| | Class 2 Device Recall EXACTAMIX Empty EVA Bag |  |
| Date Initiated by Firm | September 18, 2019 |
| Create Date | October 21, 2019 |
| Recall Status1 |
Terminated 3 on February 16, 2021 |
| Recall Number | Z-0167-2020 |
| Recall Event ID |
83804 |
| 510(K)Number | K900585 |
| Product Classification |
Set, i.V. Fluid transfer - Product Code LHI
|
| Product | EXACTAMIX Empty EVA (Ethylene Vinyl Acetate) TPN Bag with the following product code and sizes:
1. H938737, 250mL;
2. H938738, 500mL;
3. H938739, 1000mL;
4. H938740, 2000mL;
5. H938741, 3000mL
Product Usage: The EXACTAMIX Empty EVA bags are used as the destination container in compounding of large-volume sterile solutions for further administration to a single patient. The empty bags are used at pharmacies and compounding centers by pharmacists and pharmacy technicians to compound multi-ingredient solutions, as prescribed by a physician. |
| Code Information |
1. UDI Number: 00085412477213, Lot Numbers (60060391, 60061809, 60061816, 60061819, 60063137, 60063138, 60063141, 60063142, 60063143, 60063146, 60063147, 60063148, 60064545, 60064546, 60064547, 60064548, 60064549, 60064551, 60064552, 60064553, 60064554, 60066253, 60066254, 60066255, 60066256, 60066257, 60066258, 60068559, 60070835, 60070836, 60070837, 60074866, 60074955, 60074956, 60074959, 60075125, 60075170, 60080371, 60080372, 60080373, 60080374, 60080375, 60080376, 60081617, 60081618, 60081621, 60081622, 60081649, 60081650, 60081655, 60082887, 60083516, 60083519, 60084461, 60084462, 60084465, 60084466, 60084471, 60084472, 60084475, 60085739, 60085740, 60085741, 60085742, 60085743, 60085744, 60085745, 60086979, 60086980, 60086981, 60086982, 60086985, 60086986, 60088180, 60088181, 60088182, 60088183, 60088185, 60088186, 60089784, 60089785, 60089786, 60089787, 60089788, 60089789, 60091137, 60091138, 60091139, 60091140, 60091141, 60091142, 60092384, 60092407, 60092408, 60092409, 60092410, 60092412, 60093834, 60093835, 60093836, 60093839, 60093841, 60095438, 60095445, 60095446, 60095447, 60096697, 60096698, 60096699, 60096700, 60096701, 60096702, 60097821, 60097822, 60097823, 60097824, 60097825, 60098859, 60098860, 60098861, 60098863, 60098864, 60099739, 60099740, 60099741, 60099742, 60099743, 60099744, 60099745, 60101085, 60101086, 60101099, 60101100, 60101103, 60101104, 60102461, 60102462, 60102463, 60102464, 60102466, 60102467, 60102468, 60103937, 60103938, 60103939, 60103941, 60103942, 60104859, 60105197, 60105198, 60105199, 60105200, 60105201, 60105379, 60105380, 60105389, 60105390, 60105393, 60105394, 60105395, 60106875, 60106877, 60106878, 60106879, 60106880, 60106881, 60108259, 60108260, 60108261, 60108262, 60108263, 60108264, 60108265, 60109537, 60109555, 60109556, 60109561, 60109562, 60109575, 60109576, 60110723, 60110725, 60110726, 60110727, 60110728, 60112087, 60113326, 60113329, 60113331, 60113332, 60113736, 60113737, 60113787, 60113788, 60113790, 60113791, 60114853, 60114855, 60114856, 60115234, 60115237, 60115238, 60115239, 60115240, 60115241, 60115242, 60116219, 60116835, 60116836, 60116843, 60116844, 60116851, 60116857, 60118036, 60118039, 60118040, 60118041, 60118042, 60119280, 60119332, 60119333, 60119375, 60119376, 