| | Class 2 Device Recall Angiodynamics DURAFLOW |  |
| Date Initiated by Firm | March 18, 2024 |
| Date Posted | May 03, 2024 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1768-2024 |
| Recall Event ID |
94349 |
| 510(K)Number | K101843 |
| Product Classification |
Catheter, hemodialysis, implanted - Product Code MSD
|
| Product | DURAMAX STACKED TIP 28CM STR. BASIC KIT, 1-UP PG- Indicated for use in attaining Long-Term vascular access for Hemodialysis and Apheresis.
Product Number: H787103028041 |
| Code Information |
UDI-DI: 15051684011846
Lot Numbers: 5805840
|
Recalling Firm/ Manufacturer |
Angiodynamics, Inc. 603 Queensbury Ave Queensbury NY 12804-7619
|
| For Additional Information Contact | SAME 518-798-1215 |
Manufacturer Reason for Recall | Tri-Ball Tunneler may contain loosely affixed metal shavings, potential for the metal shaving to embolize, possibly requiring removal via surgical procedure. |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | Angiodynamics issued Recall Notifications Packages via Federal Express to 1 Domestic
Consignee on 2024-03-18 . Letter states reason for recall, health risk and action to take:
IMMEDIATELY
o Locate the affected product in your Inventory, cease any and all distribution segregate in a secure location pending return.
o Forward a copy of this recall notification to any locations to which you have distributed affected product.
2. Complete and return the Reply Verification Tracking Form.
" Promptly complete, sign, and return the enclosed Reply Verification Tracking Form (even if you do not have any product to return), following the directions on this page and the Reply Verification Tracking Form.
Email Reply Verification Tracking Form (preferred): recall@angiodynamics.com
Fax Reply Verification Tracking Form:
Attn: Hemodialysis Catheter Kits with Tri-Ball Tunneller
Fax number 1-855-273-0519
" Call AngioDynamics Customer Service at 1-800-772-6446 between 8:00 a.m. and 7:00 p.m. (Monday Friday: Eastern Standard Time) to obtain a replacement or credit for your returned product.
Package any product that is being returned in an appropriate shipping box.
" Write the RMA number on the RMA label (provided on the Recall Verification Tracking Form) and affix the label to the outside of the shipping box.
" Seal the box and return via your preferred carrier to:
AngioDynamics, Inc.
24 Native Drive
Queensbury, NY 12804
Attn: Hemodialysis Catheter Kits with Tri-Ball Tunneller |
| Quantity in Commerce | 30 units |
| Distribution | VA |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|