| | Class 2 Device Recall Philips Respironics V680 Ventilator |  |
| Date Initiated by Firm | November 02, 2020 |
| Date Posted | January 22, 2021 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0917-2021 |
| Recall Event ID |
86697 |
| 510(K)Number | K082660 |
| Product Classification |
Ventilator, continuous, minimal ventilatory support, facility use - Product Code MNT
|
| Product | Philips Respironics V680 Ventilator Part Number 850011 - Product Usage: intended to be used only with various combinations of Respironics-recommended patient circuits, interfaces (masks), humidifiers, and other accessories. |
| Code Information |
Model: V680 Part Number: 850011 V680 Ventilator, Finished Serial Numbers: 201009427 201009428 201009429 201009493 201009494 201009495 201009496 201009498 201009499 201009660 201009661 201009662 201009663 201009664 201009665 201009666 201009667 201009668 201009669 201009670 201009671 201009672 201009673 201009675 201009676 201009677 201009678 201009679 201009674 201008743 201008765 201008766 201008770 201008771 201008776 201008785 201008786 201008805 201009497 201008801 201008741 201008744 201008760 201008764 201008767 201008768 201008772 201008775 201008783 201008808 201008762 201008773 201008781 201008809 201008807 100124645 100124646 100125009 100125011 100125012 100125013 100125014 100125148 100125149 100125150 100125152 100125153 100125154 100125155 201008740 201008742 201008774 201008777 201008802 201008804 100125541 100125544 201008780 201008800 201008803 100125504 100125505 100125509 100125512 100125542 100125545 100125550 100125551 100125552 201008784 100125508 100125546 100125549 100125010 100125516 100125517 100125151 100125515 100125547 201008782 100125513 100125510 100129145 201010488 201010551 201010627 100129141 100129142 100129143 201010254 201010535 201010546 201010674 100126254 100129144 100134076 201010428 201010482 201010490 201010536 201010548 201010620 201010622 201010675 100125518 100129147 100134077 201010255 201010360 201010434 201010485 201010538 201010623 201011218 100125553 100129146 100129148 100134073 100134074 100134075 100134078 100134079 100134180 100134181 100134182 100134183 201010251 201010252 201010256 201010257 201010258 201010259 201010426 201010427 201010429 201010430 201010438 201010484 201010491 201010537 201010539 201010540 201010544 201010545 201010547 201010550 201010552 201010553 201010624 201010625 201010626 201011216 201011217 201011220 100134184 201010439 201010483 100134337 100135409 100125543 100134335 100135396 100135398 100135400 100135401 100135402 100135404 100135405 100135406 100135407 100135631 100135633 100135634 100135635 100135636 100135638 100135639 100135630 100135637 100135642 100135644 100135650 201011219 100135640 100135641 100135643 100135647 100135652 100135397 100135645 100135646 100135648 100135649 100135651 100135653 100135654 100135655 100137426 100137427 100137428 100137429 100137430 100137431 100137433 100137434 100137436 100137437 100137438 100137439 100137440 100137444 100137446 100137447 100137448 100137449 100137450 100138009 100138010 100138011 100138012 100138016 100138017 100138018 100138019 100138508 100138510 100135403 100135632 100137435 100137442 100137443 100137445 100138013 100138080 100138082 100138507 100137432 100137441 100138008 100138081 100138083 100138086 100138087 100138503 100138512 100138513 100138514 100138515 100138522 100138084 100138085 100138509 100138511 100138517 100138518 100138519 100138520 100138521 100138705 100138707 100138814 100138816 100138818 100138943 100139067 100138506 100138516 100138819 100138940 100138941 100138942 100139064 100139066 100139068 100139069 100139070 100139071 100139074 100139075 100139077 100139078 100139079 100139080 100139082 100139083 100139084 100139087 100140238 100140239 100140300 100140301 100140302 100140303 100140304 100140305 100140310 100140311 201012739 201013230 201013231 100139076 100139081 100139088 100140308 100140315 100140317 201012733 201012735 201013213 201013220 201013221 201013224 201013225 201013226 201013228 201013229 100138504 100139072 100139085 100140312 100140314 100140316 201012732 201012734 201012736 201013209 201013210 201013211 201013212 201013215 201013223 201013227 201013233 201013234 201013235 100139065 100139073 100140309 201012738 201013214 201013236 100139086 100140306 100140313 100142426 100142428 100142429 100142431 100142432 100142434 100142437 100142460 201013217 201013218 201013219 201013222 100135399 100135408 100142637 100142427 100142436 100142438 201013216 201013232 100142634 100142435 100142638 100142640 100142641 100142635 100142636 100142642 201013271 201013272 201013275 201013273 201013276 201013279 201013720 201013725 201013747 100142639 201013274 201013277 201014176 