• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Axcel / Axcel

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Axcel / Axcelsee related information
Date Initiated by FirmNovember 08, 2023
Date PostedJanuary 19, 2024
Recall Status1 Open3, Classified
Recall NumberZ-0764-2024
Recall Event ID 93567
510(K)NumberK030906 
Product Classification Light, surgical, ceiling mounted - Product Code FSY
ProductMaquet Axcel / Axcel + OR Light Systems, Model Numbers and Catalog Numbers AXL 5001 - ARD567501961C, AXL 5001S - ARD567505961C, AXL 5002 - ARD567501962C, AXL 5002S - ARD567505962C, AXL 5501 - ARD567501960A, AXL 5501 - ARD567501960C, AXL 5501S - ARD567505960C, AXL+5001 DF - ARD567502961C, AXL+5001 SF - ARD567502861C, AXL+5301 DF - ARD567502971C, AXL+5501 DF - ARD567502960C, AXL+5501 SF - ARD567502860C, AXL+5531 DF - ARD567502930C, AXL50 RLC - ARDRLC309000C, AXL50 RLC - ARDRLC309006C, AXL5003 - ARD567505968C, AXL5003 US - ARD567505969C
Code Information All systems affected
FEI Number 3013876692
Recalling Firm/
Manufacturer
Getinge Usa Sales Inc
1 Geoffrey Way
Wayne NJ 07470-2035
For Additional Information ContactJessica Minaya
973-709-7634
Manufacturer Reason
for Recall
A potential for a light system to fall in the operating room.
FDA Determined
Cause 2
Equipment maintenance
ActionAn URGENT MEDICAL DEVICE CORRECTION letter dated 11/8/23 was sent to customers. Actions to be taken by the customer: Our records indicate that you have been identified as having purchased at least one model of one of these product ranges. Please identify your affected units using the referenced devices on page 1 and follow the instructions below. Please distribute this notice and forward the link below to access manuals to anyone within your organization who may need to use the manuals, and especially to your service or service provider who performs your maintenance. Light system fixing and other replacements: Regarding: - the fixing of the light systems: suspensions fixing screws, adapter fixing screws, bushing fixing screws - the brake screws - the safety segments - the batteries, a dedicated instruction to illustrate and facilitate understanding of what to replace and what to order for replacement has been issued. The instruction is available on a dedicated space on the web portal https://www.getinge.com/int/campaigns/maintenance-and-service-on-or-light-system/, and provided upon request by emailing MSA808092.sw@getinge.com Should you have questions or require additional information, please do not hesitate to contact Maquet SAS/Getinge or your local representative. Please complete and sign the attached MEDICAL DEVICE CORRECTION - RESPONSE FORM (page 7) to acknowledge that you have received this notification. Return the completed form to Maquet/Getinge by e-mailing a scanned copy to or-lights2023.qrc@getinge.com or by faxing the form to (866) 350-0897. If you are a distributor who has shipped any affected products to customers, please forward this document to their attention for appropriate action. Actions to be taken by Maquet SAS / Getinge: As the majority of customers may not have regular contact with Getinge services or authorized Getinge representatives and therefore may not be aware of current maintenance recommendations, Maquet SAS/Getinge is providin
Quantity in Commerce236793 units in total
DistributionUS Nationwide. Global Distribution.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = FSY
-
-