Date Initiated by Firm |
December 15, 2020 |
Create Date |
January 07, 2021 |
Recall Status1 |
Open3, Classified |
Recall Number |
Z-0802-2021 |
Recall Event ID |
86946 |
Product Classification |
Prosthesis, hip, semi-constrained (metal uncemented acetabular component) - Product Code KWA
|
Product |
Pinnacle Cups, Part numbers 121701048 121701050 121701052 121701054 121701056 121701058 121701060 121701062 121711048 121711052 121712050 121712052 121712056 121722048 121722050 121722052 121722054 121722056 121722058 121722060 121722062 121731048 121731050 121731052 121731054 121731056 121731058 121731060 121731062 121732048 121732050 121732052 121732054 121732056 121732058 121732060 121732062 121732064 - Product Usage: The PINNACLE CUP devices are part of the Pinnacle Hip Solution for the anatomic reconstruction of the hip joint, which promotes prosthetic joint load and function. |
Code Information |
Lots 8591362 8591361 8591363 8592831 8594131 8593445 8593245 8592830 8596365 8594132 8594461 8596511 8596371 8596372 8596369 8596495 8596510 8597345 8597338 8597347 8597351 8594149 8597340 8597339 8597465 8592165 8596366 8597466 8592829 8597892 8598955 8597898 8598954 8597341 8597346 8597463 8598953 8599439 8594152 8598421 8597342 8599734 8594155 8598956 8599793 8599801 8599791 8599800 8599448 8599792 8599447 8598945 8599657 8599655 8599790 8599669 8599656 8599894 8600091 8599895 8592317 8599668 8599902 8599803 8599796 8600146 8600148 8600173 8600147 8599665 8599797 8599667 8599664 8599670 8600170 8599663 8599798 8601003 8599899 8601399 8601004 8600169 8601641 8599666 8600172 8601411 8601005 8601634 8601656 8601633 8601644 8601002 8601635 8602769 8600174 8601409 8602929 8599659 8602837 8602915 8602914 8602844 8602930 8602836 8602835 8602838 8602833 8602818 8602839 8602779 8602843 8602913 8602926 8602925 8601414 8602771 8600175 8599608 8600115 8599810 8602815 8602793 8603685 8603631 8604745 8603681 8604756 8602778 8604784 8603663 8604758 8604759 8604744 8602785 8604754 8605058 8606232 8605620 8605059 8605619 8605621 8605624 8606223 8605615 8605071 8605616 8605069 8605061 8605622 8606231 8605070 8605062 8604761 8606227 8605618 8605617 8606229 8606469 8604753 8605074 8606470 8606959 8606956 8607746 8608272 8607772 8608040 8607755 8605066 8608045 8608071 8608044 8608294 8608041 8605065 8608308 8608293 8608303 8608661 8608648 8608669 8610116 8608606 8610340 8610105 8610330 8610337 8610115 8610106 8610850 8610338 8608580 8608312 8607759 8608581 8605073 8607748 8608313 8608573 8610858 8610870 8610873 8611591 8611592 8612926 8611624 8611659 8613062 8613060 8613068 8613061 8611614 8613368 8613065 8613371 8613385 8613370 8613064 8613067 8613066 8613480 8613367 8613483 8613063 8612955 8610111 8613506 8613457 8611656 8611658 8611655 8615242 8616171 8615241 8615213 8616176 8616175 8614807 8616172 8616154 8616179 8617537 8616183 8616173 8616177 8616178 8617532 8616156 8616157 8614823 8618203 8618233 8617533 8617551 8618236 8618232 8614824 8618204 8618202 8617553 8617552 8613594 8618231 8618343 8618336 8617580 8613610 8613596 8617562 8613609 8616346 8616344 8618333 8617531 8617567 8616345 8613612 8618355 8618337 8620112 8620113 8621218 8621219 8621202 8617539 8621198 8616347 8617563 8617538 8621908 8621897 8618351 8621904 8621898 8620368 8621525 8620082 8620078 8621945 8621199 8621947 8621940 8621943 8623248 8621213 8623260 8623370 8623369 8623489 8612964 8611660 8623241 8621899 8623240 8621214 8621197 8623520 8622685 8623434 8612949 8623491 8623390 8618342 8612951 8623448 8623389 8623265 8621938 8623542 8623261 8623376 8623541 8623450 8624345 8623474 8624160 8624217 8623476 8624346 8624343 8624340 8624342 8624344 8621211 8625746 8624341 8625772 8625777 8623452 8621210 8622497 8623475 8623447 8617540 8618341 8625747 8625751 8625752 8624348 8625771 8625760 8624164 8624162 8623381 8623453 8623545 8625773 8623454 8623391 8625759 8623375 8626275 8626264 8625735 8626287 8626278 8625761 8625758 8626274 8624005 8627746 8623439 8623443 8623549 8627766 8628221 8623493 8623444 8626277 8623504 8629122 8628016 8623552 8623553 8629126 8628208 8629124 8623466 8621939 8623546 8629443 8629438 8623555 8629425 8629429 8629426 8623478 8629428 8627780 8629431 8629427 8631578 8629421 8631577 8628695 8631910 8629423 8629422 8631911 8629424 8631942 8632367 8631939 8632547 8632636 8632366 8632545 8632638 8632637 8632541 8623548 8631943 8631566 8632615 8631583 8632527 8632543 8632621 8632529 8632540 8629120 8628051 8631891 8632546
8624004 8624006 8632609 8633126 8629108 8633113 8633711 8632533 8628046 8633637 8628673 8629110 8633638 8633130 8633131 8633124 8628690 8628060 8628674 8633634 8633703 8633133 8634179 8634135 8634196 8634197 8634156 8634608 8634190 8634195 8634649 8634648 8634787 8634643 8634646 8634650 8634142 8634194 8634594 8634607 8634188 8634801 8634140 8624001 8634797 8634287 8628693 8634786 8634138 8634629 8635284 8633633 8634796 8635091 8635099 8635054 8634791 8635057 8636602 8635283 8636601 8635092 8634802 8634807 8636645 8636600 8636614 8636603 8637696 8636598 8636615 8635058 8636641 8634788 8637697 8636613 8637965 8634633 8635103 8637964 8637699 8637963 8637693 8637698 8637695 8637694 8637686 8637700 8637975 8634626 8634647 8634165 8634638 8634166 8634624 8634640 8634628 8634154 8639057 8639072 8637977 8639641 8639055 8639056 8639657 8638400 8639656 8637718 8632526 8634286 8639658 8639907 8639906 8639654 8639632 8639927 8639896 8639897 8639926 8640011 8640012 8639631 8624002 8639642 8639883 8637719 8638403 8639930 8639912 8639916 8640043 8639898 8639911 8640094 8639633 8640115 8639917 8640551 8641633 8641612 8641611 8640092 8639934 8639933 8640099 8640659 8641613 8640743 8641621 8639909 8640052 8640018 8634634 8640054 8641617 8640050 8640055 8641843 8641620 8640093 8642290 8640047 8641837 8641842 8640051 8640660 8640561 8641610 8641638 8640559 8640562 8642302 8642301 8642298 8641829 8642977 8641637 8642299 8642967 8642300 8642978 8642982 8642485 8642303 8629416 8643873 8642981 8642965 8643880 8643878 8643879 8644446 8643881 8644445 8642287 8644447 8642979 8644449 8642980 8644442 8644469 8644448 8644450 8644470 8644466 8644467 8644849 8644844 8644845 8645053 8644846 8644847 8645060 8645102 8645239 8645527 8645177 8644848 8645532 8645533 8647078 8647077 8632633 8645244 8645542 8645243 8645241 8645242 8647075 8647073 8647248 8645183 8632622 8647145 8645073 8647252 8647249 8647250 8647251 8647854 8647851 8645075 8647074 8645074 8645179 8645186 8647852 8647850 8647862 8647864 8647859 8647076 8649008 8647853 8648199 8649006 8649009 8645543 8649007 8649830 8649854 8650226 8650227 8649855 8650452 8650451 8650200 8650201 8650199 8650202 8650203 8649856 8649857 8649849 8650244 8650208 8649846 8649860 8650207 8650211 8649848 8650447 8651824 8649863 8650454 8650228 8649847 8650229 8650428 8635102 8653310 8653337 8653308 8649888 8651832 8651835 8651830 8653309 8652736 8653322 8653321 8651803 8650213 8651827 8651829 8650212 8651801 8653311 8651802 8651800 8651815 8651804 8651805 8651833 8654878 8654790 8654775 8654317 8654325 8654326 8654319 8654322 8654867 8654784 8654320 8654873 8654870 8653344 8654796 8654813 8655000 8654880 8654872 8654865 8651808 8653689 8653692 8653690 8651926 8651831 8654333 8654776 8654787 8656863 8655304 8654321 8654318 8654324 8654877 8654774 8655346 8656868 8655293 8655312 8655341 8656862 8655332 8655337 8655305 8657720 8657691 8655299 8657717 8655349 8658044 8658046 8658420 8658047 8658045 8658424 8658050 8658070 8659260 8658419 8658071 8658423 8658049 8655343 8655302 8658069 8658068 8658073 8658076 8659272 8658074 8659257 8655303 8658421 8658432 8658048 8654785 8658412 8658433 8659274 8659256 8659433 8659552 8659263 8658437 8659375 8659435 8661588 8659376 8658417 8658436 8661590 8660226 8658416 8659554 8658435 8659360 8655294 8659381 8659378 8659363 8653688 8661587 8663052 8640560 8663057 8663049 8662999 8663901 8667103 8665073 8667884 8658425 8668823 8668824 8667881 8667882 8668802 8667435 8668803 8668806 8662506 8668313 8668317 8670283 8670284 8670286 8670340 8670278 8670280 8670279 8670289 8670339 8670350 8670322 8670282 8670358 8663047 8670281 8670321 8670695 8670697 8670722 8670696 8670694 8670699 8645105 8651849 8645104 8648117 8668884 8642377 8670218 8648132 8670199 8645103 8668889 8668890 8651847 8648138 8670217 8670201 8672989 8672544 8672973 8671234 8674590 8671235 8671241 8674591 8671231 8671237 8674612 8675721 8675724 8675725 8675723 8675722 8675177 8676417 8676420 8676412 8676418 8668883 8642383 8676878 8676881 8676902 8676892 8676879 8676891 8642382 8676882 8642381 8670204 8674076 8668874 8670203 8648134 8645106 8653679 8674103 8674116 8674114 8674101 8670202 8674107 8641856 8641854 8651848 8674086 8674115 8674100 8670205 8668892 8668893 8670228 8674084 8674088 8674089 8668887 8642378 8670206 8674102 8674087 8674083 8648136 8645110 8645107 8676908 8677438 8677437 8676907 8678562 8670215 8678569 8678563 8648137 8641858 8641857 8674110 8640274 8674085 8642380 8676921 8677451 8678962 8670198 8667952 8640261 8668872 8645113 8667954 8641855 8670216 8640273 8642379 8641859 8676910 8676911 8676984 8678568 8676985 8677170 8670207 8677162 8677161 8677165 8670208 8678566 8679318 8679319 8679553 8677160 8677180 8680271 8677172 8677163 8677174 8677185 8680272 8677166 8677171 8680275 8677169 8679358 8678565 8680276 8680279 8680278 8677164 8680293 8680273 8680232 8681879 8681294 8680263 8681293 8680233 8681329 8681304 8681402 8681292 8684030 8684031 8684026 8684025 8682212 8684021 8681308 8679555 8684024 8684057 8684323 8684027 8685422 8684023 8685421 8684321 8686369 8684059 8686368 8685441 8687888 8686367 8685443 8685423 8685439 8686398 8686370 8684347 8684058 8687882 8687881 8682222 8681770 8681771 8682229 8682228 8680292 8680312 8680274 8677173 8681775 8682227 8682225 8680288 8682226 8684105 8687821 8681795 8681780 8686403 8686399 8687832 8687834 8680300 8687831 8682230 8682244 8680301 8680294 8682245 8690154 8690153 8690152 8685021 8687833 8682224 8690145 8690149 8690150 8690158 8687877 8687034 8681794 8680308 8682220 8690155 8682217 8682218 8682219 8688817 8687032 8681777 8685022 8684106 8687822 8684099 8682223 8684366 8690730 8681784 8690733 8687824 8690701 8687828 8685029 8685024 8681786 8680277 8685028 8687829 8670209 8690732 8690741 8690738 8690739 8690742 8684360 8684359 8680298 8684100 8690720 8682243 8684115 8681781 8681798 8690731 8680299 8688756 8690157 8680296 8681782 8681779 8684364 8680309 8685023 8690740 8681783 8681796 8684117 8684371 8684098 8687033 8684370 8684367 8690673 8688814 8684101 8684382 8684374 8687880 8688811 8687879 8692681 8688755 8690147 8686401 8680307 8690684 8682242 8687029 8688752 8694560 8695222 8692851 8692682 8695217 8694570 8696203 8696214 8690690 8695221 8691902 8695225 8692876 8692877 8694563 8692874 8692878 8694556 8690674 8690151 8690166 8690668 8697330 8695340 8693162 8690689 8690143 8695344 8690683 8695224 8695220 8692680 8697329 8697869 8697319 8692849 8690932 8697898 8693163 8697866 8694565 8692850 8695223 8690902 8694562 8690930 8686400 8690901 8690908 8697334 8695349 8695347 8693992 8697327 8697333 8693999 8697328 8697332 8693991 8697318 8696293 8692897 8696292 8690728 8692889 8684103 8685027 8682231 8697364 8692888 8697928 8697363 8697382 8690708 8684373 8696278 8684372 8697381 8697932 8699305 8697933 8692890 8697362 8684377 8697934 8684104 8680314 8690931 8684368 8684113 8696291 8697377 8697936 8684376 8684375 8690710 8684102 8688846 8681785 8697929 8681778 8680297 8690711 8684097 8692906 8692907 8699337 8699312 8699307 8697931 8697374 8697378 8696328 8696288 8692905 8702366 8697930 8700705 8700703 8688850 8693150 8693147 8697904 8695350 8699274 8702333 8702312 8699273 8702306 8692679 8700670 8702916 8693148 8698246 8702324 8702317 8697903 8702285 8695345 8700668 8702271 8702276 8702908 8702278 8698256 8702286 8698257 8702309 8702321 8684381 8699338 8702377 8678567 8699310 8699309 8699299 8697380 8700704 8696290 8702367 8692651 8692652 8690709 8692882 8692902 8699306 8697365 8702368 8692908 8692904 8702883 8699311 8692901 8703314 8702380 8702376 8702863 8702369 8702370 8697358 8702876 8702365 8702372 8700702 8702371 8700707 8702868 8703317 8703312 8702870 8702871 8703316 8702869 8703313 8702332 8702313 8703259 8699281 8696202 8697902 8702340 8703301 8688753 8702907 8705013 8703357 8702277 8702897 8695218 8698255 8705014 8702319 8703261 8695219 8702320 8695351 8703272 8702322 8700669 8697896 8704967 8706235 8697315 8697863 8697865 8697882 8704974 8704969 8705004 8702898 8705003 8702341 8704972 8690938 8690936 8706210 8707596 8707595 8706299 8707597 8702893 8706236 8702318 8707779 8707784 8702865 8707783 8702884 8690691 8708557 8708554 8703311 8702395 8703315 8680295 8707778 8708558 8708556 8707776 8707781 8708679 8708680 8708550 8691903 8690910 8702917 8703293 8702279 8703292 8702314 8703299 8707802 8691931 8695348 8690937 8698247 8702281 8690168 8703966 8702299 8697868 8704970 8703264 8704973 8706237 8704966 8702273 8703303 8702325 8700654 8700653 8707570 8708527 8703942 8707836 8707826 8707569 8706222 8707807 8703277 8700657 8700645 8708528 8703300 8707588 8708536 8707835 8711378 8708513 8708510 8708511 8707833 8705007 8711394 8711377 8708514 8707589 8711371 8707590 8711396 8707843 8707571 8712284 8708509 8711398 8708480 8711397 8711374 8713204 8712202 8709063 8712198 8713203 8709066 8713206 8709064 8712263 8712197 8712260 8712262 8709065 8709062 8709069 8709059 8709058 8713207 8712200 8711343 8713891 8713894 8713915 8713895 8713885 8713882 8713886 8713912 8713883 8713890 8704980 8711355 8707808 8711356 8705012 8711368 8707794 8712285 8712299 8711373 8707837 8708479 8707810 8707809 8712303 8712267 8705008 8705011 8712266 8712304 8712265 8705009 8707811 8708512 8713828 8711393 8713811 8714097 8714348 8713805 8714096 8712298 8713874 8712476 8703262 8714781 8708529 8702308 8712302 8707834 8711395 8714355 8716070 8713881 8716066 8713810 8714359 8713873 8713804 8714104 8714357 8714103 8714101 8716067 8716072 8714356 8715040 8714358 8717454 8719357 8715906 8718506 8716059 8714099 8716065 8717471 8716074 8719241 8714098 8716073 8716071 8712477 8719268 8714105 8717465 8717693 8714100 8707812 8704978 8709105 8708481 8719307 8712272 8719283 8716106 8719337 8714102 8717695 8719284 8717682 8719339 8719294 8719864 8719293 8719341 8722013 8719865 8722036 8719299 8722009 8722014 8706221 8719895 8722003 8719912 8719334 8722034 8719848 8714367 8719913 8722024 8719896 8722023 8719297 8722030 8714077 8719866 8722016 8716076 8719900 8719327 8722004 8719899 8714086 8719897 8719247 8719871 8719859 8719860 