• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Atellica CH 930 Analyzer

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Atellica CH 930 Analyzersee related information
Date Initiated by FirmFebruary 27, 2020
Create DateApril 24, 2020
Recall Status1 Terminated 3 on October 25, 2021
Recall NumberZ-1797-2020
Recall Event ID 85268
510(K)NumberK151767 
Product Classification Electrode, ion specific, sodium - Product Code JGS
ProductAtellica CH 930 Analyzer, Siemens Material Number 11067000 with software version v1.21.0 SP1 or lower
Code Information UDI: 00630414002163; All systems with software version v1.21.0 SP1 or lower  Serial Numbers: CM01702 CM01708 CM00184 CM00196 CM01051 CM01467 CM00770 CM00815 CM00983 CM01053 CM00802 CM01004 CM01903 CM00526 CM00635 CM00809 CM00733 CM00735 CM01045 CM01050 CM01052 CM01059 CM00420 CM00632 CM00633 CM00209 CM00210 CM01799 CM01590 CM01612 CM01047 CM01048 CM00859 CM00865 CM01624 CM01626 CM00829 CM00830 CM00651 CM00654 CM01603 CM01608 CM01695 CM01701 CM00749 CM00786 CM01567 CM01569 CM01682 CM01683 CM00773 CM01431 CM01474 CM01583 CM01598 CM01602 CM00148 CM00151 CM00152 CM01907 CM01738 CM01739 CM01105 CM00799 CM00800 CM01823 CM01917 CM01409 CM01410 CM01588 CM01600 CM01274 CM01306 CM00947 CM00948 CM01660 CM01661 CM00812 CM00828 CM00820 CM01593 CM01594 CM01585 CM01599 CM01058 CM00421 CM00606 CM00331 CM00333 CM01133 CM01149 CM01619 CM01191 CM01200 CM00769 CM00858 CM01237 CM01259 CM00797 CM00798 CM01589 CM01614 CM00426 CM00427 CM00701 CM00704 CM01264 CM01278 CM01391 CM01512 CM00611 CM00612 CM01735 CM00643 CM00646 CM00831 CM00832 CM01761 CM01772 CM01727 CM01740 CM01745 CM00315 CM00316 CM00123 CM00181 CM01031 P-P1-07 CM00147 CM00432 CM00297 CM00430 CM00780 CM01296 CM00268 CM00271 CM00265 CM00267 CM01759 CM01760 CM01579 CM01611 CM01139 CM00607 CM00625 CM01699 CM01716 CM00884 CM00888 CM00577 CM00631 CM01650 CM00459 CM00510 CM01441 CM01343 CM01394 CM00973 CM00977 CM01586 CM01616 CM01620 CM01621 CM00634 CM00637 CM01400 CM01406 CM00626 CM00627 CM01344 CM01345 CM00418 CM00419 CM00724 CM00628 CM00630 CM00958 CM00979 CM00971 CM00481 CM00482 CM01066 CM01118 CM01923 CM01675 CM01241 CM01253 CM01242 CM01255 CM01247 CM01268 CM01801 CM01528 CM01529 CM00864 CM00867 CM00400 CM00407 CM01455 CM01459 CM00851 CM00852 CM01499 CM00330 CM00343 CM00344 CM01513 CM01553 CM01818 CM01819 CM01597 CM00905 CM00915 CM00917 CM00918 CM00457 CM00458 CM01065 CM01239 CM01245 CM00639 CM00640 CM00746 CM00747 CM00748 CM01694 CM01696 CM00182 CM00183 CM01401 CM01423 CM00422 CM00423 CM00424 CM00425 CM00781 CM01576 CM00833 CM00834 CM00957 CM01016 CM00201 CM00621 CM01556 CM01566 CM01580 CM01581 CM01582 CM01800 CM01333 CM00723 CM00734 CM01767 CM01768 CM01318 CM01717 CM01773 CM01399 CM01293 CM01707 CM00768 CM00959 CM00988 CM01577 CM00229 CM00246 CM00715 CM00721 CM01072 CM00613 CM00614 CM01034 CM01068 CM01017 CM01653 CM00743 CM00744 CM01061 CM01062 CM00810 CM01362 CM00853 CM00856 