| | Class 1 Device Recall Cardiosave |  |
| Date Initiated by Firm | February 07, 2023 |
| Date Posted | March 09, 2023 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1147-2023 |
| Recall Event ID |
91603 |
| 510(K)Number | K181122 |
| Product Classification |
System, balloon, intra-aortic and control - Product Code DSP
|
| Product | Cardiosave Hybrid
Model Nos. 0998-00-0800-31
0998-00-0800-32
0998-00-0800-33
0998-00-0800-34
0998-00-0800-35
0998-00-0800-45
0998-00-0800-52
0998-00-0800-53
0998-00-0800-55
0998-00-0800-65
0998-UC-0800-31*
0998-UC-0800-33*
0998-UC-0800-52*
0998-UC-0800-53*
0998-00-0800-55*
* -- not distributed in US |
| Code Information |
Model UDI
0998-00-0800-31 10607567109053
0998-00-0800-32 10607567111117
0998-00-0800-33 10607567109008
0998-00-0800-34 10607567111940
0998-00-0800-35 10607567109107
0998-00-0800-45 10607567108421
0998-00-0800-52 10607567108438
0998-00-0800-53 10607567108391
0998-00-0800-55 10607567108414
0998-00-0800-65 10607567113432
0998-UC-0800-31* N/A
0998-UC-0800-33* N/A
0998-UC-0800-52* N/A
0998-UC-0800-53* N/A
0998-00-0800-55* N/A
All serial numbers |
| FEI Number |
3001418283
|
Recalling Firm/ Manufacturer |
Datascope Corp. 1300 Macarthur Blvd Mahwah NJ 07430-2052
|
| For Additional Information Contact | Ms. Allison Jean Kaplan 973-709-7779 |
Manufacturer Reason for Recall | An unexpected shutdown of the IABP may occur due to loss of communication between the Executive Processor PCBA and the Video Generator PCBA. This may result in an interruption to therapy which may threaten the hemodynamic stability of the supported patient as the user is left unaware to the status of the Cardiosave IABP. |
FDA Determined Cause 2 | Software design |
| Action | On February 7, 2023, the firm notified customers of 4 issues affecting the IABP via an urgent Medical Device Correction letter, which included the present issue.
Users were instructed to do the following:
1. If you experience an unexpected shutdown of Cardiosave IABP during therapy, or a frozen or black screen, use another IABP to continue therapy. Until an alternative IABP is located you may attempt to restart the IABP. If the IABP remains non-operational, immediately remove from the patient care environment for further product evaluation.
2. If your device remains inoperable, please contact your service representative to identify the cause and take the necessary actions required.
The firm is developing a software correction to address the issue. Once available, a Datascope/Getinge service representative will contact you to schedule the installation of the updated software. This work will be done at no cost to your facility.
If you have any questions, please contact your Datascope/Getinge representative or call the Datascope/Getinge Technical Support at 1-888-943-8872, options 4, 2, 1, Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone). |
| Quantity in Commerce | 4793 (US); 4407 (OUS) |
| Distribution | Worldwide Distribution: US (nationwide) including states of: AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, NV, OH, OK, OR, PA, SC, SD, TN, TX, VA, WA, WI, and WV; and OUS (Foreign) countries of: ALBANIA, ALGERIA, ANGOLA, ARGENTINA, AUSTRALIA, AZERBAIJAN, BAHRAIN, BANGLADESH, BARBADOS, BELRUS, BELGIUM, BOSNIA AND HERZEGOVINA, BRAZIL, BRUNEI, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, C¿te d'Ivoire, CROATIA, CZECH REPUBLIC, ECUADOR, EGYPT, EL SALVADOR, ESTONIA, FINLAND, FRANCE,GERMANY,GHANA,GUATEMALA,HONG KONG,HUNGARY,ICELAND,INDIA,INDONESIA
IRAN, IRAQ, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KENYA, KOSOVO, KUWAIT, LATVIA, LEBANON, LIBYA, LITHUANIA, LUXEMBOURG, MACEDONIA, MALAYSIA, MEXICO, MOROCCO,
MYANMAR, NETHERLANDS, NEW ZEALAND, NIGERIA, NORWAY, OMAN, PAKISTAN, PALESTINE, PERU, PHILIPPINES, POLAND, PORTUGAL, QATAR, RUSSIA, SAUDI ARABIA,
SERBIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SRI LANKA, SWEDEN, SWITZERLAND, SYRIA, TAIWAN, TANZANIA, THAILAND, TRINIDAD AND TOBAGO
TUNISIA, TURKEY, TURKMENISTAN, UKRAINE, UNITED ARAB EMIRATES, UNITED KINGDOM
UZBEKISTAN, VIETNAM, YEMEN AND ZAMBIA.
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = DSP
|
|
|
|