201014178 201014179 201014182 201014183 201014211 201014213 201013713 201013717 201013749 201013887 201014185 201014214 201013890 201014177 201014181 201014184 201014196 201014198 201014199 201014215 201014216 201014217 201014218 201014219 201013709 201013710 201013711 201013712 201013714 201013715 201013716 201013724 201013744 201013748 201013750 201013753 201013888 201013889 201013891 201013893 201014197 201014340 201014180 201014324 201014342 201014388 201014195 100146372 100146374 201014322 201014323 201014326 201014328 201014386 100146369 100146370 100146371 100146373 100146873 201014321 201014325 201014327 201014341 201014343 201014344 100147209 100146442 100146875 100146375 100146876 100147212 201014544 100147217 201014534 201014536 201014542 201014600 201014538 100147215 100146376 100146377 100146440 100146443 100146872 100147218 100147300 201014329 100146441 100146869 100147219 201014543 100146871 100147211 100147301 201014537 201014539 100146867 201014384 201014385 201014535 201014545 100146378 100146379 100146868 100146870 100146874 100147210 100147216 201014330 201014387 201014541 201010431 201010489 100147214 201010486 201013746 201010487 100147208 100147213 201010432 201010549 201010621 201011215 100129149 100147302 100148513 100148515 201010433 100154345 100153251 100153256 100154348 201015021 100153252 100154346 100154349 100154353 100156200 100156201 100160490 201015022 100154721 100154722 100153451 100154696 100154698 100153257 100153448 100153449 100154352 201015023 100153447 100154697 100154699 100154724 100156105 100156202 100156204 100156205 100160493 100160494 100160495 100160496 100160962 100160975 100160663 100153450 100154347 100154351 100154354 100156206 100160489 100160492 100160498 100160664 100160665 100160666 100160668 100160918 100160919 100160960 100160964 100160965 100160966 100160968 100160969 100160970 100160971 100160972 100160973 100160974 100160976 100160977 100160978 100160979 100160980 100160981 100160982 100160983 100160984 100161100 100161101 100161102 100161103 100161107 100161109 100161110 100161112 100154350 100160667 100160961 100160967 100161105 100161108 100161111 100161113 100161114 100161115 100161116 100161117 100161119 100161120 100160963 100161104 100161106 100161121 100161122 100161845 100161846 100161847 100161848 100162258 100162300 100162303 100162259 100162257 100162301 100161849 100162256 100162311 100162312 100154725 100162255 100162302 100162305 100162306 100162307 100162308 100163248 100162314 100162304 100162313 100163045 100163249 100162310 100163049 100163050 100163051 100163054 100163064 100163066 100163067 100163068 100163242 100163245 100163250 100163357 100163358 100163359 100163420 100163421 100163782 100163784 100163787 100163791 100162309 100163065 100163244 100163246 100163247 100163792 100163047 100163052 100163241 100163786 100163789 100163783 100163048 100163785 100163790 100163793 100163794 100163795 201016701 201016702 201016705 100163788 100165980 100165981 100165982 100165983 100165984 100165985 100165986 100165987 100165988 100165989 100165990 100165991 100165992 201016703 201016704 201016719 201016920 201016922 201016923 100163053 201016926 201017222 100168214 100168219 100168263 100168274 100168282 100168817 100168867 100168868 100168869 100168870 100168874 100168818 100169142 100168814 100168872 100168875 100169144 100169146 100169148 100169380 100168812 100168813 100168815 100168816 100168819 100168866 100168873 100169141 100169143 100169145 100169147 100169149 100169150 100169151 100169152 100169153 100169154 100169155 100169156 100169157 100169158 100169159 100170298 100170301 100170302 100170379 100170440 100170441 100170442 100170447 100169381 100170300 100170444 100169500 100170299 100170443 100170445 100170446 100170449 100170450 100170451 100170452 100170453 100170454 100170455 100170457 100170448 100170456 100170461 100174095 100174096 100174097 100174098 100174099 100174261 100174262 100170460 100174260 100174263 100174392 100174395 100170458 100170459 100174264 100174393 100174394 100174396 100174397 100174399 100174503 100174504 100174506 100175290 100174398 100174505 100175288 100175361 100175362 100175363 100175364 100175365 100175367 100175369 100175371 100175334 100175366 100174507 100174510 100175284 100175286 100175289 100175291 100175292 100175325 100175326 100175327 100175330 100175331 100175332 100175333 100175335 100175337 100175339 100175368 100175372 100175373 100175773 100175774 100175776 100175778 100174502 100175293 100175338 100175360 100175374 100175775 100175777 100175779 100176180 100176181 100176441 100176442 100176443 100175329 100176444 100176445 100176451 100176657 100176800 100176802 100176803 100176805 100176807 100176811 100176825 100176826 100176827 