Inc 6640 Eli Whitney Dr Ste 200 Columbia MD 21046-2626
|
| For Additional Information Contact | Thomas Gilbert PhD 410-715-1700 Ext. 8509 |
Manufacturer Reason for Recall | The devices were released from a lot for which one sample failed to meet the company' s tensile strength specification for lot release. |
FDA Determined Cause 2 | Nonconforming Material/Component |
| Action | Urgent Medical Device Recall notification letters dated 2/14/19 were sent to customers. |
| Quantity in Commerce | 1 |
| Distribution | US nationwide distribution in the state of CA. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = FTM
|