Date Initiated by Firm | June 04, 2021 |
Create Date | August 06, 2021 |
Recall Status1 |
Completed |
Recall Number | Z-2110-2021 |
Recall Event ID |
88073 |
510(K)Number | K173084 |
Product Classification |
Pump, infusion - Product Code FRN
|
Product | Spectrum IQ Infusion System with Dose IQ Safety Software, Product Code 3570009. for the controlled administration of fluids. |
Code Information |
All Serial Numbers |
Recalling Firm/ Manufacturer |
Baxter Healthcare Corporation 1 Baxter Pkwy Deerfield IL 60015-4625
|
For Additional Information Contact | 800-422-9837 |
Manufacturer Reason for Recall | System errors were noted following changes the customers implemented to the configuration of their network and server systems. Gateway Server System performance anomalies resulted in excessive Spectrum IQ pump connectivity errors. |
FDA Determined Cause 2 | Software design |
Action | Baxter initiated the recall by letter on 06/04/2021. The letter described the problem and the hazard involved, and requested the consignee Baxter Technical Assistance in advance of making changes to your network or server infrastructure supporting the pumps or the Baxter Gateway, as indicated in Baxter's Gateway Server Installation Handoff document which customers receive upon installation.
The firm is developing a software patch for all Spectrum IQ infusion pumps. |
Quantity in Commerce | 130135 units |
Distribution | United States (including Puerto Rico), Canada, Bahamas, Bermuda, Guyana, and Trinidad and Tobago. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
|
510(K) Database | 510(K)s with Product Code = FRN
|