• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 1 Device Recall GE Healthcare Incubator Carestation CS1

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 1 Device Recall GE Healthcare Incubator Carestation CS1see related information
Date Initiated by FirmJune 06, 2023
Date PostedOctober 31, 2024
Recall Status1 Completed
Recall NumberZ-0141-2025
Recall Event ID 95456
510(K)NumberK152809 K213553 
Product Classification Incubator, neonatal - Product Code FMZ
ProductGE Healthcare Giraffe Incubator Carestation CS1, Infant Incubator. Model Numbers: 1) 2082844-002-01040417; 2) 2082844-002-01044956; 3) 2082844-002-01058424; 4) 2082844-002-01069017; 5) 2082844-002-01089666; 6) 2082844-002-01090474; 7) 2082844-002-01091269; 8) 2082844-002-01093489; 9) 2082844-002-01113262; 10) 2082844-002-01113711; 11) 2082844-002-01123085; 12) 2082844-002-01123918; 13) 2082844-002-01128966; 14) 2082844-002-01133042; 15) 2082844-002-01146447; 16) 2082844-002-100205; 17) 2082844-002-100749; 18) 2082844-002-100752; 19) 2082844-002-100995; 20) 2082844-002-101425; 21) 2082844-002-523548; 22) 2082844-002-529928; 23) 2082844-002-594800; 24) 2082844-002-643544; 25) 2082844-002-925841; 26) 2082844-002-965540; 27) 2082844-002-982916; 28) 2082844-002-983420; 29) 2082844-002-993905; 30) 2082844-002-995018; 31) 2082844-002-995799; 32) 2082844-002-998943; Radiant infant warmer
Code Information 1) 2082844-002-01040417, UDI: (01)00840682116855(21)TAAC70003(11)230106, (01)00840682116855(21)TAAC70004(11)230109, (01)00840682116855(21)TAAC70011(11)230110, (01)00840682116855(21)TAAC70012(11)230110, (01)00840682116855(21)TAAC70013(11)230111, (01)00840682116855(21)TAAC70015(11)230112, (01)00840682116855(21)TAAC70016(11)230112, (01)00840682116855(21)TAAC70020(11)230112, (01)00840682116855(21)TAAC70022(11)230112, (01)00840682116855(21)TAAC70024(11)230112, (01)00840682116855(21)TAAC70147(11)230317, (01)00840682116855(21)TAAC70148(11)230317, (01)00840682116855(21)TAAC70149(11)230317, (01)00840682116855(21)TAAC70150(11)230317, (01)00840682116855(21)TAAC70152(11)230320, (01)00840682116855(21)TAAC70153(11)230320, (01)00840682116855(21)TAAC70155(11)230322, (01)00840682116855(21)TAAC70158(11)230322, (01)00840682116855(21)TAAC70159(11)230323, (01)00840682116855(21)TAAC70160(11)230323, (01)00840682116855(21)TAAC70161(11)230323, (01)00840682116855(21)TAAC70162(11)230323, (01)00840682116855(21)TAAC70163(11)230324, (01)00840682116855(21)TAAC70164(11)230324, (01)00840682116855(21)TAAC70165(11)230324, (01)00840682116855(21)TAAC70166(11)230324, (01)00840682116855(21)TAAC70167(11)230327, (01)00840682116855(21)TAAC70169(11)230327, (01)00840682116855(21)TAAC70170(11)230327, (01)00840682116855(21)TAAC70171(11)230328, (01)00840682116855(21)TAAC70177(11)230330, (01)00840682116855(21)TAAC70178(11)230330, (01)00840682116855(21)TAAC70179(11)230331, (01)00840682116855(21)TAAC70181(11)230331, (01)00840682116855(21)TAAC70182(11)230331, (01)00840682116855(21)TAAC70184(11)230331, (01)00840682116855(21)TAAC70192(11)230405, (01)00840682116855(21)TAAC70193(11)230405, (01)00840682116855(21)TAAC70196(11)230405, (01)00840682116855(21)TAAC70197(11)230405, (01)00840682116855(21)TAAC70200(11)230407, (01)00840682116855(21)TAAC70201(11)230407, (01)00840682116855(21)TAAC70202(11)230407, (01)00840682116855(21)TAAC70203(11)230407, (01)00840682116855(21)TAAC70204(11)230407, (01)00840682116855(21)TAAC70205(11)230407, (01)00840682116855(21)TAAC70208(11)230411, (01)00840682116855(21)TAAC70