| Class 2 Device Recall Philips Respironics V60 Ventilator | |
Date Initiated by Firm | October 30, 2020 |
Create Date | January 22, 2021 |
Recall Status1 |
Completed |
Recall Number | Z-0892-2021 |
Recall Event ID |
86646 |
510(K)Number | K082660 |
Product Classification |
Ventilator, continuous, minimal ventilatory support, facility use - Product Code MNT
|
Product | Philips Respironics V60 Ventilator Part Number R1053618 - Product Usage: intended to be used only with various combinations of Respironics-recommended patient circuits, interfaces (masks), humidifiers, and other accessories. |
Code Information |
Model: V60 Ventilator, Part Number R1053618 V60 Ventilator, Japan Opt: CFLEX,AVAPS Serial Numbers: 100039155 100054624 100054738 |
Recalling Firm/ Manufacturer |
Respironics California, LLC 2271 Cosmos Ct Carlsbad CA 92011-1517
|
For Additional Information Contact | 800-722-9377 |
Manufacturer Reason for Recall | Ventilator may experience potential issues with the backup battery failures: 1) if the battery is allowed to over discharge, or 2) if the ventilator is stored for an extended period of time without being plugged into AC power. |
FDA Determined Cause 2 | Device Design |
Action | On 10/30/2020, the firm sent an "URGENT-MEDICAL DEVICE CORRECTION" Notification Letter via priority mail informing customer that the Ventilators may experience two potential issues with the battery (see table below for additional
information).
1. The ventilator may exhibit a high priority Check Vent: Battery Failed alarm if the battery is allowed to over discharge.
2. The battery may fail to charge if the ventilator is stored for an extended period of time without being plugged into AC power.
If the ventilator exhibits a high priority Battery Failed alarm while connected to
AC power, the V60 ventilator will annunciate an audible and visual alarm. The
ventilator will continue to deliver prescribed therapy; however, the audible and
visual alarm will continue until the device is powered down; therefore, the patient
should be transferred to an alternate source of ventilation as soon as one
becomes available. The hazard associated with transferring a patient to an
alternate source of ventilation is a temporary loss of therapy leading to potential
hypercarbia or hypoxemia (moderate).
If AC power fails and the backup battery is not functioning, an audible and visual
alarm will annunciate for at least two minutes. The ventilator will no longer be
operative resulting in a total loss of therapy. The patient will require an alternate
source of ventilation. The hazard associated with a ventilator shutdown is a total
loss of therapy leading to potential hypercarbia or hypoxemia (severe).
Customers are instructed:
Follow the instructions in the V60 user manual and the instructions below to
reduce any risk associated with potential battery issues.
1) Avoid allowing the ventilator battery to become completely discharged.
Otherwise, the battery may become over-discharged and require long
recharge times of up to 16 hours or more. The over-discharged condition
may permanently damage the battery so that it is unable to recharge. To
prevent the occurrence o |
Quantity in Commerce | 3 systems |
Distribution | Worldwide distribution - U.S. Nationwide distribution including in the states of AK, AL, AR, AS, AZ, CA, CO, CT, DC, DE, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE,NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of Afghanistan, Albania, Algeria, Andorra, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bhutan, Bolivia, Brazil, Bulgaria, Burkina Faso, Cambodia, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Ethiopia, Finland, France, French Guiana, French Polynesia, Germany, Ghana, Gibraltar, Greece, Guatemala, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jersey, Jordan, Kazakhstan, Kenya, Latvia, Lebanon, Libya, Luxembourg, Malaysia, Maldives, Mali, Malta, Mexico, Monaco, Mongolia, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Caledonia, New Zealand, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Fed., Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, ST. Pier Miquel, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, Venezuela, Vietnam, Wallis Futuna, and Zambia. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
|
510(K) Database | 510(K)s with Product Code = MNT
|
|
|
|