8722015 8719333 8722012 8719863 8722037 8714080 8714085 8722029 8719867 8722031 8723055 8722011 8719295 8722035 8722033 8719340 8717692 8717804 8719246 8717806 8719245 8719326 8719324 8714081 8714082 8719342 8719243 8721987 8721988 8721989 8719927 8719356 8719926 8721991 8722949 8719928 8722948 8722943 8719923 8724027 8719924 8721981 8723872 8722941 8719925 8725120 8724028 8723871 8723866 8723048 8714084 8717697 8719238 8719869 8719242 8721999 8714083 8714079 8714076 8725047 8723893 8719335 8723923 8723031 8725044 8719325 8723911 8719244 8717805 8719237 8717694 8719298 8725088 8723924 8725053 8726040 8717698 8723904 8719267 8719306 8717467 8719901 8726001 8719873 8725086 8721993 8725052 8722010 8719239 8725079 8725075 8725077 8725087 8719336 8725045 8726000 8726043 8726211 8704992 8721998 8709106 8690934 8726002 8725025 8725027 8725029 8722018 8722001 8719870 8726041 8727759 8726008 8725031 8719296 8716075 8723060 8727752 8726203 8719301 8726009 8723922 8725030 8726044 8723912 8719872 8723050 8727761 8726219 8723059 8723054 8725050 8727715 8727760 8728843 8727745 8726216 8728844 8728866 8726053 8726010 8727746 8725109 8723052 8717683 8728861 8728862 8726222 8730155 8728865 8727709 8728831 8727705 8728832 8727710 8728833 8723062 8730175 8726039 8723057 8726042 8719868 8719338 8723903 8730825 8730824 8723899 8730808 8728849 8725110 8726032 8723905 8728848 8725046 8726212 8725051 8730150 8728847 8726215 8717466 8726204 8723907 8725048 8726054 8723906 8730184 8730807 8730168 8730185 8728842 8728867 8730167 8714779 8719898 8726011 8731250 8717696 8731249 8731240 8714780 8731252 8733377 8733376 8733534 8733375 8731222 8731226 8733535 8716098 8723032 8728874 8728830 8703359 8730891 8730915 8730907 8730890 8730857 8731351 8731615 8730868 8733924 8730862 8730841 8730842 8733374 8733996 8735355 8730856 8730916 8733385 8731354 8735354 8735156 8733373 8735164 8735912 8735356 8735165 8736696 8735930 8735166 8735934 8736697 8735913 8735942 8735933 8735932 8735914 8735931 8735365 8735952 8730870 8730872 8733402 8733923 8731352 8730831 8731339 8735838 8735854 8733397 8730917 8735876 8730864 8735875 8733925 8731355 8730869 8735884 8735858 8736692 8735929 8736722 8736721 8735357 8736704 8735937 8735155 8736720 8736705 8737153 8736880 8739283 8739256 8739293 8736706 8736907 8739266 8736731 8735951 8736879 8739287 8736895 8736698 8739258 8739255 8739257 8736693 8739267 8735936 8739259 8739268 8739265 8740382 8740383 8740380 8736719 8735928 8739260 8741748 8736915 8731251 8740384 8741722 8736723 8741719 8740368 8740385 8739262 8736714 8735358 8741747 8741720 8741741 8741716 8741718 8742758 8742332 8742772 8742313 8742337 8742219 8742339 8742336 8742773 8742333 8742774 8742759 8742763 8742335 8742762 8742221 8742314 8723030 8742218 8742222 8740369 8743092 8742642 8735881 8735392 8742644 8735390 8735856 8733532 8739317 8739329 8735853 8740335 8735376 8736931 8735849 8735173 8735169 8735852 8736923 8735837 8739330 8739327 8730840 8735855 8746071 8733922 8735883 8735869 8730829 8730863 8735859 8733398 8746070 8733528 8735379 8746506 8746044 8746060 8741757 8746493 8746492 8747102 8747505 8746491 8747086 8746505 8747520 8747506 8743182 8747518 8746502 8735138 8746046 8743094 8747513 8748090 8748089 8742923 8747574 8743093 8742334 8742641 8748152 8748057 8748154 8748119 8748732 8748730 8747519 8748067 8748122 8748121 8748065 8746048 8748069 8748153 8748120 8736930 8735391 8740357 8747091 8748068 8748070 8743184 8740358 8746043 8746061 8747090 8748142 8748061 8746045 8748145 8748963 8748063 8748965 8743183 8748075 8748904 8748074 8748885 8748966 8748964 8748741 8742234 8748742 8743181 8748903 8748962 8748905 8739323 8747530 8747525 8735877 8735378 8739328 8747526 8735170 8735380 8735377 8735839 8743213 8735840 8748975 8730830 8746072 8740360 8740334 8746068 8743214 8747531 8746075 8746074 8746254 8748976 8735882 8747533 8735844 8748881 8748935 8748961 8751156 8751161 8751157 8751168 8751162 8751154 8751167 8754262 8751691 8751191 8754266 8752535 8752533 8752534 8752566 8751674 8752520 8751196 8755278 8751689 8751182 8752562 8755276 8755937 8751181 8755251 8755901 8754573 8755879 8755933 8754247 8755919 8751695 8752564 8751677 8751688 8755880 8751671 8752517 8755914 8752519 8752556 8754580 8752512 8755934 8749276 8755917 8755280 8757333 8748880 8757340 8757334 8738967 8754578 8754251 8752565 8755900 8757314 8757328 8755275 8755893 8757341 8751678 8757365 8757316 8757343 8754237 8757326 8757380 8752568 8757327 8757381 8754579 8757347 8757335 8757385 8757329 8757330 8757373 8751183 8757383 8757338 8752505 8757319 8756012 8758100 8755920 8755279 8755899 8751672 8757658 8757384 8752521 8757389 8757346 8757315 8751676 8755894 8757382 8758107 8758102 8758101 8758082 8751690 8755267 8751694 8758083 8757345 8758089 8755277 8758098 8751673 8751670 8758084 8758105 8755268 8751184 8758090 8758091 8758087 8758108 8758092 8758081 8758103 8758086 8749278 8757339 8758112 8757325 8755888 8761584 8757331 8758088 8758085 8757379 8761591 8755921 8757657 8751679 8755881 8761592 8761586 8761588 8757375 8755455 8754223 8761620 8754224 8758016 8755437 8751131 8761631 8751132 8758232 8754230 8748978 8754231 8751146 8758015 8757413 8755451 8748882 8754238 8757423 8755435 8755436 8755454 8758230 8757406 8749293 8762963 8758231 8758036 8733400 8740359 8756077 8747079 8757374 8735845 8763793 8757424 8763797 8763005 8763948 8763915 8763006 8763919 8762999 8763003 8763010 8763821 8763011 8763914 8762290 8755995 8763917 8766913 8763790 8763791 8764504 8766908 8763795 8763794 8763947 8763946 8763944 8761589 8763822 8751185 8761573 8763913 8763820 8762263 8762287 8763916 8763017 8762272 8763824 8763874 8762289 8766972 8762285 8763002 8763008 8763918 8763009 8763868 8763825 8763007 8762274 8761587 8763876 8763945 8763001 8768507 8768532 8766915 8766921 8763912 8763012 8761585 8763016 8766920 8733390 8766923 8763943 8766947 8766930 8763875 8762288 8769054 8757344 8763942 8764501 8763796 8764500 8768525 8770112 8768493 8770109 8770110 8763955 8770113 8770107 8770193 8768912 8768911 8770187 8766929 8768941 8770153 8770154 8770854 8771175 8770841 8770853 8770106 8770826 8770025 8770186 8768942 8770808 8763792 8771176 8763873 8762278 8771185 8774585 8763823 8770855 8774593 8771160 8774592 8774580 8774584 8763004 8774587 8774550 8774557 8768935 8770115 8768936 8774556 8766916 8775407 8774594 8768527 8775406 8771184 8774579 8774588 8770840 8771177 8775411 8763887 8770189 8768913 8770811 8776956 8776959 8775452 8775495 8776958 8776957 8770024 8775494 8768907 8770827 8768509 8768508 8774586 8775410 8768495 8771162 8762297 8775486 8770108 8775898 8768529 8777651 8768524 8776428 8776673 8770114 8768496 8775419 8770073 8775409 8775487 8776617 8775484 8775481 8775408 8762284 8735836 8770825 8768934 8776730 8780744 8776425 8770852 8775485 8763954 8771186 8780745 8780767 8775476 8780768 8777660 8780773 8780628 8776725 8780772 8780630 8776423 8780769 8770075 8781767 8781766 8781764 8776944 8782561 8782563 8782871 8782562 8782872 8782870 8782874 8782873 8776406 8775865 8771178 8782824 8782838 8782840 8775405 8782843 8776616 8775482 8783350 8776421 8775864 8774555 8783352 8781815 8780758 8783351 8782533 8780759 8782836 8783304 8782539 8782866 8783757 8782841 8776343 8785897 8783313 8781765 8785891 8782864 8785895 8783320 8774549 8785694 8785914 8782839 8782844 8785697 8787174 8776614 8781786 8786488 8787202 8786674 8787176 8782538 8786053 8787178 8788314 8782837 8787204 8787179 8788852 8785696 8783746 8788315 8786057 8786673 8786502 8786667 8786496 8786060 8786676 8783346 8787175 8787200 8786495 8788855 8783749 8789053 8786505 8787205 8788853 8785698 8781787 8789049 8788850 8789051 8783334 8782540 8788374 8780655 8787180 8789165 8783353 8788859 8788318 8777673 8787177 8788298 8786681 8789163 8789164 8785913 8786052 8787201 8786683 8785894 8782863 8783324 8787214 8786694 8786682 8782865 8790030 8786685 8786700 8786703 8785898 8786684 8789039 8789147 8783305 8789148 8786695 8790031 8792622 8789864 8785702 8792589 8789048 8786668 8781782 8788854 8786494 8789050 8786501 8793432 8782821 8789862 8789170 8793021 8793116 8793484 8792612 8793115 8793699 8793698 8793702 8792621 8792623 8776855 8789867 8776854 8793112 8793703 8793187 8793111 8792620 8793701 8789848 8789162 8793423 8780636 8793417 8780743 8793709 8789167 8782517 8789847 8789916 8776615 8775897 8792574 8789837 8789912 8775483 8789059 8781781 8782845 8792600 8793726 8788856 8792994 8793052 8789915 8787173 8789914 8788316 8794651 8792578 8793727 8793436 8793729 8792601 8793708 8796486 8787203 8789168 8789052 8789846 8796488 8793435 8788307 8793193 8793018 8793728 8793438 8789841 8789845 8796487 8786063 8789856 8793023 8789169 8796485 8793020 8788851 8793434 8796476 8780653 8792588 8794649 8793433 8789166 8796927 8793022 8796477 8793733 8793460 8789865 8789866 8793745 8795275 8793704 8794671 8794662 8795282 8794673 8794666 8793700 8795276 8794664 8796641 8795278 8794672 8796766 8793483 8795277 8798563 8793747 8796776 8798792 8796924 8796923 8798552 8798536 8796926 8794653 8793710 8798553 8793019 8798541 8796919 8793452 8793416 8799139 8793739 8794663 8798946 8794652 8783344 8798968 8798969 8798963 8798975 8798971 8799251 8796489 8799151 8798962 8799155 8799154 8798964 8799245 8781780 8799238 8799227 8799226 8800147 8800748 8800136 8793044 8800763 8798814 8800150 8800142 8799246 8801747 8800749 8800764 8769053 8794676 8801745 8800153 8789868 8800762 8800768 8801799 8801812 8801800 8798978 8798979 8801754 8801781 8801746 8801758 8801780 8801751 8801783 8801750 8801786 8801787 8801752 8801785 8801795 8796767 8801811 8802851 8802849 8802846 8801810 8801808 8802850 8802794 8801777 8801748 8803027 8802793 8802792 8801788 8803036 8803031 8803045 8802796 8802797 8803053 8802795 8804715 8803044 8803046 8803052 8805016 8804749 8804750 8804748 8776410 8804758 8803029 8802798 8805015 8805019 8803051 8805465 8805453 8805463 8805464 8805437 8803050 8805803 8805789 8805438 8805451 8805425 8805424 8804753 8805813 8805806 8805811 8805810 8807146 8807131 8807157 8807156 8805843 8807138 8807130 8805839 8807179 8805840 8805831 8805830 8807180 8805838 8807178 8807422 8807442 8807441 8807431 8807444 8807443 8807145 8807404 8808135 8808130 8808265 8807407 8808236 8808134 8808256 8807388 8808257 8808254 8808255 8808259 8807410 8809250 8807445 8809251 8807423 8810008 8808767 8808258 8808725 8805804 8805805 8808768 8810258 8810263 8808755 8808727 8808756 8811098 8811116 8810236 8808757 8810522 8810254 8810253 8811084 8810259 8793188 8810505 8810496 8810514 8810516 8811059 8810504 8808758 8811117 8810007 8811090 8811089 8810511 8811083 8810252 8810507 8811904 8811100 8811099 8811101 8811085 8812347 8805780 8812344 8812345 8811901 8811928 8811927 8811060 8811094 8811093 8811903 8810014 8811902 8812343 8810510 8811926 8810515 8812346 8812348 8812327 8801809 8811992 8811981 8810279 8811954 8812002 8812001 8811993 8811997 8811998 8811994 8811935 8811936 8811934 8811938 8814963 8814263 8812696 8814332 8812694 8814323 8815670 8815924 8815669 8814304 8812326 8812350 8815921 8814839 8812761 8815949 8815944 8815948 8812760 8816816 8815947 8816785 8817586 8817399 8817400 8817388 8817410 8817427 8817424 8817237 8817398 8817386 8817233 8817402 8815923 8814945 8815922 8817420 8817425 8812357 8814305 8815666 8817603 8817228 8817421 8815932 8817422 8812736 8817223 8817426 8817389 8817411 8817390 8817431 8817221 8818036 8817234 8817428 8817910 8817222 8817905 8817906 8817904 8818037 8817902 8817407 8818022 8818031 8817227 8817423 8817903 8817412 8818030 8815933 8817673 8817672 8817675 8817216 8817893 8817895 8807158 8812358 8818353 8814946 8817674 8821784 8817587 8818350 8818340 8817669 8817671 8822006 8817676 8822007 8818352 8817217 8822284 8822285 8822287 8818023 8822018 8814253 8822027 8817408 8822014 8822013 8822740 8822269 8822751 8822763 8822019 8822031 8822286 8817211 8822278 8822741 8817215 8822030 8823016 8823003 8823004 8823017 8823018 8822868 8817599 8822012 8823520 8823521 8823021 8823516 8822869 8823515 8823020 8822766 8821782 8822870 8823607 8822765 8823600 8822756 8823517 8822760 8823603 8823556 8823608 8823587 8823604 8822759 8823605 8823910 8823557 8823553 8822764 8823918 8823921 8823924 8823564 8823562 8823554 8824163 8822274 8824162 8823916 8823909 8823923 8823561 8823922 8823917 8823585 8823914 8823586 8823555 8824164 8824192 8823565 8822847 8823915 8822757 8824368 8823522 8824022 8824020 8824021 8824190 8824351 8824345 8824367 8824348 8824349 8824347 8824353 8823019 8826047 8824355 8826046 8824350 8824191 8826360 8826366 8826079 8826370 8826363 8826364 8826361 8826362 8826546 8828185 8826559 8824354 8824356 8826562 8824352 8826556 8826558 8824364 8826547 8826065 8826557 8826048 8828801 8827820 8826543 8828822 8827821 8829082 8828807 8826067 8826050 8826064 8828791 8827819 8828804 8828790 8828805 8826560 8830093 8829011 8830532 8830526 8830533 8830104 8830531 8830549 8829087 8828802 8826066 8830571 8828180 8826561 8830557 8828806 8830570 8830547 8830544 8830546 8829086 8830527 8830857 8830855 8830530 8830529 8830528 8830111 8830550 8830881 8830548 8830556 8830109 8830875 8830879 8834889 8830866 8834880 8830551 8829647 8830743 8830103 8830573 8830567 8830569 8830863 8831364 8830572 8830568 8831338 8830746 8834924 8831337 8831334 8830083 8831336 8830894 8830747 8830891 8830084 8835303 8834892 8835302 8831372 8830895 8830561 8835304 8835911 8830563 8834878 8834919 8835550 8834877 8835571 8830755 8834888 8835570 8835569 8835938 8835942 8835536 8831374 8835956 8835573 8835957 8835274 8831384 8835279 8835269 8836643 8829651 8834917 8835912 8835305 8830749 8836300 8835280 8836329 8835951 8836655 8835554 8835275 8834882 8834881 8836327 8830110 8835577 8830744 8830745 8836293 8836661 8835271 8836878 8835537 8836883 8835962 8836877 8836868 8835286 8835278 8835287 8836869 8837111 8836874 8836875 8836641 8834918 8830880 8834890 8836876 8836873 8835553 8835578 8834911 8836656 8836652 8834910 8836328 8835313 8835927 8836654 8836653 8835314 8831335 8835934 8835933 8836326 8837330 8840108 8830759 8837197 8837340 8837104 8837079 8840091 8837106 8840092 8837136 8840155 8840599 8837118 8841029 8841094 8839905 8841057 8840615 8841031 8837125 8841037 8840608 8837116 8841093 8840616 8841028 8841345 8841589 8807129 8841466 8841468 8840617 8841469 8841467 8837145 8840109 8841097 8840158 8841315 8837168 8841058 8841095 8837331 8840157 8840609 8842606 8842604 8840110 8841962 8842593 8841963 8841979 8842596 8842660 8842599 8842594 8841940 8841980 8842597 8842654 8841945 8841978 8841938 8842598 8842664 8842702 