| | Class 2 Device Recall Exactech Equinoxe |  |
| Date Initiated by Firm | March 06, 2024 |
| Date Posted | March 29, 2024 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1412-2024 |
| Recall Event ID |
94092 |
| 510(K)Number | K113309 |
| Product Classification |
Prosthesis, shoulder, semi-constrained, metal/polymer cemented - Product Code KWS
|
| Product | Exactech Equinoxe CAGE GLENOID, CEMENTED, alpha curvature, Mates with:
a) 38, 41, 44 head, Small, Item Number 314-13-02,
b) 38, 41, 44 head, Medium, Item Number 314-13-03,
c) 38, 41, 44 head, Large, Item Number 314-13-04; Shoulder Arthroplasty |
| Code Information |
a) Item Number 314-13-02, UDI/DI 10885862172679, Serial Numbers: 2334190, 2334196, 2334178, 2334192, 2334193, 2330945, 2177477, 2177488, 2330864, 2330867, 2177487, 2330948, 2339049, 2452141, 2334179, 2334191, 2177486, 2334180, 2334181, 2334184, 2177485, 2330866, 2330869, 2330863, 2330949, 2330947, 2474424, 2479342, 2334204, 2330950, 2330951, 2334189, 2474441, 2474433, 2177481, 2460075, 2460078, 2177494, 2452137, 2330870, 2334185, 2452125, 2339045, 2339047, 2177484, 2334200, 2334194, 2334202, 2339052, 2177492, 2460068, 2474428, 2474435, 2474440, 2330865, 2541672, 2522113, 2452126, 2622872, 2330946, 2479332, 2479341, 2541668, 2334201, 2622842, 2479331, 2522100, 2622864, 2479325, 2479324, 2615282, 2594866, 2594857, 2334183, 2452142, 2622850, 2615269, 2452130, 2541662, 2334198, 2479346, 2594861, 2339048, 2522106, 2622824, 2622879, 2622818, 2330944, 2615262, 2479333, 2474439, 2622856, 2334182, 2460076, 2460072, 2460067, 2330943, 2460070, 2460071, 2541670, 2452132, 2177490, 2594858, 2622820, 2452138, 2334187, 2339053, 2622873, 2622838, 2622823, 2594860, 2622835, 2615271, 2522095, 2622816, 2460077, 2622826, 2622867, 2334197, 2522109, 2334177, 2622883, 2522105, 2622847, 2479347, 2634054, 2330868, 2474427, 2541663, 2452143, 2594872, 2615261, 2177495, 2622863, 2522111, 2474426, 2622854, 2622839, 2622865, 2522097, 2601953, 2334188, 2622881, 2634058, 2177483, 2634050, 2634053, 2474437, 2479343, 2622840, 2541673, 2622861, 2622855, 2615260, 2541671, 2622845, 2522112, 2339050, 2339044, 2339051, 2634052, 2479336, 2330871, 2460079, 2622817, 2541669, 2615275, 2622860, 2479319, 2615264, 2452136, 2541665, 2479329, 2622827, 2541676, 2452131, 2622834, 2474442, 2615280, 2615277, 2622880, 2479348, 2622825, 2479339, 2622836, 2479320, 2479351, 2334195, 2474423, 2474425, 2479338, 2474434, 2452139, 2594870, 2460065, 2177479, 2177482, 2479345, 2474438, 2622859, 2622837, 2622862, 2622853, 2615268, 2594859, 2622870, 2522108, 2622828, 2522101, 2522102, 2811926, 2700812, 2700821, 2811910, 2700819, 2770310, 2700813, 2770334, 2811984, 2689275, 2811929, 2770324, 2811914, 2811928, 2811978, 2770317, 2770307, 2689272, 2770316, 2770337, 2777060, 2811975, 2770336, 2689265, 2811923, 2811982, 2770303, 2700818, 2777339, 2770301, 2770330, 2777052, 2777342, 2811892, 2811901, 2777340, 2700814, 2770297, 2811919, 2777343, 2700808, 2689273, 2811995, 2811988, 2770309, 2777057, 2700809, 2811899, 2689269, 2700817, 2689270, 2811920, 2777055, 2770306, 2811904, 2811994, 2689276, 2770327, 2770314, 2777344, 2777053, 2777076, 2811970, 2689277, 2811897, 2811900, 2811891, 2811906, 2479323, 2177480, 2622858, 2334175, 2330942, 2474430, 2474429, 2622830, 2594865, 2622843, 2479326, 2622848, 2622819, 2622846, 2452144, 2615273, 2334199, 2460069, 2541664, 2622851, 2622841, 2460074, 2622884, 2634048, 2594869, 2479340, 2622878, 2479344, 2474431, 2622822, 2594863, 2615274, 2594862, 2334203, 2615278, 2594867, 2594868, 2339046, 2634049, 2479327, 2522099, 2594856, 2615266, 2479349, 2622832, 2479330, 2177493, 2615270, 2460073, 2452140, 2479350, 2479317, 2594864, 2893190, 2893189, 2893201, 2969469, 2811911, 2811903, 3556623, 2811922, 2893179, 2893210, 2770302, 2811925, 3588544, 3687003, 2770299, 2777073, 2992533, 2992522, 2992532, 2992544, 2777338, 2992519, 2893197, 3564166, 3636197, 2992503, 3564148, 3623108, 3623262, 3623260, 2914370, 3659299, 3714252, 3623250, 3714255, 2992512, 3636185, 2914401, 3556624, 2992534, 2893199, 2777059, 2770308, 2893188, 3623253, 2770298, 2914375, 2777337, 3636199, 2777068, 2992531, 3564158, 3623242, 3588542, 3636184, 3556620, 3636188, 2777058, 3636179, 3623240, 2992510, 2811979, 3564165, 2811908, 3556633, 2811989, 2992506, 3556638, 3588530, 3556621, 3636195, 3636250, 3659297, 2777062, 3623273, 3588525, 2969477, 2893192, 3636245, 3577208, 3686983, 3577214, 3556622, 2992502, 2969484, 2811976, 2811977, 2811990, 2777078, 2969479, 2770333, 2969487, 2992524, 2992527, 3623244, 3683868, 3683881, 3588539, 2893200, 2893191, 3588528, 2770305, 2893178, 2914394, 2689274, 2811913, 2811985, 2811924, 2777336, 3556636, 2992540, 2777077, 3623232, 3683871, 2811918, 3714277, 2811972, 2992536, 2992517, 3564154, 3588537, 3564155, 2770331, 3683875, 2893206, 2777071, 2893212, 3564151, 3623241, 3588527, 3556628, 3588536, 2811909, 3577203, 2811983, 2811991, 2914396, 2992523, 2811912, 3564152, 3659301, 3659315, 2992546, 3686988, 3686981, 3686980, 2992529, 2770318, 2811898, 2811917, 3564144, 2811902, 3564163, 3588531, 3623141, 2777063, 2992541, 2811980, 2700816, 2811996, 3556625, 3564153, 2969472, 3623233, 3683876, 3683886, 3623249, 3683863, 3588532, 2777075, 3556618, 2770319, 2689266, 2992499, 2914402, 2992515, 3659296, 2969468, 2969486, 2777054, 3564161, 2914378, 3623120, 3714270, 2914372, 3623107, 3577213, 3623239, 2992514, 3556634, 2969476, 3577217, 2914397, 3556635, 3623256, 2992497, 2969470, 3686986, 2777066, 2992528, 2992507, 2893177, 3683867, 3636239, 3636235, 2992509, 2992535, 3623109, 2811993, 2893194, 2811921, 2811986, 2992537, 2811888, 3577215, 3623278, 2969473, 2914379, 2992539, 3686995, 2770315, 3683862, 3588543, 2893202, 3588535, 2811905, 2777056, 2893193, 2770328, 2893198, 3636233, 2700810, 2914399, 3686989, 3686993, 2992545, 3623251, 3636182, 3577216, 3623277, 2893185, 2893180, 3577204, 3588538, 3683864, 2914377, 2777061, 3556630, 2914393, 2969474, 2811907, 3577219, 2777069, 2811890, 2914390, 3564157, 2777072, 3588524, 2992525, 3623248, 3636232, 3636240, 3623246, 3623264, 3683880, 2811974, 3636252, 3636246, 3686994, 2914371, 2914373, 2914381, 2914383, 2914385, 2914395, 3623114, 3636178, 2770326, 3564164, 3623103, 2770320, 3623238, 3636237, 3714286, 3636196, 3623236, 3686996, 3636249, 3556637, 2452129, 2334176, 2452127, 2452128, 2541667, 2615272, 2452133, 2474432, 2479334, 2177478, 2460066, 2634046, 2634059, 2522104, 2522098, 2479337, 3917688, 3917671, 2689268, 3797306, 3636198, 2914386, 3714280, 3910148, 2700815, 4026356, 2914384, 4049202, 3797272, 3868609, 3714242, 3917672, 2770312, 2992508, 4149744, 3683874, 4108638, 2969478, 3714266, 2811889, 3714239, 4099795, 3797273, 3577212, 3797282, 3917674, 4054492, 3797312, 4005358, 4108640, 4153670, 4182735, 3797304, 3797303, 3714268, 2992501, 2914387, 3623099, 3623118, 4032036, 4075948, 3623265, 3714285, 4121963, 4153677, 4108637, 4153683, 3623276, 3623275, 2893196, 3556631, 3837437, 3910190, 3837444, 4049216, 4149745, 3714251, 2969475, 2811981, 3577220, 2992516, 3837427, 4027063, 4049205, 3837430, 3714238, 3623105, 2914398, 3868620, 3910187, 4054504, 4041871, 4108649, 3797274, 3868614, 2969480, 2777064, 3623117, 2811895, 2992542, 3623112, 4091830, 3636243, 3659314, 3868617, 3623143, 3868652, 3917678, 3910157, 3917683, 4026361, 4054493, 3837439, 3868606, 4108646, 4182729, 3714244, 3623110, 3687000, 3623231, 3837424, 4018035, 3564150, 2770323, 3714279, 3683882, 3623270, 3868627, 4005359, 3683884, 2914388, 3659309, 4150348, 3910168, 4091573, 3623116, 3683865, 3797277, 3797281, 3868607, 3837446, 3910144, 3623252, 3714271, 2914374, 4049210, 4032034, 4054496, 4054498, 4075950, 4122650, 3623127, 4026346, 3868651, 3623135, 4026359, 3623137, 2770332, 3636192, 4027060, 3659303, 3659306, 3868616, 3636247, 3623106, 3556627, 3636228, 2969471, 3797314, 2893187, 3683870, 3910154, 4026355, 3797307, 4078828, 4121984, 3623125, 3564160, 3623268, 3623247, 3623254, 3910191, 4150343, 4159930, 4026347, 3837436, 3623243, 3636238, 3588533, 3714276, 4091829, 2992504, 3714258, 3683872, 3910183, 3910164, 4122651, 4149748, 4164644, 3623097, 3797297, 2811896, 2992521, 3868640, 3910149, 2893204, 3797283, 3910184, 3623098, 4027062, 4066121, 4075959, 4108643, 4066123, 4108642, 4121961, 3623100, 3623267, 2777335, 3714265, 3623140, 3687002, 3714248, 3556626, 3797310, 3623119, 3797302, 3683866, 3714267, 2992543, 3623139, 3797275, 3636244, 3868619, 3636200, 4075953, 4150336, 3623121, 3623266, 3659307, 3868647, 3687001, 3868611, 3910181, 3714282, 4063693, 3623130, 3623279, 3577210, 3714253, 3837441, 2777067, 3837428, 3623237, 3868654, 3837438, 2893209, 3837442, 3868608, 2893207, 4026351, 4041869, 3917687, 3837434, 4091827, 2969485, 3910171, 3564162, 3917684, 3623128, 3623131, 3797309, 3797287, 3868646, 4096972, 4149746, 4193209, 3797295, 3714263, 3917690, 3917667, 3910185, 4026364, 4075949, 3910182, 3636248, 3837445, 3910174, 4091569, 2914389, 2992511, 4026367, 4108650, 3623272, 3623113, 3714261, 3714243, 3868634, 3577205, 3868631, 3686987, 3868623, 3910173, 4018034, 4026349, 3925145, 3623123, 3686999, 3910189, 3623132, 4078824, 3910176, 3865649, 3865652, 3865648, 3865651, 3865650, 3925144, 3925143, 3865654, 3925142, 3865653, 2777074, 3623138, 3910156, 3910167, 4103638, 3659312, 2893182, 3623095, 3659302, 3623104, 4005360, 2992520, 3659313, 3564145, 3714241, 2770329, 3868625, 3556639, 4108641, 2777070, 2811893, 3868622, 2770325, 3917675, 3659308, 3910150, 3910152, 3623259, 2893186, 3837440, 3797289, 3683878, 3659298, 3623102, 3577207, 4026352, 3577202, 2770300, 3714274, 3683877, 3797308, 2992498, 3868615, 3868645, 3910153, 3636194, 3868643, 4018036, 3797276, 4049207, 4041868, 4075958, 4049204, 4091567, 4103640, 4153666, 4054494, 2811987, 3868621, 3564146, 4121966, 3910170, 3714281, 3714284, 3868618, 3714247, 3636251, 3837443, 3556629, 3797311, 3714245, 3868638, 3917680, 3917685, 3868629, 3868630, 3868639, 3868653, 4096968, 4096969, 4142039, 4142040, 4142041, 4142047, 4142049, 3714283, 3797269, 3797271, 3797279, 3797288, 3797291, 3797298, 3868636, 3868644, 3910178, 4153668, 4193211, 2893195, 3683869, 3683883, 3910159, 3714246, 3686992, 3623096, 3714272, 3797296, 3797301, 2893211, 4005353, 4096971, 4142043, 4171128, 3714254, 3910188, 3686984, 3714264, 3714259, 3714257, 3686997, 3686998, 3868637, 3623142, 3714237, 3917686, 3683879, 4005354, 2622866, 2622875, 2522110, 2622876, 2615267, 2615265, 2330952, 2615281, 2522103, 2452134, 2622869, 2479328, 2541666, 2622871, 2622882, 2622815, 2622821, 3636202, 3577206, 4224355, 4268139, 4268130, 3623245, 4108639, 4338475, 4338506, 2992526, 3797317, 4322912, 4121978, 4343674, 4502103, 3868649, 4343698, 3636201, 4193206, 4150346, 4338497, 4054501, 4150350, 4150351, 4242476, 4256865, 4193212, 3659304, 4242481, 4256862, 4280848, 4332482, 4063694, 4091574, 4049208, 4439549, 4268149, 4343673, 4545870, 4193199, 4153674, 4228090, 4182724, 4228094, 4256875, 4054500, 3917676, 4121976, 4121967, 4164643, 3917677, 4159931, 4242474, 4502094, 4332512, 4193194, 4153665, 4149750, 4149742, 4149741, 4332479, 4332520, 4228080, 4193221, 4153673, 2689279, 3623111, 4049214, 4298139, 2992530, 4193204, 4298147, 4343661, 4343699, 4531850, 4190753, 4091824, 4256876, 4256860, 4228089, 4224351, 4343652, 4268158, 4280851, 4332499, 4298134, 4121977, 4428868, 4343654, 4502097, 4469375, 4159932, 4193214, 4228085, 4075952, 4280841, 4338494, 4322933, 4193195, 4054503, 4121971, 4268168, 4153672, 4256869, 3797293, 3868642, 4026362, 4041872, 4228096, 4268146, 4343653, 4298142, 4332380, 4428871, 4332381, 4182730, 4121980, 4338489, 4121982, 4121965, 4121983, 4185996, 4164639, 4268135, 4268151, 4256877, 4256859, 4150344, 4298140, 4256880, 4153678, 4332390, 4332374, 4332517, 4332514, 4428880, 4121975, 4049211, 4149749, 4343700, 4428869, 4078825, 4268153, 4439543, 4005355, 2914382, 4032033, 4091825, 4338499, 2770321, 4239209, 3623271, 4428879, 4216357, 4239208, 4280844, 4332372, 4332378, 4531854, 4091823, 4256872, 4280843, 4150330, 4193215, 4224352, 3564147, 4268132, 3623134, 4280849, 4103639, 4256871, 4343677, 4531851, 4159927, 4268164, 4103643, 4099796, 4182734, 4332491, 4332515, 3623126, 4193219, 4242480, 4332497, 4338484, 4469377, 4182731, 4193202, 4224183, 4121979, 4224353, 4268156, 4332480, 4054497, 4256857, 4502089, 4332496, 4428875, 4298149, 4428870, 4428881, 4469354, 4108644, 4159934, 4054499, 4338486, 2969483, 4469362, 4150345, 4121974, 4193200, 4298135, 3714262, 3623258, 3636181, 3910186, 3910163, 3868626, 4228077, 3837423, 4338487, 4545859, 3910172, 3686982, 3868613, 4224189, 4332386, 4026350, 4343667, 4298137, 4099797, 4150334, 4193207, 4091575, 4091576, 4268144, 4268161, 4298145, 4332382, 4256866, 2893205.2811973, 4159926, 4054495, 4026363, 4027064, 4054502, 4049215, 4332387, 3917673, 4322906, 4091831, 4469369, 4338490, 4439562, 4032035, 4091572, 