• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 1 Device Recall VariSoft infusion set

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 1 Device Recall VariSoft infusion setsee related information
Date Initiated by FirmOctober 11, 2023
Date PostedNovember 21, 2023
Recall Status1 Open3, Classified
Recall NumberZ-0232-2024
Recall Event ID 93269
510(K)NumberK162812 
Product Classification Set, administration, intravascular - Product Code FPA
ProductVariSoft infusion set, Model Number 1002827
Code Information GTIN 05705244018709, Serial Numbers: 5388357, 5388362, 5388372, 5388366; GTIN 05705244018693, Serial Numbers: 5388357, 5388362, 5388372, 5388366
Recalling Firm/
Manufacturer
Unomedical A/S
Aaholmvej 1 - 3
OSTED
LEJRE Denmark
For Additional Information ContactMr. Henrik Gyllun
45 48 16 7000
Manufacturer Reason
for Recall
The Varisoft infusion set is used in conjunction with tandem insulin pumps. The defect reported was damage to the connector piece during manufacturing. This damage results in the connector becoming detached more easily from the insulin set than expected, requiring less force to disconnect than intended. Disconnection, when not detected, for example during sleep, leads to missed basal dosing. which in turn can result in elevated blood glucose and ketone level (nocturnal hyperglycemia). Diabetic ketoacidosis.
FDA Determined
Cause 2
Process change control
ActionUnomedical a/s issued an Urgent Field Safety Notice to its sole consignee on 10/11/2023 via email. The notice explained the issue, potential risk to patient, and requested the distributor to quarantine all affected product pending return and to notify their customers.
Quantity in Commerce8028 units
DistributionUS Nationwide distribution in the state of CA.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = FPA
-
-