CM01257 CM01258 CM01663 CM01686 CM01176 CM01724 CM00523 CM00524 CM00288 CM00289 CM01432 CM01433 CM01700 CM01617 CM00528 CM00529 CM00756 CM00757 CM00650 CM00653 CM00310 CM00525 CM00545 CM00647 CM00663 CM01395 CM01396 CM00783 CM00784 CM00208 CM00211 CM00272 CM00273 CM01106 CM01107 CM01094 CM01126 CM00449 CM00448 CM01218 CM01297 CM00361 CM00362 CM00202 CM00203 CM00293 CM00693 CM00696 CM00697 CM01645 CM01651 CM01659 CM01662 CM00965 CM00986 CM01657 CM01673 CM00803 CM00804 CM00805 CM00847 CM00962 CM00985 CM00964 CM00970 CM00982 CM01013 CM00963 CM00974 CM01057 CM01069 CM01077 CM01079 CM01081 CM01082 CM01084 CM01085 CM01086 CM01087 CM01090 CM01092 CM01093 CM01096 CM01097 CM01098 CM01099 CM01100 CM01110 CM01120 CM00953 CM01697 CM00163 CM01003 CM00527 CM00729 CM00730 CM00221 CM00222 CM00717 CM01280 CM01666 CM01630 CM01632 CM01520 CM01152 CM01163 CM01300 CM01151 CM00949 CM00664 CM00666 CM00261 CM01162 CM01170 CM00906 CM00903 CM00814 CM00684 CM01631 CM01637 CM01102 CM01165 CM01169 CM01196 CM00886 CM01023 CM01024 CM01369 CM01371 CM01374 CM01387 CM01351 CM01370 CM01372 CM01384 CM00848 CM00414 CM00415 CM00942 CM01073 CM01074 CM00996 CM00328 CM00462 CM00465 CM00467 CM01033 CM00312 CM01456 CM01064 CM01448 CM00304 CM00305 CM00409 CM01028 CM00279 CM00280 CM01019 CM00236 CM00303 CM00314 CM00329 CM00410 CM00413 CM00306 CM00655 CM00659 CM00667 CM00668 CM00755 CM00763 CM00764 CM00767 CM00788 CM00796 CM00807 CM00939 CM01055 CM01720 CM00902 CM00787 CM00821 CM01726 CM01729 CM01018 CM01041 CM01075 CM01536 CM01728 CM01734 CM00284 CM00285 CM01730 CM01185 CM01475 CM00461 CM00808 CM00731 CM00732 CM00762 CM01403 CM01430 CM00555 CM00237 CM00238 CM00239 CM00416 CM01178 CM01054 CM01112 CM00745 CM00455 CM01557 CM00874 CM01208 CM01291 CM01752 CM01043 CM01177 CM01438 CM01485 CM01130 CM01140 CM01314 CM01138 CM01146 CM01894 CM01698 CM00149 CM01470 CM00277 CM00835 CM00840 CM01187 CM01236 CM01355 CM01336 CM01199 CM01897 CM00932 CM00933 CM01325 CM00761 CM00661 CM00676 CM00951 CM00665 CM00765 CM00924 CM00925 CM01453 CM01454 CM01510 CM01584 CM00346 CM00675 CM00677 CM00725 CM00736 CM00896 CM00515 CM00516 CM00518 CM00519 CM00517 CM00509 CM00514 CM01123 CM01125 CM01128 CM01135 CM01416 CM01417 CM00197 CM00198 CM00160 CM00161 CM00167 CM00169 CM01172 CM01295 CM01610 CM01613 CM00992 CM00256 CM01192 CM00885 CM00862 CM00283 CM00278 CM00920 CM00922 CM00923 CM00929 CM01119 CM01121 CM01211 CM01210 CM01207 CM01231 CM01469 CM01483 CM01487 CM01488 CM00994 CM00999 CM01000 CM01030 CM01150 CM01181 CM00997 CM00998 CM01001 CM01002 CM01153 CM01164 CM01180 CM01652 CM01680 CM01691 CM01690 CM01015 CM01755 CM00488 CM00493 CM00494 CM01413 CM01414 CM01269 CM01279 CM00501 CM00556 CM00242 CM00244 CM00934 CM00969 CM01578 CM01605 CM00706 CM00722 CM00679 CM01111 CM01115 CM00817 CM01462 CM00433 CM00946 CM00961 CM00240 CM00258 CM00503 CM00601 CM01326 CM01327 CM01134 CM01142 CM01390 CM01392 CM01412 CM01445 CM00558 CM00935 CM00936 CM00940 CM00497 CM00687 