60119377, 60119380, 60119381, 60119766, 60119781, 60119782, 60119783, 60119785, 60119786, 60120927, 60120928, 60120929, 60120930, 60120931, 60120932, 60120933, 60122722, 60122723, 60122724, 60122725, 60122726, 60122727, 60122728, 60124002, 60124003, 60124004, 60124639, 60124640, 60124641, 60124642, 60124643, 60124644, 60125998, 60125999, 60126000, 60126001, 60126002, 60126003, 60127183, 60127184, 60127185, 60127186, 60127188, 60127189, 60128809, 60128810, 60128849, 60128850, 60128851, 60129801, 60129802, 60129976, 60130029, 60130030, 60130785, 60130786, 60131001, 60131885, 60131886, 60131888, 60131889, 60131909, 60131910, 60131911, 60131913, 60132221, 60132222, 60132223, 60132893, 60132894, 60132895, 60132896, 60132897, 60132898, 60132899, 60132919, 60133465, 60133466, 60133467, 60133468, 60133469, 60134183, 60134184, 60134185, 60134214, 60134215, 60134216, 60134217, 60134218, 60135981, 60135982, 60135983, 60137150, 60137151, 60137152, 60137153, 60137154, 60137155, 60137708, 60137711, 60137823, 60137824, 60137825, 60138236, 60138353, 60138354, 60138359, 60139315, 60139318, 60139319, 60139320, 60140557, 60141754, 60142539, 60142540, 60143045, 60143046, 60143047, 60143048, 60143217, 60143218, 60143219, 60143220, 60143950, 60143951, 60143952, 60143953, 60143954, 60143955, 60145050, 60145444, 60145445, 60145446, 60145447, 60145448, 60145449, 60145450, 60148481, 60149035, 60149036, 60149037, 60149038, 60150445, 60151728, 60151740, 60152505, 60152686, 60152688, 60153376, 60153630, 60153633, 60154180, 60154253, 60154254, 60154259, 60154260, 60154261, 60154306, 60154838, 60154839, 60155183, 60155184, 60155190, 60155195, 60155196, 60155991, 60155993, 60156255, 60156256, 60156257, 60156259, 60157135, 60157136, 60157137, 60157138, 60158786, 60160737, 60160738, 60160741, 60160994, 60161581, 60161583, 60161584, 60161585, 60161587, 60164153, 60164154, 60164155, 60164579, 60165794, 60165795, 60165870, 60165871, 60165872, 60165873, 60166265, 60166266, 60166267, 60166268, 60167288, 60167289, 60167290) Expiration Dates: 04/30/2020 12/31/2021; 2. UDI Number: 00085412477220, Lot Numbers (60049663, 60049703, 60051310, 60052592, 60052693, 60053851, 60053854, 60053855, 60055340, 60055342, 60056343, 60057621, 60057622, 60057623, 60057624, 60057625, 60057626, 60059214, 60059219, 60059220, 60059229, 60059230, 60059241, 60059242, 60060402, 60060423, 60060424, 60060425, 60066392, 60066393, 60066394, 60066395, 60068172, 60068197, 60068198, 60068199, 60068200, 60069106, 60069109, 60069110, 60069112, 60069113, 60070839, 60070840, 60070843, 60070844, 60072120, 60072121, 60072122, 60072125, 60074016, 60074988, 60074991, 60074992, 60075126, 60075172, 60075175, 60075769, 60077611, 60077612, 60077613, 60077614, 60077615, 60079002, 60080311, 60080312, 60080317, 60080318, 60080325, 60080329, 60081682, 60081683, 60081684, 60081685, 60081687, 60082892, 60082895, 60082896, 60084620, 60084629, 60085746, 60085747, 60085748, 60085749, 60085750, 60085751, 60085752, 60086993, 60086994, 60086995, 60086996, 60086997, 60086998, 60088173, 60088175, 60088176, 60088177, 60088178, 60088179, 60089796, 60089797, 60089798, 60091153, 60091154, 60091156, 60091157, 60091158, 60091159, 60092420, 60092423, 