• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Medline Catheter connector

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Medline Catheter connectorsee related information
Date Initiated by FirmOctober 16, 2024
Date PostedNovember 22, 2024
Recall Status1 Open3, Classified
Recall NumberZ-0509-2025
Recall Event ID 95615
Product Classification Spinal epidural anesthesia kit - Product Code OFT
ProductMEDLINE procedure kits labeled as follows: 1) COMBINED SPINAL AND EPIDURAL, REF DYNJRA1268B; 2) COMBINED SPINAL AND EPIDURAL, REF DYNJRA1268C; 3) FEMORAL BLOCK TRAY, REF DYNJRA1739A; 4) FEMORAL BLOCK TRAY, REF DYNJRA1739B; 5) 18G CPNB 20G STYLETED CATHETER, REF DYNJRA1869; 6) CONT EPID 18G NDL/20G CATH, REF DYNJRA9035; 7) TRAY 18G CPNB 20G STYLETED, REF DYNJRA9040; 8) CATHETER REINF W/ CONNECT OT, REF EPIREINF19O; 9) CATHETER REINF WITH CONNECTOR, REF EPIREINF29C; 10) CONTINUOUS EPIDURAL UPDATED, REF PAIN0384B; 11) CONT EPI 20G W/HUSTEAD, REF PAIN1069C; 12) CONTINUOUS EPIDURAL, REF PAIN1271A; 13) TRAY, CONT EPI 20G NYL W/ HUST, REF PAIN1281B; 14) CONTINUOUS EPIDURAL TRAY, REF PAIN1317; 15) 20G CONT. EPID TRAY, REF PAIN1353B; 16) CONTINUOUS EPIDURAL TRAY, REF PAIN1358; 17) TRAY, CONT EPI 20G NYL W/ HUST, REF PAIN1361; 18) JHAC CONT EPIDURAL, REF PAIN1437A; 19) CONTINUOUS EPIDURAL TRAY, REF PAIN1453B; 20) EPIDURAL CATH KIT, REF PAIN1536A; 21) CONTINUOUS EPIDURAL TRAY, REF PAIN1547; 22) EPIDURAL TRAY, REF PAIN1558; 23) CONTINUOUS EPIDURAL, REF PAIN1581A; 24) CONTINUOUS EPIDURAL TRAY, REF PAIN1587A; 25) CONTINUOUS EPIDURAL TRAY, REF PAIN1587B; 26) CONTINUOUS EPIDURAL TRAY, REF PAIN1605A; 27) CONTINUOUS EPIDURAL TRAY, REF PAIN1636; 28) EPIDURAL CATHETERIZATION KIT, REF PAIN1644A; 29) CONTINUOUS EPIDURAL TRAY, REF PAIN1667A; 30) CONT. EPIDURAL TRAY - NYLON, REF PAIN1709; 31) CONT. EPIDURAL TRAY - NYLON, REF PAIN1709A; 32) CONTINUOUS EPIDURAL, REF PAIN1766A; 33) PEDIATRIC EPIDURAL TRAY, REF PAIN1806; 34) EPIDURAL TRAY IMF 103130, REF PAIN1807; 35) CONTINUOUS EPIDURAL TRAY, REF PAIN1816; 36) CONTINUOUS EPIDURAL, REF PAIN1844; 37) CSE TRAY, REF PAIN1845; 38) BWH L&D EPIDURAL KIT, REF PAIN1846; 39) 17G CONT. EPI - MRI SAFE, REF PAIN1854; 40) 17G L&D CONT. EPIDURAL, REF PAIN1855; 41) CONTINUOUS/CSE TRAY, REF PAIN1856; 42) CONTINUOUS/CSE TRAY, REF PAIN1856A; 43) CONTINUOUS EPIDURAL TRAY, REF PAIN1864; 44) CONTINUOUS EPIDURAL TRAY, REF PAIN1864A; 45) CONTINUOUS EPIDURAL TRAY, REF PAIN1878; 46) CSE TRAY, REF PAIN1885; 47) TEAM HEALTH EPIDURAL, REF PAIN1887; 