• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 1 Device Recall medtronic

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 1 Device Recall medtronicsee related information
Date Initiated by FirmMay 10, 2023
Date PostedJune 16, 2023
Recall Status1 Open3, Classified
Recall NumberZ-1814-2023
Recall Event ID 92328
PMA NumberP980016S653 
Product Classification Implantable cardioverter defibrillator (non-CRT) - Product Code LWS
ProductICD-VR DVMD3D1 PRIMO MRI, Model Number DVMD3D1; Implantable Cardioverter Defibrillators
Code Information GTIN 00763000001520, Lot Serial Numbers: CWK604305S, CWK601554S, CWK601693S, CWK601762S, CWK603310S, CWK600674S, CWK605298S, CWK605119S, CWK605216S, CWK605052S, CWK605244S, CWK605024S, CWK605176S, CWK605108S, CWK605326S, CWK605015S, CWK605215S, CWK605100S, CWK605123S, CWK604958S, CWK605229S, CWK605502S, CWK604624S, CWK604625S, CWK604396S, CWK605478S, CWK604413S, CWK604933S, CWK605391S, CWK601948S, CWK601949S, CWK601950S, CWK601953S, CWK601961S, CWK601965S, CWK601966S, CWK601970S, CWK601971S, CWK601972S, CWK601974S, CWK601976S, CWK602568S, CWK602617S, CWK602627S, CWK602632S, CWK602691S, CWK602827S, CWK602843S, CWK602870S, CWK601348S, CWK603369S, CWK604247S, CWK604852S, CWK603773S, CWK601726S, CWK601802S, CWK602031S, CWK602325S, CWK602330S, CWK602332S, CWK602336S, CWK602338S, CWK602339S, CWK602355S, CWK602357S, CWK602577S, CWK602645S, CWK602652S, CWK603257S, CWK603258S, CWK603319S, CWK603334S, CWK603926S, CWK603957S, CWK603969S, CWK604118S, CWK604120S, CWK604127S, CWK604307S, CWK604310S, CWK604313S, CWK604321S, CWK605197S, CWK605198S, CWK605207S, CWK605208S, CWK605209S, CWK605210S, CWK605211S, CWK605212S, CWK605449S, CWK605470S, CWK605473S, CWK603235S, CWK603237S, CWK603242S, CWK601913S, CWK602871S, CWK603546S, CWK604544S, CWK604577S, CWK604848S, CWK602774S, CWK603194S, CWK604685S, CWK602551S, CWK604090S, CWK603535S, CWK603951S, CWK601375S, CWK602172S, CWK602462S, CWK602548S, CWK602639S, CWK602643S, CWK602649S, CWK602858S, CWK603058S, CWK603090S, CWK603249S, CWK603289S, CWK603291S, CWK603389S, CWK603437S, CWK603473S, CWK603557S, CWK603727S, CWK603796S, CWK603812S, CWK603923S, CWK603958S, CWK603982S, CWK604107S, CWK604344S, CWK604446S, CWK604543S, CWK604684S, CWK604693S, CWK604845S, CWK604869S, CWK604900S, CWK605126S, CWK605257S, CWK605302S, CWK605396S, CWK605403S, CWK601041S, CWK604110S, CWK602795S, CWK603823S, CWK600763S, CWK601523S, CWK601614S, CWK602461S, CWK602467S, CWK602484S, CWK602723S, CWK603728S, CWK603017S, CWK604545S, CWK600791S, CWK604373S, CWK604362S, CWK601283S, CWK602667S, CWK603581S, CWK603590S, CWK603315S, CWK602456S, CWK603222S, CWK604085S, CWK602110S, CWK604164S, CWK600678S, CWK601465S, CWK600758S, CWK603943S, CWK604574S, CWK604838S, CWK603792S, CWK603852S, CWK602950S, CWK601804S, CWK601915S, CWK602610S, CWK602611S, CWK602743S, CWK602895S, CWK604316S, CWK604634S, CWK602111S, CWK602271S, CWK603221S, CWK603519S, CWK603302S, CWK603729S, CWK603730S, CWK603731S, CWK603224S, CWK605263S, CWK601946S, CWK602931S, CWK603085S, CWK604165S, CWK604167S, CWK603726S, CWK600557S, CWK600889S, CWK601479S, CWK601919S, CWK602103S, CWK602335S, CWK603481S, CWK604368S, CWK602903S, CWK603016S, CWK603390S, CWK603594S, CWK600851S, CWK600419S, CWK600513S, CWK600995S, CWK602121S, CWK603394S, CWK603311S, CWK603579S, CWK604415S, CWK601163S, CWK602354S, CWK602452S, CWK603641S, CWK604602S, CWK601678S, CWK600848S, CWK601482S, CWK602090S, CWK602129S, CWK602273S, CWK602475S, CWK602572S, CWK602914S, CWK603716S, CWK604177S, CWK605463S, CWK600579S, CWK603236S, CWK601359S, CWK601387S, CWK601516S, CWK601566S, CWK601592S, CWK602414S, CWK602469S, CWK602485S, CWK602607S, CWK603014S, CWK603040S, CWK603063S, CWK603081S, CWK603167S, CWK603352S, CWK603355S, CWK603439S, CWK604111S, CWK604128S, CWK604466S, CWK604541S, CWK604598S, CWK604910S, CWK605163S, CWK601079S, CWK601174S, CWK602407S, CWK603411S, CWK603461S, CWK603520S, CWK603800S, CWK603861S, CWK604018S, CWK604029S, CWK604030S, CWK604471S, CWK604473S, CWK604474S, CWK604881S, CWK605007S, CWK605104S, CWK605162S, CWK600295S, CWK600389S, CWK600458S, CWK600484S, CWK602586S, CWK603320S, CWK603324S, CWK603572S, CWK603718S, CWK603733S, CWK604009S, CWK604038S, CWK604087S, CWK604098S, CWK604552S, CWK604571S, CWK604651S, CWK604895S, CWK604929S, CWK602660S, CWK602089S, CWK602114S, CWK602910S, CWK603837S, CWK602664S, CWK604233S, CWK600431S, CWK600481S, CWK603254S, CWK603259S, CWK603697S, CWK603814S, CWK604468S, CWK604586S, CWK604851S, CWK603244S, CWK604121S, CWK604576S, CWK604585S, CWK604780S, CWK605134S, CWK605181S, CWK605182S, CWK601159S, CWK604558S, CWK600635S, CWK600636S, CWK600637S, CWK600638S, CWK600639S, CWK600640S, CWK603176S, CWK603297S, CWK603305S, CWK603878S, CWK604814S, CWK604875S, CWK603053S, CWK600774S, CWK600910S, CWK601307S, CWK600456S, CWK600720S, CWK601015S, CWK601500S, CWK601569S, CWK601662S, CWK602342S, CWK602637S, CWK602647S, CWK602798S, CWK602818S, CWK602819S, CWK602982S, CWK603163S, CWK603283S, CWK603397S, CWK603562S, CWK604505S, CWK604837S, CWK601251S, CWK601718S, CWK600058S, CWK600503S, CWK600506S, CWK600539S, CWK601351S, CWK601741S, CWK601801S, CWK603225S, CWK600461S, CWK600504S, CWK600505S, CWK600509S, CWK600538S, CWK600861S, CWK600924S, CWK600723S, CWK604590S, CWK604984S, CWK605167S, CWK601356S, CWK601486S, CWK601496S, CWK602108S, CWK603454S, CWK602887S, CWK602888S, CWK602890S, CWK602959S, CWK602960S, CWK603219S, CWK603469S, CWK604169S, CWK604174S, CWK604447S, CWK603005S, CWK603006S, CWK603447S, CWK602370S, CWK603462S, CWK603507S, CWK604568S, CWK604800S, CWK605140S, CWK600017S, CWK600025S, CWK600026S, CWK600036S, CWK600037S, CWK600038S, CWK600040S, CWK600051S, CWK600052S, CWK600071S, CWK600073S, CWK600202S, CWK600255S, CWK600269S, CWK600272S, CWK600282S, CWK600323S, CWK600380S, CWK600493S, CWK600836S, CWK601148S, CWK601188S, CWK601277S, CWK601311S, CWK601411S, CWK601412S, CWK601413S, CWK601469S, CWK601481S, CWK601501S, CWK601514S, CWK601561S, CWK601599S, CWK601722S, CWK602546S, CWK602547S, CWK602693S, CWK602855S, CWK602856S, CWK602879S, CWK603150S, CWK603272S, CWK603979S, CWK603980S, CWK604078S, CWK604403S, CWK604649S, CWK604674S, CWK605012S, CWK600888S, CWK600908S, CWK600583S, CWK600584S, CWK600799S, CWK600800S, CWK600915S, CWK601320S, CWK601418S, CWK601589S, CWK601590S, CWK602241S, CWK602531S, CWK602532S, CWK602533S, CWK602865S, CWK602896S, CWK602897S, CWK602976S, CWK602977S, CWK603533S, CWK604276S, CWK605065S, CWK604550S, CWK600392S, CWK600393S, CWK600394S, CWK600427S, CWK600467S, CWK600491S, CWK600659S, CWK600707S, CWK600770S, CWK600849S, CWK600890S, CWK600926S, CWK601016S, CWK601067S, CWK601072S, CWK601073S, CWK601286S, CWK601419S, CWK601438S, CWK601910S, CWK601911S, CWK602035S, CWK602104S, CWK602164S, CWK602402S, CWK602403S, CWK602648S, CWK602657S, CWK602701S, CWK603279S, CWK603363S, CWK603364S, CWK603368S, CWK603470S, CWK603472S, CWK603774S, CWK603882S, CWK604103S, CWK600237S, CWK600238S, CWK600460S, CWK600462S, CWK600517S, CWK600522S, CWK600543S, CWK603145S, CWK603146S, CWK603403S, CWK603589S, CWK603635S, CWK603466S, CWK603467S, CWK604176S, CWK605018S, CWK605115S, CWK600027S, CWK600032S, CWK600033S, CWK600034S, CWK600035S, CWK600222S, CWK600230S, CWK600241S, CWK600333S, CWK600398S, CWK600446S, CWK600482S, CWK600485S, CWK600549S, CWK600645S, CWK600944S, CWK601005S, CWK601006S, CWK601063S, CWK601064S, CWK601136S, CWK601836S, CWK601850S, CWK601851S, CWK601852S, CWK601869S, CWK602763S, CWK602771S, CWK602899S, CWK602900S, CWK602901S, CWK602968S, CWK603087S, CWK603195S, CWK603382S, CWK603819S, CWK604654S, CWK604709S, CWK604901S, CWK601448S, CWK601449S, CWK601450S, CWK601452S, CWK601453S, CWK601583S, CWK601584S, CWK601585S, CWK601586S, CWK601587S, CWK602028S, CWK602941S, CWK602942S, CWK602943S, CWK602944S, CWK602945S, CWK604386S, CWK604387S, CWK604388S, CWK604524S, CWK604799S, CWK604808S, CWK604823S, CWK604830S, CWK605057S, CWK605058S, CWK605059S, CWK605089S, CWK605275S, CWK602636S, CWK603010S, CWK603213S, CWK603696S, CWK605131S, CWK605132S, CWK601832S, CWK603326S, CWK603512S, CWK600020S, CWK600055S, CWK600324S, CWK600996S, CWK601553S, CWK601765S, CWK602482S, CWK602872S, CWK603358S, CWK603564S, CWK603719S, CWK603961S, CWK603976S, CWK604008S, CWK604070S, CWK604338S, CWK604453