4164640, 4428877, 3797305, 4242477, 3797294, 3910158, 4343682, 4469373, 4164636, 4078827, 4256858, 4268134, 4343665, 3577218, 4121969, 4338483, 4075947, 4193197, 3910169, 3910192, 4502104, 3910165, 3910155, 4075951, 3588523, 4164641, 4153682, 3868624, 3868650, 4545881, 4256868, 4428885, 3917668, 3659310, 4268141, 3623101, 3686985, 4150337, 4268143, 3623269, 3868612, 4228084, 4502084, 4239210, 3588546, 3910180, 2770304, 3868633, 4224357, 2893181, 4256873, 4224181, 3623133, 3623261, 3868628, 4103641, 4228088, 4032037, 4149738, 4332473, 3686991, 4322935, 4256881, 4338507, 3686979, 3636180, 4171131, 3797290, 4171135, 4121968, 4343675, 4332371, 4228095, 4298138, 4224356, 4343687, 4193220, 4182732, 4153680, 2914392, 3910177, 4322907, 4164642, 4103642, 4193201, 3868641, 4091568, 4322917, 2914391, 3623234, 4027061, 4078826, 4332368, 4268150, 4026344, 4099800, 4343685, 3636186, 4153679, 3714249, 3797278, 4242478, 3623257, 3636242, 4121981, 4343671, 3588540, 4256874, 4026345, 4153681, 4343666, 4531843, 3623274, 4153684, 3636189, 3577211, 4268163, 2893203, 4099794, 4150338, 4041873, 3910179, 2914380, 4108647, 4469360, 2777065, 4343694, 4343655, 2992500, 4343656, 4298130, 3797316, 4242473, 4182722, 4193203, 4256861, 4322936, 4343664, 4322934, 4159933, 4469359, 4343691, 4343657, 4332477, 4343651, 3910160, 3714256, 4343692, 3636241, 4256870, 4224182, 4224188, 4268145, 4268157, 4268160, 4298132, 4322922, 4332519, 4343659, 4332389, 4469355, 4322924, 4439552, 4268123, 4268147, 4280839, 4332495, 4332500, 4332521, 4332385, 4332511, 4343658, 4343662, 4439538, 4439555, 4469356, 4469368, 4531846, 4439539, 4439540, 4190754, 4228078, 4228081, 4193208, 4193205, 4171133, 4193216, 4228086, 4171134, 4224186, 4322923, 4322914, 4224187, 4268159, 4005351, 4142045, 2893183, 3797300, 3797315, 4216354, 4268167, 4298136, 3714260, 4186002, 4322925, 4228083, 4193218, 4580522, 4580518, 4580542, 4644621, 4644572, 2522096, 2479321, 4332522, 3636183, 4545865, 4268148, 4338477, 4542821, 4545863, 3714273, 4439559, 4332367, 3868635, 3868632, 4531861, 3868648, 4338482, 4531847, 4122652, 4428882, 4026357, 3588541, 4091570, 4439550, 4542818, 4542802, 4439553, 4150341, 4531849, 4150332, 4343695, 4502093, 4343684, 4542801, 4185998, 4256863, 4343697, 4469358, 2770311, 4542807, 3636231, 4108648, 4343688, 4268131, 4545861, 4228091, 3623136, 3797313, 4216356, 4149747, 4322909, 4268137, 4185997, 4190755, 4439556, 4542797, 4298131, 4280845, 3564159, 4075954, 4026360, 4150327, 4338504, 4239207, 4343647, 4545869, 4545875, 4322928, 3917669, 4332391, 4343689, 4322915, 4542798, 4224354, 3714278, 4153676, 4338496, 4332494, 4542819, 3910147, 4150333, 4439558, 4268166, 3636190, 4428873, 2893184, 4502081, 4338500, 4545866, 3910146, 4542816, 4531858, 4439551, 4150329, 4531863, 4531862, 4542810, 4545874, 4268162, 2893208, 4332388, 4502096, 4469374, 4502092, 4502083, 4338476, 3636229, 4428865, 4338481, 4332501, 4005356, 4332505, 4502090, 4502088, 2689278, 4026358, 3797299, 3910145, 4332366, 4502098, 4268127, 2969481, 4531852, 4531853, 3623124, 4075955, 4542814, 4545857, 4332507, 4502082, 4545858, 4193193, 4502087, 2811971, 4322929, 4322932, 4531844, 4343649, 4332370, 3659300, 3917679, 4542806, 4545873, 2777341, 3636234, 4428876, 4268125, 4545860, 4185999, 4182733, 4332489, 4171132, 4428886, 4159928, 3686990, 3910151, 4439544, 4545879, 4096970, 3588545, 4542812, 3556619, 4298143, 4428888, 3910193, 2811992, 4322911, 4150335, 4005357, 4332369, 2914376, 4332510, 3623235, 3577209, 4338478, 4091826, 3910162, 4469367, 4256878, 3837431, 4343670, 4542813, 4142044, 4005349, 4256879, 4280842, 4228093, 4428872, 4531864, 3925141, 4224350, 4228087, 4332475, 4149737, 4108645, 4268154, 4149740, 3623129, 4439545, 4428883, 3623230, 4322916, 4439541, 4343648, 4428887, 4542820, 3910161, 4066122, 4343679, 4332373, 4502086, 4531860, 4268122, 4268124, 3797284, 4531857, 4224349, 4005350, 4332513, 4121960, 4531855, 4332384, 4502105, 4545867, 3623255, 2689267, 4026354, 4099799, 3683885, 4026343, 4545862, 3659311, 4041874, 4322908, 3564149, 4343676, 4542815, 4224185, 4439560, 4469361, 4332478, 4542808, 4338493, 4542817, 3917682, 4005352, 3910175, 4280850, 4545880, 3797286, 4332379, 4322921, 4332376, 4332502, 4228079, 4531841, 4224184, 4545877, 4343693, 4224358, 4332383, 4171130, 4343650, 4439557, 4542805, 4644608, 4761982, 4775452, 4580525, 4644591, 4811690, 4983282, 4761990, 4993513, 4811702, 4669143, 4735121, 4775457, 4775462, 4795083, 4811684, 4644586, 4950873, 4826677, 4775471, 4644618, 4669131, 4826690, 4896822, 4993517, 4950856, 4580523, 4580528, 4669133, 4795061, 4684328, 4756784, 4888720, 4756794, 4826689, 4811708, 5068276, 4644612, 4644601, 4580539, 4580524, 4644585, 4684321, 4756788, 4761977, 4756783, 4761981, 4826693, 4897256, 4669136, 4947162, 4826698, 5013054, 4947136, 5137144, 4644595, 4756772, 5073114, 4826674, 4756790, 5041043, 4795067, 4669127, 4950883, 4811712, 4897270, 5073097, 4669129, 5137141, 4795074, 4826685, 4775448, 4735120, 4775468, 4826681, 4861719, 4756778, 4950857, 4963676, 4826696, 4963674, 4580537, 4580519, 4684329, 4897283, 4684316, 4947161, 4947163, 5137152, 4644590, 4684319, 5068279, 5073121, 4983288, 5073109, 5068283, 4861718, 4950863, 4580527, 4669126, 4644604, 4669139, 4775467, 5041057, 4897257, 4950871, 5068294, 4669124, 4644619, 4795089, 4684332, 4669140, 4847765, 4644597, 4826683, 4861733, 4947157, 4896843, 4897262, 5068298, 4896834, 4861734, 4896826, 4847787, 4896844, 4756781, 4756791, 4847780, 4644603, 4896832, 5068300, 4580533, 4644609, 4669142, 4761972, 4888716, 4811703, 5068286, 4993521, 4861724, 4775461, 4826686, 4811693, 4861716, 4993511, 4756787, 4580520, 4775449, 4775472, 4669125, 4826701, 5068275, 4756793, 4669141, 4580521, 4896819, 4644589, 4644613, 4669123, 4775458, 4826682, 4795062, 5073120, 4896827, 4947152, 5068293, 4795063, 4811698, 4684339, 4684333, 4811689, 4993512, 4861735, 4897254, 4993527, 5073116, 4811709, 4795084, 4580529, 4993520, 4756786, 4826691, 4811699, 4644578, 5068303, 4644587, 4644607, 4811688, 4795065, 4950859, 4795068, 4897274, 4795085, 4896825, 4963666, 4775475, 4775469, 5137163, 4897271, 4644581, 4684331, 4761975, 4947149, 5073101, 4669120, 4580538, 4644576, 4684317, 4684327, 4847773, 4811706, 4811694, 4950866, 5013051, 4580535, 4963692, 4896824, 4861725, 4669121, 4644616, 4826700, 4861731, 4826694, 4644605, 4580532, 4580541, 4644584, 4888721, 4896821, 4847768, 4756773, 4669138, 4826678, 4896841, 4775474, 4644583, 4669135, 4756777, 5073124, 4795076, 4669122, 4826688, 4947156, 4775460, 4811683, 4775459, 4795075, 4847779, 4861728, 4811711, 4847774, 4896833, 4861729, 4896823, 4950855, 4947139, 4963685, 4983271, 4950867, 4861743, 5041051, 5133977, 4963667, 5133999, 4684335, 4756795, 4684325, 5068299, 4756774, 4983273, 5073096, 4847786, 4983279, 4644579, 4644593, 4644592, 4756782, 4756789, 4775456, 4775450, 4795081, 4795082, 4847776, 4983284, 4963665, 4963681, 4761973, 4644573, 4580534, 4761976, 4993503, 4580536, 4644599, 4644596, 4795077, 4684323, 4684320, 4761994, 4811697, 4896845, 4897273, 4897276, 4897280, 4811691, 4847788, 4963677, 4761988, 4775455, 4993519, 4847777, 4847771, 4761989, 4268140, 4026365, 3588534, 4268126, 4338491, 4182727, 4338488, 4542811, 3636230, 4469357, 2914400, 4228082, 4469363, 3868610, 4268152, 4322926, 4164637, 4164635, 4026366, 4332498, 4469365, 4338474, 4502091, 4469372, 4332490, 3868655, 3556632, 3917670, 4049209, 4322930, 3577201, 2969482, 4228092, 4242479, 4332483, 3837432, 4332492, 4332488, 4502101, 4322918, 4280838, 4545876, 4121973, 4332493, 4332516, 4091571, 4298146, 4069142, 4439561, 4150328, 4343672, 4182723, 4322927, 2992518, 4338498, 4439546, 4343681, 3588526, 4280846, 3837426, 4343686, 3714250, 4502100, 3797292, 4545872, 3714275, 4150331, 4193198, 4343680, 4121964, 4332481, 4280847, 4193217, 3917689, 3797285, 4669137, 5208236, 4896836, 4947138, 5546939, 4847761, 4644606, 5325283, 4993531, 5339822, 5068278, 5180241, 4993518, 5319253, 5068284, 5180236, 4897260, 4963694, 5339833, 5507176, 5180223, 5228452, 5309824, 4684318, 5240328, 5228455, 5546951, 4826699, 4993504, 5433408, 5133995, 4644580, 4947158, 5310210, 4993528, 5180248, 4983266, 5073105, 4983275, 5470168, 4947137, 5310226, 5328029, 5310821, 5404226, 5339824, 4580530, 4947144, 5068295, 5180252, 5073117, 5197038, 5240333, 5228447, 5180230, 5228441, 5228445, 5228437, 5328036, 5137161, 5309836, 5352708, 4896816, 5328027, 5180267, 5228433, 5310834, 5309835, 5479369, 5479366, 5339828, 5309843, 4861720, 4847778, 5319241, 5507186, 4861715, 5180221, 4756775, 4644598, 4888722, 4756776, 4950854, 4947143, 4775473, 5228434, 4847763, 5228440, 5228436, 5483531, 5309830, 5479360, 4861741, 5310215, 4983286, 5319251, 5133978, 5137143, 4795070, 5328033, 5310238, 4993505, 5137151, 4983278, 5228451, 5404199, 4983263, 5339834, 5546934, 5339829, 5309827, 4888718, 4644617, 5228435, 4963678, 4761980, 4950865, 5068291, 5208224, 5228426, 5328034, 5339832, 4795069, 5137160, 4993516, 4993526, 4963693, 5240342, 5073113, 5208221, 4963686, 5310817, 5433402, 5041044, 5068274, 5134002, 4993507, 5068281, 4847769, 5197047, 4684336, 5240325, 5310224, 5309832, 4950880, 4896840, 5339827, 4950870, 4847766, 5483521, 5180265, 5041053, 4795080, 5180253, 5068302, 5479348, 5354600, 5073106, 5041045, 5133982, 5073123, 5309828, 5309833, 4861740, 5197044, 5328028, 5507177, 5470155, 5433398, 5470173, 5352697, 5328009, 5339815, 5354608, 5352692, 5240334, 4756796, 5137154, 5354605, 5133974, 4963688, 4897277, 5197031, 4947150, 5483519, 5507199, 5180225, 4896842, 5073110, 4826675, 5479363, 5310233, 4826702, 5546935, 5180245, 5133980, 5339821, 5180272, 4644577, 5319231, 5240327, 4826697, 5073108, 5068290, 4963671, 5433391, 5433395, 5180271, 4950862, 4847785, 4761985, 5339835, 4950869, 5041034, 5180234, 5228446, 4811707, 5319240, 5208234, 5310225, 5325261, 5310844, 5507180, 5404204, 5470144, 5339816, 4795064, 5133993, 4993522, 4888715, 4775465, 5479361, 5470147, 5180244, 4847781, 4950879, 5208214, 5208223, 5404202, 5546946, 5483534, 5310230, 5325267, 4756780, 5197036, 4795072, 4684322, 5228432, 5133990, 5309848, 5339836, 5137148, 4669132, 5197026, 5470145, 4897266, 4861727, 5310840, 4644575, 5310234, 5133998, 4897267, 4896839, 4847770, 5339837, 5073122, 4669144, 5133975, 4963682, 5404206, 5470159, 4795073, 4795087, 5470172, 5180256, 4795079, 5404225, 4950877, 5133989, 5137158, 5404198, 4963668, 4947141, 5068292, 5133996, 5310842, 5339830, 5228450, 5433409, 5041054, 5068287, 5180222, 5309825, 5310232, 5319250, 5483525, 4847767, 5228453, 5041052, 5433411, 5310221, 5068277, 5180275, 5180264, 5228444, 5404208, 5483536, 5433418, 5507205, 5240347, 5546936, 5404201, 5068282, 5240337, 5041041, 5068285, 5208233, 5228429, 5470158, 5470146, 4897259, 4983269, 5240345, 4811700, 4896831, 5134001, 4947159, 5133987, 5180257, 4950861, 5352709, 4947146, 4756792, 4861738, 4826692, 5180261, 5433393, 5328013, 5328014, 5470160, 5354614, 5433407, 4861742, 5041059, 5339818, 5433405, 4947160, 4644594, 5073100, 5133976, 5208226, 5137150, 4811705, 5328021, 5352699, 4861717, 4983274, 4947134, 5073107, 4775453, 5133986, 5352700, 4983272, 5208235, 5352702, 5310227, 4811685, 5507197, 5309826, 5041042, 5228448, 5240324, 5180279, 5309834, 5328018, 5433403, 5339826, 4847760, 4993515, 4847762, 5352694, 5041035, 5180262, 5208210, 5352704, 5404224, 5180251, 5208222, 5328019, 5328035, 5352706, 5309847, 5404203, 5546949, 4775446, 5470157, 4993506, 4761987, 5325282, 5180246, 5180242, 5325274, 5325275, 5325277, 5433396, 5180235, 5137146, 5208211, 5325260, 5240330, 5325262, 4896818, 5041050, 4811696, 5180254, 5328015, 5352703, 5339842, 5180260, 5180231, 5433416, 5479359, 5309838, 4963670, 5180280, 5325272, 5325269, 4897265, 5208218, 4896820, 5325280, 5352693, 4684337, 5073112, 5649890, 5652788, 5652800, 5652805, 5652795, 5602761, 5649885, 5652792, 5652813, 5652791, 5649882, 5649891, 5649887, 5580389, 5580396, 5580409, 5580394, 5580406, 5580392, 5652803, 5652816, 5580401, 5602741, 5649893, 5602746, 5602732, 5580386, 5602763, 5580404, 5424411, 5424400, 5424388, 5424384, 5424399, 5424398, 5424410, 5424408, 5424385, 5424396, 3636187, 3623263, 4150342, 4469366, 4193192, 4159929, 4149739, 4150347, 4469364, 4322910, 4338492, 3910166, 4545864, 4150349, 4182725, 4322913, 4268133, 4531859, 4428878, 4153664, 4121962, 4338495, 4182726, 4026353, 4428866, 4545878, 4531845, 4439548, 4469376, 4268142, 4193213, 4343663, 4343683, 4142046, 4121972, 4531848, 3837433, 4164634, 4322920, 4338502, 4542799, 3683873, 5319252, 5228427, 4983270, 5319258, 5310830, 5310831, 5041040, 4983283, 5041037, 5470164, 5470165, 5339839, 5546944, 5180274, 5068288, 5310229, 5470170, 5319255, 5328026, 5339814, 4896828, 5197024, 5310839, 5310828, 4826695, 5479347, 5483533, 5180255, 4826673, 5483520, 5180238, 4669134, 5507179, 5240340, 5433390, 5339838, 5041036, 5068289, 5507194, 5404211, 