209(11)230411, (01)00840682116855(21)TAAC70210(11)230411, (01)00840682116855(21)TAAC70211(11)230414, (01)00840682116855(21)TAAC70212(11)230414, (01)00840682116855(21)TAAC70213(11)230414, (01)00840682116855(21)TAAC70214(11)230414, (01)00840682116855(21)TAAC70215(11)230417, (01)00840682116855(21)TAAC70216(11)230417, (01)00840682116855(21)TAAC70217(11)230417, (01)00840682116855(21)TAAC70218(11)230417, (01)00840682116855(21)TAAC70223(11)230418, (01)00840682116855(21)TAAC70224(11)230418, (01)00840682116855(21)TAAC70225(11)230418, (01)00840682116855(21)TAAC70226(11)230418, (01)00840682116855(21)TAAC70227(11)230418, (01)00840682116855(21)TAAC70228(11)230418, (01)00840682116855(21)TAAC70229(11)230419, (01)00840682116855(21)TAAC70242(11)230421, (01)00840682116855(21)TAAC70243(11)230421, (01)00840682116855(21)TAAC70244(11)230421, (01)00840682116855(21)TAAC70245(11)230421, (01)00840682116855(21)TAAC70246(11)230421, (01)00840682116855(21)TAAC70247(11)230421; 2) 2082844-002-01044956, UDI: (01)008406821168552(21)TAA23140007SA(11)230404, (01)008406821168552(21)TAA23140008SA(11)230404; 3) 2082844-002-01058424, UDI: (01)00840682116855(21)TAAC70142(11)230213; 4) 2082844-002-01069017, UDI: (01)008406821168552(21)TAA23110001SA(11)230316, (01)008406821168552(21)TAA23110002SA(11)230316, (01)008406821168552(21)TAA23110003SA(11)230316; 5) 2082844-002-01089666, UDI: (01)00840682116855(21)TAAC70124(11)230201; 6) 2082844-002-01090474, UDI: (01)00840682116855(21)TAAC70126(11)230201, (01)00840682116855(21)TAAC70128(11)230206, (01)00840682116862(21)TAAC70132(11)230206, (01)00840682116855(21)TAAC70133(11)230206, (01)00840682116855(21)TAAC70125(11)230201, (01)00840682116855(21)TAAC70127(11)230206, (01)00840682116855(21)TAAC70129(11)230206, (01)00840682116855(21)TAAC70131(11)230206, (01)00840682116855(21)TAAC70136(11)230206, (01)00840682116855(21)TAAC70137(11)230206; 7) 2082844-002-01091269, UDI: (01)008406821168552(21)TAA23090001SA(11)230228 228; 8) 2082844-002-01093489, UDI: (01)00840682116855(21)TAAC70187(11)230403, (01)00840682116855(21)TAAC70188(11)230403, (01)00840682116855(21)TAAC70189(11)230403, (01)00840682116855(21)TAAC70190(11)230403, (01)00840682116855(21)TAAC70191(11)230403; 9) 2082844-002-01113262, UDI: (01)008406821168552(21)TAA23050010SA(11)230130, (01)008406821168552(21)TAA23050011SA(11)230130, (01)008406821168552(21)TAA23050012SA(11)230130, (01)008406821168552(21)TAA23050013SA(11)230130; 10) 2082844-002-01113711, UDI: (01)008406821168552(21)TAA23130005SA(11)230328, (01)008406821168552(21)TAA23130007SA(11)230329, (01)008406821168552(21)TAA23130008SA(11)230329, (01)008406821168552(21)TAA23130009SA(11)230329; 11) 2082844-002-01123085, UDI: (01)008406821168552(21)TAA23190005SA(11)230513, (01)008406821168552(21)TAA23190006SA(11)230513, (01)008406821168552(21)TAA23190007SA(11)230513, (01)008406821168552(21)TAA23190008SA(11)230513, (01)008406821168552(21)TAA23190009SA(11)230513, (01)008406821168552(21)TAA23190010SA(11)230513; 12) 2082844-002-01123918, UDI: (01)008406821168552(21)TAA23130001SA(11)230328; 13) 2082844-002-01128966, UDI: (01)008406821168552(21)TAA23040022SA(11)230127; 14) 2082844-002-01133042, UDI: (01)008406821168552(21)TAA23130003SA(11)230328, (01)008406821168552(21)TAA23130004SA(11)230328, (01)008406821168552(21)TAA23130006SA(11)230329; 15) 2082844-002-01146447, UDI: (01)008406821168552(21)TAA23100001SA(11)230307; 16) 2082844-002-100205, UDI: (01)008406821168552(21)TAA23030030SA(11)230121; 17) 2082844-002-100749, UDI: (01)00840682116855(21)TAAC70272(11)230426; 18) 2082844-002-100752, UDI: (01)00840682116855(21)TAAC70075(11)230123, (01)00840682116855(21)TAAC70083(11)230124, (01)00840682116855(21)TAAC70084(11)230124, (01)00840682116855(21)TAAC70086(11)230125, (01)00840682116855(21)TAAC70087(11)230125; 19) 2082844-002-100995, UDI: (01)008406821168552(21)TAA23090003SA(11)230228, (01)008406821168552(21)TAA23090004SA(11)230228, (01)008406821168552(21)TAA23090005SA(11)230228, (01)008406821168552(21)TAA23090006SA(11)230228, (01)008406821168552(21)TAA23090002SA(11)230228, (01)008406821168552(21)TAA23090007SA(11)230228; 20) 2082844-002-101425, UDI: (01)008406821168552(21)TAA23190003SA(11)230512, (01)008406821168552(21)TAA23160001SA(11)230419, (01)008406821168552(21)TAA23170001SA(11)230424, (01)008406821168552(21)TAA23130031SA(11)230331, (01)008406821168552(21)TAA23130032SA(11)230331, (01)008406821168552(21)TAA23130033SA(11)230331, (01)008406821168552(21)TAA23130034SA(11)230331, (01)008406821168552(21)TAA23130035SA(11)230331, (01)008406821168552(21)TAA23130036SA(11)230331, (01)008406821168552(21)TAA23130021SA(11)230330, (01)00840682116855(21)TAAC70143(11)230224, (01)00840682116855(21)TAAC70144(11)230224, (01)00840682116855(21)TAAC70145(11)230224, (01)008406821168552(21)TAA23130025SA(11)230331, (01)008406821168552(21)TAA23130026SA(11)230331, (01)008406821168552(21)TAA23130027SA(11)230331, (01)008406821168552(21)TAA23130028SA(11)230331, (01)008406821168552(21)TAA23130029SA(11)230331, (01)008406821168552(21)TAA23130030SA(11)230331, (01)00840682116855(21)TAAC70146(11)230313, (01)008406821168552(21)TAA23140001SA(11)230403, (01)008406821168552(21)TAA23140002SA(11)230403, (01)008406821168552(21)TAA23130022SA(11)230330, (01)008406821168552(21)TAA23130023SA(11)230330, (01)008406821168552(21)TAA23130024SA(11)230330, (01)008406821168552(21)TAA23160002SA(11)230419; 21) 2082844-002-523548, UDI: (01)00840682116855(21)TAAW62043(11)170130; (01)00840682116855(21)TAAZ70271(11)200224; 22) 2082844-002-529928, UDI: (01)008406821168552(21)TAA23130002SA(11)230328; 23) 2082844-002-594800, UDI: (01)00840682116855(21)TAAB70352(11)220520; 24) 2082844-002-643544, UDI: (01)008406821168552(21)TAA23030023SA(11)230120, (01)008406821168552(21)TAA23030024SA(11)230120, (01)008406821168552(21)TAA23030025SA(11)230120; 25) 2082844-002-925841, UDI: (01)00840682116855(21)TAAB70547(11)220815; 26) 2082844-002-965540, UDI: (01)00840682116855(21)TAAB70617(11)220826, (01)00840682116855(21)TAAB70665(11)220928, (01)00840682116855(21)TAAB70666(11)220928, (01)00840682116855(21)TAAB70670(11)220929, (01)00840682116855(21)TAAB70671(11)220929, (01)00840682116855(21)TAAB70681(11)221003, (01)00840682116855(21)TAAB70691(11)221004, (01)00840682116855(21)TAAB70702(11)221005, (01)00840682116855(21)TAAB70703(11)221005, (01)00840682116855(21)TAAB70732(11)221010; 27) 2082844-002-982916, UDI: (01)00840682116855(21)TAAC70033(11)230113, (01)00840682116855(21)TAAC70037(11)230116, (01)00840682116855(21)TAAC70044(11)230116, (01)00840682116855(21)TAAC70052(11)230119, (01)00840682116855(21)TAAC70058(11)230119, (01)00840682116855(21)TAAC70059(11)230119, (01)00840682116855(21)TAAC70060(11)230119, (01)00840682116855(21)TAAC70065(11)230120, (01)00840682116855(21)TAAC70067(11)230123, (01)00840682116855(21)TAAC70068(11)230123, (01)00840682116855(21)TAAB70608(11)220824; 28) 2082844-002-983420, UDI: (01)00840682116855(21)TAAC70076(11)230123, (01)00840682116855(21)TAAC70078(11)230124, (01)00840682116855(21)TAAC70079(11)230124, (01)00840682116855(21)TAAC70082(11)230124, (01)00840682116855(21)TAAC70089(11)230125, (01)00840682116855(21)TAAC70090(11)230125, (01)00840682116855(21)TAAC70092(11)230126, (01)00840682116855(21)TAAC70093(11)230126; 29) 2082844-002-993905, UDI: (01)008406821168552(21)TAA23040004SA(11)230123, (01)008406821168552(21)TAA23040005SA(11)230123, (01)008406821168552(21)TAA23040006SA(11)230124, (01)008406821168552(21)TAA23050014SA(11)230130; 30) 2082844-002-995018, UDI: (01)00840682116855(21)TAAB70551(11)220815, (01)00840682116855(21)TAAB70558(11)220819; 31) 2082844-002-995799, UDI: (01)008406821168552(21)TAA23040019SA(11)230127, (01)008406821168552(21)TAA23040020SA(11)230127, (01)008406821168552(21)TAA23190004SA(11)230513, (01)008406821168552(21)TAA23160007SA(11)230420, (01)008406821168552(21)TAA23160009SA(11)230420, (01)008406821168552(21)TAA23130015SA(11)230329, (01)008406821168552(21)TAA23130016SA(11)230329, (01)008406821168552(21)TAA23130017SA(11)230329, (01)008406821168552(21)TAA23130018SA(11)230329, (01)008406821168552(21)TAA23130020SA(11)230329, (01)008406821168552(21)TAA23160003SA(11)230419, (01)008406821168552(21)TAA23160004SA(11)230419, (01)008406821168552(21)TAA23160005SA(11)230419, (01)008406821168552(21)TAA23160006SA(11)230419, (01)008406821168552(21)TAA23160008SA(11)230420, (01)008406821168552(21)TAA23160010SA(11)230420, (01)008406821168552(21)TAA23160011SA(11)230420, (01)008406821168552(21)TAA23200003SA(11)230515, (01)008406821168552(21)TAA23200004SA(11)230515, (01)008406821168552(21)TAA23200005SA(11)230515, (01)008406821168552(21)TAA23200006SA(11)230515, (01)008406821168552(21)TAA23200007SA(11)230515, (01)008406821168552(21)TAA23200008SA(11)230515, (01)008406821168552(21)TAA23200009SA(11)230515, (01)008406821168552(21)TAA23200010SA(11)230515, (01)008406821168552(21)TAA23130019SA(11)230329; 32) 2082844-002-998943, UDI: (01)00840682116855(21)TAAC70050(11)230119, (01)00840682116855(21)TAAC70053(11)230119, (01)00840682116855(21)TAAC70110(11)230130
FEI Number 3014220748
Recalling Firm/
Manufacturer
DATEX--OHMEDA, INC.
9900 W Innovation Dr
Wauwatosa WI 53226-4856
For Additional Information ContactGE HealthCare Service
800-437-1171
Manufacturer Reason
for Recall
Prior to current established manufacturing practices, higher levels of airborne formaldehyde exposure from newly manufactured incubators could be seen during clinical use that would rapidly decrease over a week. Affected customers were notified in June 2023 to run incubators for a week to potentially off gas any residual airborne formaldehyde. The general scientific consensus is that sensory irritation of the eyes is the most sensitive endpoint with respect to exposure to airborne formaldehyde and is the adverse effect clinically experienced at the lowest concentration.
FDA Determined
Cause 2
Component design/selection
ActionGE HealthCare notified its consignees of the issue by letter beginning on 06/06/2023 using a traceable means. The notice explained the issue of potentially higher levels of formaldehyde which rapidly decrease of a week, and instructions on how to accelerate the rate of dissipation if the unit had not yet been put into clinical use. For questions: contact 1-800-437-1171 or your local GE HealthCare Representative
Quantity in Commerce224 units
Distributionworldwide
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
510(K) Database510(K)s with Product Code = FMZ
510(K)s with Product Code = FMZ
-
-