| | Class 2 Device Recall Compella |  |
| Date Initiated by Firm | February 14, 2022 |
| Create Date | March 23, 2022 |
| Recall Status1 |
Terminated 3 on July 02, 2026 |
| Recall Number | Z-0820-2022 |
| Recall Event ID |
89644 |
| Product Classification |
Bed, flotation therapy, powered - Product Code IOQ
|
| Product | Compella Bariatric Bed System |
| Code Information |
All lots/serial numbers
UDI information:
Compella Bariatric Bed - 00887761000902
|
| FEI Number |
1417572
|
Recalling Firm/ Manufacturer |
Baxter Healthcare Corporation
1 Baxter Pkwy
Deerfield
IL
60015-4625
|
Manufacturer Reason for Recall | Potential damage to the power cord may cause the power conductors to be exposed, thereby creating a risk of shock from the power mains. |
FDA Determined Cause 2 | Labeling design |
| Action | Customer notification letters were mailed beginning 02/21/2022.
A localized customer letter will be mailed (with signature service) to all first-level consignees by Sedgwick. The letter will inform the customer of the issue, reason for Field Action and provide risk mitigations. The customer will be instructed to sign acknowledgement form to confirm receipt and return to a dedicated email address.
Distributors will be asked to forward to their end users. Distributors will be provided with instructions on communication with their end users.
All consignees will be required to return a signed acknowledgment form and confirm that they read and understand the issue.
Distributors will be asked to forward to their end users. Distributors will be provided instruction on communication with their end users.
|
| Quantity in Commerce | 5186 units |
| Distribution | Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahamas, Bahrain, Belgium, Bermuda, Brazil, Cameroon, Canada, Estonia, Finland, France, Georgia, Germany, Greece, Hungary, India, Ireland, Italy, Jamaica, Jordan, Kuwait, Mexico, Netherlands, New Zealand, Norway, Oman, Poland, Qatar, Romania, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, Thailand, Turkmenistan, UAE, and UK. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
|
|
|