CM00691 CM00928 CM00931 CM00694 CM00703 CM00890 CM01703 CM00170 CM00171 CM00172 CM01472 CM01405 CM01407 CM01415 CM01265 CM01331 CM01440 CM01359 CM01360 CM01672 CM01688 CM01689 CM00569 CM00570 CM00582 CM00485 CM00495 CM00568 CM01209 CM01233 CM01179 CM01232 CM01250 CM01252 CM00499 CM00543 CM00443 CM01020 CM00452 CM00496 CM00135 CM00140 CM01503 CM01537 CM01541 CM00498 CM01622 CM00476 CM00792 CM00557 CM00620 CM01715 CM00154 CM00877 CM00880 CM00921 CM01515 CM01530 CM01532 CM00950 CM00776 CM01526 CM00442 CM00891 CM00894 CM01756 CM01757 CM01029 CM00857 CM00132 CM00134 CM01337 CM01341 CM00816 CM00403 CM00404 CM00405 CM00345 CM00349 CM00350 CM00153 CM00439 CM00311 CM00649 CM01408 CM01411 CM01618 CM01559 CM00137 CM00139 CM00188 CM00189 CM00190 CM00487 CM00926 CM01267 CM01275 CM01288 CM01294 CM00227 CM01032 CM01422 CM00738 CM00739 CM00740 CM00741 CM00742 CM01038 CM01039 CM00259 CM00662 CM01427 CM01439 CM01664 CM01665 CM00843 CM00844 CM01159 CM01160 CM01166 CM01743 CM01744 CM00751 CM00502 CM00276 CM00434 CM01466 CM00536 CM01465 CM00270 CM01812 CM01366 CM01234 CM01719 CM01381 CM01382 CM01385 CM01389 CM00754 CM01136 CM01380 CM01722 CM01647 CM01892 CM01495 CM01816 CM01310 CM01500 CM01171 CM01329 CM01365 CM01025 CM01721 CM01379 CM01501 CM01473 CM00760 CM00294 CM00351 CM01480 CM00165 CM01502 CM01353 CM01540 CM01544 CM01302 CM01644 CM01308 CM01354 CM01494 CM01103 CM01108 CM01010 CM00357 CM00369 CM00381 CM01426 CM01561 CM01504 CM01514 CM01562 CM01324 CM01224 CM00573 CM01320 CM00539 CM00540 CM00537 CM00538 CM00580 CM00603 CM00616 CM00431 CM01251 CM00133 CM00447 CM00456 CM00463 CM00454 CM00388 CM00385 CM00685 CM00753 CM00759 CM00775 CM01342 CM01347 CM00218 CM00219 CM00220 CM00705 CM00383 CM00617 CM00618 CM01542 CM00576 CM01558 CM01217 CM00605 CM00615 CM00166 CM00176 CM00177 CM00178 CM00391 CM00453 CM00444 CM00930 CM00952 CM01350 CM01358 CM00402 CM00440 CM00450 CM00563 CM00597 CM00254 CM01349 CM00897 CM00900 CM00910 CM00911 CM00912 CM00913 CM00466 CM00469 CM00777 CM00823 CM00260 CM01442 CM01443 CM01014 CM00195 CM00225 CM00226 CM00224 CM00882 CM00854 CM00253 CM00252 CM00384 CM00214 CM00217 CM00207 CM00145 CM00619 CM00600 CM00484 CM00486 CM01220 CM00870 CM00806 CM00468 CM01229 CM01436 CM01446 CM01450 CM01452 CM01063 CM01071 CM01076 CM00771 CM00868 CM00872 CM00608 CM00836 CM01303 CM01281 CM01311 CM01322 CM01299 CM01373 CM01364 CM01282 CM01388 CM01340 CM00689 CM00726 CM00130 CM00871 CM01533 CM00472 CM00473 CM00839 CM01709 CM00173 CM00174 CM00175 CM00441 CM01490 CM01491 CM00406 CM01713 CM00282 CM00281 CM01143 CM01145 CM01486 CM01157 CM01156 CM00247 CM01357 CM01867 CM00863 CM00893 CM01868 CM00670 CM01056 CM01778 CM01779 CM01361 CM01460 CM01204 CM00699 CM00483 CM00710 CM01027 CM01078 CM01060 CM01080 CM01091 CM01088 CM01710 CM00474 CM00475 CM00674 CM01841 CM00199 CM00200 CM01363 CM01316 CM00711 CM00712 CM00714 CM00899 CM00901 CM00908 CM00909 CM00919 CM00938 CM01629 CM00602 CM00883 CM00895 CM00194 