100174501 100176182 100176812 100176813 100176821 100175328 100175336 100176446 100176447 100176448 100176449 100176658 100176759 100176808 100176809 100176817 100176822 100176824 100176450 100176814 100176820 100160661 100177146 100179320 100179996 100179998 100179999 100180003 100180004 100180005 100180001 100180006 100180007 100180011 100180021 100180321 100180325 100180012 100180013 100180013 100180008 100180009 100179819 100180000 100180002 100180010 100180014 100180015 100180016 100180017 100180018 100180019 100180020 100180024 100180323 100180324 100180326 100177145 100179997 100180328 100176275 100176276 100176277 100176278 100176279 100176810 100177140 100177141 100177142 100177143 100177144 100177147 100177148 100177149 100179310 100179311 100179312 100179313 100179314 100179315 100179316 100179317 100179318 100179319 100179324 100161118 100180022 100180320 100180322 100180327 100180916 100180917 100180918 100180919 100181640 100181641 100181642 100181644 100181645 100181646 100181647 100181649 100181652 100181653 100182076 100182074 100182078 100180915 100181643 100182289 100181648 100181650 100181651 100182073 100182075 100182077 100182081 100182290 100182292 100182294 100182295 100182297 100182298 100182299 100182300 100182301 100182302 100182303 100182304 100182305 100182306 100182307 100183045 100183046 100183047 100183049 100183050 100183057 100183059 100183117 100183304 100183355 100183048 100183051 100183054 100183055 100183056 100183114 100183119 100183303 100183305 100183310 100183311 100183312 100183313 100183317 100183318 100183354 100183356 100183359 100183360 100183362 100183053 100183115 100183302 100183307 100183309 100183319 100183353 100183400 100183401 100183403 100183705 100183612 100183713 100183118 100183308 100183538 100183553 100183556 100183557 100183559 100183566 100183604 100183605 100183606 100183607 100183611 100183700 100183702 100183703 100183704 100183706 100183707 100183708 100183709 100183710 100183711 100183714 100183715 100183716 100183717 100183718 100183721 100183724 100183841 100183306 100183527 100183529 100183530 100183532 100183535 100183572 100183608 100183610 100183842 100183843 100183847 100183849 100183850 100183052 100183058 100183358 100183361 100183402 100183524 100183525 100183526 100183528 100183531 100183533 100183534 100183536 100183539 100183554 100183555 100183558 100183560 100183565 100183570 100183603 100183701 100183720 100183844 100183848 100183116 100183537 100183561 100183562 100183563 100183564 100183569 100183571 100183601 100183609 100183722 100183846 100183856 100183845 100183851 100183854 100183567 100183568 100183602 100183719 100183723 100183840 100183852 100183853 100183301 100183357 100183855 100183858 100183857 201020755 201020752 201020756 201020758 100183859 100183860 201020751 201020753 201020754 201021222 201021223 201021224 201021225 201021226 201021227 201021228 201021229 201021230 201021231 201021232 201021233 201021234 201021235 201021236 201021237 201021238 201021239 201021260 201021261 201021262 201021263 201021265 201021267 201021268 201021269 201021271 201020757 201021264 201021272 100190530 100190531 100190532 100190537 100190786 201021266 201021475 201021478 201021479 201021480 100190785 100190529 100190533 100191248 100191320 100192523 201021473 201021474 201021487 100191242 100191243 100191244 100191245 100191246 100191247 100191249 100191250 100191251 100191252 100191253 100191254 100191255 100191256 100191257 100191259 100191321 100191324 100191325 100191326 100192522 100190534 100190535 100190536 100190780 100190782 100190783 100191323 100192526 100192524 100192525 100192531 100193335 100193336 100190538 100190781 100190784 100190787 100191322 100192527 100192528 100192529 100192530 100192533 100192535 100193320 100193321 100193323 100193326 100193327 100193329 100193330 100193331 100193332 100193333 100193334 100191258 100192532 100193322 100193324 100193325 100193337 100193339 100193341 100193342 100193343 100194369 100193340 100193344 201021476 201021477 201021481 201021482 201021483 201021484 201021485 201021486 100190539 201021270 100195367 100195384 100194425 100195300 100195370 100195372 100195373 100195378 100195379 100195387 201022774 201022779 201023060 201023063 100194376 100195309 100195375 201023061 100190788 100194372 100194375 100194377 100194420 100194421 100195303 100195376 100195382 100195385 201023062 201023069 100192534 100194374 100194378 100194422 100194423 100194427 100194428 100194432 100195301 100195302 100195304 100195307 100195368 100195369 100195371 100195374 100195377 100195380 100195381 201022776 100194368 100194373 100194379 100194424 100194426 100194429 100195308 201023066 201023070 201023071 201023072 100194431 100195383 201022773 201022778 