S, CWK604582S, CWK604966S, CWK604989S, CWK605133S, CWK605343S, CWK600531S, CWK600532S, CWK600561S, CWK600588S, CWK600590S, CWK600610S, CWK600820S, CWK600897S, CWK601180S, CWK601404S, CWK601454S, CWK601503S, CWK601715S, CWK602483S, CWK602534S, CWK603241S, CWK603266S, CWK603435S, CWK603577S, CWK603592S, CWK603720S, CWK603766S, CWK603973S, CWK604043S, CWK604401S, CWK605087S, CWK605448S, CWK602826S, CWK604101S, CWK604025S, CWK601066S, CWK601137S, CWK601161S, CWK601309S, CWK601549S, CWK601594S, CWK601629S, CWK601729S, CWK601753S, CWK601755S, CWK601776S, CWK601785S, CWK601789S, CWK601835S, CWK601849S, CWK601873S, CWK601874S, CWK601875S, CWK601877S, CWK601878S, CWK601905S, CWK601944S, CWK602049S, CWK602051S, CWK602055S, CWK602092S, CWK602128S, CWK602150S, CWK602227S, CWK602284S, CWK602362S, CWK602390S, CWK602583S, CWK602708S, CWK602724S, CWK602740S, CWK602746S, CWK602768S, CWK602785S, CWK602935S, CWK602989S, CWK602990S, CWK603037S, CWK603165S, CWK603245S, CWK603346S, CWK603388S, CWK603402S, CWK603560S, CWK603563S, CWK603583S, CWK603679S, CWK603741S, CWK603810S, CWK604339S, CWK604435S, CWK604436S, CWK604454S, CWK604548S, CWK604894S, CWK604931S, CWK604965S, CWK605190S, CWK605330S, CWK605434S, CWK601617S, CWK602429S, CWK604217S, CWK604331S, CWK604677S, CWK600354S, CWK600355S, CWK600546S, CWK600934S, CWK601070S, CWK601187S, CWK602624S, CWK603226S, CWK603227S, CWK603273S, CWK603671S, CWK603734S, CWK604133S, CWK604234S, CWK604269S, CWK604708S, CWK605091S, CWK605172S, CWK605173S, CWK601824S, CWK602124S, CWK603261S, CWK603373S, CWK604364S, CWK602119S, CWK604999S, CWK600413S, CWK600866S, CWK603098S, CWK603099S, CWK604179S, CWK600963S, CWK601526S, CWK600023S, CWK600024S, CWK600409S, CWK600410S, CWK600411S, CWK600412S, CWK600641S, CWK600676S, CWK600698S, CWK603749S, CWK600767S, CWK600911S, CWK601047S, CWK600283S, CWK600284S, CWK600285S, CWK600286S, CWK600608S, CWK600615S, CWK600616S, CWK600746S, CWK600747S, CWK600991S, CWK600992S, CWK601405S, CWK601406S, CWK601817S, CWK602420S, CWK602755S, CWK603022S, CWK603023S, CWK603152S, CWK603153S, CWK603680S, CWK603921S, CWK603922S, CWK604205S, CWK604221S, CWK604723S, CWK605010S, CWK604589S, CWK604978S, CWK605272S, CWK605384S, CWK604173S, CWK600604S, CWK601061S, CWK604425S, CWK600363S, CWK600364S, CWK600365S, CWK600366S, CWK600367S, CWK600368S, CWK600494S, CWK600495S, CWK600496S, CWK600497S, CWK600498S, CWK600708S, CWK600709S, CWK600710S, CWK600712S, CWK600713S, CWK602023S, CWK602024S, CWK602263S, CWK602264S, CWK602265S, CWK602266S, CWK602267S, CWK602358S, CWK602359S, CWK602361S, CWK602537S, CWK602538S, CWK602539S, CWK600007S, CWK600054S, CWK600056S, CWK600057S, CWK600059S, CWK600062S, CWK600063S, CWK600065S, CWK600172S, CWK600174S, CWK600191S, CWK600192S, CWK600204S, CWK600205S, CWK600225S, CWK600226S, CWK600264S, CWK600265S, CWK600275S, CWK600276S, CWK600277S, CWK600311S, CWK600312S, CWK600326S, CWK600327S, CWK600329S, CWK600330S, CWK600331S, CWK600332S, CWK600336S, CWK600338S, CWK600358S, CWK600359S, CWK600360S, CWK600361S, CWK600386S, CWK600388S, CWK600390S, CWK600391S, CWK600399S, CWK600401S, CWK600459S, CWK600512S, CWK600617S, CWK600629S, CWK600630S, CWK600631S, CWK600632S, CWK600633S, CWK600667S, CWK600668S, CWK600669S, CWK600670S, CWK600671S, CWK600727S, CWK600728S, CWK600765S, CWK600766S, CWK600772S, CWK600773S, CWK600830S, CWK600833S, CWK600862S, CWK600867S, CWK600883S, CWK600884S, CWK600885S, CWK600898S, CWK600899S, CWK600900S, CWK600901S, CWK600902S, CWK600903S, CWK600936S, CWK600937S, CWK600938S, CWK600939S, CWK600988S, CWK600989S, CWK601009S, CWK601011S, CWK601020S, CWK601038S, CWK601040S, CWK601153S, CWK601154S, CWK601175S, CWK601176S, CWK601233S, CWK601234S, CWK601294S, CWK601295S, CWK601336S, CWK601337S, CWK601346S, CWK601382S, CWK601388S, CWK601416S, CWK601417S, CWK601463S, CWK601477S, CWK601478S, CWK601492S, CWK601493S, CWK601504S, CWK601505S, CWK601507S, CWK601557S, CWK601558S, CWK601559S, CWK601675S, CWK601676S, CWK601735S, CWK601736S, CWK601737S, CWK601746S, CWK601747S, CWK601792S, CWK601853S, CWK601854S, CWK601947S, CWK602038S, CWK602039S, CWK602093S, CWK602094S, CWK602137S, CWK602144S, CWK602178S, CWK602255S, CWK602345S, CWK602348S, CWK602378S, CWK602380S, CWK602446S, CWK602466S, CWK602500S, CWK602540S, CWK602613S, CWK602616S, CWK602662S, CWK602663S, CWK602750S, CWK602773S, CWK602816S, CWK602873S, CWK603038S, CWK603210S, CWK603280S, CWK603312S, CWK603456S, CWK603516S, CWK603542S, CWK603713S, CWK603903S, CWK603905S, CWK604068S, CWK604102S, CWK604125S, CWK604789S, CWK605037S, CWK605336S, CWK602958S, CWK603228S, CWK603506S, CWK603538S, CWK604168S, CWK600457S, CWK600518S, CWK600520S, CWK600567S, CWK600743S, CWK600744S, CWK600745S, CWK600748S, CWK600750S, CWK600751S, CWK600752S, CWK600753S, CWK600768S, CWK600769S, CWK600777S, CWK600805S, CWK600806S, CWK600807S, CWK600811S, CWK600812S, CWK600813S, CWK600814S, CWK600817S, CWK600818S, CWK600819S, CWK600822S, CWK600823S, CWK600824S, CWK600825S, CWK600826S, CWK600827S, CWK600828S, CWK600838S, CWK600843S, CWK600844S, CWK600870S, CWK600871S, CWK600872S, CWK600873S, CWK600874S, CWK600875S, CWK600876S, CWK600880S, CWK600881S, CWK600882S, CWK600886S, CWK600906S, CWK600907S, CWK600928S, CWK600961S, CWK600962S, CWK600967S, CWK600968S, CWK600969S, CWK600970S, CWK600977S, CWK600979S, CWK600980S, CWK600981S, CWK600993S, CWK600994S, CWK601021S, CWK601022S, CWK601023S, CWK601024S, CWK601057S, CWK601068S, CWK601069S, CWK601081S, CWK601083S, CWK601151S, CWK601165S, CWK601166S, CWK601167S, CWK601168S, CWK601170S, CWK601185S, CWK601186S, CWK601191S, CWK601193S, CWK601197S, CWK601198S, CWK601199S, CWK601200S, CWK601201S, CWK601225S, CWK601226S, CWK601227S, CWK601228S, CWK601229S, CWK601230S, CWK601231S, CWK601232S, CWK601237S, CWK601238S, CWK601239S, CWK601240S, CWK601241S, CWK601243S, CWK601244S, CWK601245S, CWK601246S, CWK601247S, CWK601248S, CWK601250S, CWK601253S, CWK601258S, CWK601259S, CWK601260S, CWK601261S, CWK601262S, CWK601272S, CWK601273S, CWK601274S, CWK601275S, CWK601276S, CWK601279S, CWK601287S, CWK601288S, CWK601291S, CWK601297S, CWK601298S, CWK601312S, CWK601313S, CWK601314S, CWK601318S, CWK601319S, CWK601343S, CWK601344S, CWK601357S, CWK601386S, CWK601389S, CWK601391S, CWK601395S, CWK601396S, CWK601397S, CWK601399S, CWK601403S, CWK601407S, CWK601408S, CWK601410S, CWK601414S, CWK601422S, CWK601423S, CWK601424S, CWK601425S, CWK601432S, CWK601434S, CWK601435S, CWK601436S, CWK601439S, CWK601445S, CWK601447S, CWK601462S, CWK601470S, CWK601471S, CWK601472S, CWK601474S, CWK601475S, CWK601476S, CWK601483S, CWK601485S, CWK601490S, CWK601520S, CWK601525S, CWK601534S, CWK601535S, CWK601536S, CWK601537S, CWK601538S, CWK601539S, CWK601540S, CWK601541S, CWK601542S, CWK601543S, CWK601544S, CWK601545S, CWK601546S, CWK601547S, CWK601572S, CWK601573S, CWK601574S, CWK601576S, CWK601577S, CWK601579S, CWK601580S, CWK601591S, CWK601607S, CWK601608S, CWK601609S, CWK601610S, CWK601615S, CWK601623S, CWK601624S, CWK601626S, CWK601628S, CWK601632S, CWK601634S, CWK601635S, CWK601636S, CWK601637S, CWK601638S, CWK601639S, CWK601640S, CWK601641S, CWK601642S, CWK601643S, CWK601644S, CWK601645S, CWK601646S, CWK601647S, CWK601648S, CWK601649S, CWK601650S, CWK601651S, CWK601652S, CWK601653S, CWK601665S, CWK601672S, CWK601679S, CWK601680S, CWK601682S, CWK601683S, CWK601684S, CWK601685S, CWK601686S, CWK601688S, CWK601689S, CWK601690S, CWK601716S, CWK601717S, CWK601725S, CWK601734S, CWK601757S, CWK601758S, CWK601759S, CWK601766S, CWK601771S, CWK601774S, CWK601777S, CWK601778S, CWK601779S, CWK601780S, CWK601784S, CWK601786S, CWK601787S, CWK601788S, CWK601805S, CWK601807S, CWK601809S, CWK601818S, CWK601820S, CWK601821S, CWK601822S, CWK601826S, CWK601827S, CWK601828S, CWK601830S, CWK601831S, CWK601841S, CWK601876S, CWK601881S, CWK601907S, CWK601959S, CWK601960S, CWK601962S, CWK601963S, CWK601964S, CWK601967S, CWK601968S, CWK601969S, CWK601973S, CWK601975S, CWK602064S, CWK602072S, CWK602073S, CWK602074S, CWK602075S, CWK602076S, CWK602095S, CWK602115S, CWK602116S, CWK602118S, CWK602252S, CWK602272S, CWK602274S, CWK602275S, CWK602277S, CWK602296S, CWK602306S, CWK602307S, CWK602312S, CWK602316S, CWK602331S, CWK602340S, CWK602341S, CWK602347S, CWK602349S, CWK602364S, CWK602365S, CWK602366S, CWK602367S, CWK602