| Date Initiated by Firm | February 16, 2021 |
| Create Date | April 13, 2021 |
| Recall Status1 |
Terminated 3 on December 08, 2022 |
| Recall Number | Z-1417-2021 |
| Recall Event ID |
87575 |
| Product Classification |
Culture media, non-selective and non-differential - Product Code JSG
|
| Product | MedSchenker Smart Transport Medium (STM) System: CULTURE MEDIA, NON-SELECTIVE AND NON-DIFFERENTIAL - MULTIPURPOSE CULTURE MEDIUM - Product Usage: intended for the collection and transport of clinical specimens containing viruses, from the collection site to the testing laboratory. I |
| Code Information |
Serial #: STM-UNV-0001-Q121, Lot #: JAH161, KAG076 |
| FEI Number |
3015989005
|
Recalling Firm/ Manufacturer |
Vectornate USA Inc 10 Industrial Ave Ste 4 Mahwah NJ 07430-2284
|
| For Additional Information Contact | Shreya Patil 201-393-2973 |
Manufacturer Reason for Recall | Discoloration and turbidity were present in the specified lots of STMs because of Longer Monsoon season at the South Korean Manufacturing Facility. |
FDA Determined Cause 2 | Process control |
| Action | The firm did not give out a public statement of a recall or recall strategy as it only affected one client and was specific to two identified lots. the firm did notify the customer of the issue via phone and email. |
| Quantity in Commerce | 130,000 devices |
| Distribution | US Nationwide distribution in the state of NJ. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|