201023065 201023074 201023076 201023440 201023447 201023449 100194371 100194430 201023064 201023067 201023068 201023075 201023077 201023439 201023443 201023448 100194370 201023073 201023445 201023450 201023451 201023441 201023454 201023457 201022775 201022777 201023442 201023444 201023452 201023453 201023455 201023459 201023460 100193338 100195306 201023446 201023456 201023458 201023461 201024840 201024841 201024842 201024845 201024846 201024847 201024848 201024849 201024850 201024851 201024844 201024852 201024853 201025424 201026081 201026084 201024843 201024854 201024855 201024856 201024857 201024858 201024859 201025408 201025409 201025410 201025411 201025412 201025413 201025415 201025416 201025425 201025429 201025430 201025431 201025433 201025434 201025435 201025436 201025437 201025839 201026082 201026083 201026085 201026086 201026087 201026088 201026089 201026090 201026093 201025427 201026080 201026092 201026094 201026095 201026097 201026098 201026099 201026100 201026484 201026482 201026483 201026487 201025414 201026091 201026102 201026480 201026481 201026491 201026492 201026494 201026103 201026485 201026488 201026490 201026101 201025426 201026500 201026677 201025417 201025418 201025419 201025422 201025423 201025428 201026489 201026495 201026497 201026498 201026502 201026674 201026675 201026676 201026678 201027007 201027009 201027011 201027012 201027014 201027015 201027016 201027017 201027023 100155777 100155778 100155779 100156203 100160499 100155776 100160660 100176804 100176806 100176815 100176816 100176818 100176819 201025420 201026486 201026493 201026499 201026501 201026503 201026504 201026505 201026506 201026507 201026508 201026509 201027004 201027005 201027006 201027008 201027010 201027018 201027459 201027460 201027461 201027462 201027022 201027028 201027724 201027727 201026496 201027013 201027020 201027025 201027027 201027456 201027457 201027458 201027464 201027721 201027723 201027725 201027726 201027728 201027729 201027720 201026096 201027024 201027465 201027854 201028145 201027679 201027732 201027853 201027858 201027859 201027861 201027862 201028156 201028157 201027463 201027846 201027852 201028142 201025421 201028147 201028440 201027849 201028161 201028443 201028447 201027733 201027848 201028144 201027860 201027026 201027842 201027730 201027019 201027731 201027734 201027739 201027840 201027843 201027844 201027845 201027847 201027850 201027856 201027857 201028148 201028149 201028150 201028154 201028155 201028159 201028450 201028454 201028457 201028461 201028462 201028463 201028588 201027021 201028446 201028458 201028460 201027865 201028146 201028151 201028152 201028153 201028162 201028163 201028439 201028442 201028444 201028445 201028451 201028453 201028455 201028456 201028141 201028158 201028448 201028449 201027867 201028140 201028143 201028160 201028441 201028459 201028586 201028587 201028591 201027738 201027851 201028452 201028592 201028595 201028596 201028597 201028598 201028747 201028599 201028744 201028745 201030354 201030355 201028742 201030358 201030600 201030602 201028741 201028746 201028750 201030357 201030359 201030601 201030604 201030606 201030608 201030611 201030607 201030610 201028740 201030614 201031225 201031229 201031230 201031231 201030603 201030605 201030609 201030615 201031224 201031228 201031233 201031234 201031236 201031443 201031227 201031232 201031235 201031239 201032045 201031237 201031238 201031441 201031442 201032046 201032047 201032048 201031440 201032044 201032291 201032293 201032442 201032443 201028589 201028590 201028593 201028594 201032151 201032152 201032290 201032309 201032440 100218839 100218840 100218841 201032153 201032294 201032298 201032311 201032441 201032445 201032447 100218818 100218820 100218822 100218823 100218824 100218825 100218826 100218828 100218829 100219117 100219119 100219120 100219121 201031226 201032049 201032050 201032051 201032052 201032053 201032147 201032148 201032149 201032150 201032289 201032296 201032297 201032310 201032312 201032313 201032315 201032316 201032444 100218821 100218836 100218817 100219037 100219322 100219325 100219332 100220764 100220765 100220768 201032777 201032778 201032844 100219036 201032779 201032780 100218837 100219038 100219327 100219330 100219331 100220766 100220767 100220769 100220770 100220771 201032446 201032781 201032782 100218827 100219031 100219034 100219320 100219321 100220775 100220780 100220784 100220787 201032314 100218835 100219030 100219033 100219122 100219329 100220773 100220781 100221889 100221899 100218833 100220788 100221892 100222822 100218834 100219028 100221900 100222838 100221908 100222836 100222841 100224287 100224288 100222833 100222840 100225758 100224294 100224293 100224296 100225757 100221890 100221896 100221898 100222821 100222831 100227938 100219029 