CM00191 CM00876 CM00873 CM00308 CM00309 CM00146 CM01309 CM01313 CM00801 CM01554 CM01564 CM00464 CM01420 CM01421 CM01457 CM01731 CM01732 CM01733 CM00125 CM01008 CM01298 CM01451 CM01129 CM01522 CM00993 CM01009 CM01026 CM01228 CM01175 CM01681 CM01687 CM01437 CM01428 CM01429 CM01383 CM01386 CM00117 CM00673 CM00412 CM00193 CM01338 CM01036 CM01005 CM01748 CM01012 CM01212 CM01284 CM01292 CM01011 CM01286 CM01201 CM00968 CM01260 CM01346 CM00916 CM01367 CM01213 CM01287 CM01434 CM01435 CM01646 CM01339 CM01477 CM01676 CM01677 CM01678 CM01424 CM01425 CM01718 CM01725 CM01193 CM01198 CM00943 CM00955 CM00954 CM01202 CM01147 CM01131 CM01418 CM01419 CM01007 CM01221 CM00249 CM01285 CM01658 CM01604 CM00966 CM01393 CM00914 CM00713 CM00230 CM00551 CM01116 CM00347 CM00334 CM00945 CM01307 CM01315 CM01305 CM00698 CM00778 CM01623 CM01633 CM00287 CM00286 CM00335 CM00505 CM00504 CM01227 CM00506 CM00507 CM00508 CM00987 CM01021 CM01534 CM01538 CM00332 CM00336 CM00337 CM00708 CM00709 CM00554 CM00559 CM00561 CM00562 CM00838 CM00648 CM01506 CM01498 CM01492 CM01489 CM00491 CM00492 CM01493 CM00489 CM00758 CM00234 CM00233 CM00547 CM00549 CM00550 CM00552 CM00553 CM00560 CM00530 CM00629 CM00359 CM00360 CM00490 CM00941 CM00967 CM01197 CM01206 CM00317 CM00318 CM00319 CM00320 CM00321 CM00323 CM00324 CM00325 CM00326 CM00150 CM00136 CM01335 CM01356 CM00215 CM00216 CM00720 CM01158 CM01194 CM01319 CM01640 CM01328 CM01484 CM01625 CM00564 CM00824 CM00825 CM01124 CM00386 CM00363 CM00396 CM01674 CM00248 CM00257 CM00251 CM00366 CM01458 CM01519 CM01517 CM00245 CM01521 CM01919 CM00658 CM00813 CM00512 CM00904 CM01668 CM01669 CM01188 CM01189 CM00541 CM00669 CM00671 CM00678 CM00542 CM01270 CM01277 CM01317 CM01330 CM00688 CM00875 CM00398 CM00429 CM00428 CM00477 CM00478 CM00479 CM00752 CM00975 CM01044 CM01109 CM01114 CM00989 CM00990 CM00389 CM00390 CM01321 CM01497 CM01524 CM01525 CM00340 CM00339 CM00520 CM00521 CM00716 CM00991 CM01035 CM00793 CM00794 CM00937 CM00131 CM00500 CM00338 CM00342 CM01545 CM00826 CM00827 CM00795 CM00255 CM00275 CM00300 CM00263 CM00264 CM00692 CM00737 CM00451 CM00927 CM01552 CM01565 CM00566 CM00587 CM00594 CM00599 CM00604 CM00686 CM00690 CM00700 CM00702 CM00707 CM00978 CM00980 CM01225 CM01226 CM01246 CM01203 CM01248 CM01254 CM01261 CM01273 CM00179 CM01348 CM01468 CM01447 CM00159 CM00156 CM00158 CM01776 CM01777 CM01154 CM01161 CM01332 CM01334 CM00879 CM00887 CM01792 CM00341 CM00575 CM00204 CM00791 CM00822 CM01655 CM01746 CM00298 CM00984 CM01574 CM00850 CM00790 CM00842 CM00845 CM01527 CM01543 CM01368 CM00683 CM01186 CM01271 CM00356 CM00355 CM00438 CM00437 CM00435 CM00436 CM00574 CM00585 CM00589 CM00586 CM00590 CM00583 CM01670 CM01671 CM01615 CM01654 CM01667 CM01679 CM00682 CM00262 CM00241 CM00866 CM01083 CM00638 CM00206 CM01648 CM01649 CM01518 CM01523 CM01183 CM01205 CM01040 CM01042 CM00141 CM00143 CM00565 CM00571 CM00598 CM00591 CM01222 CM00186 CM00187 GGCCEP23 GGCCEP4 CM00578 CM00588 CM00296 CM00960 CM00972 CM01022 CM00595 CM01132 CM01290 CM01266 CM01549 CM00727 CM01535 CM00785 CM01046 CM00511 CM00728 CM01693 CM01737 CM00513 CM01505 CM01141 CM00353 CM00354 
