| | Class 2 Device Recall COOK MEDICAL |  |
| Date Initiated by Firm | November 03, 2023 |
| Date Posted | December 01, 2023 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0459-2024 |
| Recall Event ID |
93326 |
| 510(K)Number | K171223 |
| Product Classification |
Stents, drains and dilators for the biliary ducts - Product Code FGE
|
| Product | QUANTUM TTC BILIARY BALLOON DILATOR, REF QBD-4X3, Order Number G22654; used to dilate strictures of the biliary tree |
| Code Information |
UDI/DI 00827002226548 (each), 10827002226545 (case), Lot Numbers: W4720848, W4721792, W4721795, W4722159, W4722981, W4723304, W4723846, W4724692, W4737576, W4738717, W4742525, W4743193, W4743605, W4743606, W4743607, W4743608, W4744022, W4744441, W4748043, W4748946, W4749748, W4750264, W4750265, W4750910, W4751772, W4751906, W4751914, W4753360, W4753361, W4753362, W4753365, W4753816, W4755138, W4756901, W4756903, W4756905, W4757732, W4758647, W4759511, W4759976, W4760715, W4761738, W4761740, W4761935, W4762150 |
| FEI Number |
1037905
|
Recalling Firm/ Manufacturer |
Wilson-Cook Medical Inc.
4900 Bethania Station Rd
Winston Salem
NC
27105-1203
|
| For Additional Information Contact | Cook Medical Customer Relations 800-457-4500 |
Manufacturer Reason for Recall | Nonconforming devices while manufactured correctly, do not comply to their design requirements. Potential harms associated with balloon detachment are endoscopic retrieval of an impacted or non-impacted object, bleeding caused by the retrieval device, or the incident can occur without injury and the object is left to pass naturally. |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | Cook Medical issued an URGENT: MEDICAL DEVICE REMOVAL notice on 11/03/2023 by mail. The notice explained the problem with the device, risk to health, and requested the following actions be taken:
Customer: Cease use, quarantine, and return all affected devices. If devices were further distributed, those to whom the devices were sent must be notified.
Distributor: Quarantine all affected devices, notify your customers to whom you distributed devices. Return all devices to Cook Medical. |
| Quantity in Commerce | 553 devices |
| Distribution | US |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = FGE
|
|
|
|