| | Class 1 Device Recall MiniMed 670G, 770G, and 780G Insulin Pumps |  |
| Date Initiated by Firm | July 31, 2024 |
| Date Posted | October 03, 2024 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0003-2025 |
| Recall Event ID |
95121 |
| PMA Number | P160017 |
| Product Classification |
Automated insulin dosing device system, single hormonal control - Product Code OZP
|
| Product | MiniMed Insulin Pump Device. MiniMed 670G, REF: MMT-1580, MMT-1581, MMT-1582, MMT-1780, MMT-1781, MMT-1782, MMT-1740, MMT-1741, MMT-1742, MMT-1760, MMT-1761, MMT-1762;
MiniMed 770G, REF: MMT-1880, MMT-1881, MMT-1882, MMT-1890, MMT-1891, MMT-1892;
MiniMed 780G, REF: MMT-1884, MMT-1885, MMT-1886, MMT-1894, MMT-1895, MMT-1896 |
| Code Information |
REF/UDI-DI(GTIN):
MMT-1780/00763000190453, 00643169946057, 00763000365882, 00763000096984, 00763000283445, 763000283513, 763000283506, 763000365882, 763000283520, 76300028351303, 763000393786, 763000166526, 76300028351301, 643169939202, 763000190460, 763000090203, 763000256821, 643169939219, 763000072537, 763000365912, 00763000090203, 00763000190460, 00763000283520, 00763000393786;
MMT-1781/00763000187415, 00763000283568, 00763000283544, 00763000404536, 00763000283551, 00763000283537, 00763000365899, 00763000371937, 00763000225575, 00763000225582, 00763000235765, 00763000187378, 00763000140601, 763000240707, 763000187378, 763000140601, 763000283544;
MMT-1782/00763000187422, 00763000235772, 00763000283599, 00763000264727, 00763000404543, 00763000283582, 00763000371944, 00763000365905, 00763000283575, 00643169957640, 00763000157463, 00763000225612, 763000225629, 763000225612, 763000264727;
MMT-1741/00763000408916, 00763000408930;
MMT-1742/00763000408947, 00763000250232, 00763000408954;
MMT-1760/763000278427, 763000283445, 643169946057, 763000190453, 763000096984, 643169946040, 00643169946057, 00763000190453, 00763000283445, 763000367084, 763000179649, 763000395858, 00763000096984;
MMT-1761/763000187392, 00763000165451, 00763000165420, 00763000165468, 00763000165444, 00763000165437, 00763000257309, 00763000165413, 00763000289065, 00763000289089, 00763000289041, 00763000367657, 00763000289195, 00763000289096, 00763000289072, 00763000289058, 00763000324797, 00763000374518, 00763000374532, 00763000374525, 00763000374556, 00763000399849, 00763000399856, 00763000374587, 763000367657;
MMT-1762/763000264710, 00763000264710, 00763000165475, 00763000165482, 00763000165499, 00763000165505, 00763000165512, 00763000165529, 00763000165536, 00763000165543, 00763000165550, 00763000203313, 00763000216481, 00763000252076, 00763000283476, 00763000289102, 00763000289119, 00763000289126, 00763000289133, 00763000289140, 00763000289164, 00763000289171, 00763000289188, 00763000324780, 00763000374594, 00763000374600, 00763000374624, 00763000399986;
MMT-1880/00763000414344, 00763000545604, 00763000439859, 763000578640, 763000586195, 76300058619502, 763000439866, 763000413750, 76300057864002, 763000414344, 763000545604, 763000640095, 763000439859, 76300058619501, 00763000413750, 00763000586188;
MMT-1881/00763000416546, 00763000439903, 00763000629892, 763000416546;
MMT-1882/00763000416553, 00763000439910, 00763000545505, 00763000629908;
MMT-1890/763000545604, 763000414344, 763000439859, 763000640095, 00763000414344, 00763000439859, 00763000545604, 76300054560402, 00763000640095, 76300064009502;
MMT-1891/00763000642242, 00763000456634, 00763000642235, 00763000418151, 00763000482596;
MMT-1892/00763000416553, 00763000439910, 00763000545505, 00763000629908;
MMT-1884/00763000414344, 76300081937801, 76300081938501, 763000819378, 763000819385, 00763000586195, A7630007585401, 00763000758547, 00763000439866, 00763000578640, 00763000439859, 00763000545604, 00763000640095, 00763000819392, 763000545604, 00763000758530, 00763000504762;
MMT-1885/00763000384258, 00763000521523, 00763000442187, 00763000521509, 00763000578671, 00763000639228, 00763000421939, 00763000597900, 00763000854942, 00763000854829, A7630006392201, 763000384272, 763000521523, 763000854829;
MMT-1886/00763000384265, 00763000441449, 00763000521530, 00763000521516, 00763000578688, 00763000442194, 00763000639235, 763000384289, 763000521530, 76300044240801, 763000688165, 76300063923501;
MMT-1894/76300086566501, 76300075854701, 763000758547, 763000819392, 763000865665, 00763000758547, 00763000819392;