5354593, 5208231, 5404222, 4795086, 5180269, 5325266, 5507196, 5319234, 5309831, 5228443, 5309846, 5310835, 4826676, 5354589, 5309837, 5483524, 4684338, 5197042, 5197027, 4963687, 4761991, 5197030, 5546932, 5180228, 5352707, 5180250, 5319233, 5208232, 5208216, 5328020, 5507185, 4811704, 5328037, 4761996, 5309849, 4963669, 5197050, 5470150, 4761983, 5041039, 5546942, 5310220, 5137140, 4684334, 5339819, 5404216, 5133988, 5546927, 4756779, 5433394, 5180258, 5507201, 4847775, 5470161, 5197051, 5310837, 5310235, 4947155, 5310231, 5352705, 5479368, 5546928, 5310816, 5507174, 5197043, 5137156, 5310214, 5041060, 5133991, 4983287, 5404212, 4897282, 5041047, 5137145, 5404213, 5068301, 5137147, 5354597, 5339831, 5352695, 5507190, 5433401, 5546947, 5352698, 5479352, 5339841, 5354609, 4861739, 5319243, 5507204, 5433414, 5433419, 5208213, 5208212, 5483522, 4826679, 5507198, 5354604, 5180232, 5319230, 5319237, 5208225, 5479353, 5197052, 5197023, 5483518, 5319245, 5479370, 5309851, 5354601, 5328010, 4993508, 4644588, 5133983, 5354587, 4811710, 4950874, 5404214, 5470167, 5309823, 5180263, 5470156, 5133994, 5507182, 4861737, 5479354, 5208219, 5208230, 5197029, 4847764, 5197028, 5197037, 5073115, 4847782, 5180229, 5309841, 5470166, 4963680, 5133984, 5310832, 4993514, 5310822, 4811701, 5433410, 5137164, 5310239, 5137157, 5546950, 4795078, 4580531, 5310223, 5325268, 5483538, 5433400, 5433412, 5073119, 5137155, 5354596, 5470153, 5546952, 5507188, 5404221, 5339817, 5328012, 5354586, 5319238, 5228449, 5354606, 5068280, 5228428, 5546925, 5309840, 5325279, 5197033, 5180233, 5228430, 5137149, 5507203, 5240326, 4761995, 4947153, 5240329, 4950876, 4861723, 4775463, 5310845, 5180247, 5483527, 5339825, 5328023, 4644620, 5546938, 4795071, 5354594, 5404210, 5354611, 5433404, 5309844, 5507183, 4761992, 4897275, 5328016, 5180273, 5546943, 5310824, 5073103, 5546945, 5240344, 5546926, 5546953, 5240346, 5310818, 5354592, 4684324, 5319236, 5479346, 5354590, 5325270, 4963673, 5479350, 5354615, 5479362, 5507195, 5325263, 5404200, 5325264, 5580400, 6055319, 5820908, 6043711, 6036342, 5950779, 5820876, 5950745, 6058255, 6055344, 6058269, 5820914, 5956398, 5950751, 5956406, 6003185, 6043714, 6102033, 6043728, 5580388, 6068613, 5820890, 6262828, 5950752, 6130377, 6055326, 5649886, 5649884, 5599194, 5950740, 6058237, 6055298, 6055289, 5599207, 6068614, 5950741, 6068635, 5599211, 6043727, 6068624, 6036345, 5820895, 5599208, 5602731, 5820913, 5599212, 5820884, 5913028, 6003191, 6055294, 5956391, 6055328, 6055351, 6102032, 6150634, 5652789, 5950783, 6183217, 5950742, 5599189, 5599213, 5820907, 6055342, 5602736, 5913026, 5956384, 5956388, 6058247, 6179534, 6102023, 6055347, 5820880, 6043708, 6058249, 6058234, 6150638, 5649881, 6055354, 5820920, 5843498, 6003189, 5580413, 5820898, 5820930, 5956399, 5820915, 6055356, 6055395, 6055286, 6068617, 5599205, 5599182, 6058252, 5843503, 5599198, 6055403, 6055334, 6055329, 5602757, 5913031, 5820874, 5649875, 5652790, 5580403, 6058254, 5649888, 5950774, 5820917, 5843505, 5913034, 5820919, 5602733, 6058231, 5599185, 5950747, 6043720, 6043722, 5950732, 6183228, 6055284, 6102004, 5820902, 6150637, 5602754, 5649870, 5580399, 5652799, 5652793, 5956390, 5602755, 5602760, 5580391, 5820911, 5820901, 5950758, 5649892, 6036343, 6055375, 6055343, 5950786, 6055373, 6102017, 5950755, 6068636, 6058245, 5950769, 6102040, 6102012, 6183225, 5602748, 6058253, 5580408, 5820900, 5950773, 6055332, 6058235, 5820906, 5649865, 5820873, 5843504, 5843511, 6003183, 6068641, 5956392, 5580405, 6043723, 5913033, 5580385, 5599197, 5913035, 5950777, 6003182, 5950744, 6043704, 6058222, 6058226, 6068621, 5580395, 5820888, 6055318, 5820931, 6058221, 5820899, 5843507, 5956404, 6055350, 6102009, 6102026, 6238053, 5580393, 5602762, 5950787, 5950761, 6043721, 5652804, 6238071, 5649867, 5602752, 5843501, 6055372, 6179522, 5820881, 5843500, 5820883, 5652801, 6058233, 5950760, 6058258, 6055376, 6055300, 5843509, 5820882, 5913036, 6055397, 6055327, 5820905, 6150647, 5599181, 6102035, 6055301, 6058243, 6055307, 6102022, 6055310, 5820875, 5950782, 6003187, 5843510, 5950750, 5652796, 5602745, 5950768, 5950784, 5599209, 5602749, 5820878, 5820924, 5602764, 6055295, 6102039, 6055341, 6003179, 6055391, 6058240, 6183241, 5602737, 5580398, 6055400, 6237393, 5820926, 5580387, 5956402, 5652808, 5652807, 5652797, 5580411, 5913032, 6055387, 6058251, 5649866, 6150640, 5820927, 5820918, 6003195, 6055335, 6036346, 6055331, 5602739, 5950728, 5580410, 6130400, 6237397, 5580384, 5843506, 6055358, 5950731, 6055313, 6055337, 6055394, 5820879, 6003184, 5913030, 6055359, 6055282, 5950736, 6058268, 6150639, 6150623, 6068629, 6130389, 6058225, 6238055, 6058238, 5599199, 6130375, 5649879, 6055348, 5599214, 5956397, 6102006, 6130396, 6102020, 6055396, 5950737, 5950738, 6055308, 6055314, 6055315, 6055320, 6055321, 6055346, 6055352, 6150635, 6150649, 5599186, 5599192, 6102028, 6179507, 6179525, 5950781, 6055345, 5820896, 5599204, 6068633, 5599188, 5599203, 5950748, 5599206, 6237389, 5599201, 5820886, 6043718, 5950770, 5820887, 6043701, 5599210, 6068631, 5599180, 5599190, 5950734, 5599196, 6043729, 6043705, 6036348, 5599179, 6043726, 6043716, 5820885, 6043724, 6043725, 5820891, 5424386, 5424391, 5424402, 5424397, 4322905, 4150339, 4332474, 4542800, 4531840, 4343669, 4338479, 4343678, 4469378, 4338505, 4164638, 4338501, 4298148, 4428884, 4439542, 4186001, 5483528, 5433397, 5483544, 5483526, 5354610, 5483543, 4896829, 5180224, 4950878, 5319247, 5328025, 5310212, 5240335, 5339843, 4826680, 5483541, 4861744, 4897272, 4644582, 5310228, 4861722, 5197034, 4775447, 4897261, 4983264, 4644610, 5404217, 5310820, 5319249, 5483523, 5507193, 5354607, 5546924, 5546930, 5310825, 4963679, 5310826, 5310841, 4983276, 5507189, 5546929, 5310836, 5404223, 5073118, 5228431, 5483540, 4775454, 5354602, 5180237, 5180240, 5479367, 4983277, 5328022, 5208209, 5546948, 5240336, 5483539, 5073104, 4963690, 5483530, 5339823, 5180259, 5240341, 4896835, 4644600, 4963691, 5483537, 5470151, 4950864, 4761986, 4761974, 5310823, 4811687, 5479343, 5479351, 5328017, 5470152, 5325281, 5354598, 5843502, 6300652, 6439498, 6251136, 6424033, 6590998, 6591020, 6036341, 6448669, 6238056, 6043706, 6036344, 6448661, 6238048, 5950754, 6550346, 6237383, 6262840, 6238067, 5649877, 6386249, 6238070, 6043717, 6590984, 6336730, 6102042, 6238046, 6300650, 6237392, 6557608, 5820889, 6390548, 6262839, 6300658, 6055357, 6179516, 6273525, 6130376, 6424032, 6150648, 6439562, 6465994, 6465990, 6557567, 6390524, 6466007, 6550376, 6557574, 6557592, 6424087, 6550380, 6557583, 6130385, 6237395, 6102005, 6341991, 6300672, 6390532, 6550345, 6386242, 6068626, 6179524, 6390554, 6055360, 6390549, 6424039, 6345338, 6550342, 6424054, 6390531, 6130391, 6130387, 6262819, 6439487, 5913037, 6336707, 6325744, 6386237, 6262833, 5599187, 6003178, 5649871, 6273536, 6550338, 6068615, 6238054, 6055280, 6341988, 6439554, 6342013, 6150652, 6237404, 6251145, 5956389, 6273531, 6386238, 6386232, 6390537, 6390521, 5950735, 6550332, 6550373, 6251143, 6550372, 6102002, 6130381, 6424070, 6550357, 6439571, 6251152, 6238060, 6237391, 6300659, 6550362, 5950771, 6557589, 6557551, 6102031, 6300653, 6251131, 6237388, 5956385, 6251134, 6273523, 6390545, 6251139, 6390539, 6424088, 6390522, 6439495, 6386250, 6590990, 5820904, 6342005, 6058256, 6342008, 6448657, 5820893, 6238069, 6342015, 6102041, 6341995, 6273550, 6300678, 6390556, 6466009, 5602742, 6150644, 6055369, 6183242, 6130393, 6150631, 6336719, 6336714, 6003188, 6448679, 6336724, 6557547, 6342014, 6557566, 6591007, 6102008, 6336713, 6386228, 6150629, 6273539, 5649873, 6150627, 6055291, 6055283, 6055368, 6055303, 6300668, 6273535, 6238073, 6183232, 6336731, 6345339, 6386254, 6386226, 6424055, 6440740, 6439565, 6439567, 6448683, 6591014, 6251157, 6273543, 6273546, 6130401, 5820928, 6336718, 6238049, 6336712, 6424047, 6439556, 6557579, 6448684, 5649864, 6003181, 6238064, 6550377, 6068625, 6102036, 5602738, 6550341, 6557571, 6237405, 6439474, 6179515, 6390561, 6130384, 6237403, 6550365, 6439566, 6043709, 6390551, 5599195, 6179531, 5843499, 6557555, 6550358, 6237401, 6237400, 6238050, 6251138, 6237399, 6251137, 6251135, 6251159, 6251133, 6300657, 6390550, 6386239, 6386244, 6390540, 6262836, 6439553, 6439561, 6439551, 6439485, 6390547, 5602743, 6424051, 6550355, 6466011, 6550356, 6448682, 6055285, 6300664, 5820894, 6251142, 6179521, 6251153, 6179509, 6424066, 6550351, 6439471, 6465995, 6557596, 5599184, 6130398, 5652812, 6179527, 6238063, 6237394, 6390543, 6390553, 6448685, 6336710, 6342001, 6058261, 6300670, 5580407, 6150641, 6273548, 6179513, 6251150, 6386236, 6325747, 6043702, 6341996, 6262821, 6130388, 6325748, 5950749, 6102024, 6251154, 6055379, 6036340, 6102013, 6424042, 6300667, 5602735, 6336708, 5950762, 6055325, 6055317, 6130390, 6237410, 6325746, 6386243, 6055309, 6424049, 6262842, 6183216, 5820910, 6300674, 6390535, 6557576, 6068640, 6130394, 6043715, 6036349, 6465992, 6439550, 6336703, 6465991, 6557558, 6557546, 5652794, 6102021, 6251141, 6300660, 6262835, 6342011, 6386246, 6390552, 6591018, 6273541, 6262824, 6325749, 6390536, 6273542, 6341999, 6342003, 6550360, 6550366, 6130395, 6424041, 6342000, 6591003, 6150630, 5652814, 6150642, 6150632, 6183219, 6262825, 6342007, 6390523, 6439575, 6055361, 6150645, 6557611, 5820892, 6055336, 6386235, 6466015, 5820877, 6341989, 6448672, 6300675, 6424035, 6068627, 6390546, 5652815, 6386253, 6300655, 6179535, 6300654, 6058262, 6300666, 6102015, 6251132, 6262827, 6424044, 6251147, 6390533, 6465998, 6386230, 6386229, 6439478, 6439497, 6466002, 6439481, 6439476, 6448680, 6557554, 6439573, 6130378, 6055302, 6390538, 6550370, 6750255, 6262838, 6390541, 6550381, 6439470, 6439473, 6273522, 6273524, 6336726, 6341992, 6341998, 6342009, 6342010, 6424059, 6466004, 6466012, 6466014, 6557577, 6424040, 6557572, 6055364, 6390527, 6055384, 6055383, 6055385, 6130380, 6130379, 6273534, 6439572, 6068619, 6179529, 6336715, 6055401, 6336722, 6179526, 6102007, 5956401, 6439577, 6390562, 6390518, 6102016, 6003190, 5950739, 6055333, 6238062, 6262847, 6300665, 6055312, 6251148, 6262823, 6150650, 6439558, 6273533, 6237412, 6068639, 6055281, 6336720, 6179517, 5424404, 4469371, 4502095, 4502102, 5319246, 5073095, 5180266, 5180226, 5328032, 4963689, 4897279, 5309839, 5404205, 4993529, 5483529, 5133973, 4993525, 4756785, 5309850, 5310829, 5354591, 4897264, 5507206, 5325265, 4888723, 5137159, 5328031, 4896837, 5228438, 5507191, 4861721, 5319256, 5507192, 5197048, 4947154, 5354603, 4735119, 5197041, 5073102, 5546940, 4897269, 5180227, 5319242, 4580540, 5310843, 5310236, 5133992, 4761993, 4950881, 5310237, 5310827, 5319229, 5137162, 5137139, 5479358, 4983285, 5240338, 4861730, 4947145, 4775451, 5507200, 4644602, 5041061, 5479356, 5208220, 4761984, 5328038, 5479357, 5325276, 4963672, 6439552, 6754861, 6466005, 5820903, 6819449, 6754859, 6386255, 6055299, 6819831, 6342017, 6550391, 5820925, 6424057, 6439469, 6251130, 6439488, 6386252, 6439570, 6796272, 6557549, 6819422, 5950729, 6802451, 6796261, 5599183, 6796250, 6813480, 6448663, 6754857, 6550359, 5599200, 6557609, 6055339, 6750230, 6055330, 6424080, 6262843, 6390544, 6590999, 6590993, 5602734, 6130397, 6557586, 6754883, 6591001, 6557559, 6750223, 6750224, 6750243, 6796259, 6750226, 6820228, 6820227, 6796279, 6262820, 6557604, 6183224, 6055378, 5950772, 6796266, 6557600, 6819826, 6853886, 6043710, 6853881, 6055365, 6058224, 6807135, 6424062, 6853885, 6750248, 6238045, 6750253, 6557588, 6424079, 6300671, 6466013, 6820219, 6448658, 6550367, 6550368, 6796253, 6796257, 6802452, 6448660, 6754867, 6853865, 6448673, 6820223, 6557599, 6557581, 6750238, 6819839, 6813481, 6055296, 6557561, 6439483, 6819801, 6796249, 6796275, 6465986, 6336723, 6807138, 6273529, 6750228, 6591015, 6424045, 6819400, 6819819, 6796262, 6819451, 6819828, 6853876, 6591019, 5913029, 6439574, 6466001, 5956383, 6557548, 6590992, 6336711, 6819438, 6055316, 6754866, 5820922, 6754865, 6590987, 6802457, 6802450, 6819399, 6819423, 6853869, 6342006, 6754884, 5602740, 6750220, 6754869, 6853862, 6796260, 6550354, 6068628, 6386240, 6591011, 6750249, 6813482, 6750257, 6750258, 6424037, 6390526, 6813472, 6591017, 6424068, 6557605, 6300651, 6796252, 6448659, 6336709, 6796251, 6183227, 6300661, 6807134, 6055292, 6853880, 6550344, 6796270, 6819431, 6550363, 6102027, 6465996, 6439568, 6058228, 6130399, 6130383, 6238068, 6179528, 6819450, 6550334, 6796269, 6853879, 6591016, 6550375, 6853868, 6754879, 6550369, 6750221, 6819418, 5913027, 6819790, 6251146, 5820912, 6590994, 6819412, 6179536, 6273551, 6439484, 6590988, 6819402, 6390519, 6102037, 6058270, 6448670, 6590995, 6754885, 6550347, 6750242, 6754880, 6754886, 6819440, 6819404, 6819786, 6796274, 6424078, 6796268, 6853860, 6819792, 6754877, 6819834, 6819812, 6550350, 6819409, 6557568, 6750251, 