 Class 2 Device Recall Medlinesee related information
Date Initiated by FirmApril 08, 2024
Date PostedSeptember 06, 2024
Recall Status1 Open3, Classified
Recall NumberZ-2988-2024
Recall Event ID 94583
Product Classification Neurological tray - Product Code OJG
ProductMedline Convenience kits, labeled as: 1) NERVE BLOCK TRAY-LF, Pack Number DYNJRA0377B; 2) PERIPHERAL NERVE BLOCK, Pack Number PAIN0137
Code Information DYNJRA0377B, Lot Number 19DDB791 ; DYNJRA0377B, Lot Number 21KDA359 ; PAIN0137, Lot Number 21CLA312; PAIN0137, Lot Number 21ALA608; PAIN0137, Lot Number 20LLA425; PAIN0137, Lot Number 20JLA143; PAIN0137, Lot Number 20HLA191; PAIN0137, Lot Number 20GLA104; PAIN0137, Lot Number 20CLA627; PAIN0137, Lot Number 20BLA634; PAIN0137, Lot Number 20ALA243; PAIN0137, Lot Number 21CLA312; PAIN0137, Lot Number 21ALA608; PAIN0137, Lot Number 20LLA425; PAIN0137, Lot Number 20JLA143; PAIN0137, Lot Number 20HLA191; PAIN0137, Lot Number 20GLA104; PAIN0137, Lot Number 20CLA627; PAIN0137, Lot Number 20BLA634; PAIN0137, Lot Number 20ALA243
Recalling Firm/
Manufacturer
MEDLINE INDUSTRIES, LP - Northfield
3 Lakes Dr
Northfield IL 60093-2753
For Additional Information ContactKarin Johnson
886-359-1704
Manufacturer Reason
for Recall
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.
FDA Determined
Cause 2
Under Investigation by firm
ActionMedline Industries LP issued a recall notice to its consignees on 04/08/2024 via letter delivered via USPS and email. The notice explained the issue, provided a link to for information pertaining to the risk involved https://www.fda.gov/medical-devices/safety-communications/update-evaluating-plastic-syringes-made-china-potential-device-failures-fda-safety-communication) and request these syringes within the convenience be discarded, except for circumstances where no alternative syringes are readily available. If these syringes must be used, closely monitor for leaks, breakage, and any other problems, and report any quality issues to Medline and the FDA. Distributors were directed to notify their customers. The firm issued a second recall notice (an expansion) on 05/02/2024 using the same methods.
DistributionWorldwide distribution.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.