8842711 8841964 8842658 8842663 8842662 8841961 8841470 8842707 8842713 8842693 8843132 8842701 8842692 8843130 8842723 8842607 8842724 8842610 8843307 8843149 8842703 8843127 8846622 8843299 8842160 8843298 8843140 8843136 8843150 8846886 8846615 8846619 8846906 8842694 8846916 8846888 8846914 8846966 8846964 8846965 8846927 8847816 8847110 8847109 8847107 8847599 8847592 8848369 8848371 8848355 8848467 8848357 8848673 8847726 8847705 8847706 8849245 8848977 8848697 8849101 8848707 8848675 8849488 8849253 8849522 8848695 8849494 8848577 8848674 8849495 8849479 8849518 8849489 8849513 8849523 8849485 8847725 8848385 8849530 8849242 8848386 8849527 8849241 8849528 8849531 8849544 8849819 8849821 8849820 8849517 8849823 8849834 8849520 8847723 8849822 8849831 8850139 8850177 8850176 8849836 8849844 8849486 8849843 8849543 8850140 8849842 8849835 8849545 8850183 8852835 8852839 8852834 8850181 8852846 8852849 8852827 8853168 8853171 8852848 8853178 8852850 8852847 8852836 8853170 8852844 8852840 8852828 8852832 8853443 8853169 8853166 8853444 8853448 8853179 8853167 8852831 8853294 8853447 8853442 8853451 8853450 8853452 8853256 8854026 8854009 8853272 8853273 8854025 8854003 8854008 8854013 8854491 8854833 8854835 8854826 8854877 8854843 8854873 8854832 8854888 8854875 8854836 8854876 8854823 8854844 8854824 8854872 8854007 8856186 8856182 8856183 8854871 8854522 8854027 8854473 8855032 8854002 8854827 8854855 8854492 8854856 8854508 8854900 8856566 8856609 8856488 8856570 8856604 8856562 8856606 8856605 8856492 8856574 8859444 8859428 8856573 8854509 8854899 8854901 8859475 8859477 8853457 8856184 8854513 8854510 8859476 8860443 8860424 8860053 8860059 8860439 8860438 8860252 8861724 8862500 8859882 8859463 8861725 8862689 8860520 8862522 8861687 8862796 8862807 8862509 8854845 8856608 8862496 8859887 8859881 8856489 8862499 8859895 8859913 8862794 8862793 8862680 8862808 8861685 8862505 8862497 8862779 8861683 8856607 8862781 8862892 8860518 8860517 8859445 8862488 8856572 8862514 8859446 8859885 8859899 8865315 8862891 8862901 8862685 8862504 8862893 8859457 8859432 8859894 8862523 8862902 8862812 8862679 8862783 8862894 8865580 8856564 8862897 8862489 8865570 8865316 8865302 8860251 8859906 8859892 8862795 8859900 8856498 8859458 8859888 8861723 8866068 8865304 8865303 8866074 8859893 8856490 8865581 8866075 8865587 8866207 8866204 8865582 8866303 8866308 8866304 8866305 8866898 8866307 8866198 8865578 8866306 8865583 8862797 8866197 8861688 8861722 8867427 8867436 8867303 8867608 8867428 8867607 8867437 8867298 8867806 8867800 8867435 8867803 8867815 8867807 8867810 8867297 8867814 8870955 8867809 8867808 8867602 8870967 8870966 8870947 8870945 8870941 8871901 8870950 8870948 8870942 8870946 8870943 8872273 8871897 8871885 8872272 8871898 8872270 8872267 8872278 8871902 8871886 8871882 8871889 8871881 8872271 8870949 8873471 8873074 8873435 8873462 8873448 8873454 8873446 8873456 8873445 8873470 8873443 8873460 8873457 8873444 8873461 8873450 8873900 8876308 8873956 8873449 8876320 8876318 8873952 8873484 8877334 8877332 8876293 8873899 8873447 8873895 8877653 8877865 8877867 8877866 8873962 8877894 8878902 8878723 8879143 8879142 8879144 8877829 8877827 8877825 8877320 8878724 8879086 8878753 8879094 8879095 8877828 8877333 8877319 8877345 8879801 8881348 8877824 8877826 8881349 8879800 8881624 8881625 8883462 8883464 8883465 8883473 8884721 8884754 8878752 8884749 8884755 8884750 8883445 8883504 8883481 8883486 8883446 8883482 8883503 8884742 8885595 8883485 8883483 8885611 8885610 8884741 8886109 8885160 8884739 8883484 8884735 8884734 8884740 8886053 8884738 8886046 8886061 8885602 8885601 8886056 8886063 8886047 8886055 8878783 8883644 8886103 8889788 8889794 8886216 8886220 8886223 8889473 8886116 8883665 8886214 8886210 8886225 8886219 8889496 8884769 8891136 8890294 8891137 8890852 8889797 8890853 8884733 8891220 8886215 8889474 8889495 8889472 8889471 8878781 8885619 8886106 8890970 8883663 8886107 8883657 8883664 8891236 8891141 8890855 8891138 8891398 8891407 8891221 8890854 8889787 8889790 8891396 8890890 8885625 8889484 8891235 8878782 8890937 8890886 8890889 8891399 8892063 8892601 8891139 8892451 8890284 8892448 8890885 8892452 8892470 8892469 8892475 8892064 8892060 8892476 8889688 8889494 8890977 8892202 8886211 8890976 8892170 8886217 8892203 8892171 8889497 8891400 8892162 8892602 8892599 8895886 8893356 8893347 8895761 8895746 8892618 8895760 8893348 8892467 8895910 8892468 8893354 8895762 8895915 8893343 8895911 8893337 8895898 8895751 8890887 8895897 8892495 8885161 8895916 8895753 8895901 8895752 8895754 8892204 8895767 8892499 8895750 8895902 8896432 8895766 8896437 8896444 8893355 8892598 8886224 8893338 8896939 8897787 8896942 8896940 8895784 8897737 8897738 8896436 8897735 8897786 8895768 8898207 8896335 8896347 8898203 8895884 8896902 8896904 8895896 8898199 8896352 8896430 8895755 8896336 8898189 8898205 8896878 8896903 8898348 8893357 8898188 8898183 8898182 8896423 8898349 8900448 8898357 8896420 8898174 8893363 8895903 8898210 8898173 8896907 8896906 8896422 8898356 8900447 8898209 8902830 8902829 8902773 8897706 8896879 8892496 8898201 8898184 8896431 8902777 8897694 8898202 8902758 8896346 8902766 8896421 8898345 8902802 8896905 8902776 8897736 8891224 8895769 8897795 8902767 8896927 8896456 8896929 8885620 8895885 8903339 8896419 8902764 8902867 8902868 8902869 8898494 8897796 8896455 8896932 8898499 8897768 8896933 8898428 8902939 8902903 8898429 8902911 8898437 8902801 8903345 8891216 8890873 8897739 8902953 8904011 8905165 8902904 8904974 8904986 8902763 8904973 8908193 8903658 8903657 8900463 8903344 8902759 8902922 8903955 8904992 8902772 8897769 8902938 8902913 8905161 8902907 8905060 8902909 8902954 8903961 8905034 8905066 8902910 8903963 8902914 8904008 8904002 8905167 8908140 8904004 8904007 8898498 8902908 8902906 8902940 8902967 8910551 8904001 8905163 8908172 8905164 8910549 8904003 8905033 8902921 8910999 8909641 8910125 8908432 8908408 8892603 8908184 8910124 8908602 8911031 8911090 8908192 8896353 8913971 8913970 8911689 8909659 8910479 8910903 8911679 8911082 8911089 8911041 8908186 8908187 8911664 8911083 8908404 8910916 8909660 8885612 8913969 8910930 8911091 8902902 8914655 8910932 8908087 8908416 8908431 8909642 8909658 8909661 8910933 8913963 8911678 8908604 8911030 8892604 8910929 8910939 8908597 8910902 8911092 8911029 8914641 8911665 8908183 8908606 8916425 8916514 8916497 8916494 8916214 8916435 8916434 8916515 8916420 8916498 8916496 8916477 8916506 8914967 8914002 8914670 8910568 8909735 8916557 8908171 8908156 8909738 8908141 8909737 8910538 8909751 8910537 8908142 8905064 8905058 8915014 8903962 8915021 8911023 8911012 8916859 8905065 8910569 8914300 8914288 8910542 8911529 8914671 8915019 8914286 8911549 8916854 8908157 8922122 8911528 8922121 8922118 8922254 8922120 8922229 8922227 8922230 8922458 8916959 8921552 8921467 8910539 8922231 8922123 8910543 8922255 8910550 8915022 8908139 8915958 8908158 8911540 8892186 8921480 8909769 8921479 8922282 8905160 8916856 8911016 8911009 8915957 8911530 8911531 8910541 8916855 8916563 8910540 8909734 8915018 8911539 8911532 8916860 8915013 8909736 8923219 8923686 8923683 8923846 8923220 8923223 8923056 8923999 8923996 8926053 8926060 8924000 8923997 8923847 8923995 8923685 8923849 8923994 8926051 8923035 8921475 8924199 8926365 8926367 8926369 8923684 8924198 8926052 8923998 8921476 8924196 8926055
8926366 8916572 8922192 8916573 8911024 8923051 8923216 8921551 8924200 8923660 8926888 8926781 8911010 8926757 8923877 8926782 8914287 8916861 8908148 8923876 8916862 8908147 8908153 8911541 8916863 8905059 8916857 8928034 8928036 8928035 8923231 8927761 8928037 8927958 8929153 8928047 8929433 8928048 8929131 8923848 8923036 8929130 8926819 8926818 8929128 8928068 8926796 8929152 8926797 8927769 8927962 8927759 8929432 8927771 8927760 8927960 8911087 8932418 8926793 8911011 8932413 8929518 8929129 8892187 8932416 8929517 8932419 8932410 8927971 8929154 8926794 8932411 8926795 8928963 8922191 8923631 8916844 8911527 8910548 8905168 8916560 8929449 8929471 8929450 8928984 8927959 8929421 8929516 8932409 8933392 8933422 8933394 8929515 8926817 8932417 8935250 8932414 8933393 8927765 8933423 8908149 8909739 8916548 8929470 8926759 8935089 8922200 8933608 8935097 8935628 8935627 8935442 8935236 8908154 8853441 8929519 8935444 8932420 8933395 8935235 8935228 8935237 8935810 8935232 8935234 8935802 8935809 8927972 8926889 8935806 8935415 8933597 8916436 8935813 8938442 8938432 8938433 8938498 8935448 8938447 8938430 8935836 8935815 8935447 8935801 8935414 8935812 8935814 8935835 8935834 8935811 8935816 8938492 8939244 8938499 8940476 8940290 8938503 8935443 8939240 8938446 8935098 8933590 8935629 8935625 8940402 8941115 8941113 8941091 8941605 8938521 8941116 8940407 8935852 8941090 8941096 8938518 8940403 8938522 8940400 8935850 8941107 8940580 8941612 8941114 8938489 8939245 8938491 8939213 8941531 8941506 8941525 8940475 8935231 8940275 8940274 8941527 8941530 8941526 8938490 8941505 8935446 8940581 8941602 8941504 8940273 8940276 8941524 8939233 8938488 8940270 8940272 8939228 8916981 8935467 8933424 8942064 8942065 8939241 8923933 8940405 8944917 8944918 8941932 8941921 8941917 8941936 8944919 8935851 8941918 8941604 8944914 8940404 8940578 8940579 8941931 8941934 8941933 8944920 8938520 8947797 8947798 8941935 8946305 8946304 8938519 8941095 8946281 8944915 8944913 8946308 8944909 8946306 8941093 8941919 8949108 8948184 8940291 8946661 8926890 8946630 8948181 8949150 8946634 8949103 8949141 8948182 8948188 8948193 8949102 8948187 8951098 8949100 8951082 8948194 8948189 8948201 8951100 8948196 8951099 8947843 8949134 8948180 8948162 8947845 8948186 8949104 8949101 8952031 8948185 8948183 8938429 8952589 8952032 8946307 8935116 8952026 8951632 8951634 8952028 8952582 8952579 8952952 8941092 8952588 8952587 8951114 8947817 8952565 8952941 8949123 8949118 8952944 8941930 8952946 8952936 8954256 8953396 8952016 8952945 8953430 8952548 8952547 8952546 8952550 8954221 8954306 8953360 8951108 8935626 8952025 8952549 8953428 8954622 8954257 8954213 8951107 8949121 8954212 8952566 8949147 8953411 8957266 8957264 8957265 8954641 8952943 8954216 8952585 8952027 8953367 8954307 8954314 8954648 8954644 8954315 8956908 8957267 8956907 8954313 8954304 8948200 8954217 8953372 8954316 8957822 8957821 8957817 8941920 8957823 8953373 8952951 8953427 8957819 8956912 8953368 8953359 8951113 8957258 8957244 8947799 8933600 8938517 8957257 8957810 8957802 8954291 8959820 8954285 8959824 8957369 8957381 8957372 8959818 8959823 8956905 8954284 8954279 8959821 8953966 8959819 8953967 8957373 8953970 8954272 8957370 8959781 8959822 8953962 8953977 8953963 8941937 8959825 8954281 8957367 8954653 8959510 8958714 8957801 8958716 8952695 8947828 8954273 8953976 8947832 8947829 8949120 8960752 8957382 8960748 8958749 8951112 8953978 8953979 8957812 8959549 8958726 8958727 8947831 8960345 8959534 8958751 8959535 8958735 8954288 8958750 8959513 8953981 8954287 8953969 8957813 8961975 8958715 8947833 8960208 8960320 8961979 8960250 8959511 8961974 8959548 8960244 8959551 8964500 8964503 8964861 8960235 8952707 8957387 8954290 8964864 8954286 8958752 8964866 8960249 8953982 8960314 8960326 8960222 8960325 8960329 8965413 8953980 8959537 8960232 8960319 8960346 8960332 8959547 8957255 8959552 8960328 8960199 8960347 8959536 8964904 8954289 8964999 8965419 8964998 8965421 8964888 8960317 8964876 8965420 8966063 8964865 8964880 8966068 8965422 8964877 8964884 8960254 8964894 8964882 8960348 8964900 8966054 8964868 8966077 8964881 8964878 8966075 8968427 8966092 8960252 8965424 8966059 8966093 8965423 8965377 8966028 8964867 8960318 8966074 8966076 8966100 8966097 8966098 8957292 8957293 8953425 8960760 8953424 8960761 8960744 8957269 8964516 8968455 8966124 8968456 8964863 8960759 8964517 8964518 8960745 8966118 8965381 8957380 8970606 8966123 8966121 8960742 8965473 8968449 8959833 8959512 8959828 8959829 8964527 8964902 8961954 8960751 8957256 8961951 8954657 8959831 8964505 8960204 8969705 8970498 8968442 8971227 8968437 8971480 8968414 8964506 8969709 8971456 8971454 8971458 8959830 8952710 8959832 8971453 8971452 8970483 8970481 8971491 8970495 8971489 8968399 8971688 8970488 8971498 8971512 8971457 8971685 8971494 8970474 8971455 8970486 8970475 8971495 8971497 8966122 8965474 8965475 8971234 8971496 8968468 8971700 8965472 8971714 8959826 8971699 8971265 8971229 8968463 8971266 8966117 8968450 8971703 8970613 8971251 8971260 8971704 8976403 8975400 8971228 8970499 8975398 8975402 8975429 8976407 8971521 8971661 8971680 8975431 8971475 8971520 8968457 8977547 8969716 8975415 8975428 8971500 8975412 8975413 8977849 8971676 8977850 8975410 8970489 8975411 8971662 8977826 8977844 8977528 8977533 8977862 8977861 8975474 8977845 8971663 8968415 8971279 8966120 8977561 8977829 8975386 8976793 8976792 8971280 8977562 8971702 8977813 8971712 8977773 8976803 8976794 8978165 8976797 8976800 8978168 8978171 8971713 8978164 8977834 8975391 8971259 8978167 8978308 8977557 8977847 8978305 8977555 8978309 8977556 8978100 8978871 8978102 8978098 8979437 8979433 8978317 8977841 8978860 8978858 8978096 8977554 8979432 8978062 8971307 8978099 8984162 8978873 8971273 8978872 8984180 8971274 8978306 8978311 8979434 8971697 8971252 8978109 8971308 8971275 8971306 8984784 8984781 8984184 8984795 8978108 8985482 8977842 8979436 8984181 8985530 8978178 8977571 8985525 8978923 8984764 8984796 8985533 8978179 8978925 8985521 8979441 8985476 8985987 8985988 8985483 8977568 8985486 8978870 8984183 8978951 8985539 8985950 8977569 8984812 8977570 8984814 8984811 8978948 8984813 8978936 8985949 8986003 8988173 8985531 8985522 8984182 8985517 8984815 8985488 8984763 8988189 8988174 8988028 8977840 8985942 8978861 8985529 8988033 8985930 8986016 8988031 8988032 8985523 8985934 8988030 8985515 8985933 8986018 8986014 8988034 8985927 8984810 8985943 8986013 8985973 8987104 8985956 8985538 8987131 8991157 8991148 8985955 8987138 8988204 8987132 8988203 8987105 8985974 8991149 8987148 8991159 8987149 8991521 8987133 8988637 8988611 8988638 8987153 8987130 8987139 8987147 8987135 8988360 8988361 8985972 8987094 8987093 8987152 8987146 8987114 8987151 8985526 8988035 8988039 8988624 8987129 8987134 8988577 8988038 8987092 8988359 8987183 8988393 8992148 8988358 8992140 8992146 8995565 8995301 8987182 8992147 8988394 8941094 8978952 8987136 8985527 8992156 8992155 8996459 8995573 8996458 8996456 8996457 