6750250, 6550339, 6750236, 6819424, 6557556, 6150628, 5599193, 6802465, 6439494, 6796242, 6807137, 6819428, 6179512, 6796265, 6813483, 6058260, 6819441, 6557594, 6754856, 6262834, 6183240, 6813476, 5652810, 5602747, 6424065, 6055297, 6055371, 6386247, 6043713, 6754878, 6325740, 6439569, 6251156, 6819407, 6819415, 6853888, 6150636, 6819403, 6819820, 6179533, 6300673, 6819830, 6390529, 6262829, 6754871, 6796258, 6058265, 6796245, 6424067, 6183218, 6557565, 6424075, 6750229, 6754858, 6448676, 6557545, 6424048, 6424063, 6273527, 6590997, 6550336, 6819405, 6819442, 6853864, 6055305, 6796271, 6466010, 6754873, 6819434, 6819425, 6754864, 6465999, 6557580, 6802466, 6424086, 6448674, 6386248, 6262837, 6557564, 6819421, 6424081, 6439559, 6819414, 6820217, 6557557, 6754855, 5649874, 6345342, 6819810, 6273547, 6819802, 6055389, 6055340, 6750247, 6550353, 6590989, 6754870, 6424034, 6819785, 6819813, 6819808, 5956400, 6819799, 6238072, 5950743, 6853873, 6819401, 6819416, 6300679, 6003192, 5950766, 6058248, 6439563, 6557560, 6183234, 6819798, 6557602, 6130402, 6439472, 6820212, 6345341, 6550337, 6439477, 6068632, 6557593, 6068618, 6130373, 6424043, 6237409, 6390525, 6336704, 6439555, 6336725, 6819832, 6439549, 6179511, 6853861, 6557595, 6055402, 6424072, 6390558, 6325742, 5652806, 6036347, 6448681, 6754854, 6557550, 6150625, 6557552, 5820897, 6591005, 6754882, 5580402, 5820909, 6922254, 7004111, 7013153, 6999855, 6999823, 6999879, 6999851, 7035175, 7004115, 6999867, 6999868, 6999850, 7013135, 7013136, 7013144, 6999843, 6999852, 7013100, 7013097, 7051981, 7051982, 6999838, 7013150, 6999862, 7035174, 6999827, 7051957, 6999837, 7051951, 7051983, 7004445, 7013124, 7050799, 6922252, 6999866, 7035173, 6922247, 6922242, 7013108, 7013141, 7051961, 6999832, 6999828, 7013099, 6922257, 6999863, 7051954, 6922234, 7004114, 7013151, 7051978, 7051977, 7051968, 6999841, 7013116, 7004117, 7013201, 6922259, 6999822, 6999839, 7013123, 7050794, 7035184, 7013110, 6922239, 6999834, 6922258, 6922240, 6999835, 7013117, 7035177, 7051984, 7050788, 6922230, 7004110, 7013132, 7013133, 6999870, 6999873, 6999876, 7013094, 7013125, 7013140, 7013145, 7013196, 6922244, 5424405, 5424395, 5424403, 5424389, 5328024, 5433392, 5404218, 5133985, 5325273, 5354613, 5507202, 5479355, 5404227, 5339820, 5479364, 5319257, 5310833, 5197040, 5180243, 5310211, 5319254, 5310838, 5328011, 5310219, 5208215, 5404209, 5197025, 5208217, 5137142, 4993510, 5470162, 4888719, 5404207, 4950872, 5133997, 5041048, 5470171, 5479349, 5483535, 5325278, 5073099, 5319232, 5652817, 5956396, 6750235, 6819809, 6448675, 5602765, 6130392, 6424073, 5649883, 6557578, 6055386, 6102011, 6853874, 6300669, 5956395, 6439490, 6819796, 6853871, 6058241, 5649872, 6183223, 6055382, 6819818, 6102018, 6386234, 5843508, 6550352, 6557597, 6466006, 6424069, 6262845, 6819800, 6853884, 6819787, 6102019, 6058264, 6750237, 6390520, 6055381, 6813474, 6058227, 6813479, 6819795, 6853859, 6273528, 6439493, 6055279, 6853863, 5602759, 6055324, 6796254, 6448665, 5652798, 6750225, 6390555, 6262818, 6237407, 6390528, 6237411, 5649868, 6273537, 6273538, 6273526, 6390530, 6796264, 6183231, 6819793, 6055276, 6819842, 6819410, 5602750, 6796248, 6853867, 6819797, 6853889, 6819439, 6183222, 6813473, 5652811, 5950785, 6754860, 6754876, 5649876, 6550384, 6055363, 5950746, 6750245, 5950757, 6325745, 6550333, 6557585, 6820213, 6591006, 6819822, 6183215, 6819840, 6819788, 6557598, 6336716, 6819419, 6853854, 6819836, 6448668, 6342012, 6557569, 6238061, 6802453, 6796247, 6424031, 6424082, 6055277, 6439486, 6068638, 6796256, 6819435, 6819791, 6802455, 6819829, 6590991, 6591010, 6802454, 6819448, 6424058, 6439489, 6336717, 6058250, 6557562, 6750256, 6386251, 6102029, 6424060, 6802461, 6550335, 5652802, 6550388, 6251158, 6802458, 6819794, 6820226, 6796241, 6853887, 6550371, 6819825, 6853882, 6820222, 6237390, 6820220, 6820215, 6820216, 6802468, 6796277, 5950730, 6448677, 6150643, 6819841, 6591009, 5950753, 6448666, 6439564, 6754887, 6448662, 6043719, 6590986, 6819814, 6590985, 6819784, 6390560, 7051971, 7051975, 6999824, 7051974, 7050792, 6999864, 7013096, 6922223, 7004120, 7051980, 7051956, 7051985, 7035188, 7051953, 7051967, 7013139, 7035178, 7050797, 6999842, 7004118, 6922249, 7051952, 7004129, 6999836, 7013146, 7051959, 7004128, 7050796, 6999833, 7013152, 6999826, 7051970, 7035187, 7013199, 7051955, 7004121, 7050800, 7013147, 6922231, 7013107, 7013131, 7013128, 6999825, 6922235, 7013148, 7051972, 7051966, 7004112, 7050802, 6999847, 7035183, 7004458, 7013129, 6922227, 6922255, 7013138, 7004122, 7013105, 7050801, 6999853, 7004456, 7004455, 7013104, 7013102, 6922246, 6922224, 7004446, 7013106, 7004450, 7050791, 7035182, 5470169, 6238057, 6055278, 6550382, 6819411, 6439557, 6102025, 6055293, 6550386, 6550385, 6591008, 6068637, 6796276, 6754863, 6237406, 6262846, 6055399, 6341997, 6819445, 6055288, 6819432, 6183214, 6819429, 6557607, 6550340, 6465989, 6750259, 6819408, 6819398, 6424071, 6237408, 6251140, 6796278, 6150633, 6813475, 6439480, 6424076, 6183221, 5820921, 5950780, 6439576, 6819803, 6251149, 6068634, 6550389, 6262831, 6754888, 6557591, 6341993, 6999872, 6999831, 6999871, 6999830, 7051962, 7013115, 6999857, 7004123, 6922245, 6999878, 6922232, 6999848, 7035181, 6922221, 7004457, 7004454, 6999829, 7013195, 7004444, 7050793, 6922236, 7004453, 7013111, 6058263, 6754881, 6819417, 7051965, 6999861, 7013121;
b) Item Number 314-13-03, UDI/DI 10885862172686, Serial Numbers: 2330938, 2330933, 2333431, 2330931, 2452157, 2333425, 2330924, 2330937, 2330922, 2452146, 2474418, 2474406, 2333441, 2333417, 2406568, 2406579, 2452145, 2330934, 2449053, 2460052, 2460048, 2460050, 2522143, 2452159, 2406577, 2333439, 2333436, 2474420, 2452173, 2333419, 2330941, 2460047, 2406571, 2333424, 2330935, 2330939, 2452167, 2204809, 2204799, 2333421, 2333435, 2406585, 2406583, 2406566, 2474415, 2333432, 2406576, 2452154, 2460051, 2474396, 2330923, 2330927, 2406580, 2452175, 2452166, 2204794, 2204811, 2204801, 2204805, 2204808, 2204795, 2204792, 2204797, 2204807, 2204791, 2204803, 2204793, 2204800, 2330940, 2330932, 2333423, 2406586, 2406582, 2406573, 2449056, 2452152, 2452169, 2452150, 2499297, 2406569, 2406587, 2479355, 2204810, 2333427, 2330930, 2330929, 2449054, 2333434, 2460054, 2330936, 2452147, 2406581, 2499284, 2333420, 2460046, 2333430.2333428, 2656410, 2499303, 2499293, 2499307, 2522123, 2542387, 2333415, 2622915, 2542399, 2622922, 2522138, 2522124, 2522127, 2622888, 2622891, 2330925, 2333416, 2622919, 2656405, 2665799, 2474402, 2656396, 2622904, 2542398, 2542390, 2474393, 2608566, 2608563, 2656404, 2656342, 2656411, 2665798, 2522140, 2452149, 2406589, 2452160, 2656352, 2608569, 2656334, 2622913, 2665802, 2656387, 2656398, 2499289, 2499308, 2452174, 2622889, 2474388, 2460049, 2452156, 2622897, 2622907, 2542396, 2665809, 2665810, 2449055, 2452151, 2452170, 2406578, 2452172, 2622923, 2474387, 2452164, 2406572, 2474385, 2522137, 2608561, 2499285, 2522135, 2474403, 2542391, 2622895, 2622901, 2656406, 2608575, 2522141, 2474401, 2608567, 2622905, 2622890, 2406575, 2522136, 2656355, 2542400, 2622903, 2474389, 2656354, 2656348, 2622911, 2474394, 2474413, 2474400, 2522119, 2656395, 2522120, 2656353, 2452161, 2522132, 2333437, 2499296, 2656401, 2608562, 2622908, 2406570, 2452158, 2608573, 2542395, 2608564, 2406584, 2474395, 2622899, 2656421, 2474416, 2656428, 2474422, 2479365, 2474414, 2608558, 2608565, 2333426, 2474407, 2656393, 2622893, 2542389, 2622924, 2479366, 2622886, 2656391, 2622894, 2656417, 2406567, 2622916, 2449057, 2542397, 2522139, 2665808, 2474405, 2542394, 2452163, 2656422, 2499295, 2656412, 2656346, 2665805, 2656400, 2608574, 2474392, 2608576, 2499288, 2499304, 2499286, 2479357, 2479361, 2542392, 2474391, 2608568, 2608570, 2622896, 2656350, 2622885, 2656357, 2656389, 2656394, 2474421, 2608571, 2333440, 2333429, 2474408, 2656390, 2656427, 2452165, 2474390, 2522115, 2479363, 2522130, 2522145, 2479352, 2608559, 2608578, 2479364, 2622921, 2522129, 2499283, 2499298, 2460053, 2499291, 2522121, 2499290, 2452162, 2474384, 2452171, 2333422, 2333418, 2452153, 2608557, 2479359, 2656407, 2474386, 2474409, 2608560, 2622912, 2622914, 2622920, 2522133, 2522134, 2689046, 2777376, 2777374, 2816759, 2770391, 2689051, 2812021, 2777380, 2770390, 2777358, 2770426, 2770388, 2812017, 2689058, 2770413, 2700835, 2777370, 2777363, 2777379, 2777372, 2816752, 2812024, 2700837, 2689053, 2770396, 2777388, 2777401, 2700839, 2812007, 2777387, 2812003, 2812012, 2770399, 2770378, 2689049, 2770397, 2777367, 2812014, 2700840, 2816773, 2689037, 2770386, 2812016, 2700825, 2689043, 2689052, 2816757, 2770392, 2770418, 2811998, 2770403, 2689041, 2770385, 2700832, 2689060, 2777364, 2770425, 2777355, 2700833, 2770412, 2770404, 2770424, 2812005, 2777413, 2812018, 2816762, 2689050, 2700828, 2770423, 2770383, 2700830, 2689054, 2770389, 2770380, 2770419, 2770422, 2777346, 2770416, 2700827, 2812026, 2689062, 2777354, 2777356, 2777390, 2893285, 2893289, 2777348, 2474412, 2452148, 2333438, 2665807, 2474383, 2522142, 2622900, 2204796, 2333433, 2522144, 2656420, 2656403, 2656386, 2542401, 2622906, 2656388, 2656384, 2656338, 2522116, 2522122, 2522117, 2479353, 2499302, 2656335, 2665806, 2499294, 2522125, 2656356, 2499299, 2522118, 2474411, 2656339, 2656424, 2656419, 2665803, 2622892, 2656358, 2499287, 2656337, 2479360, 2479362, 2474399, 2656426, 2499300, 2656415, 2656340, 2665813, 2479356, 2479354, 2665795, 2522114, 2622918, 2665812, 2665800, 2474410, 2656392, 2452155, 2542393, 2622925, 2499292, 2542388, 2656416, 2893277, 3593936, 3622886, 2893298, 3622896, 2816778, 2777409, 2893280, 2952033, 3593895, 2777366, 2951998, 2812000, 2951978, 3577298, 3587565, 3587563, 3668970, 3674047, 3674058, 3593939, 3593887, 3674054, 2777373, 2951982, 2689059, 2777407, 2777377, 2770420, 3622877, 3668984, 2951981, 2969452, 2812025, 3577285, 2816772, 3593940, 3587575, 2951996, 3622869, 3622894, 3668948, 3587559, 3622879, 3622897, 2969453, 3577309, 3593956, 3593955, 3622843, 2969440, 3622821, 3593890, 2969430, 3577299, 2969429, 2812001, 2893290, 2893282, 2893287, 2969463, 3577305, 3622833, 3693564, 3622836, 3673827, 3750101, 3673833, 2770405, 2816763, 2816756, 2952032, 3577294, 2952054, 2777345, 3587564, 3610023, 3577290, 3577302, 3674048, 2893293, 2969427, 2893276, 3577279, 3610015, 2969462, 2969431, 2952050, 2816775, 3564184, 3610029, 2952059, 3622854, 3622866, 3673848, 3750118, 2951999, 2969461, 2777406, 2777347, 2816768, 2952047, 2770409, 3577301, 3622849, 3577310, 3674044, 3693588, 3668980, 2812022, 2952045, 3622812, 3622814, 3587566, 2700842, 2969422, 2777371, 2952038, 3587571, 3622888, 3668982, 2812011, 2777375, 2777368, 2969460, 3564183, 3564181, 3564175, 3593948, 3622837, 3622845, 2689042, 2812020, 2816754, 3622858, 2777389, 2770421, 2816758, 2969437, 2969423, 2893291, 3693554, 2812004, 3693565, 3577307, 3674045, 3673843, 3673838, 2777360, 3587574, 2816777, 2952043, 3622860, 2816771, 3587580, 2777412, 2951977, 3577282, 3587573, 3587569, 2952057, 3622883, 2893281, 2952061, 3622829, 3577281, 3577293, 3622839, 2700829, 2689063, 2777400, 2969442, 3610024, 2689055, 2770408, 3593933, 2816774, 2700834, 2777411, 2770393, 2951980, 2969465, 2969467, 3587567, 3622828, 2951988, 3668968, 3587570, 3674067, 3622863, 3593894, 2952058, 3622832, 2777381, 2969459, 3564169, 3693552, 2816764, 3610019, 3750121, 2777357, 3577284, 3673850, 3674055, 3693569, 2770394, 2952035, 3577303, 3593950, 3610022, 2893292, 3622840, 2893270, 2893296, 3577280, 3593943, 2893297, 3587572, 3622835, 2770415, 3622815, 3693547, 3668963, 3693559, 3668960, 3593944, 2777405, 3668971, 3693581, 3577300, 2811999, 2969436, 3622868, 3750098, 2969425, 2770382, 2777352, 2816765, 2816767, 2893284, 3593938, 3593949, 2770407, 2969435, 3750115, 2816760, 2893288, 2951983, 3564167, 2770398, 3564176, 2951993, 2893283, 3674062, 2777391, 2812009, 3577283, 2952041, 2969445, 2969450, 3593942, 2951991, 2689061, 3593892, 3593891, 3610016, 3564173, 2812008, 2951992, 2952048, 2952040, 2777398, 2816780, 3610018, 2777393, 3564168, 2777384, 3622885, 3587561, 2816755, 2893272, 2816781, 3564171, 3593953, 3674056, 3622873, 3693583, 2816761, 2770414, 2969451, 2777404, 2893271, 2952037, 3564182, 2816769, 2816779, 3587557, 2969455, 2951995, 3668954, 3622891, 3577276, 3564172, 3668939, 3693563, 2689044, 3673832, 2777396, 3610030, 3668957, 3750114, 3622892, 2969433, 3622881, 2893294, 2689048, 2700831, 2770411, 2893273, 2952053, 3587568, 3622841, 3668976, 3668986, 2816770, 2689057, 3622819, 2951989, 3622880, 2969426, 3622811, 3577277, 3668952, 2770401, 2893279, 2969454, 2952049, 2952046, 2952034, 2952001, 2951975, 3577288, 3577304, 3577289, 3668938, 3610006, 3668959, 3673842, 2893286, 2893295, 3577306, 2952036, 3622875, 2951997, 2969464, 2770395, 2777365, 2812015, 2770417, 3622893, 2770379, 2969428, 3577287, 