• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 1 Device Recall CADD Disposables

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 1 Device Recall CADD Disposablessee related information
Date Initiated by FirmDecember 09, 2022
Date PostedJanuary 30, 2023
Recall Status1 Open3, Classified
Recall NumberZ-0966-2023
Recall Event ID 91392
Product Classification Set, administration, intravascular - Product Code FPA
ProductCADD Administration Set, List Numbers: 1) REF 21-7300-24, 2) REF 21-7301-24, 3) REF 21-7302-24, 4) REF 21-7308-24, 5) REF 21-7309-24, 6) REF 21-7310-24, 7) REF 21-7322-24, 8) REF 21-7323-24, 9) REF 21-7324-24, 10) REF 21-7333-24, 11) REF 21-7343-24, 12) REF 21-7346-24, 13) REF 21-7349-24, 14) REF 21-7359-24, 15) REF 21-7363-24, 16) REF 21-7383-24, 17) REF 21-7391-24, 18) REF 21-7394-24, 19) REF 21-7600-24, 20) REF 21-7609-24, 21) REF 21-7624-24, 22) REF 21-7649-24, 23) REF 21-7301-24JP, 24) REF 21-7302-24JP, 25) REF 21-7308-24JP, 26) REF 21-7309-24JP, 27) REF 21-7322-24JP, 28) REF 21-7359-24JP, 29) REF 21-7394-24JP, 30) REF 21-7600-24JP, 31) REF 21-7609-24JP, 32) REF 21-7624-24JP, 33) REF 21-7649-24JP. Computerized Ambulatory Delivery Device
Code Information 1) REF 21-7300-24, UDI/DI 30610586027196, Lot Number: 3630772 through 4334107; 2) REF 21-7301-24, UDI/DI 30610586027219, Lot numbers: 3630747 through 4334119; 3) REF 21-7302-24, UDI/DI 30610586027233, Lot numbers: 3617363 through 4334138; 4) REF 21-7308-24, UDI/DI 30610586032374, Lot numbers: 4053922 through 4334076; 5) REF 21-7309-24, UDI/DU 30610586032381, Lot numbers: 4062405 though 4334085; 6) REF 21-7310-24, UDI/DU 30610586032398, Lot numbers: 4062404 through 4330874; 7) REF 21-7322-24, UDI/DI 30610586029640, Lot numbers: 3776375 through 4334324; 8) REF 21-7323-24, UDI/DI 30610586039618, Lot numbers: 3776373 through 4334325; 9) REF 21-7324-24, UDI/DI 30610586029657, Lot numbers: 3773527 through 4334328; 10) REF 21-7333-24, UDI/DI 30610586043233, Lot numbers: 3776362 through 3984144; 11) REF 21-7343-24, UDI/DI 35019517191824, Lot numbers: 3965344 through 4334338; 12) REF 21-7346-24, UDI/DI 35019517161957, Lot numbers: 3776356 through 4320785; 13) REF 21-7349-24, UDI/DI 35019517184239, Lot numbers: 3926579 through 4339160; 14) REF 21-7359-24, UDI/DI 30610586027400, Lot numbers: 3776315 through 4334339; 15) REF 21-7363-24, UDI/DI 35019517161964, Lot numbers: 3773412 through 4334343; 16) REF 21-7383-24, UDI/DI 30610586027479, Lot numbers: 3780549 through 3971523; 17) REF 21-7391-24, UDI/DI 30610586027516, Lot numbers: 3773276 through 4334351; 18) REF 21-7394-24, UDI/DI 30610586027530, Lot number: 3774739 through 4339178; 19) REF 21-7600-24, UDI/DI 30610586044001, Lot numbers: 4084914 through 4329633; 20) REF 21-7609-24, UDI/DI 30610586044018, Lot Numbers: 4072200 through 4334090; 21) REF 21-7624-24, UDI/DI 30610586044742, Lot numbers: 4092506 through 4309481; 22) REF 21-7649-24, UDI/DI 35019517184246, Lot numbers: 4076410 through 4334357. 