8996460 8996433 8996431 8996434 8996938 8997271 8995307 8997272 8996461 8995574 8997295 8996483 8996473 8997474 8997473 9001281 9001284 9001300 9001288 9000572 9001286 9001295 9001293 9001289 9001292 9001287 9001294 8997306 8997513 8996482 8996484 8996481 8998436 8996449 8996450 8988207 8995308 8995309 8995563 8995564 8995562 8995561 8995560 9002963 9002998 8996485 8998415 8998413 8998414 9002398 9002396 9002387 9002397 9002386 8996486 9002985 9003429 9002997 9003661 9003703 9003432 9007780 9007792 8988206 9003437 9003438 8998435 9003356 9007785 8996455 9003353 9007781 9007824 9003355 9003434 9007827 9003433 9007764 9008546 9007828 9008553 9008557 9008552 9003004 9008572 9008560 9007765 9002395 9002399 9008558 9008102 9003435 9009447 9009446 9009417 9009406 9009394 9009387 9003436 9007825 9007768 9009868 9008564 9008571 9009420 9007763 9009403 9008621 9011228 9009388 9011262 9007809 9002390 9011439 9009735 9009717 9009739 9011440 9008077 9009734 9008079 9011408 9008066 9009716 9008096 9009718 9002400 9008083 9008097 9008099 9008068 9008100 9008081 9008076 9002394 9009722 9008055 9008080 9008095 9011441 9009736 9009710 9009738 9011447 9008078 9011409 9011459 9011467 9011793 9011183 9011443 9008069 9013245 9011457 9011654 9011463 9011189 9011179 9013414 9011235 9013431 9009853 9011181 9011465 9013415 9013418 9013427 9011653 9013428 9009708 9013429 9011229 9013494 9013430 9011165 9011268 9011794 9009852 9011643 9003696 9013432 9011798 9008101 9011166 9008098 9011167 9011173 9013617 9013441 9009854 9013435 9013434 9011242 9013443 9013460 9011175 9013462 9013364 9011458 9013367 9013380 9013449 9011170 9013798 9013797 9013379 9013378 9013450 9013447 9011646 9009737 9013440 9008067 9011164 9013433 9011781 9013366 9013362 9013374 9013369 9013496 9013363 9017797 9018305 9017827 9018303 9009740 9009848 9018304 9017821 9013368 9017828 9013371 9013375 9018306 9017813 9013370 9017796 9022258 9011464 9011792 9011677 9009711 9011461 9011460 9008103 9013295 9011445 9013084 9018274 9013064 9013082 9011675 9018294 9013081 9009712 9013062 9013061 9008056 9018277 9013243 9018276 9011442 9013063 9013210 9013187 9013244 9013242 9018270 9018269 9008093 9013068 9018295 9011462 9013083 9013231 9011243 9018268 9013205 9013192 9013206 9013193 9013085 9011446 9013066 9013067 9013296 9011241 9013232 8995305 9013240 9013186 9018302 9013211 9008566 9013238 9013189 9022335 9013190 9018283 9018278 9018273 9018271 9018284 9013241 9013191 9013239 9013654 9023378 9018267 9018266 9018282 9017799 9018281 9023357 9025927 9023376 9023362 9003005 9025928 9023377 9018298 9023363 9018280 9018275 9018279 9013188 9024799 9024798 9024800 9026223 9026236 9025969 9025970 9025971 9025972 9026178 9023170 9017820 9023169 9023161 9024754 9023224 9023195 9023456 9026179 9024755 9024757 9030170 9029369 9025937 9023184 9017814 9017818 9029327 9023190 9030171 9023183 9025934 9024756 9030747 9029315 9030213 9026183 9030219 9030749 9030751 9030216 9030211 9030130 9030215 9030745 9026226 9026230 9026225 9026224 9026229 9030210 9030209 9030756 9032778 9032777 9030750 9030722 9030724 9030218 9032787 9032782 9030748 9032801 9030196 9030723 9033322 9032792 9032096 9032798 9032795 9032796 9026227 9026231 9030735 9026235 9030222 9031554 9030223 9032791 9026228 9030734 9032794 9033465 9033464 9030214 9026222 9033451 9033449 9030217 9030221 9033321 9032799 9033448 9026184 9033408 9032781 9032667 8938431 9029335 9029381 9029385 9029313 9025931 9029366 9033942 9032720 9032703 9030128 9032672 9032666 9032670 9032646 9037711 9032704 9030156 9031464 9033424 9032671 9029439 9036330 9037712 9033410 9036336 9033937 9037713 9037682 9032647 9033941 9033943 9030188 9036353 9033425 9040001 9033303 9037683 9037725 9036337 9040011 9040028 9040002 9040000 9037716 9033935 9040569 9040570 9040665 9040004 9040940 9040005 9040664 9040567 9040003 9036342 9039998 9036358 9040937 9040571 9040951 9040027 9040935 9036343 9031467 9037714 9037099 9034433 9030746 9034210 9034209 9033442 9016173 9034451 9033450 9040038 9034481 9033320 9034199 9034406 9034371 9034450 9034198 9034412 9041102 9032779 9034370 9034201 9041103 9044147 9034479 9034218 9034200 9033319 9034373 9033411 9040572 9033318 9033317 9033316 9034410 9040037 9040938 9034409 9033441 9033315 9040043 9040691 9033463 9033312 9033310 9041122 9041105 9034374 9034211 9034386 9034432 9034190 9032784 9034382 9038070 9032789 9037715 9040566 9033433 9034448 9034217 9034445 9034447 9044908 9040692 9040040 9040039 9034180 9033309 9033308 9034369 9040019 9034449 9031471 9032725 9034387 9034452 9041124 9044149 9043723 9043678 9044897 9034482 9034367 9033313 9034220 9043727 9034381 9034380 9040706 9041119 9045906 9041108 9041117 9041125 9041121 9034375 9041120 9043722 9044905 9034368 9041118 9043650 9045230 9043726 9041104 9034378 9041106 9041109 9034405 9034379 9034376 9034385 9034431 9029322 9044907 9041661 9041111 9040613 9032661 9033431 9037681 9032724 9043648 9044899 9043679 9045324 9047188 9047178 9045317 9045316 9040991 9041697 9045315 9033301 9044119 9033432 9044107 9040953 9045312 9046937 9047679 9047187 9047177 9040617 9047602 9048177 9048190 9047176 9044118 9040618 9040568 9038071 9047598 9047675 9045229 9040987 9046935 9047678 9051465 9048174 9048188 9047635 9046938 9046936 9047600 9047674 9047603 9047585 9051429 9054148 9053264 9051397 9054116 9053245 9051323 9054120 9045319 9053257 9034446 9052839 9052286 9034179 9052306 9047653 9052294 9053309 9047811 9052320 9053303 9045254 9047021 9040042 9047011 9047030 9052284 9045233 9045242 9044898 9045879 9047650 9045907 9052301 9055573 9047812 9041123 9052305 9055625 9055597 9054443 9045244 9053283 9052290 9045911 9031470 9054452 9054453 9052311 9045241 9044904 9045246 9043649 9055582 9047065 9055599 9055598 9045251 9047012 9054445 9053310 9055626 9047836 9055830 9047646 9055827 9055828 9047828 9047832 9047644 9054455 9055581 9043651 9052287 9052300 9052289 9045243 9052304 9054450 9052298 9052297 9047066 9047834 9047833 9052291 9043647 9047029 9055829 9047825 9047064 9047844 9055847 9045248 9047843 9034377 9047028 9047829 9047195 9053295 9053296 9052288 9044903 9044895 9045903 9047027 9047022 9047645 9045256 9047019 9047036 9045231 9045255 9047827 9045245 9052307 9056325 9045232 9056286 9055214 9055237 9055215 9047842 9040589 9040587 9055377 9054146 9056302 9055240 9055803 9055801 9052817 9039999 9051342 9056344 9055816 9056332 9058085 9061423 9055883 9052740 9055802 9055239 9051428 9047601 9055211 9041107 9055247 9055191 9055863 9058148 9054131 9053266 9059103 9055882 9056304 9054110 9055865 9063263 9051462 9052739 9054118 9063262 9054101 9055894 9054080 9055376 9054151 9055817 9055793 9055808 9054081 9055821 9058117 9056343 9055222 9055807 9055238 9053258 9054109 9064794 9063274 9064762 9055190 9054111 9064764 9051463 9054133 9052296 9047037 9053308 9047020 9064763 9059495 9052314 9052313 9052309 9052312 9064426 9059493 9052324 9052285 9061160 9061192 9063327 9044894 9064429 9070040 9064820 9064808 9064927 9059494 9059387 9052325 9059383 9052310 9061193 9059498 9063325 9059497 9059496 9045250 9061157 9064805 9052326 9061179 9064404 9064816 9064915 9064804 9064907 9064914 9056867 9061158 9064917 9061163 9056285 9070048 9047017 9063318 9064427 9064833 9059388 9055832 9055831 9064384 9047648 9040952 9064815 9046293 9064389 9055236 9058146 9063248 9064817 9070708 9055622 9071548 9040015 9070032 9070041 9064916 9064928 9064911 9070034 9063308 9056868 9064832 9054454 9063303 9052323 9064432 9063317 9070047 9070733 9070033 9070042 9070725 9055846 9064803 9059384 9063302 9070717 9071637 9070714 9071550 9061195 9070043 9064929 9070707 9070177 9071573 9059396 9071654 9059386 9070702 9064428 9070036 9064405 9064807 9064403 9052327 9072505 9073194 9071529 9071528 9070690 9070686 9071577 9073195 9072513 9070695 9071545 9064402 9045240 9055589 9071665 9071570 9072497 9065635 9071544 9069859 9072498 9071576 9071549 9071533 9065639 9070715 9072500 9071534 9037061 9073190 9070175 9056303 9073261 9061164 9063326 9056865 9070049 9072565 9059492 9070039 9065641 9070731 9034413 9061156 9071661 9071663 9058732 9055231 9077334 9033296 9077331 9071546 9073331 9078201 9072569 9077330 9078200 9078233 9073330 9073341 9063328 9072585 9072553 9040705 9073329 9073310 9072564 9071668 9070675 9045247 9056866 9070035 9073337 9078218 9064910 9078202 9072571 9034372 9044887 9078234 9073340 9071666 9052299 9070045 9072580 9071664 9071653 9073332 9073309 9077338 9070732 9078217 9073339 9070723 9072584 9077332 9072576 9072582 9072586 9073338 9070050 9073328 9061161 9072577 9077349 9070051 9072588 9056329 9070729 9071667 9063304 9070724 9070728 9064806 9070735 9064410 9072573 9063324 9070740 9078204 9072572 9072579 9072570 9072583 9073260 9077279 9078170 9078163 9073257 9072578 9077301 9044896 9070044 9073188 9077280 9073249 9077284 9077283 9072574 9072587 9070687 9059491 9078164 9071553 9077287 9080134 9078161 9073196 9073255 9077288 9071547 9055814 9064898 9078169 9080866 9073189 9077282 9080885 9078165 9078196 9080887 9073192 9084793 9080868 9080888 9080117 9084794 9073193 9077278 9084792 9084771 9085606 9078158 9080864 9085599 9085610 9071575 9080919 9080184 9080924 9078410 9080922 9080181 9080917 9073311 9080185 9080920 9080206 9080204 9080911 9080935 9080913 9080944 9080925 9080197 9085151 9080179 9080186 9080912 9080178 9078409 9080923 9080183 9080918 9077302 9089127 9080182 9080198 9080232 9088683 9080938 9080936 9080886 9085603 9078191 9071594 9078188 9078187 9078189 9085672 9085671 9088699 9088695 9080933 9089577 9085674 9045877 9080890 9089610 9088694 9085673 9089568 9088660 9088658 9089575 9088662 9078171 9080187 9089574 9064909 9089765 9089124 9089758 9090485 9089784 9089785 9085143 9089755 9085147 9090486 9090493 9085152 9090560 9090567 9090550 9089760 9090568 9089567 9090548 9080793 9085566 9090482 9090490 9090545 9095328 9080889 9090542 9089762 9089154 9034181 9051464 9090578 9047874 9073197 9080942 9089757 9089759 9089770 9085145 9089786 9090479 9095327 9089761 9089769 9090477 9095314 9095336 9095647 9095339 9090561 9090489 9095333 9095340 9089787 9095315 9090483 9090476 9090475 9088661 9090487 9090543 9078166 9080169 9090551 9089771 9090577 9090715 9089576 9090656 9090652 9090492 9090693 9090447 9090650 9090680 9088663 9095689 9095007 9088689 9097067 9090491 9095019 9095709 9090718 9095011 9097065 9090696 9088650 9090694 9095006 9090682 9095045 9097064 9095704 9090651 9095701 9097092 9095072 9097062 9090681 9095068 9095703 9095030 9095008 9098178 9095036 9098218 9090576 9095335 9090478 9090488 9095025 9098263 9098219 9097022 9097066 9098681 9098671 9097021 9098662 9097000 9097001 9098262 9098261 9100100 9098245 9098258 9098225 9090546 9095085 9090547 9052295 9095646 9095332 9098260 9088687 9090549 9095702 9098673 9097046 9097017 9097095 9095074 9098667 9097002 9080205 9098228 9098224 9090698 9095721 9095078 9098644 9100063 9090697 9100062 9100010 9100075 9101154 9097039 9101141 9100080 9098222 9101138 9089607 9098226 9098229 9100083 9101213 9098273 9104689 9101181 9100987 9104694 9098237 9101133 9100101 9100070 9100985 9098672 9100986 9100079 9095687 9101144 9101220 9101143 9097044 9097019 9100087 9098234 9095334 9098679 9101217 9101135 9100069 9098238 9101137 9101219 9098668 9104683 9101150 9100086 9101134 9102509 9104148 9104657 9104001 9104685 9104011 9098233 9104714 9105661 9104159 9104012 9104160 9104154 9104688 9104686 9104014 9110887 9111746 9104009 9100045 9100953 9106336 9111294 9100974 9111301 9111290 9101043 9110832 9111747 9100951 9106376 9098231 9111296 9100976 9111297 9101155 9098664 9100085 9106338 9101151 9101183 9110642 9097047 9112997 9095685 9100084 9100973 9112237 9104671 9101142 9112201 9112202 9110714 9105660 9104008 9112998 9104162 9100064 9101182 9110826 9110717 9110840 9105353 9106335 9110716 9110770 9105354 9110707 9102511 9111748 9090649 9110772 9110782 9110789 9115429 9110646 9110783 9119743 9121316 9110662 9121315 9110738 9101304 9110653 9115485 9119749 9110665 9095723 9122149 9100024 9104153 9119751 9119747 9110784 9095073 9119794 9119748 9119786 9119746 9123638 9119795 9110742 9110741 9119817 9121245 9121331 9119801 9119761 9110796 9121221 9121330 9115493 9110777 9095024 9117166 9117165 9117143 9117130 9117124 9117128 9115825 9117078 9117164 9116868 9098680 9111315 9124083 9117095 9117077 9115842 9114504 9114503 9124084 9124086 9111284 9117079 9098230 9097018 9104672 9100088 9117142 9114479 9113005 9113006 9114502 9097024 9098674 9097025 9106333 9112242 9101145 9096999 9111309 9101147 9098666 9112255 9101192 9113012 9113011 9112221 9114500 9112241 9106337 9113010 9102510 9111293 9112220 9101149 9111291 9101139 9117116 9115826 9117122 9104687 9102512 9101186 9098217 9101305 9100091 9124098 9123641 9112995 9119765 9102386 9119775 9111282 9128818 9128829 9128831 9106375 9125063 9125062 9124195 9128840 9124203 9128830 9124204 9125064 9101157 9119806 9128879 9130039 9128838 9128851 9128888 9128816 9119677 9119802 9119760 9124197 9119785 9128860 9119779 9112226 9125065 9128725 9128713 9101185 9119869 9117133 9117131 9117123 9128827 9125631 9119676 9119778 9125598 9125595 9129574 9128711 9125602 9128731 9129581 9125607 9125604 9125597 9111295 9120669 9125606 9125600 9125603 9125610 9120615 9106371 9104658 9129573 9120619 9125481 9111316 9114473 9101184 9117100 9111310 9125596 9120620 9114488 9100090 9104676 9101148 9114480 9112253 9106369 9106374 9111300 9117167 9117072 9104675 9095331 9116869 9112254 9114499 9116867 9095686 9113191 9104693 9125601 9097040 9120616 9125605 9116864 9117134 9117075 9120668 9129585 9117074 9116862 9101146 9097023 9100074 9101140 9106368 9117115 9125609 9125593 9124106 9128826 9129575 9098665 9128884 9119816 9124196 9130042 9125630 9125608 9125617 9129582 9115841 9125618 9128874 9125649 9111283 9128894 9128747 9128750 9128719 9128718 9124206 9128751 9128755 9128817 9112240 9116865 9117070 9138355 9138353 9130052 9137342 9117252 9115830 9138360 9138354 9128839 9119919 9137344 9128939 9138359 9128707 9128717 9128919 9138449 9138365 9128828 9132450 9132478 9137778 9125626 9132479 9137784 9137781 9134571 9134511 9137795 9134514 9134513 9134573 9137782 9134518 9125619 9125581 9125629 9129584 9125628 9088664 9101308 9095047 9101307 9130097 9100095 9097020 9117099 9134569 9115840 9125582 9101193 9120665 9116861 9134704 9134560 9125501 9125588 9125594 9125625 9125613 9132480 9132481 9125591 9134525 9117073 9125589 9120617 9125492 9125627 9125512 9125502 9134575 9134512 9125506 9137786 9134572 9134574 9125507 9125650 9134568 9125508 9137777 9125491 9125590 9137759 9137785 9134557 9134509 