| Date Initiated by Firm | July 20, 2021 |
| Create Date | September 08, 2021 |
| Recall Status1 |
Terminated 3 on March 15, 2023 |
| Recall Number | Z-2437-2021 |
| Recall Event ID |
88444 |
| 510(K)Number | K162449 |
| Product Classification |
System, ablation, microwave and accessories - Product Code NEY
|
| Product | Solero Generator US PG, Item No. H78712740000US0, Catalog No. 12740000US |
| Code Information |
Serial Nos. QBY0002693
QBY0002376
QBY0002765
QBY0002380
QBY0002763
QBY0002662
QBY0002570
QBY0002683
QBY0002551
QBY0002618
QBY0002006
QBY0002293
QBY0002578
QBY0002666
QBY0002378
QBY0002375
QBY0002283
QBY0002539
QBY0002762
QBY0002681
QBY0002690
QBY0002389
QBY0002518
QBY0002670
QBY0002395
QBY0002687
QBY0002613
QBY0002664
QBY0002759
QBY0002780
|
Recalling Firm/ Manufacturer |
Angiodynamics, Inc. 603 Queensbury Ave Queensbury NY 12804-7619
|
| For Additional Information Contact | Saleem Cheeks 518-795-1174 |
Manufacturer Reason for Recall | Specific serial numbers of the Solero MTA Generator require servicing to upgrade the software to help reduce the incidence of Error 0001, which can occur during system start-up. |
FDA Determined Cause 2 | Software design |
| Action | On July 20, 2021, the firm initiated the recall by issuing Urgent Voluntary Medical Device Recall letters to affected consignees.
Customers were informed of the software upgrade. Customers were asked to return their generators to AngioDynamics for service. The firm's Hardware Service Department will contact you to schedule the upgrade, which may include the shipment of a replacement generator to your facility if needed to be used during the upgrade of your devices.
In the interim, while this upgrade is being scheduled, you may continue to use the Solero generator and are instructed to adhere to the following Warning in the current product labeling (User Manual):
WARNING: Do not initiate the procedure/anesthesia until the applicator has been connected, primed, and the generator status bar indicates "Ready". |
| Quantity in Commerce | 26 US, 4 OUS |
| Distribution | Domestic distribution Nationwide. Foreign distribution worldwide. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = NEY
|