| Date Initiated by Firm | November 11, 2019 |
| Create Date | January 03, 2020 |
| Recall Status1 |
Terminated 3 on September 10, 2020 |
| Recall Number | Z-0776-2020 |
| Recall Event ID |
84405 |
| 510(K)Number | K182201 |
| Product Classification |
Screw, fixation, bone - Product Code HWC
|
| Product | Stratum MDS Screw 3.5x46mm ST
Part Number: STRM-MDS-3546ST |
| Code Information |
Lot Number: M1266 UDI: 00817701023159 |
| FEI Number |
3009540749
|
Recalling Firm/ Manufacturer |
Nextremity Solutions
6210 N Buffalo St
Warsaw
IN
46580
|
| For Additional Information Contact | 732-683-9305 |
Manufacturer Reason for Recall | Components 2mm shorter than indicated on the package label |
FDA Determined Cause 2 | Packaging process control |
| Action | Nextremity Solutions issued Urgent Medical Device Recall letter on 11/11/19 stating reason for recall, health risk and action to take: do not use, quarantine and return to Zimmer Biomet representative who will remove product, Contact customer service at 574-371-3071 between 8am -5pm(EST) M-F. |
| Quantity in Commerce | 87 units |
| Distribution | Nationwide |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
| 510(K) Database | 510(K)s with Product Code = HWC
|