9137783 9134507 9134586 9117080 9134587 9098227 9101136 9134529 9125651 9113004 9134539 9134570 9134515 9134321 9134517 9134582 9134527 9137798 9137801 9137764 9134522 9120618 9137799 9137345 9134521 9134274 9134524 9134537 9134516 9142492 9137411 9141619 9134565 9125509 9142746 9137797 9125500 9142443 9125615 9128715 9137410 9137351 9137265 9141614 9137758 9137341 9137416 9138285 9143123 9137270 9137266 9137269 9142745 9134318 9137800 9137271 9134534 9134519 9134540 9134561 9134508 9134271 9134520 9114501 9134536 9117071 9117157 9134269 9047046 9045910 9134320 9146524 9132477 9047034 9117076 9146584 9146525 9146520 9125612 9134559 9125499 9106377 9134510 9101306 9101309 9146527 9134272 9146529 9134535 9146585 9134528 9134276 9148440 9146531 9137757 9134563 9134273 9134653 9134581 9146528 9148445 9134619 9134317 9134361 9141610 9146530 9142435 9148443 9134620 9148439 9134526 9146583 9134523 9148442 9148446 9134362 9134288 9147128 9137763 9146526 9146521 9134270 9146572 9146570 9134652 9137776 9149285 9137457 9147129 9149280 9149994 9134618 9137773 9146536 9149297 9148444 9134564 9146522 9146523 9149282 9149248 9134289 9149262 9142434 9146578 9150001 9137775 9141618 9128815 9146541 9149263 9149997 9137415 9146544 9138450 9142744 9142431 9137304 9143121 9142494 9142497 9149281 9149269 9149264 9150013 9149277 9150006 9149996 9149249 9137772 9150000 9150004 9150005 9149995 9142496 9141616 9149284 9142437 9137774 9137282 9146571 9142436 9149230 9149993 9148448 9149283 9150012 9148447 9149276 9146540 9146542 9146538 9150010 9137303 9138413 9138429 9128716 9117096 9142489 9149265 9148441 9149296 9134734 9137409 9150009 9134735 9150011 9149279 9151098 9151129 9151117 9149278 9149998 9151097 9151118 9149243 9151296 9149268 9156578 9154017 9149273 9151390 9151096 9149245 9146543 9151421 9150008 9151119 9156538 9146537 9151133 9151134 9151238 9151128 9156520 9151116 9156579 9151423 9156537 9149271 9156519 9156542 9149244 9151392 9154018 9151389 9156516 9151394 9157389 9156514 9156522 9149274 9157402 9157413 9149246 9157403 9141589 9134287 9157385 9143124 9156515 9149275 9151393 9156517 9149267 9151229 9146539 9156863 9158230 9158227 9158263 9159090 9157449 9156864 9158282 9157391 9157447 9158276 9158284 9158255 9157375 9151240 9157404 9156521 9158297 9157386 9157374 9156518 9157388 9157387 9134567 9157446 9151239 9137762 9151433 9134275 9151422 9149270 9151420 9151391 9149247 9151419 9090481 9120621 9156539 9159066 9151460 9156560 9157377 9156561 9157366 9160443 9157417 9157371 9160500 9160450 9160448 9159609 9159065 9160449 9160505 9157390 9160496 9159608 9160399 9158342 9158432 9158372 9163134 9160741 9158306 9160856 9160468 9160434 9163229 9160400 9160401 9160469 9159562 9159596 9160403 9159559 9163222 9158249 9160501 9163221 9163224 9159560 9159554 9158252 9159557 9163183 9165836 9163228 9163223 9167179 9163152 9158253 9160791 9160855 9160441 9158302 9151132 9158365 9165816 9151297 9163149 9149272 9167178 9163219 9159564 9163230 9157378 9159565 9160465 9156870 9159563 9163215 9159566 9166663 9166665 9166664 9160436 9166667 9166695 9166714 9164576 9164575 9166666 9164609 9164573 9165837 9166715 9164574 9164610 9164403 9165785 9167208 9167763 9165896 9165838 9167772 9165835 9167762 9167784 9165811 9165892 9165804 9165801 9167777 9169692 9159597 9163220 9164611 9169151 9166674 9165925 9167787 9165924 9169738 9169663 9171585 9169660 9167790 9169682 9167769 9165800 9169665 9171633 9166677 9166678 9170704 9164612 9169145 9164583 9169149 9164608 9170711 9159672 9164584 9169146 9164616 9164606 9164582 9169150 9172294 9169017 9164577 9170707 9169018 9164615 9166675 9164614 9164567 9159710 9169659 9175543 9175546 9175526 9175545 9165931 9175999 9171608 9167169 9165802 9171637 9175519 9166676 9170739 9164388 9164361 9172281 9171724 9175536 9177115 9172259 9170726 9175032 9170721 9170734 9170718 9175033 9175023 9170719 9177041 9174997 9170731 9170712 9175034 9170715 9170725 9170733 9164620 9170730 9170772 9170728 9176007 9169658 9175549 9176013 9176012 9175548 9176038 9175007 9171695 9175542 9176022 9176015 9175986 9172262 9175544 9178915 9176010 9180059 9178898 9178899 9178896 9175520 9178897 9171607 9176009 9175550 9175552 9179045 9179997 9170720 9178861 9171723 9177079 9179020 9179802 9177770 9177790 9177866 9174996 9172280 9179043 9178862 9177792 9179803 9177769 9178924 9179717 9177791 9177090 9179712 9179759 9179715 9177979 9179804 9177780 9177813 9177795 9177863 9179713 9176998 9179756 9177771 9177531 9179749 9177794 9177862 9177867 9177864 9177782 9176977 9177950 9179716 9177565 9177530 9177570 9177793 9177567 9177949 9179758 9179754 9177865 9177815 9177814 9177569 9177878 9164453 9177879 9177812 9179755 9179714 9177803 9185882 9177783 9159715 9186830 9176986 9170782 9170708 9177817 9176978 9177787 9176982 9177981 9185853 9183245 9177652 9180175 9176980 9183239 9179056 9180104 9186408 9187014 9179022 9183273 9179726 9180178 9183238 9183230 9180034 9178936 9187116 9188847 9189903 9188910 9185881 9164618 9185848 9179748 9183296 9185855 9185851 9185850 9188908 9186790 9177553 9187113 9159731 9187115 9178074 9178075 9187114 9185870 9185919 9185837 9187013 9192092 9192026 9189925 9187012 9188536 9180196 9189914 9183274 9177558 9192018 9183246 9192024 9180179 9177825 9188535 9187030 9185902 9192697 9189889 9189886 9187031 9185901 9194645 9186788 9185889 9194646 9169147 9189791 9185927 9177983 9192698 9185965 9179727 9189887 9189909 9185869 9188906 9185890 9186789 9185836 9177984 9179811 9188157 9170732 9166700 9190407 9190562 9189905 9189904 9195186 9194658 9186791 9185914 9190408 9185831 9185860 9194660 9189892 9190433 9189898 9194632 9189908 9177880 9190538 9190435 9185915 9194665 9190410 9177552 9159673 9185849 9177948 9190409 9194655 9185858 9186794 9194589 9194659 9185846 9194657 9187112 9188911 9185970 9190561 9189902 9190586 9177639 9194631 9194669 9196220 9188227 9194185 9194652 9194194 9192078 9190434 9177076 9194775 9195170 9195137 9194202 9195168 9194215 9199247 9195165 9188156 9195169 9190585 9196219 9194193 9195857 9199335 9199337 9195849 9194189 9199338 9192091 9194778 9172273 9185857 9185856 9172284 9189901 9195213 9188916 9188912 9186016 9196810 9190419 9199333 9196888 9199723 9190515 9201316 9200950 9198393 9201291 9196841 9195205 9194651 9201831 9195862 9201546 9199336 9195209 9196804 9194183 9194733 9197978 9194214 9201541 9195848 9200059 9195851 9194734 9183233 9194188 9200057 9196839 9194187 9190539 9186787 9201835 9201293 9201287 9190437 9202932 9201834 9201322 9194201 9196825 9196803 9199276 9186402 9195204 9196814 9201311 9196812 9196853 9196844 9201288 9205635 9201314 9201833 9194779 9201296 9201550 9195861 9199258 9201568 9201882 9201540 9195863 9192076 9201549 9205625 9200065 9201893 9195135 9208600 9204525 9207331 9201297 9207361 9204526 9201564 9208608 9204514 9208609 9200069 9207333 9201569 9201334 9208611 9207337 9207356 9207329 9207336 9196889 9201556 9196826 9201294 9205633 9202824 9206156 9208625 9205115 9200951 9196893 9202821 9196887 9208626 9190436 9201853 9201861 9202899 9201870 9201828 9196892 9201310 9202896 9206148 9202886 9206140 9202931 9202903 9208633 9206150 9206497 9206141 9196813 9208639 9196243 9205644 9190516 9205636 9208627 9206143 9208619 9190523 9199248 9206491 9202913 9202923 9206155 9208641 9194590 9205594 9190575 9202924 9208642 9212176 9205114 9208612 9208617 9205639 9208634 9202895 9208621 9212965 9212153 9208615 9208618 9212163 9212974 9212963 9212154 9213030 9212177 9212162 9202857 9207179 9196902 9205631 9192699 9212268 9212281 9212164 9213006 9212172 9212966 9213012 9212976 9212171 9212174 9212152 9206149 9199326 9202900 9212851 9208620 9212982 9213031 9212175 9201845 9208616 9213042 9206151 9202861 9208645 9206494 9208635 9192696 9212973 9213011 9213007 9204518 9190514 9212266 9213855 9213868 9213078 9212267 9213877 9213426 9213100 9213084 9213427 9213447 9213854 9213034 9212981 9213240 9202820 9212983 9212975 9212165 9204515 9213035 9212989 9212985 9212988 9213002 9212979 9212980 9213067 9214215 9186816 9212288 9212289 9213009 9213241 9213243 9213043 9212932 9208622 9215123 9212184 9212984 9212931 9212852 9212185 9215184 9213099 9215259 9213438 9213068 9215524 9213432 9213434 9213886 9213869 9215260 9223594 9215522 9207334 9212287 9223595 9213016 9208601 9212155 9204524 9223599 9223597 9201551 9197954 9223603 9212193 9213008 9208613 9212972 9215691 9212929 9224497 9224870 9224482 9212930 9213019 9213014 9224795 9215697 9224498 9224898 9226184 9215687 9226025 9215688 9212853 9226024 9202823 9215696 9224502 9224798 9224897 9224794 9215999 9226183 9213013 9215693 9224483 9224850 9226555 9226027 9212278 9224878 9224847 9224797 9224877 9215208 9212833 9213023 9224899 9224873 9215692 9224796 9227135 9224848 9227118 9215261 9199260 9215523 9227214 9213119 9227050 9226030 9215677 9227044 9206142 9213433 9227047 9212286 9224958 9223604 9215516 9215695 9213446 9224903 9213227 9227227 9224881 9227333 9213033 9212303 9233784 9227344 9227335 9224938 9225244 9225245 9233758 9225243 9213415 9227107 9227049 9225248 9227102 9235683 9227101 9235676 9235912 9227140 9224907 9235673 9227134 9235680 9227699 9205645 9199329 9227099 9224499 9227685 9235679 9195207 9195208 9201315 9199277 9199262 9227108 9227286 9236814 9199328 9227315 9205640 9227321 9236579 9227092 9227316 9227662 9236577 9227320 9236816 9224853 9236578 9236913 9227283 9227141 9236580 9236822 9236607 9236911 9227287 9227289 9227212 9236551 9199253 9227698 9213425 9233858 9233868 9233770 9233867 9233805 9233746 9233806 9233744 9233739 9233764 9233795 9233741 9233743 9233854 9237917 9233852 9233863 9233768 9233869 9233864 9233877 9233740 9213844 9237919 9235927 9227677 9233861 9233808 9235677 9236817 9195136 9233757 9227343 9225224 9236573 9236821 9224947 9235674 9235925 9233865 9237922 9227142 9233876 9237930 9237933 9237934 9237907 9237970 9237903 9237902 9236915 9236540 9233862 9237916 9202897 9213020 9237896 9241038 9237975 9236586 9237908 9237901 9237979 9233771 9208646 9236820 9241254 9202898 9241218 9237897 9241258 9241219 9238275 9241257 9237904 9241260 9238276 9241264 9241248 9241255 9237884 9237980 9236587 9241240 9241046 9236541 9237885 9241259 9241251 9241228 9238271 9241022 9227094 9241250 9227133 9237936 9237920 9241710 9241765 9241734 9237927 9224883 9227091 9241262 9192089 9241751 9213083 9241700 9237964 9241740 9241750 9241737 9241701 9233866 9241702 9241709 9241733 9242516 9242522 9243077 9242537 9243062 9241727 9241749 9243819
9205630 9243826 9241711 9241263 9233932 9227711 9215676 9237923 9233830 9238303 9238307 9236936 9238281 9236937 9238294 9238314 9242554 9238296 9233849 9238287 9241603 9242548 9242577 9238306 9238279 9238309 9243088 9242561 9238280 9227048 9242560 9213867 9238328 9241284 9215521 9243820 9243821 9244673 9247679 9213445 9244688 9244655 9247695 9247696 9247677 9247681 9244672 9241587 9244005 9247669 9243093 9244000 9241679 9241673 9233787 9243081 9243999 9248169 9241678 9242552 9243083 9248166 9244007 9243984 9243983 9247692 9243836 9247967 9248230 9243995 9248292 9243997 9248151 9248150 9197979 9248149 9241672 9248189 9243985 9248190 9248021 9250741 9250738 9250691 9250687 9250688 9250743 9250696 9250697 9250745 9250753 9250698 9250744 9250695 9250701 9250685 9250692 9248231 9250678 9250754 9250757 9250702 9250739 9250679 9250801 9250709 9248088 9250797 9248061 9250802 9250800 9160464 9250714 9251002 9251001 9253381 9251000 9253377 9253379 9253376 9253895 9247668 9253380 9253521 9253323 9250809 9253322 9253834 9253882 9253888 9253889 9253531 9254544 9253887 9253523 9253392 9254958 9253831 9253532 9253838 9250985 9253812 9253837 9255079 9254882 9254867 9254292 9254923 9254865 9254883 9255063 9254545 9254936 9253840 9253847 9253843 9254538 9255956 9255960 9255958 9255954 9253844 9253891 9253894 9253520 9255957 9253530 9253902 9255952 9255967 9255955 9255969 9255970 9255599 9254291 9254866 9257924 9254864 9257921 9255600 9243986 9254966 9258069 9258072 9258075 9256054 9256052 9258212 9258231 9258232 9258071 9258070 9258214 9258959 9258230 9258213 9258956 9258958 9255601 9253309 9201548 9258929 9254270 9258946 9258943 9257925 9257931 9259030 9253312 9257918 9259029 9259038 9259040 9259039 9260611 9259032 9260243 9260612 9259049 9206490 9260244 9260629 9206489 9260631 9257919 9259031 9260713 9260714 9258955 9260137 9260229 9259120 9258957 9260149 9259116 9260897 9260145 9260896 9198395 9260147 9260153 9196894 9202856 9205593 9196840 9201323 9196899 9263639 9263638 9264053 9259048 9260913 9263941 9260725 9260905 9263939 9260720 9260904 9259110 9259133 9264054 9259107 9264705 9205131 9264166 9264167 9260630 9259130 9253313 9201335 9257920 9260150 9258964 9260732 9260920 9264156 9263940 9263950 9263946 9264159 9264699 9265179 9260919 9260721 9265180 9264155 9259119 9260744 9260934 9260739 9260695 9260907 9264157 9264158 9264169 9260933 9260921 9263942 9263944 9260922 9264170 9260916 9264174 9264171 9264698 9265971 9265981 9264706 9265973 9266361 9260902 9260915 9266347 9265943 9263974 9265937 9263949 9263945 9255609 9260148 9255633 9265911 9264172 9263948 9265972 9265979 9264168 9265944 9266681 9266687 9266348 9265938 9264181 9265978 9260918 9266688 9265982 9266686 9266689 9265182 9264702 9267451 9266961 9267452 9267004 9266962 9266684 9268442 9268463 9269848 9268441 9268444 9269847 9269842 9270248 9269850 9268445 9269852 9269843 9268461 9268462 9269846 9269845 9269844 9268443 9270247 9268464 9269849 9269854 9270246 9270436 9270435 9270389 9270438 9270439 9270443 9269853 9267453 9267434 9266974 9271531 9266945 9267003 9270241 9271606 9271600 9266975 9269855 9271478 9267435 9267005 9267438 9267002 9270249 9269851 9271512 9270448 9272324 9271748 9270441 9270388 9272329 9271741 9272331 9270440 9271528 9270449 9270446 9271747 9268454 9271744 9270445 9270432 9271760 9270452 9271530 9271517 9271746 9272276 9271485 9270447 9270444 9271738 9272269 9272264 9271745 9271763 9271739 9266944 9272163 9267440 9272165 9267439 9272571 9273188 9270437 9273696 9273020 9271514 9272277 9273028 9269102 9270434 9272578 9273019 9271511 9272312 9271486 9272296 9273027 9271764 9268449 9271761 9272294 9273699 9273691 9273690 9273697 9273692 9272299 9270442 9273213 9273214 9269076 9273988 9272576 9275073 9275072 9273176 9273191 9273993 9276294 9273029 9272573 9273174 9273178 9273175 9275101 9273177 9273189 9273992 9275093 9273708 9273030 9273716 9273199 9273964 9272293 9276291 9273961 9276879 9273194 9273237 9276292 9273963 9276880 9274004 9273279 9273994 9276877 9274348 9276878 9273278 9274003 9273236 9273238 9273996 9276355 9274006 9274361 