FEI Number 2432235
Recalling Firm/
Manufacturer
Siemens Healthcare Diagnostics, Inc.
511 Benedict Ave
Tarrytown NY 10591-5005
For Additional Information ContactRanjan Sadarangani
914-631-8000
Manufacturer Reason
for Recall
Editing an assay Test Definition and switching to another assay s Test Definition screen without saving the changes using the save button on the screen, may corrupt the settings of the assay s test definition, and Auto-Rerun or Auto-Dilution results may generate a result of Zero (0) for quantitative assays or Negative for qualitative assays if a well of the reagent pack is not calibrated.
FDA Determined
Cause 2
Under Investigation by firm
ActionAn Urgent Medical Device Correction notification letter dated 2/27/20 was sent to customers. The purpose of this communication is to inform you of the issues listed in Table 2 below, that can occur on the Atellica CH 930 Analyzer, installed with Atellica Solution software (SW) versions v1.21.0 SP1 or lower and provide instructions on actions your laboratory must take. These issues will be corrected in SW v1.22.0 and v1.21.11, which will be available soon. Actions to be Taken by the Customer The following actions must be taken until your system has been updated to software version v1.22.0 (or V1.21.11 for Atellica Solution with decapper modules) or higher which resolves the issues listed above. 1. When making modifications to an assays Test Definition settings on the Setup>Test Definition>CH Test Definition screen, always follow the instructions outlined in the Atellica Solution online help 1.20 instructions Editing Assay Identification Parameters in CH Test Definition. Each Test Definition must be saved using the Save button on the CH Test Definition screen before starting modifications on another assays Test Definition. After changing any setting on the Definition, Calculation, or Calibration tabs of the CH Test Definition screen, select Save and then OK. Verify that QC results are not affected, and that results, units, and flags are reported correctly for any Test Definition that has been updated before running patient samples. 2. Ensure all reagent packs have a valid calibration (all wells) before running samples. This information is available on the screen Calibration > Calibration Overview. Refer to the Atellica Solution online help 1.20 instructions, Viewing Assay Reagent Calibration Order Status. When available, software v1.22.0 and/or v1.21.11 will be delivered as follows: For systems running software v1.19.0 or higher, that are connected to the Smart Remote Service (SRS), the software will be delivered electronically
Quantity in Commerce1807
DistributionUS Nationwide. Global distribution.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = JGS
-
-