| | Class 2 Device Recall Medline |  |
| Date Initiated by Firm | April 08, 2024 |
| Date Posted | August 27, 2024 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-2632-2024 |
| Recall Event ID |
94582 |
| Product Classification |
Tray, surgical - Product Code LRP
|
| Product | Medline procedure packs containing Shenli syringes, labeled as follows:
1) NEURO ANGIO PACK, Part Number DYNDA1431A;
2) CVC INSERT BUNDLE 16CM, Part Number ECVC3260;
3) CENTRAL LINE INSERTION KIT, Part Number ECVC7440;
4) CENTRAL LINE INSERTION KIT, Part Number ECVC7445;
5) CENTRAL LINE INSERTION KIT W/20CM CATH, Part Number ECVC7445A;
6) VANTEX 7FR 3L 20CM CVC BUNDLE US, Part Number ECVC8045;
7) CVC DOUBLE LUMEN 16CM, Part Number ECVC8105 |
| Code Information |
DYNDA1431A, Lot Number 2019021190;
ECVC3260, Lot Number 2019091850;
ECVC3260, Lot Number 2019120950;
ECVC7440, Lot Number 2021071950;
ECVC7440, Lot Number 2021092150;
ECVC7445, Lot Number 2021052850;
ECVC7445, Lot Number 2021062250;
ECVC7445A, Lot Number 2022012850;
ECVC7445A, Lot Number 2023010390;
ECVC8045, Lot Number 2023031590;
ECVC8105, Lot Number 2023112190 |
| FEI Number |
1417592
|
Recalling Firm/ Manufacturer |
MEDLINE INDUSTRIES, LP - Northfield 3 Lakes Dr Northfield IL 60093-2753
|
| For Additional Information Contact | Karin Johnson 886-359-1704 |
Manufacturer Reason for Recall | Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health. |
FDA Determined Cause 2 | Nonconforming Material/Component |
| Action | Medline Industries LP issued a recall notice to its consignees on 04/08/2024 via letter delivered via USPS and email. The notice explained the issue, provided a link to for information pertaining to the risk involved https://www.fda.gov/medical-devices/safety-communications/update-evaluating-plastic-syringes-made-china-potential-device-failures-fda-safety-communication) and request these syringes within the convenience be discarded, except for circumstances where no alternative syringes are readily available. If these syringes must be used, closely monitor for leaks, breakage, and any other problems, and report any quality issues to Medline and the FDA. Distributors were directed to notify their customers. The firm issued a second recall notice (an expansion) on 05/02/2024 using the same methods. |
| Quantity in Commerce | 1,679,067 units in total |
| Distribution | Worldwide distribution. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|