9275136 9275132 9273709 9276876 9276343 9273700 9275128 9271757 9273701 9272292 9277431 9277404 9277076 9275138 9277446 9276346 9273229 9273031 9276293 9273704 9277421 9276357 9277476 9275133 9277423 9275137 9273722 9277410 9277477 9276364 9277424 9277478 9273960 9276301 9274005 9277420 9278419 9275071 9276356 9273962 9280245 9273968 9278425 9278418 9278417 9278424 9273978 9273977 9280908 9277058 9278126 9277419 9273975 9272295 9273980 9282290 9280909 9273974 9276300 9276903 9280226 9276897 9276902 9276881 9278420 9276299 9278125 9278358 9278427 9279960 9273972 9280659 9276894 9276901 9280246 9276904 9280866 9276898 9276298 9277658 9280663 9278363 9276900 9278357 9280662 9278372 9273981 9275150 9278370 9280146 9279258 9276933 9276895 9276347 9276363 9277259 9277422 9277265 9283373 9280145 9279278 9280153 9279259 9276928 9283350 9277469 9283357 9277659 9280147 9278371 9283014 9283015 9279261 9276896 9278431 9278369 9274011 9278367 9287128 9278373 9283337 9287149 9278432 9287150 9283329 9287130 9278435 9269864 9287132 9201572 9278365 9278364 9279953 9287152 9283336 9283800 9283816 9273982 9283334 9287148 9280664 9287131 9283339 9283795 9283815 9283818 9274012 9287783 9287771 9283798 9278433 9283359 9287789 9283338 9287169 9277409 9287782 9287188 9287129 9283360 9287205 9283361 9287204 9283351 9287787 9283362 9287109 9288614 9288566 9289281 9287133 9276932 9278450 9287836 9287833 9287832 9287117 9287785 9276930 9287780 9289863 9289865 9289905 9287119 9287151 9288623 9287837 9278377 9278376 9287153 9289876 9290012 9288624 9287786 9289866 9289874 9287846 9287110 9287845 9287154 9288619 9287840 9287868 9287866 9283330 9287116 9289872 9287869 9278434 9289873 9287865 9290362 9287125 9287849 9283323 9290361 9289875 9282280 9289904 9287850 9280661 9290364 9288305 9288323 9205130 9290363 9287876 9287779 9287874 9279952 9289877 9283834 9273228 9289916 9283817 9283364 9283838 9288722 9288724 9287252 9288375 9287905 9287904 9287946 9288728 9283833 9288723 9287250 9283835 9287854 9294888 9287856 9287878 9293962 9294891 9292970 9283837 9294897 9287873 9294305 9295096 9287167 9294306 9283322 9290105 9292977 9294900 9287855 9294889 9295086 9295090 9288332 9295154 9295156 9295089 9295631 9295633 9276899 9295628 9287233 9295155 9295630 9295088 9294073 9296191 9287948 9295359 9294031 9293973 9295380 9294087 9296200 9287232 9289906 9294052 9287912 9296201 9287918 9287242 9287915 9293963 9295372 9288725 9296807 9296207 9296212 9295369 9296824 9293964 9287916 9296203 9294016 9283363 9287917 9296208 9295373 9296809 9294015 9296817 9295101 9296206 9296189 9287189 9276929 9296831 9301163 9287914 9299329 9287913 9299328 9296163 9287794 9295613 9295611 9295610 9296133 9295614 9296134 9296129 9296135 9296131 9295657 9301176 9301177 9301376 9301375 9301823 9295627 9301374 9297634 9300108 9296168 9295660 9295091 9296176 9297721 9301175 9296171 9303375 9296187 9297713 9303374 9297632 9288331 9303405 9297635 9302706 9299410 9299418 9299409 9299437 9301160 9301134 9301166 9299341 9299375 9301140 9299333 9299343 9301135 9303918 9303376 9295157 9300128 9297615 9303671 9296149 9297608 9297617 9301822 9303404 9297610 9303382 9297724 9303673 9295612 9296177 9297611 9297630 9297609 9300109 9297680 9301178 9301820 9297633 9297614 9297612 9303894 9295658 9303893 9303903 9297719 9297706 9297631 9297722 9297723 9297734 9297623 9303711 9303723 9304552 9296153 9303722 9297684 9303904 9303905 9297642 9297641 9297616 9297639 9296169 9304805 9296178 9303719 9304549 9297640 9302704 9303674 9304565 9303672 9296167 9296162 9301169 9301139 9299346 9299367 9304814 9304818 9304809 9299373 9299423 9294034 9194191 9302703 9294059 9297683 9304797 9297720 9300115 9307560 9301824 9307559 9307562 9297652 9307561 9307835 9295100 9308998 9309192 9308045 9308044 9308380 9308047 9307971 9308043 9310194 9309219 9309195 9308048 9307445 9307450 9309166 9307436 9309014 9307963 9309193 9299401 9310190 9308397 9310195 9310703 9310187 9309010 9310579 9307965 9309013 9307448 9307806 9307808 9307809 9307558 9307446 9310529 9309165 9307449 9307444 9309160 9307804 9308037 9310539 9309167 9307803 9310528 9309175 9308997 9310537 9307810 9310541 9310545 9310547 9310543 9310538 9310542 9307452 9309163 9307807 9310540 9308381 9309018 9310954 9310948 9307798 9313315 9309221 9313286 9313289 9313366 9308383 9309176 9310544 9313935 9313940 9309194 9313936 9313938 9303403 9313937 9313290 9313988 9313953 9313959 9313922 9313939 9313989 9310546 9313984 9313991 9304551 9313941 9313983 9314083 9314005 9314004 9314084 9314006 9313975 9314171 9313288 9313965 9314101 9314082 9314119 9314136 9314344 9314125 9314121 9314120 9314189 9299428 9314114 9314215 9314050 9313356 9314052 9313999 9314115 9316778 9314086 9314008 9314000 9314001 9316819 9313996 9315697 9314018 9314003 9314265 9313998 9314002 9314021 9314020 9314266 9314131 9317082 9314126 9314195 9317759 9316817 9317777 9313997 9314071 9315709 9313995 9314026 9317740 9317032 9314017 9317046 9315692 9317738 9314010 9316821 9314085 9314070 9314009 9314051 9317023 9317732 9317737 9318399 9317027 9302702 9314011 9317031 9314025 9317026 9316818 9321284 9318371 9321173 9314012 9319737 9319758 9321175 9314127 9321174 9314132 9314201 9317764 9317763 9317088 9319756 9318370 9318406 9322424 9322422 9322429 9314032 9321281 9317736 9321278 9318397 9316844 9321277 9322432 9323093 9314110 9314112 9319660 9319757 9319637 9319638 9323594 9323609 9318378 9316847 9323118 9323103 9323600 9319639 9318386 9316848 9323817 9318385 9314033 9323550 9322460 9323534 9323542 9323533 9323538 9318422 9323537 9323535 9314031 9318389 9322431 9318380 9318420 9318404 9318357 9323699 9318381 9317735 9318419 9322443 9323557 9323541 9318388 9322441 9323525 9323556 9323809 9318383 9318376 9318356 9323811 9323543 9323555 9323521 9322433 9323707 9318384 9323706 9321276 9318398 9318408 9322444 9323810 9322430 9318424 9318351 9323818 9323102 9323839 9323738 9323526 9318423 9316820 9323854 9314034 9326673 9323114 9323746 9326667 9323840 9323828 9323835 9323833 9323524 9329214 9326465 9329167 9323850 9314087 9329168 9323834 9323737 9323832 9323618 9323795 9326464 9326678 9329966 9323117 9329213 9330503 9330512 9323808 9330547 9329271 9329283 9330546 9330539 9329221 9331018 9330498 9331061 9331050 9330555 9329282 9330565 9329173 9329226 9329270 9329281 9330543 9331019 9330541 9330561 9329181 9330558 9331017 9329172 9329224 9330537 9329175 9330538 9330548 9329272 9331038 9330540 9330544 9331033 9331032 9331932 9331023 9331035 9331036 9329874 9331078 9323710 9331037 9331081 9331077 9331919 9331034 9331055 9323709 9330542 9331915 9335066 9335067 9323708 9331914 9335065 9331908 9318382 9330535 9330560 9331920 9330513 9335634 9318387 9335104 9335632 9335660 9335654 9329955 9322481 9335638 9331079 9335729 9336144 9335737 9335728 9336142 9335731 9336139 9335743 9336136 9335735 9335738 9337847 9337895 9336135 9336145 9335733 9337902 9336141 9335739 9336137 9335694 9335051 9337897 9335079 9335740 9337904 9338686 9330507 9336062 9338073 9336066 9337898 9337896 9337899 9337906 9337822 9337903 9337886 9337907 9337858 9337862 9337860 9337861 9335704 9337882 9337857 9337859 9337863 9337901 9339154 9338623 9339156 9338626 9338800 9335695 9338773 9335696 9338803 9338781 9337824 9335736 9338801 9336134 9337900 9337823 9338782 9337883 9338772 9337825 9335727 9339170 9337884 9339153 9337887 9337888 9338805 9335734 9338806 9338804 9339155 9337817 9337920 9339404 9337929 9335726 9338792 9338793 9338807 9335732 9336061 9323604 9322442 9335747 9335166 9338802 9201292 9343777 9340013 9341359 9341487 9341438 9342992 9340012 9339955 9339936 9342971 9341522 9341447 9341534 9340376 9345815 9340369 9340381 9340342 9340387 9340377 9340395 9342322 9338839 9344607 9340405 9340341 9343768 9340357 9343774 9340399 9340326 9340396 9340379 9343775 9340384 9340327 9340283 9343765 9340378 9340372 9340375 9340359 9340382 9343767 9340350 9342317 9340365 9345828 9343770 9340404 9342321 9344593 9340285 9342320 9344618 9342993 9340340 9345850 9340358 9342996 9341374 9346382 9340398 9342983 9345849 9341523 9346374 9341334 9341489 9340397 9340338 9345760 9345825 9340356 9345826 9343756 9340394 9340374 9342327 9342298 9339405 9344595 9340349 9339406 9340406 9347089 9344598 9344589 9344594 9345759 9341488 9342326 9350951 9347114 9346384 9347115 9345818 9340351 9343755 9340447 9346434 9340407 9346439 9345831 9346431 9345827 9343769 9344599 9340455 9342297 9342323 9340388 9346432 9346433 9341375 9341341 9350905 9352974 9341343 9341344 9352975 9347081 9350904 9350929 9350910 9352994 9341346 9341362 9350930 9341333 9350931 9351546 9350941 9341496 9341376 9351619 9351622 9346662 9346417 9346532 9340354 9346416 9351414 9346518 9344592 9351407 9340366 9351332 9351410 9346446 9352552 9351428 9343185 9351333 9351433 9344596 9346440 9351412 9351331 9346663 9343773 9346509 9351335 9346661 9340353 9345830 9352593 9346445 9351429 9351339 9346447 9351338 9351336 9351411 9352592 9346448 9351330 9344601 9351337 9352554 9351389 9352555 9351413 9351430 9346507 9352595 9351531 9351396 9351340 9351427 9346501 9351390 9351400 9351432 9351334 9352594 9352950 9351630 9341448 9346521 9351628 9351629 9351550 9355077 9353000 9352951 9355233 9355090 9351399 9354489 9355086 9352976 9352953 9355252 9355253 9355239 9355084 9355264 9355241 9338842 9354435 9355240 9355087 9355097 9355078 9355332 9351251 9351252 9355297 9352596 9351397 9355331 9351243 9351401 9351244 9355300 9355322 9351393 9355325 9355334 9355323 9355335 9355294 9355284 9351388 9355330 9355298 9355321 9355295 9355291 9355299 9355293 9351392 9355292 9355080 9355333 9359290 9358802 9358918 9358919 9359294 9359311 9351398 9359316 9358804 9359315 9359313 9359314 9341499 9340339 9360246 9355303 9355272 9358871 9360587 9358872 9358874 9358891 9360725 9360726 9358870 9358869 9355304 9358922 9358873 9358936 9355305 9360728 9355306 9355324 9358935 9358929 9358890 9360608 9360582 9361497 9361508 9360599 9361509 9360598 9360609 9360616 9361504 9362279 9361503 9361505 9361506 9360681 9360658 9360729 9360680 9360692 9360700 9361490 9362266 9362400 9361514 9362466 9352952 9362295 9362407 9360691 9362290 9362463 9362278 9360684 9361517 9361513 9361507 9360682 9362277 9360601 9361518 9362429 9362426 9360740 9358921 9365866 9362882 9362875 9365872 9365865 9362880 9365870 9362406 9362419 9362420 9362890 9362881 9362417 9362292 9362883 9362891 9362452 9362468 9362500 9362772 9362482 9362481 9362497 9362498 9362451 9362776 9362774 9362791 9362775 9362450 9362792 9362779 9362467 9362795 9362469 9362765 9362847 9362484 9362849 9365826 9362848 9365856 9362804 9365794 9362797 9362796 9362763 9364570 9367606 9365795 9364598 9364507 9362811 9362855 9364594 9363275 9362850 9362805 9362856 9364601 9364506 9364586 9364513 9362773 9364476 9365799 9363273 9365812 9364478 9364479 9364614 9364599 9369547 9365829 9365802 9369469 9362810 9367662 9365828 9367650 9362483 9363277 9369472 9365801 9365830 9364612 9369473 9363274 9365545 9365778 9364595 9364602 9364611 9370065 9369453 9370079 9370066 9370064 9362889 9364613 9367660 9364753 9367624 9369475 9369470 9369483 9369521 9369478 9369481 9369484 9362851 9364518 9369477 9364517 9363276 9363278 9370067 9370078 9365858 9370080 9370068 9362499 9364516 9362416 9365546 9373256 9367632 9369494 9360683 9370086 9362418 9362778 9367625 9369543 9369542 9369544 9369493 9369476 9370811 9370809 9370041 9373252 9373254 9355301 9373255 9373250 9370019 9373379 9373416 9370826 9370825 9373249 9369540 9370520 9370542 9370485 9369474 9370420 9370516 9369529 9369482 9367626 9370514 9370543 9373397 9373384 9369454 9373383 9373373 9370824 9373382 9374226 9374229 9374790 9374205 9374224 9375259 9374764 9375260 9370497 9374772 9374771 9374864 9374865 9374853 9370513 9370517 9374768 9370519 9374867 9367631 9374849 9370512 9374661 9370510 9370518 9375389 9375435 9375410 9375436 9376365 9375379 9375380 9375371 9375414 9375372 9376361 9376360 9375434 9375378 9375370 9376376 9375374 9374225 9376344 9376400 9376375 9376357 9376343 9375446 9375397 9376349 9380448 9375574 9380450 9376362 9380447 9380411 9375428 9376350 9380407 9375712 9380876 9375638 9375588 9375701 9374859 9374852 9375635 9375587 9375702 9375704 9374879 9375711 9375585 9375715 9374890 9374886 9380802 9375717 9375700 9374881 9380885 9375713 9380816 9380437 9380421 9380427 9380480 9381704 9381727 9380483 9380787 9382099 9380429 9381718 9380780 9380786 9375719 9375720 9381710 9375722 9374880 9375716 9375705 9375751 9369468 9367630 9382646 9382645 9375723 9374827 9381953 9375675 9381987 9383975 9374798 9380819 9374866 9382108 9382307 9382001 9388898 9388863 9381989 9382308 9382571 9382005 9381991 9375682 9381980 9382630 9380801 9380793 9375679 9382629 9381979 9381935 9367629 9375714 9374885 9388982 9375680 9381959 9375681 9374826 9382724 9388981 9388963 9384264 9374889 9388971 9375724 9388962 9381952 9375676 9389001 9388979 9389012 9389010 9389011 9374891 9388993 9382012 9388987 9389000 9388975 9389009 9388983 9380426 9388969 9388978 9375640 9380860 9382727 9382725 9384334 9384282 9384286 9381967 9384247 9382726 9380862 9374888 9381972 9380857 9384253 9381937 9384308 9384281 9382703 9384303 9389054 9392113 9392133 9393324 9393439 9384335 9392082 9380861 9375639 9381968 9375721 9384252 9389062 9393449 9393854 9388973 9382036 9393471 9389064 9380875 9393863 9384304 9393901 9393910 9384296 9394085 9384310 9393900 9384267 9394220 9382702 9394138 9381960 9393918 9384332 9384248 9392141 9392124 9394084 9394083 9394195 9393911 9394162 9394089 9381981 9393904 9394164 9393922 9394180 9393905 9384322 9394231 9382699 9394228 9394087 9393906 9394192 9393913 9380856 9394198 9393912 9384317 9384268 9394194 9384313 9384311 9382687 9394165 9394211 9382689 9384345 9382686 9394157 9384330 9394226 9394139 9384292 9384323 9394171 9394229 9384297 9384249 9381936 9394142 9394152 9384343 9384318 9384309 9394090 9375678 9394172 9394227 9384254 9397739 9394140 9400541 9394149 9375677 9397738 9394187 9394217 9394145 9395065 9384331 9399056 9384293 9382714 9397736 9400547 9400545 9400568 9400540 9394144 9399065 9400539 9400548 9399055 9397737 9394150 9394151 9394147 9399741 9396805 9394584 9394578 9396873 9394576 9393838 9394178 9396865 9400532 9396876 9394575 9402591 9398794 9398796 9396780 9396904 9399060 9398793 9394230 9398797 9394197 9395066 9394146 9384319 9384312 9382729 9396818 9396782 9396783 9392123 9402767 9402766 9402684 9402634 9396866 9402665 9402761 9394207 9395097 9395134 9384344 9395088 9395067 9400550 9395026 9395087 9400531 9395072 9395055 9395038 9395024 9395122 9402208 9395105 9395028 