| Class 2 Device Recall Spectral CT |  |
Date Initiated by Firm | November 30, 2023 |
Date Posted | January 12, 2024 |
Recall Status1 |
Open3, Classified |
Recall Number | Z-0725-2024 |
Recall Event ID |
93640 |
510(K)Number | K203020 |
Product Classification |
System, x-ray, tomography, computed - Product Code JAK
|
Product | Spectral CT 7500, Software Version 5.0.0.X. Computed Tomography X-Ray System. |
Code Information |
Model No. 728333 and 728340; Unique Device Identifier (UDI) (01)00884838101111(21)10148, (01)00884838101111(21)10158, (01)00884838101111(21)10160, (01)00884838101111(21)10162, (01)00884838101111(21)10163, (01)00884838101111(21)10165, (01)00884838101111(21)10166, (01)00884838101111(21)10167, (01)00884838111103(21)397002, (01)00884838111103(21)397003, (01)00884838111103(21)397004, (01)00884838111103(21)397005, (01)00884838111103(21)397006, (01)00884838111103(21)397007, (01)00884838111103(21)397008, (01)00884838111103(21)397009, (01)00884838111103(21)397010, (01)00884838111103(21)397011, (01)00884838111103(21)397013, (01)00884838111103(21)397014, (01)00884838111103(21)397015, (01)00884838101111(21)10021, (01)00884838101111(21)10034, (01)00884838101111(21)10035, (01)00884838101111(21)10036, (01)00884838101111(21)10038, (01)00884838101111(21)10039, (01)00884838101111(21)10042, (01)00884838101111(21)10048, (01)00884838101111(21)10049, (01)00884838101111(21)10052, (01)00884838101111(21)10055, (01)00884838101111(21)10057, (01)00884838101111(21)10067, (01)00884838101111(21)10068, (01)00884838101111(21)10069, (01)00884838101111(21)10081, (01)00884838101111(21)10082, (01)00884838101111(21)10085, (01)00884838101111(21)10087, (01)00884838101111(21)10088, (01)00884838101111(21)10089, (01)00884838101111(21)10091, (01)00884838101111(21)10092, (01)00884838101111(21)10093, (01)00884838101111(21)10096, (01)00884838101111(21)10098, (01)00884838101111(21)10100, (01)00884838101111(21)10102, (01)00884838101111(21)10103, (01)00884838101111(21)10104, (01)00884838101111(21)10105, (01)00884838101111(21)10106, (01)00884838101111(21)10107, (01)00884838101111(21)10108, (01)00884838101111(21)10109, (01)00884838101111(21)10111, (01)00884838101111(21)10112, (01)00884838101111(21)10113, (01)00884838101111(21)10116, (01)00884838101111(21)10119, (01)00884838101111(21)10120, (01)00884838101111(21)10125, (01)00884838101111(21)10126, (01)00884838101111(21)10127, (01)00884838101111(21)10128, (01)00884838101111(21)10131, (01)00884838101111(21)10132, (01)00884838101111(21)10133, (01)00884838101111(21)10134, (01)00884838101111(21)10136, (01)00884838101111(21)10137, (01)00884838101111(21)10138, (01)00884838101111(21)10140, (01)00884838101111(21)10141, (01)00884838101111(21)10142, (01)00884838101111(21)10143, (01)00884838101111(21)10144, (01)00884838101111(21)10145, (01)00884838101111(21)10147, (01)00884838101111(21)10149, (01)00884838101111(21)10151, (01)00884838101111(21)10152, (01)00884838101111(21)10168, (01)00884838101111(21)10171, (01)00884838111103(21)397016, (01)00884838111103(21)397019, (01)00884838111103(21)397018, (01)00884838111103(21)397017, (01)00884838111103(21)397020, (01)00884838111103(21)397021, (01)00884838101111(21)10182, (01)00884838101111(21)10173, (01)00884838101111(21)10174, (01)00884838111103(21)397022, (01)00884838101111(21)10186, (01)00884838111103(21)397024, (01)00884838111103(21)397025, (01)00884838101111(21)10187, (01)00884838111103(21)397027, (01)00884838111103(21)397028, (01)00884838101111(21)10193, (01)00884838101111(21)10155, (01)00884838101111(21)10156, (01)00884838101111(21)10159, (01)00884838101111(21)10161, (01)00884838101111(21)10164, (01)00884838111103(21)397012, (01)00884838101111(21)10013, (01)00884838101111(21)10014, (01)00884838101111(21)10015, (01)00884838101111(21)10016, (01)00884838101111(21)10017, (01)00884838101111(21)10018, (01)00884838101111(21)10019, (01)00884838101111(21)10020, (01)00884838101111(21)10022, (01)00884838101111(21)10023, (01)00884838101111(21)10024, (01)00884838101111(21)10025, (01)00884838101111(21)10031, (01)00884838101111(21)10032, (01)00884838101111(21)10033, (01)00884838101111(21)10040, (01)00884838101111(21)10041, (01)00884838101111(21)10043, (01)00884838101111(21)10044, (01)00884838101111(21)10045, (01)00884838101111(21)10046, (01)00884838101111(21)10047, (01)00884838101111(21)10050, (01)00884838101111(21)10051, (01)00884838101111(21)10053, (01)00884838101111(21)10056, (01)00884838101111(21)10058.