Date Initiated by Firm | October 22, 2018 |
Create Date | May 21, 2019 |
Recall Status1 |
Terminated 3 on August 25, 2022 |
Recall Number | Z-1279-2019 |
Recall Event ID |
82753 |
510(K)Number | K121792 |
Product Classification |
System, x-ray, tomography, computed - Product Code JAK
|
Product | NeuViz 64 Multi-slice CT Scanner System
Product Usage: The Multi-Slice CT Scanner System can be used as a whole body computed tomography x-ray system featuring a continuously rotating x-ray tube and detector array. The acquired x-ray
transmission data is reconstructed by computer into cross-sectional images of the body from either the same axial plane taken at different angles or spiral planes taken at different angles. |
Code Information |
N64E120022E; N64E130039E; N641130051 E; N64E130062E; N64E130063E; N64E130066E;1 N641140077E; N64E140116E; N641140117E; N641140118E; N641140119E; N641140120E; N64EN150043E; N64E150121E; N64E150125E; N641150129E; N641150131 E; N641150136E; N641N160066E; N641N160067E; N64EN160068E; N64EN160071 E; N64EN160082E; N64EN160085E; N64EN160086E; N641N1 70093E; N641N1701 05E; N641N1 701 16E; N641N170120E; N641N170127E; N641N170128E; N641N170129E; N641N180156E; N641N180157E; N641N180161E; N641N180162E; N641N180205E |
Recalling Firm/ Manufacturer |
Neusoft Medical Systems Co., Ltd. No. 177-1 Cuangxin Road Hunnan District Shenyang China
|
For Additional Information Contact | 024-23358311 |
Manufacturer Reason for Recall | Software defect: For Helical scan with ClearView function, when small arc (uninterrupted scan) occurs during scan,
there will be some electromagnetic interference in the system which may affect the control signal of the flying focal spot, resulting in an incorrect flying focal spot sequence. If the system does not receive error, it would continue to complete the scan with the incorrect flying focal spot sequence, having an effect on the reconstruction of some images, resulting in image loss.
|
FDA Determined Cause 2 | Software design |
Action | A customer notification letter will be sent to affected customers by email. All the affected systems will be corrected with the upgraded software in the field free of charge. |
Quantity in Commerce | 37 |
Distribution | US Nationwide Distribution. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = JAK
|