3564179, 3622878, 3622887, 3622890, 3668944, 3693551, 3674066, 3693549, 3668973, 3610007, 3610010, 3674057, 3622872, 3668969, 3668975, 3622899, 3693573, 3693550, 3622853, 2777394, 2969424, 2969439, 3564177, 2656336, 2656344, 2499301, 2665797, 2656385, 2656397, 2474382, 2474417, 2656413, 2656341, 2608577, 2656408, 2406588, 2406574, 2665804, 2656409, 2204804, 2330928, 4153705, 4096974, 3668950, 3593952, 3668964, 3587558, 3750132, 4142011, 3898062, 3750135, 3668985, 2969441, 4005370, 2770410, 3587579, 4049222, 3750105, 3622842, 3750131, 3593889, 4005368, 3917602, 4049230, 4153693, 3593884, 3835191, 3917593, 3835206, 2812002, 3750095, 2811997, 3674049, 3668987, 4063682, 4063683, 4091580, 4066125, 3898082, 4020753, 4122672, 4091592, 3750108, 3797260, 4063686, 4164647, 3668967, 3587562, 3835178, 4005361, 4091581, 3587581, 4032052, 3797266, 3668945, 3622834, 3917610, 3917594, 4049225, 4059228, 3693548, 3835168, 2700823, 3835224, 3693576, 3835205, 3835192, 3898071, 3898067, 3917615, 4020772, 4091598, 4091586, 4063689, 4108487, 3693568, 3835195, 4020751, 3835225, 4049233, 4108467, 4108468, 4142022, 4142013, 4185978, 3593954, 3835207, 3835238, 3835155, 3750092, 3750102, 3898075, 3898070, 3917599, 4063677, 4091588, 4091584, 2951973, 3593893, 3668953, 2969438, 3622818, 4122490, 4091593, 3668978, 3673836, 3898063, 3835208, 4164649, 3898098, 3835189, 3898064, 3835186, 4020760, 4032038, 4091578, 3835209, 4005371, 4153703, 3835173, 3674064, 2969449, 2969448, 3835169, 3898088, 3673841, 3917607, 3673835, 3985479, 4150363, 4005367, 3797235, 3917600, 3797265, 3898105, 3610020, 3985491, 3917596, 4049232, 4108481, 3835248, 3668943, 4032044, 4091577, 3917619, 4122665, 3898083, 2952039, 2969432, 3898059, 4049219, 4091585, 3797230, 3668983, 4150356, 4020763, 3673849, 3693577, 3797221, 3797219, 3835204, 3673839, 3835198, 3835180, 3917617, 3835162, 4005362, 3835185, 4049217, 2969458, 4032056, 3917635, 4005364, 4059226, 3564174, 2952052, 3622813, 3898073, 4078829, 4078831, 3673845, 4108484, 4150381, 3985480, 2777410, 3898103, 4128039, 3674046, 4142023, 2777385, 3668946, 2951994, 3622848, 3668941, 3622856, 3668977, 4049223, 4032047, 2951990, 2951984, 3985478, 2777408, 3668956, 4063681, 3693560, 3985486, 3797232, 3587578, 3917612, 3622820, 3622831, 3835161, 3622855, 3622861, 3917633, 3898084, 3898060, 3985477, 3673837, 4122656, 3797229, 2777397, 3835202, 3835219, 4103632, 3985484, 4099806, 3564178, 3593896, 3898086, 3797247, 3668940, 3898092, 3622838, 3898097, 3835213, 4049236, 4049220, 3668972, 3797259, 3835200, 3750107, 4032053, 3622870, 3693590, 3593888, 3835211, 3593941, 3917621, 4066126, 4063679, 4091832, 4049234, 2770402, 3668965, 3593886, 3835181, 3622884, 3835232, 3835223, 4049237, 3898068, 2777399, 3835175, 3673828, 2812023, 4078839, 3610028, 3898074, 3835215, 4005365, 4005376, 3577291, 2969443, 3668942, 3797255, 3622822, 3898087, 3674052, 3668947, 3835246, 4054486, 3610011, 4020767, 3693553, 3622898, 2952056, 2777402, 2969444, 3750125, 3835174, 3835166, 4020750, 3693571, 4059229, 4078836, 4078838, 3622871, 2689039, 3564185, 3693567, 3835199, 2812006, 3797254, 4122659, 3917603, 3835237, 4078837, 3750106, 3835216, 4049221, 3622847, 3835188, 3622865, 3898066, 3835156, 3622810, 4063688, 3985485, 3610008, 3898080, 3622876, 3610017, 4128029, 2952055, 3622867, 4020762, 3797263, 3593945, 4128038, 4150376, 4150371, 4193245, 3797225, 3797233, 3593947, 3693585, 3835240, 3797240, 4153710, 3577297, 3577308, 4054490, 4128037, 3622826, 3797231, 3898091, 3835170, 3593957, 4020769, 3835231, 3593934, 3917627, 3593937, 4063685, 4108471, 3797267, 4150358, 4150373, 4142024, 3674065, 4020771, 3622824, 4185982, 3835234, 4032042, 2689056, 3610013, 4108479, 2689040, 3593885, 3797237, 3797243, 3593935, 3917632, 2893269, 4153698, 3750138, 3835201, 3622823, 4063676, 4150372, 3835249, 3750096, 3797252, 3835151, 3835160, 4023792, 3835227, 2952042, 2952060, 2952051, 4078835, 4032041, 3917614, 3622864, 3622851, 3985482, 3985487, 4049229, 3917601, 4049235, 3750112, 3917641, 3693578, 3917634, 3835163, 3750110, 3673830, 3835220, 3898061, 4091833, 3797250, 4005363, 4091591, 4150375, 3917622, 4059227, 3564170, 3835179, 4092806, 4128030, 4122671, 3622895, 3898065, 4099805, 3693587, 3797253, 3864551, 3864552, 3925140, 3835152, 3797244, 3835233, 3693566, 3750111, 2816776, 3750134, 3835230, 4122657, 3750109, 3693584, 3693586, 3864548, 3622859, 3864549, 3864547, 3864546, 3925137, 3925136, 3797228, 3864550, 3925139, 3925138, 2893275, 4171123, 4171126, 4171118, 4171115, 3797261, 3622817, 3797242, 3917598, 3750103, 3750124, 4122669, 3917611, 3673840, 3797227, 3622830, 3917620, 4122662, 3917618, 3835157, 3674059, 2969466, 2816753, 3668966, 2812013, 3674050, 3622857, 3835228, 3835217, 3835190, 3835197, 3898085, 3898081, 3898069, 3917626, 3898072, 3917638, 3917636, 3985490, 3898096, 4005366, 4020764, 4063680, 3622852, 4108489, 3693562, 4005373, 3750130, 4096979, 2777362, 3835235, 2770384, 4099803, 4122676, 3593946, 4122661, 3693561, 3587577, 3668955, 3797249, 3750104, 3835153, 3835167, 3835193, 3835221, 3898093, 3898079, 2952044, 3673847, 4020770, 3564180, 4096982, 4096977, 4122653, 4122655, 3898100, 3577292, 3898076, 2777383, 2777369, 3835159, 2777414, 3622827, 3797239, 3917624, 4108480, 3835243, 3668961, 3587576, 3835214, 4091834, 2689038, 3797238, 3917629, 3898090, 3917631, 3797264, 3797268, 4049224, 4020768, 3668951, 3797262, 3622862, 3750122, 3835158, 2777386, 2816782, 3835218, 3693558, 3693543, 3693556, 3750116, 3693557, 3797248, 3797258, 3835242, 4171117, 4171120, 3750139, 3750120, 3750133, 3750137, 3750113, 3797226, 3797245, 3898095, 3898099, 4020754, 2893278, 2770427, 3693574, 4018033, 3673834, 3750117, 3750119, 3668974, 3622846, 3674063, 3917597, 3835165, 3673844, 4108475, 3693555, 3674053, 3797257, 3693546, 3610027, 3610012, 4108470, 2656402, 2522126, 2656351, 2622910, 2622898, 2656343, 2656349, 2452168, 2499305, 2622917, 2330926, 4142012, 3622874, 3917605, 4265508, 3750140, 4449397, 4507244, 4122677, 4280825, 4280816, 4449227, 4265520, 4032043, 4216307, 4032051, 4193229, 4265511, 4449264, 4232402, 2951987, 4142025, 4449368, 3835244, 4091582, 4182747, 4216308, 4265486, 4265490, 4449384, 4505482, 4449383, 4507276, 3835210, 4280806, 4122670, 4449355, 4523876, 4239015, 4171124, 4398490, 4449232, 4528053, 4523877, 4216300, 4232400, 4449354, 4193222, 3750094, 4265497, 4232387, 4239016, 4280815, 4376544, 4376532, 4142007, 4193232, 4250697, 4265498, 4298124, 4398482, 4528080, 4507258, 4528077, 4265516, 4449365, 4449382, 4265519, 4449261, 4528055, 4142008, 4153688, 4256902, 4322356, 4376535, 4398478, 4265518, 4507275, 4142031, 4322364, 4505491, 4280836, 4507284, 4507262, 4528060, 4216317, 4142010, 4208641, 4216312, 4256911, 4239002, 4398473, 4449386, 4398468, 4398485, 4505503, 4049228, 4322339, 4528083, 4190758, 4128033, 4505459, 4507248, 3693580, 4091835, 4078834, 2969457, 4376508, 4507280, 4185989, 4224302, 4376509, 4216302, 4507257, 4032039, 3674051, 3673851, 4505476, 4265496, 4049231, 4232399, 4232386, 4256882, 4041857, 4232388, 4322353, 4142014, 4322329, 4041858, 4153692, 4142026, 4398470, 4449233, 4153699, 4159911, 4208640, 3835176, 4265501, 4232393, 4507290, 4376527, 4122674, 4449269, 4216294, 4322355, 4185993, 4182749, 4449234, 4505504, 4398467, 4528048, 4398461, 4505460, 4449390, 4096983, 4142015, 4322348, 4449385, 4507289, 4528071, 4185980, 4032049, 4091597, 4193244, 4265484, 4150365, 4185986, 4449372, 4528059, 4449376, 4063692, 4507282, 4063691, 4142016, 4398488, 2969446, 4153694, 4232389, 4142021, 4265482, 4322315, 4449236, 4449248, 4398483, 4507260, 4507277, 3750136, 3750129, 4265485, 4376549, 4376546, 4376503, 4505507, 4232394, 4322321, 4142029, 3577295, 4322327, 4216297, 4398472, 4449222, 3797251, 4265524, 4153695, 4449369, 4376529, 4280823, 3750123, 4091589, 4128036, 4108476, 4528067, 4078830, 4150360, 4239014, 4265480, 4193224, 4054489, 4128023, 4256895, 4322351, 4398460, 3917608, 3985489, 4398491, 4398455, 4449360, 4505474, 4376502, 3985483, 2893274, 4193241, 4449224, 4298121, 4280803, 4376538, 4280807, 4376534, 4239000, 4256892, 4265507, 4298108, 4280829, 4280813, 4322365, 4398479, 4449357, 4505485, 4507247, 4142030, 4398487, 4142036, 4376522, 4449228, 4449262, 4216290, 4150367, 4103631, 3985488, 4505473, 4091590, 4108472, 4238998, 4032048, 3898077, 4239010, 4322313, 3898078, 4182744, 4193236, 4208642, 4224300, 4449387, 4265517, 4063690, 4150380, 4159913, 4216292, 4150362, 4280837, 4322360, 4265526, 4216301, 3898094, 4142018, 4216298, 4091583, 4376518, 4298116, 4265525, 2777351, 4298110, 2689047, 4122663, 3835250, 4528045, 4449400, 4122675, 4096976, 4232379, 4208646, 4265479, 4265483, 4208639, 4256887, 4185988, 3835241, 4153696, 4193226, 4193240, 3917595, 4193234, 4216315, 4265527, 4224301, 4280808, 3917628, 3835212, 4449238, 4005372, 4108491, 4091596, 4142034, 4256909, 4280834, 4265481, 4232380, 2969456, 4376550, 4322358, 4449221, 4449358, 4505478, 4528042, 4238993, 4505489, 4505465, 4507293, 4078832, 4398489, 4032040, 4159910, 4122660, 4322363, 4376497, 4449244, 4298113, 4182745, 4280830, 4193243, 4185977, 4150361, 4322345, 4376547, 4216289, 4507269, 4280818, 4398458, 4507245, 4528062, 4528072, 4059225, 4153702, 4193227, 4265494, 2812019, 4528037, 4376501, 4265505, 4298119, 4376528, 4216316, 3835177, 4238995, 4376524, 4224306, 4505481, 4449257, 4150379, 4153706, 4128028, 4185979, 4505467, 4449260, 4256885, 4193246, 4322368, 4505464, 4224303, 4108469, 4190756, 4063678, 4239007, 4505468, 4523869, 4193230, 4164646, 3668981, 4193237, 3622882, 4193239, 2969434, 4256899, 4193231, 4280811, 4298129, 4216288, 4280833, 4376498, 4398481, 4376554, 4142019, 4523873, 4505505, 4449374, 4523872, 4505496, 4505502, 4193223, 4091579, 3835171, 3835203, 4128034, 4153687, 4523855, 3693570, 4032050, 4322322, 4449395, 4322317, 4505462, 4153708, 3917609, 3622825, 4507288, 4063687, 2770381, 4150366, 4185994, 4185984, 4041855, 4171122, 4185990, 4182742, 4128027, 2689064, 3835154, 4216319, 4256907, 4265510, 3693544, 4153686, 4505494, 4449353, 4265492, 4528075, 4528070, 4153691, 4239013, 4505495, 4507268, 4449250, 4449359, 4216310, 4171125, 4280809, 4449364, 4256908, 4239001, 4449377, 4108477, 4449394, 4449363, 4449253, 3985476, 4239004, 4128031, 4216291, 4239003, 4239012, 4216306, 4216303, 4280831, 4265523, 4096978, 4122654, 4005375, 4507283, 4507291, 4108464, 4185985, 4041856, 4322361, 4020761, 4398459, 4376552, 4523853, 4322324, 4322328, 4449373, 4265493, 4280810, 4298127, 2951976, 4108486, 4193225, 3577286, 4280822, 4091594, 4078833, 4150364, 4216305, 4224308, 4232401, 4322362, 4322318, 4322341, 4376507, 4449391, 4507256, 4256891, 4041851, 4256900, 4449392, 4376519, 4398465, 4322359, 3674060, 3593882, 4232396, 4032045, 4239009, 2951985, 4232391, 4265513, 4238994, 4280820, 4265512, 3673831, 4122664, 4096973, 3898089, 4128042, 3750093, 4150370, 4280835, 4232374, 4507272, 4216313, 4250695, 4298115, 4108473, 4122658, 4049226, 3587560, 4041859, 4122489, 4265502, 4193228, 4128041, 4322352, 4185981, 3835229, 2700841, 4398475, 4122668, 4216318, 4182750, 4376504, 4280812, 4376499, 3917625, 4208647, 3835187, 4164645, 3835183, 4096975, 4449378, 4265487, 4232376, 4232381, 4256889, 4298125, 4298126, 4376537, 4449351, 4449256, 4376530, 4298111, 4298112, 4298128, 4298117, 4298118, 4232382, 4322320, 4376521, 4449393, 4449380, 4182736, 3917606, 4265509, 4108474, 4208644, 4376545, 4108478, 4182740, 4041853, 4232392, 4298123, 4232375, 3835196, 4265488, 4153707, 4376516, 4376541, 4182737, 4232373, 4232384, 4256903, 4018031, 4449235, 4018032, 3917592, 4020759, 4256886, 3917616, 4020765, 3835164, 4256884, 4232383, 4298122, 4208645, 4507274, 4612896, 4657296, 4612898, 4634097, 4634093, 4651607, 4612900, 4612889, 4634168, 4612899, 4634102, 4634174, 4651585, 4651588, 4651594, 4657311, 4634083, 4634092, 2522128, 3668962, 3797222, 3622889, 4398480, 4376496, 3668979, 4376543, 3577278, 4528074, 4280821, 4298109, 3917613, 4041854, 4224304, 4150368, 4142027, 4150377, 4099802, 4449356, 4322326, 4505470, 3610014, 4265504, 4096980, 4505471, 4193233, 4322334, 4523863, 4280827, 3835184, 4505469, 4507261, 4398469, 4449237, 4507278, 4128040, 3668958, 3693575, 4298114, 4398452, 4142032, 3593883, 4153690, 4398466, 4185987, 3797220, 4505490, 4216314, 4376506, 4528057, 4528068, 3835182, 3610026, 4238997, 4528047, 4193242, 4142017, 4449231, 4449367, 4376523, 4171116, 4376514, 4528044, 4528069, 4280826, 4063684, 4528046, 4523875, 4190757, 4523860, 4449263, 3898104, 4376548, 4049218, 4265489, 3835251, 4449255, 4398476, 4322323, 4507251, 2816766, 2969447, 4505506, 4507252, 4507263, 4523854, 4239005, 4507292, 3673829, 4505487, 4528079, 4528051, 3750097, 4376511, 4449243, 4224305, 4322338, 4449366, 4032055, 4528061, 4232377, 4507250, 4505466, 4256888, 4239011, 3917630, 4322354, 4528041, 4449226, 4523866, 4449258, 4322367, 4449266, 4505497, 4054488, 