| | Class 1 Device Recall Engstrom Carestation, Engstrm PRO, CARESCAPE R860 |  |
| Date Initiated by Firm | April 18, 2022 |
| Date Posted | June 06, 2022 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1227-2022 |
| Recall Event ID |
90076 |
| 510(K)Number | K111116 K142679 |
| Product Classification |
Ventilator, continuous, facility use - Product Code CBK
|
| Product | Replacement back up batteries distributed on or after April 1, 2019, for CARESCAPE R860, Engstrom Carestation and Engstrom PRO Ventilators. |
| Code Information |
Field Replacement Unit kits PN: 1009-5682-000-S (BTRY SEALED LEAD ACID RECHARGEABLE 12V) and PN: 5856787-S (BTRY SEALED LEAD ACID RECHARGEABLE 12V PAIR).
|
Recalling Firm/ Manufacturer |
GE Healthcare, LLC 3000 N Grandview Blvd Waukesha WI 53188-1615
|
| For Additional Information Contact | GE Healthcare Service 1800-437-1171 |
Manufacturer Reason for Recall | Insufficient battery backup power resulting in premature shutdown of the ventilator when not connected to AC mains power supply. |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | The firm, GE Healthcare, sent an "URGENT MEDICAL DEVICE CORRECTION" letter the consignees via mail on 04/18/2022. The letter describes the product, problem and actions to be taken. The customers instructed to take the following actions:
1. You can continue to use the affected ventilators while the ventilator is connected
to an AC mains power source that is supported by backup emergency power.
2. If it is absolutely necessary to use the ventilator by relying on the battery (such
as during required transport where there are no other options), ensure you follow
standard clinical practice of having a readily accessible means of appropriate
alternative ventilation (for example a bag-valve system) and personnel with the
capability to administer this alternative means at all times.
3. Immediately after receiving this communication, perform the Battery
Performance Test as described in Appendix A. Replace the batteries when
necessary, before patient use.
4. When not in patient use, it is recommended that the device always remains
connected to the AC mains power source to prevent battery discharge and
degradation, even when in storage.
5. It is recommended that the Battery Performance Test be completed every
three months as described in Appendix A.
6. If the device has been in storage for over three months, perform the Battery
Performance Test as described in Appendix A prior to use.
7. The backup batteries must be replaced at a minimum every three years.
8. Complete the attached Medical Device Notification Acknowledgement Response
form and send to FMI34126RC.BATTERY@ge.co
If you have any questions or concerns regarding this notification, please contact
GE Healthcare Service at 1-800-437-1171 or your local Service Representative. |
| Quantity in Commerce | 88,379 batteries |
| Distribution | Worldwide Distribution: US (nationwide) to AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI and OUS (internationally) to Australia, Austria, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, China, Columbia, Cote D'Ivoire, Mexico, Uruguay, United Kingdom, United Arab Emirates, Turkey, Thailand, Tanzania, Sweden, Spain, Singapore, Saudi Arabia, Russia, Romania, Portugal, Poland, Philippines, Panama, Omar, Norway, Nigeria, Nicaragua, New Zealand, Netherlands, Malaysia, Latvia, Kuwait, Korea, Kenya, Kazakhstan, Japan, Italy, Ireland, Indonesia, India, Hungary, Germany, France, Finland, Estonia, Egypt, Ecuador, Denmark, Czech Republic. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = CBK 510(K)s with Product Code = CBK
|
|
|
|