• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 1 Device Recall medtronic

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 1 Device Recall medtronicsee related information
Date Initiated by FirmMay 10, 2023
Date PostedJune 16, 2023
Recall Status1 Open3, Classified
Recall NumberZ-1803-2023
Recall Event ID 92328
PMA NumberP980016S551 
Product Classification Implantable cardioverter defibrillator (non-CRT) - Product Code LWS
ProductICD-VR DVFB1D1 VISIA AF MRI US DF1, Model Number DVFB1D1; Implantable Cardioverter Defibrillators
Code Information GTIN 00643169837799, Lot Serial Numbers: CWG208686H, CWG211264H, CWG205930H, CWG208117H, CWG210956H, CWG210958H, CWG211814H, CWG212178H, CWG208135H, CWG209489H, CWG209599H, CWG212702H, CWG214349H, CWG214379H, CWG206584H, CWG208347H, CWG206776H, CWG206895H, CWG211478H, CWG211709H, CWG212079H, CWG212183H, CWG212836H, CWG205283H, CWG205284H, CWG205288H, CWG207805H, CWG207808H, CWG207809H, CWG207810H, CWG207811H, CWG207812H, CWG208949H, CWG209030H, CWG209031H, CWG209285H, CWG209646H, CWG209659H, CWG209823H, CWG210928H, CWG210929H, CWG211866H, CWG212840H, CWG212841H, CWG212842H, CWG212843H, CWG212844H, CWG212846H, CWG212847H, CWG212851H, CWG212853H, CWG212854H, CWG213620H, CWG213622H, CWG213623H, CWG206242H, CWG206868H, CWG209799H, CWG212914H, CWG205298H, CWG205299H, CWG205450H, CWG205777H, CWG206838H, CWG206972H, CWG208335H, CWG208502H, CWG209280H, CWG209978H, CWG209979H, CWG209981H, CWG209982H, CWG209983H, CWG211070H, CWG211071H, CWG211072H, CWG211696H, CWG212566H, CWG212568H, CWG212569H, CWG213285H, CWG213505H, CWG213756H, CWG214277H, CWG214324H, CWG204936H, CWG205368H, CWG206168H, CWG208188H, CWG208467H, CWG213778H, CWG206589H, CWG206593H, CWG209988H, CWG210659H, CWG213726H, CWG213727H, CWG213728H, CWG208549H, CWG213157H, CWG206078H, CWG206763H, CWG207072H, CWG207829H, CWG207979H, CWG209086H, CWG209271H, CWG209763H, CWG211506H, CWG213628H, CWG214299H, CWG206015H, CWG206115H, CWG206347H, CWG206398H, CWG206864H, CWG207060H, CWG207230H, CWG208431H, CWG208825H, CWG208869H, CWG209569H, CWG212666H, CWG207012H, CWG209624H, CWG205839H, CWG206775H, CWG208030H, CWG208084H, CWG208184H, CWG208744H, CWG209781H, CWG210362H, CWG210390H, CWG211194H, CWG211301H, CWG211817H, CWG212533H, CWG212552H, CWG214390H, CWG207167H, CWG204969H, CWG206871H, CWG209061H, CWG210696H, CWG211027H, CWG211028H, CWG208153H, CWG207295H, CWG207567H, CWG207715H, CWG213540H, CWG208792H, CWG214152H, CWG211554H, CWG212531H, CWG206039H, CWG206385H, CWG212687H, CWG213369H, CWG205033H, CWG205702H, CWG207326H, CWG207802H, CWG209619H, CWG211314H, CWG213372H, CWG205592H, CWG205593H, CWG206639H, CWG207720H, CWG208487H, CWG209060H, CWG209579H, CWG209695H, CWG211030H, CWG212172H, CWG212518H, CWG213735H, CWG213737H, CWG213798H, CWG203850H, CWG205238H, CWG205289H, CWG205375H, CWG205802H, CWG205924H, CWG206181H, CWG206353H, CWG206360H, CWG206383H, CWG206427H, CWG206521H, CWG206556H, CWG206696H, CWG206934H, CWG207015H, CWG207020H, CWG207021H, CWG207137H, CWG207506H, CWG207838H, CWG207988H, CWG208087H, CWG208350H, CWG208360H, CWG208499H, CWG208670H, CWG208802H, CWG208882H, CWG209143H, CWG209386H, CWG209415H, CWG209739H, CWG209766H, CWG210141H, CWG210159H, CWG210350H, CWG210555H, CWG210656H, CWG211125H, CWG211134H, CWG211606H, CWG211780H, CWG212686H, CWG212810H, CWG212881H, CWG214000H, CWG205772H, CWG210406H, CWG211159H, CWG205371H, CWG212802H, CWG206888H, CWG208887H, CWG210637H, CWG210851H, CWG211578H, CWG211587H, CWG212907H, CWG205150H, CWG206204H, CWG212634H, CWG213780H, CWG206723H, CWG206724H, CWG207939H, CWG208937H, CWG208938H, CWG208939H, CWG210954H, CWG210955H, CWG213846H, CWG213928H, CWG214069H, CWG214323H, CWG214346H, CWG207278H, CWG211826H, CWG211919H, CWG205679H, CWG207239H, CWG207615H, CWG207767H, CWG208897H, CWG211455H, CWG209893H, CWG210576H, CWG211210H, CWG207914H, CWG207916H, CWG211024H, CWG210897H, CWG205031H, CWG205177H, CWG205359H, CWG205360H, CWG205361H, CWG205363H, CWG205365H, CWG205366H, CWG206657H, CWG206659H, CWG206660H, CWG207622H, CWG207623H, CWG207625H, CWG207626H, CWG207627H, CWG207629H, CWG207630H, CWG207631H, CWG207632H, CWG207633H, CWG207634H, CWG207635H, CWG207641H, CWG207672H, CWG208917H, CWG208919H, CWG208921H, CWG208922H, CWG208923H, CWG208924H, CWG208925H, CWG208926H, CWG210013H, CWG210014H, CWG210016H, CWG210017H, CWG210018H, CWG210019H, CWG210020H, CWG210022H, CWG210023H, CWG210026H, CWG210028H, CWG210030H, CWG210657H, CWG210704H, CWG210706H, CWG210707H, CWG210711H, CWG210712H, CWG210713H, CWG210714H, CWG210715H, CWG210717H, CWG210718H, CWG211486H, CWG212317H, CWG212319H, CWG212320H, CWG212321H, CWG212323H, CWG212324H, CWG212327H, CWG212328H, CWG213590H, CWG213591H, CWG213595H, CWG213596H, CWG213597H, CWG213598H, CWG213599H, CWG206240H, CWG210249H, CWG211774H, CWG212262H, CWG212450H, CWG207150H, CWG208118H, CWG211245H, CWG211340H, CWG205082H, CWG206462H, CWG209359H, CWG209360H, CWG206917H, CWG207301H, CWG208083H, CWG208197H, CWG208873H, CWG209147H, CWG209952H, CWG210188H, CWG211445H, CWG212188H, CWG212660H, CWG213352H, CWG213492H, CWG206275H, CWG206643H, CWG210785H, CWG207054H, CWG208616H, CWG209218H, CWG209543H, CWG210191H, CWG210459H, CWG210465H, CWG210998H, CWG211195H, CWG211297H, CWG212070H, CWG214001H, CWG214018H, CWG214132H, CWG209305H, CWG213050H, CWG204892H, CWG205197H, CWG206374H, CWG213699H, CWG213820H, CWG207962H, CWG209481H, CWG210314H, CWG211272H, CWG211545H, CWG214022H, CWG209727H, CWG206009H, CWG206893H, CWG211193H, CWG213049H, CWG213137H, CWG213669H, CWG213986H, CWG214145H, CWG214148H, CWG207101H, CWG208652H, CWG210384H, CWG211622H, CWG212697H, CWG205101H, CWG206506H, CWG208150H, CWG211630H, CWG213306H, CWG208803H, CWG209255H, CWG209376H, CWG213769H, CWG214004H, CWG205435H, CWG207272H, CWG209085H, CWG211391H, CWG212650H, CWG205801H, CWG211268H, CWG211221H, CWG213796H, CWG211124H, CWG213938H, CWG214118H, CWG208228H, CWG209265H, CWG209954H, CWG211399H, CWG211574H, CWG213177H, CWG206571H, CWG207852H, CWG210984H, CWG211322H, CWG211481H, CWG213895H, CWG210225H, CWG211523H, CWG207571H, CWG210464H, CWG211800H, CWG205513H, CWG205551H, CWG206081H, CWG206333H, CWG206822H, CWG208106H, CWG208438H, CWG211501H, CWG211605H, CWG212191H, CWG212527H, CWG206064H, CWG207716H, CWG210746H, CWG213902H, CWG205201H, CWG206450H, CWG211424H, CWG212469H, CWG205494H, CWG208338H, CWG208841H, CWG208933H, CWG209471H, CWG209884H, CWG210412H, CWG210426H, CWG210445H, CWG210557H, CWG211177H, CWG212816H, CWG213178H, CWG213531H, CWG207169H, CWG209220H, CWG211242H, CWG212140H, CWG206229H, CWG207401H, CWG209141H, CWG209503H, CWG213517H, CWG209136H, CWG205173H, CWG205833H, CWG206012H, CWG206829H, CWG209272H, CWG211250H, CWG213275H, CWG213308H, CWG213662H, CWG208061H, CWG208661H, CWG206711H, CWG206712H, CWG207702H, CWG207721H, CWG207882H, CWG212565H, CWG213663H, CWG213664H, CWG205971H, CWG209810H, CWG211823H, CWG204930H, CWG208757H, CWG208758H, CWG208760H, CWG211331H, CWG211976H, CWG211980H, CWG212581H, CWG212582H, CWG214344H, CWG214345H, CWG205369H, CWG206035H, CWG206172H, CWG207303H, CWG211553H, CWG211737H, CWG205738H, CWG207083H, CWG207264H, CWG207266H, CWG208650H, CWG208944H, CWG209180H, CWG210917H, CWG211857H, CWG212507H, CWG212562H, CWG213718H, CWG205700H, CWG205710H, CWG206832H, CWG206836H, CWG206837H, CWG207745H, CWG209012H, CWG209013H, CWG209014H, CWG212498H, CWG212415H, CWG205997H, CWG210515H, CWG211144H, CWG211145H, CWG211146H, CWG211557H, CWG212357H, CWG212358H, CWG206995H, CWG207535H, CWG209128H, CWG209729H, CWG211219H, CWG211770H, CWG213018H, CWG204931H, CWG204932H, CWG204934H, CWG205001H, CWG205625H, CWG205627H, CWG206523H, CWG206591H, CWG206994H, CWG207959H, CWG208325H, CWG208951H, CWG209440H, CWG210126H, CWG210127H, CWG210532H, CWG211259H, CWG211260H, CWG211791H, CWG211891H, CWG213158H, CWG213898H, CWG214097H, CWG214178H, CWG214260H, CWG214343H, CWG214447H, CWG209050H, CWG207181H, CWG211913H, CWG212868H, CWG204905H, CWG205290H, CWG205451H, CWG209736H, CWG209901H, CWG210556H, CWG214227H, CWG212590H, CWG207830H, CWG213876H, CWG205497H, CWG208739H, CWG210154H, CWG210469H, CWG211318H, CWG211397H, CWG212039H, CWG212437H, CWG213278H, CWG214342H, CWG206670H, CWG206672H, CWG209192H, CWG210046H, CWG211715H, CWG212334H, CWG212335H, CWG213440H, CWG213441H, CWG204944H, CWG205276H, CWG205583H, CWG205584H, CWG205590H, CWG205596H, CWG205599H, CWG205600H, CWG205687H, CWG206569H, CWG206644H, CWG206956H, CWG207097H, CWG207186H, CWG207188H, CWG207204H, CWG207236H, CWG207246H, CWG207284H, CWG207469H, CWG208443H, CWG209482H, CWG209506H, CWG209507H, CWG209509H, CWG209510H, CWG209511H, CWG209588H, CWG209625H, CWG209778H, CWG212483H, CWG212484H, CWG212485H, CWG212486H, CWG212487H, CWG212488H, CWG212493H, CWG212494H, CWG207042H, CWG209449H, CWG211834H, CWG212805H, CWG212811H, CWG207096H, CWG210742H, CWG212189H, CWG208465H, CWG213968H, CWG205749H, CWG207220H, CWG209900H, CWG211431H, CWG214414H, CWG214415H, CWG206543H, CWG211717H, CWG212617H, CWG206131H, CWG208352H, CWG208673H, CWG209571H, CWG209926H, CWG211726H, CWG211840H, CWG212621H, CWG213128H, CWG214166H, CWG209319H, CWG211429H, CWG205077H, CWG206230H, CWG209052H, CWG210433H, CWG205304H, CWG207844H, CWG212057H, CWG212758H, CWG213339H, CWG213458H, CWG208535H, CWG206379H, CWG206435H, CWG208134H, CWG207104H, CWG214253H, CWG207048H, CWG207466H, CWG209660H, CWG214303H, CWG209347H, CWG211851H, CWG213382H, CWG205019H, CWG205729H, CWG207667H, CWG210527H, CWG211216H, CWG211430H, CWG211796H, CWG212120H, CWG205757H, CWG211761H, CWG206202H, CWG206928H, CWG209010H, CWG211450H, CWG212396H, CWG206106H, CWG205642H, CWG207156H, CWG208169H, CWG209067H, CWG210240H, CWG211760H, CWG212945H, CWG208482H, CWG210745H, CWG211535H, CWG212020H, CWG213085H, CWG205472H, CWG205560H, CWG205753H, CWG208656H, CWG209115H, CWG209677H, CWG209877H, CWG209992H, CWG210121H, CWG210402H, CWG210505H, CWG210612H, CWG212027H, CWG205656H, CWG209199H, CWG209108H, CWG205510H, CWG206235H, CWG206953H, CWG207109H, CWG209603H, CWG210262H, CWG211638H, CWG211934H, CWG212593H, CWG206545H, CWG206068H, CWG214157H, CWG206350H, CWG205837H, CWG207299H, CWG207360H, CWG207918H, CWG208014H, CWG209617H, CWG209948H, CWG210644H, CWG211333H, CWG211365H, CWG211784H, CWG212355H, CWG212926H, CWG213421H, CWG207034H, CWG208285H, CWG212713H, CWG209377H, CWG207073H, CWG212093H, CWG213036H, CWG206447H, CWG206967H, CWG206974H, CWG210220H, CWG210323H, CWG210544H, CWG211094H, CWG211619H, CWG213734H, CWG213887H, CWG214179H, CWG214254H, CWG214399H, CWG214400H, CWG214445H, CWG214503H, CWG214036H, CWG205905H, CWG207919H, CWG211882H, CWG210279H, CWG211902H, CWG212009H, CWG214095H, CWG212060H, CWG208323H, CWG209404H, CWG213186H, CWG208321H, CWG213097H, CWG205644H, CWG205678H, CWG209916H, CWG211970H, CWG206924H, CWG208076H, CWG210495H, CWG212128H, CWG213837H, CWG205919H, CWG208864H, CWG209527H, CWG210250H, CWG210572H, CWG214073H, CWG214183H, CWG210399H, CWG211531H, CWG212934H, CWG213347H, CWG213945H, CWG207184H, CWG210798H, CWG213027H, CWG207127H, CWG208520H, CWG212790H, CWG212003H, CWG205165H, CWG205715H, CWG207589H, CWG211226H, CWG212102H, CWG213056H, CWG205016H, CWG205105H, CWG205115H, CWG205159H, CWG205444H, CWG205445H, CWG205479H, CWG206119H, CWG206145H, CWG206228H, CWG206253H, CWG206261H, CWG206271H, CWG206475H, CWG206527H, CWG206626H, CWG206904H, CWG207242H, CWG207444H, CWG207468H, CWG207511H, CWG207512H, CWG207592H, CWG207740H, CWG207806H, CWG207817H, CWG207893H, CWG208039H, CWG208112H, CWG208272H, CWG208678H, CWG208988H, CWG208989H, CWG209104H, CWG209222H, CWG209232H, CWG209244H, CWG209362H, CWG209501H, CWG209647H, CWG209792H, CWG209990H, CWG210163H, CWG210289H, CWG210599H, CWG210610H, CWG210611H, CWG210686H, CWG210842H, CWG211118H, CWG211156H, CWG211179H, CWG211547H, CWG211640H, CWG211687H, CWG211778H, CWG211785H, CWG212025H, CWG212026H, CWG212315H, CWG212316H, CWG212578H, CWG212579H, CWG212736H, CWG212773H, CWG212782H, CWG212922H, CWG212931H, CWG213013H, CWG213042H, CWG213172H, CWG213197H, CWG213313H, CWG213503H, CWG213854H, CWG213973H, CWG214025H, CWG214026H, CWG214079H, CWG214228H, CWG214328H, CWG214351H, CWG214371H, CWG214419H, CWG205292H, CWG211248H, CWG212165H, CWG205229H, CWG205310H, CWG205763H, CWG206234H, CWG206935H, CWG207051H, CWG207069H, CWG207417H, CWG210502H, CWG211447H, CWG212230H, CWG212700H, CWG213800H, CWG205009H, CWG210896H, CWG211140H, CWG211422H, CWG212768H, CWG213323H, CWG214127H, CWG205213H, CWG207593H, CWG212661H, CWG211986H, CWG204945H, CWG207017H, CWG207158H, CWG207267H, CWG207323H, CWG207971H, CWG208419H, CWG209478H, CWG211610H, CWG211747H, CWG211977H, CWG211978H, CWG212192H, CWG213247H, CWG213388H, CWG205260H, CWG212971H, CWG205760H, CWG207244H, CWG207355H, CWG208109H, CWG214382H, CWG205236H, CWG205987H, CWG207411H, CWG208609H, CWG209413H, CWG209460H, CWG210692H, CWG211663H, CWG213185H, CWG213343H, CWG213400H, CWG214232H, CWG211163H, CWG212440H, CWG213661H, CWG205470H, CWG205570H, CWG206423H, CWG207357H, CWG207368H, CWG207397H, CWG207953H, CWG208858H, CWG208902H, CWG208904H, CWG209430H, CWG210422H, CWG210451H, CWG211222H, CWG211223H, CWG211711H, CWG212167H, CWG212647H, CWG213150H, CWG205418H, CWG205419H, CWG205421H, CWG205423H, CWG206243H, CWG208282H, CWG208584H, CWG208585H, CWG208963H, CWG208964H, CWG208965H, CWG210369H, CWG210371H, CWG210841H, CWG210844H, CWG210882H, CWG210883H, CWG210884H, CWG211004H, CWG211006H, CWG211007H, CWG211008H, CWG211009H, CWG211011H, CWG211014H, CWG211016H, CWG211765H, CWG211833H, CWG212495H, CWG212729H, CWG212756H, CWG213011H, CWG213022H, CWG211139H, CWG206270H, CWG207525H, CWG207942H, CWG211571H, CWG208932H, CWG205910H, CWG210638H, CWG213750H, CWG205524H, CWG206074H, CWG206222H, CWG207613H, CWG211522H, CWG212540H, CWG212696H, CWG208451H, CWG211975H, CWG212350H, CWG212979H, CWG210144H, CWG211202H, CWG212209H, CWG204890H, CWG209833H, CWG205643H, CWG205944H, CWG206394H, CWG206646H, CWG208200H, CWG209790H, CWG210302H, CWG210560H, CWG210775H, CWG211488H, CWG211990H, CWG212675H, CWG213129H, CWG208601H, CWG206500H, CWG207159H, CWG207977H, CWG209417H, CWG211321H, CWG212241H, CWG205090H, CWG204900H, CWG204901H, CWG204904H, CWG204906H, CWG204908H, CWG204909H, CWG204910H, CWG204911H, CWG205356H, CWG205386H, CWG205387H, CWG205388H, CWG205389H, CWG206515H, CWG206516H, CWG206518H, CWG206519H, CWG206520H, CWG207934H, CWG207935H, CWG207936H, CWG207937H, CWG207938H, CWG208542H, CWG208544H, CWG208545H, CWG208547H, CWG208548H, CWG208773H, CWG208774H, CWG210630H, CWG210631H, CWG210632H, CWG210634H, CWG210635H, CWG210636H, CWG212282H, CWG212283H, CWG212284H, CWG212285H, CWG212286H, CWG212287H, CWG212288H, CWG212304H, CWG212306H, CWG212307H, CWG212308H, CWG212309H, CWG212311H, CWG212314H, CWG214221H, CWG214333H, CWG214348H, CWG214360H, CWG214376H, CWG205157H, CWG207978H, CWG208731H, CWG209205H, CWG209513H, CWG210357H, CWG210588H, CWG212893H, CWG213822H, CWG206154H, CWG208330H, CWG208483H, CWG209020H, CWG209056H, CWG211473H, CWG214492H, CWG206315H, CWG209412H, CWG212064H, CWG205900H, CWG210580H, CWG211274H, CWG214006H, CWG208385H, CWG210857H, CWG207248H, CWG209365H, CWG210667H, CWG211157H, CWG206719H, CWG207528H, CWG208480H, CWG208667H, CWG210048H, CWG210233H, CWG205912H, CWG214121H, CWG205569H, CWG207422H, CWG208388H, CWG211010H, CWG211458H, CWG213789H, CWG214159H, CWG213230H, CWG205648H, CWG205649H, CWG207174H, CWG207189H, CWG208156H, CWG208231H, CWG208318H, CWG208838H, CWG208970H, CWG209975H, CWG209976H, CWG209977H, CWG211379H, CWG212796H, CWG213426H, CWG205038H, CWG205089H, CWG205269H, CWG205529H, CWG207233H, CWG207263H, CWG209306H, CWG209514H, CWG211262H, CWG211942H, CWG207404H, CWG207901H, CWG210122H, CWG211767H, CWG211973H, CWG212158H, CWG212977H, CWG212991H, CWG214053H, CWG214146H, CWG214321H, CWG205228H, CWG210873H, CWG210927H, CWG211371H, CWG212909H, CWG213282H, CWG213412H, CWG213777H, CWG213906H, CWG213926H, CWG214409H, CWG205187H, CWG205266H, CWG208378H, CWG209178H, CWG212923H, CWG214063H, CWG214483H, CWG209848H, CWG210007H, CWG210434H, CWG210757H, CWG211822H, CWG211903H, CWG212911H, CWG206960H, CWG207160H, CWG212127H, CWG214249H, CWG214368H, CWG205704H, CWG208082H, CWG208681H, CWG209932H, CWG212663H, CWG213037H, CWG213179H, CWG213513H, CWG213717H, CWG213298H, CWG210430H, CWG212910H, CWG207403H, CWG206354H, CWG212908H, CWG214286H, CWG210720H, CWG206839H, CWG210088H, CWG212747H, CWG213489H, CWG214150H, CWG205571H, CWG208095H, CWG213405H, CWG208581H, CWG208942H, CWG209096H, CWG211706H, CWG212210H, CWG212559H, CWG212746H, CWG214122H, CWG214252H, CWG210916H, CWG208664H, CWG209744H, CWG206510H, CWG207392H, CWG209573H, CWG209743H, CWG210400H, CWG211702H, CWG212546H, CWG213316H, CWG206307H, CWG206376H, CWG208263H, CWG211423H, CWG211649H, CWG211984H, CWG212777H, CWG212817H, CWG213984H, CWG206948H, CWG205246H, CWG209155H, CWG210252H, CWG211665H, CWG213478H, CWG205026H, CWG205449H, CWG205452H, CWG207092H, CWG207093H, CWG208405H, CWG214171H, CWG206768H, CWG213934H, CWG205463H, CWG205464H, CWG205582H, CWG206393H, CWG206782H, CWG207281H, CWG207676H, CWG207677H, CWG207678H, CWG207679H, CWG207680H, CWG207681H, CWG207719H, CWG207850H, CWG208977H, CWG208978H, CWG208979H, CWG208980H, CWG208981H, CWG208982H, CWG208984H, CWG209683H, CWG210075H, CWG210246H, CWG210876H, CWG210880H, CWG211724H, CWG212464H, CWG213823H, CWG205437H, CWG208457H, CWG208907H, CWG209431H, CWG210436H, CWG210490H, CWG210589H, CWG210815H, CWG210830H, CWG211910H, CWG212014H, CWG212606H, CWG212951H, CWG213062H, CWG213076H, CWG213191H, CWG213987H, CWG214052H, CWG209442H, CWG211267H, CWG211983H, CWG212234H, CWG212964H, CWG212651H, CWG213617H, CWG205694H, CWG205758H, CWG205771H, CWG208380H, CWG209676H, CWG209837H, CWG211133H, CWG211779H, CWG212053H, CWG212259H, CWG212918H, CWG213122H, CWG213417H, CWG205256H, CWG205998H, CWG205999H, CWG206000H, CWG206158H, CWG206160H, CWG206162H, CWG206163H, CWG206358H, CWG208479H, CWG208854H, CWG208855H, CWG209196H, CWG209206H, CWG209237H, CWG209444H, CWG209445H, CWG209446H, CWG209448H, CWG209867H, CWG209967H, CWG210564H, CWG210565H, CWG210890H, CWG210892H, CWG211756H, CWG211762H, CWG211945H, CWG210293H, CWG209416H, CWG209636H, CWG209913H, CWG210212H, CWG210313H, CWG211716H, CWG213390H, CWG206254H, CWG206616H, CWG210990H, CWG211099H, CWG211208H, CWG214214H, CWG214465H, CWG205842H, CWG206486H, CWG207081H, CWG207407H, CWG207965H, CWG207966H, CWG209062H, CWG209063H, CWG209534H, CWG209696H, CWG209905H, CWG209935H, CWG209980H, CWG210152H, CWG210153H, CWG210340H, CWG211032H, CWG211033H, CWG211801H, CWG205906H, CWG208312H, CWG208411H, CWG210646H, CWG213345H, CWG213637H, CWG214266H, CWG212021H, CWG213941H, CWG212695H, CWG214084H, CWG209135H, CWG208178H, CWG209925H, CWG212019H, CWG213304H, CWG213368H, CWG213094H, CWG213116H, CWG213130H, CWG213135H, CWG213156H, CWG213516H, CWG213767H, CWG214131H, CWG214498H, CWG214499H, CWG210624H, CWG206306H, CWG207216H, CWG209113H, CWG205060H, CWG205137H, CWG205169H, CWG205192H, CWG206929H, CWG207218H, CWG208975H, CWG208976H, CWG209019H, CWG209023H, CWG211520H, CWG212174H, CWG212223H, CWG213470H, CWG213471H, CWG213473H, CWG213475H, CWG205734H, CWG206429H, CWG206815H, CWG208122H, CWG208260H, CWG209512H, CWG210253H, CWG213335H, CWG212975H, CWG214137H, CWG205949H, CWG209077H, CWG211299H, CWG211324H, CWG212398H, CWG212942H, CWG213393H, CWG214164H, CWG214184H, CWG214384H, CWG210115H, CWG206147H, CWG206758H, CWG210199H, CWG210546H, CWG212101H, CWG213214H, CWG214051H, CWG205471H, CWG205581H, CWG205936H, CWG206652H, CWG206823H, CWG207105H, CWG207771H, CWG208121H, CWG208638H, CWG208953H, CWG210592H, CWG213365H, CWG214209H, CWG206552H, CWG206968H, CWG208433H, CWG208572H, CWG209409H, CWG210472H, CWG205043H, CWG205111H, CWG205352H, CWG205984H, CWG206130H, CWG206890H, CWG207061H, CWG208000H, CWG208284H, CWG208406H, CWG209615H, CWG210025H, CWG211154H, CWG211292H, CWG211353H, CWG211493H, CWG211502H, CWG211542H, CWG212121H, CWG212657H, CWG213105H, CWG213749H, CWG214058H, CWG212883H, CWG207730H, CWG206617H, CWG206675H, CWG209502H, CWG210055H, CWG210377H, CWG210625H, CWG210626H, CWG210627H, CWG210628H, CWG213602H, CWG214210H, CWG213792H, CWG213962H, CWG206378H, CWG208849H, CWG207583H, CWG208015H, CWG209598H, CWG211294H, CWG212705H, CWG213561H, CWG213776H, CWG205443H, CWG205804H, CWG206344H, CWG209962H, CWG210583H, CWG210622H, CWG211283H, CWG211335H, CWG206678H, CWG207684H, CWG207688H, CWG207694H, CWG207695H, CWG209011H, CWG209887H, CWG206468H, CWG212032H, CWG214199H, CWG207000H, CWG207041H, CWG207889H, CWG208409H, CWG208961H, CWG209197H, CWG209229H, CWG210330H, CWG211428H, CWG212218H, CWG212960H, CWG213310H, CWG213864H, CWG214160H, CWG214173H, CWG214229H, CWG214417H, CWG205403H, CWG205799H, CWG206827H, CWG209291H, CWG212708H, CWG212709H, CWG213924H, CWG205207H, CWG206021H, CWG213091H, CWG207544H, CWG205838H, CWG205657H, CWG210483H, CWG205568H, CWG206203H, CWG206348H, CWG207814H, CWG207815H, CWG208680H, CWG208993H, CWG208996H, CWG208997H, CWG210899H, CWG211310H, CWG211748H, CWG211755H, CWG211788H, CWG212292H, CWG212293H, CWG213487H, CWG213488H, CWG210804H, CWG212891H, CWG205689H, CWG205826H, CWG207963H, CWG207964H, CWG208074H, CWG209429H, CWG210810H, CWG210811H, CWG211021H, CWG211022H, CWG213738H, CWG213739H, CWG206482H, CWG208042H, CWG209230H, CWG211328H, CWG212490H, CWG212672H, CWG206777H, CWG206900H, CWG211252H, CWG211376H, CWG212072H, CWG203848H, CWG206916H, CWG208842H, CWG205754H, CWG206288H, CWG206975H, CWG207151H, CWG208114H, CWG209369H, CWG209777H, CWG213217H, CWG205844H, CWG205845H, CWG210541H, CWG210623H, CWG214258H, CWG205901H, CWG207071H, CWG207290H, CWG213041H, CWG206438H, CWG211757H, CWG207223H, CWG207698H, CWG207699H, CWG207700H, CWG207701H, CWG209243H, CWG210306H, CWG210359H, CWG210807H, CWG211035H, CWG212529H, CWG212530H, CWG213741H, CWG213742H, CWG208415H, CWG208503H, CWG209987H, CWG210453H, CWG211813H, CWG212423H, CWG212570H, CWG212639H, CWG213350H, CWG213845H, CWG213857H, CWG213977H, CWG214478H, CWG212656H, CWG213874H, CWG206781H, CWG208064H, CWG209133H, CWG210263H, CWG210834H, CWG205897H, CWG206460H, CWG206882H, CWG206883H, CWG208240H, CWG205756H, CWG211093H, CWG205376H, CWG205377H, CWG205378H, CWG206835H, CWG206857H, CWG211947H, CWG211953H, CWG211957H, CWG207112H, CWG209110H, CWG209339H, CWG213108H, CWG213329H, CWG213946H, CWG208889H, CWG209630H, CWG209723H, CWG212574H, CWG214162H, CWG209674H, CWG209964H, CWG211387H, CWG212640H, CWG212107H, CWG211581H, CWG213914H, CWG206570H, CWG207025H, CWG205727H, CWG208660H, CWG209488H, CWG212721H, CWG213909H, CWG214274H, CWG209574H, CWG205305H, CWG206265H, CWG206317H, CWG207891H, CWG209223H, CWG209283H, CWG209423H, CWG210772H, CWG211041H, CWG211808H, CWG213821H, CWG210005H, CWG213009H, CWG204982H, CWG205927H, CWG206415H, CWG206416H, CWG206478H, CWG206535H, CWG206969H, CWG208235H, CWG209772H, CWG210691H, CWG211586H, CWG211654H, CWG213601H, CWG213817H, CWG205112H, CWG205538H, CWG209699H, CWG210256H, CWG213935H, CWG214110H, CWG205522H, CWG205523H, CWG206661H, CWG206663H, CWG206664H, CWG206665H, CWG207697H, CWG209002H, CWG209003H, CWG209005H, CWG209006H, CWG209007H, CWG209008H, CWG209508H, CWG209813H, CWG210593H, CWG210594H, CWG210595H, CWG210602H, CWG211832H, CWG211845H, CWG212264H, CWG212265H, CWG212266H, CWG212561H, CWG212808H, CWG212812H, CWG212822H, CWG212850H, CWG211249H, CWG212658H, CWG213149H, CWG206685H, CWG211171H, CWG205868H, CWG205870H, CWG206272H, CWG206821H, CWG208862H, CWG209081H, CWG209623H, CWG209897H, CWG210207H, CWG211967H, CWG213371H, CWG207743H, CWG208120H, CWG208396H, CWG210245H, CWG212534H, CWG213045H, CWG204985H, CWG204986H, CWG204987H, CWG205712H, CWG206530H, CWG208818H, CWG210296H, CWG210679H, CWG211074H, CWG211969H, CWG213816H, CWG214080H, CWG205233H, CWG208481H, CWG209124H, CWG212204H, CWG213804H, CWG204970H, CWG210494H, CWG210920H, CWG205325H, CWG205527H, CWG205528H, CWG206339H, CWG206343H, CWG206808H, CWG206809H, CWG206817H, CWG206818H, CWG206819H, CWG207226H, CWG207257H, CWG207559H, CWG207832H, CWG207833H, CWG207834H, CWG207835H, CWG207836H, CWG207839H, CWG209111H, CWG209295H, CWG209859H, CWG209861H, CWG209862H, CWG209864H, CWG210688H, CWG210738H, CWG210741H, CWG213542H, CWG213543H, CWG209949H, CWG210875H, CWG211589H, CWG213334H, CWG204907H, CWG205139H, CWG205143H, CWG205195H, CWG205200H, CWG205498H, CWG205785H, CWG206052H, CWG206212H, CWG206396H, CWG206489H, CWG206613H, CWG206899H, CWG206922H, CWG206923H, CWG206926H, CWG207014H, CWG207228H, CWG207566H, CWG207619H, CWG207820H, CWG207821H, CWG207981H, CWG208435H, CWG209385H, CWG209424H, CWG209657H, CWG209664H, CWG209731H, CWG210530H, CWG210771H, CWG211015H, CWG211220H, CWG211352H, CWG211789H, CWG211804H, CWG212154H, CWG212969H, CWG213389H, CWG205877H, CWG206469H, CWG206473H, CWG210049H, CWG212701H, CWG206913H, CWG209769H, CWG211968H, CWG205129H, CWG205311H, CWG207489H, CWG214096H, CWG214437H, CWG208189H, CWG210189H, CWG212774H, CWG210753H, CWG213508H, CWG205087H, CWG208136H, CWG208361H, CWG208529H, CWG210431H, CWG213166H, CWG205562H, CWG206008H, CWG206109H, CWG207191H, CWG207396H, CWG207957H, CWG208001H, CWG208365H, CWG208399H, CWG208400H, CWG209642H, CWG210651H, CWG210831H, CWG213588H, CWG205214H, CWG207006H, CWG208247H, CWG211798H, CWG213671H, CWG205095H, CWG205199H, CWG205614H, CWG206036H, CWG206151H, CWG206177H, CWG206649H, CWG207089H, CWG207648H, CWG207796H, CWG207898H, CWG208633H, CWG208718H, CWG209026H, CWG209539H, CWG209941H, CWG209971H, CWG210244H, CWG210365H, CWG210393H, CWG210687H, CWG210939H, CWG210947H, CWG211117H, CWG212016H, CWG212103H, CWG212781H, CWG205577H, CWG204978H, CWG206698H, CWG207329H, CWG208181H, CWG208821H, CWG209089H, CWG209468H, CWG209751H, CWG210682H, CWG211256H, CWG206970H, CWG207433H, CWG211732H, CWG213207H, CWG213229H, CWG214072H, CWG212076H, CWG207013H, CWG207545H, CWG207549H, CWG205327H, CWG205742H, CWG205824H, CWG206406H, CWG206459H, CWG207033H, CWG207587H, CWG208223H, CWG208477H, CWG209882H, CWG209883H, CWG210374H, CWG211326H, CWG211677H, CWG212637H, CWG212856H, CWG212871H, CWG212872H, CWG213790H, CWG204984H, CWG205274H, CWG211480H, CWG205834H, CWG207180H, CWG211956H, CWG213578H, CWG205074H, CWG208027H, CWG208468H, CWG209342H, CWG211418H, CWG211609H, CWG212229H, CWG213028H, CWG206075H, CWG208132H, CWG210475H, CWG214091H, CWG205145H, CWG205174H, CWG208692H, CWG205006H, CWG207149H, CWG207195H, CWG205645H, CWG206129H, CWG208079H, CWG208175H, CWG209667H, CWG210056H, CWG211525H, CWG211997H, CWG212375H, CWG207225H, CWG208025H, CWG208081H, CWG209343H, CWG210394H, CWG210484H, CWG212455H, CWG212456H, CWG212457H, CWG212460H, CWG213955H, CWG207458H, CWG208577H, CWG209568H, CWG210510H, CWG211505H, CWG214187H, CWG205034H, CWG205131H, CWG205846H, CWG206019H, CWG206260H, CWG207132H, CWG209673H, CWG210392H, CWG208098H, CWG208252H, CWG209608H, CWG209930H, CWG211408H, CWG211621H, CWG212557H, CWG213974H, CWG214268H, CWG205787H, CWG206090H, CWG206405H, CWG206772H, CWG211682H, CWG214230H, CWG207194H, CWG211413H, CWG213023H, CWG213883H, CWG205934H, CWG207372H, CWG211676H, CWG214398H, CWG205222H, CWG205829H, CWG206083H, CWG207886H, CWG208648H, CWG210386H, CWG211588H, CWG212963H, CWG206120H, CWG206174H, CWG206980H, CWG210872H, CWG212041H, CWG212075H, CWG212789H, CWG213457H, CWG205127H, CWG206517H, CWG208088H, CWG211388H, CWG213183H, CWG213273H, CWG205843H, CWG206846H, CWG211153H, CWG213111H, CWG214100H, CWG214423H, CWG211794H, CWG209151H, CWG207340H, CWG212054H, CWG213881H, CWG205961H, CWG206622H, CWG207531H, CWG206441H, CWG206703H, CWG207770H, CWG207772H, CWG208256H, CWG208653H, CWG209595H, CWG210481H, CWG212332H, CWG213192H, CWG213437H, CWG214098H, CWG214147H, CWG206218H, CWG206816H, CWG208072H, CWG209234H, CWG209391H, CWG209438H, CWG210736H, CWG211325H, CWG211558H, CWG211603H, CWG211795H, CWG212879H, CWG213939H, CWG207803H, CWG208275H, CWG212588H, CWG208473H, CWG209307H, CWG210886H, CWG206906H, CWG208871H, CWG210989H, CWG211086H, CWG205235H, CWG205811H, CWG205812H, CWG209536H, CWG209791H, CWG210809H, CWG211475H, CWG211872H, CWG211884H, CWG213811H, CWG214014H, CWG208703H, CWG212472H, CWG213676H, CWG214248H, CWG205057H, CWG210787H, CWG206010H, CWG210372H, CWG214099H, CWG209167H, CWG210673H, CWG211568H, CWG213101H, CWG213243H, CWG214129H, CWG205013H, CWG205130H, CWG207731H, CWG213849H, CWG205663H, CWG211526H, CWG210901H, CWG205258H, CWG205848H, CWG205974H, CWG208776H, CWG208822H, CWG208823H, CWG206357H, CWG208407H, CWG209149H, CWG211441H, CWG206175H, CWG206507H, CWG207039H, CWG209314H, CWG210996H, CWG206366H, CWG207464H, CWG207872H, CWG207944H, CWG211044H, CWG211046H, CWG213683H, CWG214242H, CWG206274H, CWG206551H, CWG207507H, CWG207972H, CWG208437H, CWG208566H, CWG209259H, CWG209614H, CWG209869H, CWG211235H, CWG206461H, CWG209268H, CWG210919H, CWG212691H, CWG206979H, CWG209337H, CWG212392H, CWG212750H, CWG207842H, CWG208815H, CWG210410H, CWG210669H, CWG213958H, CWG208177H, CWG212114H, CWG213905H, CWG206421H, CWG210971H, CWG214285H, CWG206023H, CWG206282H, CWG207282H, CWG207425H, CWG212411H, CWG212500H, CWG212509H, CWG210457H, CWG214239H, CWG212126H, CWG214101H, CWG206689H, CWG208596H, CWG209239H, CWG210297H, CWG205147H, CWG205480H, CWG206266H, CWG206618H, CWG206910H, CWG207024H, CWG207390H, CWG208355H, CWG209357H, CWG211085H, CWG212085H, CWG212866H, CWG213442H, CWG205338H, CWG205339H, CWG208008H, CWG209279H, CWG210539H, CWG210780H, CWG211510H, CWG212944H, CWG205980H, CWG209103H, CWG209164H, CWG209261H, CWG211178H, CWG214493H, CWG207067H, CWG209485H, CWG205564H, CWG208066H, CWG213075H, CWG213673H, CWG206108H, CWG205454H, CWG205579H, CWG205623H, CWG206386H, CWG206824H, CWG207142H, CWG207491H, CWG208290H, CWG210197H, CWG210235H, CWG211524H, CWG211683H, CWG212553H, CWG213020H, CWG213608H, CWG214103H, CWG206006H, CWG208033H, CWG210792H, CWG212068H, CWG206568H, CWG206853H, CWG208877H, CWG209187H, CWG211410H, CWG211575H, CWG211994H, CWG212614H, CWG206440H, CWG206490H, CWG207563H, CWG209531H, CWG210744H, CWG210846H, CWG210874H, CWG211120H, CWG211291H, CWG211616H, CWG211839H, CWG206494H, CWG208190H, CWG210946H, CWG212804H, CWG212821H, CWG214250H, CWG214504H, CWG206247H, CWG209939H, CWG210173H, CWG210174H, CWG210175H, CWG210176H, CWG211097H, CWG211098H, CWG206220H, CWG207126H, CWG207441H, CWG209469H, CWG205809H, CWG206305H, CWG206577H, CWG208359H, CWG211143H, CWG212083H, CWG212084H, CWG213175H, CWG212993H, CWG205983H, CWG206559H, CWG209428H, CWG210803H, CWG211003H, CWG205176H, CWG205594H, CWG206854H, CWG206855H, CWG206858H, CWG207744H, CWG207747H, CWG207930H, CWG207932H, CWG209730H, CWG211005H, CWG211138H, CWG211396H, CWG212180H, CWG212348H, CWG212502H, CWG212503H, CWG212845H, CWG213828H, CWG206005H, CWG208286H, CWG210404H, CWG210460H, CWG211281H, CWG211484H, CWG212835H, CWG213851H, CWG211645H, CWG205790H, CWG206986H, CWG207095H, CWG207146H, CWG207455H, CWG208023H, CWG208900H, CWG209461H, CWG209669H, CWG211536H, CWG212877H, CWG213139H, CWG213993H, CWG214381H, CWG209796H, CWG207044H, CWG205966H, CWG206221H, CWG208351H, CWG210185H, CWG210419H, CWG211377H, CWG207251H, CWG208884H, CWG209216H, CWG209961H, CWG212086H, CWG214237H, CWG205410H, CWG205411H, CWG206673H, CWG206674H, CWG207837H, CWG210653H, CWG208021H, CWG208070H, CWG208115H, CWG211056H, CWG211538H, CWG207008H, CWG207428H, CWG211451H, CWG211452H, CWG212982H, CWG211346H, CWG211695H, CWG206529H, CWG207037H, CWG207087H, CWG207221H, CWG207764H, CWG210029H, CWG211084H, CWG212447H, CWG214169H, CWG205181H, CWG208201H, CWG208268H, CWG208426H, CWG210378H, CWG210808H, CWG211060H, CWG212144H, CWG212913H, CWG213125H, CWG213141H, CWG213199H, CWG213228H, CWG214403H, CWG212722H, CWG208202H, CWG209908H, CWG211378H, CWG206091H, CWG209580H, CWG210409H, CWG211280H, CWG212236H, CWG213634H, CWG207154H, CWG209105H, CWG211151H, CWG213255H, CWG205847H, CWG214446H, CWG205068H, CWG205646H, CWG206718H, CWG208179H, CWG210418H, CWG212564H, CWG213252H, CWG207562H, CWG209745H, CWG210251H, CWG210543H, CWG211344H, CWG212749H, CWG205750H, CWG205954H, CWG207637H, CWG213376H, CWG213918H, CWG205344H, CWG206962H, CWG207778H, CWG209001H, CWG209750H, CWG211213H, CWG211374H, CWG212139H, CWG212921H, CWG214040H, CWG214261H, CWG208916H, CWG213687H, CWG213900H, CWG214247H, CWG206582H, CWG207611H, CWG208159H, CWG205709H, CWG205711H, CWG205821H, CWG206133H, CWG207200H, CWG207498H, CWG208097H, CWG208867H, CWG209541H, CWG210208H, CWG210941H, CWG212929H, CWG213103H, CWG209741H, CWG213500H, CWG214074H, CWG207436H, CWG207780H, CWG211693H, CWG204996H, CWG205106H, CWG205891H, CWG206246H, CWG207090H, CWG207664H, CWG207754H, CWG208314H, CWG208493H, CWG208719H, CWG209252H, CWG210182H, CWG210195H, CWG210216H, CWG210315H, CWG210351H, CWG210420H, CWG210985H, CWG212664H, CWG212779H, CWG212873H, CWG212876H, CWG213017H, CWG213913H, CWG214049H, CWG205832H, CWG207209H, CWG211563H, CWG205239H, CWG206425H, CWG209991H, CWG211499H, CWG212714H, CWG205010H, CWG205194H, CWG205555H, CWG207709H, CWG207712H, CWG208234H, CWG208551H, CWG209388H, CWG209589H, CWG210150H, CWG210151H, CWG211600H, CWG211750H, CWG212171H, CWG213004H, CWG205206H, CWG205621H, CWG206901H, CWG209724H, CWG212347H, CWG213302H, CWG212589H, CWG213068H, CWG213925H, CWG212005H, CWG213102H, CWG210877H, CWG211401H, CWG213801H, CWG206452H, CWG206694H, CWG207892H, CWG208340H, CWG209839H, CWG210855H, CWG211928H, CWG212059H, CWG212715H, CWG213078H, CWG213656H, CWG213863H, CWG214397H, CWG214436H, CWG205005H, CWG205071H, CWG207503H, CWG205392H, CWG206100H, CWG206912H, CWG207335H, CWG209029H, CWG209202H, CWG211705H, CWG211731H, CWG205576H, CWG212544H, CWG205348H, CWG205573H, CWG205669H, CWG206780H, CWG206925H, CWG206988H, CWG206989H, CWG207056H, CWG207561H, CWG207599H, CWG207707H, CWG207794H, CWG207795H, CWG207797H, CWG208749H, CWG208750H, CWG208753H, CWG208756H, CWG210922H, CWG210923H, CWG210924H, CWG210925H, CWG211471H, CWG211739H, CWG212616H, CWG212806H, CWG213826H, CWG207585H, CWG211449H, CWG205251H, CWG205988H, CWG206054H, CWG206096H, CWG206479H, CWG207283H, CWG207342H, CWG207348H, CWG207448H, CWG207449H, CWG207578H, CWG207933H, CWG208185H, CWG208262H, CWG208522H, CWG208538H, CWG208610H, CWG208811H, CWG209028H, CWG209112H, CWG209152H, CWG209331H, CWG210186H, CWG210218H, CWG210303H, CWG210467H, CWG210519H, CWG210520H, CWG210606H, CWG210608H, CWG210858H, CWG210859H, CWG210940H, CWG211166H, CWG211167H, CWG211168H, CWG211359H, CWG211382H, CWG211930H, CWG213460H, CWG213577H, CWG213593H, CWG213813H, CWG206118H, CWG207026H, CWG207331H, CWG210183H, CWG211000H, CWG211870H, CWG212899H, CWG213138H, CWG213567H, CWG213920H, CWG211718H, CWG211818H, CWG208444H, CWG208445H, CWG210910H, CWG214102H, CWG208375H, CWG210889H, CWG211561H, CWG211211H, CWG212274H, CWG205012H, CWG205776H, CWG206128H, CWG206217H, CWG207157H, CWG207253H, CWG207261H, CWG207801H, CWG207943H, CWG208304H, CWG208835H, CWG209080H, CWG209371H, CWG211375H, CWG211444H, CWG211964H, CWG212345H, CWG212439H, CWG212547H, CWG212772H, CWG212795H, CWG212938H, CWG212966H, CWG213165H, CWG213952H, CWG214134H, CWG214141H, CWG214307H, CWG214378H, CWG214404H, CWG214461H, CWG214486H, CWG214505H, CWG205685H, CWG206411H, CWG206813H, CWG207102H, CWG207410H, CWG207565H, CWG207569H, CWG208103H, CWG208404H, CWG208567H, CWG209132H, CWG210737H, CWG211656H, CWG212074H, CWG212648H, CWG212807H, CWG205508H, CWG207018H, CWG207293H, CWG209169H, CWG209456H, CWG209540H, CWG210082H, CWG210492H, CWG210643H, CWG211582H, CWG211927H, CWG212162H, CWG212537H, CWG212539H, CWG206213H, CWG207077H, CWG208875H, CWG211278H, CWG205148H, CWG206944H, CWG208301H, CWG209722H, CWG211320H, CWG205306H, CWG205574H, CWG206093H, CWG211583H, CWG213833H, CWG205681H, CWG207313H, CWG209121H, CWG210921H, CWG213674H, CWG205291H, CWG206940H, CWG208699H, CWG210633H, CWG213026H, CWG213493H, CWG207165H, CWG207332H, CWG209840H, CWG210863H, CWG212792H, CWG213224H, CWG214007H, CWG205721H, CWG206932H, CWG207866H, CWG209101H, CWG209818H, CWG210316H, CWG211776H, CWG211782H, CWG212865H, CWG213145H, CWG207398H, CWG207967H, CWG211674H, CWG211713H, CWG211843H, CWG204997H, CWG205208H, CWG209217H, CWG210881H, CWG211279H, CWG211999H, CWG206127H, CWG210678H, CWG211215H, CWG211753H, CWG211881H, CWG204928H, CWG204929H, CWG210164H, CWG212047H, CWG212887H, CWG205402H, CWG205429H, CWG205918H, CWG209340H, CWG210177H, CWG210405H, CWG211053H, CWG211054H, CWG211055H, CWG210462H, CWG212164H, CWG213055H, CWG205603H, CWG205607H, CWG205622H, CWG205629H, CWG207738H, CWG207781H, CWG207782H, CWG207783H, CWG207785H, CWG207787H, CWG207788H, CWG207789H, CWG207790H, CWG207791H, CWG207792H, CWG207793H, CWG211963H, CWG205049H, CWG205062H, CWG205707H, CWG207823H, CWG209158H, CWG209299H, CWG209544H, CWG210342H, CWG211771H, CWG212117H, CWG205056H, CWG206141H, CWG206532H, CWG207057H, CWG208065H, CWG208880H, CWG209091H, CWG210480H, CWG213153H, CWG213370H, CWG213626H, CWG213770H, CWG214050H, CWG212732H, CWG205788H, CWG205394H, CWG208496H, CWG209235H, CWG211598H, CWG213249H, CWG208382H, CWG213120H, CWG207084H, CWG207989H, CWG209289H, CWG209537H, CWG209928H, CWG212442H, CWG209207H, CWG205631H, CWG211212H, CWG213490H, CWG213890H, CWG213916H, CWG206211H, CWG208606H, CWG208850H, CWG209238H, CWG210227H, CWG211234H, CWG211690H, CWG213919H, CWG214016H, CWG205114H, CWG205827H, CWG206909H, CWG210086H, CWG211136H, CWG211174H, CWG212066H, CWG212182H, CWG212543H, CWG212679H, CWG213110H, CWG206609H, CWG208449H, CWG210397H, CWG210548H, CWG211119H, CWG211722H, CWG212669H, CWG213152H, CWG209320H, CWG207533H, CWG207994H, CWG205155H, CWG207725H, CWG212055H, CWG205025H, CWG208559H, CWG209453H, CWG213832H, CWG204993H, CWG206981H, CWG208420H, CWG211276H, CWG212163H, CWG206104H, CWG210255H, CWG210773H, CWG210906H, CWG213019H, CWG213119H, CWG205438H, CWG205963H, CWG206364H, CWG206800H, CWG206999H, CWG207510H, CWG207570H, CWG208046H, CWG208091H, CWG209269H, CWG209421H, CWG210373H, CWG210603H, CWG210829H, CWG211062H, CWG211232H, CWG211239H, CWG212703H, CWG213001H, CWG213148H, CWG213385H, CWG213502H, CWG213815H, CWG209815H, CWG210221H, CWG206560H, CWG207249H, CWG209650H, CWG212720H, CWG211207H, CWG205223H, CWG205224H, CWG206620H, CWG209759H, CWG209762H, CWG206954H, CWG207607H, CWG211461H, CWG212619H, CWG212980H, CWG207534H, CWG206038H, CWG206047H, CWG206150H, CWG206223H, CWG206224H, CWG206304H, CWG207905H, CWG207997H, CWG207998H, CWG208804H, CWG209575H, CWG209933H, CWG210052H, CWG210838H, CWG210944H, CWG211214H, CWG212024H, CWG212755H, CWG213114H, CWG213204H, CWG213764H, CWG213858H, CWG214009H, CWG214067H, CWG214411H, CWG213810H, CWG213812H, CWG213929H, CWG207685H, CWG208006H, CWG208047H, CWG209548H, CWG209767H, CWG210375H, CWG212984H, CWG213080H, CWG206605H, CWG206699H, CWG207969H, CWG208294H, CWG209137H, CWG209396H, CWG211981H, CWG213988H, CWG206226H, CWG206356H, CWG209168H, CWG212089H, CWG208863H, CWG205110H, CWG205530H, CWG205532H, CWG205534H, CWG205535H, CWG205536H, CWG206721H, CWG206751H, CWG206753H, CWG206754H, CWG206755H, CWG208453H, CWG208454H, CWG208455H, CWG208456H, CWG209686H, CWG209689H, CWG209690H, CWG209691H, CWG209692H, CWG209693H, CWG209697H, CWG209858H, CWG212132H, CWG212133H, CWG212269H, CWG212270H, CWG212272H, CWG212273H, CWG212520H, CWG212521H, CWG212522H, CWG212524H, CWG213328H, CWG213720H, CWG213723H, CWG207529H, CWG208412H, CWG208666H, CWG210789H, CWG211987H, CWG213205H, CWG213991H, CWG213992H, CWG214281H, CWG208569H, CWG209927H, CWG211105H, CWG212106H, CWG214168H, CWG205075H, CWG205473H, CWG205567H, CWG205585H, CWG205716H, CWG205774H, CWG206695H, CWG207010H, CWG207011H, CWG207043H, CWG207779H, CWG210034H, CWG210257H, CWG212649H, CWG213269H, CWG213355H, CWG204990H, CWG205542H, CWG205768H, CWG206608H, CWG206826H, CWG207007H, CWG207123H, CWG207234H, CWG209277H, CWG209403H, CWG209521H, CWG210196H, CWG210408H, CWG213326H, CWG213327H, CWG205404H, CWG206094H, CWG206290H, CWG206409H, CWG206627H, CWG206897H, CWG206990H, CWG207029H, CWG207353H, CWG208518H, CWG208659H, CWG209227H, CWG209716H, CWG209940H, CWG210237H, CWG210770H, CWG211191H, CWG211261H, CWG211489H, CWG212235H, CWG212263H, CWG212504H, CWG212906H, CWG213295H, CWG213296H, CWG213319H, CWG213860H, CWG213995H, CWG214028H, CWG214089H, CWG214325H, CWG214361H, CWG214372H, CWG205399H, CWG206178H, CWG206714H, CWG206814H, CWG207462H, CWG207975H, CWG209661H, CWG209775H, CWG211122H, CWG211584H, CWG212160H, CWG212377H, CWG207656H, CWG207996H, CWG208311H, CWG209492H, CWG210824H, CWG211541H, CWG212359H, CWG212361H, CWG212362H, CWG205008H, CWG205676H, CWG205677H, CWG205951H, CWG205952H, CWG207493H, CWG207494H, CWG207509H, CWG207573H, CWG207714H, CWG207818H, CWG207869H, CWG207961H, CWG208162H, CWG208163H, CWG209292H, CWG209515H, CWG209780H, CWG210248H, CWG210641H, CWG210862H, CWG210936H, CWG211020H, CWG211604H, CWG211836H, CWG212834H, CWG212947H, CWG213415H, CWG213431H, CWG213589H, CWG213818H, CWG206866H, CWG208199H, CWG209262H, CWG205053H, CWG205054H, CWG205381H, CWG205382H, CWG206773H, CWG207298H, CWG207690H, CWG207984H, CWG208883H, CWG209666H, CWG209707H, CWG210143H, CWG210348H, CWG212152H, CWG213358H, CWG213807H, CWG205448H, CWG205483H, CWG205485H, CWG205486H, CWG206484H, CWG206485H, CWG206732H, CWG206733H, CWG206734H, CWG206735H, CWG206740H, CWG206741H, CWG207746H, CWG209097H, CWG209752H, CWG209753H, CWG209754H, CWG209755H, CWG209756H, CWG209757H, CWG210202H, CWG210203H, CWG210205H, CWG212870H, CWG213366H, CWG213512H, CWG205820H, CWG207328H, CWG210345H, CWG205055H, CWG205146H, CWG205468H, CWG205701H, CWG206180H, CWG206538H, CWG207080H, CWG207120H, CWG207164H, CWG207542H, CWG207555H, CWG208295H, CWG210044H, CWG210311H, CWG210341H, CWG210658H, CWG210661H, CWG210797H, CWG210850H, CWG210934H, CWG211290H, CWG212225H, CWG212555H, CWG212627H, CWG212645H, CWG212646H, CWG212889H, CWG213346H, CWG213361H, CWG213408H, CWG213688H, CWG207423H, CWG210549H, CWG214088H, CWG209203H, CWG213067H, CWG213808H, CWG213223H, CWG205204H, CWG205295H, CWG206549H, CWG206920H, CWG209288H, CWG209437H, CWG209559H, CWG210260H, CWG210800H, CWG211327H, CWG213071H, CWG206503H, CWG206508H, CWG208509H, CWG205921H, CWG208265H, CWG204897H, CWG205465H, CWG206412H, CWG206476H, CWG207019H, CWG208665H, CWG213209H, CWG213283H, CWG213721H, CWG211734H, CWG207908H, CWG207909H, CWG207911H, CWG208224H, CWG207219H, CWG208005H, CWG214408H, CWG206367H, CWG206368H, CWG208424H, CWG208425H, CWG209865H, CWG211772H, CWG212595H, CWG211544H, CWG205044H, CWG205917H, CWG207125H, CWG207658H, CWG207659H, CWG207816H, CWG208373H, CWG208531H, CWG209193H, CWG210545H, CWG210763H, CWG212778H, CWG212829H, CWG213922H, CWG214220H, CWG214482H, CWG207103H, CWG207479H, CWG207652H, CWG207851H, CWG208026H, CWG208278H, CWG213061H, CWG210895H, CWG212338H, CWG208552H, CWG208620H, CWG208626H, CWG208629H, CWG208714H, CWG209296H, CWG210976H, CWG210977H, CWG210978H, CWG210980H, CWG210981H, CWG212246H, CWG212248H, CWG212249H, CWG212250H, CWG212251H, CWG212253H, CWG212255H, CWG212935H, CWG212939H, CWG213556H, CWG213558H, CWG213559H, CWG213560H, CWG213562H, CWG213564H, CWG205118H, CWG205124H, CWG205962H, CWG205996H, CWG206055H, CWG206400H, CWG206687H, CWG206921H, CWG206930H, CWG207140H, CWG207496H, CWG208016H, CWG208063H, CWG208183H, CWG208539H, CWG209138H, CWG209348H, CWG209397H, CWG209621H, CWG205011H, CWG205219H, CWG205270H, CWG205496H, CWG207117H, CWG208501H, CWG208565H, CWG208623H, CWG209405H, CWG209419H, CWG211643H, CWG207858H, CWG210849H, CWG204899H, CWG206381H, CWG207130H, CWG209549H, CWG211329H, CWG211880H, CWG212897H, CWG213932H, CWG205035H, CWG205268H, CWG206466H, CWG209557H, CWG209620H, CWG210072H, CWG210073H, CWG211332H, CWG211417H, CWG211497H, CWG211539H, CWG212351H, CWG212352H, CWG212353H, CWG212882H, CWG213272H, CWG213579H, CWG214508H, CWG205500H, CWG207074H, CWG207192H, CWG207260H, CWG209526H, CWG209561H, CWG210414H, CWG210591H, CWG211991H, CWG204994H, CWG205557H, CWG206077H, CWG209402H, CWG213276H, CWG207822H, CWG207825H, CWG207828H, CWG212794H, CWG206757H, CWG208936H, CWG206472H, CWG211454H, CWG208416H, CWG210581H, CWG210700H, CWG212397H, CWG212798H, CWG214156H, CWG213311H, CWG213544H, CWG207138H, CWG207170H, CWG208164H, CWG209118H, CWG210456H, CWG212360H, CWG213044H, CWG214387H, CWG212930H, CWG212652H, CWG206303H, CWG206623H, CWG207274H, CWG207946H, CWG208669H, CWG209122H, CWG209211H, CWG211472H, CWG210639H, CWG205102H, CWG205431H, CWG205994H, CWG206063H, CWG206414H, CWG206439H, CWG206501H, CWG207046H, CWG207445H, CWG207465H, CWG208101H, CWG209609H, CWG209671H, CWG210531H, CWG212884H, CWG205092H, CWG208657H, CWG205140H, CWG205141H, CWG205186H, CWG205682H, CWG205736H, CWG205764H, CWG206448H, CWG206456H, CWG206747H, CWG206761H, CWG206919H, CWG207240H, CWG207309H, CWG207406H, CWG207409H, CWG207733H, CWG207907H, CWG207954H, CWG208010H, CWG208052H, CWG208248H, CWG209070H, CWG209071H, CWG209590H, CWG210104H, CWG210441H, CWG210463H, CWG210485H, CWG210885H, CWG211330H, CWG211929H, CWG211985H, CWG212170H, CWG212678H, CWG213773H, CWG205355H, CWG205616H, CWG205618H, CWG205619H, CWG207717H, CWG211853H, CWG205667H, CWG205675H, CWG206380H, CWG206514H, CWG206784H, CWG206785H, CWG206786H, CWG206787H, CWG208782H, CWG208783H, CWG208784H, CWG208786H, CWG208787H, CWG209426H, CWG210347H, CWG210391H, CWG210421H, CWG210586H, CWG210660H, CWG210671H, CWG211100H, CWG211101H, CWG211883H, CWG212970H, CWG213279H, CWG213395H, CWG213438H, CWG213445H, CWG213446H, CWG213447H, CWG213448H, CWG213449H, CWG213548H, CWG209182H, CWG212763H, CWG213635H, CWG214432H, CWG206638H, CWG211492H, CWG211701H, CWG213571H, CWG214272H, CWG214406H, CWG212290H, CWG205854H, CWG209162H, CWG205226H, CWG206630H, CWG212159H, CWG206558H, CWG211703H, CWG205185H, CWG206362H, CWG207235H, CWG209794H, CWG212069H, CWG205250H, CWG205495H, CWG205857H, CWG206002H, CWG206003H, CWG207742H, CWG207756H, CWG207759H, CWG207760H, CWG207761H, CWG209496H, CWG209566H, CWG209783H, CWG209946H, CWG210165H, CWG210166H, CWG210167H, CWG210168H, CWG210169H, CWG210170H, CWG210171H, CWG210443H, CWG210511H, CWG210848H, CWG211749H, CWG211838H, CWG210136H, CWG214428H, CWG209399H, CWG210596H, CWG211723H, CWG211948H, CWG213109H, CWG213967H, CWG214109H, CWG214123H, CWG213787H, CWG208463H, CWG210982H, CWG212033H, CWG214085H, CWG209208H, CWG211562H, CWG205218H, CWG205220H, CWG207215H, CWG208035H, CWG206927H, CWG212864H, CWG206110H, CWG212145H, CWG205215H, CWG207055H, CWG209758H, CWG210458H, CWG206029H, CWG207434H, CWG208796H, CWG208003H, CWG209552H, CWG209986H, CWG210009H, CWG210994H, CWG213880H, CWG214337H, CWG205313H, CWG207617H, CWG209344H, CWG211527H, CWG212949H, CWG205938H, CWG207424H, CWG213795H, CWG214076H, CWG205041H, CWG205212H, CWG205632H, CWG205641H, CWG207065H, CWG209945H, CWG211498H, CWG213912H, CWG212751H, CWG213089H, CWG214182H, CWG206982H, CWG213380H, CWG205735H, CWG207970H, CWG209583H, CWG206971H, CWG204958H, CWG205397H, CWG206610H, CWG206941H, CWG206957H, CWG207430H, CWG208166H, CWG208253H, CWG208299H, CWG208313H, CWG208947H, CWG209670H, CWG209733H, CWG210640H, CWG210879H, CWG210993H, CWG210995H, CWG211479H, CWG211519H, CWG211809H, CWG211850H, CWG212123H, CWG212643H, CWG213258H, CWG213407H, CWG213429H, CWG213430H, CWG213433H, CWG213894H, CWG213915H, CWG214393H, CWG205342H, CWG205343H, CWG205725H, CWG205726H, CWG206073H, CWG206139H, CWG206601H, CWG206653H, CWG206987H, CWG209828H, CWG209829H, CWG211253H, CWG212470H, CWG212471H, CWG212474H, CWG212478H, CWG212479H, CWG212860H, CWG213482H, CWG213483H, CWG213484H, CWG213485H, CWG213752H, CWG205324H, CWG208646H, CWG210552H, CWG213603H, CWG206522H, CWG212029H, CWG214452H, CWG205103H, CWG205690H, CWG206112H, CWG206245H, CWG206918H, CWG207413H, CWG207481H, CWG207575H, CWG207750H, CWG207755H, CWG207757H, CWG208127H, CWG208370H, CWG208562H, CWG210287H, CWG210439H, CWG211079H, CWG211080H, CWG211081H, CWG211082H, CWG211469H, CWG211874H, CWG211887H, CWG212149H, CWG212628H, CWG213468H, CWG213604H, CWG213684H, CWG213838H, CWG214422H, CWG214507H, CWG205003H, CWG205084H, CWG205170H, CWG207686H, CWG210160H, CWG210162H, CWG213104H, CWG213624H, CWG213850H, CWG205168H, CWG206579H, CWG208054H, CWG209250H, CWG209258H, CWG209353H, CWG209547H, CWG211155H, CWG213541H, CWG208137H, CWG208138H, CWG208139H, CWG208140H, CWG209706H, CWG213730H, CWG213731H, CWG208037H, CWG208062H, CWG209663H, CWG212073H, CWG205267H, CWG205904H, CWG208466H, CWG205520H, CWG206424H, CWG208439H, CWG212718H, CWG213763H, CWG206206H, CWG210416H, CWG205259H, CWG205307H, CWG205314H, CWG205337H, CWG205652H, CWG205654H, CWG205658H, CWG205659H, CWG205660H, CWG205661H, CWG206334H, CWG206359H, CWG206791H, CWG206830H, CWG206841H, CWG206842H, CWG206843H, CWG206844H, CWG206847H, CWG206849H, CWG206850H, CWG207807H, CWG207860H, CWG209246H, CWG209870H, CWG209871H, CWG210951H, CWG210952H, CWG211019H, CWG211457H, CWG211566H, CWG211612H, CWG211744H, CWG211858H, CWG211892H, CWG212090H, CWG212091H, CWG212092H, CWG212237H, CWG212301H, CWG212408H, CWG212410H, CWG212458H, CWG212874H, CWG213254H, CWG213301H, CWG213435H, CWG213452H, CWG213549H, CWG213550H, CWG213551H, CWG213552H, CWG213649H, CWG214047H, CWG214226H, CWG214315H, CWG214317H, CWG214427H, CWG214456H, CWG205133H, CWG209475H, CWG212915H, CWG205395H, CWG205507H, CWG206287H, CWG210219H, CWG212418H, CWG205823H, CWG208296H, CWG206285H, CWG206293H, CWG206296H, CWG206702H, CWG206705H, CWG206706H, CWG206708H, CWG206709H, CWG209016H, CWG209025H, CWG212111H, CWG212289H, CWG212394H, CWG213520H, CWG213521H, CWG213523H, CWG213524H, CWG213525H, CWG213527H, CWG213528H, CWG213529H, CWG206176H, CWG206933H, CWG209576H, CWG209953H, CWG210193H, CWG211123H, CWG211317H, CWG211357H, CWG214316H, CWG205798H, CWG207271H, CWG204967H, CWG204968H, CWG205566H, CWG205572H, CWG211476H, CWG211886H, CWG213274H, CWG213632H, CWG214405H, CWG205745H, CWG206244H, CWG213196H, CWG210761H, CWG212734H, CWG214334H, CWG204902H, CWG204999H, CWG207349H, CWG210027H, CWG210129H, CWG211185H, CWG211190H, CWG211236H, CWG211590H, CWG211694H, CWG212000H, CWG212002H, CWG212078H, CWG212207H, CWG212247H, CWG212550H, CWG213176H, CWG213610H, CWG213893H, CWG213949H, CWG214116H, CWG214120H, CWG208376H, CWG210911H, CWG211989H, CWG213112H, CWG214265H, CWG206446H, CWG207141H, CWG208130H, CWG209632H, CWG210900H, CWG211432H, CWG212551H, CWG214071H, CWG214241H, CWG209830H, CWG209831H, CWG211425H, CWG208682H, CWG205308H, CWG205481H, CWG205482H, CWG205665H, CWG205888H, CWG205916H, CWG206624H, CWG206640H, CWG206641H, CWG207232H, CWG207687H, CWG207693H, CWG208017H, CWG208462H, CWG208967H, CWG208968H, CWG209435H, CWG209436H, CWG209985H, CWG210756H, CWG211993H, CWG212577H, CWG212857H, CWG212952H, CWG213033H, CWG213215H, CWG208579H, CWG211721H, CWG211924H, CWG205794H, CWG209009H, CWG213059H, CWG205027H, CWG205929H, CWG206337H, CWG207114H, CWG207546H, CWG209221H, CWG209443H, CWG209594H, CWG210325H, CWG211875H, CWG211943H, CWG212051H, CWG213034H, CWG213969H, CWG214197H, CWG204980H, CWG205193H, CWG208525H, CWG209903H, CWG212130H, CWG213099H, CWG206595H, CWG207375H, CWG208198H, CWG208604H, CWG209487H, CWG213882H, CWG210290H, CWG210517H, CWG210518H, CWG210676H, CWG210817H, CWG211689H, CWG211807H, CWG212620H, CWG208429H, CWG206349H, CWG205505H, CWG209973H, CWG206812H, CWG210209H, CWG212707H, CWG212863H, CWG213052H, CWG205221H, CWG205604H, CWG206642H, CWG206795H, CWG207660H, CWG208527H, CWG209109H, CWG209613H, CWG210664H, CWG210665H, CWG210836H, CWG211047H, CWG211263H, CWG211898H, CWG211899H, CWG212099H, CWG212499H, CWG212710H, CWG213325H, CWG213378H, CWG213650H, CWG213951H, CWG214485H, CWG210579H, CWG210950H, CWG211827H, CWG212061H, CWG213729H, CWG214125H, CWG213425H, CWG214048H, CWG214216H, CWG208528H, CWG205894H, CWG206419H, CWG211466H, CWG204917H, CWG204920H, CWG204922H, CWG204923H, CWG204925H, CWG204926H, CWG204927H, CWG206451H, CWG206454H, CWG206872H, CWG206873H, CWG206874H, CWG206875H, CWG206876H, CWG206878H, CWG206880H, CWG208293H, CWG208570H, CWG209072H, CWG209464H, CWG209466H, CWG209467H, CWG209997H, CWG210000H, CWG210001H, CWG210003H, CWG211130H, CWG211940H, CWG212600H, CWG212623H, CWG212624H, CWG213331H, CWG213332H, CWG213333H, CWG213805H, CWG212098H, CWG211312H, CWG211347H, CWG210063H, CWG210401H, CWG211548H, CWG205873H, CWG207296H, CWG208094H, CWG209267H, CWG209532H, CWG204921H, CWG206014H, CWG206116H, CWG207552H, CWG207646H, CWG209309H, CWG212635H, CWG212638H, CWG213367H, CWG213704H, CWG208029H, CWG210690H, CWG212933H, CWG213216H, CWG205957H, CWG208277H, CWG208427H, CWG210491H, CWG211040H, CWG211243H, CWG212673H, CWG213507H, CWG214273H, CWG207343H, CWG207320H, CWG214208H, CWG208504H, CWG210618H, CWG213867H, CWG214509H, CWG205428H, CWG206372H, CWG207139H, CWG207364H, CWG207718H, CWG208236H, CWG209273H, CWG209311H, CWG209401H, CWG209626H, CWG210300H, CWG210988H, CWG213751H, CWG207300H, CWG210778H, CWG211915H, CWG209761H, CWG210113H, CWG210470H, CWG210819H, CWG210986H, CWG211196H, CWG211237H, CWG211642H, CWG212028H, CWG213029H, CWG213835H, CWG214264H, CWG205277H, CWG205278H, CWG205279H, CWG205280H, CWG205281H, CWG205282H, CWG208845H, CWG210128H, CWG210130H, CWG210132H, CWG212198H, CWG212199H, CWG212200H, CWG212202H, CWG213961H, CWG208331H, CWG208397H, CWG209612H, CWG209812H, CWG210794H, CWG212443H, CWG213545H, CWG213557H, CWG205189H, CWG205384H, CWG205878H, CWG206279H, CWG206388H, CWG206420H, CWG206445H, CWG206949H, CWG207207H, CWG207259H, CWG207354H, CWG207443H, CWG207541H, CWG207590H, CWG207903H, CWG208104H, CWG208165H, CWG208651H, CWG208677H, CWG208928H, CWG208940H, CWG209114H, CWG209411H, CWG209633H, CWG209911H, CWG209959H, CWG210335H, CWG210407H, CWG210534H, CWG210854H, CWG211063H, CWG211064H, CWG211065H, CWG211067H, CWG211068H, CWG211069H, CWG211265H, CWG211266H, CWG211440H, CWG212276H, CWG212277H, CWG212279H, CWG212280H, CWG212281H, CWG212563H, CWG212694H, CWG212901H, CWG212978H, CWG213155H, CWG213411H, CWG213509H, CWG213547H, CWG213642H, CWG213643H, CWG213644H, CWG213645H, CWG213646H, CWG213647H, CWG213648H, CWG213668H, CWG213771H, CWG213799H, CWG213904H, CWG214010H, CWG214151H, CWG214218H, CWG214271H, CWG214312H, CWG214320H, CWG214336H, CWG214362H, CWG205765H, CWG209717H, CWG210802H, CWG213839H, CWG209134H, CWG209150H, CWG209212H, CWG210489H, CWG210621H, CWG211513H, CWG212001H, CWG207530H, CWG209452H, CWG210228H, CWG212031H, CWG210275H, CWG210864H, CWG212766H, CWG213309H, CWG204992H, CWG213262H, CWG204975H, CWG211273H, CWG206433H, CWG207451H, CWG209923H, CWG210214H, CWG210230H, CWG210760H, CWG214401H, CWG206298H, CWG208608H, CWG210011H, CWG211158H, CWG211666H, CWG214257H, CWG211039H, CWG212674H, CWG213660H, CWG214469H, CWG208945H, CWG210822H, CWG211949H, CWG208885H, CWG212007H, CWG205136H, CWG206059H, CWG206390H, CWG206648H, CWG207052H, CWG208733H, CWG209082H, CWG209610H, CWG210004H, CWG210112H, CWG211560H, CWG211601H, CWG213187H, CWG213842H, CWG208328H, CWG210324H, CWG206370H, CWG206716H, CWG207075H, CWG207336H, CWG208022H, CWG209301H, CWG209936H, CWG211181H, CWG211569H, CWG212134H, CWG214003H, CWG214440H, CWG211974H, CWG206534H, CWG210987H, CWG211366H, CWG211395H, CWG212719H, CWG213088H, CWG214176H, CWG205447H, CWG205883H, CWG205992H, CWG206132H, CWG206297H, CWG207338H, CWG213025H, CWG205915H, CWG207210H, CWG208319H, CWG209361H, CWG209721H, CWG212215H, CWG206946H, CWG210061H, CWG210320H, CWG209606H, CWG209896H, CWG212040H, CWG205923H, CWG206170H, CWG211251H, CWG211959H, CWG211966H, CWG213154H, CWG213501H, CWG213000H, CWG212642H, CWG206255H, CWG212626H, CWG212629H, CWG208086H, CWG211503H, CWG206058H, CWG206292H, CWG206978H, CWG207247H, CWG210361H, CWG211057H, CWG212419H, CWG212421H, CWG213640H, CWG213641H, CWG207110H, CWG207446H, CWG212740H, CWG212880H, CWG213259H, CWG206060H, CWG210870H, CWG208269H, CWG210224H, CWG205135H, CWG208068H, CWG208386H, CWG209528H, CWG210385H, CWG210776H, CWG210827H, CWG211161H, CWG211206H, CWG211955H, CWG211992H, CWG212220H, CWG212233H, CWG212538H, CWG212596H, CWG212767H, CWG213118H, CWG213235H, CWG213349H, CWG213609H, CWG213878H, CWG213886H, CWG214015H, CWG214106H, CWG214224H, CWG214355H, CWG214501H, CWG211042H, CWG207270H, CWG209637H, CWG209827H, CWG212726H, CWG213834H, CWG207691H, CWG213972H, CWG206095H, CWG208007H, CWG208856H, CWG210396H, CWG210425H, CWG208422H, CWG209712H, CWG210232H, CWG210529H, CWG210701H, CWG212655H, CWG206637H, CWG207831H, CWG208367H, CWG209297H, CWG211660H, CWG208891H, CWG208892H, CWG208893H, CWG208895H, CWG209822H, CWG212445H, CWG212446H, CWG212731H, CWG213534H, CWG205748H, CWG205970H, CWG208148H, CWG214421H, CWG214466H, CWG211224H, CWG214322H, CWG208546H, CWG210236H, CWG206258H, CWG208622H, CWG205808H, CWG208078H, CWG208327H, CWG213241H, CWG213281H, CWG207152H, CWG210432H, CWG206779H, CWG207974H, CWG208241H, CWG208337H, CWG208492H, CWG208732H, CWG208827H, CWG209394H, CWG209517H, CWG209631H, CWG210181H, CWG211201H, CWG211204H, CWG212698H, CWG212869H, CWG212996H, CWG213261H, CWG213406H, CWG213758H, CWG209194H, CWG210825H, CWG213074H, CWG207459H, CWG209330H, CWG213336H, CWG207579H, CWG211889H, CWG212467H, CWG212784H, CWG205624H, CWG206280H, CWG213173H, CWG213342H, CWG214429H, CWG214430H, CWG208460H, CWG211859H, CWG213307H, CWG205426H, CWG210223H, CWG210389H, CWG210845H, CWG210867H, CWG213219H, CWG213555H, CWG213666H, CWG213885H, CWG214276H, CWG214484H, CWG205227H, CWG204974H, CWG206580H, CWG207155H, CWG212797H, CWG213621H, CWG206166H, CWG207662H, CWG208264H, CWG209027H, CWG209672H, CWG210473H, CWG210698H, CWG210840H, CWG210861H, CWG210869H, CWG211746H, CWG211877H, CWG212136H, CWG212258H, CWG212728H, CWG212940H, CWG213396H, CWG213427H, CWG213636H, CWG214105H, CWG211258H, CWG214338H, CWG214434H, CWG206810H, CWG209860H, CWG210790H, CWG212760H, CWG205795H, CWG212604H, CWG204988H, CWG206027H, CWG210826H, CWG213035H, CWG214207H, CWG214439H, CWG205793H, CWG205909H, CWG206998H, CWG207431H, CWG209824H, CWG210563H, CWG211700H, CWG211952H, CWG212801H, CWG213189H, CWG207172H, CWG211673H, CWG212941H, CWG213113H, CWG214463H, CWG205797H, CWG206788H, CWG208067H, CWG208080H, CWG209634H, CWG210769H, CWG210835H, CWG213436H, CWG207819H, CWG211533H, CWG213164H, CWG213937H, CWG205807H, CWG209931H, CWG205651H, CWG205653H, CWG206492H, CWG207537H, CWG209462H, CWG209999H, CWG211162H, CWG212888H, CWG213292H, CWG214234H, CWG214468H, CWG205309H, CWG205549H, CWG206325H, CWG206340H, CWG206748H, CWG208800H, CWG208941H, CWG211555H, CWG212175H, CWG212177H, CWG213299H, CWG213499H, CWG206089H, CWG211633H, CWG204937H, CWG206310H, CWG208356H, CWG208790H, CWG208859H, CWG209533H, CWG209605H, CWG209937H, CWG210478H, CWG210694H, CWG210728H, CWG211386H, CWG211668H, CWG205028H, CWG211380H, CWG211412H, CWG213392H, CWG214353H, CWG212943H, CWG207863H, CWG210968H, CWG211691H, CWG205601H, CWG205706H, CWG206443H, CWG206681H, CWG206801H, CWG206869H, CWG207285H, CWG207917H, CWG208393H, CWG208768H, CWG209880H, CWG209881H, CWG209995H, CWG211048H, CWG211049H, CWG211270H, CWG211763H, CWG211906H, CWG212501H, CWG212743H, CWG213638H, CWG213675H, CWG214165H, CWG214222H, CWG214383H, CWG214470H, CWG209083H, CWG209256H, CWG211608H, CWG213070H, CWG213242H, CWG213856H, CWG205196H, CWG205416H, CWG205606H, CWG205609H, CWG205610H, CWG205612H, CWG205613H, CWG206137H, CWG206802H, CWG206803H, CWG206804H, CWG206805H, CWG206806H, CWG207206H, CWG207276H, CWG207337H, CWG207476H, CWG207540H, CWG207609H, CWG207880H, CWG207920H, CWG207921H, CWG207922H, CWG207923H, CWG207925H, CWG207926H, CWG207927H, CWG207928H, CWG208111H, CWG208203H, CWG208737H, CWG208795H, CWG210124H, CWG210155H, CWG210157H, CWG210158H, CWG210449H, CWG210680H, CWG210747H, CWG210969H, CWG210972H, CWG210975H, CWG211183H, CWG211225H, CWG211806H, CWG211828H, CWG211852H, CWG211958H, CWG212049H, CWG212067H, CWG212097H, CWG212197H, CWG212294H, CWG212295H, CWG212296H, CWG212297H, CWG212298H, CWG212429H, CWG212505H, CWG212576H, CWG212583H, CWG212584H, CWG212585H, CWG212692H, CWG212919H, CWG212920H, CWG212937H, CWG212990H, CWG213024H, CWG213201H, CWG213253H, CWG213265H, CWG213743H, CWG213744H, CWG213762H, CWG213931H, CWG213957H, CWG214140H, CWG214211H, CWG214213H, CWG207610H, CWG212712H, CWG207128H, CWG204963H, CWG204965H, CWG205867H, CWG206215H, CWG206289H, CWG206797H, CWG206798H, CWG207749H, CWG207879H, CWG208316H, CWG208834H, CWG205875H, CWG210125H, CWG210762H, CWG212030H, CWG213032H, CWG206326H, CWG209747H, CWG212369H, CWG213195H, CWG213233H, CWG205317H, CWG206793H, CWG207121H, CWG208872H, CWG209808H, CWG210084H, CWG210615H, CWG213903H, CWG213996H, CWG205149H, CWG205830H, CWG205995H, CWG208697H, CWG209811H, CWG212110H, CWG212541H, CWG212670H, CWG212981H, CWG213875H, CWG214438H, CWG205893H, CWG206505H, CWG208255H, CWG208398H, CWG211073H, CWG211289H, CWG211437H, CWG212138H, CWG212607H, CWG213005H, CWG213827H, CWG213943H, CWG214201H, CWG214374H, CWG214392H, CWG205955H, CWG206286H, CWG208024H, CWG208113H, CWG209092H, CWG209317H, CWG210383H, CWG212196H, CWG213587H, CWG214117H, CWG205733H, CWG207673H, CWG206264H, CWG206403H, CWG206658H, CWG206769H, CWG207068H, CWG207079H, CWG207985H, CWG207991H, CWG208102H, CWG208280H, CWG208432H, CWG208448H, CWG209328H, CWG209770H, CWG210585H, CWG210674H, CWG210839H, CWG211434H, CWG211752H, CWG213515H, CWG205889H, CWG206495H, CWG209838H, CWG211337H, CWG213847H, CWG207202H, CWG207316H, CWG209542H, CWG206401H, CWG211229H, CWG207881H, CWG208215H, CWG208603H, CWG209247H, CWG210114H, CWG213658H, CWG205703H, CWG206105H, CWG206939H, CWG207190H, CWG207461H, CWG207513H, CWG207661H, CWG208107H, CWG209204H, CWG209551H, CWG210286H, CWG210609H, CWG212671H, CWG204981H, CWG205434H, CWG206428H, CWG206464H, CWG206586H, CWG207027H, CWG207289H, CWG207564H, CWG209148H, CWG209173H, CWG210479H, CWG210486H, CWG210684H, CWG211302H, CWG212071H, CWG212157H, CWG212776H, CWG212950H, CWG213169H, CWG213227H, CWG207439H, CWG210427H, CWG206636H, CWG212187H, CWG204942H, CWG204943H, CWG204946H, CWG204948H, CWG205037H, CWG205079H, CWG205128H, CWG205164H, CWG205210H, CWG205628H, CWG205630H, CWG206028H, CWG206111H, CWG206573H, CWG206574H, CWG206575H, CWG206859H, CWG206860H, CWG207344H, CWG207775H, CWG207951H, CWG207956H, CWG211112H, CWG211113H, CWG211114H, CWG211935H, CWG213373H, CWG213374H, CWG213443H, CWG213444H, CWG207078H, CWG207532H, CWG212813H, CWG211528H, CWG213383H, CWG205761H, CWG205884H, CWG207442H, CWG209375H, CWG208442H, CWG208317H, CWG211916H, CWG212665H, CWG205188H, CWG205633H, CWG205635H, CWG205637H, CWG205638H, CWG205639H, CWG205640H, CWG207621H, CWG207704H, CWG207705H, CWG207706H, CWG207710H, CWG208193H, CWG208227H, CWG208230H, CWG208617H, CWG208621H, CWG209874H, CWG210057H, CWG210134H, CWG210137H, CWG210138H, CWG210139H, CWG211628H, CWG212449H, CWG212451H, CWG212452H, CWG212454H, CWG212459H, CWG212461H, CWG212463H, CWG212468H, CWG205515H, CWG206996H, CWG207182H, CWG208143H, CWG211433H, CWG213439H, CWG207243H, CWG207437H, CWG208915H, CWG208972H, CWG209054H, CWG209942H, CWG210983H, CWG211939H, CWG211982H, CWG212567H, CWG213171H, CWG213322H, CWG213450H, CWG213565H, CWG213653H, CWG213654H, CWG213655H, CWG214019H, CWG214020H, CWG214021H, CWG214087H, CWG214094H, CWG205781H, CWG212004H, CWG206167H, CWG213428H, CWG205991H, CWG207654H, CWG210535H, CWG204919H, CWG212654H, CWG213115H, CWG206526H, CWG206566H, CWG210053H, CWG213619H, CWG213942H, CWG210477H, CWG211282H, CWG211917H, CWG212190H, CWG212341H, CWG212867H, CWG213348H, CWG214448H, CWG211132H, CWG206252H, CWG211427H, CWG211467H, CWG212100H, CWG213194H, CWG209233H, CWG210933H, CWG209349H, CWG210363H, CWG210992H, CWG211304H, CWG213907H, CWG205455H, CWG205540H, CWG205541H, CWG206690H, CWG206691H, CWG206692H, CWG206704H, CWG208851H, CWG208946H, CWG208948H, CWG209789H, CWG210083H, CWG210905H, CWG211001H, CWG211002H, CWG211351H, CWG212260H, CWG212261H, CWG212322H, CWG205014H, CWG206562H, CWG206612H, CWG208695H, CWG209805H, CWG209834H, CWG211658H, CWG213127H, CWG213948H, CWG214243H, CWG207203H, CWG207383H, CWG208258H, CWG210343H, CWG210448H, CWG210903H, CWG211390H, CWG211647H, CWG211699H, CWG211707H, CWG212632H, CWG213180H, CWG213182H, CWG213190H, CWG213401H, CWG214093H, CWG214193H, CWG214269H, CWG211846H, CWG209324H, CWG206205H, CWG207624H, CWG211496H, CWG211869H, CWG212830H, CWG214357H, CWG205004H, CWG205556H, CWG206033H, CWG206726H, CWG207751H, CWG208770H, CWG209685H, CWG210949H, CWG205920H, CWG206207H, CWG205925H, CWG214068H, CWG213757H, CWG210321H, CWG210813H, CWG212056H, CWG212956H, CWG206581H, CWG205172H, CWG206084H, CWG206497H, CWG206629H, CWG207058H, CWG207620H, CWG208908H, CWG211414H, CWG211710H, CWG213077H, CWG206389H, CWG208195H, CWG207671H, CWG211460H, CWG212113H, CWG205519H, CWG206727H, CWG206728H, CWG208217H, CWG211241H, CWG211409H, CWG212342H, CWG213477H, CWG213570H, CWG214335H, CWG205301H, CWG206635H, CWG206725H, CWG207550H, CWG205948H, CWG206375H, CWG207347H, CWG208145H, CWG205456H, CWG205467H, CWG205531H, CWG205533H, CWG205539H, CWG207726H, CWG207728H, CWG209793H, CWG210065H, CWG210066H, CWG210067H, CWG210068H, CWG210070H, CWG210071H, CWG212129H, CWG212496H, CWG213572H, CWG213573H, CWG213574H, CWG213575H, CWG213921H, CWG214023H, CWG213814H, CWG205109H, CWG205232H, CWG213098H, CWG208524H, CWG211404H, CWG214115H, CWG207557H, CWG211286H, CWG211495H, CWG213181H, CWG213244H, CWG205662H, CWG207063H, CWG212022H, CWG205814H, CWG208336H, CWG205818H, CWG207843H, CWG207949H, CWG209286H, CWG209484H, CWG212723H, CWG212739H, CWG213612H, CWG214163H, CWG209504H, CWG209505H, CWG207643H, CWG210078H, CWG210234H, CWG212745H, CWG214327H, CWG208761H, CWG208764H, CWG206680H, CWG207088H, CWG209765H, CWG205088H, CWG205096H, CWG210417H, CWG211090H, CWG210547H, CWG205903H, CWG206963H, CWG207168H, CWG211515H, CWG212959H, CWG214219H, CWG212988H, CWG214364H, CWG205180H, CWG205943H, CWG206668H, CWG206863H, CWG207859H, CWG207862H, CWG209032H, CWG209033H, CWG209516H, CWG210292H, CWG210308H, CWG211036H, CWG211334H, CWG211799H, CWG211933H, CWG211946H, CWG212184H, CWG212571H, CWG212609H, CWG212814H, CWG213394H, CWG211758H, CWG205240H, CWG208533H, CWG211370H, CWG205353H, CWG204896H, CWG205469H, CWG205591H, CWG206455H, CWG206700H, CWG206707H, CWG207222H, CWG207727H, CWG208225H, CWG208641H, CWG209088H, CWG209847H, CWG211141H, CWG213625H, CWG207035H, CWG207618H, CWG208051H, CWG209564H, CWG205683H, CWG206373H, CWG204976H, CWG213979H, CWG211316H, CWG211360H, CWG211367H, CWG213510H, CWG207358H, CWG210080H, CWG206945H, CWG206148H, CWG206952H, CWG210847H, CWG210149H, CWG211148H, CWG211617H, CWG210616H, CWG211641H, CWG212783H, CWG206007H, CWG206583H, CWG205085H, CWG205783H, CWG206042H, CWG206387H, CWG206886H, CWG207588H, CWG208009H, CWG208192H, CWG208362H, CWG208630H, CWG209157H, CWG209332H, CWG209694H, CWG209710H, CWG209785H, CWG210823H, CWG211521H, CWG211556H, CWG211842H, CWG212631H, CWG212788H, CWG205408H, CWG205692H, CWG207241H, CWG209364H, CWG209798H, CWG209850H, CWG212799H, CWG207576H, CWG209389H, CWG209816H, CWG207086H, CWG212010H, CWG207847H, CWG206103H, CWG206363H, CWG206992H, CWG209263H, CWG213416H, CWG213518H, CWG209195H, CWG210291H, CWG209254H, CWG209281H, CWG209550H, CWG209451H, CWG207001H, CWG207804H, CWG209172H, CWG209495H, CWG210452H, CWG210499H, CWG210865H, CWG213057H, CWG213092H, CWG213514H, CWG205550H, CWG205849H, CWG209715H, CWG212532H, CWG205477H, CWG205840H, CWG205972H, CWG208160H, CWG209363H, CWG210860H, CWG211369H, CWG211841H, CWG211860H, CWG211888H, CWG211904H, CWG212379H, CWG212605H, CWG212636H, CWG208133H, CWG208173H, CWG208812H, CWG209120H, CWG209749H, CWG210766H, CWG213238H, CWG213312H, CWG206903H, CWG209142H, CWG209270H, CWG209641H, CWG211453H, CWG211704H, CWG211879H, CWG209057H, CWG209058H, CWG209814H, CWG209843H, CWG210598H, CWG211205H, CWG212936H, CWG213494H, CWG213495H, CWG213496H, CWG205956H, CWG207480H, CWG209131H, CWG210853H, CWG213318H, CWG214476H, CWG213170H, CWG213936H, CWG205320H, CWG205321H, CWG206022H, CWG207377H, CWG208886H, CWG210288H, CWG210360H, CWG210620H, CWG211209H, CWG211311H, CWG212155H, CWG212682H, CWG213785H, CWG213793H, CWG213994H, CWG206540H, CWG209107H, CWG209242H, CWG213123H, CWG208715H, CWG212549H, CWG205459H, CWG208911H, CWG209917H, CWG209920H, CWG210446H, CWG211738H, CWG212015H, CWG212826H, CWG214223H, CWG214255H, CWG214394H, CWG214425H, CWG214450H, CWG205116H, CWG205120H, CWG205231H, CWG205354H, CWG205617H, CWG206102H, CWG206498H, CWG206619H, CWG206738H, CWG209325H, CWG210454H, CWG210507H, CWG210898H, CWG213008H, CWG213058H, CWG213090H, CWG213424H, CWG213944H, CWG214032H, CWG214037H, CWG214039H, CWG205543H, CWG205544H, CWG205545H, CWG205546H, CWG205548H, CWG206729H, CWG206730H, CWG207362H, CWG207722H, CWG207723H, CWG210261H, CWG210273H, CWG210730H, CWG210731H, CWG210732H, CWG210733H, CWG210734H, CWG212242H, CWG212243H, CWG212244H, CWG214034H, CWG214035H, CWG205121H, CWG205203H, CWG205319H, CWG205822H, CWG207890H, CWG208704H, CWG208705H, CWG208707H, CWG208708H, CWG208742H, CWG209004H, CWG209350H, CWG209351H, CWG209582H, CWG209586H, CWG209649H, CWG209675H, CWG210039H, CWG210040H, CWG210042H, CWG210043H, CWG210045H, CWG210190H, CWG210749H, CWG211494H, CWG212372H, CWG213866H, CWG214188H, CWG214330H, CWG206291H, CWG207256H, CWG207941H, CWG208843H, CWG210526H, CWG214326H, CWG206531H, CWG208613H, CWG209425H, CWG209817H, CWG210116H, CWG210133H, CWG210767H, CWG211110H, CWG211681H, CWG212554H, CWG209523H, CWG210368H, CWG209821H, CWG210198H, CWG211088H, CWG211089H, CWG211091H, CWG211203H, CWG211293H, CWG212515H, CWG212516H, CWG205674H, CWG206576H, CWG206936H, CWG207523H, CWG207548H, CWG208305H, CWG210685H, CWG211296H, CWG211579H, CWG211620H, CWG211876H, CWG212211H, CWG212219H, CWG212653H, CWG213852H, CWG205953H, CWG211708H, CWG211900H, CWG211926H, CWG211938H, CWG212042H, CWG212741H, CWG205882H, CWG209264H, CWG209708H, CWG206267H, CWG209153H, CWG209174H, CWG209711H, CWG210423H, CWG210812H, CWG211931H, CWG205932H, CWG208152H, CWG209154H, CWG209714H, CWG210497H, CWG212343H, CWG212346H, CWG212354H, CWG212608H, CWG213305H, CWG213872H, CWG214225H, CWG214431H, CWG214474H, CWG205370H, CWG205046H, CWG210601H, CWG205385H, CWG205503H, CWG205800H, CWG206422H, CWG210642H, CWG213569H, CWG207639H, CWG210258H, CWG204989H, CWG206049H, CWG207668H, CWG208216H, CWG205216H, CWG205475H, CWG209015H, CWG205565H, CWG210332H, CWG213053H, CWG214013H, CWG214062H, CWG209282H, CWG214011H, CWG205070H, CWG205767H, CWG205859H, CWG206259H, CWG207107H, CWG208270H, CWG208315H, CWG208408H, CWG209316H, CWG209321H, CWG210050H, CWG210076H, CWG210542H, CWG210887H, CWG211083H, CWG211309H, CWG211470H, CWG211474H, CWG211740H, CWG212023H, CWG212239H, CWG212365H, CWG212366H, CWG212370H, CWG212371H, CWG212615H, CWG212716H, CWG212878H, CWG213340H, CWG213611H, CWG213613H, CWG213755H, CWG213976H, CWG214104H, CWG214161H, CWG206365H, CWG206684H, CWG211825H, CWG213251H, CWG205575H, CWG205620H, CWG213391H, CWG210791H, CWG211692H, CWG209806H, CWG206155H, CWG207275H, CWG208675H, CWG210554H, CWG212413H, CWG213237H, CWG213288H, CWG213670H, CWG213772H, CWG213784H, CWG214158H, CWG214280H, CWG214301H, CWG214481H, CWG206113H, CWG207334H, CWG208526H, CWG212689H, CWG207311H, CWG205810H, CWG208685H, CWG209300H, CWG211339H, CWG211364H, CWG206585H, CWG207091H, CWG208557H, CWG210703H, CWG210960H, CWG213966H, CWG214175H, CWG214441H, CWG205262H, CWG208055H, CWG209820H, CWG210498H, CWG214059H, CWG205881H, CWG210304H, CWG205347H, CWG205935H, CWG207594H, CWG209410H, CWG209177H, CWG210695H, CWG213031H, CWG213277H, CWG206044H, CWG213107H, CWG205318H, CWG212612H, CWG213403H, CWG213616H, CWG208395H, CWG206092H, CWG207122H, CWG210265H, CWG208413H, CWG210398H, CWG212062H, CWG213781H, CWG207038H, CWG208950H, CWG209578H, CWG213506H, CWG213841H, CWG209895H, CWG211627H, CWG211792H, CWG212905H, CWG205030H, CWG205061H, CWG205072H, CWG205179H, CWG206278H, CWG206746H, CWG206942H, CWG207453H, CWG207536H, CWG208170H, CWG208999H, CWG209000H, CWG210339H, CWG210538H, CWG210561H, CWG211050H, CWG211051H, CWG211052H, CWG212587H, CWG214081H, CWG206276H, CWG210550H, CWG211618H, CWG212948H, CWG208906H, CWG205257H, CWG205430H, CWG205717H, CWG205806H, CWG205890H, CWG205907H, CWG205914H, CWG206087H, CWG206196H, CWG206332H, CWG206342H, CWG206544H, CWG206564H, CWG206701H, CWG207098H, CWG207143H, CWG207145H, CWG207153H, CWG207438H, CWG208474H, CWG208536H, CWG208676H, CWG208696H, CWG208711H, CWG208712H, CWG208713H, CWG209090H, CWG209144H, CWG209315H, CWG209553H, CWG209801H, CWG209914H, CWG209969H, CWG210333H, CWG211233H, CWG211908H, CWG214457H, CWG205042H, CWG205982H, CWG206072H, CWG206361H, CWG206550H, CWG211686H, CWG210868H, CWG205740H, CWG207059H, CWG208182H, CWG208239H, CWG208671H, CWG209875H, CWG211465H, CWG211965H, CWG207279H, CWG206682H, CWG211559H, CWG211783H, CWG212058H, CWG214244H, CWG205698H, CWG207185H, CWG209312H, CWG205122H, CWG206355H, CWG207124H, CWG207177H, CWG207670H, CWG208876H, CWG214045H, CWG209878H, CWG209879H, CWG210828H, CWG211507H, CWG211436H, CWG207254H, CWG208208H, CWG208644H, CWG209224H, CWG209474H, CWG209849H, CWG210997H, CWG211285H, CWG212008H, CWG213220H, CWG208510H, CWG206277H, CWG210294H, CWG214082H, CWG207304H, CWG212917H, CWG209228H, CWG205409H, CWG205518H, CWG205134H, CWG209458H, CWG210455H, CWG214083H, CWG214424H, CWG209310H, CWG209530H, CWG209601H, CWG207317H, CWG211951H, CWG212997H, CWG213047H, CWG205351H, CWG209209H, CWG207421H, CWG211918H, CWG212622H, CWG210328H, CWG205138H, CWG206686H, CWG207840H, CWG208161H, CWG208530H, CWG209335H, CWG211936H, CWG213803H, CWG208927H, CWG210229H, CWG210317H, CWG206043H, CWG207286H, CWG207306H, CWG208302H, CWG208497H, CWG207367H, CWG212641H, CWG207501H, CWG208211H, CWG214078H, CWG214119H, CWG207136H, CWG207363H, CWG207474H, CWG208428H, CWG211766H, CWG213239H, CWG213753H, CWG213963H, CWG213999H, CWG207883H, CWG207884H, CWG209678H, CWG209679H, CWG211075H, CWG211076H, CWG211078H, CWG205769H, CWG205887H, CWG212137H, CWG213284H, CWG213286H, CWG213287H, CWG211602H, CWG214263H, CWG206294H, CWG207213H, CWG207539H, CWG207663H, CWG208131H, CWG209809H, CWG210145H, CWG211373H, CWG214413H, CWG212232H, CWG208339H, CWG209836H, CWG210085H, CWG209486H, CWG205913H, CWG205928H, CWG206984H, CWG212535H, CWG212603H, CWG213375H, CWG213947H, CWG208346H, CWG209190H, CWG212580H, CWG212916H, CWG212989H, CWG213410H, CWG213454H, CWG213526H, CWG213950H, CWG213701H, CWG214240H, CWG205132H, CWG207265H, CWG208597H, CWG208814H, CWG209294H, CWG209726H, CWG210999H, CWG211128H, CWG211805H, CWG212105H, CWG212203H, CWG214305H, CWG206239H, CWG209119H, CWG210702H, CWG213479H, CWG208839H, CWG210211H, CWG210329H, CWG207376H, CWG213791H, CWG211308H, CWG212613H, CWG208222H, CWG209055H, CWG209538H, CWG212681H, CWG213106H, CWG210194H, CWG206896H, CWG212601H, CWG210805H, CWG209876H, CWG213761H, CWG213819H, CWG205744H, CWG207765H, CWG210429H, CWG212558H, CWG213248H, CWG214278H, CWG205751H, CWG205400H, CWG205401H, CWG205405H, CWG205407H, CWG207131H, CWG207135H, CWG211661H, CWG211920H, CWG212151H, CWG214061H, CWG214135H, CWG214155H, CWG205636H, CWG207871H, CWG208206H, CWG208484H, CWG208934H, CWG210103H, CWG213245H, CWG214464H, CWG205357H, CWG205950H, CWG206182H, CWG206185H, CWG206188H, CWG206189H, CWG206192H, CWG207379H, CWG207380H, CWG207384H, CWG207385H, CWG207386H, CWG207387H, CWG207388H, CWG207391H, CWG207393H, CWG207394H, CWG208291H, CWG209700H, CWG209701H, CWG209702H, CWG209784H, CWG209853H, CWG209854H, CWG209855H, CWG209856H, CWG209857H, CWG211137H, CWG212226H, CWG212227H, CWG212228H, CWG212424H, CWG212427H, CWG212428H, CWG212430H, CWG212431H, CWG212434H, CWG212435H, CWG212436H, CWG213337H, CWG213356H, CWG213357H, CWG213712H, CWG206666H, CWG206040H, CWG207129H, CWG212793H, CWG207732H, CWG208611H, CWG213908H, CWG207854H, CWG211288H, CWG205125H, CWG205831H, CWG206199H, CWG206318H, CWG206320H, CWG206525H, CWG206861H, CWG207497H, CWG208075H, CWG209308H, CWG211685H, CWG212006H, CWG213989H, CWG214114H, CWG205976H, CWG214149H, CWG214259H, CWG205294H, CWG205986H, CWG206884H, CWG208116H, CWG208417H, CWG208607H, CWG210146H, CWG210551H, CWG212586H, CWG212992H, CWG213256H, CWG214044H, CWG214153H, CWG204938H, CWG214107H, CWG211790H, CWG207486H, CWG208511H, CWG207753H, CWG208807H, CWG209490H, CWG211398H, CWG213222H, CWG213843H, CWG214113H, CWG209188H, CWG209635H, CWG213320H, CWG206862H, CWG206902H, CWG207262H, CWG207574H, CWG207647H, CWG207904H, CWG208537H, CWG208709H, CWG209191H, CWG209565H, CWG209910H, CWG210801H, CWG211509H, CWG211922H, CWG212048H, CWG212462H, CWG212748H, CWG212752H, CWG212898H, CWG212912H, CWG212983H, CWG213012H, CWG213379H, CWG214033H, CWG214138H, CWG214143H, CWG214154H, CWG214367H, CWG208271H, CWG207848H, CWG208514H, CWG208515H, CWG208516H, CWG208517H, CWG208728H, CWG208772H, CWG212769H, CWG212771H, CWG212780H, CWG205899H, CWG209117H, CWG207366H, CWG205182H, CWG207703H, CWG213600H, CWG204960H, CWG204961H, CWG204962H, CWG212299H, CWG206263H, CWG209338H, CWG213048H, CWG214267H, CWG205836H, CWG209249H, CWG210496H, CWG206069H, CWG206338H, CWG209441H, CWG209718H, CWG210852H, CWG205039H, CWG206251H, CWG207682H, CWG205063H, CWG205144H, CWG205374H, CWG205981H, CWG205989H, CWG206034H, CWG206645H, CWG206898H, CWG206911H, CWG208452H, CWG210935H, CWG210943H, CWG211164H, CWG211184H, CWG212238H, CWG205007H, CWG205198H, CWG210326H, CWG210344H, CWG210617H, CWG210755H, CWG211287H, CWG211381H, CWG211403H, CWG212489H, CWG212644H, CWG212717H, CWG212954H, CWG213889H, CWG214170H, CWG211592H, CWG209787H, CWG210672H, CWG210508H, CWG206774H, CWG207472H, CWG208012H, CWG212994H, CWG212995H, CWG213081H, CWG213809H, CWG212742H, CWG206169H, CWG207371H, CWG210795H, CWG212038H, CWG212081H, CWG213964H, CWG214217H, CWG205000H, CWG206057H, CWG208558H, CWG208801H, CWG209970H, CWG209994H, CWG210387H, CWG210777H, CWG210781H, CWG213134H, CWG214038H, CWG208093H, CWG208592H, CWG204898H, CWG206107H, CWG206352H, CWG211576H, CWG205691H, CWG209556H, CWG209581H, CWG210487H, CWG210821H, CWG212122H, CWG213136H, CWG213917H, CWG211905H, CWG213338H, CWG213039H, CWG205684H, CWG208805H, CWG210047H, CWG208931H, CWG209181H, CWG210942H, CWG205335H, CWG206655H, CWG207899H, CWG209779H, CWG206463H, CWG210380H, CWG207543H, CWG213836H, CWG207495H, CWG207608H, CWG213824H, CWG206908H, CWG207669H, CWG208099H, CWG211678H, CWG213167H, CWG209737H, CWG209251H, CWG208309H, CWG213831H, CWG206377H, CWG208038H, CWG211108H, CWG211354H, CWG212245H, CWG213121H, CWG213672H, CWG207016H, CWG208281H, CWG211650H, CWG213455H, CWG214318H, CWG207268H, CWG206384H, CWG206082H, CWG207616H, CWG209095H, CWG209520H, CWG210200H, CWG208561H, CWG206528H, CWG206951H, CWG207370H, CWG207477H, CWG208058H, CWG208176H, CWG210788H, CWG211218H, CWG212599H, CWG213399H, CWG213434H, CWG213467H, CWG213927H, CWG214418H, CWG208571H, CWG210957H, CWG208586H, CWG208791H, CWG208860H, CWG209327H, CWG209455H, CWG213160H, CWG205225H, CWG207373H, CWG210051H, CWG210346H, CWG213087H, CWG213198H, CWG210909H, CWG205190H, CWG205296H, CWG205876H, CWG206149H, CWG207478H, CWG208553H, CWG209099H, CWG210299H, CWG211355H, CWG212104H, CWG212775H, CWG213131H, CWG210064H, CWG211730H, CWG212542H, CWG212825H, CWG206408H, CWG212680H, CWG206993H, CWG214388H, CWG205517H, CWG205879H, CWG206444H, CWG206467H, CWG206491H, CWG206966H, CWG207045H, CWG207649H, CWG209368H, CWG211508H, CWG211820H, CWG212711H, CWG212932H, CWG213146H, CWG213844H, CWG213983H, CWG214245H, CWG214369H, CWG214385H, CWG205946H, CWG208034H, CWG208568H, CWG209585H, CWG213563H, CWG211646H, CWG211743H, CWG213384H, CWG205175H, CWG210187H, CWG210354H, CWG210681H, CWG210750H, CWG210913H, CWG212224H, CWG212231H, CWG212861H, CWG207166H, CWG213453H, CWG210832H, CWG206977H, CWG208297H, CWG209123H, CWG205097H, CWG207305H, CWG206955H, CWG211504H, CWG212166H, CWG209886H, CWG211482H, CWG210574H, CWG205911H, CWG206938H, CWG213193H, CWG207119H, CWG213218H, CWG213232H, CWG213901H, CWG213007H, CWG209524H, CWG205816H, CWG207100H, CWG207214H, CWG208602H, CWG209275H, CWG210918H, CWG211246H, CWG210866H, CWG210366H, CWG208207H, CWG211349H, CWG212946H, CWG206889H, CWG207040H, CWG209171H, CWG213877H, CWG205346H, CWG206195H, CWG206197H, CWG214005H, CWG206512H, CWG207318H, CWG210364H, CWG211087H, CWG213975H, CWG214075H, CWG211277H, CWG210331H, CWG210504H, CWG210600H, CWG209958H, CWG214142H, CWG207116H, CWG205436H, CWG209972H, CWG206457H, CWG207193H, CWG207636H, CWG208636H, CWG206319H, CWG207492H, CWG205152H, CWG205693H, CWG205993H, CWG206257H, CWG206742H, CWG208049H, CWG208498H, CWG209126H, CWG209184H, CWG210210H, CWG210466H, CWG210590H, CWG210699H, CWG211244H, CWG211415H, CWG212185H, CWG212432H, CWG212725H, CWG213711H, CWG212556H, CWG205860H, CWG206715H, CWG207312H, CWG207435H, CWG210806H, CWG210907H, CWG213263H, CWG213746H, CWG213802H, CWG213870H, CWG214233H, CWG214340H, CWG205835H, CWG206536H, CWG206770H, CWG208204H, CWG208288H, CWG211172H, CWG211338H, CWG211659H, CWG213096H, CWG213162H, CWG213981H, CWG213754H, CWG206025H, CWG206250H, CWG206557H, CWG210930H, CWG210931H, CWG211298H, CWG204983H, CWG208819H, CWG208831H, CWG213208H, CWG213911H, CWG206153H, CWG206676H, CWG206958H, CWG207440H, CWG207460H, CWG208371H, CWG208966H, CWG210413H, CWG210575H, CWG211394H, CWG213003H, CWG205752H, CWG208261H, CWG208990H, CWG211173H, CWG213143H, CWG214319H, CWG211095H, CWG210562H, CWG213469H, CWG209545H, CWG206322H, CWG210735H, CWG209473H, CWG212088H, CWG207111H, CWG210904H, CWG213381H, CWG207683H, CWG210217H, CWG214375H, CWG206173H, CWG212676H, CWG206796H, CWG207447H, CWG209201H, CWG210239H, CWG211319H, CWG212753H, CWG212800H, CWG212875H, CWG213341H, CWG213344H, CWG213884H, CWG214433H, CWG208486H, CWG205123H, CWG205958H, CWG206341H, CWG206449H, CWG206555H, CWG207269H, CWG207308H, CWG207580H, CWG208461H, CWG210752H, CWG210915H, CWG211439H, CWG211516H, CWG211698H, CWG211769H, CWG211914H, CWG212602H, CWG207287H, CWG211594H, CWG212036H, CWG212572H, CWG212764H, CWG213970H, CWG208237H, CWG210740H, CWG205861H, CWG206477H, CWG206481H, CWG207356H, CWG209318H, CWG211890H, CWG214042H, CWG206410H, CWG207986H, CWG208214H, CWG209276H, CWG211197H, CWG213168H, CWG213703H, CWG207031H, CWG209480H, CWG211238H, CWG210774H, CWG212356H, CWG209773H, CWG208246H, CWG209166H, CWG213073H, CWG213363H, CWG211464H, CWG212737H, CWG205161H, CWG208090H, CWG210002H, CWG211303H, CWG211736H, CWG212738H, CWG213324H, CWG209705H, CWG211787H, CWG204935H, CWG205059H, CWG206045H, CWG206097H, CWG207341H, CWG207450H, CWG208124H, CWG209260H, CWG210415H, CWG211680H, CWG213226H, CWG214070H, CWG214515H, CWG207148H, CWG211228H, CWG207258H, CWG213257H, CWG209852H, CWG214380H, CWG209577H, CWG210577H, CWG211240H, CWG205664H, CWG213432H, CWG203847H, CWG207094H, CWG205457H, CWG205672H, CWG206017H, CWG206208H, CWG207179H, CWG207900H, CWG209145H, CWG209984H, CWG213051H, CWG213707H, CWG208971H, CWG211491H, CWG212173H, CWG209245H, CWG211115H, CWG211175H, CWG212858H, CWG205730H, CWG205825H, CWG208123H, CWG210231H, CWG210358H, CWG211247H, CWG211599H, CWG211664H, CWG211821H, CWG212125H, CWG213144H, CWG213163H, CWG213829H, CWG213855H, CWG205595H, CWG205796H, CWG206314H, CWG207724H, CWG209846H, CWG210179H, CWG210242H, CWG210523H, CWG212890H, CWG213657H, CWG209129H, CWG209378H, CWG211147H, CWG211306H, CWG209832H, CWG211275H, CWG211438H, CWG211500H, CWG213303H, CWG214467H, CWG205271H, CWG206284H, CWG206309H, CWG207003H, CWG209093H, CWG209782H, CWG210352H, CWG212181H, CWG208379H, CWG208693H, CWG209326H, CWG212050H, CWG213289H, CWG210037H, CWG212727H, CWG213504H, CWG209558H, CWG205975H, CWG210388H, CWG213980H, CWG208053H, CWG210278H, CWG208955H, CWG209788H, CWG211485H, CWG213797H, CWG206402H, CWG205248H, CWG205977H, CWG206164H, CWG208476H, CWG208582H, CWG210573H, CWG212610H, CWG213066H, CWG213264H, CWG213584H, CWG205723H, CWG205941H, CWG206114H, CWG209210H, CWG211109H, CWG212791H, CWG212859H, CWG213682H, CWG214251H, CWG205588H, CWG206046H, CWG206085H, CWG206825H, CWG207896H, CWG207897H, CWG208363H, CWG208793H, CWG208794H, CWG208861H, CWG210932H, CWG211180H, CWG211591H, CWG211925H, CWG212862H, CWG213206H, CWG213614H, CWG211127H, CWG213280H, CWG205265H, CWG205673H, CWG205863H, CWG206437H, CWG207913H, CWG209493H, CWG210743H, CWG211490H, CWG211729H, CWG212704H, CWG205372H, CWG205705H, CWG205722H, CWG205964H, CWG205965H, CWG205967H, CWG205968H, CWG207729H, CWG207968H, CWG208421H, CWG208441H, CWG208458H, CWG208560H, CWG208627H, CWG209563H, CWG209698H, CWG210269H, CWG210272H, CWG210274H, CWG210276H, CWG210277H, CWG210280H, CWG210284H, CWG211697H, CWG206561H, CWG206693H, CWG206831H, CWG206894H, CWG210024H, CWG212240H, CWG213200H, CWG213202H, CWG207591H, CWG209963H, CWG209130H, CWG214031H, CWG211361H, CWG212216H, CWG212591H, CWG213161H, CWG208684H, CWG210468H, CWG211865H, CWG213782H, CWG206143H, CWG206237H, CWG206331H, CWG206717H, CWG207076H, CWG207277H, CWG207395H, CWG207475H, CWG210353H, CWG210488H, CWG210512H, CWG210559H, CWG211199H, CWG211200H, CWG211231H, CWG211254H, CWG212955H, CWG213082H, CWG213317H, CWG214041H, CWG210060H, CWG206241H, CWG208249H, CWG213126H, CWG213491H, CWG213054H, CWG213351H, CWG213354H, CWG206458H, CWG208534H, CWG209684H, CWG210178H, CWG211996H, CWG214077H, CWG205263H, CWG209929H, CWG209185H, CWG210305H, CWG212902H, CWG214256H, CWG212597H, CWG214262H, CWG205670H, CWG206625H, CWG206138H, CWG207568H, CWG207800H, CWG208306H, CWG210708H, CWG212044H, CWG212448H, CWG207983H, CWG212389H, CWG212390H, CWG212395H, CWG212399H, CWG212401H, CWG212402H, CWG212885H, CWG213360H, CWG213708H, CWG213713H, CWG213714H, CWG213715H, CWG213716H, CWG206937H, CWG207163H, CWG210506H, CWG214410H, CWG214426H, CWG206483H, CWG211518H, CWG213117H, CWG205762H, CWG212573H, CWG206121H, CWG208209H, CWG205895H, CWG209170H, CWG209422H, CWG210382H, CWG213246H, CWG213418H, CWG213486H, CWG212453H, CWG205081H, CWG205209H, CWG206542H, CWG207280H, CWG207657H, CWG208690H, CWG209560H, CWG210796H, CWG211998H, CWG213466H, CWG206633H, CWG212548H, CWG212787H, CWG212376H, CWG212378H, CWG212380H, CWG205782H, CWG207350H, CWG209165H, CWG213184H, CWG213892H, CWG211182H, CWG204977H, CWG205119H, CWG205237H, CWG205273H, CWG206227H, CWG207577H, CWG208283H, CWG209140H, CWG210059H, CWG210254H, CWG210264H, CWG213084H, CWG210310H, CWG213083H, CWG208868H, CWG209094H, CWG207980H, CWG209713H, CWG213779H, CWG206165H, CWG207553H, CWG207554H, CWG209370H, CWG207108H, CWG208700H, CWG214444H, CWG210926H, CWG211468H, CWG205655H, CWG205142H, CWG205432H, CWG205453H, CWG205462H, CWG207488H, CWG207582H, CWG207827H, CWG208226H, CWG208229H, CWG208233H, CWG209500H, CWG209898H, CWG209899H, CWG209902H, CWG209904H, CWG209907H, CWG209912H, CWG209915H, CWG209919H, CWG209921H, CWG209922H, CWG210435H, CWG210447H, CWG211129H, CWG211720H, CWG212684H, CWG213095H, CWG213422H, CWG209497H, CWG207640H, CWG209944H, CWG207161H, CWG207250H, CWG210843H, CWG210937H, CWG211534H, CWG211614H, CWG205287H, CWG205487H, CWG205491H, CWG205493H, CWG205499H, CWG205502H, CWG205504H, CWG205509H, CWG205511H, CWG205512H, CWG206596H, CWG206597H, CWG206598H, CWG206599H, CWG206603H, CWG206604H, CWG207596H, CWG207597H, CWG207598H, CWG207601H, CWG207602H, CWG207604H, CWG207605H, CWG210089H, CWG210090H, CWG210091H, CWG210092H, CWG210094H, CWG210096H, CWG210097H, CWG210098H, CWG210099H, CWG210100H, CWG210101H, CWG210647H, CWG210648H, CWG210663H, CWG210722H, CWG210723H, CWG210820H, CWG212391H, CWG213667H, CWG205521H, CWG211170H, CWG213065H, CWG205586H, CWG206985H, CWG210119H, CWG210783H, CWG211142H, CWG211305H, CWG211572H, CWG213069H, CWG209084H, CWG211421H, CWG213294H, CWG207351H, CWG211725H, CWG212381H, CWG212382H, CWG212385H, CWG212386H, CWG212388H, CWG213705H, CWG213706H, CWG205817H, CWG207432H, CWG209479H, CWG205100H, CWG205650H, CWG205755H, CWG205931H, CWG205940H, CWG205942H, CWG206327H, CWG207115H, CWG207176H, CWG208899H, CWG208901H, CWG208935H, CWG209018H, CWG209021H, CWG209322H, CWG209323H, CWG209888H, CWG209889H, CWG209890H, CWG209891H, CWG209892H, CWG210105H, CWG210106H, CWG210107H, CWG210108H, CWG210109H, CWG210117H, CWG210120H, CWG210123H, CWG210327H, CWG211284H, CWG211543H, CWG211596H, CWG211653H, CWG211844H, CWG212300H, CWG212303H, CWG212305H, CWG212367H, CWG212414H, CWG212598H, CWG213016H, CWG213063H, CWG213133H, CWG213498H, CWG213630H, CWG213631H, CWG213633H, CWG213659H, CWG213677H, CWG213680H, CWG214350H, CWG214458H, CWG204973H, CWG207294H, CWG207302H, CWG207768H, CWG207769H, CWG207773H, CWG207774H, CWG208254H, CWG209894H, CWG210336H, CWG211829H, CWG211831H, CWG212046H, CWG212373H, CWG212482H, CWG212803H, CWG213472H, CWG213480H, CWG207638H, CWG205330H, CWG205331H, CWG206231H, CWG206232H, CWG206233H, CWG206496H, CWG206743H, CWG206744H, CWG206745H, CWG207482H, CWG207483H, CWG207484H, CWG207485H, CWG208220H, CWG208221H, CWG208645H, CWG208824H, CWG208870H, CWG209036H, CWG209037H, CWG209038H, CWG209042H, CWG209047H, CWG209648H, CWG209651H, CWG209652H, CWG209653H, CWG209654H, CWG209655H, CWG209656H, CWG211550H, CWG211551H, CWG211552H, CWG212506H, CWG212508H, CWG212510H, CWG212511H, CWG212513H, CWG212514H, CWG212967H, CWG213211H, CWG213212H, CWG213213H, CWG213377H, CWG213461H, CWG213462H, CWG213463H, CWG213464H, CWG214288H, CWG214289H, CWG214290H, CWG214292H, CWG214293H, CWG214294H, CWG214295H, CWG214296H, CWG214297H, CWG214298H, CWG214454H, CWG214495H, CWG214497H, CWG209562H, CWG204940H, CWG204941H, CWG204950H, CWG204951H, CWG204952H, CWG204953H, CWG204954H, CWG204955H, CWG204956H, CWG204957H, CWG205243H, CWG205244H, CWG205245H, CWG205247H, CWG205249H, CWG205252H, CWG205253H, CWG205254H, CWG207399H, CWG207400H, CWG207414H, CWG207415H, CWG207418H, CWG207420H, CWG207514H, CWG207515H, CWG207516H, CWG207517H, CWG207518H, CWG207521H, CWG207522H, CWG207524H, CWG207526H, CWG207736H, CWG207849H, CWG207853H, CWG207855H, CWG207856H, CWG207857H, CWG207861H, CWG207864H, CWG207867H, CWG207868H, CWG207870H, CWG207874H, CWG207875H, CWG207877H, CWG207878H, CWG208210H, CWG209183H, CWG209554H, CWG209965H, CWG210031H, CWG210032H, CWG210033H, CWG210035H, CWG210036H, CWG210038H, CWG210513H, CWG210619H, CWG210724H, CWG210726H, CWG210727H, CWG210764H, CWG210814H, CWG210816H, CWG210961H, CWG210962H, CWG210964H, CWG210966H, CWG210967H, CWG211025H, CWG211623H, CWG211624H, CWG211626H, CWG211631H, CWG211632H, CWG211634H, CWG211635H, CWG211636H, CWG211637H, CWG211648H, CWG212141H, CWG212142H, CWG212143H, CWG212146H, CWG212150H, CWG212161H, CWG212254H, CWG212329H, CWG212330H, CWG212331H, CWG212333H, CWG212786H, CWG212820H, CWG212827H, CWG212828H, CWG212832H, CWG213532H, CWG213535H, CWG213536H, CWG213537H, CWG213538H, CWG213539H, CWG213689H, CWG213690H, CWG213691H, CWG213693H, CWG213694H, CWG214029H, CWG214489H, CWG214490H, CWG214491H, CWG206200H, CWG207321H, CWG207345H, CWG208287H, CWG209644H, CWG208158H, CWG208276H, CWG209176H, CWG209384H, CWG209800H, CWG210509H, CWG213419H, CWG214238H, CWG208959H, CWG206426H, CWG211227H, CWG211385H, CWG209795H, CWG211103H, CWG211104H, CWG211106H, CWG212112H, CWG212063H, CWG210295H, CWG210888H, CWG211567H, CWG212017H, CWG212340H, CWG212974H, CWG210566H, CWG210567H, CWG210568H, CWG210570H, CWG210754H, CWG210758H, CWG212525H, CWG212526H, CWG213580H, CWG213583H, CWG214203H, CWG214204H, CWG214205H, CWG212374H, CWG206762H, CWG207653H, CWG208298H, CWG211348H, CWG211644H, CWG211803H, CWG211960H, CWG211988H, CWG212169H, CWG213719H, CWG212403H, CWG212404H, CWG212407H, CWG212409H, CWG212417H, CWG212425H, CWG213695H, CWG213698H, CWG212118H, CWG213985H, CWG205559H, CWG205587H, CWG205728H, CWG205852H, CWG205926H, CWG206134H, CWG206186H, CWG206248H, CWG206336H, CWG206346H, CWG206369H, CWG206656H, CWG206720H, CWG206834H, CWG206973H, CWG206976H, CWG206997H, CWG207314H, CWG207315H, CWG207381H, CWG207402H, CWG207508H, CWG207519H, CWG207572H, CWG207614H, CWG207665H, CWG207666H, CWG208576H, CWG208595H, CWG208599H, CWG208600H, CWG208655H, CWG209039H, CWG209065H, CWG209175H, CWG209226H, CWG209266H, CWG209334H, CWG209341H, CWG209498H, CWG209587H, CWG209681H, CWG209704H, CWG209720H, CWG209735H, CWG209760H, CWG209771H, CWG209825H, CWG209909H, CWG209918H, CWG209934H, CWG209950H, CWG209957H, CWG210006H, CWG210069H, CWG210283H, CWG210471H, CWG210516H, CWG211038H, CWG211045H, CWG211362H, CWG211368H, CWG211372H, CWG211411H, CWG211607H, CWG211657H, CWG211727H, CWG211932H, CWG212212H, CWG212384H, CWG212393H, CWG212420H, CWG212668H, CWG212724H, CWG212735H, CWG212962H, CWG212972H, CWG212973H, CWG212999H, CWG213046H, CWG213060H, CWG213064H, CWG213353H, CWG213364H, CWG213386H, CWG213553H, CWG213554H, CWG213566H, CWG213576H, CWG213581H, CWG213582H, CWG213585H, CWG213848H, CWG213930H, CWG214167H, CWG214177H, CWG214180H, CWG214181H, CWG214185H, CWG214189H, CWG214191H, CWG214192H, CWG214200H, CWG214206H, CWG214231H, CWG214302H, CWG214304H, CWG214308H, CWG214309H, CWG214310H, CWG214311H, CWG214313H, CWG214314H, CWG214391H, CWG214396H, CWG214459H, CWG214472H, CWG214475H, CWG214480H, CWG214496H, CWG214395H; GTIN 00763000059576, Lot Serial Numbers: CWG601901S, CWG600080S, CWG605202S, CWG601189S, CWG602331S, CWG603128S, CWG603192S, CWG604372S, CWG602791S, CWG600676S, CWG601450S, CWG605243S, CWG605244S, CWG602890S, CWG603367S, CWG603368S, CWG603369S, CWG603371S, CWG603375S, CWG605555S, CWG605556S, CWG602601S, CWG602683S, CWG601159S, CWG604025S, CWG604431S, CWG601290S, CWG601302S, CWG601644S, CWG602120S, CWG602402S, CWG602563S, CWG602776S, CWG603446S, CWG603886S, CWG603887S, CWG603888S, CWG603897S, CWG604255S, CWG605856S, CWG601020S, CWG605121S, CWG600023S, CWG600028S, CWG600029S, CWG600030S, CWG600031S, CWG602566S, CWG602933S, CWG602185S, CWG603376S, CWG603723S, CWG604609S, CWG604979S, CWG603198S, CWG605656S, CWG600318S, CWG600402S, CWG601766S, CWG603261S, CWG603378S, CWG603880S, CWG605296S, CWG605928S, CWG600441S, CWG605483S, CWG605207S, CWG600158S, CWG600233S, CWG600700S, CWG600949S, CWG601030S, CWG601458S, CWG602229S, CWG602686S, CWG602789S, CWG602920S, CWG603008S, CWG603208S, CWG603641S, CWG603826S, CWG604000S, CWG604398S, CWG604462S, CWG604478S, CWG605365S, CWG605697S, CWG600487S, CWG600494S, CWG600498S, CWG600521S, CWG600525S, CWG604189S, CWG604823S, CWG605232S, CWG605319S, CWG605554S, CWG600520S, CWG600626S, CWG600895S, CWG601179S, CWG601856S, CWG602710S, CWG603068S, CWG603135S, CWG603227S, CWG603521S, CWG603698S, CWG603789S, CWG604005S, CWG605822S, CWG600066S, CWG600643S, CWG600814S, CWG602084S, CWG602892S, CWG604445S, CWG600507S, CWG601207S, CWG602093S, CWG604826S, CWG605343S, CWG604108S, CWG605262S, CWG600682S, CWG601235S, CWG601323S, CWG601683S, CWG601001S, CWG601347S, CWG603892S, CWG603372S, CWG604008S, CWG604040S, CWG600037S, CWG600899S, CWG601221S, CWG601435S, CWG605673S, CWG605674S, CWG600006S, CWG600406S, CWG600417S, CWG600597S, CWG600777S, CWG600803S, CWG601360S, CWG601492S, CWG601725S, CWG601815S, CWG602230S, CWG602369S, CWG602412S, CWG602430S, CWG602619S, CWG602649S, CWG602792S, CWG602956S, CWG603004S, CWG603343S, CWG603350S, CWG603592S, CWG604003S, CWG605746S, CWG602404S, CWG601554S, CWG603856S, CWG606009S, CWG601015S, CWG601054S, CWG603038S, CWG604350S, CWG601591S, CWG602003S, CWG602087S, CWG603139S, CWG603520S, CWG604099S, CWG605080S, CWG601710S, CWG601957S, CWG603217S, CWG601483S, CWG605969S, CWG605550S, CWG600451S, CWG600572S, CWG603476S, CWG603692S, CWG603693S, CWG605585S, CWG605587S, CWG605588S, CWG600527S, CWG601171S, CWG601501S, CWG601733S, CWG601040S, CWG604921S, CWG600007S, CWG600945S, CWG601060S, CWG601261S, CWG602516S, CWG602795S, CWG602797S, CWG602937S, CWG603967S, CWG604047S, CWG600780S, CWG600786S, CWG600486S, CWG600532S, CWG600852S, CWG600858S, CWG601955S, CWG603254S, CWG605431S, CWG605438S, CWG602524S, CWG602967S, CWG603196S, CWG603541S, CWG603981S, CWG603986S, CWG604464S, CWG604466S, CWG604468S, CWG604572S, CWG601953S, CWG602103S, CWG602469S, CWG600330S, CWG601474S, CWG601881S, CWG602357S, CWG602855S, CWG605116S, CWG605439S, CWG600466S, CWG601277S, CWG602223S, CWG602867S, CWG603102S, CWG603883S, CWG604045S, CWG604218S, CWG604604S, CWG604729S, CWG601078S, CWG601308S, CWG601635S, CWG601689S, CWG602341S, CWG602758S, CWG602941S, CWG603165S, CWG603491S, CWG604043S, CWG604357S, CWG604011S, CWG600049S, CWG600050S, CWG601659S, CWG601741S, CWG603052S, CWG603248S, CWG603386S, CWG603524S, CWG603813S, CWG603955S, CWG600495S, CWG600869S, CWG600154S, CWG600938S, CWG601174S, CWG601359S, CWG601871S, CWG600979S, CWG601098S, CWG602326S, CWG604577S, CWG600163S, CWG600312S, CWG602075S, CWG602102S, CWG603273S, CWG603877S, CWG603922S, CWG604084S, CWG604452S, CWG602245S, CWG602321S, CWG602322S, CWG602653S, CWG604963S, CWG600419S, CWG600800S, CWG601533S, CWG602264S, CWG603281S, CWG601454S, CWG602401S, CWG602845S, CWG603152S, CWG604353S, CWG604457S, CWG601495S, CWG601065S, CWG601698S, CWG602530S, CWG602975S, CWG604257S, CWG604923S, CWG605142S, CWG603792S, CWG604097S, CWG604241S, CWG600064S, CWG600065S, CWG601142S, CWG606018S, CWG606019S, CWG600559S, CWG600757S, CWG601541S, CWG601990S, CWG602465S, CWG602472S, CWG602473S, CWG603310S, CWG605791S, CWG602164S, CWG603619S, CWG604752S, CWG602851S, CWG604971S, CWG605534S, CWG600206S, CWG601135S, CWG602547S, CWG602841S, CWG603282S, CWG603515S, CWG603569S, CWG603905S, CWG604102S, CWG604170S, CWG604212S, CWG604421S, CWG603462S, CWG604513S, CWG601615S, CWG600272S, CWG600779S, CWG602050S, CWG603084S, CWG602428S, CWG600569S, CWG601631S, CWG602934S, CWG603750S, CWG605330S, CWG601826S, CWG602066S, CWG602214S, CWG602300S, CWG605655S, CWG600245S, CWG600139S, CWG600140S, CWG600874S, CWG600876S, CWG600906S, CWG604527S, CWG604783S, CWG604871S, CWG600141S, CWG600959S, CWG600961S, CWG601416S, CWG601519S, CWG601521S, CWG602208S, CWG602268S, CWG602346S, CWG602352S, CWG603015S, CWG603294S, CWG603295S, CWG603357S, CWG603379S, CWG604030S, CWG604031S, CWG604185S, CWG604430S, CWG604469S, CWG604470S, CWG600633S, CWG601594S, CWG602015S, CWG602475S, CWG601029S, CWG602951S, CWG603078S, CWG604559S, CWG601865S, CWG604869S, CWG602680S, CWG604374S, CWG601417S, CWG601636S, CWG601867S, CWG601651S, CWG601975S, CWG602592S, CWG602974S, CWG603513S, CWG605488S, CWG604548S, CWG602794S, CWG603059S, CWG602764S, CWG604384S, CWG603005S, CWG602591S, CWG600619S, CWG600818S, CWG605352S, CWG600341S, CWG602026S, CWG602454S, CWG602729S, CWG602828S, CWG602844S, CWG604805S, CWG604443S, CWG601314S, CWG604588S, CWG604539S, CWG604717S, CWG600973S, CWG602837S, CWG603724S, CWG600177S, CWG603805S, CWG603861S, CWG605119S, CWG605367S, CWG601834S, CWG603742S, CWG604864S, CWG602170S, CWG603088S, CWG605095S, CWG605578S, CWG600136S, CWG600410S, CWG600554S, CWG600628S, CWG601049S, CWG602165S, CWG602299S, CWG602588S, CWG602597S, CWG602670S, CWG602823S, CWG603242S, CWG602508S, CWG603012S, CWG604866S, CWG605850S, CWG603548S, CWG602463S, CWG602773S, CWG600632S, CWG602486S, CWG605566S, CWG600440S, CWG600469S, CWG600957S, CWG601257S, CWG601984S, CWG602173S, CWG604064S, CWG600918S, CWG603652S, CWG604178S, CWG602846S, CWG602418S, CWG604379S, CWG602833S, CWG603011S, CWG603306S, CWG604182S, CWG604360S, CWG602445S, CWG601075S, CWG603075S, CWG604658S, CWG601687S, CWG602944S, CWG603450S, CWG604424S, CWG603370S, CWG602080S, CWG603082S, CWG604324S, CWG605794S, CWG605832S, CWG600654S, CWG600699S, CWG600261S, CWG600366S, CWG600427S, CWG600857S, CWG600965S, CWG600993S, CWG601068S, CWG601274S, CWG601318S, CWG601358S, CWG601400S, CWG601410S, CWG601420S, CWG601497S, CWG601794S, CWG601859S, CWG601909S, CWG601911S, CWG602040S, CWG602101S, CWG602109S, CWG602413S, CWG603630S, CWG604655S, CWG604688S, CWG604904S, CWG604927S, CWG604937S, CWG605048S, CWG605140S, CWG605167S, CWG605209S, CWG605211S, CWG605407S, CWG605413S, CWG601873S, CWG602439S, CWG602760S, CWG603298S, CWG603307S, CWG603311S, CWG603339S, CWG603353S, CWG603467S, CWG603522S, CWG603594S, CWG603604S, CWG603606S, CWG603780S, CWG603783S, CWG603827S, CWG603917S, CWG603918S, CWG603923S, CWG603928S, CWG604027S, CWG604028S, CWG604032S, CWG604033S, CWG604065S, CWG604136S, CWG604222S, CWG604238S, CWG604441S, CWG604451S, CWG604471S, CWG604643S, CWG605045S, CWG605046S, CWG602625S, CWG600010S, CWG600151S, CWG604535S, CWG604536S, CWG604490S, CWG601286S, CWG603216S, CWG601211S, CWG604054S, CWG604689S, CWG604690S, CWG605336S, CWG602044S, CWG602647S, CWG603036S, CWG603037S, CWG603399S, CWG605261S, CWG603787S, CWG600042S, CWG604711S, CWG601208S, CWG604146S, CWG600099S, CWG600539S, CWG600620S, CWG600902S, CWG600909S, CWG600997S, CWG601216S, CWG601493S, CWG601523S, CWG602014S, CWG603551S, CWG604479S, CWG604988S, CWG604952S, CWG600963S, CWG601206S, CWG601224S, CWG601233S, CWG602169S, CWG602250S, CWG602251S, CWG602850S, CWG603016S, CWG603017S, CWG603497S, CWG603574S, CWG603575S, CWG603576S, CWG603602S, CWG602861S, CWG605021S, CWG602082S, CWG602642S, CWG601462S, CWG605052S, CWG605544S, CWG605809S, CWG604706S, CWG605198S, CWG605252S, CWG604270S, CWG600601S, CWG601248S, CWG600523S, CWG600548S, CWG601085S, CWG601267S, CWG601623S, CWG601770S, CWG602345S, CWG602458S, CWG603224S, CWG603562S, CWG604041S, CWG604060S, CWG604130S, CWG604506S, CWG603779S, CWG602987S, CWG602484S, CWG603771S, CWG604725S, CWG600929S, CWG601570S, CWG603289S, CWG603290S, CWG603747S, CWG603748S, CWG603749S, CWG605175S, CWG605176S, CWG605177S, CWG605178S, CWG605179S, CWG605384S, CWG605385S, CWG605386S, CWG605387S, CWG605388S, CWG605596S, CWG605597S, CWG605599S, CWG605600S, CWG605601S, CWG605947S, CWG605948S, CWG605950S, CWG600454S, CWG601284S, CWG601292S, CWG601317S, CWG601663S, CWG601796S, CWG601797S, CWG601799S, CWG602079S, CWG602556S, CWG602730S, CWG602803S, CWG602804S, CWG602805S, CWG602807S, CWG602809S, CWG602905S, CWG602983S, CWG602985S, CWG602997S, CWG603166S, CWG603167S, CWG603169S, CWG603170S, CWG603171S, CWG603836S, CWG603838S, CWG603839S, CWG603841S, CWG604049S, CWG604051S, CWG604053S, CWG604258S, CWG604259S, CWG604261S, CWG604262S, CWG605945S, CWG605946S, CWG605949S, CWG603946S, CWG604319S, CWG601090S, CWG600230S, CWG600450S, CWG601107S, CWG601548S, CWG603300S, CWG602866S, CWG604260S, CWG603393S, CWG603394S, CWG601627S, CWG602012S, CWG604302S, CWG600302S, CWG604685S, CWG602205S, CWG603757S, CWG600472S, CWG600611S, CWG602350S, CWG602856S, CWG603251S, CWG600892S, CWG602596S, CWG602770S, CWG603919S, CWG605602S, CWG601755S, CWG602557S, CWG602599S, CWG603461S, CWG603795S, CWG603937S, CWG604176S, CWG605643S, CWG605858S, CWG601045S, CWG601414S, CWG602017S, CWG602069S, CWG603026S, CWG603623S, CWG603819S, CWG604494S, CWG604627S, CWG602432S, CWG603536S, CWG603517S, CWG604517S, CWG604744S, CWG605766S, CWG601823S, CWG602195S, CWG603061S, CWG603377S, CWG603943S, CWG604309S, CWG601593S, CWG602263S, CWG603042S, CWG603044S, CWG603385S, CWG603395S, CWG604015S, CWG600886S, CWG600887S, CWG600888S, CWG600889S, CWG600890S, CWG602446S, CWG604363S, CWG604516S, CWG600411S, CWG601303S, CWG602781S, CWG605158S, CWG605268S, CWG605383S, CWG603762S, CWG603791S, CWG606031S, CWG605492S, CWG603801S, CWG600778S, CWG601656S, CWG603665S, CWG604800S, CWG605125S, CWG601402S, CWG601407S, CWG604721S, CWG601568S, CWG604993S, CWG606041S, CWG600508S, CWG604842S, CWG604748S, CWG605778S, CWG604227S, CWG604738S, CWG604740S, CWG601734S, CWG602119S, CWG602604S, CWG603058S, CWG603349S, CWG603452S, CWG603685S, CWG603695S, CWG603739S, CWG603740S, CWG604103S, CWG604128S, CWG604298S, CWG604433S, CWG604435S, CWG604552S, CWG604553S, CWG605784S, CWG600481S, CWG601676S, CWG602062S, CWG604217S, CWG604220S, CWG604297S, CWG600862S, CWG601181S, CWG601281S, CWG602455S, CWG602474S, CWG602739S, CWG603159S, CWG603822S, CWG603664S, CWG603580S, CWG603584S, CWG600608S, CWG604730S, CWG600271S, CWG600327S, CWG600328S, CWG605324S, CWG601818S, CWG602504S, CWG603247S, CWG603327S, CWG604911S, CWG605156S, CWG605429S, CWG605899S, CWG601617S, CWG601653S, CWG602450S, CWG603336S, CWG603337S, CWG603482S, CWG604117S, CWG600303S, CWG600304S, CWG604762S, CWG600519S, CWG600750S, CWG600863S, CWG601750S, CWG602162S, CWG602571S, CWG602711S, CWG602714S, CWG603269S, CWG601305S, CWG602252S, CWG602244S, CWG604712S, CWG604714S, CWG604715S, CWG604716S, CWG605124S, CWG605126S, CWG605128S, CWG605129S, CWG605454S, CWG605542S, CWG605543S, CWG605547S, CWG605625S, CWG605672S, CWG605866S, CWG605867S, CWG603156S, CWG604110S, CWG604223S, CWG604224S, CWG604280S, CWG604437S, CWG604438S, CWG604525S, CWG604530S, CWG604531S, CWG605467S, CWG605868S, CWG605869S, CWG605870S, CWG600545S, CWG600901S, CWG601209S, CWG601212S, CWG601215S, CWG601262S, CWG601352S, CWG601567S, CWG601878S, CWG602433S, CWG602607S, CWG602629S, CWG602630S, CWG602802S, CWG603260S, CWG603627S, CWG603655S, CWG603656S, CWG603997S, CWG604077S, CWG604447S, CWG603668S, CWG605787S, CWG600409S, CWG604620S, CWG604503S, CWG605011S, CWG600431S, CWG601182S, CWG601312S, CWG601691S, CWG602047S, CWG604150S, CWG604537S, CWG604551S, CWG605070S, CWG605285S, CWG604821S, CWG600677S, CWG601485S, CWG601754S, CWG602198S, CWG602257S, CWG602490S, CWG602931S, CWG602932S, CWG603583S, CWG603773S, CWG605405S, CWG602917S, CWG600513S, CWG601101S, CWG601577S, CWG602620S, CWG601429S, CWG601771S, CWG602551S, CWG603957S, CWG600192S, CWG600320S, CWG600442S, CWG600680S, CWG600764S, CWG600799S, CWG600809S, CWG603658S, CWG600725S, CWG600726S, CWG602682S, CWG604934S, CWG604935S, CWG600019S, CWG600547S, CWG604964S, CWG602716S, CWG602718S, CWG600211S, CWG601393S, CWG603207S, CWG603262S, CWG604420S, CWG604388S, CWG603252S, CWG603798S, CWG605300S, CWG602035S, CWG605255S, CWG603763S, CWG605366S, CWG602703S, CWG602821S, CWG605024S, CWG600834S, CWG600835S, CWG601003S, CWG602993S, CWG601576S, CWG602572S, CWG603915S, CWG602782S, CWG600505S, CWG600998S, CWG601010S, CWG603743S, CWG604473S, CWG600323S, CWG601431S, CWG603355S, CWG603433S, CWG600253S, CWG600012S, CWG600150S, CWG600152S, CWG600535S, CWG600580S, CWG600624S, CWG600625S, CWG600861S, CWG601105S, CWG603934S, CWG602207S, CWG605839S, CWG603475S, CWG601977S, CWG603041S, CWG603333S, CWG600345S, CWG601201S, CWG601874S, CWG605302S, CWG605219S, CWG602785S, CWG603961S, CWG605233S, CWG600424S, CWG600810S, CWG601232S, CWG601379S, CWG601637S, CWG604412S, CWG604415S, CWG600447S, CWG600921S, CWG601165S, CWG603297S, CWG604751S, CWG605434S, CWG604070S, CWG600544S, CWG600931S, CWG601639S, CWG601654S, CWG603364S, CWG602494S, CWG603463S, CWG603962S, CWG604500S, CWG605389S, CWG605394S, CWG605395S, CWG605424S, CWG601489S, CWG603043S, CWG603718S, CWG603843S, CWG604089S, CWG600604S, CWG600715S, CWG601128S, CWG601546S, CWG602883S, CWG604012S, CWG604299S, CWG604592S, CWG600975S, CWG603472S, CWG605538S, CWG602593S, CWG601336S, CWG602441S, CWG604161S, CWG602088S, CWG602233S, CWG602640S, CWG600362S, CWG602121S, CWG605748S, CWG600173S, CWG601513S, CWG601544S, CWG601935S, CWG602522S, CWG602692S, CWG603504S, CWG604098S, CWG605471S, CWG604087S, CWG600174S, CWG601532S, CWG602447S, CWG603060S, CWG603238S, CWG600185S, CWG600696S, CWG601399S, CWG602910S, CWG600254S, CWG600255S, CWG600256S, CWG600262S, CWG600209S, CWG600400S, CWG601014S, CWG601504S, CWG601682S, CWG601716S, CWG602060S, CWG602151S, CWG602160S, CWG604763S, CWG603147S, CWG601590S, CWG601836S, CWG601951S, CWG603597S, CWG604290S, CWG600590S, CWG603821S, CWG602661S, CWG604845S, CWG604873S, CWG601063S, CWG601199S, CWG601217S, CWG601237S, CWG601287S, CWG601307S, CWG601338S, CWG601363S, CWG601664S, CWG601926S, CWG602213S, CWG602544S, CWG602546S, CWG602895S, CWG603555S, CWG603645S, CWG603671S, CWG603719S, CWG604300S, CWG604449S, CWG604576S, CWG604844S, CWG605448S, CWG605450S, CWG605320S, CWG605797S, CWG602085S, CWG603544S, CWG600538S, CWG602398S, CWG600587S, CWG600759S, CWG600772S, CWG600908S, CWG601509S, CWG603966S, CWG604132S, CWG604284S, CWG604566S, CWG604638S, CWG600423S, CWG600900S, CWG601246S, CWG601929S, CWG602478S, CWG602482S, CWG603034S, CWG603586S, CWG605184S, CWG603420S, CWG600014S, CWG600016S, CWG600054S, CWG600214S, CWG600234S, CWG600236S, CWG600418S, CWG600420S, CWG600482S, CWG601031S, CWG601241S, CWG601980S, CWG602853S, CWG604405S, CWG604427S, CWG604928S, CWG602924S, CWG605526S, CWG605358S, CWG600576S, CWG601525S, CWG602227S, CWG602829S, CWG601581S, CWG602801S, CWG601345S, CWG602282S, CWG604467S, CWG600435S, CWG604651S, CWG600524S, CWG600954S, CWG603174S, CWG605066S, CWG600291S, CWG600426S, CWG600458S, CWG600602S, CWG600944S, CWG602006S, CWG602991S, CWG603423S, CWG603956S, CWG602132S, CWG603272S, CWG603104S, CWG604764S, CWG600639S, CWG600645S, CWG600839S, CWG601553S, CWG605065S, CWG605068S, CWG602705S, CWG600156S, CWG600517S, CWG600860S, CWG601511S, CWG600264S, CWG604417S, CWG604395S, CWG603753S, CWG603755S, CWG600478S, CWG600766S, CWG600794S, CWG602266S, CWG602606S, CWG602658S, CWG602872S, CWG602886S, CWG602965S, CWG603335S, CWG603959S, CWG603987S, CWG604180S, CWG604183S, CWG604453S, CWG605354S, CWG602097S, CWG601817S, CWG600614S, CWG600646S, CWG600934S, CWG601149S, CWG601178S, CWG600724S, CWG600942S, CWG601931S, CWG602623S, CWG604245S, CWG602618S, CWG602888S, CWG603093S, CWG603782S, CWG604267S, CWG605292S, CWG602114S, CWG602279S, CWG602638S, CWG604072S, CWG601556S, CWG601855S, CWG602078S, CWG602219S, CWG601143S, CWG602089S, CWG602442S, CWG604755S, CWG602141S, CWG601427S, CWG601925S, CWG602727S, CWG602831S, CWG602938S, CWG603669S, CWG600635S, CWG601123S, CWG601792S, CWG601952S, CWG602140S, CWG602394S, CWG602806S, CWG602231S, CWG603045S, CWG603083S, CWG604853S, CWG601082S, CWG601713S, CWG602526S, CWG600207S, CWG601154S, CWG602870S, CWG601190S, CWG602878S, CWG602921S, CWG602663S, CWG603111S, CWG603329S, CWG604642S, CWG604943S, CWG601385S, CWG601456S, CWG602614S, CWG602748S, CWG603076S, CWG603568S, CWG603709S, CWG604135S, CWG604403S, CWG604600S, CWG604095S, CWG605433S, CWG605603S, CWG604684S, CWG605281S, CWG605753S, CWG600465S, CWG601134S, CWG601175S, CWG603340S, CWG600579S, CWG600613S, CWG603977S, CWG600641S, CWG602685S, CWG602709S, CWG603239S, CWG603259S, CWG600995S, CWG601260S, CWG602767S, CWG605297S, CWG605474S, CWG600036S, CWG605456S, CWG601271S, CWG601993S, CWG602157S, CWG603741S, CWG603979S, CWG603947S, CWG601959S, CWG603338S, CWG603847S, CWG604141S, CWG601894S, CWG605359S, CWG603490S, CWG600502S, CWG600798S, CWG600898S, CWG601522S, CWG601560S, CWG601579S, CWG601610S, CWG602249S, CWG603007S, CWG603010S, CWG601188S, CWG600437S, CWG602467S, CWG602832S, CWG602950S, CWG603540S, CWG605138S, CWG605204S, CWG605355S, CWG603707S, CWG604692S, CWG600414S, CWG602470S, CWG600394S, CWG605349S, CWG601759S, CWG604413S, CWG600522S, CWG600674S, CWG601119S, CWG602570S, CWG602624S, CWG600878S, CWG601076S, CWG602415S, CWG602585S, CWG602787S, CWG602901S, CWG602994S, CWG602995S, CWG603144S, CWG603177S, CWG604416S, CWG602228S, CWG600315S, CWG600316S, CWG603317S, CWG603318S, CWG601491S, CWG602999S, CWG603411S, CWG600274S, CWG601510S, CWG600145S, CWG600269S, CWG600425S, CWG600841S, CWG602626S, CWG605939S, CWG605616S, CWG604881S, CWG601170S, CWG601368S, CWG604835S, CWG604951S, CWG605811S, CWG600736S, CWG600742S, CWG600845S, CWG601311S, CWG601793S, CWG601795S, CWG602477S, CWG603587S, CWG603601S, CWG603866S, CWG603873S, CWG604544S, CWG602318S, CWG602234S, CWG600428S, CWG601196S, CWG603631S, CWG602112S, CWG602712S, CWG604377S, CWG604938S, CWG604829S, CWG604832S, CWG604902S, CWG602681S, CWG603380S, CWG604192S, CWG600722S, CWG601129S, CWG602561S, CWG600575S, CWG601445S, CWG602835S, CWG602344S, CWG600629S, CWG600967S, CWG600968S, CWG600990S, CWG602589S, CWG602990S, CWG603143S, CWG603240S, CWG603244S, CWG603732S, CWG604211S, CWG602720S, CWG600987S, CWG603907S, CWG601618S, CWG602309S, CWG602204S, CWG602600S, CWG603810S, CWG601828S, CWG605967S, CWG605160S, CWG601295S, CWG602323S, CWG603930S, CWG604681S, CWG605287S, CWG600398S, CWG601139S, CWG602342S, CWG600731S, CWG600732S, CWG600733S, CWG601452S, CWG601680S, CWG605246S, CWG603751S, CWG600221S, CWG600405S, CWG600496S, CWG604504S, CWG604326S, CWG600210S, CWG600952S, CWG601256S, CWG601503S, CWG601545S, CWG602491S, CWG602876S, CWG602936S, CWG602943S, CWG603274S, CWG601279S, CWG604787S, CWG603794S, CWG600562S, CWG601455S, CWG602736S, CWG604164S, CWG601813S, CWG603547S, CWG601023S, CWG601398S, CWG601918S, CWG602193S, CWG602436S, CWG603105S, CWG603106S, CWG603236S, CWG603730S, CWG604428S, CWG604414S, CWG605859S, CWG606005S, CWG600504S, CWG601050S, CWG600649S, CWG601005S, CWG604492S, CWG605284S, CWG600506S, CWG600571S, CWG600831S, CWG600950S, CWG601195S, CWG601319S, CWG601342S, CWG601430S, CWG602004S, CWG602181S, CWG602374S, CWG602632S, CWG602772S, CWG603204S, CWG603219S, CWG603245S, CWG603427S, CWG603600S, CWG603824S, CWG604144S, CWG604198S, CWG604204S, CWG604608S, CWG604564S, CWG605436S, CWG605551S, CWG605346S, CWG604728S, CWG604780S, CWG605004S, CWG605079S, CWG605430S, CWG600444S, CWG600448S, CWG600658S, CWG600711S, CWG600746S, CWG600875S, CWG601018S, CWG601043S, CWG601488S, CWG601496S, CWG601540S, CWG601677S, CWG601709S, CWG601791S, CWG602013S, CWG602048S, CWG602733S, CWG602857S, CWG602860S, CWG603428S, CWG603444S, CWG603449S, CWG603552S, CWG603666S, CWG604839S, CWG605040S, CWG600670S, CWG600748S, CWG600760S, CWG600804S, CWG601120S, CWG602397S, CWG602527S, CWG602896S, CWG603069S, CWG603232S, CWG603453S, CWG603456S, CWG604594S, CWG605174S, CWG600877S, CWG600933S, CWG600969S, CWG600972S, CWG601228S, CWG601293S, CWG601457S, CWG602578S, CWG602947S, CWG604229S, CWG604419S, CWG601197S, CWG601335S, CWG600847S, CWG601605S, CWG602116S, CWG602314S, CWG604863S, CWG605067S, CWG605092S, CWG605108S, CWG605508S, CWG600718S, CWG601077S, CWG601245S, CWG601269S, CWG601313S, CWG601614S, CWG601723S, CWG601849S, CWG602182S, CWG602448S, CWG603243S, CWG603314S, CWG603481S, CWG603493S, CWG603507S, CWG603770S, CWG604497S, CWG605382S, CWG600404S, CWG600413S, CWG600631S, CWG600668S, CWG600679S, CWG600683S, CWG600684S, CWG600752S, CWG600916S, CWG600917S, CWG601053S, CWG601355S, CWG602270S, CWG603857S, CWG604981S, CWG605445S, CWG600144S, CWG600588S, CWG600596S, CWG600640S, CWG601921S, CWG601941S, CWG602197S, CWG602497S, CWG603270S, CWG601944S, CWG602034S, CWG603206S, CWG601381S, CWG601785S, CWG602813S, CWG603903S, CWG604036S, CWG600222S, CWG602065S, CWG602220S, CWG602840S, CWG602922S, CWG602977S, CWG603691S, CWG604498S, CWG604901S, CWG604709S, CWG604747S, CWG605353S, CWG601356S, CWG603926S, CWG604619S, CWG603046S, CWG601524S, CWG604067S, CWG600175S, CWG603410S, CWG603733S, CWG601272S, CWG602190S, CWG601153S, CWG601002S, CWG601066S, CWG601365S, CWG601552S, CWG603049S, CWG601976S, CWG604985S, CWG603550S, CWG603797S, CWG604397S, CWG605159S, CWG600493S, CWG600672S, CWG601138S, CWG601412S, CWG603120S, CWG603715S, CWG604440S, CWG600712S, CWG600927S, CWG601448S, CWG602167S, CWG602238S, CWG602253S, CWG603571S, CWG604092S, CWG604932S, CWG600157S, CWG602952S, CWG602301S, CWG600958S, CWG601007S, CWG601538S, CWG601649S, CWG601811S, CWG602005S, CWG602188S, CWG602222S, CWG600180S, CWG604101S, CWG603933S, CWG602916S, CWG602708S, CWG604268S, CWG604710S, CWG604739S, CWG600473S, CWG600868S, CWG600881S, CWG601100S, CWG601202S, CWG601227S, CWG601350S, CWG601426S, CWG601810S, CWG602261S, CWG602294S, CWG602361S, CWG602884S, CWG602919S, CWG603033S, CWG603193S, CWG603202S, CWG603388S, CWG603527S, CWG603644S, CWG604547S, CWG604575S, CWG600586S, CWG601584S, CWG601858S, CWG601428S, CWG602179S, CWG602247S, CWG603391S, CWG603939S, CWG602462S, CWG602634S, CWG603995S, CWG604668S, CWG605058S, CWG605147S, CWG601470S, CWG601942S, CWG602022S, CWG602091S, CWG602515S, CWG602690S, CWG602769S, CWG603760S, CWG604075S, CWG604079S, CWG605930S, CWG601061S, CWG603969S, CWG600182S, CWG605662S, CWG601482S, CWG603063S, CWG603419S, CWG603636S, CWG603637S, CWG603867S, CWG605146S, CWG600485S, CWG601939S, CWG602679S, CWG603032S, CWG603405S, CWG603545S, CWG603000S, CWG600622S, CWG600623S, CWG601011S, CWG601339S, CWG601340S, CWG601411S, CWG601436S, CWG601437S, CWG601443S, CWG601526S, CWG602113S, CWG602183S, CWG602274S, CWG602349S, CWG602353S, CWG602403S, CWG602406S, CWG602722S, CWG602724S, CWG602725S, CWG602854S, CWG602979S, CWG604214S, CWG605568S, CWG602628S, CWG604122S, CWG605854S, CWG603458S, CWG604679S, CWG605196S, CWG605479S, CWG605614S, CWG605629S, CWG605635S, CWG600305S, CWG600322S, CWG600329S, CWG600374S, CWG600534S, CWG600599S, CWG600774S, CWG600819S, CWG600948S, CWG601364S, CWG601718S, CWG601843S, CWG602466S, CWG602543S, CWG602562S, CWG602827S, CWG602871S, CWG602899S, CWG602912S, CWG602992S, CWG603056S, CWG603319S, CWG603328S, CWG603330S, CWG603396S, CWG603425S, CWG603442S, CWG603849S, CWG604156S, CWG604207S, CWG604303S, CWG604477S, CWG604361S, CWG602379S, CWG603119S, CWG600422S, CWG601210S, CWG603766S, CWG603768S, CWG600603S, CWG600656S, CWG600675S, CWG602783S, CWG603579S, CWG604063S, CWG604385S, CWG604391S, CWG600167S, CWG601222S, CWG601721S, CWG601829S, CWG602552S, CWG602568S, CWG602651S, CWG603098S, CWG603218S, CWG600477S, CWG600491S, CWG600598S, CWG600617S, CWG600710S, CWG600928S, CWG601088S, CWG601121S, CWG601168S, CWG601253S, CWG601781S, CWG602186S, CWG602389S, CWG602435S, CWG602717S, CWG602775S, CWG603048S, CWG603183S, CWG603214S, CWG603228S, CWG603250S, CWG603302S, CWG603322S, CWG603585S, CWG603868S, CWG603935S, CWG603991S, CWG604206S, CWG604488S, CWG600453S, CWG600526S, CWG600528S, CWG600652S, CWG600866S, CWG600904S, CWG601662S, CWG601883S, CWG602147S, CWG602271S, CWG602302S, CWG602337S, CWG602383S, CWG602518S, CWG602608S, CWG602838S, CWG602908S, CWG602911S, CWG602953S, CWG602973S, CWG603523S, CWG604584S, CWG600452S, CWG600456S, CWG600541S, CWG600577S, CWG600634S, CWG600802S, CWG600910S, CWG601611S, CWG601761S, CWG602453S, CWG602796S, CWG600164S, CWG604786S, CWG605217S, CWG605449S, CWG600457S, CWG600516S, CWG600570S, CWG600637S, CWG600661S, CWG600822S, CWG600828S, CWG600833S, CWG600850S, CWG600956S, CWG601027S, CWG601041S, CWG601056S, CWG601106S, CWG601124S, CWG601291S, CWG601330S, CWG601475S, CWG601812S, CWG601934S, CWG602028S, CWG602196S, CWG602400S, CWG602424S, CWG602485S, CWG602575S, CWG603199S, CWG603257S, CWG603258S, CWG603271S, CWG603292S, CWG603537S, CWG604134S, CWG604459S, CWG605704S, CWG605926S, CWG604304S, CWG604422S, CWG604756S, CWG600563S, CWG601601S, CWG602815S, CWG603382S, CWG604616S, CWG602038S, CWG603519S, CWG604355S, CWG600015S, CWG600549S, CWG600618S, CWG600644S, CWG603859S, CWG603973S, CWG600369S, CWG600403S, CWG600433S, CWG600840S, CWG601562S, CWG603460S, CWG604708S, CWG600439S, CWG600187S, CWG600589S, CWG600709S, CWG601152S, CWG602548S, CWG601087S, CWG603180S, CWG603776S, CWG604523S, CWG605664S, CWG601288S, CWG603109S, CWG604910S, CWG604104S, CWG604266S, CWG600055S, CWG600056S, CWG604974S, CWG601940S, CWG602390S, CWG602425S, CWG602693S, CWG602903S, CWG602906S, CWG603009S, CWG603092S, CWG603230S, CWG603331S, CWG603332S, CWG603567S, CWG604142S, CWG604295S, CWG604574S, CWG604540S, CWG604687S, CWG604697S, CWG600308S, CWG600040S, CWG600479S, CWG602152S, CWG602194S, CWG602613S, CWG604129S, CWG604177S, CWG604278S, CWG604622S, CWG605117S, CWG605118S, CWG605458S, CWG605683S, CWG605684S, CWG601392S, CWG601937S, CWG602652S, CWG603509S, CWG603512S, CWG603799S, CWG603806S, CWG604586S, CWG604587S, CWG604593S, CWG605018S, CWG605154S, CWG605951S, CWG605415S, CWG600666S, CWG600782S, CWG601161S, CWG602020S, CWG602370S, CWG605682S, CWG606032S, CWG602391S, CWG603670S, CWG601705S, CWG602381S, CWG602610S, CWG602788S, CWG601289S, CWG602030S, CWG602408S, CWG604812S, CWG604896S, CWG605015S, CWG600966S, CWG601586S, CWG601746S, CWG601821S, CWG602476S, CWG602747S, CWG603301S, CWG603474S, CWG602422S, CWG600543S, CWG600708S, CWG601640S, CWG601655S, CWG601730S, CWG602298S, CWG602574S, CWG602779S, CWG604423S, CWG602368S, CWG600270S, CWG600314S, CWG600753S, CWG603938S, CWG604394S, CWG604200S, CWG601380S, CWG601384S, CWG601726S, CWG601760S, CWG602135S, CWG602329S, CWG602443S, CWG602771S, CWG603182S, CWG603878S, CWG604337S, CWG604196S, CWG604201S, CWG604323S, CWG604348S, CWG604442S, CWG604731S, CWG604777S, CWG604629S, CWG600663S, CWG601747S, CWG601930S, CWG603003S, CWG604137S, CWG604289S, CWG604362S, CWG604502S, CWG604507S, CWG604635S, CWG605269S, CWG605303S, CWG605379S, CWG605530S, CWG602296S, CWG604336S, CWG604373S, CWG605641S, CWG601612S, CWG603697S, CWG605294S, CWG601903S, CWG602431S, CWG603651S, CWG605769S, CWG604925S, CWG600734S, CWG601351S, CWG602818S, CWG600728S, CWG600855S, CWG600947S, CWG601218S, CWG601451S, CWG601882S, CWG603347S, CWG604664S, CWG601994S, CWG602137S, CWG604352S, CWG604824S, CWG600981S, CWG600982S, CWG601304S, CWG602761S, CWG605393S, CWG600333S, CWG602090S, CWG600737S, CWG604021S, CWG603477S, CWG605280S, CWG602958S, CWG602117S, CWG604986S, CWG601386S, CWG600832S, CWG602511S, CWG602545S, CWG602930S, CWG603528S, CWG600787S, CWG602133S, CWG602569S, CWG602849S, CWG604381S, CWG604661S, CWG600503S, CWG601167S, CWG601866S, CWG602189S, CWG602396S, CWG603237S, CWG603516S, CWG603532S, CWG601536S, CWG602784S, CWG603226S, CWG604686S, CWG604314S, CWG605057S, CWG605348S, CWG600018S, CWG601084S, CWG601464S, CWG603608S, CWG603609S, CWG603614S, CWG603615S, CWG604175S, CWG605350S, CWG605525S, CWG605923S, CWG605977S, CWG600005S, CWG600008S, CWG600251S, CWG601348S, CWG602440S, CWG604380S, CWG604888S, CWG601961S, CWG604768S, CWG604769S, CWG604789S, CWG601039S, CWG601624S, CWG601800S, CWG602262S, CWG602276S, CWG602281S, CWG603348S, CWG603483S, CWG603703S, CWG605645S, CWG600610S, CWG604191S, CWG605037S, CWG602287S, CWG600943S, CWG601527S, CWG600788S, CWG604784S, CWG605581S, CWG605795S, CWG600038S, CWG600043S, CWG600044S, CWG600048S, CWG600191S, CWG600978S, CWG601708S, CWG600134S, CWG603660S, CWG604285S, CWG604907S, CWG605311S, CWG605381S, CWG600776S, CWG600824S, CWG600923S, CWG601021S, CWG601325S, CWG601449S, CWG602124S, CWG602328S, CWG602444S, CWG602698S, CWG602793S, CWG602976S, CWG603246S, CWG603358S, CWG603480S, CWG603485S, CWG603514S, CWG603633S, CWG603788S, CWG604252S, CWG605327S, CWG605328S, CWG605078S, CWG600079S, CWG601840S, CWG602155S, CWG602479S, CWG602535S, CWG603398S, CWG604131S, CWG600583S, CWG601052S, CWG601418S, CWG604699S, CWG600289S, CWG600290S, CWG600293S, CWG600295S, CWG600296S, CWG601916S, CWG602166S, CWG602172S, CWG602502S, CWG603057S, CWG603354S, CWG603356S, CWG603653S, CWG603970S, CWG603971S, CWG603972S, CWG604383S, CWG604411S, CWG605837S, CWG605838S, CWG600607S, CWG601960S, CWG602726S, CWG603593S, CWG603533S, CWG600224S, CWG600208S, CWG602957S, CWG604322S, CWG601268S, CWG603096S, CWG602645S, CWG601696S, CWG603019S, CWG603495S, CWG603777S, CWG603844S, CWG603982S, CWG604683S, CWG605151S, CWG605976S, CWG600309S, CWG603607S, CWG606029S, CWG602311S, CWG600412S, CWG601310S, CWG605237S, CWG604820S, CWG600989S, CWG601130S, CWG600243S, CWG600568S, CWG600627S, CWG600903S, CWG600962S, CWG601016S, CWG601026S, CWG601095S, CWG601118S, CWG601133S, CWG605798S, CWG603538S, CWG601650S, CWG602081S, CWG602894S, CWG601244S, CWG601897S, CWG602029S, CWG604263S, CWG601229S, CWG603097S, CWG600083S, CWG602579S, CWG602968S, CWG603759S, CWG600205S, CWG600492S, CWG600555S, CWG605028S, CWG600186S, CWG600719S, CWG601954S, CWG604796S, CWG601697S, CWG600951S, CWG603118S, CWG600219S, CWG600220S, CWG601242S, CWG601884S, CWG602395S, CWG602489S, CWG602778S, CWG603021S, CWG603643S, CWG604210S, CWG600197S, CWG600200S, CWG600202S, CWG600204S, CWG601518S, CWG602176S, CWG602667S, CWG602673S, CWG602701S, CWG602706S, CWG602707S, CWG605225S, CWG600795S, CWG600796S, CWG603099S, CWG603543S, CWG605723S, CWG605724S, CWG601748S, CWG601978S, CWG603153S, CWG600429S, CWG601471S, CWG603203S, CWG602018S, CWG602049S, CWG604048S, CWG602621S, CWG605636S, CWG601000S, CWG602839S, CWG602897S, CWG604234S, CWG605480S, CWG603869S, CWG605091S, CWG600317S, CWG601753S, CWG603853S, CWG604495S, CWG604578S, CWG604583S, CWG605972S, CWG605978S, CWG601406S, CWG602573S, CWG600870S, CWG602460S, CWG605963S, CWG602468S, CWG601349S, CWG603811S, CWG604874S, CWG604962S, CWG604997S, CWG605270S, CWG601220S, CWG602072S, CWG603406S, CWG603445S, CWG603598S, CWG603611S, CWG603736S, CWG603870S, CWG604387S, CWG604216S, CWG601835S, CWG603408S, CWG603767S, CWG604082S, CWG604667S, CWG604669S, CWG604673S, CWG604674S, CWG604680S, CWG601862S, CWG601864S, CWG604665S, CWG604641S, CWG604798S, CWG604894S, CWG604933S, CWG605029S, CWG605120S, CWG605212S, CWG605228S, CWG605230S, CWG605235S, CWG605347S, CWG605473S, CWG605502S, CWG605646S, CWG605833S, CWG600388S, CWG600648S, CWG600650S, CWG600767S, CWG600775S, CWG600791S, CWG600848S, CWG600867S, CWG600911S, CWG600988S, CWG601151S, CWG601158S, CWG601166S, CWG601193S, CWG601251S, CWG601264S, CWG601297S, CWG601408S, CWG601529S, CWG601563S, CWG601569S, CWG601600S, CWG601609S, CWG601613S, CWG601619S, CWG601638S, CWG601647S, CWG601679S, CWG601711S, CWG601715S, CWG601745S, CWG601768S, CWG601814S, CWG601833S, CWG601904S, CWG601932S, CWG601974S, CWG602021S, CWG602099S, CWG602161S, CWG602259S, CWG602291S, CWG602305S, CWG602324S, CWG602348S, CWG602399S, CWG602434S, CWG602488S, CWG602509S, CWG602538S, CWG602635S, CWG602636S, CWG602655S, CWG602657S, CWG602755S, CWG602762S, CWG603110S, CWG603137S, CWG603334S, CWG603443S, CWG603468S, CWG603508S, CWG603632S, CWG603702S, CWG603958S, CWG604034S, CWG604119S, CWG604120S, CWG604184S, CWG604221S, CWG604557S, CWG605773S, CWG602199S, CWG602891S, CWG603704S, CWG603786S, CWG603860S, CWG604808S, CWG601550S, CWG601630S, CWG602451S, CWG602704S, CWG602875S, CWG602893S, CWG603222S, CWG603231S, CWG603233S, CWG603832S, CWG603921S, CWG604026S, CWG604068S, CWG604151S, CWG604987S, CWG605668S, CWG604393S, CWG601148S, CWG601621S, CWG602143S, CWG604953S, CWG602031S, CWG602737S, CWG601968S, CWG602525S, CWG602616S, CWG601092S, CWG602007S, CWG602290S, CWG603459S, CWG601162S, CWG603599S, CWG603906S, CWG604670S, CWG604700S, CWG604078S, CWG602449S, CWG605634S, CWG605622S, CWG601004S, CWG601266S, CWG601306S, CWG601566S, CWG601819S, CWG606017S, CWG604870S, CWG604895S, CWG605063S, CWG605203S, CWG605569S, CWG605624S, CWG600540S, CWG600594S, CWG601294S, CWG602236S, CWG602267S, CWG602483S, CWG603071S, CWG603366S, CWG603389S, CWG603549S, CWG603648S, CWG604080S, CWG604311S, CWG604618S, CWG604930S, CWG601847S, CWG604915S, CWG601354S, CWG601559S, CWG603455S, CWG600047S, CWG602293S, CWG603089S, CWG603908S, CWG604946S, CWG605001S, CWG605411S, CWG600143S, CWG600531S, CWG600561S, CWG601465S, CWG602687S, CWG602808S, CWG603035S, CWG603288S, CWG603996S, CWG604484S, CWG604585S, CWG604662S, CWG605016S, CWG605425S, CWG605781S, CWG600188S, CWG603168S, CWG604139S, CWG605003S, CWG605012S, CWG601752S, CWG602650S, CWG603929S, CWG604081S, CWG604581S, CWG605970S, CWG600216S, CWG604009S, CWG604039S, CWG602098S, CWG602096S, CWG602495S, CWG604884S, CWG605130S, CWG605341S, CWG605687S, CWG600960S, CWG601177S, CWG601333S, CWG601373S, CWG601658S, CWG602142S, CWG602580S, CWG602583S, CWG602586S, CWG602669S, CWG603126S, CWG603145S, CWG603535S, CWG603667S, CWG603774S, CWG603800S, CWG604073S, CWG602542S, CWG600556S, CWG601028S, CWG605772S, CWG600463S, CWG600716S, CWG600829S, CWG600937S, CWG603932S, CWG605670S, CWG600155S, CWG600235S, CWG600237S, CWG600489S, CWG601042S, CWG604519S, CWG601633S, CWG601634S, CWG602677S, CWG604770S, CWG601706S, CWG604491S, CWG602297S, CWG602959S, CWG603342S, CWG600190S, CWG601595S, CWG604649S, CWG600560S, CWG604942S, CWG605447S, CWG600011S, CWG600361S, CWG600464S, CWG600765S, CWG600773S, CWG600811S, CWG602139S, CWG602996S, CWG603095S, CWG601749S, CWG601906S, CWG603067S, CWG601875S, CWG604091S, CWG604542S, CWG604691S, CWG605101S, CWG601915S, CWG602025S, CWG603116S, CWG603640S, CWG605223S, CWG601200S, CWG601628S, CWG602702S, CWG604701S, CWG604949S, CWG604989S, CWG605314S, CWG605316S, CWG600980S, CWG601322S, CWG601820S, CWG601891S, CWG602590S, CWG602750S, CWG603209S, CWG603590S, CWG603622S, CWG603737S, CWG603848S, CWG604236S, CWG604605S, CWG600720S, CWG603024S, CWG603663S, CWG603711S, CWG601321S, CWG601949S, CWG602206S, CWG604676S, CWG604906S, CWG605592S, CWG605728S, CWG605884S, CWG600756S, CWG600940S, CWG602627S, CWG602864S, CWG602962S, CWG603062S, CWG603115S, CWG604364S, CWG601346S, CWG604066S, CWG600551S, CWG600552S, CWG601472S, CWG601933S, CWG602744S, CWG602972S, CWG605137S, CWG600298S, CWG603790S, CWG600849S, CWG601144S, CWG601145S, CWG602278S, CWG602409S, CWG602505S, CWG602842S, CWG604543S, CWG601185S, CWG604647S, CWG602423S, CWG603162S, CWG605289S, CWG602499S, CWG601969S, CWG605318S, CWG600792S, CWG601516S, CWG601756S, CWG601998S, CWG602032S, CWG602272S, CWG603823S, CWG604486S, CWG600194S, CWG600294S, CWG600430S, CWG603079S, CWG603603S, CWG601885S, CWG601539S, CWG600784S, CWG601460S, CWG604061S, CWG604554S, CWG605139S, CWG605293S, CWG600133S, CWG600137S, CWG601180S, CWG601299S, CWG601515S, CWG602239S, CWG602948S, CWG602980S, CWG603409S, CWG603441S, CWG604549S, CWG604672S, CWG604675S, CWG604955S, CWG604958S, CWG605099S, CWG605239S, CWG605309S, CWG605321S, CWG605361S, CWG605374S, CWG605375S, CWG605552S, CWG605567S, CWG605574S, CWG605607S, CWG600842S, CWG600974S, CWG601024S, CWG601126S, CWG601127S, CWG601157S, CWG601176S, CWG601252S, CWG601258S, CWG601353S, CWG601382S, CWG601383S, CWG601388S, CWG601616S, CWG601643S, CWG601693S, CWG601773S, CWG601775S, CWG601776S, CWG601777S, CWG601778S, CWG601910S, CWG602019S, CWG602058S, CWG602059S, CWG602392S, CWG602411S, CWG602517S, CWG602533S, CWG602587S, CWG603030S, CWG603146S, CWG603160S, CWG603212S, CWG603345S, CWG603591S, CWG603650S, CWG603725S, CWG603729S, CWG604425S, CWG604426S, CWG604520S, CWG604522S, CWG604956S, CWG605308S, CWG604856S, CWG605107S, CWG600449S, CWG604590S, CWG603293S, CWG603944S, CWG604160S, CWG605171S, CWG601714S, CWG601729S, CWG602095S, CWG601341S, CWG605093S, CWG601757S, CWG603846S, CWG603927S, CWG600434S, CWG604732S, CWG600591S, CWG601057S, CWG601575S, CWG601599S, CWG601702S, CWG602100S, CWG602192S, CWG602615S, CWG603029S, CWG603714S, CWG602954S, CWG603107S, CWG600052S, CWG601598S, CWG604109S, CWG605096S, CWG601033S, CWG601093S, CWG601367S, CWG601422S, CWG605638S, CWG602023S, CWG603383S, CWG605498S, CWG605461S, CWG600460S, CWG600557S, CWG600935S, CWG601213S, CWG601254S, CWG601371S, CWG601376S, CWG601490S, CWG601494S, CWG601832S, CWG602285S, CWG602410S, CWG602576S, CWG602684S, CWG602963S, CWG603351S, CWG603352S, CWG603381S, CWG604038S, CWG605680S, CWG603264S, CWG605886S, CWG605887S, CWG604399S, CWG601999S, CWG600537S, CWG604017S, CWG605113S, CWG600299S, CWG600300S, CWG601259S, CWG600743S, CWG600755S, CWG601081S, CWG601555S, CWG602111S, CWG602158S, CWG602177S, CWG602487S, CWG602918S, CWG603210S, CWG603344S, CWG603952S, CWG603965S, CWG603989S, CWG604607S, CWG601972S, CWG603566S, CWG600087S, CWG600088S, CWG600089S, CWG600747S, CWG603392S, CWG603672S, CWG603673S, CWG603675S, CWG603676S, CWG603677S, CWG603678S, CWG603679S, CWG603680S, CWG603682S, CWG603683S, CWG605618S, CWG605621S, CWG600907S, CWG601985S, CWG602073S, CWG602074S, CWG602076S, CWG602235S, CWG602858S, CWG603002S, CWG603542S, CWG605619S, CWG605620S, CWG601751S, CWG602209S, CWG602363S, CWG605475S, CWG605313S, CWG601275S, CWG602009S, CWG603502S, CWG603557S, CWG604406S, CWG601172S, CWG603101S, CWG601822S, CWG603255S, CWG602904S, CWG601920S, CWG600838S, CWG601740S, CWG604735S, CWG601186S, CWG601648S, CWG602041S, CWG602118S, CWG602438S, CWG603578S, CWG604112S, CWG604114S, CWG604630S, CWG604801S, CWG604912S, CWG604961S, CWG605035S, CWG605148S, CWG605633S, CWG601704S, CWG601827S, CWG601854S, CWG601970S, CWG602284S, CWG602347S, CWG602417S, CWG602520S, CWG602555S, CWG602766S, CWG603213S, CWG603492S, CWG603499S, CWG603511S, CWG603690S, CWG604074S, CWG602310S, CWG603256S, CWG602094S, CWG602054S, CWG600035S, CWG600287S, CWG601440S, CWG603275S, CWG603108S, CWG600017S, CWG600915S, CWG601596S, CWG604365S, CWG601852S, CWG602105S, CWG602175S, CWG603565S, CWG601719S, CWG604857S, CWG603699S, CWG604408S, CWG604545S, CWG604333S, CWG605053S, CWG601620S, CWG602288S, CWG605446S, CWG603277S, CWG604977S, CWG605136S, CWG605617S, CWG601589S, CWG603437S, CWG603885S, CWG604803S, CWG604931S, CWG605491S, CWG601641S, CWG601103S, CWG605170S, CWG601357S, CWG602037S, CWG602319S, CWG602869S, CWG603722S, CWG604568S, CWG604454S, CWG605701S, CWG602746S, CWG603077S, CWG605266S, CWG604761S, CWG604972S, CWG602184S, CWG602671S, CWG602863S, CWG602874S, CWG603308S, CWG603581S, CWG603815S, CWG603817S, CWG603945S, CWG604240S, CWG604254S, CWG605741S, CWG605981S, CWG600763S, CWG603796S, CWG603925S, CWG603940S, CWG602110S, CWG600238S, CWG600653S, CWG604013S, CWG604612S, CWG604615S, CWG602481S, CWG600529S, CWG600443S, CWG600574S, CWG600939S, CWG601046S, CWG601652S, CWG604173S, CWG604249S, CWG604277S, CWG605885S, CWG603263S, CWG603628S, CWG600769S, CWG602163S, CWG601924S, CWG602459S, CWG601764S, CWG601720S, CWG603114S, CWG603465S, CWG600836S, CWG601958S, CWG604621S, CWG604743S, CWG601250S, CWG601505S, CWG602255S, CWG601273S, CWG601889S, CWG604703S, CWG604778S, CWG604876S, CWG605631S, CWG604197S, CWG604646S, CWG605060S, CWG604016S, CWG604286S, CWG602560S, CWG600267S, CWG605072S, CWG601282S, CWG602498S, CWG603117S, CWG603363S, CWG600605S, CWG603185S, CWG604090S, CWG602665S, CWG600020S, CWG604802S, CWG604749S, CWG604750S, CWG605604S, CWG601983S, CWG603129S, CWG603189S, CWG603320S, CWG604318S, CWG605906S, CWG600518S, CWG601824S, CWG603175S, CWG603020S, CWG603361S, CWG604476S, CWG605345S, CWG601506S, CWG601507S, CWG603022S, CWG601238S, CWG601917S, CWG602332S, CWG602480S, CWG602757S, CWG603595S, CWG603895S, CWG602010S, CWG604325S, CWG605944S, CWG605789S, CWG600060S, CWG600142S, CWG600585S, CWG600595S, CWG600905S, CWG600996S, CWG601022S, CWG601419S, CWG601444S, CWG601542S, CWG601888S, CWG602216S, CWG602694S, CWG602923S, CWG602955S, CWG603001S, CWG603006S, CWG603074S, CWG603223S, CWG603365S, CWG603384S, CWG603438S, CWG603439S, CWG603642S, CWG604165S, CWG605362S, CWG603054S, CWG604758S, CWG600880S, CWG601375S, CWG603064S, CWG603654S, CWG603713S, CWG601117S, CWG604634S, CWG602691S, CWG605131S, CWG600057S, CWG600062S, CWG600566S, CWG602064S, CWG602364S, CWG602367S, CWG602539S, CWG605132S, CWG605133S, CWG605193S, CWG605194S, CWG600171S, CWG600266S, CWG600343S, CWG600352S, CWG600357S, CWG601801S, CWG601802S, CWG601803S, CWG601804S, CWG601806S, CWG601807S, CWG601808S, CWG601809S, CWG601870S, CWG601879S, CWG601880S, CWG601886S, CWG602928S, CWG603085S, CWG603113S, CWG603125S, CWG603127S, CWG603131S, CWG603197S, CWG605123S, CWG605181S, CWG605183S, CWG605185S, CWG605186S, CWG605187S, CWG605188S, CWG605189S, CWG605191S, CWG605195S, CWG605514S, CWG605515S, CWG605516S, CWG605517S, CWG605518S, CWG605519S, CWG604695S, CWG604696S, CWG600239S, CWG600240S, CWG600241S, CWG600396S, CWG600837S, CWG601695S, CWG601887S, CWG602501S, CWG605714S, CWG603976S, CWG605709S, CWG605711S, CWG605712S, CWG605713S, CWG605825S, CWG605826S, CWG605828S, CWG605829S, CWG605871S, CWG605872S, CWG605873S, CWG605874S, CWG605875S, CWG605876S, CWG605877S, CWG605878S, CWG605879S, CWG605880S, CWG602820S, CWG603132S, CWG604225S, CWG605983S, CWG605404S, CWG604058S, CWG605220S, CWG600581S, CWG600665S, CWG600843S, CWG602862S, CWG605904S, CWG600667S, CWG604720S, CWG601907S, CWG603909S, CWG600432S, CWG600664S, CWG603070S, CWG604759S, CWG600242S, CWG603469S, CWG604791S, CWG604792S, CWG601047S, CWG601163S, CWG601947S, CWG602826S, CWG603091S, CWG600964S, CWG605419S, CWG603051S, CWG605813S, CWG600263S, CWG605788S, CWG601230S, CWG602777S, CWG601606S, CWG603769S, CWG601842S, CWG601982S, CWG602248S, CWG605531S, CWG604838S, CWG604841S, CWG605482S, CWG605648S, CWG600977S, CWG600986S, CWG601048S, CWG601728S, CWG601844S, CWG601848S, CWG601869S, CWG602092S, CWG602159S, CWG602437S, CWG602666S, CWG602978S, CWG602982S, CWG605989S, CWG601285S, CWG605705S, CWG603186S, CWG606013S, CWG603596S, CWG600992S, CWG605164S, CWG601962S, CWG602819S, CWG603040S, CWG604153S, CWG604116S, CWG603151S, CWG605229S, CWG600161S, CWG601072S, CWG602144S, CWG602675S, CWG603457S, CWG603802S, CWG605974S, CWG602221S, CWG604335S, CWG601405S, CWG600360S, CWG605409S, CWG604292S, CWG604296S, CWG604772S, CWG600085S, CWG600735S, CWG603984S, CWG604163S, CWG604561S, CWG601097S, CWG601140S, CWG601239S, CWG602824S, CWG604645S, CWG604705S, CWG602780S, CWG600771S, CWG601816S, CWG602812S, CWG603466S, CWG604024S, CWG600325S, CWG600148S, CWG600705S, CWG600955S, CWG603053S, CWG604014S, CWG600805S, CWG600808S, CWG604671S, CWG600212S, CWG604235S, CWG604401S, CWG603896S, CWG605114S, CWG600438S, CWG602514S, CWG602909S, CWG602913S, CWG603287S, CWG603960S, CWG604461S, CWG605034S, CWG604334S, CWG600823S, CWG600920S, CWG601369S, CWG601439S, CWG601499S, CWG601736S, CWG602086S, CWG601739S, CWG603215S, CWG601146S, CWG602083S, CWG604969S, CWG602122S, CWG603172S, CWG602002S, CWG600288S, CWG601547S, CWG601945S, CWG603039S, CWG603150S, CWG603155S, CWG603163S, CWG601876S, CWG601895S, CWG602289S, CWG604636S, CWG605546S, CWG600558S, CWG600615S, CWG600659S, CWG600741S, CWG600749S, CWG600879S, CWG601131S, CWG601191S, CWG601433S, CWG602180S, CWG602393S, CWG602946S, CWG603181S, CWG603488S, CWG600642S, CWG601086S, CWG601585S, CWG602964S, CWG603362S, CWG600367S, CWG602280S, CWG601684S, CWG604795S, CWG604878S, CWG600499S, CWG601473S, CWG601572S, CWG601574S, CWG601686S, CWG604591S, CWG601531S, CWG602865S, CWG602988S, CWG603200S, CWG603963S, CWG604213S, CWG600994S, CWG602887S, CWG601296S, CWG603201S, CWG600983S, CWG604775S, CWG603556S, CWG603090S, CWG601694S, CWG603689S, CWG601900S, CWG601557S, CWG604733S, CWG600723S, CWG601270S, CWG601841S, CWG603539S, CWG604480S, CWG604765S, CWG600565S, CWG601461S, CWG603066S, CWG603265S, CWG603268S, CWG605539S, CWG600530S, CWG600660S, CWG605036S, CWG605399S, CWG602201S, CWG601603S, CWG602304S, CWG602602S, CWG602603S, CWG602577S, CWG602581S, CWG602594S, CWG602873S, CWG604367S, CWG602742S, CWG605363S, CWG600468S, CWG602503S, CWG602689S, CWG603309S, CWG605055S, CWG601441S, CWG604754S, CWG601017S, CWG601642S, CWG601987S, CWG604950S, CWG601114S, CWG602277S, CWG602315S, CWG602925S, CWG603013S, CWG600169S, CWG603489S, CWG603716S, CWG604831S, CWG604509S, CWG605205S, CWG605451S, CWG605845S, CWG605846S, CWG605847S, CWG605848S, CWG605849S, CWG605936S, CWG605937S, CWG602217S, CWG602416S, CWG602420S, CWG602513S, CWG602148S, CWG604436S, CWG605272S, CWG600976S, CWG601845S, CWG603804S, CWG604115S, CWG605432S, CWG603705S, CWG600391S, CWG601009S, CWG601012S, CWG601019S, CWG601837S, CWG601839S, CWG601838S, CWG602011S, CWG603809S, CWG601413S, CWG601668S, CWG601790S, CWG601943S, CWG603086S, CWG604281S, CWG604283S, CWG604532S, CWG600754S, CWG602713S, CWG601592S, CWG600825S, CWG602243S, CWG602405S, CWG603503S, CWG603721S, CWG605855S, CWG600226S, CWG600331S, CWG600415S, CWG601116S, CWG601132S, CWG601223S, CWG601438S, CWG603023S, CWG603220S, CWG603323S, CWG603432S, CWG605005S, CWG605440S, CWG605726S, CWG605727S, CWG602351S, CWG605020S, CWG600514S, CWG601701S, CWG601896S, CWG602926S, CWG603286S, CWG603891S, CWG604358S, CWG604370S, CWG604599S, CWG605820S, CWG603781S, CWG600160S, CWG601008S, CWG601234S, CWG601481S, CWG601597S, CWG603050S, CWG604625S, CWG601073S, CWG602256S, CWG605455S, CWG602989S, CWG603904S, CWG604007S, CWG600445S, CWG603130S, CWG601625S, CWG602055S, CWG602429S, CWG600356S, CWG602986S, CWG601737S, CWG603624S, CWG604836S, CWG605076S, CWG603684S, CWG605765S, CWG600573S, CWG602001S, CWG602106S, CWG600740S, CWG605942S, CWG601283S, CWG600459S, CWG604877S, CWG602070S, CWG602646S, CWG604472S, CWG600231S, CWG601070S, CWG602659S, CWG601309S, CWG602735S, CWG604482S, CWG604818S, CWG604819S, CWG600785S, CWG602969S, CWG603613S, CWG600153S, CWG600864S, CWG600953S, CWG603710S, CWG605818S, CWG603374S, CWG603978S, CWG604349S, CWG604118S, CWG605882S, CWG604799S, CWG600761S, CWG600894S, CWG603028S, CWG603065S, CWG603531S, CWG603534S, CWG604055S, CWG604172S, CWG604375S, CWG604376S, CWG600467S, CWG601851S, CWG604809S, CWG605457S, CWG605494S, CWG605497S, CWG604312S, CWG605277S, CWG601948S, CWG604737S, CWG604745S, CWG604843S, CWG605222S, CWG605231S, CWG605834S, CWG601038S, CWG601265S, CWG601409S, CWG601467S, CWG601580S, CWG601825S, CWG602016S, CWG602127S, CWG602312S, CWG602317S, CWG602631S, CWG602929S, CWG602981S, CWG603124S, CWG603387S, CWG603501S, CWG603688S, CWG604166S, CWG601366S, CWG602662S, CWG604556S, CWG605168S, CWG600865S, CWG604579S, CWG604959S, CWG605238S, CWG600252S, CWG601164S, CWG602848S, CWG603047S, CWG600533S, CWG603831S, CWG600932S, CWG602817S, CWG601912S, CWG603407S, CWG601632S, CWG601136S, CWG604858S, CWG604282S, CWG602330S, CWG600081S, CWG600228S, CWG601071S, CWG603291S, CWG603161S, CWG601301S, CWG603910S, CWG602617S, CWG603487S, CWG600630S, CWG600821S, CWG600897S, CWG601535S, CWG601660S, CWG602678S, CWG602695S, CWG603149S, CWG604957S, CWG605559S, CWG604889S, CWG604994S, CWG605286S, CWG602740S, CWG602877S, CWG603992S, CWG604123S, CWG604310S, CWG600162S, CWG600621S, CWG600738S, CWG600844S, CWG600896S, CWG600936S, CWG601079S, CWG601374S, CWG601587S, CWG601700S, CWG602024S, CWG602200S, CWG604493S, CWG602053S, CWG602639S, CWG600913S, CWG600567S, CWG600713S, CWG601125S, CWG604138S, CWG604924S, CWG604096S, CWG604456S, CWG602800S, CWG603346S, CWG604663S, CWG604496S, CWG603341S, CWG605677S, CWG603717S, CWG601344S, CWG603899S, CWG602960S, CWG601055S, CWG601064S, CWG604555S, CWG600480S, CWG600681S, CWG603830S, CWG602027S, CWG602456S, CWG605007S, CWG604922S, CWG601661S, CWG604781S, CWG604967S, CWG601403S, CWG601514S, CWG600783S, CWG603176S, CWG602063S, CWG602156S, CWG601551S, CWG603138S, CWG601300S, CWG601424S, CWG602461S, CWG606007S, CWG604734S, CWG603912S, CWG604626S, CWG603094S, CWG604186S, CWG602732S, CWG602940S, CWG605536S, CWG601401S, CWG601486S, CWG602260S, CWG603205S, CWG603253S, CWG603726S, CWG601080S, CWG602656S, CWG601758S, CWG603136S, CWG602493S, CWG603157S, CWG600335S, CWG601214S, CWG601738S, CWG602697S, CWG604822S, CWG601868S, CWG604273S, CWG604499S, CWG600751S, CWG603649S, CWG604168S, CWG605632S, CWG605731S, CWG600816S, CWG601453S, CWG602146S, CWG601320S, CWG601337S, CWG602457S, CWG602531S, CWG600488S, CWG600564S, CWG604666S, CWG604093S, CWG600930S, CWG600678S, CWG600925S, CWG602325S, CWG602715S, CWG604984S, CWG603316S, CWG605660S, CWG604980S, CWG602033S, CWG605892S, CWG600999S, CWG602327S, CWG603573S, CWG600407S, CWG600408S, CWG600550S, CWG600762S, CWG601434S, CWG600248S, CWG603188S, CWG601074S, CWG603875S, CWG604037S, CWG604321S, CWG605577S, CWG605241S, CWG603312S, CWG604682S, CWG604707S, CWG600365S, CWG600584S, CWG600606S, CWG600854S, CWG601096S, CWG601370S, CWG601657S, CWG602174S, CWG602218S, CWG602492S, CWG602507S, CWG604052S, CWG604505S, CWG604521S, CWG605650S, CWG605717S, CWG605771S, CWG602879S, CWG602885S, CWG605337S, CWG604804S, CWG604968S, CWG605042S, CWG605043S, CWG605332S, CWG605780S, CWG600704S, CWG600985S, CWG601059S, CWG601255S, CWG601500S, CWG601565S, CWG601578S, CWG601608S, CWG601646S, CWG601688S, CWG602500S, CWG602564S, CWG602567S, CWG602721S, CWG602738S, CWG602882S, CWG602915S, CWG603194S, CWG603418S, CWG603451S, CWG603812S, CWG603858S, CWG603948S, CWG603985S, CWG605333S, CWG606025S, CWG605240S, CWG602226S, CWG603018S, CWG604042S, CWG600698S, CWG604305S, CWG602889S, CWG604960S, CWG603173S, CWG602598S, CWG604637S, CWG604879S, CWG604898S, CWG604947S, CWG601187S, CWG603998S, CWG604149S, CWG604294S, CWG604329S, CWG604330S, CWG604481S, CWG604908S, CWG602881S, CWG604071S, CWG602565S, CWG605406S, CWG603112S, CWG603141S, CWG603471S, CWG603893S, CWG602303S, CWG602558S, CWG600512S, CWG601226S, CWG601442S, CWG602313S, CWG602521S, CWG603920S, CWG602077S, CWG602836S, CWG600259S, CWG602660S, CWG603397S, CWG602605S, CWG600358S, CWG601991S, CWG602286S, CWG603634S, CWG604402S, CWG604439S, CWG602554S, CWG603191S, CWG604773S, CWG604233S, CWG600313S, CWG601508S, CWG601732S, CWG602134S, CWG604169S, CWG605351S, CWG601362S, CWG603284S, CWG605689S, CWG605589S, CWG600578S, CWG601534S, CWG603894S, CWG602523S, CWG602210S, CWG602307S, CWG602799S, CWG600310S, CWG601102S, CWG601108S, CWG601109S, CWG601110S, CWG601111S, CWG601112S, CWG601113S, CWG602043S, CWG602045S, CWG602612S, CWG602471S, CWG601332S, CWG601463S, CWG602935S, CWG604069S, CWG605663S, CWG605535S, CWG603761S, CWG603916S, CWG600021S, CWG600022S, CWG600024S, CWG600025S, CWG600026S, CWG600027S, CWG601203S, CWG601204S, CWG601205S, CWG601231S, CWG601243S, CWG601315S, CWG601537S, CWG601645S, CWG601762S, CWG601763S, CWG601765S, CWG602056S, CWG602057S, CWG602224S, CWG602232S, CWG602512S, CWG602529S, CWG602532S, CWG602536S, CWG602537S, CWG602749S, CWG602751S, CWG602752S, CWG602753S, CWG602754S, CWG603241S, CWG603279S, CWG603470S, CWG603657S, CWG603863S, CWG603864S, CWG603865S, CWG604100S, CWG604595S, CWG604596S, CWG601727S, CWG602202S, CWG602203S, CWG604882S, CWG602949S, CWG602816S, CWG604905S, CWG604022S, CWG602811S, CWG602559S, CWG602582S, CWG602584S, CWG603954S, CWG602641S, CWG600346S, CWG600368S, CWG600387S, CWG600392S, CWG601767S, CWG601769S, CWG602355S, CWG602786S, CWG604659S, CWG601564S, CWG601582S, CWG602354S, CWG602637S, CWG602768S, CWG603712S, CWG604485S, CWG600321S, CWG605041S, CWG603305S, CWG602123S, CWG602126S, CWG603558S, CWG603560S, CWG603561S, CWG603563S, CWG600377S, CWG603983S, CWG602421S, CWG603402S, CWG603404S, CWG603479S, CWG604404S, CWG604970S, CWG600461S, CWG601805S, CWG602914S, CWG600070S, CWG600071S, CWG600072S, CWG600074S, CWG600075S, CWG600076S, CWG600077S, CWG600078S, CWG600090S, CWG600094S, CWG600095S, CWG600096S, CWG600100S, CWG600101S, CWG600103S, CWG600104S, CWG600105S, CWG600107S, CWG600108S, CWG600109S, CWG600113S, CWG600115S, CWG600116S, CWG600117S, CWG600118S, CWG600119S, CWG600120S, CWG600121S, CWG600123S, CWG600124S, CWG600126S, CWG600128S, CWG600129S, CWG600130S, CWG600131S, CWG600132S, CWG600170S, CWG600277S, CWG600281S, CWG600282S, CWG600285S, CWG600470S, CWG600483S, CWG600484S, CWG600685S, CWG600686S, CWG600687S, CWG600688S, CWG600689S, CWG600690S, CWG600691S, CWG600692S, CWG600693S, CWG600694S, CWG600872S, CWG600873S, CWG601147S, CWG601155S, CWG601160S, CWG601183S, CWG601192S, CWG601326S, CWG601327S, CWG601328S, CWG601394S, CWG601395S, CWG601397S, CWG601665S, CWG601666S, CWG601667S, CWG601669S, CWG601670S, CWG601671S, CWG601672S, CWG601673S, CWG601674S, CWG601675S, CWG601779S, CWG601780S, CWG601783S, CWG601784S, CWG601787S, CWG601789S, CWG601908S, CWG602212S, CWG602241S, CWG602668S, CWG602672S, CWG602763S, CWG602765S, CWG602191S, CWG602927S, CWG602961S, CWG603285S, CWG602664S, CWG601280S, CWG605253S, CWG604508S, CWG600770S, CWG602810S, CWG604797S, CWG604179S, CWG600592S, CWG603081S, CWG603889S, CWG604861S, CWG602153S, CWG602358S, CWG602359S, CWG602549S, CWG602550S, CWG602553S, CWG603744S, CWG603745S, CWG603746S, CWG604936S, CWG605082S, CWG605390S, CWG605487S, CWG605955S, CWG605956S, CWG604694S, CWG605061S, CWG605364S, CWG601044S, CWG602115S, CWG602900S, CWG603234S, CWG604162S, CWG602107S, CWG602150S, CWG602242S, CWG602540S, CWG602541S, CWG602648S, CWG602830S, CWG601013S, CWG603313S, CWG603315S, CWG602464S, CWG600542S, CWG602187S, CWG605356S, CWG604852S, CWG605049S, CWG605100S, CWG601058S, CWG601459S, CWG602125S, CWG602745S, CWG603221S, CWG603412S, CWG603417S, CWG603554S, CWG603588S, CWG603589S, CWG603646S, CWG603854S, CWG604247S, CWG604248S, CWG604301S, CWG604448S, CWG605102S, CWG603100S, CWG604272S, CWG602283S, CWG604807S, CWG600655S, CWG601091S, CWG601156S, CWG601468S, CWG601477S, CWG601478S, CWG601479S, CWG601480S, CWG601857S, CWG601938S, CWG601963S, CWG601967S, CWG601995S, CWG601996S, CWG601997S, CWG602000S, CWG602225S, CWG602320S, CWG602333S, CWG602334S, CWG602335S, CWG602336S, CWG602338S, CWG602366S, CWG602377S, CWG602378S, CWG603413S, CWG603414S, CWG603429S, CWG603430S, CWG603431S, CWG603436S, CWG604339S, CWG604343S, CWG604347S, CWG601743S, CWG603424S, CWG604346S, CWG603415S, CWG600922S, CWG601069S, CWG601219S, CWG601423S, CWG601965S, CWG602129S, CWG602339S, CWG602356S, CWG601512S, CWG600971S, CWG601498S, CWG602365S, CWG602385S, CWG601712S, CWG600609S, CWG601899S, CWG602373S, CWG601390S, CWG601964S, CWG603435S, CWG602699S, CWG604275S, CWG604276S, CWG605523S, CWG606003S, CWG606004S
FEI Number 2182208
Recalling Firm/
Manufacturer
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
8200 Coral Sea St Ne
Mounds View MN 55112-4391
For Additional Information ContactTechnical Services
800-929-4043
Manufacturer Reason
for Recall
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
FDA Determined
Cause 2
Under Investigation by firm
ActionMedtronic issued an URGENT MEDICAL DEVICE CORRECTION NOTICE to its consignees on 05/10/2023 by UPS 2-day delivery or email. Consignees are asked to forward this notification to those who need to be aware within the organization and to any location which the devices have been distributed. The notice explained the problem and the risk and requested the following patient management recommendations: - Prophylactic device replacement is NOT recommended. - Program all HV therapy pathways B>AX in all therapy zones to minimize the risk for this issue. - Prioritize reprogramming patients who have both a history of HV therapy and Rx1 programmed AX>B. - For remaining patients with AX>B programming in any HV therapy sequence, schedule (with appropriate discretion) the next follow-up for in-clinic reprogramming to minimize potential for reduced- or no-energy HV therapies to occur. - Contact Medtronic Technical Services (1-800-929-4043) or your local representative if one of the following is observed as these may be an indication of either a device or lead-related issue: o Reduced- or no-energy HV therapy is displayed in Episode Text (regardless of programmed pathway) o A persistent drop of approximately 50% in RA, RV and LV pacing lead impedance measurements as this may be an indication of increased potential for a future reduced- or no-energy therapy. Update 10/10/2023: Beginning 09/27/2023, Medtronic issued communication via letter to consignees regarding the availability of the revised IFU and SmartSync Software. Consignees are asked to forward this notification to those who need to be aware within the organization and to any location which the devices have been distributed. Consignees are asked to complete and return a confirmation certificate (or equivalent record) confirming they have received the notification.
Quantity in Commerce12099 units
DistributionUS Nationwide - Worldwide Distribution
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
PMA DatabasePMAs with Product Code = LWS
PMAs with Product Code = LWS
-
-