4322366, 3673846, 4449225, 4091587, 4449245, 4528050, 3577296, 4505477, 4523856, 4322333, 4150369, 4322331, 4507265, 4298120, 3750091, 4507271, 4505486, 2770400, 3668949, 3610021, 4505483, 4280804, 3622816, 4376513, 4376536, 3622844, 4322350, 4322337, 4523859, 4171119, 4280819, 4376520, 4232372, 4208643, 4505475, 4505500, 4185992, 4096981, 4153700, 4265522, 4505463, 4528066, 4128032, 4528076, 4505480, 4528082, 4449246, 4507285, 4153709, 4128025, 4376531, 4256896, 4256883, 4256897, 4020752, 2777349, 4322332, 4322314, 4224309, 4449381, 4528058, 4528043, 3835247, 4528081, 3797234, 2951972, 4265478, 4505461, 4523864, 4185983, 4322319, 4153697, 4376510, 4265521, 4523865, 4216295, 4523862, 4216296, 4239008, 4398462, 4150353, 4216320, 4376525, 4280805, 4280814, 4505498, 3985475, 4108465, 4091595, 3917623, 4020757, 4449249, 3898102, 4376539, 4280832, 4239006, 4449259, 4528073, 4507255, 4376515, 4507253, 4505479, 4507267, 4507259, 4523857, 4449388, 2951974, 4232398, 4182748, 4449370, 4182743, 3593951, 4159914, 4398463, 3917637, 4449223, 3797224, 4005369, 4449239, 2777382, 4238996, 4069140, 3917604, 4265514, 4376517, 4528039, 4208638, 4507286, 4256904, 3750100, 4250696, 4376500, 4528064, 4232395, 4505472, 4883505, 4787020, 4634100, 4936439, 4869246, 4634063, 4811737, 4737831, 4756930, 4787009, 4634103, 4787030, 4657289, 4657288, 4634067, 4825294, 4888664, 4657302, 4936420, 4634113, 4634057, 4612902, 4956099, 4795290, 4811739, 4634107, 4969577, 4811733, 4634159, 4634129, 4668996, 4925834, 5069814, 4888660, 5069827, 4634156, 5095353, 4634139, 4897941, 4634121, 4883510, 4634073, 4925851, 4825298, 5095351, 4612891, 5119621, 4634146, 4768566, 4651610, 4925859, 4634109, 4634142, 4756935, 4787032, 4888661, 5031152, 5119613, 4754983, 4651603, 4888676, 4825290, 4956100, 4634061, 4811738, 4925845, 5040033, 4668995, 4811720, 4756939, 4756919, 4657304, 4634066, 4754996, 4936429, 4754987, 5031144, 4651597, 4634124, 4847958, 4651587, 4634161, 4755006, 4634125, 4768572, 4737820, 5069818, 4657292, 4936419, 4897943, 4897950, 4669009, 4737834, 4612901, 4888682, 4883520, 4925850, 4969590, 4755003, 4634077, 4795305, 4634075, 4634179, 4669007, 4756927, 4657312, 4787036, 4795281, 4651589, 4756937, 4897969, 4869248, 4897960, 4969576, 4897959, 4737823, 4787021, 4847971, 4825311, 4925854, 4847952, 4936433, 5135445, 4634153, 4657295, 4768579, 4768574, 4768576, 4651593, 4847957, 4897942, 4869253, 4956119, 4925844, 5001182, 5069813, 4956112, 4956101, 4634158, 4825297, 4756928, 4754984, 4754982, 4768577, 4897940, 4847948, 4897944, 4795301, 5040031, 4956110, 4897956, 4956093, 4847962, 5069811, 5095338, 4634177, 4634104, 4657287, 4811714, 4634127, 4634108, 4634138, 4825306, 4897955, 4634143, 5040046, 4883508, 4668998, 4657283, 4768584, 4825303, 5040028, 4897961, 4897952, 4668994, 4787027, 4634163, 5069804, 5150925, 4668992, 4795302, 4657291, 4634149, 4651592, 4634096, 4651591, 4847955, 4634126, 5031158, 5040019, 5040032, 5001172, 4634084, 4634181, 4825289, 4869236, 4969588, 4634162, 4897946, 4634172, 4869240, 4657290, 4883526, 4825314, 4883511, 5040048, 5069820, 4969574, 4756925, 5001179, 4811732, 4969583, 4737818, 4634082, 4737837, 5040038, 4795303, 4811734, 4768570, 4847963, 4651606, 4634080, 4869234, 5040043, 4925841, 4768578, 4795304, 4936431, 4883518, 5069824, 4634079, 4634123, 4811723, 4651583, 4795300, 4883500, 4883512, 4634106, 4634133, 4737826, 4634087, 4634137, 4754985, 4668999, 4651604, 5001161, 4956095, 4669005, 4669000, 4755002, 4657308, 4754988, 4825316, 4787015, 4811742, 4811731, 4847969, 4825302, 4869247, 4869250, 4756932, 4936438, 5031161, 5031153, 4925848, 4634070, 4634069, 5001173, 4883524, 4883516, 4888674, 5069832, 4936441, 4657301, 4634074, 4669010, 4634114, 4737816, 4811735, 4936443, 5095340, 4634060, 4925846, 4936428, 5069825, 4634094, 4888657, 4737817, 4634178, 4657300, 4634068, 4657284, 4737813, 4669015, 4869252, 4956092, 4888667, 4869225, 4669003, 4737827, 4634116, 4657286, 4847951, 4669012, 4768580, 4795280, 4811727, 5135444, 4634089, 4768565, 4634173, 4634117, 4754998, 4612897, 4755001, 4787014, 4755005, 4756924, 4756926, 4811719, 4883509, 4883527, 4883519, 4936436, 4969581, 4969586, 4847943, 4925849, 4869226, 5031142, 5135452, 4936430, 4969566, 4969569, 4847966, 4657285, 4717097, 4634064, 4969578, 4897958, 4795277, 4925838, 4969587, 4634155, 4669001, 4651601, 4657310, 4787016, 4897964, 4969563, 4936424, 4897962, 4925837, 5119615, 4956111, 4925860, 4811721, 4756923, 4612893, 4956104, 4756933, 4787023, 4737815, 4795288, 4634145, 4936440, 5069831, 5119622, 4795296, 4956091, 4737814, 4634148, 4651611, 4657294, 4634176, 5031150, 5135442, 4634144, 4634105, 4737829, 4795283, 4897966, 4634157, 4634134, 4883513, 4634065, 4657307, 4634154, 4634059, 4811722, 5031164, 5069830, 4888655, 4669002, 4717095, 4669008, 4756922, 4883521, 4925835, 4925840, 4936423, 4768564, 4888662, 4651584, 4869229, 4869245, 4925856, 4754993, 4737822, 4897953, 4634110, 4969580, 4768562, 4869237, 4869244, 4756940, 4869241, 4657298, 4651608, 4634160, 4668991, 4768561, 4787034, 4787017, 4811729, 4883499, 4888663, 4969568, 5040027, 4754990, 4768582, 4969570, 4717096, 4634111, 4651598, 4897954, 4755000, 4737836, 4925858, 4668997, 4634131, 4787010, 4883523, 4634112, 4756920, 4651605, 4795282, 4869242, 4657305, 4634101, 4825288, 4634152, 4612904, 4768581, 4897949, 5040029, 4768567, 4768573, 4756921, 4651600, 4787024, 4956116, 4956103, 5031143, 5031146, 5150937, 4768563, 4754989, 4969571, 4768571, 4634086, 4888670, 4634136, 4847965, 4936417, 5031166, 4787013, 4787025, 5069803, 4634147, 4634141, 4768575, 4795298, 5031141, 5040035, 5095342, 4787037, 4612890, 4634098, 4787019, 4634091, 4737821, 4883506, 4634170, 4634099, 4847968, 4847942, 4634081, 4768583, 4634151, 4787026, 5001166, 4528065, 3898057, 4322340, 4398477, 2777403, 4256910, 4507246, 4122666, 4122487, 4523858, 2777350, 4398471, 4224307, 4265495, 4507279, 4449389, 4280817, 4122667, 4523867, 4507264, 4449241, 4523874, 4528054, 4128024, 4322343, 4449267, 4398454, 4507287, 4398464, 2777378, 4232403, 2951986, 4322342, 4449371, 4449379, 4376512, 3693582, 4507273, 4150359, 4185995, 4256905, 4232378, 2951979, 4256901, 4449251, 4398456, 4182738, 4122673, 4505488, 4376526, 4528056, 4122488, 4099804, 4505508, 3610009, 4376540, 3835172, 4032046, 4265515, 3674061, 4449247, 4159912, 4528049, 4216311, 4507266, 3693545, 4153701, 4182739, 4528038, 4232385, 4528085, 4756934, 4787029, 4925852, 4811736, 5222199, 5256832, 5222205, 5332629, 4825300, 5158213, 5241381, 5180307, 5241360, 5040042, 4669014, 5180308, 5040030, 5222188, 5489802, 5445396, 4756936, 5150926, 4969561, 4888668, 5533204, 5259884, 4737833, 4634130, 5411099, 5241337, 5256802, 5332661, 4888656, 5533187, 4883517, 4897967, 5222210, 5411089, 5040026, 5031149, 5259883, 4956118, 5302793, 4634128, 5489809, 5230253, 5180303, 4737830, 5180309, 4925839, 4811715, 5241384, 5533198, 5001178, 4936418, 5180342, 5158220, 5180353, 5222193, 4634062, 5150934, 5302795, 5158227, 5533222, 4936434, 5180306, 5259862, 5411096, 4795285, 5158234, 5095354, 5259891, 4847961, 5256794, 5256813, 5180328, 5241367, 5259865, 5385165, 5410429, 5445404, 5410418, 5490870, 5434673, 5337364, 5180349, 5551667, 4795293, 4969579, 5222190, 5337353, 4847967, 4936416, 4634122, 5180329, 4956117, 5180354, 5180331, 5230242, 5256819, 5434684, 5180335, 5445398, 5256815, 5135448, 5040022, 5241361, 5411088, 5533221, 4969573, 4795294, 5302792, 4869224, 5222206, 5337363, 5337359, 5337367, 5533174, 4754992, 5069807, 5180327, 5230254, 5180339, 4754986, 5069809, 5001170, 5222201, 4825291, 5241366, 5180311, 5259895, 5095356, 5222196, 5230235, 5119604, 5222211, 5230237, 5095361, 5158222, 5332640, 5411101, 5001153, 5095359, 5445400, 5222191, 5434675, 5095360, 4847954, 5385174, 5001154, 5411086, 4925857, 5119618, 4768568, 4795286, 4668993, 5489797, 5533231, 5180346, 4883515, 5150924, 5095339, 4651602, 5180317, 4936425, 4612894, 5158231, 5031157, 5150944, 5031138, 5241330, 5158228, 5332655, 5332645, 5327715, 5434667, 5302817, 5428881, 5385172, 5434686, 5410432, 5180315, 5256796, 5150927, 5150941, 5259879, 4869230, 5337683, 5222187, 5428884, 5483490, 5428882, 5434669, 5533205, 5119612, 5241368, 5483512, 5158229, 5445391, 5533181, 5135454, 5428897, 5222200, 5337655, 5434689, 5327711, 4936427, 5180358, 4888658, 5158226, 5434702, 5259894, 5150930, 4869231, 5259864, 4888677, 5259888, 4888673, 5302801, 5337365, 5337674, 5230248, 5410434, 5069823, 5180322, 5337680, 5259869, 5259872, 4754991, 5483503, 5428887, 5001171, 4825313, 5241386, 5337362, 5445411, 5150931, 4869239, 5040045, 5259874, 5241353, 5259858, 5180350, 5256830, 5337667, 5241332, 5337656, 5533206, 5410421, 5533223, 5069826, 5256823, 5095357, 4897965, 5230233, 5180347, 5241358, 4612895, 5241333, 4969567, 5302789, 5410424, 5069810, 5095362, 5230249, 5241377, 5533199, 4825301, 4956107, 5490861, 5180352, 5040039, 5256833, 5069805, 5069819, 5180334, 5428880, 5256795, 5428904, 5551669, 4669013, 5095358, 4651599, 5434693, 5241379, 4634165, 5001177, 4825307, 5259867, 4888653, 5385167, 5241362, 5158237, 5222198, 5385183, 5150938, 5337681, 5489794, 5150949, 5411083, 4756931, 5241370, 5332649, 5241343, 5150928, 5001175, 5241340, 4956115, 5150948, 5158223, 5180323, 5095364, 5302806, 4969585, 5385178, 5259866, 5533213, 5256804, 4969575, 5040021, 4825317, 4888671, 4669011, 4811717, 5180324, 4888669, 4825315, 5302802, 4956102, 5230258, 5411094, 4825309, 5069822, 4811725, 5337383, 4888659, 5302791, 4634119, 4847956, 5180314, 5180355, 5241375, 4737824, 5410419, 5445414, 5533209, 5533226, 5483515, 5158218, 5158221, 5222186, 5259889, 5259861, 5230247, 5230244, 5256806, 5337654, 5385162, 5150952, 4634072, 5489789, 5230239, 5230260, 5332648, 4888654, 5150936, 5031154, 5259860, 5259875, 5180356, 5241376, 5222197, 5241373, 5337349, 5241380, 5337658, 5337669, 5337666, 5428891, 5434697, 5434690, 5410407, 5533177, 5411091, 5489806, 5489813, 5337687, 5490847, 5337376, 5489810, 5489799, 4956097, 5241350, 4795276, 5150953, 5158214, 5230232, 5241354, 5241372, 5180302, 5095349, 4795291, 5256803, 4768585, 5180357, 4897963, 5150947, 5230240, 5337685, 5119625, 5230250, 5119619, 4811724, 5180321, 4888672, 5119608, 5119602, 4897968, 5410412, 5489808, 5434678, 5302799, 5158216, 5180345, 5337686, 5031148, 5222183, 4897951, 5150933, 5230245, 5230252, 5150929, 5489787, 5410413, 5490856, 4888666, 5489788, 5031160, 5302800, 4925842, 5158215, 5241356, 5302790, 4787011, 5385175, 5490860, 5533201, 5551662, 5040044, 5256805, 5158238, 5445415, 5385168, 5337378, 4811730, 5385171, 5332647, 5180337, 5337668, 5095343, 5040034, 5158233, 5385177, 4612903, 5150935, 5302813, 5332642, 4634171, 5445403, 5410415, 5180304, 4925855, 5483508, 5180320, 5256822, 5040047, 5158241, 5327710, 5222209, 5434700, 5302798, 5445406, 4634175, 5410436, 4612905, 4869233, 5533178, 5337361, 5001168, 5302794, 4888680, 5095367, 5031145, 5119624, 5150943, 5150945, 5259892, 5302816, 5230259, 5241352, 5302788, 5337673, 5445402, 5385166, 5434676, 5434681, 5434692, 5337688, 5489801, 4936432, 4888665, 5256809, 4936444, 5385163, 5489805, 5158219, 5490851, 5180330, 5533227, 5332652, 4795287, 4612892, 4787033, 4634088, 4634169, 5119610, 4717098, 5302805, 5230238, 4657306, 4888678, 5119601, 5241378, 5337661, 5150942, 5385170, 5410435, 4956113, 4737832, 4936437, 5259886, 5302812, 5222202, 4754997, 4795297, 5095352, 5150951, 4756938, 5158232, 5119616, 4634140, 5337369, 5533233, 5135453, 5180312, 5031163, 5119617, 5332659, 5489786, 5490850, 5385164, 5241331, 5222182, 5222192, 5259893, 5327712, 5489798, 5533189, 4925836, 5119599, 4634135, 5150940, 5256799, 5222184, 5533229, 5230256, 4787028, 4787018, 5119614, 5150932, 5158240, 5241369, 5490857, 5135449, 5069816, 4787022, 5256797, 4634078, 5428902, 4811713, 4956108, 5241359, 4869251, 4956090, 4634150, 5222204, 5332630, 5332646, 5428899, 4634071, 5411084, 5095345, 5483499, 5483504, 5589962, 5649122, 5649130, 5649143, 5589959, 5604856, 5630266, 5649150, 5649137, 5630251, 5649126, 5589950, 5649139, 5649134, 5589956, 5604846, 5604859, 5649133, 5649141, 5589964, 5589954, 5560315, 5649127, 5649146, 5560304, 5589965, 5653147, 5615343, 5630244, 5630268, 5649124, 5649123, 5604854, 5426474, 5426476, 5426464, 5426472, 5426454, 5426456, 5426468, 5426457, 5426462, 5426470, 5426459, 5426466, 5426461, 5426469, 5426475, 4108488, 4265506, 4182741, 4066124, 4322347, 4449362, 3917639, 4256906, 4216293, 4449229, 3898101, 4054487, 4142009, 4398486, 4322336, 4505493, 4150354, 4171127, 4182746, 4322344, 4054491, 4528063, 4142028, 4523870, 4256894, 4528052, 4128026, 4150352, 4153689, 4449268, 4507281, 4449252, 4216304, 4216299, 4505492, 4528036, 4376533, 4153704, 4449398, 4069141, 4528084, 4398484, 4634132, 5337371, 5241357, 5490871, 5434683, 5411103, 5337678, 5256818, 5259868, 5332631, 5095348, 4634120, 5180343, 5230246, 5483500, 5119597, 5259859, 