60092424, 60092425, 60092426, 60092427, 60094112, 60094115, 60094116, 60094117, 60094118, 60094119, 60095455, 60095456, 60095458, 60095459, 60095460, 60095461, 60096711, 60096712, 60096719, 60096720, 60096721, 60096722, 60096723, 60097834, 60097835, 60097836, 60097837, 60097838, 60097839, 60098879, 60098880, 60098888, 60098897, 60098898, 60099750, 60099751, 60099752, 60099753, 60099754, 60099755, 60099756, 60101125, 60101126, 60101127, 60101128, 60101129, 60101130, 60102476, 60102477, 60102478, 60102479, 60102480, 60102481, 60102482, 60103960, 60103961, 60103962, 60103963, 60105207, 60105208, 60105209, 60105210, 60105415, 60105416, 60105421, 60105524, 60106899, 60106900, 60106901, 60106902, 60106903, 60106904, 60108273, 60108275, 60108276, 60108277, 60108278, 60108279, 60109607, 60109608, 60109615, 60112130, 60112137, 60113232, 60113233, 60113234, 60113235, 60113236, 60115865, 60115866, 60115867, 60115868, 60115869, 60115870, 60116858, 60116871, 60116872, 60116877, 60116878, 60116883, 60116884, 60118054, 60119275, 60119276, 60119277, 60119389, 60119390, 60119391, 60119392, 60119393, 60119394, 60119395, 60119794, 60119795, 60119796, 60119799, 60119800, 60119803, 60119804, 60120941, 60120942, 60120943, 60120944, 60120945, 60120946, 60120951, 60122736, 60122737, 60124015, 60124016, 60124017, 60124018, 60124652, 60124653, 60124654, 60124655, 60124656, 60124657, 60124658, 60126012, 60130818, 60130819, 60131897, 60131898, 60131899, 60132905, 60132906, 60132907, 60132908, 60132909, 60134196, 60134199, 60134200, 60134201, 60135992, 60135993, 60135994, 60135995, 60135996, 60135997, 60137164, 60137165, 60137166, 60137167, 60137713, 60138393, 60138394, 60138402, 60138409, 60140229, 60140565, 60140566, 60141888, 60141890, 60142551, 60142552, 60143503, 60143504, 60143505, 60143506, 60145069, 60145070, 60145457, 60145458, 60145459, 60145460, 60145461, 60145462, 60145463, 60146627, 60146637, 60147002, 60147003, 60147004, 60148490, 60148491, 60148492, 60148493, 60148494, 60148495, 60148496, 60149040, 60149811, 60149812, 60149813, 60149814, 60150425, 60150426, 60150427, 60150428, 60151631, 60151632, 60151651, 60151652, 60151659, 60152677, 60152678, 60153506, 60153707, 60153708, 60154215, 60154216, 60154217, 60154856, 60154859, 60154860, 60154861, 60154862, 60154863, 60154864, 60155172, 60155173, 60155174, 60155175, 60156243, 60156245, 60157738, 60157741, 60157743, 60157744, 60158768, 60158769, 60158770, 60158774, 60159853, 60159854, 60159857, 60160719, 60160720, 60160721, 60160722, 60160723, 60160728, 60161566, 60161567, 60161568, 60161569, 60161571, 60161572, 60162914, 60162951, 60162952, 60162959, 60165790, 60165791, 60165860, 60165863, 60165864, 60165865, 60166255, 60166256, 60166257, 60166258) Expiration Dates: 02/29/2020 12/31/2021; 3. UDI Number: 00085412477237, Lot Numbers (60064180, 60064181, 60064182, 60064183, 60064184, 60064185, 60064186, 60066019, 60069084, 60069099, 60069100, 60069102, 60069103, 60069105, 60072115, 60072118, 60074015, 60074076, 60074077, 60075858, 60075899, 60080378, 60080379, 60080380, 60080381, 60080382, 60081689, 60081690, 60081693, 60081694, 60081705, 60081706, 60082903, 60083520, 60083555, 60084631, 60084633, 60084634, 60084635, 60084636, 