100220779 100221895 100221902 100221910 100222638 100219323 100220772 100221893 100218831 100218832 100218838 100219032 100219035 100219039 100219326 100220776 100220777 100220778 100220782 100220783 100220785 100221891 100221894 100221897 100221901 100221903 100221909 100221911 100221912 100222639 100222817 100222818 100222819 100222826 100222832 100224295 100225761 100226491 100226495 100226496 100226499 100226520 201032950 201032951 201032952 201032953 100226494 100227972 100227974 100226493 100220786 100229271 100229272 100229273 100229274 100229275 100229277 100229278 100229279 201032295 100229276 100229280 100229281 100229312 100229313 100229314 100229315 100229316 100229317 100229318 100229319 100229441 100229442 100229444 100229445 100229440 100229443 100229446 100230807 100230808 100230809 100230811 100230812 100230814 100230815 100230816 100230818 100230856 100231124 100230810 100230817 100230821 100230859 100231121 100231122 100231123 100231126 100231417 100231134 100231416 100231419 100230820 100231128 100231132 100231133 100231462 100231463 100231664 100231665 100231666 100231667 100231668 100231669 100230813 100230819 100230857 100231127 100231129 100231130 100231131 100229266 100231125 100231415 100229267 100231671 100231784 100227960 100231697 100231698 100231780 100232322 201033110 100222823 100229268 100224291 201033135 100222637 100227970 100229269 201033118 100222824 100222834 100224290 100226498 100227964 100227973 201033117 100225756 100225762 201033115 100221913 100222820 100224297 100227965 100227977 100219328 100222635 100222825 100222828 100222829 100222837 100224292 100225755 100225759 100225763 100226492 100227937 100227961 100227963 100227966 100227967 100230855 100230858 100231418 100231460 100231461 100231464 100231670 201033106 201033107 201033108 201033114 201033116 100222830 100222835 100227962 100227978 100222634 100222636 100227976 100229270 100227975 201033109 100221906 100231785 100232723 100231699 100232329 100232721 100232735 201033908 201034043 100231783 100231786 100232321 100232324 100232325 100232326 100232327 100232328 100232330 100232720 100232722 100232727 100232728 100232729 100232733 100232736 201033796 201033797 201033911 201033912 201033913 201033914 201034049 201034050 201034052 100232323 100232724 100232725 100232726 100232730 100232731 100232734 201033794 100232732 100238558 100238559 100238588 100238593 100238761 201033795 201033798 201033799 201033907 201033909 201033910 201033915 201033916 201034040 201034041 201034042 201034051 201034053 201034092 201034093 201034094 201034095 201034097 201034098 201034099 100221904 100231672 100238580 100238581 100238582 100238583 100238584 100238585 100238586 100238587 100238589 100238590 100238591 100238592 100238594 100238595 100238596 100238597 100238763 201034096 201034160 201034161 100225764 100238765 100238768 100238762 100238766 100238767 100238769 100238770 100238931 100238932 100238933 100238934 100238935 100238936 100238937 100238938 100238939 100239262 100239264 100239265 100239266 100239270 100239272 100239263 100239269 100239271 100239274 100239519 100239795 100239796 100239798 100239799 100240120 100240121 100238764 100239260 100239261 100239275 100239517 100240140 100240141 100239797 100240122 100240123 100240143 100239267 100240079 100240144 100239268 100239273 100239518 100240073 100240074 100240075 100240076 100240078 100240142 100241293 100241296 100241292 100241294 100241295 100241297 100241299 100241300 100241301 100241764 100240077 100241298 100241762 100241763 100241769 100241768 100241765 100241767 100241771 100241766 100241770 100242962 100242963 100242969 100242971 100242972 100242973 100243770 100243771 100243774 100243775 100243776 100243778 100244319 100242970 100243779 100244320 100244323 100244325 100244326 100244327 100244328 100244329 100244331 100243777 100243840 100244318 100244321 100244322 100244324 100244333 100244334 100244335 100244337 100244330 100244332 100244914 100244915 100244916 100244917 100244919 100244336 100244918 100244920 100244921 100244922 100244923 100244924 100244925 100244926 100244927 100244928 100244929 100244930 100244932 100244933 100245753 100245754 100245755 100243772 100244931 100245757 100245756 100245758 100245759 100245760 100245761 100245762 100246942 100246944 100246945 100246947 100246946 100246943 100246948 100246949 100246950 100246951 100246952 100246954 100246955 100246956 100246957 100246958 100246959 100246961 100247849 100247850 100247851 100247853 100247855 100246953 100246960 100247852 100247854 100247856 100247857 100247858 100247859 100247861 100247864 100247865 100247866 100247867 100248328 100248331 100248333 100247868 100247863 100248324 100248325 100248327 100248329 100248330 100248332 100248335 