• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recallsee related information
Date Initiated by FirmSeptember 24, 2020
Create DateNovember 02, 2020
Recall Status1 Terminated 3 on May 20, 2022
Recall NumberZ-0363-2021
Recall Event ID 86521
Product Classification Prosthesis, hip, semi-constrained, metal/polymer, cemented - Product Code JDI
ProductZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 22 mm, 10 degree inclined face, O.D. Cup with Spacers, 49 mm Number: 00-8065-646-22
Code Information All lots
FEI Number 1000220733
Recalling Firm/
Manufacturer
Zimmer Biomet, Inc.
1800 W Center St
Warsaw IN 46580-2304
For Additional Information Contact411 Technical Services
574-371-3071
Manufacturer Reason
for Recall
Sterilant (hydrogen peroxide) used in acetabular cup hip prosthesis component was not evaluated to confirm the biocompatibility of the residual sterilant. Cannot rule out the potential for adverse tissue reactions.
FDA Determined
Cause 2
Device Design
ActionZimmer Biomet issued URGENT MEDICAL DEVICE RECALL AND NOTICE OF US FIELD DISCONTINUATION letter to Distributors. Risk Manages, Physicianson/ 9/24/20 stating reason for recall, heath risk and action to take: 1.Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. Complete Attachment 1  Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. This form will be returned even if you do not have affected products at your facility. 4. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facilitys documentation. 5. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com. Surgeon Responsibilities: 1. Review this notification for awareness of the contents. 2. There are no specific patient monitoring instructions related to this recall that are recommended beyond your existing follow-up schedule.
Quantity in Commerce412 units total
DistributionUS Nationwide
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
-
-