5031151, 5180333, 5489785, 5095363, 5337358, 4883502, 5031155, 5490859, 5302810, 5434668, 5001169, 5445397, 5337672, 5302814, 4847946, 5410423, 5533218, 5241385, 5551673, 5241329, 5411093, 5490865, 5490846, 5411100, 5337374, 4825312, 5551683, 5332641, 5434695, 5551672, 5551671, 5490862, 5483514, 5095347, 4969572, 5533228, 5241345, 5031137, 5095350, 5411102, 5410431, 4634164, 4825305, 5490849, 5551680, 4869227, 5490864, 5434679, 5180319, 5337350, 5135451, 5241341, 5302796, 5241382, 5490844, 5445413, 5040040, 5489811, 5337379, 5241371, 5428901, 5411098, 5445416, 5489804, 5533196, 5180305, 5410416, 5180325, 5332632, 5031165, 5533183, 5256820, 5551670, 5445407, 5533175, 5385182, 5445395, 5031159, 5410408, 5337664, 5256825, 4811716, 5551686, 5135443, 5551665, 5533230, 5040036, 4936421, 5533224, 5434699, 5551684, 5256831, 5001174, 4795278, 5428903, 5490866, 5332651, 5180351, 5533212, 5180344, 5445401, 5533188, 5241383, 5256811, 5490869, 5445410, 5332654, 5337372, 5180326, 5241347, 5428888, 5489800, 5551690, 5428879, 5483509, 5434670, 5434698, 5533215, 5001162, 5533217, 5533208, 5445412, 5445390, 5410420, 5135447, 5256816, 5551677, 5040024, 5337660, 5241349, 5001176, 4883503, 5256801, 5385176, 5428898, 5428900, 5483511, 5069828, 4657309, 5337671, 5001159, 5410417, 5259877, 5259896, 5256817, 5483517, 5551682, 5551678, 4888679, 5337373, 5337380, 5489814, 5533214, 5533180, 5489792, 4925853, 5410430, 4657293, 4869232, 4897947, 4825292, 5259881, 5230255, 5489807, 5551666, 5119598, 5119623, 5095341, 5533179, 5533194, 5119611, 5445393, 5490855, 5428896, 5551676, 5259882, 5327713, 5551668, 5489795, 4897945, 4936422, 5385179, 5533193, 4956096, 5180341, 5411087, 5337375, 5337670, 5241364, 5332633, 5445394, 5411092, 5483505, 5428889, 5411085, 4737819, 5428878, 5119605, 4956109, 5332639, 5551688, 4925843, 4956098, 5551681, 4847944, 5241363, 4756929, 5533191, 5483496, 5241346, 5445405, 5241344, 5135457, 5410425, 4825293, 5337352, 5490863, 5222203, 5551675, 4811741, 5337657, 5135458, 4634167, 5302809, 5483513, 5445399, 5411079, 5332628, 4847945, 4969565, 4847947, 4787008, 5241338, 5259890, 5337356, 5222185, 5533197, 5241348, 5489796, 5241365, 5490852, 5428877, 5337663, 5337357, 4883528, 5483507, 5135450, 5241342, 5483516, 5332644, 4768569, 6051818, 5649142, 5589943, 6062541, 6062528, 6108534, 5845332, 5845346, 5959798, 6051810, 5913020, 5845357, 5630246, 5959796, 5913004, 5959787, 5604857, 5649128, 6130303, 6154595, 5941154, 5560310, 5615334, 5615346, 5630255, 5959783, 6037940, 6045202, 6045198, 5913017, 5845360, 6037942, 5845340, 6062527, 5845343, 6051778, 6068479, 6068501, 5560302, 6062511, 6068478, 6068482, 6068475, 6100679, 5845342, 5941169, 6062513, 5615348, 6068486, 6088031, 6108540, 5653119, 5941158, 6130319, 6062537, 5653122, 5589944, 5941128, 5941138, 5560312, 5959802, 5560305, 6154612, 5589946, 5913024, 6051807, 6051808, 5630270, 5649131, 6098638, 5615336, 5630263, 5941135, 5604848, 6045203, 6030087, 5941164, 6051785, 5615333, 5615341, 5941165, 6062516, 6108554, 6108552, 5653112, 6100703, 5845355, 5649147, 6051793, 5615354, 5649145, 5653111, 6037900, 5604851, 5560316, 5897622, 5913025, 5630259, 5959801, 5959775, 6051792, 6062518, 6100702, 6108545, 6108539, 6130311, 6183326, 6037913, 6030079, 6100684, 6108555, 5560308, 5604837, 6045185, 6051784, 6051826, 5941139, 5941173, 6098645, 6062547, 6244178, 5913005, 5615356, 5604863, 6051795, 6062517, 6100680, 6051804, 5630257, 5589953, 5941126, 6068483, 5604839, 5589948, 5653152, 5615359, 5941134, 5941143, 6045200, 6062538, 6068497, 6062510, 6051794, 6051803, 6108535, 5604862, 5589955, 5653108, 5845350, 6062546, 6098656, 5604850, 5649149, 5653098, 6003167, 5604852, 6098642, 6062515, 5653103, 6037935, 5653116, 6045184, 5589947, 5560290, 5913002, 5845336, 6154598, 6037908, 5589937, 6068474, 5913015, 5845349, 5589940, 5653114, 6100687, 5604847, 5589963, 5630256, 5913018, 6062531, 5560288, 5630243, 5959778, 6003176, 6037930, 6051782, 6098632, 6130332, 6108550, 6154600, 6154596, 6273210.6045190, 5615350, 6098652, 6100690, 6273229, 5845356, 5941133, 6098648, 5615355, 6154617, 6108561, 6244169, 5604840, 6062525, 5615357, 5589938, 6068484, 6068509, 6098658, 5560309, 5630242, 6037945, 6068481, 6130307, 6045191, 5653101, 6045188, 6045201, 6045183, 6037936, 5630250, 5941141, 5845341, 6108533, 5653110, 5630248, 5560313, 5653106, 5589952, 5649144, 5653107, 5845351, 5897621, 5845362, 5913010, 5959803, 5941170, 5941168, 5941171, 6003174, 6003162, 6003159, 5615345, 6037916, 6045193, 6037907, 6051813, 6051797, 6037915, 6088036, 6062514, 6098629, 6098635, 6051780, 6100696, 6062526, 6037934, 6108559, 6244157, 6244154, 6108541, 5653105, 5649148, 6037914, 5615331, 5560306, 6045196, 5589960, 6030078, 6068473, 6130312, 6045177, 5845347, 6045197, 5653099, 5941174, 6030083, 6051800, 5560287, 5653109, 5649132, 5653100, 5941155, 6051781, 5615339, 5649129, 5845339, 5604836, 5604864, 5653118, 6051809, 6037946, 5941136, 6062512, 5589942, 5959793, 5604841, 5604845, 5630254, 5941119, 5845344, 5560289, 5959779, 6003168, 6003166, 6051817, 6003158, 6037932, 6045181, 6068489, 6068490, 6088034, 6088041, 6100694, 6098630, 6098649, 6130309, 6130308, 5560294, 5630269, 5845358, 5615335, 5589958, 6098647, 6068488, 6051806, 5845335, 6051788, 5845353, 6030077, 5897624, 5941145, 6183293, 5653133, 5913006, 6037918, 6037901, 6037903, 6100698, 6183274, 6183284, 6183285, 6183312, 6183332, 5653129, 5653143, 6068504, 6068505, 6088032, 6130306, 6108537, 6108548, 6154604, 6154610, 6183289, 6271780, 5653144, 5653149, 6130320, 6244156, 6244168, 5653153, 5653124, 5653148, 5653137, 5653134, 5653155, 5653156, 5653131, 5653130, 5653141, 5653126, 5959788, 6068496, 5941162, 5604844, 5615332, 5604842, 6037920, 6037928, 5913003, 5653128, 5589961, 5604860, 6037947, 5426453, 5426473, 5426477, 5426452, 5426471, 5426455, 5426460, 5426467, 5426478, 4449361, 4449230, 4322346, 4449396, 4150357, 4505501, 4265499, 4171121, 4523871, 4150374, 4322316, 4505484, 4128035, 4142035, 4142020, 4449254, 4376542, 5040025, 5001158, 5332653, 5031139, 5483506, 5533232, 5337662, 5150950, 5180338, 4925833, 5158235, 5434672, 5158239, 5337355, 5259873, 5445408, 5230257, 5230251, 5332635, 5259885, 5489812, 5337676, 5241355, 4795299, 5533176, 5119607, 5180318, 5095344, 5411090, 5411078, 5410433, 5332638, 4634076, 5040037, 5332657, 5533207, 5337360, 5256826, 5551664, 5119603, 5241351, 5230236, 5533200, 5302804, 5490848, 5332660, 5222207, 5533190, 5259878, 5135456, 4795292, 5410422, 5040023, 5119620, 5410428, 5180348, 5119600, 5256798, 5337677, 4847960, 5222208, 5259876, 4717099, 5302803, 5031156, 5031162, 4925847, 5158224, 5158217, 5489803, 5332636, 5489791, 5158236, 5256800, 5095346, 5428895, 4787031, 4869238, 4657299, 5256810, 5410427, 5327714, 6051798, 6273233, 6062545, 6088027, 5630264, 6273211, 6037905, 6247202, 6342057, 6342063, 6250396, 6537487, 6244180, 6154613, 6247200, 6250410, 6330504, 6556404, 6062530, 6183292, 5615353, 6300705, 6271790, 5649151, 6273228, 6330513, 6338398, 6244182, 6330524, 6342059, 6338371, 6183282, 6098636, 6183278, 5630271, 6556396, 6037938, 6154597, 6250435, 6300682, 6300686, 6330508, 6183276, 6300707, 6244176, 6250432, 6330525, 6330515, 6250434, 6338395, 6338385, 6338409, 6098650, 6045195, 6271777, 5649140, 6051787, 6347942, 6537479, 6037931, 6330507, 6300698, 5615358, 6247210, 6455947, 6347947, 5604858, 6537486, 6338400, 6537533, 6300709, 5959797, 6271766, 6051820, 6392297, 6068477, 6342042, 5897623, 6300702, 6273232, 6342052, 5845331, 6330520, 6030084, 6455940, 6338376, 6068498, 6244177, 6068492, 6250395, 6345273, 6338408, 6098653, 6271768, 6537478, 6244179, 6338392, 6556366, 6130325, 6108538, 5560298, 5615342, 6330522, 6338374, 6108557, 6342043, 5941125, 5941148, 6183273, 6338373, 6392310, 6338381, 6455957, 6154601, 6154591, 5630247, 6244167, 6250408, 6537523, 6537526, 6154594, 5615340, 6347936, 6330500, 6250420, 6392291, 6244164, 6051814, 6247193, 6250399, 6247185, 6300693, 6347964, 6342064, 6455960, 6183322, 6300706, 6347955, 6247207, 6271789, 6250406, 6062533, 6330529, 6271767, 6271771, 6250423, 6330527, 5615338, 6068485, 6068493, 6347951, 6537475, 5941130, 6247213, 6051816, 6338399, 6273235, 6130318, 5649135, 6130317, 6330518, 6300704, 6455956, 6537501, 6130315, 6154590, 5959786, 6455945, 6345277, 6183304, 6338391, 6338379, 6455964, 6247209, 6250431, 6244175, 6556355, 6130314, 6345268, 6108562, 5630258, 6250413, 6537509, 6037909, 6037941, 6098655, 6342051, 6455955, 6244183, 6250391, 6250433, 6300695, 6342061, 6342047, 6271764, 6556379, 6392289, 6244163, 6271778, 6537492, 6556361, 6088047, 6068476, 6062508, 6451358, 6537531, 6347956, 6247186, 6271776, 6003161, 6250438, 5653121, 6247189, 6330519, 6392322, 6108536, 6247205, 6247206, 6250427, 6342039, 6342066, 6062535, 6345271, 6100686, 6455958, 6537507, 6330528, 6100683, 6338410, 6300685, 5630245, 5604838, 6244165, 6556401, 6108543, 6330505, 6037902, 5653117, 6273231, 6244159, 6108558, 6247192, 6271786, 6250409, 6342053, 6300684, 6342044, 6392318, 6247187, 6300697, 6250411, 6392287, 6537529, 6556358, 6597816, 6273225, 6250394, 6183280, 6342058, 6098640, 6250437, 5653102, 6062540, 6347958, 6100706, 6455962, 6558530, 5913011, 6183283, 5560314, 6183305, 6247194, 6342055, 6183327, 6537491, 6045178, 6088028, 6154615, 6455939, 6342056, 6100697, 6271788, 6051777, 6051812, 6003160, 5845330, 6347965, 6392324, 5630261, 6068508, 5959785, 6062521, 6250430, 6062522, 6098639, 6273219, 6250424, 6250414, 6108546, 6183298, 6244181, 6183299, 6244166, 6247212, 6342067, 6342060, 6537497, 6100699, 5941160, 6051821, 5913001, 6347948, 5913014, 5604853, 6250412, 6342045, 6537480, 6250440, 6330526, 6244162, 6183311, 6347961, 6250397, 6300694, 6250386, 6273221, 5630260, 6345275, 6030086, 5913013, 6250393, 5959795, 6250441, 6108553, 6154605, 6244173, 6273213, 6347943, 6455953, 5649125, 6338380, 6455951, 6537528, 6558504, 6558507, 6271785, 6271791, 6250387, 6088050, 5653154, 5653136, 5653146, 5653135, 6300692, 6045186, 6392308, 6154618, 6247199, 6100705, 6100700, 6250384, 6345270, 6250388, 5653138, 6045176, 6051822, 6330514, 6100691, 6100695, 6183323, 6250419, 6330521, 5913021, 6051790, 5913012, 6392285, 6037911, 6273214, 6154599, 6154592, 6244171, 6037921, 6455944, 6273236, 6273224, 6154607, 5653120, 6062529, 6045204, 6250415, 4449399, 4507249, 4507270, 4449240, 4847970, 5119626, 5241339, 5490858, 5533195, 5428886, 4869235, 5069808, 4756942, 4651612, 4795295, 5031140, 5551689, 5337659, 4883525, 5483488, 5001155, 5533220, 5001156, 5411080, 5337684, 5158225, 5332662, 5180310, 5256827, 5069829, 4956114, 5256808, 5180336, 5001165, 4847950, 5180301, 5180340, 4847953, 5337382, 4825295, 5230241, 5337381, 4969582, 5337665, 5551679, 5241335, 5411081, 5332656, 4634085, 5411095, 5445409, 4847964, 5490867, 4897948, 4795289, 5337370, 4634115, 5150939, 5434674, 5332637, 5256812, 4754995, 4969562, 5332650, 5031147, 6273209, 6858677, 6300680, 6556389, 6788765, 6098631, 6250401, 6778970, 6037923, 6788775, 5959792, 6858676, 6778971, 6098644, 6051786, 6558515, 6154602, 6183277, 6300699, 6749914, 6613501, 6330523, 5589957, 6271781, 6338370, 6330506, 6455959, 6779001, 6778969, 6749923, 6749909, 6749900, 6749912, 6330517, 6558554, 6788778, 5560292, 6451349, 6597831, 6392300, 6749874, 6749893, 6556348, 6273216, 6857610, 6597821, 6003177, 6098651, 6556393, 6613502, 6858671, 6597828, 6347950, 6338394, 6451363, 6051823, 6051779, 6749924, 6858667, 6250418, 6183287, 6088044, 6779012, 6030081, 6597845, 6558536, 6558518, 6838010, 6857622, 6788787, 6271773, 6300690, 6537489, 6597856, 6556381, 6455941, 6342038, 6451360, 6392313, 6788785, 6556384, 6779015, 6788781, 6597808, 6537506, 6556354, 6392311, 6749895, 6749915, 6788795, 6597846, 6273227, 6250402, 6858669, 6330511, 6098646, 6597802, 6556353, 6613503, 6068511, 6051805, 6558544, 6749883, 6597838, 6858670, 6451356, 6347952, 6088042, 5845354, 6558509, 6244172, 6597851, 6558539, 6250392, 6068494, 6778983, 6597811, 6597832, 6749876, 6597854, 6597800, 6858680, 6788791, 6347954, 6858684, 6068487, 6342048, 6597850, 6597827, 6183313, 6130321, 5845329, 6858674, 6244174, 6558543, 6247197, 6347941, 6597797, 6558541, 6537498, 6347967, 6183295, 6558502, 6788759, 6098641, 6392317, 6154616, 6003163, 6130322, 6556357, 6558546, 6455946, 6556380, 6342065, 6597798, 6108549, 6749892, 6338389, 6556369, 6455937, 6537503, 6597818, 6250405, 6003172, 6558525, 6788782, 6558551, 6556363, 6838001, 6556356, 6183310, 6338407, 6347949, 6749927, 6537482, 6003165, 6338401, 6558521, 6778998, 6451354, 6392295, 6455949, 6455948, 5630267, 5604849, 6183318, 6829524, 6779003, 6451355, 6858690, 5589966, 6558550, 6857626, 6455961, 6537514, 6749908, 6597824, 6749906, 6455950, 6857638, 6347966, 6100701, 5560299, 6556347, 5941127, 6556390, 6130328, 6779014, 6778999, 6778985, 6556374, 6749929, 6556376, 6271765, 6062544, 5649138, 6273218, 5630249, 