60084637, 60085753, 60085754, 60085756, 60085757, 60085758, 60085759, 60085760, 60087000, 60087001, 60087002, 60087003, 60087004, 60087005, 60088210, 60088217, 60088218, 60088221, 60088222, 60088225, 60088226, 60089801, 60089802, 60089803, 60089804, 60095475, 60095476, 60095477, 60095478, 60095479, 60095480, 60095481, 60096724, 60096727, 60096728, 60096729, 60096730, 60096731, 60096732, 60097841, 60097842, 60097843, 60097845, 60097846, 60098899, 60098900, 60098901, 60098903, 60098904, 60110758, 60110759, 60110760, 60110762, 60112138, 60112145, 60112146, 60112153, 60112154, 60113242, 60113245, 60115051, 60115052, 60115053, 60115054, 60115055, 60115056, 60115057, 60115253, 60115254, 60115255, 60115256, 60115257, 60115258, 60115259, 60118569, 60119382, 60119383, 60119384, 60119385, 60119386, 60119387, 60119388, 60120571, 60120572, 60120573, 60124007, 60124009, 60124010, 60124646, 60124647, 60124648, 60124649, 60124650, 60124651, 60125601, 60126004, 60126005, 60126006, 60126007, 60126010, 60127190, 60127192, 60127193, 60127194, 60132900, 60132902, 60132903, 60132904, 60134186, 60134187, 60134188, 60134189, 60134190, 60134191, 60135985, 60135986, 60135988, 60135989, 60135990, 60137712, 60138469, 60140582, 60140847, 60140848, 60140859, 60140869, 60140870, 60142171, 60142172, 60142173, 60142185, 60142547, 60142548, 60145055, 60146999, 60147000, 60147001, 60148484, 60148485, 60148486, 60148487, 60148488, 60148489, 60149800, 60149801, 60149802, 60149805, 60149806, 60151291, 60151292, 60154205, 60154206, 60154207, 60154208, 60154209, 60154210, 60154211, 60154212, 60154213, 60154214, 60155994, 60155995, 60155997, 60155998, 60156219, 60156220, 60156221, 60156222, 60156223, 60156466, 60156467, 60156468, 60157729, 60157730, 60158355, 60158359, 60158417, 60158418, 60158541, 60158749, 60158750, 60158751, 60158752, 60158754, 60158756, 60158759, 60158760, 60158792, 60158795, 60159839, 60159841, 60160682, 60160925, 60160926, 60161397, 60161398, 60161551, 60161552, 60161554, 60161555, 60161556, 60161557, 60161558, 60161559, 60162227, 60162228, 60162873, 60162874, 60162883, 60162884, 60162885, 60162909, 60162910, 60162911, 60162912, 60163509, 60164100, 60164101, 60165786, 60165787, 60165847, 60165848, 60165849, 60165850) Expiration Dates: 04/30/2020 12/31/2021; 4. UDI Number: 00085412477244, Lot Numbers (60048539, 60049820, 60056225, 60056226, 60056227, 60057619, 60059168, 60059170, 60059171, 60059176, 60060435, 60060436, 60060437, 60060438, 60060439, 60061929, 60061930, 60061931, 60061938, 60061939, 60061940, 60061941, 60061942, 60063153, 60063155, 60063156, 60063162, 60063163, 60063164, 60063165, 60063166, 60063169, 60063226, 60064558, 60064578, 60064580, 60064583, 60064584, 60064587, 60066269, 60066275, 60066277, 60066279, 60068134, 60068141, 60068142, 60068163, 60068164, 60068165, 60068166, 60068170, 60069697, 60069699, 60069700, 60069701, 60069702, 60069703, 60069704, 60070833, 60072130, 60072132, 60072134, 60072138, 60073337, 60073338, 60073375, 60073423, 60074713, 60078329, 60078333, 60078398, 60078959, 60078960, 60078961, 60078962, 60078963, 60078964, 60078965, 60078966, 60078967, 60078968, 60078969, 60078971, 60080304, 60080305, 60080306, 60080307, 60081678, 