 Class 2 Device Recall Periarticular Locking Plate Fibulasee related information
Date Initiated by FirmOctober 10, 2019
Create DateNovember 07, 2019
Recall Status1 Terminated 3 on May 19, 2020
Recall NumberZ-0305-2020
Recall Event ID 83962
510(K)NumberK150121 
Product Classification Plate, fixation, bone - Product Code HRS
ProductZPLP Fibular Plate, Item Nos. 00234701504 00234701506 00234701508 00234701510 00234701512 00234701516 00234701604 00234701606 00234701608 00234701610 00234701612 00234701614 00234701616 00234701702 00234701704 00234701706 00234701708 00234701710 00234701712 00234701714 00234701716 00234701718 00234701802 00234701804 00234701806 00234701808 00234701810 00234701812 00234701814 00234701816 00234701818 47235701704 47235701706 47235701708 47235701710 47235701712 47235701714 47235701716 47235701804 47235701806 47235701808 47235701810 47235701812 47235701814 47235701816
Code Information All product manufactured prior to January 2014.
FEI Number 1000220733
Recalling Firm/
Manufacturer
Zimmer Biomet, Inc.
1800 W Center St
Warsaw IN 46580-2304
For Additional Information ContactZimmer Biomet Customer Service
574-371-3071
Manufacturer Reason
for Recall
Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration).
FDA Determined
Cause 2
Packaging change control
ActionThe firm distributed Urgent Medical Device Correction letters to customers on October 10, 2019. The letter informed customers that the previous version of low density polyethylene (LDPE) bag could potentially adhere to the surface of metal devices. Please note, this is separate from the previous recall for highly polished implants packaged in the previous version of the LDPE bags. Customers were asked to take the following actions: 1. Review this notification and ensure that pertinent personnel are aware of the contents. 2. Ensure that backup devices for the affected products are available, as per standard orthopedic procedure planning, to reduce the likelihood of an extension of surgery time. 3. Examine devices thoroughly for adherent LDPE bag residue prior to use. Use a backup device if LDPE bag residue is detected. 4. Return any devices which display adherent LDPE bag residue to Zimmer Biomet, consistent with the standard instructions for use regarding the disposition of damaged components. 5. Complete the Certificate of Acknowledgement and return to the firm. 6. Retain a copy of the acknowledgement form with your field action records in the event of a compliance audit of your facility's documentation. If you have further questions or concerns regarding this recall, please call the firm's customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com.
Quantity in Commerce5,459,583 total devices
DistributionDistributed nationwide.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = HRS
-
-