| Date Initiated by Firm | March 25, 2021 |
| Create Date | June 01, 2021 |
| Recall Status1 |
Terminated 3 on March 20, 2023 |
| Recall Number | Z-1754-2021 |
| Recall Event ID |
87661 |
| Product Classification |
Cement, bone, vertebroplasty - Product Code NDN
|
| Product | InterV 10G Bone Access Tools/Kit, REF INTVM-FLNK. For use in Orthopedic / spinal procedures. |
| Code Information |
Lot Numbers: 11613ZV |
Recalling Firm/ Manufacturer |
Zavation 220 Lakeland Pkwy Flowood MS 39232-9552
|
| For Additional Information Contact | Quality Department 601-919-1119 |
Manufacturer Reason for Recall | Products distributed as sterile may not have been adequately sterilized |
FDA Determined Cause 2 | Vendor change control |
| Action | Zavation initially notified their consignees of the problem by email beginning on 03/24/2021 and requested they quarantine the product. The firm sent a subsequent email beginning on 04/13/2021 requesting the return of the product. The recall is being conducted to the user level. |
| Quantity in Commerce | 3 units |
| Distribution | US nationwide distribution. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|