368S, CWK602372S, CWK602375S, CWK602385S, CWK602386S, CWK602387S, CWK602388S, CWK602397S, CWK602405S, CWK602406S, CWK602438S, CWK602441S, CWK602442S, CWK602443S, CWK602451S, CWK602457S, CWK602476S, CWK602479S, CWK602487S, CWK602488S, CWK602496S, CWK602501S, CWK602506S, CWK602507S, CWK602508S, CWK602509S, CWK602510S, CWK602516S, CWK602517S, CWK602520S, CWK602530S, CWK602543S, CWK602544S, CWK602545S, CWK602550S, CWK602552S, CWK602554S, CWK602558S, CWK602560S, CWK602562S, CWK602563S, CWK602565S, CWK602566S, CWK602567S, CWK602587S, CWK602589S, CWK602590S, CWK602591S, CWK602603S, CWK602619S, CWK602683S, CWK602684S, CWK602686S, CWK602687S, CWK602688S, CWK602689S, CWK602690S, CWK602698S, CWK602699S, CWK602700S, CWK602705S, CWK602706S, CWK602744S, CWK602745S, CWK602757S, CWK602758S, CWK602759S, CWK602761S, CWK602775S, CWK602791S, CWK602792S, CWK602794S, CWK602796S, CWK602837S, CWK602847S, CWK602866S, CWK602867S, CWK602868S, CWK602869S, CWK602878S, CWK602880S, CWK602963S, CWK602964S, CWK602992S, CWK602993S, CWK603028S, CWK603029S, CWK603042S, CWK603064S, CWK603088S, CWK603089S, CWK603094S, CWK603096S, CWK603097S, CWK603147S, CWK603169S, CWK603170S, CWK603171S, CWK603172S, CWK603196S, CWK603197S, CWK603198S, CWK603215S, CWK603220S, CWK603260S, CWK603270S, CWK603271S, CWK603307S, CWK603452S, CWK603499S, CWK603500S, CWK603539S, CWK603545S, CWK603574S, CWK603575S, CWK603585S, CWK603623S, CWK603667S, CWK603668S, CWK603669S, CWK603670S, CWK603689S, CWK603690S, CWK603692S, CWK603693S, CWK603695S, CWK603698S, CWK603699S, CWK603701S, CWK603725S, CWK603750S, CWK603754S, CWK603755S, CWK603757S, CWK603759S, CWK603798S, CWK603859S, CWK603966S, CWK603968S, CWK603975S, CWK603981S, CWK603983S, CWK603984S, CWK603985S, CWK603987S, CWK603989S, CWK603990S, CWK603994S, CWK603996S, CWK604001S, CWK604002S, CWK604020S, CWK604045S, CWK600570S, CWK600571S, CWK600572S, CWK600689S, CWK600690S, CWK600691S, CWK602230S, CWK600006S, CWK600022S, CWK600044S, CWK600415S, CWK600528S, CWK600535S, CWK600552S, CWK601588S, CWK601790S, CWK602102S, CWK602580S, CWK602623S, CWK602702S, CWK602703S, CWK602932S, CWK602933S, CWK603007S, CWK603838S, CWK604113S, CWK604402S, CWK603964S, CWK604395S, CWK600210S, CWK600464S, CWK600465S, CWK600540S, CWK600544S, CWK603751S, CWK600664S, CWK601598S, CWK601772S, CWK602940S, CWK603282S, CWK603376S, CWK603436S, CWK603490S, CWK603739S, CWK603740S, CWK603879S, CWK605481S, CWK600697S, CWK600047S, CWK600060S, CWK600308S, CWK600316S, CWK600502S, CWK600541S, CWK600808S, CWK600834S, CWK600929S, CWK601017S, CWK601018S, CWK601062S, CWK601075S, CWK601077S, CWK601143S, CWK601156S, CWK601437S, CWK601515S, CWK601618S, CWK601846S, CWK602065S, CWK602224S, CWK602226S, CWK602256S, CWK602373S, CWK602422S, CWK602424S, CWK602453S, CWK602489S, CWK602528S, CWK602602S, CWK602676S, CWK602737S, CWK602738S, CWK602739S, CWK602748S, CWK602767S, CWK602852S, CWK603034S, CWK603050S, CWK603060S, CWK603061S, CWK603207S, CWK603217S, CWK603277S, CWK603406S, CWK603407S, CWK603410S, CWK603478S, CWK603479S, CWK603480S, CWK603638S, CWK603681S, CWK603891S, CWK604051S, CWK604154S, CWK604180S, CWK604347S, CWK604419S, CWK604450S, CWK604563S, CWK604653S, CWK604756S, CWK604873S, CWK605107S, CWK605142S, CWK604046S, CWK600345S, CWK600346S, CWK600562S, CWK600563S, CWK603955S, CWK603986S, CWK603995S, CWK603998S, CWK604011S, CWK604035S, CWK600569S, CWK600715S, CWK600716S, CWK600771S, CWK601401S, CWK601466S, CWK601491S, CWK601621S, CWK602162S, CWK602392S, CWK602524S, CWK602790S, CWK602985S, CWK603027S, CWK603408S, CWK603477S, CWK603511S, CWK603947S, CWK604104S, CWK604206S, CWK604346S, CWK604523S, CWK604706S, CWK604878S, CWK604915S, CWK605194S, CWK605205S, CWK603974S, CWK600260S, CWK600261S, CWK600262S, CWK600266S, CWK600267S, CWK602811S, CWK602468S, CWK603550S, CWK604972S, CWK605027S, CWK600233S, CWK600274S, CWK600290S, CWK600211S, CWK600212S, CWK600213S, CWK600214S, CWK600215S, CWK600216S, CWK600217S, CWK600218S, CWK600227S, CWK600228S, CWK600229S, CWK600347S, CWK600348S, CWK600349S, CWK600350S, CWK600646S, CWK600647S, CWK600648S, CWK600649S, CWK600650S, CWK600651S, CWK600652S, CWK600573S, CWK600574S, CWK600575S, CWK600576S, CWK600577S, CWK600578S, CWK600013S, CWK600064S, CWK600177S, CWK600240S, CWK600292S, CWK604712S, CWK605019S, CWK605020S, CWK605256S, CWK600683S, CWK600927S, CWK604728S, CWK600201S, CWK600203S, CWK600300S, CWK600542S, CWK601144S, CWK601290S, CWK601518S, CWK602384S, CWK603580S, CWK604406S, CWK604566S, CWK604669S, CWK600443S, CWK600829S, CWK600296S, CWK600435S, CWK600439S, CWK600525S, CWK600566S, CWK600568S, CWK600801S, CWK600842S, CWK600912S, CWK601427S, CWK601480S, CWK601517S, CWK601659S, CWK601661S, CWK601800S, CWK602228S, CWK602289S, CWK602337S, CWK602477S, CWK602491S, CWK602595S, CWK602621S, CWK602628S, CWK602817S, CWK603349S, CWK603451S, CWK603559S, CWK603948S, CWK604705S, CWK605442S, CWK600423S, CWK600831S, CWK601451S, CWK601455S, CWK601460S, CWK601521S, CWK602123S, CWK602515S, CWK603677S, CWK605388S, CWK600306S, CWK600319S, CWK600320S, CWK600381S, CWK600537S, CWK600603S, CWK600624S, CWK600634S, CWK600954S, CWK601065S, CWK601564S, CWK601654S, CWK601656S, CWK602046S, CWK602291S, CWK604267S, CWK601146S, CWK601147S, CWK601442S, CWK601565S, CWK602561S, CWK602820S, CWK603556S, CWK603599S, CWK604588S, CWK603747S, CWK605147S, CWK600976S, CWK600045S, CWK600046S, CWK600050S, CWK601457S, CWK601458S, CWK601459S, CWK601461S, CWK601982S, CWK602032S, CWK602033S, CWK602513S, CWK602564S, CWK602625S, CWK602626S, CWK604123S, CWK604170S, CWK604172S, CWK604315S, CWK604328S, CWK601402S, CWK600559S, CWK600609S, CWK600665S, CWK600726S, CWK600930S, CWK601444S, CWK602134S, CWK602605S, CWK602797S, CWK603161S, CWK603231S, CWK603391S, CWK603392S, CWK603549S, CWK603809S, CWK603855S, CWK603959S, CWK604013S, CWK604056S, CWK604898S, CWK605458S, CWK602749S, CWK604183S, CWK604184S, CWK604222S, CWK604224S, CWK600053S, CWK600160S, CWK600169S, CWK600171S, CWK600249S, CWK600251S, CWK600252S, CWK600281S, CWK600299S, CWK600334S, CWK600379S, CWK600405S, CWK600406S, CWK600438S, CWK600515S, CWK600516S, CWK600553S, CWK600582S, CWK600622S, CWK601347S, CWK601349S, CWK601350S, CWK601352S, CWK601409S, CWK601593S, CWK601674S, CWK601739S, CWK601740S, CWK601742S, CWK601743S, CWK602057S, CWK602105S, CWK602106S, CWK602125S, CWK602182S, CWK602323S, CWK602445S, CWK602490S, CWK602579S, CWK602601S, CWK602650S, CWK602651S, CWK602735S, CWK602736S, CWK602778S, CWK602779S, CWK602862S, CWK602991S, CWK603046S, CWK603086S, CWK603166S, CWK603613S, CWK603721S, CWK603722S, CWK603829S, CWK603830S, CWK603936S, CWK603937S, CWK604091S, CWK604263S, CWK604278S, CWK604312S, CWK604452S, CWK604572S, CWK604666S, CWK604934S, CWK604988S, CWK602058S, CWK602059S, CWK600219S, CWK600220S, CWK600243S, CWK600244S, CWK600254S, CWK600273S, CWK600302S, CWK600303S, CWK600315S, CWK600362S, CWK600416S, CWK600422S, CWK600442S, CWK600473S, CWK600475S, CWK600499S, CWK600536S, CWK600594S, CWK600605S, CWK600653S, CWK600654S, CWK600721S, CWK600840S, CWK600845S, CWK600955S, CWK600990S, CWK601004S, CWK601051S, CWK601192S, CWK601252S, CWK601280S, CWK601420S, CWK601456S, CWK601857S, CWK602100S, CWK602282S, CWK602283S, CWK602541S, CWK602668S, CWK602781S, CWK602969S, CWK603294S, CWK603353S, CWK603354S, CWK603468S, CWK603537S, CWK603770S, CWK603912S, CWK604095S, CWK604096S, CWK604265S, CWK604424S, CWK605079S, CWK605224S, CWK603786S, CWK603804S, CWK604405S, CWK605021S, CWK605066S, CWK605141S, CWK605354S, CWK604553S, CWK600585S, CWK600705S, CWK600793S, CWK600957S, CWK601328S, CWK602285S, CWK602481S, CWK602615S, CWK602875S, CWK603047S, CWK604064S, CWK604215S, CWK604404S, CWK604652S, CWK605070S, CWK600755S, CWK600756S, CWK600984S, CWK600985S, CWK600987S, CWK601028S, CWK601029S, CWK601030S, CWK601031S, CWK601032S, CWK601033S, CWK601034S, CWK601035S, CWK601036S, CWK601037S, CWK602569S, CWK602570S, CWK602571S, CWK602576S, CWK602710S, CWK602711S, CWK602712S, CWK602713S, CWK602714S, CWK602715S, CWK602716S, CWK602717S, CWK602718S, CWK602719S, CWK602915S, CWK602916S, CWK602917S, CWK602918S, CWK602919S, CWK602920S, CWK602921S, CWK602922S, CWK602924S, CWK602925S, CWK602926S, CWK602927S, CWK602928S, CWK602929S, CWK602930S, CWK603243S, CWK603247S, CWK603248S, CWK603250S, CWK603252S, CWK603253S, CWK603380S, CWK603