| | Class 2 Device Recall Siemens |  |
| Date Initiated by Firm | August 30, 2021 |
| Create Date | November 24, 2021 |
| Recall Status1 |
Terminated 3 on January 16, 2024 |
| Recall Number | Z-0289-2022 |
| Recall Event ID |
88888 |
| 510(K)Number | K190578 |
| Product Classification |
System, x-ray, tomography, computed - Product Code JAK
|
| Product | SOMATOM Edge Plus with software syngo.CT VB20
Model #10267000 |
| Code Information |
Serial Numbers:
Serial Number
122022
122098
122062
122099
122121
122146
122054
122082
122080
122020
122164
122116
122008
122036
122029
122044
122172
122074
122075
122049
122141
122167
122156
122081
122126
122129
122090
122130
122089
122085
122037
122009
122163
122064
122038
122161
122168
122158
122144
122063
122086
122035
122041
122114
122107
122101
122108
122109
122162
122125
122011
122052
122034
122057
122145
122147
122106
122061
122051
122166
122170
122092
122004
|
| FEI Number |
2240869
|
Recalling Firm/ Manufacturer |
Siemens Medical Solutions USA, Inc 40 Liberty Blvd Malvern PA 19355-1418
|
| For Additional Information Contact | SAME 610-219-4834 |
Manufacturer Reason for Recall | software syngo.CT VB20 in the installed base, with or without syngo.CT software versions VB20_SP1, SP2, SP3, or SP4, may result in workflow interruptions
resulting in the delay in diagnosis, the necessity for patient rescans, and/or the need for additional contrast media may occur |
FDA Determined Cause 2 | Software design |
| Action | Siemens Medical Solutions USA, Inc. initiated a Customer Safety Advisory Notice (CSAN) to customers via CT021/21/S on 8/30/2021. The CSAN informs the customer of the issues health risk, the new software update version VB20_SP5 to eliminate the root causes of these problems. The software update will be distributed free of charge via CT020/21/S and CT071/21/S (update is dependent on the current software version) to all affected customers.
Action to take:
Siemens Healthineers has developed software update syngo.CT VB20 SP5. This new software contains workflow improvements and solutions to the previously described software issues. Please ask your local service organization if you are interested in more detailed information about the content of the software update.
The new software, VB20_SP5, will be provided free of charge and will be distributed via Update CT020/21/S or CT071/20/S. Update CT071/20/S is only necessary if the system is being updated from software version VB20, VB20_SP1, or VB20_SP2 directly to VB20_SP5. Following the corrective action process CAPA, the cause of identified
product issues has been eliminated and a recurrence of the identified issues is prevented.
This update will be performed remotely. The remote software updates process will require approximately 45 minutes for completion and is completely automatic; however, please make sure the system and power are stable before and during the process, please do not switch off the system during the update process. Siemens highly recommends
starting the installation when the scanner will not be in use or when the necessary time for the update to be completed can be scheduled.
If you have any questions, please contact our service organization at at 1-800-888-7436.
Please review, understand, and confirm acknowledgement of the information and instructions outlined in this
Urgent Safety Advisory Notice. By signing this document, you are acknowledging that you have read and understand the content therein |
| Quantity in Commerce | 64 units |
| Distribution | US Nationwide distribution. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = JAK
|
|
|
|