60081680, 60081681, 60083767, 60084476, 60084485, 60084486, 60084489, 60084490, 60084493, 60084494, 60086987, 60086989, 60086990, 60086991, 60086992, 60088197, 60088198, 60088201, 60088202, 60088209, 60089790, 60089791, 60089792, 60089793, 60089794, 60089795, 60091145, 60091146, 60091149, 60091150, 60091151, 60091152, 60092413, 60092414, 60092415, 60092416, 60092417, 60092419, 60094105, 60094106, 60094107, 60094108, 60094109, 60094110, 60094111, 60094663, 60094664, 60095448, 60095449, 60095451, 60095452, 60095453, 60095454, 60096704, 60096705, 60096706, 60096707, 60096709, 60096710, 60097826, 60097827, 60097828, 60097829, 60097830, 60097831, 60098865, 60098866, 60098867, 60098868, 60098869, 60099746, 60099747, 60099748, 60099749, 60101112, 60101123, 60102469, 60102470, 60102471, 60102472, 60102473, 60102474, 60102475, 60103944, 60103953, 60103954, 60103955, 60103956, 60104709, 60104862, 60105212, 60105213, 60105214, 60105215, 60105216, 60105433, 60105441, 60105442, 60105443, 60106891, 60106892, 60106893, 60108269, 60108270, 60109578, 60109579, 60109580, 60109587, 60109588, 60109593, 60110730, 60110733, 60110734, 60110735, 60110737, 60112098, 60112100, 60112106, 60112107, 60113227, 60113228, 60113229, 60113230, 60113231, 60114859, 60115245, 60115247, 60115248, 60115249, 60115251, 60116895, 60116896, 60117290, 60118320, 60118321, 60119334, 60119335, 60119396, 60119398, 60119399, 60119400, 60119401, 60119402, 60119789, 60119791, 60119792, 60120952, 60120959, 60120966, 60120975, 60120976, 60122008, 60122739, 60122740, 60122741, 60124019, 60124020, 60124021, 60124022, 60124621, 60124661, 60124662, 60124663, 60124665, 60124666, 60124667, 60126016, 60126019, 60126021, 60126022, 60126025, 60127209, 60127210, 60127211, 60127212, 60127213, 60128819, 60128820, 60130043, 60130044, 60131900, 60131901, 60131902, 60131903, 60131904, 60131905, 60131931, 60131932, 60132770, 60132771, 60132772, 60132911, 60132912, 60132913, 60132914, 60132916, 60132917, 60134204, 60134210, 60134211, 60134212, 60134213, 60136001, 60136002, 60136003, 60136004, 60136005, 60136006, 60137168, 60137241, 60137242, 60137253, 60137254, 60137265, 60138411, 60138412, 60138413, 60138414, 60138415, 60139335, 60139336, 60139339, 60139340, 60140574, 60140576, 60140577, 60140578, 60141891, 60141892, 60141893, 60141894, 60141897, 60142543, 60143221, 60143222, 60143223, 60143224, 60143226, 60143956, 60143957, 60143958, 60143959, 60143960, 60143961, 60143962, 60145049, 60145465, 60145466, 60145467, 60145468, 60145469, 60147006, 60147007, 60147008, 60147009, 60147010, 60147011, 60148498, 60148499, 60148500, 60150432, 60150433, 60150434, 60150435, 60150437, 60150438, 60150460, 60151673, 60151674, 60151703, 60151704, 60151713, 60151714, 60152680, 60152681, 60152682, 60152683, 60152684, 60152685, 60154218, 60154219, 60154220, 60154221, 60154222, 60154223, 60154224, 60155176, 60155177, 60155178, 60155179, 60155180, 60155181, 60155182, 60156248, 60156250, 60156251, 60156252, 60156253, 60157745, 60157747, 60157748, 60157749, 60158775, 60158777, 60158778, 60158779, 60158780, 60158781, 60159866, 60159881, 60159882, 60159885, 60159886, 60160731, 60160732, 60160734, 60160735, 60160736, 60161575, 60161576, 60161577, 