48) SOUND ANESTHESIA EPIDURAL, REF PAIN1888; 49) CONTINUOUS EPIDURAL TRAY, REF PAIN1890; 50) CONTINUOUS EPIDURAL TRAY, REF PAIN1892; 51) UAMS CONTINUOUS EPIDURAL, REF PAIN1893; 52) TRAY,CON,EPID,19G REINF,W/TUOH, REF PAIN1894; 53) CSE TRAY, REF PAIN1898; 54) CSE TRAY, REF PAIN1904; 55) 17/25G CSE TRAY, REF PAIN1919; 56) CSE TRAY, REF PAIN1920; 57) 17/25G CSE TRAY, REF PAIN1937; 58) CONTINUOUS EPIDURAL TRAY, REF PAIN1940; 59) COMBINED SPINAL EPIDURAL TRAY, REF PAIN1958; 60) REINFORCED CATH EPIDURAL TRAY, REF PAIN1967; 61) OPEN TIP EPIDURAL CATH TRAY, REF PAIN2014; 62) TRAY,CON,EPID,19G REINF,W/TUOH, REF PAIN9025; 63) TRAY,CON,EPID,19G REINF,W/O RX, REF PAIN9026; 64) CSE 17G HUSTEAD 27G WHIT W/RX, REF PAIN9028; 65) 17G CONT EPIDURAL NYLON, REF PAIN9030; 66) TRAY,CON EPID,19G REINF OPNTIP, REF PAIN9031; 67) 19G REINF CATH LIDO/NACL ONLY, REF PAIN9034; 68) 20G NYLON CATH LIDO/NACL ONLY, REF PAIN9035; 69) 17GHUST 27GWHIT LIDO/NACL ONLY, REF PAIN9036; 70) REINIF 19G EPI CATH S STERILE, REF REPICATH19; 71) CONT. EPIDURAL TRAY, REF SAMPA0105; 72) EPIDURAL CATH PACK, REF SAMPA0106; 73) CONTINUOUS EPIDURAL TRAY, REF SAMPA0110; 74) CONTINUOUS EPIDURAL TRAY, REF SAMPA0111
Code Information Codes: 1) DYNJRA1268B, UDI/DI 10193489370485 (each) 40193489370486 (case), Lot Numbers: 20FLA512, 20HLA422, 20ILA296, 20ILA327, 20LLA223, 21ALA087, 21CLA525; 2) DYNJRA1268C, UDI/DI 10193489905830 (each) 40193489905831 (case), Lot Numbers: 21ELA224, 21ELA333, 21GLA292, 21GLA895, 21HLA924, 21ILA773, 21JLA824, 22ALA237, 22BLA899, 22DLA197, 22DLA409, 22ELA583, 22ELB120, 22FLA820, 22FLB163, 22GLB081, 22HLA566, 22ILA092, 22LLA800, 23DLA368, 23FLA942, 23GLA321, 23HLA910, 23ILA593, 23JLA268; 3) DYNJRA1739A, UDI/DI 1019348992495 (each) 40193489924955 (case), Lot Numbers: 23CBE392; 4) DYNJRA1739B, UDI/DI 10195327411237 (each) 40195327411238 (case), Lot Numbers: 23FBL394, 23GBV201, 23HBG282, 23HBG968, 23IBQ191; 5) DYNJRA1869, UDI/DI 10195327018818 (each) 40195327018819 (case), Lot Numbers: 23CBE391; 6) DYNJRA9035, UDI/DI 10884389618448 (each) 40884389618449 (case), Lot Numbers: 22GBO435, 22IBI073, 23BBB167, 23BBS677, 23CBL304, 23FBI382; 7) DYNJRA9040, UDI/DI 10193489830514 (each) 40193489830515 (case), Lot Numbers: 23CBO446, 23DBK742, 23HBG091, 23HBG092; 8) EPIREINF19O, UDI/DI 10195327222970 (each) 40195327222971 (case), Lot Numbers: 22HBB281, 23EBO266, 23GBA230, 23HBN806, 23HBR131, 23HBY420; 9) EPIREINF29C, UDI/DI 10195327114367 (each) 40195327114368 (case), Lot Numbers: 22FBU886, 22HBD608, 