| Date Initiated by Firm | September 04, 2024 |
| Date Posted | October 23, 2024 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0150-2025 |
| Recall Event ID |
95415 |
| Product Classification |
Tray, surgical - Product Code LRP
|
| Product | Centurion brand, medical procedure convenience kits, labeled as:
a) FLOOR - UMBILICAL CUT DOWN TRA, b) PLASTICS SUTURE TRAY |
| Code Information |
a) UDI/DI 40653160270929 (each), 10653160270928 (case), Lot Numbers: 23LBO858;
b) UDI/DI 40653160284346 (each), 10653160284345 (case), Lot Numbers: 23LBL653
|
| FEI Number |
1417592
|
Recalling Firm/ Manufacturer |
MEDLINE INDUSTRIES, LP - Northfield 3 Lakes Dr Northfield IL 60093-2753
|
| For Additional Information Contact | Karin Johnson 866-359-1704 |
Manufacturer Reason for Recall | It has been determined that certain Convenience kits that were packaged in breather pouch lot 323080002 may result in the potential of using non-sterile product if the vendor seal of the pouch is opened/unsealed. |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | Medline issued a MEDICAL DEVICE RECALL notice to its consignees on 09/05/2024 via email and USPS. The notice explained the issue, hazard, and requested the destruction of the affected product. Distributors were directed to notify their consignees. |
| Quantity in Commerce | 168 units |
| Distribution | Worldwide distribution - US Nationwide and the countries of Canada, Panama. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|