• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Sekisui Diagnostics OSOM BVBLUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Sekisui Diagnostics OSOM BVBLUEsee related information
Date Initiated by FirmJanuary 09, 2024
Date PostedMarch 19, 2024
Recall Status1 Open3, Classified
Recall NumberZ-1311-2024
Recall Event ID 94024
510(K)NumberK050755 
Product Classification Test, vaginal, bacterial sialidase - Product Code MXB
ProductSekisui Diagnostics Control Kit OSOM BVBLUE Bacterial Vaginosis
Code Information Catalogue Number: 184; Lot No. CK233;
FEI Number 3004386693
Recalling Firm/
Manufacturer
Mckesson Medical-Surgical Inc. Corporate Office
9954 Maryland Drive
Deep Run Iii Ste. 4000
Richmond VA 23233
For Additional Information ContactRandy King
804-553-2281
Manufacturer Reason
for Recall
Facility Critical Temperature Excursions on 08-JAN 2024 in Walk-In refrigerator due to mechanical failures, causing temperatures to fall below the validated range of 35.6 - 46.4 Degrees Fahrenheit (2 - 8 Degrees Celsius) for up to 60 minutes .
FDA Determined
Cause 2
Storage
ActionMcKesson notified customers by phone January 9, 2024 with a follow-up call January 17, 2024 instructing customers of three scenarios based upon the products they received and the investigation in conjunction with relevant manufacturer. The firm also sent Urgent Product Recall Notifications via email on February 8, 2024 instructing customers to immediately examine inventory and quarantine all affected devices, complete and return the provided response form to MMSQRCProductReviews@Mckesson.com, and notify customers who you have further distributed recalled devices to of this recall and to segregate and quarantine any devices they have on hand. Replacement products will be provided and may have the same lot number(s) as the recalled items but are not affected by the recall and are acceptable for use. Customers with questions can contact McKesson directly by phone at (800) 688-8840. Representatives are available Monday to Friday, 8:00 am to 5:00 pm PST.
Quantity in Commerce1 unit
DistributionUS Nationwide distribution in the states of AZ, PA and TX.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = MXB
-
-