22KBK655, 22NBE738, 23ABH589, 23DBK772, 23DBN448, 23JBU253; 10) PAIN0384B, UDI/DI 10193489926842 (each) 40193489926843 (case), Lot Numbers: 23BMC466, 23FMJ516, 23FMJ523, 23FMJ721, 23HMA617, 23LMB080, 23LMC430; 11) PAIN1069C, UDI/DI 1019532736489 (each) 40195327364893 (case), Lot Numbers: 23ELA637, 23GLA586, 23HLA616, 23JLB208; 12) PAIN1271A, UDI/DI 10193489619201 (each) 40193489619202 (case), Lot Numbers: 23BLA979; 13) PAIN1281B, UDI/DI 10195327288211 (each) 40195327288212 (case), Lot Numbers: 22KLA735, 23ALA157; 14) PAIN1317, UDI/DI 10888277797109 (each) 40888277797100 (case), Lot Numbers: 22GLA179, 22HLB332, 23CLA460, 23DLA279, 23ELB138, 23FLA944, 23GLA948, 23HLB106, 23ILA924, 23JLB213, 23KLB167; 15) PAIN1353B, UDI/DI 10193489336726 (each) 40193489336727 (case), Lot Numbers: 20ELA452; 16) PAIN1358, UDI/DI 10193489586749 (each) 40193489586740 (case), Lot Numbers: 22FLB088, 23CLA573; 17) PAIN1361, UDI/DI 10193489586848 (each) 40193489586849 (case), Lot Numbers: 22FLA642, 22GLA243, 23ILA612, 23JLA051; 18) PAIN1437A, UDI/DI 1019532713519 (each) 40195327135196 (case), Lot Numbers: 22EBA985, 23ABO995, 23KBH798, 23KBR650; 19) PAIN1453B, UDI/DI 1019532731038 (each) 40195327310388 (case), Lot Numbers: 22LLB007, 23ALA593, 23BLA987; 20) PAIN1536A, UDI/DI 10193489288582 (each) 40193489288583 (case), Lot Numbers: 23CLA081; 21) PAIN1547, UDI/DI 10193489813081 (each) 40193489813082 (case), Lot Numbers: 23BLB100, 23FLA069, 23JLA884; 22) PAIN1558, UDI/DI 10193489213584 (each) 40193489213585 (case), Lot Numbers: 23BMH883, 23GMA342, 23GMC058, 23IMC928; 23) PAIN1581A, UDI/DI 10195327351809 (each) 40195327351800 (case), Lot Numbers: 23BLB024, 23CLA024, 23ELB140; 24) PAIN1587A, UDI/DI 10193489336368 (each) 40193489336369 (case), Lot Numbers: 23BLA078, 23FLA277; 25) PAIN1587B, UDI/DI 10195327539511 (each) 40195327539512 (case), Lot Numbers: 23JLB184; 26) PAIN1605A, UDI/DI 1019348946814 (each) 40193489468145 (case), Lot Numbers: 22KLA748, 23BLA153; 27) PAIN1636, UDI/DI 10193489312720 (each) 40193489312721 (case), Lot Numbers: 22FLA297; 28) PAIN1644A, UDI/DI 10195327427863 (each) 40195327427864 (case), Lot Numbers: 23GLA589, 23HLA915, 23JLA225, 23KLA831; 29) PAIN1667A, UDI/DI 10193489472868 (each) 40193489472869 (case), Lot Numbers: 22FLB089, 23CLA442, 23ELA958, 23FLA945, 23HLA916, 23KLB169; 30) PAIN1709, UDI/DI 10193489816044 (each) 40193489816045 (case), Lot Numbers: 23CLA719; 31) PAIN1709A, UDI/DI 10195327411220 (each) 40195327411221 (case), Lot Numbers: 23ELA723, 23ILA927; 