| Class 2 Device Recall VITROS XT 7600 Integrated System | |
Date Initiated by Firm | March 20, 2019 |
Create Date | May 20, 2019 |
Recall Status1 |
Terminated 3 on December 08, 2020 |
Recall Number | Z-1364-2019 |
Recall Event ID |
82561 |
510(K)Number | K182063 |
Product Classification |
Analyzer, chemistry (photometric, discrete), for clinical use - Product Code JJE
|
Product | VITROS XT 7600 Integrated System, with V3.4 or 3.4.1 SW
Product Code: 6844461
Product Usage:
INTENDED USE: For in vitro diagnostic use only. The VITROS XT 7600 Integrated System is intended for use in the measurement of a variety of analytes of clinical interest |
Code Information |
Serial Number 76000108 thru current. |
Recalling Firm/ Manufacturer |
Ortho-Clinical Diagnostics 100 Indigo Creek Dr Rochester NY 14626-5101
|
Manufacturer Reason for Recall | Potential for sample fluid to be dispensed to an incorrect position on the MicroSlide, potentially leading to erroneous assay results being reported |
FDA Determined Cause 2 | Software design |
Action | Ortho Clinical Diagnostics issued on 20 March 2019, a customer letter via FedEx overnight courier and/ or ORTHO PLUS e-Communications and and/or US Postal Service Priority Mail (for PO Boxes only) customers who were shipped affected VITROS XT 7600 Systems. Software Version 3.4.2 (MOD 4) will be provided for download or DVD format. Contact Ortho Care" Technical Solutions Center at 1-800-421-3311. |
Quantity in Commerce | 85 units: US=19;OUS=66 |
Distribution | Worldwide Distribution - US Nationwide in the states of AZ, CA, IL, OH, FL, TX, MI
Foreign:
Australia, Belgium, Chile, India, Italy, Japan, South Korea, Spain, Canada,, Phillippines |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = JJE
|
|
|
|