Date Initiated by Firm | April 05, 2023 |
Date Posted | May 25, 2023 |
Recall Status1 |
Open3, Classified |
Recall Number | Z-1632-2023 |
Recall Event ID |
92179 |
Product Classification |
Calibrator, multi-analyte mixture - Product Code JLX
|
Product | Atellica CH Toxicology Calibrator (TOX CAL)-The Atellica CH Toxicology Calibrator (TOX CAL) is for in vitro diagnostic use in calibrating the Acet, ETOH and Sal assays using the Atellica CH Analyzer
Siemens Material Number (SMN): 11099440 |
Code Information |
UDi-DI: 00630414597614
Lot Numbers:
74285306 Exp. Date: 2023-05-31;
74536672 Exp Date: 2024-03-31;
74732386 Exp. Date: 2024-11-30
|
Recalling Firm/ Manufacturer |
Siemens Healthcare Diagnostics, Inc. 511 Benedict Ave Tarrytown NY 10591-5005
|
For Additional Information Contact | SAME 914-631-8000 |
Manufacturer Reason for Recall | Reassignment of the Atellica CH Toxicology Calibrators (TOX CAL) and ADVIA Chemistry ToxAmmonia Calibrators for the Atellica CH Salicylate (Sal) Assay due to a positive bias |
FDA Determined Cause 2 | Under Investigation by firm |
Action | Siemens issued Urgent Medical Device Corrections (UMDC) ACHC23-03.A.US and ACHC23-
03.A.US.CHC to US customers via FedEx beginning on 04/05/2023. An Urgent Field Safety Notices (UFSN) ACHC23-03.A.OUS and ACHC23-03.A.OUS.CHC issued to OUS Siemens reporting sites with affected customers, for distribution and regional reporting to their local competent authorities beginning on 04/05/2023.
Letter states reason for recall, health risk and action to take:
Please review this letter with your Medical Director to determine the appropriate course of
action, including for any previously generated results, if applicable.
Until the updated lot-specific value sheets are available on Document Library, keep a copy
of this letter as a reference for the updated salicylate calibrator values.
Perform the instructions provided in Additional Information section below.
Complete and return the Field Correction Effectiveness Check Form attached to this letter
within 30 days.
If you have received any complaints of illness or adverse events associated with the
products listed in Table 1, immediately contact your local Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative. |
Quantity in Commerce | 6546 units US; 3681 units OUS |
Distribution | US Nationwide and Worldwide Distribution
Foreign:
Argentina
Australia
Austria¿¿¿¿¿¿¿¿
Bahrain¿¿¿¿¿¿¿¿
Bangladesh
Belgium¿¿¿¿¿¿¿¿
Bosnia¿Herzeg.¿
Brazil
Canada
Chile
China
Colombia
Czech¿Republic¿
Egypt¿¿¿¿¿¿¿¿¿¿
Estonia¿¿¿¿¿¿¿¿
Finland¿¿¿¿¿¿¿¿
France¿¿¿¿¿¿¿¿¿
Georgia¿¿¿¿¿¿¿¿
Germany¿¿¿¿¿¿¿¿
Greece¿¿¿¿¿¿¿¿¿
Hong Kong
Hungary¿¿¿¿¿¿¿¿
India
Israel¿¿¿¿¿¿¿¿¿
Italy¿¿¿¿¿¿¿¿¿¿
Latvia¿¿¿¿¿¿¿¿¿
Malaysia
Mexico
Netherlands¿¿¿¿
Norway¿¿¿¿¿¿¿¿¿
Pakistan¿¿¿¿¿¿¿
Poland¿¿¿¿¿¿¿¿¿
Portugal¿¿¿¿¿¿¿
Republic of Korea
Saudi¿Arabia¿¿¿
Serbia¿¿¿¿¿¿¿¿¿
Slovakia¿¿¿¿¿¿¿
South Africa
Spain¿¿¿¿¿¿¿¿¿¿
Sweden¿¿¿¿¿¿¿¿¿
Taiwan
Turkey¿¿¿¿¿¿¿¿¿
United¿Kingdom¿
Uruguay
Vietnam
|
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|