100248337 100248338 100248336 100248341 100248342 100248343 100248344 100248345 100248346 100248347 100248348 100249203 100247862 100249206 100249208 100248326 100248340 100249204 100249207 100249210 100249212 100249213 100249802 100248339 100249209 100249211 100249214 100249215 100249216 100249217 100249801 100249803 100249804 100249805 100249806 100249807 100249205 100250065 100250066 100250069 100250070 100250071 100250073 100250067 100250068 100250072 100250074 100250076 100250077 100250079 100250481 100250483 100250486 100250078 100250482 100250484 100250485 100249809 100249815 100249816 100250487 100249800 100249808 100249810 100249811 100249812 100249813 100249814 100249817 100249818 100249819 100250064 100250488 100250489 100250490 100251604 100251607 100251608 100251610 100251611 100251612 100251614 100251603 100251606 100253109 100253110 100253113 100253114 100253124 100251609 100251615 100253115 100253116 100253119 100253126 100253323 100251613 100253111 100253125 100253299 100251605 100253117 100253118 100253320 100253330 100253333 100253336 100253340 100253329 100253321 100253335 100253338 100253128 100253327 100253121 100253112 100253325 100253332 100253341 100253322 100253324 100253123 100253127 100255957 100257488 100257490 100257496 100257500 100258060 100257491 100253337 100255874 100255875 100257487 100257493 100257495 100257497 100257498 100257499 100258061 100258063 100258065 100258066 100258067 100258068 100257494 100257501 100258064 100260223 100260220 100260222 100260616 100260224 100260625 100260221 100260227 100260618 100260619 100260620 100260621 100260622 100260623 100260624 100260626 100260627 100260628 100260629 100260630 100260631 100260632 100260633 100260634 100260635 100261843 100261844 100261845 100261846 100261847 100261848 100261849 100261850 100261851 100261852 100261853 100261854 100261855 100261856 100261857 100261900 100261901 100261902 100260228 100260229 100260617 100263995 100263996 100263997 100264020 100264023 100264024 100264026 100264279 100264022 100264025 100263992 100263993 100263994 100263998 100263999 100264021 100264300 100264301 100264302 100264303 100264561 100264562 100264563 100264564 100264565 100264567 100264571 100264566 100264568 100264569 100264570 100264572 100264573 100264574 100264575 100264576 100264780 100264785 100264577 100264579 100264781 100264578 100265625 100265629 100265630 100265634 100265639 100265642 100264782 100265631 100265635 100265637 201035019 201035334 201035360 201035224 100265626 100265627 100265638 100265641 201034919 201035085 201034950 201034946 201035197 201035222 201035226 201035225 201035324 201035331 201035362 201034917 100265632 100265633 201034945 201035015 201035198 201035199 201035221 201035320 201035321 201035333 201034949 201035330 201035016 201035017 201035081 201035084 201035337 201035082 201035339 201035363 201034918 201035223 201035323 100264784 100265636 201035018 201034951 201034952 201034954 201035322 201035328 201035335 201035336 201035338 201035365 201035329 100265624 100265628 100265640 100265643 100265644 100265645 100276770 100276771 100276772 100276774 100276776 100276777 100276778 100276779 100276780 100276784 100276786 100276788 100276789 100277103 100277104 100277105 100277106 100277107 100277108 100277109 100277111 100277115 100277116 100277118 100277120 201034947 100276773 100276775 100277117 100277121 100276781 100276782 100277122 100277127 100276785 100277102 100277112 100277113 100277114 100277119 100277123 100277124 100277125 100277128 100277129 100277131 100277164 100277165 100277166 100277167 100277168 100277169 100277170 100277171 100277172 100277173 100277174 100277175 100277176 100277178 100277320 100277321 100277322 100277323 100277327 100277328 100277329 100277330 100277613 100277616 100276787 100277110 100277126 100277612 100277614 100277615 100277618 100277619 100277620 100277177 100277325 100277326 100277333 100276783 100277130 100277331 100277332 100277623 100277624 201035083 201035332 201035361 100264783 100279988 100279987 100279992 100281272 100281460 100277179 100277324 100277617 100277622 100277625 100277626 100279971 100279972 100279977 100279978 100279976 100281271 100281273 100281276 201035080 100277628 100277631 100279973 100279975 100279979 100279981 100279982 100279983 100279984 100279985 100279986 100279989 100279990 100279991 100279993 100279996 100279997 100279998 100279999 100281461 201037287 100279994 100281278 100280000 201037311 201037293 201037307 100277627 100281279 201037279 201037286 201037308 201037380 201037292 100281277 201037272 201037273 201037278 201037285 201037288 201037289 201037290 201037291 201037310 201037312 201037381 201037405 201037407 201037409 201037410 201037271 201037404 201037408 100277621 100277629 