9395096 9399064 9399059 9395073 9402210 9400549 9402209 9402194 9395098 9395061 9402191 9395133 9402832 9395079 9395070 9395071 9394219 9400577 9395123 9395109 9395108 9395062 9395121 9400576 9402190 9402814 9402836 9395027 9403506 9403495 9402819 9403523 9403676 9395136 9403501 9408465 9384287 9403498 9403502 9402189 9395104 9402821 9400560 9408466 9403512 9403665 9402833 9395080 9395025 9403499 9394218 9408463 9403497 9408446 9403671 9402837 9402820 9408432 9402838 9402810 9402834 9408436 9403505 9402835 9409849 9402822 9408419 9403677 9408430 9408195 9403646 9410277 9402892 9402908 9409820 9408464 9409853 9402893 9408650 9408639 9403647 9403650 9402897 9409864 9408643 9396883 9408647 9409854 9410279 9403661 9408194 9402896 9403683 9410281 9403651 9409870 9411158 9409852 9410524 9411070 9408637 9408447 9410496 9402894 9411069 9411162 9411735 9410507 9410523 9410535 9410546 9396794 9410545 9403504 9400530 9410553 9411496 9411479 9410519 9411477 9402813 9408433 9411512 9410527 9410542 9411519 9411495 9410569 9410568 9411532 9410541 9410543 9410538 9411478 9411488 9410544 9411480 9410552 9409829 9403511 9411533 9409806 9410525 9411530 9409821 9411524 9411531 9403500 9395137 9402839 9412394 9412380 9408420 9410556 9412556 9412733 9396915 9408644 9403645 9409879 9411736 9402899 9412741 9412555 9409860 9412737 9412738 9415021 9411764 9415338 9415323 9412749 9415325 9412561 9412564 9415715 9396789 9415703 9411738 9415714 9415702 9415832 9412559 9410555 9411768 9411759 9412560 9412415 9412730 9411739 9409876 9416185 9415034 9415718 9416186 9415826 9415712 9415043 9415840 9412393 9411476 9415660 9415676 9415042 9415054 9415830 9416051 9412399 9415711 9417647 9416363 9416049 9417864 9416348 9417866 9417643 9416197 9409878 9417645 9417863 9417668 9417884 9417685 9416333 9417865 9411513 9396791 9412563 9412420 9409871 9417659 9415663 9415835 9409875 9418443 9416075 9412404 9416374 9412401 9417596 9415067 9412400 9417580 9416731 9417539 9417542 9415679 9417543 9416038 9417566 9418654 9409830 9416375 9416085 9416069 9417497 9412406 9415680 9415083 9417494 9415659 9415055 9418645 9415044 9417790 9412410 9417791 9421961 9417538 9422028 9418644 9415082 9422372 9421969 9417572 9417557 9421954 9416084 9416370 9412407 9422029 9417567 9417569 9423399 9423408 9417568 9422362 9421980 9416050 9421956 9416055 9421962 9415677 9416072 9416066 9415675 9421992 9416191 9422027 9415709 9423411 9423339 9423325 9423304 9423308 9421979 9423309 9423305 9423367 9424323 9424301 9426282 9426026 9426024 9423330 9423366 9424296 9423347 9423335 9422396 9423352 9418607 9424283 9410566 9418592 9426284 9418602 9415726 9422026 9422401 9424290 9422407 9423307 9423346 9425972 9426277 9422402 9423306 9425955 9425963 9421993 9425933 9424255 9425978 9426940 9425982 9425957 9425981 9425959 9426942 9425930 9424257 9423413 9424220 9425958 9415673 9426948 9425973 9424219 9426945 9426944 9425940 9425980 9426946 9412395 9425962 9426994 9424259 9426939 9426959 9423382 9426967 9426943 9426995 9426958 9410518 9425976 9425977 9400574 9424253 9429985 9429992 9415658 9430013 9426956 9430012 9395106 9423409 9415046 9415047 9430854 9424258 9417789 9423317 9418606 9423318 9424240 9423315 9423301 9426954 9430037 9430035 9424243 9430036 9429997 9429996 9429991 9429990 9430862 9429844 9429989 9426419 9430236 9430006 9426501 9430243 9426432 9429742 9430195 9429762 9432614 9429743 9432619 9430798 9430190 9430237 9424311 9429841 9430229 9430192 9429870 9429845 9429741 9429840 9426418 9430268 9429758 9430787 9426430 9426427 9433121 9429866 9426420 9432612 9433122 9429757 9431627 9429867 9431650 9433375 9426480 9423348 9430797 9433372 9431658 9433115 9431628 9433051 9430182 9433403 9433398 9433397 9433108 9431660 9432630 9433401 9431635 9437568 9437582 9424286 9426426 9430351 9432632 9437788 9433954 9433404 9429917 9437791 9433400 9437793 9433107 9433395 9433405 9430828 9431657 9433399 9432629 9433396 9433952 9437581 9430861 9424284 9437567 9432816 9437786 9438849 9437796 9433099 9438016 9430860 9426509 9408441 9426510 9432817 9422395 9429906 9431661 9437809 9433083 9433087 9438087 9438025 9438852 9438854 9438851 9437797 9437810 9438038 9437585 9432828 9438856 9395037 9431643 9438060 9433098 9437798 9437806 9433109 9438844 9438021 9437593 9438104 9431644 9426947 9437586 9437785 9438641 9438853 9440305 9437787 9438082 9437792 9432825 9438081 9438102 9440290 9438103 9437594 9433408 9432815 9438638 9440351 9429907 9430373 9429911 9440968 9432631 9440289 9440768 9440650 9438453 9438472 9440649 9445397 9440794 9440644 9439743 9438455 9442081 9438454 9440784 9440772 9442073 9438456 9440304 9432810 9432656 9442077 9438449 9438439 9438482 9440963 9438448 9438483 9438484 9438475 9438477 9438431 9438452 9444138 9440972 9440793 9440983 9440971 9440967 9445447 9444215 9442080 9438466 9437794 9432824 9438105 9438639 9442079 9440978 9438521 9433110 9438399 9444218 9202822 9431663 9432831 9437811 9448659 9448622 9448652 9448790 9449760 9454675 9450428 9453289 9453290 9453294 9450417 9449761 9449762 9453299 9449759 9453297 9449763 9450422 9453298 9449751 9453292 9452470 9452473 9452456 9454164 9452484 9452491 9452471 9452468 9452500 9452474 9454177 9449105 9449114 9455124 9454152 9454181 9452508 9452501 9454120 9452481 9455349 9450457 9448791 9438101 9446120 9454172 9448822 9450011 9449112 9448818 9454147 9452502 9452489 9454180 9454179 9452532 9454173 9452513 9448836 9452503 9456560 9444288 9449103 9449143 9448816 9449115 9449785 9448819 9449786 9450015 9449113 9449139 9449108 9449131 9456867 9456873 9456886 9449123 9456870 9456877 9456866 9456881 9449111 9456878 9456876 9456874 9449118 9456556 9456588 9449159 9456593 9446209 9450423 9456845 9455350 9453303 9450425 9453302 9438648 9446222 9456539 9441884 9456841 9456959 9446217 9456961 9456958 9456880 9452461 9450500 9453300 9455342 9455361 9453293 9450418 9456543 9461606 9457980 9455351 9455262 9461605 9455261 9456960 9455360 9450424 9455359 9461636 9456815 9461604 9457984 9457987 9457979 9456810 9456824 9461765 9448792 9457976 9461645 9461651 9453301 9461642 9429999 9461616 9458010 9461775 9462038 9462015 9457727 9458005 9462016 9461674 9456869 9461682 9461683 9458002 9457735 9461685 9461681 9462063 9456586 9456865 9452533 9462011 9461678 9461679 9455355 9456890 9462323 9461648 9461627 9464923 9463129 9462991 9462301 9448658 9440742 9445622 9441882 9445600 9449755 9445602 9449781 9456875 9462021 9456888 9456571 9456887 9456862 9445605 9462009 9456856 9449094 9449756 9456562 9462054 9449132 9446097 9462988 9462051 9464910 9462006 9458007 9462049 9457733 9464911 9465551 9462099 9452514 9465537 9461769 9465560 9430001 9457729 9456871 9465550 9462132 9465437 9465621 9450010 9462241 9462071 9465623 9462012 9456587 9462224 9456589 9462031 9457730 9461700 9450016 9462327 9456569 9433069 9452507 9448623 9464946 9468881 9468883 9464974 9470110 9468890 9462042 9438083 9469033 9440350 9438644 9450456 9453291 9448625 9462987 9445928 9465505 9470114 9464929 9438033 9470103 9465479 9470097 9465516 9465608 9468885 9470115 9465609 9465509 9464976 9465508 9468891 9470108 9468889 9465778 9469942 9461675 9458001 9461680 9462944 9457734 9468917 9461758 9461756 9471223 9461759 9462068 9456858 9469027 9461703 9471380 9469947 9463003 9470205 9438032 9471877 9471382 9471755 9471886 9471768 9469959 9471764 9471222 9472185 9472184 9463155 9463138 9469953 9471769 9473123 9465629 9471770 9462020 9469960 9462078 9473143 9471762 9471752 9471780 9465534 9465541 9473215 9471224 9462018 9473139 9473162 9471765 9473144 9469066 9465454 9473209 9473116 9473146 9473117 9452506 9473211 9474117 9465510 9471840 9470109 9465506 9470117 9452469 9470492 9464947 9473196 9473838 9473053 9471843 9465493 9465495 9464971 9465607 9462328 9469034 9471829 9468884 9465498 9470104 9473194 9470496 9471847 9475292 9474130 9468877 9470116 9470490 9471815 9468871 9470098 9471819 9471337 9470107 9474128 9470113 9475220 9464970 9472393 9470118 9473829 9471818 9465512 9472391 9474126 9471846 9465511 9475684 9450009 9475660 9469047 9473837 9477199 9465487 9471824 9471845 9471830 9457723 9475500 9457724 9469026 9464912 9462040 9458000 9462098 9478218 9462075 9475675 9465439 9477198 9465513 9469983 9471352 9475226 9475679 9457737 9480658 9471798 9470123 9475291 9471813 9465614 9472380 9471804 9471787 9481021 9480983 9481012 9480987 9481109 9482848 9480900 9481112 9481115 9480988 9480984 9475502 9480943 9483069 9480985 9477205 9477161 9483137 9482847 9480344 9477175 9483176 9480634 9481674 9470495 9481668 9480664 9483170 9480632 9483149 9475222 9480654 9477180 9477184 9471828 9475306 9470497 9475305 9478200 9480655 9480336 9471346 9480770 9480854 9480673 9480325 9480326 9475228 9481672 9478190 9471339 9480341 9482917 9471822 9475486 9480660 9480335 9480749 9480317 9470106 9463135 9474132 9475232 9465476 9465494 9474142 9480319 9480769 9480725 9480724 9482934 9482919 9485699 9481671 9469057 9470119 9482920 9475207 9473079 9465477 9482933 9472392 9477186 9485703 9485689 9483702 9426952 9470105 9486511 9485849 9477085 9485851 9486682 9485850 9464977 9485811 9475221 9485810 9470124 9477190 9482914 9474133 9480752 9480633 9485826 9485456 9471799 9485724 9486680 9485827 9472390 9487952 9483178 9485725 9464975 9473830 9480318 9470196 9487028 9471338 9471814 9475218 9480766 9475223 9487933 9475308 9487021 9470484 9488534 9462990 9465497 9488600 9487930 9487951 9485692 9471825 9470494 9487941 9471353 9486683 9483172 9485809 9474125 9483572 9487939 9485459 9474134 9487927 9477185 9480768 9488602 9485860 9485825 9480810 9470195 9488515 9445899 9429998 9440741 9431662 9440956 9446223 9445388 9438476 9432620 9448834 9440761 9448804 9438401 9445621 9448063 9438515 9448693 9444289 9444287 9448809 9438447 9448832 9430179 9438792 9444153 9440970 9445386 9448835 9444291 9486433 9490558 9448692 9487932 9488514 9488512 9487936 9490559 9490546 9491977 9488551 9488583 9490537 9491978 9488550 9483063 9485931 9492004 9492436 9488603 9492025 9465514 9488556 9488591 9491990 9491998 9485957 9483709 9492029 9490535 9488590 9492853 9491979 9492024 9486510 9492002 9489245 9486475 9492027 9475208 9487961 9485727 9488542 9492908 9486414 9492946 9492941 9491996 9492012 9490553 9492948 9491999 9492843 9492026 9492835 9492014 9492955 9491989 9490560 9493894 9485728 9493895 9489251 9473845 9469060 9471831 9489193 9480340 9493008 9493015 9471816 9493944 9493892 9471821 9492375 9471820 9492914 9493012 9489216 9493945 9492830 9489195 9492829 9495066 9494898 9493005 9493943 9487935 9489185 9493951 9495862 9490755 9495860 9492378 9488540 9495865 9495851 9481669 9494895 9438052 9488535 9493950 9488610 9488604 9471844 9480801 9465515 9471827 9488513 9487023 9487024 9487926 9488609 9488541 9496610 9487019 9489217 9464950 9489204 9490423 9488536 9496584 9470493 9487928 9487938 9487929 9487934 9492013 9492821 9487960 9483166 9488507 9490426 9496608 9487937 9488538 9492367 9497332 9493011 9471354 9497331 9497320 9488539 9497333 9492387 9495063 9497322 9488529 9496607 9497298 9497321 9485863 9496612 9485862 9494907 9465491 9487942 9497330 9497342 9496619 9492820 9496601 9492377 9494924 9465499 9497407 9495068 9471351 9488516 9490529 9497408 9494892 9496578 9494891 9497402 9497404 9497403 9490434 9473195 9496602 9497294 9497323 9477183 9496611 9490530 9490402 9471826 9471357 9473192 9497386 9480337 9471832 9465492 9472384 9492373 9480751 9490435 9497366 9497359 9483150 9497285 9497347 9496670 9496663 9497607 9497926 9496650 9495874 9496666 9497271 9496658 9496657 9497933 9497280 9500369 9497935 9500156 9497354 9500140 9500164 9500152 9500151 9500145 9452477 9501323 9501328 9471805 9493006 9486689 9485721 9501325 9500366 9495549 9493920 9497611 9501330 9502310 9497616 9502315 9493926 9502309 9502307 9496343 9497262 9496667 9496640 9502334 9502516 9502330 9502513 9502534 9502314 9492443 9502537 9502514 9502345 9503080 9502512 9502515 9493902 9497923 9503079 9503078 9488561 9502342 9502525 9503068 9503142 9503087 9502319 9503064 9497995 9497629 9497994 9497991 9500340 9497998 9497401 9500206 9501689 9501315 9501312 9501293 9501686 9501685 9500294 9500336 9490420 9498006 9474129 9501693 9490524 9500327 9498009 9470485 9470489 9501692 9492913 9492810 9482935 9500329 9492812 9464973 9490433 9492372 9490432 9492834 9474124 9493014 9490523 9498000 9480748 9490425 9492388 9492384 9492371 9492811 9487025 9503176 9485822 9487029 9487032 9491800 9492814 9495067 9489205 9469058 9477182 9489214 9492376 9492809 9489194 9489206 9487030 9503170 9489196 9483171 9470120 9471355 9501317 9469065 9480874 9470487 9473841 9500326 9501313 9503179 9503177 9497668 9503169 9505148 9412392 9503180 9506106 9491783 9501299 9490437 9500207 9500335 9490403 9480323 9497679 9490419 9492385 9501690 9507289 9500203 9497999 9464972 9493009 9507290 9501314 9505154 9503522 9469056 9487026 9485716 9485686 9416054 9503040 9506172 9506138 9506144 9506171 9506163 9500295 9495847 9507322 9501316 9501297 9507321 9506146 9506164 9507325 9505171 9506130 9505155 9506161 9502293 9508601 9508603 9503141 9508541 9508542 9503469 9508597 9507694 9508640 9508580 9503555 9508596 9508628 9508560 9508636 9505559 9507689 9508643 9505567 9508559 9506292 9506932 9503566 9506927 9508594 9507313 9506124 9508430 9506139 9503562 9508427 9508406 9506134 9508639 9507317 9502495 9508622 9507299 9500296 9502499 9507324 9500205 9506131 9500204 9500202 9508431 9512715 9505572 9503523 9508435 9505557 9507696 9511177 9505563 9508451 9512712 9503556 9507693 9506296
