 Class 2 Device Recall Medlinesee related information
Date Initiated by FirmApril 08, 2024
Date PostedAugust 28, 2024
Recall Status1 Open3, Classified
Recall NumberZ-2739-2024
Recall Event ID 94581
Product Classification Circuit, breathing (w connector, adaptor, y piece) - Product Code CAI
ProductMedline procedural kits labeled as: 1) ANESTHESIA CIRCUIT PEDS-LF, Pack Number DYNJAA0327F; 2) ANESTHESIA CIRCUIT PEDS-LF, Pack Number DYNJAA0327G
Code Information DYNJAA0327F , Lot Number 20LBV329 ; DYNJAA0327G , Lot Number 21FBM358 ; DYNJAA0327G , Lot Number 22CBY412 ; DYNJAA0327G , Lot Number 22GBZ157 ; DYNJAA0327G , Lot Number 22JBR268 ; DYNJAA0327G , Lot Number 23BBL309 ; DYNJAA0327G , Lot Number 23HBU238 ; DYNJAA0327G , Lot Number 23KBB605 ; DYNJAA0327G , Lot Number 24CBJ211 ;
FEI Number 1417592
Recalling Firm/
Manufacturer
MEDLINE INDUSTRIES, LP - Northfield
3 Lakes Dr
Northfield IL 60093-2753
For Additional Information ContactKarin Johnson
886-359-1704
Manufacturer Reason
for Recall
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.
FDA Determined
Cause 2
Under Investigation by firm
ActionMedline Industries LP issued a recall notice to its consignees on 04/08/2024 via letter delivered via USPS and email. The notice explained the issue, provided a link to for information pertaining to the risk involved https://www.fda.gov/medical-devices/safety-communications/update-evaluating-plastic-syringes-made-china-potential-device-failures-fda-safety-communication) and request the syringes within the convenience be discarded, except for circumstances where no alternative syringes are readily available. If these syringes must be used, closely monitor for leaks, breakage, and any other problems, and report any quality issues to Medline and the FDA. Distributors were directed to notify their customers. The firm issued a second recall notice (an expansion) on 05/02/2024 using the same methods.
DistributionWorldwide
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
-
-