381S, CWK603383S, CWK603385S, CWK604152S, CWK604153S, CWK604155S, CWK604156S, CWK604161S, CWK604360S, CWK604380S, CWK604383S, CWK604384S, CWK604385S, CWK604697S, CWK604698S, CWK604699S, CWK604700S, CWK604701S, CWK604794S, CWK604797S, CWK604798S, CWK604816S, CWK604817S, CWK604818S, CWK604819S, CWK604820S, CWK604821S, CWK604857S, CWK603026S, CWK603488S, CWK603600S, CWK603840S, CWK604397S, CWK604398S, CWK604565S, CWK603322S, CWK600832S, CWK600943S, CWK602966S, CWK603162S, CWK603344S, CWK604549S, CWK604668S, CWK604877S, CWK605102S, CWK605124S, CWK603553S, CWK604839S, CWK605278S, CWK600031S, CWK600039S, CWK600041S, CWK600042S, CWK600206S, CWK600420S, CWK600421S, CWK600788S, CWK600858S, CWK600859S, CWK600860S, CWK600914S, CWK600932S, CWK600933S, CWK600947S, CWK600948S, CWK600949S, CWK600950S, CWK600951S, CWK601008S, CWK601149S, CWK601157S, CWK601196S, CWK601271S, CWK602350S, CWK602353S, CWK602391S, CWK602644S, CWK602741S, CWK602986S, CWK603051S, CWK603052S, CWK603065S, CWK603586S, CWK604417S, CWK604547S, CWK604648S, CWK604702S, CWK603540S, CWK603988S, CWK603992S, CWK601306S, CWK601882S, CWK601977S, CWK603318S, CWK603476S, CWK603724S, CWK604003S, CWK604081S, CWK604986S, CWK601431S, CWK601571S, CWK602126S, CWK602159S, CWK602416S, CWK602418S, CWK602850S, CWK603639S, CWK603911S, CWK604418S, CWK604676S, CWK602593S, CWK602594S, CWK602646S, CWK603596S, CWK603781S, CWK604690S, CWK604847S, CWK601533S, CWK601767S, CWK602149S, CWK602762S, CWK604209S, CWK600675S, CWK600586S, CWK602906S, CWK603155S, CWK603558S, CWK604333S, CWK602356S, CWK602809S, CWK603043S, CWK603907S, CWK604743S, CWK600436S, CWK600454S, CWK600580S, CWK600694S, CWK600704S, CWK600804S, CWK600974S, CWK601315S, CWK601550S, CWK601920S, CWK602432S, CWK602470S, CWK602851S, CWK603179S, CWK603375S, CWK603518S, CWK603673S, CWK603978S, CWK604372S, CWK604375S, CWK604673S, CWK604678S, CWK600424S, CWK602229S, CWK602278S, CWK602343S, CWK603459S, CWK603672S, CWK604340S, CWK604377S, CWK604511S, CWK604636S, CWK605137S, CWK605151S, CWK605437S, CWK600695S, CWK601595S, CWK600693S, CWK600784S, CWK600916S, CWK601014S, CWK601914S, CWK602048S, CWK602132S, CWK602270S, CWK602454S, CWK602514S, CWK602642S, CWK603204S, CWK604322S, CWK601058S, CWK601783S, CWK603637S, CWK601619S, CWK601620S, CWK601745S, CWK601797S, CWK601806S, CWK601813S, CWK601814S, CWK601815S, CWK601816S, CWK601819S, CWK601908S, CWK602067S, CWK602245S, CWK602246S, CWK602321S, CWK602381S, CWK602459S, CWK602497S, CWK602498S, CWK602559S, CWK602777S, CWK603164S, CWK603202S, CWK604027S, CWK604089S, CWK604325S, CWK604335S, CWK604034S, CWK603502S, CWK604330S, CWK602937S, CWK603343S, CWK603790S, CWK603808S, CWK604277S, CWK603304S, CWK603464S, CWK600869S, CWK603184S, CWK603412S, CWK602894S, CWK603238S, CWK603465S, CWK603663S, CWK603763S, CWK604119S, CWK603953S, CWK601567S, CWK601880S, CWK602666S, CWK603031S, CWK603080S, CWK603199S, CWK604109S, CWK601562S, CWK601794S, CWK601921S, CWK602404S, CWK602474S, CWK603805S, CWK604735S, CWK602526S, CWK604371S, CWK602885S, CWK602886S, CWK602983S, CWK603379S, CWK604317S, CWK601843S, CWK603008S, CWK603484S, CWK603880S, CWK604441S, CWK602902S, CWK600428S, CWK600696S, CWK601329S, CWK601872S, CWK602249S, CWK602765S, CWK602936S, CWK602961S, CWK603083S, CWK604000S, CWK604126S, CWK604129S, CWK604210S, CWK604443S, CWK605258S, CWK605491S, CWK602529S, CWK603434S, CWK604023S, CWK604063S, CWK604069S, CWK604675S, CWK601497S, CWK601498S, CWK601508S, CWK601631S, CWK601633S, CWK601719S, CWK601748S, CWK601752S, CWK601803S, CWK601811S, CWK601488S, CWK602041S, CWK602575S, CWK602578S, CWK603769S, CWK603949S, CWK604041S, CWK604556S, CWK604813S, CWK604835S, CWK604871S, CWK604882S, CWK604960S, CWK605290S, CWK605492S, CWK605493S, CWK601310S, CWK601855S, CWK602107S, CWK602131S, CWK602344S, CWK602682S, CWK602527S, CWK603503S, CWK604039S, CWK605283S, CWK603445S, CWK603762S, CWK601390S, CWK603960S, CWK603977S, CWK604115S, CWK603768S, CWK600554S, CWK601489S, CWK601763S, CWK602034S, CWK602324S, CWK602472S, CWK603340S, CWK603341S, CWK600846S, CWK601317S, CWK601775S, CWK601916S, CWK603914S, CWK602091S, CWK600655S, CWK600700S, CWK600701S, CWK600702S, CWK601155S, CWK601502S, CWK602237S, CWK602248S, CWK602486S, CWK602521S, CWK602654S, CWK602685S, CWK602970S, CWK603062S, CWK603190S, CWK603275S, CWK603351S, CWK603360S, CWK603789S, CWK603887S, CWK604116S, CWK604117S, CWK604122S, CWK604211S, CWK604242S, CWK604449S, CWK604451S, CWK604686S, CWK604694S, CWK604887S, CWK605276S, CWK602379S, CWK601353S, CWK601563S, CWK602784S, CWK602363S, CWK602465S, CWK602707S, CWK603398S, CWK603534S, CWK603738S, CWK604283S, CWK604630S, CWK604683S, CWK604749S, CWK600868S, CWK604235S, CWK600839S, CWK603292S, CWK601171S, CWK601602S, CWK601751S, CWK601892S, CWK601904S, CWK602040S, CWK602232S, CWK600921S, CWK600923S, CWK602127S, CWK604166S, CWK605097S, CWK603013S, CWK603685S, CWK600782S, CWK602783S, CWK603548S, CWK603678S, CWK603813S, CWK603889S, CWK604570S, CWK604639S, CWK600194S, CWK604088S, CWK600626S, CWK603386S, CWK601495S, CWK602845S, CWK604343S, CWK603460S, CWK604231S, CWK600809S, CWK601879S, CWK603842S, CWK602987S, CWK601296S, CWK604314S, CWK604252S, CWK604687S, CWK602695S, CWK603036S, CWK603274S, CWK603582S, CWK600681S, CWK600803S, CWK601732S, CWK602812S, CWK602848S, CWK603309S, CWK604324S, CWK604682S, CWK605230S, CWK605494S, CWK600724S, CWK601178S, CWK601293S, CWK602389S, CWK603463S, CWK600725S, CWK601668S, CWK601731S, CWK601979S, CWK602042S, CWK602088S, CWK602326S, CWK602803S, CWK602984S, CWK603327S, CWK604188S, CWK604284S, CWK604803S, CWK604890S, CWK605195S, CWK601781S, CWK602334S, CWK602371S, CWK602631S, CWK600785S, CWK603175S, CWK603269S, CWK603916S, CWK600714S, CWK600722S, CWK600796S, CWK601263S, CWK602849S, CWK602884S, CWK603048S, CWK603576S, CWK603630S, CWK603899S, CWK604842S, CWK601918S, CWK603509S, CWK600430S, CWK600621S, CWK600741S, CWK601140S, CWK602360S, CWK603335S, CWK604805S, CWK603492S, CWK602892S, CWK600591S, CWK602079S, CWK604094S, CWK600754S, CWK601664S, CWK601728S, CWK602382S, CWK602853S, CWK603457S, CWK604573S, CWK604640S, CWK600441S, CWK600719S, CWK600905S, CWK601467S, CWK602136S, CWK603399S, CWK603493S, CWK603543S, CWK604696S, CWK605253S, CWK605483S, CWK604366S, CWK603449S, CWK600479S, CWK600487S, CWK600666S, CWK600692S, CWK600760S, CWK600789S, CWK600790S, CWK600815S, CWK600913S, CWK600982S, CWK601082S, CWK601152S, CWK601181S, CWK601552S, CWK602635S, CWK603015S, CWK603180S, CWK605227S, CWK602697S, CWK604692S, CWK604559S, CWK600593S, CWK600931S, CWK600426S, CWK603893S, CWK600453S, CWK604854S, CWK601025S, CWK601150S, CWK602235S, CWK600742S, CWK600547S, CWK601795S, CWK602400S, CWK603286S, CWK600558S, CWK600004S, CWK603025S, CWK600448S, CWK601135S, CWK601871S, CWK605213S, CWK603049S, CWK600592S, CWK601798S, CWK603177S, CWK603946S, CWK604178S, CWK604555S, CWK604583S, CWK600835S, CWK600892S, CWK603843S, CWK605480S, CWK603178S, CWK603191S, CWK603333S, CWK603806S, CWK603850S, CWK603954S, CWK604323S, CWK604367S, CWK604689S, CWK604844S, CWK604883S, CWK600677S, CWK603323S, CWK603664S, CWK604363S, CWK600680S, CWK601756S, CWK602120S, CWK604044S, CWK600999S, CWK601000S, CWK601001S, CWK601002S, CWK601978S, CWK602231S, CWK602909S, CWK603919S, CWK602988S, CWK603846S, CWK603264S, CWK604058S, CWK604320S, CWK604635S, CWK605226S, CWK600014S, CWK602247S, CWK602967S, CWK602377S, CWK602810S, CWK603675S, CWK600011S, CWK600015S, CWK600288S, CWK600307S, CWK600351S, CWK600794S, CWK600821S, CWK601080S, CWK601773S, CWK600847S, CWK601236S, CWK601282S, CWK601622S, CWK602369S, CWK603561S, CWK603784S, CWK605356S, CWK605472S, CWK601078S, CWK602066S, CWK602523S, CWK602588S, CWK603044S, CWK603413S, CWK603849S, CWK604239S, CWK604691S, CWK601612S, CWK602709S, CWK604516S, CWK600623S, CWK601812S, CWK603200S, CWK600301S, CWK600425S, CWK600490S, CWK601179S, CWK602458S, CWK603230S, CWK600852S, CWK602112S, CWK600781S, CWK602600S, CWK603212S, CWK605285S, CWK605286S, CWK603071S, CWK602659S, CWK603458S, CWK603193S, CWK600759S, CWK603290S, CWK600893S, CWK601158S, CWK601235S, CWK602585S, CWK602949S, CWK604073S, CWK605240S, CWK603173S, CWK600587S, CWK600718S, CWK600841S, CWK601519S, CWK602238S, CWK602259S, CWK602352S, CWK602525S, CWK602782S, CWK603009S, CWK603059S, CWK603154S, CWK603345S, CWK603396S, CWK603591S, CWK603