Date Initiated by Firm | June 09, 2022 |
Create Date | July 20, 2022 |
Recall Status1 |
Terminated 3 on February 08, 2024 |
Recall Number | Z-1422-2022 |
Recall Event ID |
90471 |
510(K)Number | K002585 |
Product Classification |
Transferrin, antigen, antiserum, control - Product Code DDG
|
Product | Soluble Transferrin Receptors (STFR), Catalog Number TF10159 (Research Use Only). IVD |
Code Information |
UDI-DI (GTIN): 05055273215564 All lots within expiry |
Recalling Firm/ Manufacturer |
Randox Laboratories Ltd. 55 Diamond Road Crumlin (North) Ireland
|
Manufacturer Reason for Recall | A customer reported a problem with the reagent and calibrator. Randox Laboratories carried out an investigation and confirmed the issue. The complaint is pertaining to atypical calibration curves that can results in non-numerical results. This affects Siemens Atellica CH analysers only. Customer have reported an atypical calibration curve that results in the values around the low QC range to produce a non-numerical result. It appears to be related to the bottle values of the calibrator producing a curve shape that is too linear for the logitlog calculation to calculate. This issue can only be identified in the Atellica graphical view of the calibration. In most of the cases when this has occurred, it is picked up by the QC generating a Math Error, however, there are reports of QC shifts after calibration without the Math Error flagged that may be related to this issue. Results reported with STFR Level 1, with a concentration less than 1.7mg/L, using TF10159 and TF10161 can be underestimated by up to 30%. This issue has been reported with reagent lots 575475 and 562691, calibrator lots 537725 and 562696. No other complaints received. |
FDA Determined Cause 2 | Under Investigation by firm |
Action | On June 9, 2022, the firm notified customers of the recall via email. The issue only affects Siemens Atellica CH analyzer users. No action is required if the assay is being run on other platforms.
The issue will be resolved in Atellica Solution Software (SW) version 1.25.3. Until all Atellica CH analyzers in the laboratory are updated to SW 1.25.3, the assay can be run by Open Channel. Please contact applications@randox.com for an Atellica CH Open Channel application sheet. If the assay is not run by Open Channel prior to installation of SW version 1.25.3, the assay must be re-calibrated once 1.25.3 is installed.
Customers were instructed to review results generated with the affected batches in line with the clinical profile of the patient. |
Quantity in Commerce | 12 kits (US) |
Distribution | US Nationwide distribution in the state of IL. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = DDG
|