201037274 201037275 201037276 201037284 201037309 201037382 100277630 201037277 100279974 100288469 100288471 100288472 100288473 100288474 100288475 100288476 100288477 100288478 100288479 100288580 100288582 100288583 100288585 100288586 100288587 100288588 100289770 100289771 100289772 100289773 100289776 100289779 201037313 100288584 100289774 100289777 100289778 100289781 100290714 100290719 100289768 100289775 100289782 100289783 100290710 100290711 100290712 100290713 100290718 100290720 100290721 100290723 100290715 100290716 100290717 100290726 100290724 100290725 100291598 100291627 100289769 100290728 100291621 100291630 100291631 100292461 100290727 100292455 100292456 100292464 100292465 100292466 100292469 100292721 100291620 100291624 100291626 100292462 100292467 100292720 100290729 100291599 100291629 100291633 100291634 100291637 100292458 100292459 100292468 100291622 100291628 100291632 100291623 100291625 100292460 100292463 100292719 100296195 100296194 100292723 100296198 100296858 100296860 100296864 100297020 100297026 100297028 100297029 100296199 100296197 100296851 100296854 100296855 100297021 100296853 100298489 100298488 100298490 100298491 100296865 100297626 100298500 100298526 100297033 100297024 100297030 100296857 100297022 100298486 100298525 100297630 100298482 100297031 100297628 100297631 100297633 100297634 100297635 100298496 100298527 100298529 100298532 100298533 100298535 100298620 100301143 100298485 100298487 100298530 100298484 100298536 100301086 100298531 100297025 100296196 100301088 100301089 100301090 100301097 100301099 100301093 100302050 100302051 100301095 100302052 100302054 100301092 100301094 100301141 100301142 100297027 100297627 100298534 100301087 100301091 100301096 100301140 100298539 100301144 100302049 100304941 100302057 100302058 100302059 100304602 100305132 100305134 100298481 100301085 100302055 100301098 100302060 100304601 100304942 100304976 100304977 100304978 100304979 100305133 100305340 100305342 100305345 100305346 100305347 100305348 100305351 100305352 100305353 100306068 100305349 100306061 100306064 100306070 100302053 100305341 100305343 100305344 100305350 100305354 100305939 100306060 100306062 100306063 100306065 100306066 100306067 100306069 100306071 100306072 100306073 100306074 100306075 100307211 100307213 100307214 100307215 100307216 100307217 100307218 100307541 100307545 100308426 100308428 100308431 100308432 100308433 100308435 100308436 100308437 100308439 100308440 100308517 100308600 100308602 100308605 100307219 100307540 100307544 100308424 100308429 100308434 100308441 100308518 100308609 100307212 100307542 100307546 100307547 100307557 100308425 100308427 100307210 100307548 100308601 100308607 100308608 100308612 100310181 100310182 100310183 100308604 100289780 100290722 100291635 100292457 100296869 100307543 100308438 100308519 100308603 100308606 100308610 100308611 100308613 100308614 100309856 100309857 100309859 100310180 100310184 100310186 100310187 100310188 100310189 100310190 100310191 100310192 100311610 100311611 100311612 100311615 100311619 100312022 100312023 100312024 100312025 100312196 100312026 100312027 100311614 100312020 100312193 100312195 100312191 100312194 100309858 100312192 100309855 100312021 100312320 100312197 100312198 100311616 100311618 100314606 100314860 100314863 100314864 100314866 100314867 100314870 100314874 100314613 100311613 100314607 100314608 100314609 100314611 100314614 100314617 100314862 100314868 100314872 100314610 100314612 100314615 100311617 100314618 100314865 100315000 100315001 100315002 100315004 100315005 100316027 100316035 100316036 100316037 100314616 100314759 100314861 100314871 100314873 100316029 100316031 100316032 100316033 100314869 100315003 100316209 100316215 100316218 100316221 100316028 100316034 100316205 100316212 100316216 100316217 201038853 201038855 201038858 201038951 201038953 201038950 201038940 201038949 201038854 201038943 201038945 201038946 100316200 100316211 201038850 201038851 201038852 201038859 201038948 100314619 100316030 100316038 100316203 100316204 100316213 100316224 100316202 100316206 100316207 100316208 100316214 100316219 100316220 100316222 100316223 100316225 100316227 100324359 100324460 201038849 201038857 201038941 201038947 201038954 201038955 201038956 201038957 100324351 100324354 100316229 100324353 100324356 100324357 100324461 100324464 100324466 100324467 100324469 100328720 100324470 100328674 100328675 100328676 100328715 100328716 100328717 100328718 100328721 100328761 100328781 100328782 100328783 100329188 100328719 100328780 100328784 100329191 100329192 100329424 100329428 100329437 100329612 100329196 100329433 100324358 100324468 100328677 