6558524, 6829512, 6597847, 6597805, 6749920, 6778990, 6250403, 5845328, 5653123, 6068510, 6130304, 6597812, 6778996, 6558534, 5845345, 6537525, 6250421, 6338382, 6556383, 6613510, 6749916, 5653115, 6597830, 6749878, 5959800, 6108551, 6154609, 6857614, 6100688, 6342054, 6597817, 6558549, 6088033, 6088046, 6250436, 6778979, 6613505, 6857628, 6342046, 6613495, 5653113, 5560303, 6613506, 6037929, 6857635, 6556406, 6537520, 6451353, 6788767, 6062542, 6556365, 6597804, 5941147, 6098643, 6068480, 6338393, 6556395, 6597849, 6597806, 6857616, 6051825, 6779002, 6062519, 5959804, 5845352, 6749898, 6030085, 6558527, 6183303, 6342062, 6392302, 6451364, 6537527, 6537512, 6556352, 6597855, 6779008, 6779018, 6558538, 6250398, 6558505, 6183302, 6273226, 6537519, 6392307, 6597837, 6749928, 6556392, 6537494, 6597795, 6778991, 6451362, 6451365, 6247198, 6749896, 6788761, 6392294, 6051799, 6037948, 6062523, 5941144, 6749889, 6597834, 6597823, 6779004, 6858695, 6749897, 6749911, 6347946, 6338403, 6003171, 6451366, 6597803, 6247195, 6100681, 6045187, 6045205, 6788757, 6098633, 6537496, 6100685, 6330510, 6788771, 6392314, 5615352, 5653140, 6250389, 5941176, 6392290, 5959781, 6392320, 6338387, 6051791, 6051819, 6250407, 6558513, 6558533, 6273223, 6558529, 6037949, 5913019, 6558512, 6037906, 6273207, 5604861, 6749891, 6558532, 6045182, 6455954, 6250417, 6778964, 6338405, 6244155, 5630265, 7009077, 6992092, 6975927, 6975932, 6975902, 6975898, 7037036, 6975919, 6916002, 7037018, 7032746, 6992084, 6916018, 6975901, 6915999, 7009073, 6993611, 7032727, 7037023, 6970351, 6980916, 6916008, 6992100, 6980925, 6975930, 6916010, 6915997, 6916011, 6970367, 6970358, 6970362, 6975909, 6980933, 6980912, 7009075, 6993605, 6993607, 6993610, 6980921, 6975914, 6975921, 6980920, 6980922, 6980924, 6980914, 7032715, 5332643, 4847949, 5410426, 5135455, 5533186, 5180316, 5259897, 5428890, 4969564, 5222189, 5337368, 5434682, 5069817, 5230234, 5069806, 5385181, 5410414, 5445389, 5385169, 5489790, 5135446, 5483493, 5483489, 5490868, 5222194, 5490853, 6537515, 6183307, 6597793, 6778989, 6183319, 6392286, 5913009, 6749901, 6088051, 6556350, 6778984, 5630252, 6271769, 6556367, 6098657, 6749904, 5560307, 6788764, 6597825, 6558522, 5845334, 6183329, 6858688, 6062532, 6558545, 6130324, 6250442, 6271783, 6392312, 6749913, 6088048, 6858673, 6749877, 6858679, 6392306, 6537505, 5941137, 6130330, 6749894, 6788776, 6858696, 6338383, 6749884, 6088037, 6537508, 6778977, 5941131, 6451367, 6788770, 6778966, 6778995, 6247196, 6247191, 6300703, 6779009, 5941140, 6613509, 6778980, 6338378, 6108547, 6597796, 6537518, 6838004, 6537530, 6273222, 5959776, 6537532, 5959791, 6271792, 6558540, 6088029, 6183317, 6068495, 6300696, 6857613, 6556370, 6273220, 6749887, 6250428, 6858668, 6558500, 6556360, 6558496, 6247203, 6556405, 5560293, 6558514, 6455943, 6778981, 5959777, 5615351, 6858678, 6537504, 6062524, 6788763, 6558506, 6749882, 5653145, 6558523, 6088025, 6857621, 6857615, 5560297, 6338377, 6858686, 6779016, 6857633, 6098637, 6250425, 6778987, 6597841, 5941150, 6273208, 6068499, 6338372, 6062543, 5649136, 5959782, 6556371, 6858687, 6613513, 5653104, 6613511, 6558498, 5897620, 6749905, 5941142, 6342041, 6537490, 6342050, 5941163, 6100693, 6788788, 6330509, 6788766, 6788768, 6392321, 6749910, 6455935, 6597794, 6556394, 6338396, 6558520, 6244160, 6597815, 6788769, 6338386, 5941146, 6778982, 6088035, 6537474, 6088022, 6250422, 6858685, 5653125, 6558497, 6244161, 6558495, 6558499, 6749888, 5653139, 6558531, 6455936, 6858665, 6273234, 5941177, 6788786, 5959789, 7000541, 7000542, 7000553, 7000562, 7037030, 7037034, 7032739, 7000572, 6970348, 6975900, 7000575, 7037015, 7032734, 6980918, 7009070, 6916000, 7032744, 6980909, 6975897, 7032764, 7009078, 7037043, 6993613, 7000561, 6992083, 7032749, 7009076, 6980930, 7009063, 7000537, 7000577, 6980928, 7000567, 7000564, 7000574, 6993608, 6992098, 7032757, 6980923, 6993599, 7032710, 6993606, 6992082, 7009071, 6992093, 6980934, 7000548, 6916015, 7032760, 6975916, 7037026, 7032731, 6992101, 7009068, 6980932, 7032743, 6970352, 7000539, 6992096, 7000555, 6975920, 6975923, 7000581, 7009062, 6993612, 7037007, 7032720, 6992085, 7032758, 6970357, 7000576, 7000556, 7000536, 7032721, 7009064, 7009067, 6916007, 6975899, 7000531, 7037013, 7032708, 7032726, 7032754, 7032762, 7032763, 7032766, 6916017, 7000532, 6916003, 7000529, 7032713, 6916005, 7032718, 6980910, 7032724, 7000584, 7000563, 5426465, 6537481, 6556375, 6556385, 6556391, 6597809, 6858693, 6154593, 6779011, 5589939, 6778965, 6183290, 6273217, 6347959, 6130323, 6558517, 6045179, 6392292, 6154589, 6037904, 6183288, 6597822, 6778975, 6778960, 6392309, 6749875, 6392315, 6558501, 6030080, 6345272, 6779006, 5560295, 6537517, 6273230, 6451350, 6778993, 6537477, 6558548, 6271779, 6597807, 5845337, 6556349, 6068502, 6154606, 6597839, 6338375, 6788760, 6556402, 6556378, 6558511, 6098634, 6392299, 6778967, 6558535, 6749903, 6183328, 6037943, 6556362, 6779000, 6338402, 6556368, 6975911, 7032709, 7032712, 7000558, 7037002, 6992088, 6980907, 7036994, 6916009, 7000583, 7032714, 7000566, 7032751, 7037035, 7009069, 7037014, 7000530, 7037024, 6993601, 6980915, 6970366, 6970355, 6970360, 6980908, 7000545, 7032737, 6970350, 7032755, 6788779, 5941167, 5941132, 6537500, 6558519, 7032728;
c) Item Number 314-13-04, UDI/DI 10885862172709, Serial Numbers: 2204772, 2204779, 2204774, 2460084, 2406473, 2336200, 2204788, 2330962, 2504205, 2204784, 2336215, 2204777, 2330966, 2204781, 2406476, 2336210, 2336214, 2474451, 2622751, 2622784, 2204783, 2460099, 2665816, 2336198, 2522063, 2479368, 2330965, 2330969, 2336207, 2336211, 2479372, 2336216, 2504215, 2504210, 2336219, 2622753, 2622767, 2504201, 2689133, 2777415, 2777421, 2777423, 2777417, 2777420, 2460080, 2522061, 2622750, 2522074, 2522057, 2479377, 2522067, 2474446, 2665823, 2336205, 2460081, 2522073, 2336199, 2622765, 2330953, 2474447, 2665840, 2665829, 2330956, 2622759, 2504196, 2622754, 2665851, 2460095, 2622758, 2622764, 2522058, 2336224, 2665826, 2504202, 2689147, 2951897, 2689132, 2975696, 2975730, 2689135, 2689142, 2816793, 2816787, 2770117, 2975729, 2951905, 2975705, 2689134, 2816795, 2689136, 3601863, 3628992, 2951904, 2816803, 2777419, 3642412, 3642416, 2204786, 2479379, 2479383, 2522070, 2460085, 2622776, 2665839, 2336203, 2665846, 2336223, 2622775, 2665842, 2460088, 2522065, 2522071, 2665834, 2665841, 2479370, 2479375, 2504199, 2336213, 2689145, 2951891, 2689127, 2816785, 2951890, 3917703, 3577177, 2770119, 3917711, 3588516, 2951900, 2689129, 3588522, 3628972, 2951898, 2777424, 2975712, 2975726, 2975704, 3601868, 2951883, 3588510, 3642425, 3642423, 3588503, 3917737, 3917740, 3917735, 3628991, 3628993, 4150401, 2975703, 3864541, 3925468, 4122486, 4122484, 2951889, 3917730, 2975725, 2689125, 2689131, 2689138, 2816807, 3917702, 3628968, 3628989, 3628980, 3628963, 3628969, 3628974, 3628978, 3628982, 3642427, 3917695, 3917700, 2504213, 2622756, 2622755, 2330959, 2330971, 2504203, 2504206, 2336218, 2622773, 2522072, 2622781, 2951899, 4224146, 3917719, 3917720, 4193254, 4141945, 4447555, 2816812, 4150397, 3628996, 2975713, 3629009, 4150400, 4141948, 3588517, 2951901, 2777416, 3917709, 4193259, 3601866, 3917692, 4141952, 3629008, 3628984, 2816792, 4141938, 2951911, 4224347, 4150388, 2975693, 4509410, 3917731, 4122485, 2975706, 3577176, 2816789, 2975718, 2689121, 4256856, 4265558, 4298065, 4256853, 4265532, 4338528, 3628981, 4338517, 3917712, 3628988, 4193248, 3917722, 2522064, 2622779, 2622783, 2665820, 2336204, 2665844, 2504216, 2665853, 2665849, 2479378, 2479369, 2665832, 2460082, 4150385, 2816805, 2975728, 4447556, 4193260, 4338533, 4338512, 4265545, 4224345, 2816784, 2816783, 2816790, 4265549, 2975709, 3628995, 4338511, 3628976, 4141951, 4141950, 4150403, 3577184, 3917696, 3917713, 4509412, 3588515, 3629011, 4447542, 4447564, 4509417, 2975699, 4509420, 3628987, 3601872, 4141947, 4338531, 4150382, 3925467, 4150399, 2689124, 2951909, 4265547, 4572626, 4666339, 4666320, 4825377, 4572641, 4666328, 4572636, 4572645, 4825369, 4867572, 4942194, 5154267, 4825372, 5038272, 4888866, 4572627, 4924818, 5080542, 2622763, 3588507, 4509401, 4265559, 4338524, 2975719, 4338534, 3917723, 4338510, 3917721, 4265556, 4141939, 3917707, 4256850, 4150393, 4338532, 3577178, 4141957, 2975723, 3642410, 2951910, 4338539, 4193262, 5291564, 5332706, 4888840, 4572631, 4867574, 5332708, 5361241, 5038270, 4924820, 5038265, 5249348, 5361214, 5038263, 5038268, 4924824, 5038260, 4924817, 5065708, 4888859, 5065723, 4867589, 5065731, 5065726, 5038278, 5065724, 5332716, 5154287, 4997556, 4997553, 4997551, 4924821, 4924812, 5065718, 4924819, 4888845, 5291554, 5249334, 5038264, 5287709, 5291560, 5332704, 5332711, 5556719, 5576116, 5556722, 4509397, 4224344, 4447545, 4256854, 4338522, 4224343, 4349383, 4265534, 4265550, 4265561, 3642413, 3629004, 4141949, 3917727, 5065710, 5287717, 4942186, 4825360, 5154275, 5291556, 4666336, 5287702, 5291570, 5351555, 5291567, 5154283, 5065725, 5249339, 4942192, 5291574, 5361230, 5287698, 4888860, 5361225, 4888838, 5249332, 5154289, 5332694, 5351528, 5065716, 5361235, 4942197, 5065728, 5065732, 5038271, 5154288, 5332715, 4942178, 4867584, 5351545, 5593395, 6039299, 5556705, 5556696, 6039324, 5593414, 5851486, 5556702, 5943296, 6086340, 5582403, 5556704, 6088167, 5582402, 6088169, 5556709, 6039310, 6086313, 5943293, 6051673, 6088173, 6039304, 6051668, 6086315, 5582412, 6086314, 5593394, 5593409, 6088154, 6088155, 6088144, 6088158, 6086307, 6086337, 6087870, 6087879, 6295743, 5593402, 5593406, 6068649, 6295729, 6295744, 4193251, 4265533, 4150386, 4150387, 4265548, 4338516, 4265560, 5154270, 5287720, 5249328, 5351551, 5510483, 4572640, 5287713, 5361223, 4924814, 4942187, 4825366, 4867585, 5943295, 5582392, 6086318, 5582406, 6342079, 6237986, 6068645, 6088162, 6068644, 6513981, 6039311, 6039314, 5556720, 6237997, 6051671, 5593416, 5593417, 6295733, 6507937, 6466891, 6214879, 6087851, 6237979, 4509415, 5351532, 5361229, 5287701, 4924826, 5038273, 5291582, 5249327, 5154284, 4825367, 4888856, 5065722, 5351544, 5154273, 5154291, 5287696, 5249337, 5510477, 5582418, 5943304, 6507934, 6295737, 6295727, 6214878, 6507923, 6051676, 6051667, 6466896, 6342037, 6507930, 6399484, 6051670, 5556703, 6507940, 6610953, 6087885, 5851489, 6087867, 5556710, 6940054, 6931880, 6931407, 6931397, 6990711, 6990719, 6990724, 6925908, 6925913, 6925903, 6925898, 6931398, 6931404, 6931405, 6931406, 6931418, 5080545, 5510494, 5332713, 4997552, 5287693, 5065711, 5351541, 5291561, 6237982, 5576106, 6342077, 6610958, 6237989, 5943305, 6237984, 6039298, 5943303, 6399493, 6039312, 7019743, 6931895, 7017171, 6931416, 6925902, 5332701, 6088160, 6087849, 6542766, 5956457, 6087858, 6466895, 6940064 |
| FEI Number |
1038671
|
Recalling Firm/ Manufacturer |
Exactech, Inc. 2320 Nw 66th Ct Gainesville FL 32653-1630
|
| For Additional Information Contact | Thien Doan 352-377-1140 |
Manufacturer Reason for Recall | The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH). |
FDA Determined Cause 2 | Nonconforming Material/Component |
| Action | Beginning 03/06/2024, Exactech issued an URGENT MEDICAL DEVICE RECALL notice and an URGENT DEAR HEALTHCARE PROFESSIONAL COMMUNICATION to its consignees via email. The recall notice explained the issue, clinical impact, and requested the following actions be taken:
"" CAREFULLY REVIEW THIS RECALL NOTIFICATION to ensure that you fully understand the issue identified, the recall strategy, and all actions required.
" IMMEDIATELY IDENTIFY AND QUARANTINE any of the subject devices in your inventory and/or customer s inventory listed Product Scope Listing.
" EXTEND THE DESCRIPTION OF ISSUE AND CLINICAL IMPACT as described in the recall notification to your accounts that may have this product in their possession.
" COMPLETE AND RETURN the attached Recall Inventory Response Form to Exactech via email at recalls@exac.com.
The communication to healthcare providers referenced the FDA Safety Communication issued on 01/16/2024, explained the issue, potential health risks, and suggested patient monitoring for those with implanted affected devices for potential device wear, failure, or bone loss and should consider performing X-rays to further evaluate the patient and their implanted device if you suspect a failed device.
Contact your local Exactech Representative if you have any questions regarding this communication. |
| Quantity in Commerce | 11031 units |
| Distribution | Nationwide and Australia, Austria, Belgium, Brazil, Chile, Canada , China, Colombia, France, Germany, Great Britain, Greece, Guatemala, India, Italy, Japan, Korea, Luxembourg, Malaysia, Netherlands, New Zealand, Portugal, Singapore, South Africa, Spain, Sweden, Switzerland, Thailand |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = KWS
|
|
|
|