(01)00884838101111(21)10059, (01)00884838101111(21)10060, (01)00884838101111(21)10061, (01)00884838101111(21)10062, (01)00884838101111(21)10063, (01)00884838101111(21)10065, (01)00884838101111(21)10066, (01)00884838101111(21)10070, (01)00884838101111(21)10071, (01)00884838101111(21)10072, (01)00884838101111(21)10074, (01)00884838101111(21)10075, (01)00884838101111(21)10076, (01)00884838101111(21)10077, (01)00884838101111(21)10078, (01)00884838101111(21)10079, (01)00884838101111(21)10080, (01)00884838101111(21)10084, (01)00884838101111(21)10086, (01)00884838101111(21)10090, (01)00884838101111(21)10094, (01)00884838101111(21)10095, (01)00884838101111(21)10097, (01)00884838101111(21)10099, (01)00884838101111(21)10101, (01)00884838101111(21)10110, (01)00884838101111(21)10114, (01)00884838101111(21)10115, (01)00884838101111(21)10117, (01)00884838101111(21)10118, (01)00884838101111(21)10121, (01)00884838101111(21)10122, (01)00884838101111(21)10123, (01)00884838101111(21)10124, (01)00884838101111(21)10129, (01)00884838101111(21)10130, (01)00884838101111(21)10135, (01)00884838101111(21)10139, (01)00884838101111(21)10146, (01)00884838101111(21)10150, (01)00884838101111(21)10153, (01)00884838101111(21)396001, (01)00884838101111(21)396003, (01)00884838101111(21)10170, (01)00884838101111(21)10169, (01)00884838101111(21)10172, (01)00884838101111(21)10177, (01)00884838101111(21)10179, (01)00884838111103(21)397023, (01)00884838101111(21)10176, (01)00884838101111(21)10178, (01)00884838101111(21)10184, (01)00884838101111(21)10183, (01)00884838101111(21)10180, (01)00884838101111(21)10181, (01)00884838101111(21)10185, (01)00884838101111(21)10188, (01)00884838101111(21)10191, (01)00884838101111(21)10190, (01)00884838101111(21)10192, (01)00884838101111(21)10175, (01)00884838101111(21)10189; Serial No. 10148, 10158, 10160, 10162, 10163, 10165, 10166, 10167, 397002, 397003, 397004, 397005, 397006, 397007, 397008, 397009, 397010, 397011, 397013, 397014, 397015, 10021, 10034, 10035, 10036, 10038, 10039, 10042, 10048, 10049, 10052, 10055, 10057, 10067, 10068, 10069, 10081, 10082, 10085, 10087, 10088, 10089, 10091, 10092, 10093, 10096, 10098, 10100, 10102, 10103, 10104, 10105, 10106, 10107, 10108, 10109, 10111, 10112, 10113, 10116, 10119, 10120, 10125, 10126, 10127, 10128, 10131, 10132, 10133, 10134, 10136, 10137, 10138, 10140, 10141, 10142, 10143, 10144, 10145, 10147, 10149, 10151, 10152, 10168, 10171, 397016, 397019, 397018, 397017, 397020, 397021, 10182, 10173, 10174, 397022, 10186, 397024, 397025, 10187, 397027, 397028, 10193, 10155, 10156, 10159, 10161, 10164, 397012, 10013, 10014, 10015, 10016, 10017, 10018, 10019, 10020, 10022, 10023, 10024, 10025, 10031, 10032, 10033, 10040, 10041, 10043, 10044, 10045, 10046, 10047, 10050, 10051, 10053, 10056, 10058, 10059, 10060, 10061, 10062, 10063, 10065, 10066, 10070, 10071, 10072, 10074, 10075, 10076, 10077, 10078, 10079, 10080, 10084, 10086, 10090, 10094, 10095, 10097, 10099, 10101, 10110, 10114, 10115, 10117, 10118, 10121, 10122, 10123, 10124, 10129, 10130, 10135, 10139, 10146, 10150, 10153, 396001, 396003, 10170, 10169, 10172, 10177, 10179, 397023, 10176, 10178, 10184, 10183, 10180, 10181, 10185, 10188, 10191, 10190, 10192, 10175, 10189. |
Recalling Firm/ Manufacturer |
Philips North America Llc 222 Jacobs St Cambridge MA 02141-2289
|
For Additional Information Contact | Philips Customer Care Solutions 1-800-722-9377 |
Manufacturer Reason for Recall | Multiple software issues that affect device functionality. |
FDA Determined Cause 2 | Software design |
Action | The recalling firm mailed consignees URGENT Medical Device Correction notifications dated 11/30/23. Consignees are asked to identify affected devices by their UDI-DI, of which affected devices are identified as having a UDI-DI of either 00884838101111 or
00884838111103, and a Product Code of either 728333 or 728340. Consignees may continue to use affected devices until a permanent solution is installed utilizing the precautions outlined in Table 2 of the notice. Consignees are asked to complete and return the provided response form no later than 30 days from receipt of the recall notification. Additionally, the recall notification is to be circulated among all users, so they are aware of the software issues. Philips will contact consignees to schedule a time for a field service engineer to visit your site and install the solution. Consignees with any questions or concerns are to contact Philips Customer Care Solutions Center at 1-800-722-9377. |
Quantity in Commerce | 197 units |
Distribution | Domestic: Nationwide Distribution; Foreign: Australia, Austria, Belgium, China, Czech Republic, Denmark, France, Germany, Israel, Italy, Japan, Malaysia, Netherlands, Norway, Panama, Palestine, Portugal, South Korea, Spain, Switzerland, Thailand, United Kingdom. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
510(K) Database | 510(K)s with Product Code = JAK
|
|
|
|