23) REF 21-7301-24JP, UDI/DI N/A, Lot Numbers: 3630747 through 4334113; 24) REF 21-7302-24JP, UDI/DI N/A, Lot Numbers: 3617363 through 4329630; 25) REF 21-7308-24JP, UDI/DI N/A, Lot Numbers: 4053922 through 4334075; 26) REF 21-7309-24JP, UDI/DI N/A, Lot Numbers: 4062405 through 4330870; 27) REF 21-7322-24JP, UDI/DI N/A, Lot Numbers: 3776375 through 4334324; 28) REF 21-7359-24JP, UDI/DI N/A, Lot Numbers: 3776315 through 4334339; 29) REF 21-7394-24JP, UDI/DI N/A, Lot Numbers: 3774739 through 4339178; 30) REF 21-7600-24JP, UDI/DI N/A, Lot Numbers: 4084914 through 4329633; 31) REF 21-7609-24JP, UDI/DI N/A, Lot Numbers: 4072200 through 4334090; 32) REF 21-7624-24JP, UDI/DI N/A, Lot Numbers: 4092506 through 4309481; 33) REF 21-7649-24JP, UDI/DI N/A, Lot Numbers: 4076410 through 4334357.
FEI Number 3012307300
Recalling Firm/
Manufacturer
Smiths Medical ASD Inc.
6000 Nathan Ln N
Minneapolis MN 55442-1690
For Additional Information ContactSedgwick
866-562-5931
Manufacturer Reason
for Recall
There are two potential issues with CADD Infusion System Infusion Sets: (1) Lack of Delivery or Underdelivery related to Tubing Occlusion and (2) False "No Disposable Attached (NDA)" Alarms
FDA Determined
Cause 2
Under Investigation by firm
ActionThe firm issued an URGENT MEDICAL DEVICE CORRECTION notice on 12/09/2022 by mail. Actions for Clinicians and Patients: . The pump may not adequately detect the cassette before or during an infusion due to this issue, and an alarm will be triggered. If a pump displays an NDA alarm, the user can try to resolve the alarm by repositioning the CADD Medication Cassette Reservoir while connected to the pump, repositioning the reservoir by disconnecting from the pump and reattaching it to the pump, or replacing the reservoir. . Or the user can remove the reservoir from the pump and push the plastic ridge highlighted in the circle towards the arch on the reservoir. . For patients requiring life sustaining medications, we recommend use of alternative CADD infusion sets. Contact pharmacy to discuss availability of alternative CADD infusion sets. Due to availability and specific patient situations, clinicians may consider switching patients to an alternative pump. . If the user cannot resolve the NDA alarm, replace the cassette reservoir, though the issues may recur if that product is also affected by this recall. . Clinicians, share this letter with your homecare patients and educate them about the actions above. Complete and return the Response Form via fax at 866-338-4443 or email to smithsmedical5286@sedgwick.com. . DISTRIBUTORS: Immediately forward this notice to your customers them. Have them contact Sedgwick 866-562-5931(M-F, 8am-5pm ET) to obtain a response form. Global Complaint Management, globalcomplaints@smiths-medical.com 1-(866)-216-8806, Report adverse events or product complaints: Technical Assistance (800)-258-5361 . Pharmacists: . Immediately identify affected products in your possession. Separate and label them to avoid use for life sustaining medications. . For use with life sustaining medications, contact Smiths Medical customer service (800)-258-5361 for info on obtaining alternative CADD infusion sets. On 06/12/2023, another notification
Quantity in Commerce47595495 units
DistributionWorldwide Distribution: US (nationwide) and OUS countries of: CH, CA, ES, FR, AT, BE, IT, DK, NO, EE, DE, NL, AU, GB, SE, PL, SA, IN, IL, PT, CZ, IE, NZ, LU, SI, ZA, BM, SG, MQ, CL, LB, MY, HU, JP, TW, and CN
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
-
-