60161578, 60162972) Expiration Dates: 02/29/2020 11/30/2021; 5. UDI Number: 00085412477251, Lot Numbers (60052524, 60056338, 60056340, 60057627, 60064588, 60064591, 60064592, 60064593, 60064597, 60064598, 60064599, 60064603, 60066282, 60066283, 60066287, 60066288, 60066290, 60066293, 60073296, 60073319, 60073331, 60074720, 60078409, 60078972, 60078973, 60078974, 60078975, 60082899, 60082900, 60091160, 60091161, 60091162, 60091163, 60091164, 60091165, 60091167, 60092428, 60092429, 60092430, 60092431, 60092433, 60092434, 60094123, 60094124, 60094125, 60099758, 60099760, 60099761, 60099762, 60101133, 60101134, 60102483, 60102484, 60102486, 60102487, 60102488, 60102489, 60104659, 60104860, 60105202, 60105203, 60105204, 60105205, 60105396, 60105403, 60105405, 60105406, 60105409, 60105410, 60106906, 60106915, 60106916, 60106917, 60106919, 60108443, 60108444, 60108445, 60108446, 60108447, 60108448, 60110492, 60110503, 60110504, 60110505, 60110506, 60110507, 60110508, 60117112, 60117114, 60117115, 60117117, 60118045, 60119884, 60120935, 60120937, 60120940, 60122732, 60122733, 60130038, 60131890, 60131892, 60131893, 60131894, 60132224, 60132232, 60132233, 60137157, 60137158, 60137159, 60137160, 60137161, 60137162, 60138360, 60138373, 60138374, 60138377, 60138378, 60146556, 60146557, 60147815, 60147816, 60147817, 60147818, 60147962, 60147963, 60147964, 60151605, 60151606, 60151611, 60151612, 60152667, 60152668, 60152669, 60152670, 60159843, 60159844, 60159849, 60159850, 60160713) Expiration Dates: 03/31/2020 11/30/2021 |
| FEI Number |
1417572
|
Recalling Firm/ Manufacturer |
Baxter Healthcare Corporation 1 Baxter Pkwy Deerfield IL 60015-4625
|
| For Additional Information Contact | 800-422-9837 |
Manufacturer Reason for Recall | Potential ability to leak once used for compounding. |
FDA Determined Cause 2 | Process control |
| Action | Baxter Healthcare Corporation notified customers on about 09/18/2019, via "Urgent Product Recall" letter. The letter informed customers that certain lot numbers of the EXACTAMIX Empty EVA Bags have the potential ability to leak once used for compounding.
Customers were instructed to locate and remove all affected product lots from your facility, contact Baxter Healthcare Center for Service to arrange for return and credit, complete and return the provided Baxter customer reply form via fax to 224-270-5457 or scanning and e-mailing it to fca@baxter.com, and notify customers or affected personnel/departments if the product was further distributed. For customers who purchased affected product from a distributor, contact your distributor for return and credit.
Questions can be directed to Baxter Corporate Product Surveillance at 800-437-5176, between the hours of 8:00 am and 5:00 pm Central Time, Monday through Friday. |
| Quantity in Commerce | 11,389,800 units total |
| Distribution | Worldwide distribution - UD Nationwide including the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY.
The countries of Canada, Australia, New Zealand, Singapore, Korea, Taiwan, Thailand, Hong Kong, Brazil, Columbia, Mexico, Chile, Uruguay, Dominican Republic, Panama, Bermuda. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = LHI
|
|
|
|