941S, CWK604219S, CWK604391S, CWK604884S, CWK605259S, CWK603018S, CWK603892S, CWK604245S, CWK603073S, CWK603148S, CWK603192S, CWK601727S, CWK602618S, CWK603816S, CWK603054S, CWK603055S, CWK604259S, CWK604876S, CWK605236S, CWK600642S, CWK602421S, CWK605143S, CWK603474S, CWK601581S, CWK602512S, CWK602874S, CWK602834S, CWK605128S, CWK600857S, CWK602661S, CWK604680S, CWK604037S, CWK601606S, CWK601796S, CWK602063S, CWK602518S, CWK602891S, CWK602898S, CWK603448S, CWK604084S, CWK604575S, CWK600627S, CWK601316S, CWK601601S, CWK602133S, CWK605150S, CWK603450S, CWK602522S, CWK603151S, CWK603714S, CWK601464S, CWK602047S, CWK602158S, CWK602294S, CWK602640S, CWK603308S, CWK603401S, CWK603350S, CWK603761S, CWK604311S, CWK604853S, CWK604580S, CWK603578S, CWK603791S, CWK602399S, CWK602734S, CWK602329S, CWK604642S, CWK605352S, CWK601393S, CWK602044S, CWK602473S, CWK602854S, CWK603182S, CWK603330S, CWK603909S, CWK604365S, CWK605233S, CWK605261S, CWK603039S, CWK603188S, CWK603189S, CWK603504S, CWK604327S, CWK604626S, CWK600068S, CWK600075S, CWK600245S, CWK600257S, CWK600455S, CWK600521S, CWK600550S, CWK600703S, CWK601655S, CWK601663S, CWK601677S, CWK601720S, CWK603765S, CWK601555S, CWK602841S, CWK603446S, CWK603453S, CWK603203S, CWK603299S, CWK603633S, CWK604358S, CWK604359S, CWK604444S, CWK604445S, CWK605238S, CWK605239S, CWK605300S, CWK605301S, CWK600699S, CWK600853S, CWK600854S, CWK601177S, CWK603281S, CWK605241S, CWK602742S, CWK603715S, CWK602351S, CWK602760S, CWK603084S, CWK603234S, CWK603240S, CWK603337S, CWK603619S, CWK603788S, CWK604255S, CWK604497S, CWK604836S, CWK605266S, CWK605460S, CWK602766S, CWK603296S, CWK603384S, CWK602638S, CWK603024S, CWK601415S, CWK601548S, CWK602257S, CWK605138S, CWK601551S, CWK603262S, CWK603890S, CWK600685S, CWK600686S, CWK600717S, CWK600786S, CWK600964S, CWK600965S, CWK600966S, CWK600973S, CWK600978S, CWK600986S, CWK600998S, CWK601003S, CWK601039S, CWK601289S, CWK601951S, CWK602151S, CWK602225S, CWK602417S, CWK602678S, CWK603057S, CWK603339S, CWK603347S, CWK603409S, CWK603776S, CWK603897S, CWK603898S, CWK604448S, CWK604935S, CWK605295S, CWK605296S, CWK604336S, CWK604258S, CWK602612S, CWK600740S, CWK600440S, CWK603822S, CWK604628S, CWK600761S, CWK600778S, CWK600783S, CWK600850S, CWK600891S, CWK600904S, CWK600925S, CWK600971S, CWK600983S, CWK601052S, CWK601194S, CWK601195S, CWK601255S, CWK601256S, CWK601257S, CWK601266S, CWK601267S, CWK601270S, CWK601441S, CWK601484S, CWK601575S, CWK601603S, CWK601744S, CWK601862S, CWK601867S, CWK601868S, CWK602052S, CWK602068S, CWK602071S, CWK602097S, CWK602240S, CWK602262S, CWK602333S, CWK602376S, CWK602448S, CWK602449S, CWK602450S, CWK602478S, CWK602573S, CWK602574S, CWK602606S, CWK602658S, CWK602674S, CWK602675S, CWK602696S, CWK602722S, CWK602764S, CWK602787S, CWK602788S, CWK602830S, CWK602831S, CWK602833S, CWK602881S, CWK602893S, CWK602904S, CWK602905S, CWK602946S, CWK602948S, CWK602978S, CWK603011S, CWK603012S, CWK603066S, CWK603067S, CWK603156S, CWK603157S, CWK603159S, CWK603160S, CWK603208S, CWK603209S, CWK603211S, CWK603216S, CWK603218S, CWK603284S, CWK603295S, CWK603298S, CWK603300S, CWK603301S, CWK603393S, CWK603405S, CWK603487S, CWK603521S, CWK603523S, CWK603524S, CWK603525S, CWK603526S, CWK603527S, CWK603528S, CWK603529S, CWK603530S, CWK603686S, CWK603687S, CWK603688S, CWK603742S, CWK603745S, CWK603746S, CWK603756S, CWK603771S, CWK603848S, CWK603867S, CWK603868S, CWK603870S, CWK603871S, CWK603873S, CWK603874S, CWK603876S, CWK603877S, CWK604066S, CWK604092S, CWK604093S, CWK604097S, CWK604106S, CWK604342S, CWK604345S, CWK604353S, CWK604357S, CWK604426S, CWK604427S, CWK604428S, CWK604429S, CWK604430S, CWK604431S, CWK604432S, CWK604433S, CWK604434S, CWK604520S, CWK604600S, CWK604633S, CWK604638S, CWK604688S, CWK604859S, CWK604860S, CWK604861S, CWK604862S, CWK604863S, CWK604904S, CWK605082S, CWK605202S, CWK605203S, CWK605248S, CWK605249S, CWK605250S, CWK605251S, CWK605252S, CWK605331S, CWK605332S, CWK605333S, CWK605335S, CWK605349S, CWK605350S, CWK605394S, CWK605399S, CWK605400S, CWK604257S, CWK600447S, CWK603732S, CWK604390S, CWK605382S, CWK602165S, CWK601823S, CWK600067S, CWK602656S, CWK600321S, CWK600340S, CWK600372S, CWK600471S, CWK600534S, CWK600548S, CWK601050S, CWK601355S, CWK601839S, CWK601847S, CWK604108S, CWK604332S, CWK604809S, CWK604843S, CWK604911S, CWK601019S, CWK602251S, CWK602419S, CWK603584S, CWK603844S, CWK603881S, CWK604197S, CWK604440S, CWK604472S, CWK604071S, CWK600335S, CWK602447S, CWK600795S, CWK600942S, CWK602641S, CWK602463S, CWK603004S, CWK604632S, CWK603795S, CWK603306S, CWK603588S, CWK604272S, CWK604581S, CWK600658S, CWK605328S, CWK600619S, CWK601749S, CWK602043S, CWK604587S, CWK604695S, CWK602883S, CWK603565S, CWK604067S, CWK600810S, CWK601027S, CWK601799S, CWK602268S, CWK602444S, CWK602608S, CWK603531S, CWK604578S, CWK601760S, CWK603076S, CWK602167S, CWK603498S, CWK603717S, CWK604560S, CWK600757S, CWK601560S, CWK601582S, CWK601691S, CWK602280S, CWK602604S, CWK603787S, CWK603817S, CWK603820S, CWK604074S, CWK604163S, CWK604256S, CWK604629S, CWK604739S, CWK605149S, CWK603045S, CWK604006S, CWK601845S, CWK605462S, CWK602437S, CWK604124S, CWK603374S, CWK603378S, CWK603593S, CWK601494S, CWK602495S, CWK600687S, CWK602549S, CWK602655S, CWK602844S, CWK605439S, CWK601838S, CWK602160S, CWK602161S, CWK600028S, CWK600029S, CWK600030S, CWK600195S, CWK600196S, CWK600197S, CWK600198S, CWK600199S, CWK600382S, CWK600383S, CWK600385S, CWK600729S, CWK600730S, CWK600731S, CWK600732S, CWK600733S, CWK600734S, CWK600735S, CWK600736S, CWK600737S, CWK600738S, CWK602276S, CWK602279S, CWK602281S, CWK602288S, CWK602290S, CWK602292S, CWK602293S, CWK602295S, CWK602298S, CWK602301S, CWK602303S, CWK602305S, CWK602308S, CWK602309S, CWK602310S, CWK602311S, CWK602313S, CWK602314S, CWK602315S, CWK602322S, CWK603835S, CWK601172S, CWK601173S, CWK601182S, CWK601530S, CWK601531S, CWK601791S, CWK601793S, CWK602553S, CWK602839S, CWK603232S, CWK603233S, CWK604399S, CWK604400S, CWK604464S, CWK604465S, CWK604530S, CWK604531S, CWK604858S, CWK601190S, CWK600514S, CWK601354S, CWK601600S, CWK602269S, CWK603902S, CWK605398S, CWK602804S, CWK602805S, CWK602806S, CWK602807S, CWK602808S, CWK603313S, CWK603317S, CWK603365S, CWK603366S, CWK603367S, CWK603551S, CWK603552S, CWK603811S, CWK603818S, CWK603821S, CWK604189S, CWK604193S, CWK604254S, CWK604279S, CWK604352S, CWK604354S, CWK604355S, CWK604370S, CWK604378S, CWK604379S, CWK604381S, CWK604382S, CWK604554S, CWK605281S, CWK605282S, CWK603377S, CWK603573S, CWK601059S, CWK601071S, CWK600012S, CWK600043S, CWK600232S, CWK600234S, CWK600236S, CWK600253S, CWK600287S, CWK600193S, CWK600200S, CWK600231S, CWK600235S, CWK600256S, CWK600258S, CWK600476S, CWK600452S, CWK600787S, CWK602244S, CWK600291S, CWK600356S, CWK600507S, CWK604282S, CWK600066S, CWK600178S, CWK600186S, CWK600209S, CWK600223S, CWK600242S, CWK600293S, CWK600294S, CWK600377S, CWK600378S, CWK600384S, CWK600417S, CWK600418S, CWK600500S, CWK600555S, CWK600556S, CWK600672S, CWK600673S, CWK603168S, CWK603752S, CWK603967S, CWK604028S, CWK604525S, CWK602971S, CWK602972S, CWK602973S, CWK603513S, CWK603515S, CWK603772S, CWK604010S, CWK604047S, CWK604261S, CWK604664S, CWK604956S, CWK604963S, CWK605136S, CWK605170S, CWK605184S, CWK605191S, CWK605192S, CWK604053S, CWK600341S, CWK600342S, CWK600343S, CWK600661S, CWK600662S, CWK600663S, CWK600779S, CWK600780S, CWK601189S, CWK601394S, CWK601398S, CWK601400S, CWK601768S, CWK601769S, CWK601770S, CWK601837S, CWK601840S, CWK601844S, CWK602287S, CWK602786S, CWK602829S, CWK602907S, CWK602908S, CWK603149S, CWK603482S, CWK603483S, CWK603597S, CWK603684S, CWK603924S, CWK604538S, CWK604983S, CWK600248S, CWK600339S, CWK600352S, CWK600614S, CWK600625S, CWK600684S, CWK600909S, CWK600960S, CWK600997S, CWK601045S, CWK601046S, CWK601053S, CWK601142S, CWK601160S, CWK601284S, CWK601308S, CWK601342S, CWK601487S, CWK601829S, CWK602087S, CWK602995S, CWK603832S, CWK604052S, CWK604105S, CWK604268S, CWK604356S, CWK604414S, CWK604496S, CWK604663S, CWK605076S, CWK605188S, CWK605189S, CWK600529S, CWK601056S, CWK604423S, CWK604260S, CWK601426S, CWK600048S, CWK600049S, CWK600069S, CWK600159S, CWK600166S, CWK600221S, CWK600224S, CWK600239S, CWK600263S, CWK600279S, CWK600298S, CWK600305S, CWK600317S, CWK600344S, CWK600373S, CWK600374S, CWK600375S, CWK600396S, CWK600400S, CWK600403S, CWK600404S, CWK600408S, CWK600432S, CWK600450S, CWK600472S, CWK600480S, CWK600492S, CWK600511S, CWK600551S, CWK600597S, CWK600600S, CWK600601S, CWK600613S, CWK600643S, CWK600679S, CWK600706S, CWK600764S, CWK600775S, CWK600797S, CWK600798S, CWK600877S, CWK600878S, CWK600879S, CWK600894S, CWK600918S, CWK600919S, CWK600920S, CWK600952S, CWK600953S, CWK600958S, CWK601026S, CWK601054S, CWK601076S, CWK601164S, CWK601264S, CWK601285S, CWK601303S, CWK601324S, CWK601421S, CWK601443S, CWK601468S, CWK601556S, CWK601578S, CWK601657S, CWK601658S, CWK601666S, CWK601669S, CWK601670S, CWK601671S, CWK601750S, CWK601858S, CWK601859S, CWK601860S, CWK601861S, CWK601923S, CWK601945S, CWK602082S, CWK602083S, CWK602084S, CWK602168S, CWK602169S, CWK602396S, CWK602409S, CWK602410S, CWK602426S, CWK602427S, CWK602428S, CWK602471S, CWK602535S, CWK602596S, CWK602597S, CWK602614S, CWK602622S, CWK602629S, CWK602679S, CWK602680S, CWK602681S, CWK602776S, CWK602799S, CWK602800S, CWK602824S, CWK602861S, CWK602863S, CWK602864S, CWK602952S, CWK602953S, CWK602955S, CWK602980S, CWK602981S, CWK603041S, CWK603069S, CWK603185S, CWK603186S, CWK603331S, CWK603356S, CWK603395S, CWK603475S, CWK603505S, CWK603510S, CWK603569S, CWK603782S, CWK603825S, CWK603865S, CWK603997S, CWK603999S, CWK604004S, CWK604016S, CWK604076S, CWK604080S, CWK604083S, CWK604100S, CWK604225S, CWK604271S, CWK604274S, CWK604350S, CWK604351S, CWK604411S, CWK604412S, CWK604551S, CWK604562S, CWK604564S, CWK604595S, CWK604597S, CWK604643S, CWK604662S, CWK604801S, CWK604906S, CWK604907S, CWK605017S, CWK605135S, CWK605166S, CWK605247S, CWK605267S, CWK605459S, CWK605461S, CWK600010S, CWK601506S, CWK602054S, CWK602599S, CWK603803S, CWK604622S, CWK600018S, CWK600019S, CWK600250S, CWK600268S, CWK600271S, CWK600325S, CWK600612S, CWK600941S, CWK600975S, CWK601049S, CWK601616S, CWK602036S, CWK602117S, CWK602253S, CWK602598S, CWK602725S, CWK602726S, CWK602727S, CWK603443S, CWK603614S, CWK603665S, CWK603863S, CWK603864S, CWK604618S, CWK604620S, CWK605218S, CWK604621S, CWK602173S, CWK602174S, CWK603624S, CWK603629S, CWK603631S, CWK603636S, CWK603642S, CWK603847S, CWK603952S, CWK604262S, CWK604264S, CWK604409S, CWK604410S, CWK604679S, CWK604962S, CWK604990S, CWK605038S, CWK605069S, CWK605075S, CWK605169S, CWK600328S, CWK600387S, CWK601183S, CWK601184S, CWK602821S, CWK603267S, CWK601723S, CWK601922S, CWK602261S, CWK603183S, CWK603541S, CWK603554S, CWK603555S, CWK604054S, CWK604182S, CWK604658S, CWK605464S, CWK602030S, CWK602053S, CWK603144S, CWK604019S, CWK600322S, CWK601754S, CWK603201S, CWK603404S, CWK604982S, CWK605183S, CWK605269S, CWK605270S, CWK602135S, CWK602234S, CWK603920S, CWK602286S, CWK602633S, CWK602634S, CWK603444S, CWK604408S, CWK604961S, CWK604024S, CWK604033S, CWK600371S, CWK600466S, CWK601721S, CWK602327S, CWK603824S, CWK604614S, CWK605146S, CWK605353S, CWK605496S, CWK605497S, CWK604791S, CWK605287S, CWK605288S, CWK605289S, CWK605339S, CWK605340S, CWK605372S, CWK603256S, CWK600564S, CWK603712S, CWK604014S, CWK603091S, CWK603336S, CWK603618S, CWK605120S, CWK600207S, CWK600449S, CWK600628S, CWK601055S, CWK602374S, CWK603486S, CWK603491S, CWK603826S, CWK603833S, CWK603970S, CWK604657S, CWK604713S, CWK600246S, CWK600247S, CWK600297S, CWK600463S, CWK600508S, CWK600618S, CWK601428S, CWK602609S, CWK603764S, CWK603918S, CWK604416S, CWK600016S, CWK600190S, CWK601292S, CWK603547S, CWK603777S, CWK604980S, CWK603079S, CWK604005S, CWK604185S, CWK603372S, CWK603702S, CWK605022S, CWK605036S, CWK605500S, CWK600611S, CWK600478S, CWK601527S, CWK601528S, CWK601529S, CWK601597S, CWK602423S, CWK602464S, CWK602751S, CWK602752S, CWK602754S, CWK602769S, CWK602772S, CWK602815S, CWK602934S, CWK603030S, CWK603032S, CWK603033S, CWK603205S, CWK603251S, CWK604022S, CWK603440S, CWK603441S, CWK603442S, CWK602630S, CWK604218S, CWK604072S, CWK604237S, CWK604919S, CWK604921S, CWK602692S, CWK604061S, CWK604223S, CWK605006S, CWK604422S, CWK604787S, CWK604969S, CWK604970S, CWK600208S, CWK600304S, CWK600402S, CWK600437S, CWK601138S, CWK601568S, CWK601909S, CWK602056S, CWK602581S, CWK602582S, CWK604229S, CWK604042S, CWK604048S, CWK604055S, CWK604059S, CWK604249S, CWK604253S, CWK604710S, CWK604920S, CWK605405S, CWK602877S, CWK602923S, CWK603359S, CWK603370S, CWK603371S, CWK603496S, CWK603497S, CWK603793S, CWK603794S, CWK604526S, CWK604527S, CWK604533S, CWK604534S, CWK604856S, CWK605095S, CWK605501S, CWK602665S, CWK603839S, CWK604232S, CWK601681S, CWK602250S, CWK602436S, CWK602503S, CWK602720S, CWK603019S, CWK603206S, CWK603265S, CWK603807S, CWK603883S, CWK604079S, CWK605031S, CWK605040S, CWK603074S, CWK602756S, CWK604131S, CWK604601S, CWK604711S, CWK605246S, CWK605305S, CWK600005S, CWK600526S, CWK600602S, CWK601265S, CWK601446S, CWK601605S, CWK601627S, CWK602493S, CWK602789S, CWK603020S, CWK603021S, CWK603485S, CWK603620S, CWK603780S, CWK603862S, CWK601863S, CWK601864S, CWK601865S, CWK602243S, CWK602328S, CWK602504S, CWK602555S, CWK602556S, CWK602557S, CWK602859S, CWK602860S, CWK603092S, CWK603174S, CWK603239S, CWK603361S, CWK603674S, CWK603735S, CWK603736S, CWK603799S, CWK603869S, CWK603872S, CWK603906S, CWK603993S, CWK604060S, CWK604270S, CWK604349S, CWK604407S, CWK604656S, CWK604903S, CWK604905S, CWK605067S, CWK605090S, CWK605165S, CWK605225S, CWK605262S, CWK605274S, CWK605447S, CWK600863S, CWK600864S, CWK601042S, CWK601043S, CWK601044S, CWK601084S, CWK601202S, CWK601203S, CWK601204S, CWK601299S, CWK601300S, CWK601301S, CWK601302S, CWK601304S, CWK601509S, CWK601510S, CWK601511S, CWK601512S, CWK601513S, CWK601952S, CWK601954S, CWK601955S, CWK601956S, CWK601957S, CWK601958S, CWK602653S, CWK602669S, CWK602670S, CWK602671S, CWK602672S, CWK602673S, CWK602728S, CWK602729S, CWK602730S, CWK602731S, CWK602732S, CWK603522S, CWK603612S, CWK602027S, CWK602037S, CWK602070S, CWK602239S, CWK602802S, CWK604619S, CWK605063S, CWK605293S, CWK601856S, CWK601870S, CWK602045S, CWK602077S, CWK602101S, CWK602122S, CWK602130S, CWK602145S, CWK602147S, CWK602148S, CWK602152S, CWK602153S, CWK602154S, CWK602166S, CWK602170S, CWK602177S, CWK602233S, CWK602236S, CWK602242S, CWK602254S, CWK602258S, CWK602260S, CWK602393S, CWK602394S, CWK602395S, CWK602398S, CWK602401S, CWK602408S, CWK602411S, CWK602412S, CWK602413S, CWK602415S, CWK602439S, CWK602440S, CWK602822S, CWK602823S, CWK602882S, CWK604159S, CWK604528S, CWK601596S, CWK601673S, CWK601730S, CWK602183S, CWK600469S, CWK600470S, CWK600524S, CWK600527S, CWK600545S, CWK600606S, CWK600607S, CWK600656S, CWK600657S, CWK600749S, CWK600816S, CWK600855S, CWK600856S, CWK600895S, CWK600896S, CWK600917S, CWK601074S, CWK601912S, CWK602025S, CWK602026S, CWK602085S, CWK602086S, CWK602492S, CWK602494S, CWK602542S, CWK603056S, CWK603268S, CWK603566S, CWK603570S, CWK603571S, CWK604207S, CWK604420S, CWK604421S, CWK604437S, CWK604438S, CWK604522S, CWK604831S, CWK604832S, CWK605083S, CWK605096S, CWK604112S, CWK604306S, CWK604655S, CWK604782S, CWK600074S, CWK600076S, CWK600309S, CWK600353S, CWK600376S, CWK600519S, CWK600533S, CWK600598S, CWK600688S, CWK600935S, CWK601012S, CWK601145S, CWK601305S, CWK601499S, CWK601524S, CWK601866S, CWK602435S, CWK602753S, CWK603494S, CWK603962S, CWK600070S, CWK600072S, CWK600278S, CWK600310S, CWK600397S, CWK600414S, CWK600445S, CWK600477S, CWK600489S, CWK600510S, CWK600589S, CWK600599S, CWK600739S, CWK601761S, CWK601764S, CWK602965S, CWK603827S, CWK605029S, CWK605044S, CWK604250S, CWK603567S, CWK603568S, CWK603938S, CWK603939S, CWK604542S, CWK604644S, CWK604824S, CWK604825S, CWK604827S, CWK604892S, CWK605139S, CWK605145S, CWK605200S, CWK604893S, CWK604667S, CWK604829S, CWK604826S, CWK604646S, CWK601321S, CWK601325S, CWK601326S, CWK601327S, CWK601330S, CWK601331S, CWK601332S, CWK601369S, CWK601370S, CWK601371S, CWK602300S, CWK603602S, CWK603710S, CWK603711S, CWK604194S, CWK604722S, CWK604724S, CWK604868S, CWK604916S, CWK605177S, CWK605178S, CWK601334S, CWK601335S, CWK601338S, CWK601339S, CWK602302S, CWK602304S, CWK602317S, CWK602318S, CWK604461S, CWK604462S, CWK604463S, CWK604591S, CWK604721S, CWK604865S, CWK604917S, CWK604918S, CWK604922S, CWK605311S, CWK605312S, CWK605313S, CWK605360S, CWK605362S, CWK605363S, CWK604752S, CWK603077S, CWK603093S, CWK604593S, CWK604736S, CWK605179S, CWK605180S, CWK605307S, CWK605308S, CWK605309S, CWK605310S, CWK605317S, CWK605381S, CWK602142S, CWK602180S, CWK602299S, CWK602997S, CWK602998S, CWK602999S, CWK603000S, CWK603001S, CWK603082S, CWK603109S, CWK603605S, CWK603606S, CWK604187S, CWK604470S, CWK604596S, CWK604714S, CWK604867S, CWK605365S, CWK601333S, CWK601340S, CWK601341S, CWK601364S, CWK601365S, CWK601366S, CWK601367S, CWK601368S, CWK601372S, CWK601373S, CWK601385S, CWK601429S, CWK601430S, CWK601981S, CWK602062S, CWK602138S, CWK602140S, CWK602143S, CWK602171S, CWK602175S, CWK602179S, CWK603002S, CWK603003S, CWK603072S, CWK603075S, CWK603103S, CWK603104S, CWK603105S, CWK603106S, CWK603107S, CWK603108S, CWK603113S, CWK603607S, CWK603608S, CWK603610S, CWK603611S, CWK603616S, CWK603801S, CWK603802S, CWK603894S, CWK603896S, CWK604195S, CWK604196S, CWK604389S, CWK604469S, CWK604864S, CWK604924S, CWK604925S, CWK605186S, CWK605187S, CWK605358S, CWK605359S, CWK601322S, CWK601323S, CWK601358S, CWK601360S, CWK601361S, CWK601362S, CWK601363S, CWK601374S, CWK601376S, CWK601377S, CWK601378S, CWK601380S, CWK601381S, CWK601384S, CWK602060S, CWK602061S, CWK602176S, CWK602319S, CWK602320S, CWK602938S, CWK603100S, CWK603101S, CWK603102S, CWK603111S, CWK603112S, CWK603603S, CWK603604S, CWK603617S, CWK603703S, CWK603704S, CWK603705S, CWK603706S, CWK603708S, CWK603709S, CWK603900S, CWK603930S, CWK603931S, CWK603933S, CWK604190S, CWK604191S, CWK604192S, CWK604198S, CWK604199S, CWK604201S, CWK604458S, CWK604459S, CWK604460S, CWK604715S, CWK604716S, CWK604717S, CWK604718S, CWK604719S, CWK605185S, CWK605314S, CWK605315S, CWK605316S, CWK605367S, CWK605368S, CWK605370S, CWK605378S, CWK605379S, CWK605380S, CWK602939S, CWK602962S, CWK602996S, CWK603070S, CWK604537S, CWK604866S, CWK605306S, CWK603601S, CWK604615S, CWK604616S, CWK604725S, CWK604726S, CWK605373S, CWK605374S, CWK605375S, CWK605377S, CWK601379S, CWK602956S, CWK602957S, CWK603095S, CWK603110S, CWK604592S, CWK604923S, CWK603068S, CWK604186S, CWK604467S, CWK605366S, CWK604457S, CWK603748S, CWK604057S, CWK604158S, CWK604273S, CWK604275S, CWK604308S, CWK604455S, CWK604703S, CWK605196S, CWK604793S, CWK603632S, CWK605152S, CWK605237S, CWK604561S, CWK602536S, CWK602511S, CWK602838S, CWK602801S, CWK603501S, CWK600395S, CWK600956S, CWK601630S, CWK603760S, CWK603965S, CWK605361S, CWK602951S, CWK603834S, CWK604241S, CWK604244S, CWK602770S, CWK602733S, CWK604214S, CWK604246S, CWK604226S, CWK604833S, CWK604065S, CWK604032S, CWK602505S, CWK602825S, CWK602828S, CWK602832S, CWK602835S, CWK602836S, CWK602840S, CWK602842S, CWK602846S, CWK602857S, CWK602975S, CWK602979S, CWK603035S, CWK603078S, CWK603158S, CWK603263S, CWK603276S, CWK603278S, CWK603285S, CWK603293S, CWK603314S, CWK603316S, CWK603321S, CWK603325S, CWK603332S, CWK603338S, CWK603342S, CWK603357S, CWK603495S, CWK603514S, CWK603536S, CWK603609S, CWK603615S, CWK603621S, CWK603625S, CWK603627S, CWK603628S, CWK603676S, CWK603691S, CWK603743S, CWK603744S, CWK603758S, CWK603767S, CWK603775S, CWK603778S, CWK603779S, CWK603828S, CWK603831S, CWK603836S, CWK603841S, CWK603845S, CWK603851S, CWK603856S, CWK603857S, CWK603866S, CWK603875S, CWK603904S, CWK603913S, CWK603917S, CWK603927S, CWK603928S, CWK603929S, CWK603945S, CWK603950S, CWK603956S, CWK604017S, CWK604026S, CWK604036S, CWK604099S, CWK604227S, CWK604230S, CWK604236S, CWK604238S, CWK604280S, CWK604281S, CWK604481S, CWK604482S, CWK604483S, CWK604484S, CWK604485S, CWK604486S, CWK604487S, CWK604488S, CWK604490S, CWK604491S, CWK604492S, CWK604493S, CWK604494S, CWK604495S, CWK604498S, CWK604499S, CWK604500S, CWK604501S, CWK604502S, CWK604503S, CWK604504S, CWK604507S, CWK604508S, CWK604512S, CWK604514S, CWK604515S, CWK604517S, CWK604518S, CWK604519S, CWK604737S, CWK604738S, CWK604740S, CWK604741S, CWK604742S, CWK604744S, CWK604747S, CWK604748S, CWK604750S, CWK604751S, CWK604753S, CWK604754S, CWK604755S, CWK604757S, CWK604761S, CWK604762S, CWK604763S, CWK604767S, CWK604768S, CWK604769S, CWK604770S, CWK604771S, CWK604772S, CWK604773S, CWK604774S, CWK604775S, CWK604776S, CWK604790S, CWK604792S, CWK604795S, CWK604840S, CWK604841S, CWK605337S, CWK605364S, CWK605369S, CWK605376S, CWK605383S, CWK605389S, CWK605390S, CWK605426S, CWK605427S, CWK605428S, CWK605429S, CWK605430S, CWK602502S, CWK602780S, CWK603287S, CWK603753S, CWK603910S, CWK604007S, CWK604015S, CWK604077S, CWK604086S, CWK604243S, CWK604248S, CWK604506S, CWK604513S, CWK604529S, CWK604532S, CWK604760S, CWK604974S, CWK605334S, CWK602455S, CWK602460S, CWK602889S, CWK602947S, CWK602954S, CWK603707S, CWK604475S, CWK604521S, CWK604240S, CWK605348S, CWK602425S, CWK602913S, CWK603925S, CWK600488S, CWK600530S, CWK600837S, CWK600945S, CWK601392S, CWK601433S, CWK601687S, CWK601692S, CWK604604S, CWK604606S, CWK604607S, CWK604609S, CWK604610S, CWK604612S, CWK604613S, CWK604671S, CWK604672S, CWK603329S, CWK603362S, CWK603387S, CWK603471S, CWK603595S, CWK603598S, CWK603622S, CWK603626S, CWK603634S, CWK603666S, CWK603723S, CWK603783S, CWK603785S, CWK603884S, CWK603885S, CWK603886S, CWK603888S, CWK603895S, CWK603901S, CWK603934S, CWK603935S, CWK603942S, CWK603971S, CWK603972S, CWK604040S, CWK604050S, CWK604075S, CWK604132S, CWK604157S, CWK604160S, CWK604171S, CWK604200S, CWK604202S, CWK604204S, CWK604208S, CWK604212S, CWK604213S, CWK604216S, CWK604220S, CWK604251S, CWK604476S, CWK604477S, CWK604478S, CWK604479S, CWK604480S, CWK604489S, CWK604509S, CWK604510S, CWK604539S, CWK604540S, CWK604567S, CWK604605S, CWK604608S, CWK604611S, CWK604727S, CWK604729S, CWK604730S, CWK604731S, CWK604732S, CWK604733S, CWK604734S, CWK604745S, CWK604746S, CWK604758S, CWK604759S, CWK604764S, CWK604765S, CWK604766S, CWK604781S, CWK604784S, CWK604785S, CWK604786S, CWK604788S, CWK605073S, CWK605074S, CWK605106S, CWK605112S, CWK605393S, CWK605395S, CWK605397S, CWK605404S, CWK605454S, CWK605455S, CWK604536S, CWK604031S, CWK604082S, CWK604162S, CWK604228S, CWK604266S, CWK604309S, CWK604319S, CWK604326S, CWK604392S, CWK604393S, CWK604439S, CWK604456S, CWK604535S, CWK604546S, CWK604557S, CWK604569S, CWK604594S, CWK604617S, CWK604627S, CWK604647S, CWK604665S, CWK604707S, CWK604720S, CWK604783S, CWK604806S, CWK604807S, CWK604811S, CWK604822S, CWK604846S, CWK604850S, CWK604872S, CWK604879S, CWK604880S, CWK604885S, CWK604886S, CWK604896S, CWK604897S, CWK604899S, CWK604902S, CWK604909S, CWK604912S, CWK604913S, CWK604914S, CWK604926S, CWK604927S, CWK604928S, CWK604930S, CWK604932S, CWK604957S, CWK604964S, CWK604967S, CWK604968S, CWK604971S, CWK604985S, CWK604987S, CWK604991S, CWK604992S, CWK604993S, CWK604998S, CWK605004S, CWK605005S, CWK605008S, CWK605009S, CWK605011S, CWK605013S, CWK605023S, CWK605028S, CWK605030S, CWK605039S, CWK605043S, CWK605045S, CWK605047S, CWK605053S, CWK605054S, CWK605055S, CWK605056S, CWK605078S, CWK605081S, CWK605084S, CWK605086S, CWK605088S, CWK605093S, CWK605094S, CWK605098S, CWK605099S, CWK605101S, CWK605109S, CWK605113S, CWK605114S, CWK605129S, CWK605144S, CWK605148S, CWK605153S, CWK605154S, CWK605159S, CWK605168S, CWK605171S, CWK605174S, CWK605175S, CWK605193S, CWK605199S, CWK605201S, CWK605204S, CWK605228S, CWK605235S, CWK605243S, CWK605245S, CWK605254S, CWK605255S, CWK605260S, CWK605264S, CWK605265S, CWK605271S, CWK605273S, CWK605277S, CWK605284S, CWK605294S, CWK605299S, CWK605303S, CWK605304S, CWK605327S, CWK605338S, CWK605341S, CWK605342S, CWK605344S, CWK605346S, CWK605347S, CWK605351S, CWK605355S, CWK605357S, CWK605371S, CWK605387S, CWK605401S, CWK605433S, CWK605435S, CWK605438S, CWK605440S, CWK605441S, CWK605444S, CWK605466S, CWK605467S, CWK605469S, CWK605474S, CWK605475S, CWK605479S, CWK605504S, CWK605505S, CWK605431S, CWK605432S, CWK605450S, CWK605451S, CWK605452S, CWK605453S, CWK604584S, CWK604802S, CWK605105S, CWK605242S, CWK605297S; GTIN 00763000001537, Lot Serial Numbers: CWK601704S, CWK601902S, CWK601933S, CWK603123S, CWK603416S, CWK600085S, CWK601111S, CWK602206S, CWK603648S, CWK600091S, CWK601992S, CWK604143S, CWK605509S, CWK601710S, CWK601985S, CWK601103S, CWK601104S, CWK601132S, CWK603418S, CWK603420S, CWK600114S, CWK601207S, CWK601700S, CWK601709S, CWK601898S, CWK602019S, CWK603425S, CWK603659S, CWK600077S, CWK601703S, CWK601900S, CWK602203S, CWK600148S, CWK600167S, CWK600175S, CWK600188S, CWK601088S, CWK601096S, CWK601110S, CWK601115S, CWK601116S, CWK601120S, CWK601134S, CWK601891S, CWK601931S, CWK602211S, CWK603649S, CWK603654S, CWK603658S, CWK604296S, CWK601127S, CWK601128S, CWK601129S, CWK601218S, CWK601895S, CWK600092S, CWK601108S, CWK601109S, CWK601114S, CWK601131S, CWK602003S, CWK602004S, CWK603128S, CWK603432S, CWK604297S, CWK604298S, CWK600087S, CWK601089S, CWK602209S, CWK605422S, CWK603118S, CWK602191S, CWK602195S, CWK604144S, CWK600127S, CWK600128S, CWK600134S, CWK600139S, CWK600142S, CWK601086S, CWK601090S, CWK601091S, CWK601102S, CWK601106S, CWK601122S, CWK601206S, CWK601216S, CWK601925S, CWK601943S, CWK601994S, CWK601996S, CWK602007S, CWK602022S, CWK602198S, CWK602221S, CWK603120S, CWK603132S, CWK603135S, CWK603419S, CWK603431S, CWK603662S, CWK604136S, CWK604150S, CWK604292S, CWK605418S, CWK605517S, CWK605518S, CWK600079S, CWK600143S, CWK600158S, CWK601123S, CWK602220S, CWK603646S, CWK602217S, CWK604943S, CWK604944S, CWK604945S, CWK605414S, CWK601696S, CWK603131S, CWK604938S, CWK601889S, CWK601890S, CWK601932S, CWK601989S, CWK602013S, CWK602207S, CWK603114S, CWK603426S, CWK603655S, CWK604286S, CWK604304S, CWK602214S, CWK601893S, CWK604293S, CWK601702S, CWK601936S, CWK601883S, CWK603124S, CWK603422S, CWK604950S, CWK602010S, CWK602014S, CWK603657S, CWK601101S, CWK601219S, CWK601697S, CWK601699S, CWK601997S, CWK603122S, CWK602200S, CWK604303S, CWK601121S, CWK603140S, CWK603423S, CWK604288S, CWK604290S, CWK602218S, CWK604135S, CWK604139S, CWK604141S, CWK604142S, CWK601991S, CWK601215S, CWK602223S, CWK601085S, CWK603117S, CWK601220S, CWK601708S, CWK602001S, CWK603660S, CWK601707S, CWK601713S, CWK602000S, CWK604138S, CWK601125S, CWK601927S, CWK602011S, CWK602012S, CWK602188S, CWK602189S, CWK603119S, CWK603427S, CWK603647S, CWK605413S, CWK602192S, CWK601105S, CWK601937S, CWK600094S, CWK600096S, CWK602008S, CWK602015S, CWK602202S, CWK604953S, CWK602213S, CWK604936S, CWK604946S, CWK601212S, CWK601897S, CWK602006S, CWK603141S, CWK603429S, CWK604285S, CWK601208S, CWK601209S, CWK601210S, CWK601214S, CWK601885S, CWK601928S, CWK601941S, CWK602208S, CWK604148S, CWK604291S, CWK601706S, CWK601901S, CWK601939S, CWK601983S, CWK601986S, CWK602184S, CWK602201S, CWK604140S, CWK604151S, CWK605406S, CWK601117S, CWK601124S, CWK601701S, CWK603138S, CWK603428S, CWK604941S, CWK600118S, CWK600138S, CWK600082S, CWK601100S, CWK602018S, CWK600120S, CWK601098S, CWK601211S, CWK601695S, CWK601896S, CWK601984S, CWK602005S, CWK602216S, CWK604147S, CWK605420S, CWK603115S, CWK604295S, CWK604299S, CWK604301S, CWK602187S, CWK604294S, CWK605421S, CWK605408S; GTIN 00763000611958, Lot Serial Numbers: CWK605663S, CWK605659S, CWK606048S, CWK605596S, CWK606051S, CWK605791S, CWK605641S, CWK605647S, CWK605865S, CWK605627S, CWK605839S, CWK606046S, CWK605584S, CWK605878S, CWK605642S, CWK605629S, CWK605907S, CWK605675S, CWK605676S, CWK605677S, CWK605678S, CWK605679S, CWK605842S, CWK605843S, CWK605844S, CWK605845S, CWK605846S, CWK605797S, CWK605594S, CWK605644S, CWK605808S, CWK605794S, CWK605685S, CWK605646S, CWK605636S, CWK605597S, CWK605658S, CWK605795S, CWK605796S, CWK605872S, CWK605897S, CWK605587S, CWK605779S, CWK605792S, CWK605876S, CWK605652S, CWK605798S, CWK605834S, CWK605603S, CWK605602S, CWK605589S, CWK605674S, CWK605708S, CWK605686S, CWK605687S, CWK605689S, CWK605690S, CWK605691S, CWK605610S, CWK605655S, CWK606176S, CWK605611S, CWK605803S, CWK605927S, CWK605807S, CWK605612S, CWK605618S, CWK605634S, CWK605635S, CWK605672S, CWK605761S, CWK605648S, CWK605625S, CWK605633S, CWK605790S, CWK605890S, CWK605682S, CWK605619S, CWK605640S, CWK605692S, CWK605693S, CWK605698S, CWK605785S, CWK605786S, CWK605811S, CWK605933S, CWK605934S, CWK605662S, CWK605617S, CWK606127S, CWK606128S, CWK605684S, CWK605609S, CWK605653S, CWK605669S, CWK605670S, CWK605591S, CWK605643S, CWK605645S, CWK605661S, CWK605901S, CWK605871S, CWK605866S, CWK605893S, CWK605605S, CWK605783S, CWK605836S, CWK605837S, CWK605880S, CWK605701S, CWK605702S, CWK605703S, CWK605704S, CWK605921S, CWK605592S, CWK605735S, CWK605736S, CWK605737S, CWK605738S, CWK605739S, CWK605770S, CWK605771S, CWK605750S, CWK605751S, CWK605752S, CWK605753S, CWK605765S, CWK605766S, CWK605767S, CWK605768S, CWK605769S, CWK605758S, CWK605759S, CWK605760S, CWK605762S, CWK605747S, CWK605749S, CWK605757S, CWK605755S, CWK605756S, CWK606136S, CWK605323S, CWK605325S, CWK605585S, CWK605586S, CWK605588S, CWK605590S, CWK605595S, CWK605598S, CWK605599S, CWK605600S, CWK605601S, CWK605604S, CWK605606S, CWK605607S, CWK605608S, CWK605613S, CWK605615S, CWK605620S, CWK605623S, CWK605624S, CWK605628S, CWK605637S, CWK605638S, CWK605639S, CWK605649S, CWK605650S, CWK605651S, CWK605654S, CWK605656S, CWK605660S, CWK605671S, CWK605680S, CWK605681S, CWK605695S, CWK605699S, CWK605706S, CWK605707S, CWK605709S, CWK605764S, CWK605777S, CWK605780S, CWK605781S, CWK605788S, CWK605789S, CWK605799S, CWK605800S, CWK605801S, CWK605802S, CWK605804S, CWK605806S, CWK605809S, CWK605810S, CWK605812S, CWK605835S, CWK605838S, CWK605847S, CWK605864S, CWK605867S, CWK605869S, CWK605870S, CWK605875S, CWK605877S, CWK605879S, CWK605881S, CWK605882S, CWK605883S, CWK605884S, CWK605885S, CWK605886S, CWK605887S, CWK605891S, CWK605894S, CWK605895S, CWK605898S, CWK605900S, CWK605902S, CWK605904S, CWK605905S, CWK605908S, CWK605909S, CWK605910S, CWK605911S, CWK605912S, CWK605913S, CWK605918S, CWK605919S, CWK605922S, CWK605923S, CWK605924S, CWK605925S, CWK605928S, CWK605932S, CWK605955S, CWK606033S, CWK606035S, CWK606039S, CWK606041S, CWK606045S, CWK606047S, CWK606049S, CWK606050S, CWK606052S, CWK606054S, CWK606055S, CWK606057S, CWK606060S, CWK606061S, CWK606062S, CWK606063S, CWK606077S, CWK606078S, CWK606079S, CWK606058S, CWK606059S, CWK606064S, CWK606065S, CWK606066S, CWK606074S, CWK606076S, CWK605892S, CWK605906S, CWK605920S, CWK606034S
FEI Number 2182208
Recalling Firm/
Manufacturer
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
8200 Coral Sea St Ne
Mounds View MN 55112-4391
For Additional Information ContactTechnical Services
800-929-4043
Manufacturer Reason
for Recall
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
FDA Determined
Cause 2
Under Investigation by firm
ActionMedtronic issued an URGENT MEDICAL DEVICE CORRECTION NOTICE to its consignees on 05/10/2023 by UPS 2-day delivery or email. Consignees are asked to forward this notification to those who need to be aware within the organization and to any location which the devices have been distributed. The notice explained the problem and the risk and requested the following patient management recommendations: - Prophylactic device replacement is NOT recommended. - Program all HV therapy pathways B>AX in all therapy zones to minimize the risk for this issue. - Prioritize reprogramming patients who have both a history of HV therapy and Rx1 programmed AX>B. - For remaining patients with AX>B programming in any HV therapy sequence, schedule (with appropriate discretion) the next follow-up for in-clinic reprogramming to minimize potential for reduced- or no-energy HV therapies to occur. - Contact Medtronic Technical Services (1-800-929-4043) or your local representative if one of the following is observed as these may be an indication of either a device or lead-related issue: o Reduced- or no-energy HV therapy is displayed in Episode Text (regardless of programmed pathway) o A persistent drop of approximately 50% in RA, RV and LV pacing lead impedance measurements as this may be an indication of increased potential for a future reduced- or no-energy therapy. Update 10/10/2023: Beginning 09/27/2023, Medtronic issued communication via letter to consignees regarding the availability of the revised IFU and SmartSync Software. Consignees are asked to forward this notification to those who need to be aware within the organization and to any location which the devices have been distributed. Consignees are asked to complete and return a confirmation certificate (or equivalent record) confirming they have received the notification.
Quantity in Commerce5215 units
DistributionUS Nationwide - Worldwide Distribution
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
PMA DatabasePMAs with Product Code = LWS
PMAs with Product Code = LWS
-
-