• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall ORTHOFIX

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall ORTHOFIXsee related information
Date Initiated by FirmSeptember 27, 2018
Create DateNovember 14, 2019
Recall Status1 Terminated 3 on February 01, 2022
Recall NumberZ-0390-2020
Recall Event ID 84004
510(K)NumberK955848 
Product Classification Pin, fixation, threaded - Product Code JDW
ProductORTHOFIX Catalogue Number: ref: 99-92501, PREFIX II DIAPHYSEAL COMPLETE KIT STERILE SHORT SCREWS, RX ONLY, UDI: (01) 18033509855948
Code Information Lot #  B1207286, B1207428, B1211384, B1215263  *Not marketed in US
FEI Number 3002807827
Recalling Firm/
Manufacturer
Orthofix Srl
Via Delle Nazioni 9 O 73
Veronaverona Italy
Manufacturer Reason
for Recall
As a result of complaint investigations it was determined that batches of drill bits have a geometrical detail that does not correspond to the intended design. This could result in lower cutting performance.
FDA Determined
Cause 2
Device Design
ActionOn 10/01/2018, Orthofix provided distributor, doctors, and hospitals a urgent medical device recall notification via FedEx overnight carrier . The letter asked consignees to take the following actions: 1. Immediately cease any further distribution of the affected devices; 2. Notify hospitals, surgeons and others to whom you may have distributed affected devices that such devices cannot be used and must be returned to Orthofix according to the instructions below; 3. Complete the Acknowledgement Form indicating that you have received this notification. It should be returned by email or fax to the contact details on the attached form. 4. Subsequent Distribution: If your firm has provided or sold the affected products to other users, please provide a list of those users to the undersigned and forward this notification to them to assure prompt action. 5. Product Return Instructions: To return the devices affected by this action, please follow these instructions: A - Check your inventory for the devices affected by this action; B - Complete the attached Acknowledgement Form with the quantity you are returning; C - Contact Orthofix Customer Service at (800) 266-3349 for return instructions and replacement information
DistributionUS: AZ, CA,FL,GA,HI,IA,IL,IN,KY,LA,MD,MI,MO,MS,NC,NY,OH,OK,OR,PA,SC,TN,TX,UT,VA,WA, WI, WV, Puerto Rico OUS: Kuwait, Australia, Saudi Arabia, United Arab Emirates, Ireland, Japan, Spain, Poland, Switzerland, Singapore, Peru, Brazil, Germany, Great Britain, France, Italy, Australia, South Africa, Belgium, Chile, The Netherlands, Lithuania, Sri Lanka, Denmark, Israel
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
510(K) Database510(K)s with Product Code = JDW
-
-