MMT-1895/00763000411411, 00763000734435, 00763000411473, 00763000411466, 00763000541569, 00763000411480, 00763000420703, 00763000411428, 00763000596217, 00763000411404, 00763000533861, 00763000411381, 00763000411503, 00763000596187, 00763000411398, 00763000596194, 00763000411374, 00763000411442, 00763000411435, 00763000411497, 00763000411459, 00763000596224, 00763000411633, 00763000596231, 00763000639266, 00763000384272, 00763000521547, 00763000732585, 763000734435;
MMT-1896/763000521554,00763000411510, 00763000411565, 00763000411640, 00763000411541, 00763000411589, 00763000734442, 00763000411527, 00763000411626, 00763000411602, 00763000411558, 00763000523305, 00763000411572, 00763000411596, 00763000914127, 00763000411619, 00763000411534, 00763000734459, 00763000544317, 00763000913250, 00763000533878, 00763000521554, 00763000384289, 00763000639273, 763000734442;
All serial numbers. |
| FEI Number |
3003166194
|
Recalling Firm/ Manufacturer |
Medtronic MiniMed, Inc.
18000 Devonshire St
Northridge
CA
91325-1219
|
| For Additional Information Contact | Medtronic 24-Hr Technical Support 800-646-4633 |
Manufacturer Reason for Recall | Insulin pumps that have been dropped, bumped, or experienced physical impact may have damage to internal electrical components, which may cause reduced pump battery life. Since the defect reduces the battery life overall, it shortens the time between low battery alarms and increases the frequency with which the user must insert a new battery. Further, the defect may also deliver the low battery alarm when there is significantly less time before the battery completely runs out, from the expected up to 10-hour buffer the low alert is supposed to signal, leading to device powering down and stopping insulin delivery unexpectedly. Per the firm, a subset of pumps with the defect showed the low alert alarm was delivered ~2.5 hours before the device shut off, however, patient reports indicate the time could be even shorter. Replacing the battery will not resolve this issue and the short battery life and the truncated low battery alert lead time will continue to occur. Early battery depletion could result in power loss and insulin delivery could be interrupted, resulting in under delivery of insulin potentially leading to hyperglycemia and/or DKA. |
FDA Determined Cause 2 | Device Design |
| Action | On 7/31/24, Safety Alert notices were sent to customers, healthcare professionals, distributors who were told the following:
1) Always pay attention to alerts and alarms displayed on your pump.
- Low Battery Pump alert will display when your pump has up to 10 hours of battery life left.
- Replace Battery alert will display when the pump has less than 30 minutes of battery life left. If the battery is not replaced within 10 minutes, a siren will sound and repeat once every minute until the Replace Battery Now alarm is displayed. At that time, the pump will stop insulin delivery. If the battery is not replaced within 10 minutes of this alarm, a siren will sound, and the pump will shut down.
2) If your pump has been dropped, bumped, or has experienced physical impact, pay special attention to any pump alerts and alarms, including low battery alerts, as they may occur earlier than expected.
3) Pump's battery level is on the status bar located on the pump home screen.
4) Refer to the pump user guide for instructions on how to replace the battery.
5) If you notice any significant changes in battery life or need additional troubleshooting assistance, the firm's Technical Support line for further help.
6) Always have extra new AA lithium or alkaline batteries or fully charged NiMH
batteries available, along with your other emergency kit supplies.
7) Complete and return the confirmation form by visiting https://info.medtronicdiabetes.com/battery-statusacknowledgement
In addition, distribution partners and service providers were provided with a patient letter they could use to inform customers/pump users.
If you have further questions or need assistance, please call the firm's 24-Hour Technical Support line at 800-646-4633, option 1.
|
| Quantity in Commerce | 526,558 |
| Distribution | US: CT, MI, PA, WA, IA, NY, ND, AZ, TX, OH, NC, AL, MN, IN, NJ, KY, UT, CA, FL, VA, MS, NM, NV, TN, GA, MA, NH, OK, VT, IL, ME, SC, LA, WY, RI, SD, KS, WI, MD, CO, DE, AR, AK, ID, MO, NE, WV, MT, OR, DC, HI, VI, PR.
OUS: Worldwide
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| PMA Database | PMAs with Product Code = OZP
|
|
|
|