| | Class 2 Device Recall Medline |  |
| Date Initiated by Firm | June 27, 2024 |
| Date Posted | September 11, 2024 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-3113-2024 |
| Recall Event ID |
95128 |
| Product Classification |
Plastic surgery and accessories kit - Product Code FTN
|
| Product | Medline convenience kits containing semi-rigid suction liners labeled as follows:
a) MINOR BREAST, Pack Number DYNJ908104B;
b) PLASTIC MAJOR, Pack Number DYNJ908106B;
c) ABDOMINOPLASTY W/LIPO SUCTION, Pack Number DYNJ909946;
d) PLASTIC, Pack Number DYNJ909951 |
| Code Information |
a) DYNJ908104B: UDI/DI case 40195327507115, UDI/DI each 10195327507114, Lot Numbers: 23LBT869, 24BBQ730;
b) DYNJ908106B: UDI/DI case 40195327507139, UDI/DI each 10195327507138, Lot Numbers: 24ABB087, 24CBM739, 24EBH234;
c) DYNJ909946: UDI/DI case 40195327512652, UDI/DI each 10195327512651, Lot Numbers: 23LBD934, 24DBB806;
d) DYNJ909951: UDI/DI case 40195327512904, UDI/DI each 10195327512903, Lot Numbers: 23LBD932, 24ABP740
|
| FEI Number |
1417592
|
Recalling Firm/ Manufacturer |
MEDLINE INDUSTRIES, LP - Northfield
3 Lakes Dr
Northfield
IL
60093-2753
|
Manufacturer Reason for Recall | Due to reported adverse events, Medline Industries, LP. is issuing a recall on specific SKUs and lots of 1500 cc semi-rigid liners manufactured from 09/2023-11/2023 due to the potential loss of suction or low suction during intermittent suction use. |
FDA Determined Cause 2 | Process control |
| Action | Medline Industries issued a MEDICAL DEVICE RECALL notice to its consignees on 06/27/2024 via US mail and email. The notice explained the problem with the device, potential risk to the patient, and requested the consignee immediately check their stock for the affected item number and the affected lot numbers and destroy any affected inventory. Distributors were directed to notify their customers of the recall and request the affected products be destroyed.
For questions: contact the Recall Department at 866-359-1704 or recalls@medline.com. |
| Quantity in Commerce | 350 units |
| Distribution | Worldwide distribution - US Nationwide and the countries of Canada, Panama. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|