9505571 9503554 9508444 9507706 9508585 9503567 9506122 9512708 9513157 9506300 9514664 9512690 9512716 9512728 9512691 9505592 9511165 9438646 9430003 9438850 9448681 9430848 9431659 9438643 9448651 9445441 9441872 9429969 9437795 9448001 9444136 9444137 9438400 9440774 9448068 9445465 9438642 9438647 9444836 9438790 9432655 9430000 9448653 9515158 9511191 9514663 9514679 9512736 9514676 9512698 9512720 9505562 9515172 9515760 9512689 9515770 9512700 9512719 9512704 9512722 9496579 9500130 9508591 9505566 9508428 9480316 9503521 9506133 9512724 9512718 9514660 9514678 9503482 9480767 9511206 9515177 9515744 9515817 9516736 9515745 9512717 9515786 9516740 9515833 9515831 9515167 9516734 9512738 9512739 9516741 9515822 9515757 9515178 9515170 9512673 9511166 9515761 9515799 9508595 9515802 9512730 9511237 9516745 9515800 9519986 9519965 9515803 9515825 9515798 9520304 9520255 9515807 9520252 9506165 9517637 9515161 9501708 9516735 9520297 9515787 9519982 9515827 9520305 9520257 9515793 9515844 9511257 9512725 9520293 9515771 9520310 9515847 9520302 9517638 9515297 9515313 9515299 9512275 9512277 9506125 9515321 9512274 9515343 9490525 9521193 9523346 9508434 9512253 9506143 9508623 9515318 9515360 9523358 9512273 9523342 9520234 9523347 9515317 9521183 9520269 9512276 9523352 9524196 9520281 9520271 9514876 9523343 9519921 9520284 9519920 9523348 9516344 9520225 9515298 9524813 9524235 9512262 9523345 9515397 9515266 9520233 9520272 9515826 9521980 9521946 9521977 9520737 9520728 9520736 9515801 9515816 9521979 9520735 9516739 9521939 9521974 9512272 9515750 9512263 9512254 9524823 9515349 9519919 9515306 9521192 9474127 9524195 9520282 9516348 9523357 9512261 9523688 9524187 9530431 9514654 9523687 9516345 9515830 9520242 9514661 9520283 9520240 9530395 9519918 9530486 9530388 9530430 9515319 9515311 9519926 9521198 9514652 9530488 9438645 9531280 9530432 9520221 9515308 9520241 9524182 9516347 9514653 9524181 9515342 9524188 9515404 9509811 9449116 9530400 9503042 9477187 9535138 9535087 9534819 9511228 9521985 9485861 9521984 9490424 9521936 9521955 9535137 9535431 9530475 9535373 9535372 9535374 9515310 9530387 9534780 9522052 9522974 9531580 9521945 9535444 9534635 9534811 9530480 9531570 9534874 9521992 9531557 9531582 9534889 9534856 9534862 9531576 9534852 9522014 9530487 9524225 9534620 9534618 9522012 9531010 9531583 9522987 9534877 9534851 9534820 9534830 9522001 9535433 9521953 9534821 9531584 9521958 9521954 9531581 9521986 9531561 9534630 9462300 9530474 9534853 9537259 9522988 9537266 9534628 9540091 9534855 9531560 9535451 9534875 9534815 9540115 9534854 9539966 9534623 9530481 9530393 9530472 9521203 9535376 9536935 9536938 9539965 9535364 9536936 9535088 9536972 9492832 9540016 9536989 9540010 9540012 9536973 9536985 9540014 9540013 9540017 9540026 9536974 9531273 9540937 9536988 9535089 9465496 9468925 9536958 9540938 9540011 9539998 9475224 9536957 9537206 9537212 9540913 9537242 9515400 9535424 9540051 9540056 9537343 9543276 9543200 9543190 9537317 9534876 9534872 9537200 9511225 9540114 9534812 9543280 9537270 9540131 9543263 9543776 9531559 9545121 9537211 9537195 9545117 9516733 9522054 9545120 9508629 9543211 9543790 9543826 9537324 9543291 9543784 9543201 9543298 9538171 9545119 9524824 9536987 9535375 9543993 9543217 9543221 9545118 9545087 9536986 9536937 9536991 9543283 9543213 9543988 9546120 9546141 9543777 9546712 9546131 9543212 9546743 9543994 9540090 9543274 9540031 9537001 9546737 9546129 9546734 9541639 9543202 9543989 9549252 9546735 9546745 9546124 9543788 9546741 9546142 9549553 9549230 9543214 9543297 9546119 9546130 9512245 9549243 9546744 9543818 9543859 9540015 9536978 9543855 9540032 9549241 9543290 9546747 9536887 9546748 9545069 9543999 9549238 9540120 9549223 9552682 9539968 9549691 9549099 9549681 9549683 9540086 9540914 9540085 9512654 9545116 9551950 9545094 9545099 9536997 9540033 9539979 9543210 9543786 9549684 9545106 9549114 9549113 9549115 9549097 9549709 9549098 9545095 9537005 9536984 9549123 9549110 9543856 9553627 9540025 9543998 9543857 9536942 9543851 9537000 9546733 9551944 9549687 9546137 9540135 9545100 9549694 9551934 9549688 9546746 9546732 9543854 9553663 9553658 9546731 9553662 9549112 9546736 9553659 9545088 9553660 9543858 9554324 9551943 9537004 9554326 9543849 9551942 9554344 9554342 9554343 9553626 9545055 9554319 9537255 9546138 9449160 9440969 9537261 9554308 9543275 9549710 9551928 9554325 9553665 9536941 9555980 9554309 9551926 9554338 9551931 9543279 9550470 9543865 9537202 9535418 9541588 9546742 9536992 9549122 9543850 9536977 9537362 9521970 9553725 9551923 9552673 9554330 9551924 9555977 9558213 9550471 9558494 9554312 9558486 9540084 9554337 9554282 9554314 9558214 9537361 9558212 9558215 9558370 9543987 9543204 9545057 9560177 9543990 9531594 9554281 9560183 9560176 9554246 9558484 9558368 9543216 9552695 9551922 9555979 9554286 9551927 9552707 9535422 9540092 9543827 9560187 9560184 9560185 9554311 9558383 9561165 9545056 9551930 9553722 9554304 9555961 9564314 9558485 9560186 9554316 9563090 9558386 9563856 9552706 9551949 9543860 9483089 9555962 9549111 9553723 9553715 9549700 9549699 9566964 9570400 9570404 9564308 9571382 9570403 9563079 9562748 9566963 9560838 9560887 9561156 9554284 9549697 9563083 9562728 9571647 9562758 9571370 9571371 9571646 9562723 9570405 9570406 9571385 9562750 9571648 9558369 9562727 9570053 9570070 9570183 9566965 9570051 9566936 9563080 9554331 9554313 9558384 9554287 9571649 9564310 9570182 9570407 9554244 9570184 9570052 9558483 9551933 9543828 9571861 9570180 9570181 9570202 9570056 9570222 9570071 9570201 9554329 9566895 9566896 9570072 9574219 9566935 9574218 9574244 9574220 9574213 9574743 9574239 9574238 9574214 9574742 9570055 9570200 9570214 9571857 9574561 9570215 9575497 9575987 9575499 9570210 9575995 9574559 9575233 9575498 9575997 9576301 9576305 9576312 9570054 9576299 9575989 9575986 9576330 9576328 9575988 9575994 9576333 9576327 9576329 9575231 9576313 9570211 9575230 9576304 9576302 9576300 9576303 9576436 9575232 9576311 9576332 9576000 9571858 9576331 9577409 9576528 9577441 9576529 9577416 9576534 9577439 9577440 9576532 9576517 9577426 9576536 9577417 9576454 9576516 9577436 9577435 9576435 9576531 9577410 9576523 9577658 9576522 9576452 9577437 9577654 9577656 9577721 9576535 9577657 9577678 9577443 9577438 9577425 9577424 9577650 9577415 9577659 9577677 9577725 9578490 9577653 9576533 9577661 9576455 9578733 9577652 9577727 9577724 9578502 9578992 9578503 9578483 9577728 9574739 9574245 9574243 9578482 9578700 9577651 9577726 9578504 9578486 9577649 9578480 9577442 9577660 9577682 9578514 9578705 9578485 9578706 9578956 9578734 9578484 9577681 9579000 9578960 9578876 9578979 9578516 9578735 9578754 9578961 9578738 9578699 9578964 9578481 9578871 9577423 9578694 9580743 9578962 9578513 9570223 9578693 9578065 9574250 9577986 9578064 9577979 9577976 9578068 9578063 9578049 9577980 9578886 9578067 9577975 9577969 9578050 9578061 9577978 9574753 9577955 9577983 9574748 9575823 9577984 9578899 9574755 9574741 9574754 9578891 9574749 9574737 9574746 9578087 9574249 9574750 9578890 9578070 9577960 9578903 9578088 9578889 9578888 9582390 9582407 9578993 9582408 9579800 9578901 9578930 9578909 9577987 9578904 9578986 9578066 9577961 9578900 9574751 9578892 9577956 9578203 9575782 9574745 9575822 9578989 9581479 9582404 9580755 9582415 9578707 9580789 9580742 9578701 9582401 9584096 9584098 9584094 9583771 9583753 9578910 9584095 9578060 9574756 9574744 9577977 9583754 9583764 9574738 9584097 9584108 9584127 9584150 9578737 9583322 9578963 9580785 9585728 9585036 9578708 9584155 9580721 9582417 9580758 9585703 9578687 9582254 9583331 9584147 9585699 9585778 9585777 9575771 9582400 9578916 9578902 9580720 9580771 9582252 9582402 9580711 9580764 9585773 9578069 9585772 9585775 9585774 9585776 9583332 9584113 9584093 9583742 9582253 9585731 9583306 9578987 9584112 9585706 9585014 9585015 9584130 9580713 9585698 9585085 9585068 9585083 9585722 9584119 9587712 9582416 9587713 9584128 9584109 9583743 9580765 9583770 9585705 9583752 9585462 9587700 9586148 9584129 9582393 9587709 9584086 9580725 9580733 9587699 9587711 9584087 9584111 9586151 9586134 9585075 9583334 9578959 9586132 9580730 9585730 9580748 9588080 9585050 9585055 9585054 9586137 9583309 9585013 9590051 9582413 9580726 9582248 9580772 9582258 9582392 9589994 9582259 9580716 9584085 9585040 9584122 9586144 9585037 9584121 9585701 9584151 9578922 9587701 9580719 9586142 9592647 9586133 9588078 9585460 9584083 9588097 9584131 9588096 9586139 9586146 9592597 9586145 9586165 9592595 9584084 9588095 9591545 9588084 9586136 9591538 9588483 9586150 9586140 9588482 9586141 9586131 9587710 9592598 9588091 9584110 9584154 9588071 9586163 9584114 9584092 9588488 9583765 9592646 9588079 9591535 9586164 9589570 9595016 9588070 9590044 9585043 9591558 9591793 9593952 9584186 9586152 9587722 9587721 9588487 9588089 9597721 9589613 9591813 9590040 9580724 9592637 9590012 9590036 9584187 9590015 9585035 9589917 9590013 9588523 9588522 9585079 9597720 9590062 9589617 9590014 9586147 9588083 9580767 9591791 9591794 9589679 9591812 9597715 9592638 9588504 9592594 9595015 9597987 9585022 9591792 9597724 9589991 9591476 9582263 9597722 9585087 9592633 9588525 9597980 9589986 9593941 9590412 9583308 9593953 9589659 9592613 9589987 9585711 9590045 9589997 9589660 9580723 9600669 9590415 9589988 9589671 9585704 9582257 9591795 9599788 9600623 9597716 9603264 9585030 9591478 9603172 9593935 9603268 9592641 9591466 9588524 9592285 9597501 9601410 9592284 9592640 9603284 9603151 9603296 9588503 9592621 9592286 9597511 9597517 9603322 9603806 9603808 9603807 9601852 9601417 9601853 9605963 9603787 9603769 9584188 9579804 9603775 9603150 9603785 9605962 9603803 9603792 9608733 9601411 9603288 9603870 9603148 9580731 9582262 9591581 9603168 9603261 9606735 9599773 9597982 9606712 9589992 9597714 9606738 9600654 9597713 9597562 9589673 9585021 9589677 9605559 9591580 9605562 9605404 9597978 9597983 9600611 9597565 9600652 9597977 9585020 9605462 9589674 9606720 9600650 9600625 9605550 9589996 9580718 9605554 9599780 9606632 9603781 9600613 9603786 9608720 9608736 9608739 9603289 9608734 9603789 9595653 9610003 9595649 9603784 9608747 9606631 9603782 9608721 9608718 9603776 9603797 9608726 9592614 9608722 9608717 9603292 9608755 9603793 9603790 9588090 9591557 9606719 9605595 9601415 9591590 9605597 9608738 9592622 9603795 9605961 9592642 9603794 9605414 9605596 9582249 9589566 9603788 9585080 9606736 9589995 9600651 9591598 9605467 9589676 9590039 9603259 9593968 9603921 9593967 9592605 9608719 9608746 9601854 9603323 9605969 9609612 9603796 9612461 9597524 9609607 9603791 9612459 9612957 9613117 9612988 9613122 9613118 9603783 9603873 9606696 9605421 9606711 9597981 9606778 9597563 9605466 9605526 9589680 9605564 9603260 9605420 9606675 9606676 9605537 9605548 9611593 9589989 9608754 9605555 9612460 9605538 9611586 9605415 9606721 9606779 9612659 9611600 9597723 9600624 9605429 9593942 9609611 9580717 9603871 9605527 9605533 9612661 9606754 9611595 9597564 9605534 9605427 9605478 9605524 9597525 9590035 9590038 9611591 9605487 9606722 9600612 9612911 9606700 9605489 9611590 9611636 9605523 9611640 9606657 9613328 9611602 9605490 9606704 9611594 9613329 9606705 9613121 9613119 9612931 9603872 9613985 9614619 9605488 9614950 9614585 9614961 9601416 9614974 9613986 9595659 9614986 9617704 9606727 9618419 9614991 9611631 9611596 9618359 9614982 9618544 9618418 9614620 9618363 9618367 9618406 9617701 9618365 9618876 9611683 9618421 9614729 9617702 9605582 9605584 9611668 9617705 9618364 9611629 9618352 9611679 9618865 9618864 9605417 9611609 9618490 9618832 9618827 9618861 9618829 9611669 9618845 9617706 9618826 9618489 9618846 9611627 9618879 9618512 9618349 9611682 9618522 9618877 9618371 9618493 9611632 9618358 9618420 9618417 9606710 9605553 9590413 9585018 9605612 9618351 9595654 9613285 9595647 9603768 9620429 9618475 9620448 9618444 9618480 9603887 9618476 9618474 9618962 9618482 9619297 9614973 9619513 9618958 9624337 9619511 9619508 9625197 9618957 9625223 9619507 9618481 9618960 9614960 9625179 9619518 9617511 9613120 9627061 9603890 9619283 9614610 9625231 9618477 9618469 9617518 9624245 9620430 9624233 9624244 9625178 9627058 9627054 9624343 9625187 9624536 9624342 9627059 9627060 9623170 9625210 9624231 9619284 9624239 9618329 9614978 9618322 9625222 9624323 9625221 9603892 9627056 9627057 9618484 9625217 9619287 9627055 9625211 9625213 9617517 9627170 9620036 9618828 9618597 9618595 9618583 9618589 9618594 9618596 9619517 9617524 9618834 9620447 9618598 9618836 9620074 9618856 9613116 9618473 9618874 9618565 9613123 9624858 9624851 9624860 9618548 9618499 9625245 9629485 9618531 9627186 9618558 9618564 9618853 9618501 9618860 9625244 9618883 9618528 9624542 9625243 9625263 9629440 9618873 9618502 9618852 9618570 9625249 9628391 9618574 9618835 9629432 9618497 9618880 9628424 9620449 9629386 9628396 9623181 9628429 9618496 9625265 9627189 9618833 9624857 9618884 9618530 9628495 9611630 9628420 9624854 9629435 9628544 9614984 9629452 9619549 9625186 9625271 9629434 9619285 9617730 9619282 9619286 9617725 9619504 9624230 9618511 9628550 9629489 9605614 9618539 9628500 9628394 9618508 9624226 9629490 9620450 9620042 9624293 9630524 9625188 9618572 9629433 9618505 9628549 9627188 9625274 9619551 9629491 9625262 9628397 9618510 9625248 9623172 9614967 9618507 9614949 9628433 9629443 9618538 9625182 9629208 9631094 9625190 9630523 9624234 9614966 9617509 9630276 9625237 9631098 9623171 9619280 9619288 9631092 9619279 9619289 9618537 9619547 9618961 9625212 9630277 9619550 9631459 9630271 9631460 9631461 9635068 |
Recalling Firm/ Manufacturer |
DePuy Orthopaedics, Inc. 700 Orthopaedic Dr Warsaw IN 46582-3994
|
For Additional Information Contact |
Customer service 800-255-2500
|
Manufacturer Reason for Recall |
Certain Pinnacle Cup devices may potentially exhibit an oversized "minor diameter", which could lead the Apex HE to thread through the shell of the cup without stopping or to protrude internally as a result of "cross-threading". If cross-threading occurs, physicians will not feel the positive stop of the Apex HE against the cup.
|
FDA Determined Cause 2 |
Process control |
Action |
On December 15, 2020, the firm distributed Urgent Medical Device Correction letters to customers. The letter informed customers that there is a potential for an oversized "minor diameter" on certain Pinnacle Cup devices. The correction is intended to advise physicians of the oversize condition of the hole in the cup, which could lead the Apex HE to thread through the shell of the cup without stopping or to protrude internally as a result of cross-threading. This may occur if the Apex HE is manipulated or toggled while threading it through the hole of certain Pinnacle Cup devices. If cross-threading occurs, physicians will not feel the positive stop of the Apex HE against the cup.
Customers were asked to review the information in the letter and complete and return the included customer response form. If any of the affected products have been forwarded to another facility, please contact that facility and provide a copy of the recall notification to the relevant personnel. Post a copy of the notice in a visible area for awareness.
If you have any questions, please contact your Sales Consultant. |
Quantity in Commerce |
14161 lots (10 units/lot) |
Distribution |
Worldwide distribution - US Nationwide distribution. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|