| | Class 2 Device Recall CENTURION MEDICAL PRODUCTS Surgical convenience kits |  |
| Date Initiated by Firm | December 07, 2022 |
| Date Posted | February 10, 2023 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1127-2023 |
| Recall Event ID |
91483 |
| Product Classification |
Circumcision tray - Product Code OHG
|
| Product | Surgical convenience kits labeled as:
CENTURION MEDICAL PRODUCTS CIRCUMCISION TRAY, Reorder Number CIT3345 |
| Code Information |
Reorder Number CIT3345, GTIN 40653160102374, Lot # 22HMH649
|
| FEI Number |
1417592
|
Recalling Firm/ Manufacturer |
MEDLINE INDUSTRIES, LP - Northfield
3 Lakes Dr
Northfield
IL
60093-2753
|
| For Additional Information Contact | Karin Johnson 886-359-1704 |
Manufacturer Reason for Recall | The recall is due to the perforated bag not allowing for outside of the kit wrapping to be sterile, therefore, only the contents inside the kits wrapping are sterile. |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | The firm sent a correction notice to its direct consignees on 12/07/2022 by US mail and email. The notice explained the issue and requested the following:
"Please check you stock for affected kits and over-labeling will be sent to place on kits to ensure that all labeling is accurate. No kits are to be returned back to Medline. All components are not affected. Our records indicate that these were used as a component in the list of kit(s) purchased from Medline Industries, LP."
"If you are a distributor, or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Have the customers return any affected product to you for collection and return to Medline Industries, LP."
"If you have any questions, please contact the Recall Department at 866-359-1704." |
| Quantity in Commerce | 144 units |
| Distribution | Nationwide |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|