100328678 100328722 100329187 100329420 100329609 100329614 100329617 100328699 100329434 100329079 100329611 100328679 100329422 100329425 100329431 100328698 100329421 100329423 100329615 100329430 100324352 100324465 100329190 100329194 100329195 100329426 100329435 100329438 100329607 100324463 100329189 100329427 100329613 100329616 100329429 100329436 100329608 100329610 100329618 100329620 100329621 100346820 100346821 100346822 100346823 100346824 100346825 100346826 100346827 100346828 100346829 100346830 100346831 100346832 100346833 100346835 100346836 100346837 100346838 100346839 100346840 100346841 100346842 100346844 100347194 100347195 100347196 100347197 100347198 100347199 100347200 100347204 100347205 100347207 100347208 100347209 100347328 100347329 100347332 100347337 100347338 100347339 100347340 100347342 100347343 100347344 100347347 100347350 100347356 100347359 100347360 100347361 100347362 100347367 100347368 100347369 100347370 100329619 100346834 100346843 100347201 100347202 100347206 100347330 100347331 100347333 100347334 100347341 100347345 100347346 100347348 100347351 100347352 100347353 100347354 100347355 100347357 100347358 100347364 100347365 100347366 100347371 100347203 |
| FEI Number |
2518422
|
Recalling Firm/ Manufacturer |
Respironics California, LLC 2271 Cosmos Ct Carlsbad CA 92011-1517
|
| For Additional Information Contact | 800-722-9377 |
Manufacturer Reason for Recall | Potential issues with ventilator navigation ring interface and liquid ingress:
1) The navigation ring may become inoperative or operate intermittently;
2) The navigation ring may cause ventilator parameter values to change erratically during setting entries and adjustments. |
FDA Determined Cause 2 | Device Design |
| Action | On 11/02/2020, the firm sent an "URGENT-MEDICAL DEVICE CORRECTION" Notification Letter via priority mail informing customer that the V60 Ventilators may experience two potential issues with the navigation ring interface (see table below for additional information).
1. The navigation ring may become inoperative or begin operating intermittently.
2. The navigation ring may not behave as expected, making it difficult for the user to change ventilator settings via the touchscreen.
and if clinicians accept an incorrect setting and leaves the patient on the ventilator without responding to the resulting alarms, the patient could sustain serious injuries.
Customers are instructed:
1. When cleaning V60 ventilators, do not allow cleaning agents to pool on the
ventilator surfaces. This is especially important on the device interface.
2. Check setting changes before and after accepting the changes.
3. If the navigation ring becomes inoperative or operates intermittently during
clinical use, use the touchscreen for all settings changes and navigation,
and contact Philips to schedule service following clinical use.
4. If the navigation ring interferes with your ability to adjust therapy settings via
touchscreen, remove the ventilator from service and contact Philips for
repair.
and to please review the following sections of the user manual:
1. Chapter 10: Care and maintenance. To prevent possible damage to the
ventilator when cleaning, use only those cleaning and disinfecting agents
listed, and do not drip or spray any liquids directly onto any ventilator
surface.
2. Chapter 1: Warnings, cautions, and notes. During clinical use of the
ventilator, monitor patients for changes in respiratory status with each
ventilator setting adjustment. Use an external oxygen monitor to verify
oxygen concentration, and have an alternative means of ventilation available
whenever a ventilator is in use.
Customers are also instructed to complete, sign and return the Acknowledgement and Receipt |
| Quantity in Commerce | 3,453 systems |
| Distribution | Worldwide distribution - US Nationwide including Guam and Puerto Rico; Afghanistan, Albania, Algeria, Andorra, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bhutan, Bolivia, Brazil, Bulgaria, Burkina Faso, Cambodia, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Ethiopia, Finland, France, French Guiana, French Polynesia, Germany, Ghana, Gibraltar, Greece, Guatemala, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jersey, Jordan, Kazakhstan, Kenya, Latvia, Lebanon, Libya, Luxembourg, Malaysia, Maldives, Mali, Malta, Mexico, Monaco, Mongolia, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Caledonia, New Zealand, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Fed., Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, ST. Pier Miquel, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, Venezuela, Vietnam, Wallis Futuna, and Zambia. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = MNT
|
|
|
|