32) PAIN1766A, UDI/DI 10195327361488 (each) 40195327361489 (case), Lot Numbers: 23FLB123, 23GLA313; 33) PAIN1806, UDI/DI 10195327134563 (each) 40195327134564 (case), Lot Numbers: 22ELA290, 22FLB111; 34) PAIN1807, UDI/DI 1019532713502 (each) 40195327135028 (case), Lot Numbers: 22FLB112, 23DLA445; 35) PAIN1816, UDI/DI 10195327145453 (each) 40195327145454 (case), Lot Numbers: 22GLA509, 22GLA531; 36) PAIN1844, UDI/DI 10195327207748 (each) 40195327207749 (case), Lot Numbers: 23BLA981, 23ELB126, 23ILA404, 23ILB016; 37) PAIN1845, UDI/DI 10195327210823 (each) 40195327210824 (case), Lot Numbers: 23ELA640; 38) PAIN1846, UDI/DI 10195327211639 (each) 40195327211630 (case), Lot Numbers: 23DLA446; 39) PAIN1854, UDI/DI 10195327225858 (each) 40195327225859 (case), Lot Numbers: 22JLB193, 22LLA402; 40) PAIN1855, UDI/DI 10195327226015 (each) 40195327226016 (case), Lot Numbers: 22LLA455, 23CLA867; 41) PAIN1856, UDI/DI 10195327228996 (each) 40195327228997 (case), Lot Numbers: 22JLB162, 22LLA409; 42) PAIN1856A, UDI/DI 10195327338015 (each) 40195327338016 (case), Lot Numbers: 23HLB303; 43) PAIN1864, UDI/DI 10195327238766 (each) 40195327238767 (case), Lot Numbers: 22HLB077, 22LLA731, 23FLA012; 44) PAIN1864A, UDI/DI 10195327517090 (each) 40195327517091 (case), Lot Numbers: 23LLA277; 45) PAIN1878, UDI/DI 10195327257842 (each) 40195327257843 (case), Lot Numbers: 23ALA348, 23ELA724; 46) PAIN1885, UDI/DI 10195327269159 (each) 40195327269150 (case), Lot Numbers: 22LLA664, 23DLA281, 23ELA641; 47) PAIN1887, UDI/DI 10195327272852 (each) 40195327272853 (case), Lot Numbers: 22KLA920; 48) PAIN1888, UDI/DI 10195327272913 (each) 40195327272914 (case), Lot Numbers: 23BLA151, 23BLB105; 49) PAIN1890, UDI/DI 10195327282387 (each) 40195327282388 (case), Lot Numbers: 23BLB106; 50) PAIN1892, UDI/DI 10195327283049 (each) 40195327283040 (case), Lot Numbers: 23BLB026, 23ELB077, 23ILA928; 51) PAIN1893, UDI/DI 10195327283865 (each) 40195327283866 (case), Lot Numbers: 23BLA283, 23GLA624, 23HLA803; 52) PAIN1894, UDI/DI 10195327283896 (each) 40195327283897 (case), Lot Numbers: 23BLB107, 23ELA725, 23GLA370; 53) PAIN1898, UDI/DI 10195327297596 (each) 40195327297597 (case), Lot Numbers: 23BLA284, 23BLA726; 54) PAIN1904, UDI/DI 10195327312336 (each) 40195327312337 (case), Lot Numbers: 23ELB079, 23GLA947, 23HLA774; 55) PAIN1919, UDI/DI 10195327342807 (each) 40195327342808 (case), Lot Numbers: 23CLA461, 23GLB137, 23HLA653, 23ILA462; 56) PAIN1920, UDI/DI 10195327346072 (each) 40195327346073 (case), Lot Numbers: 23FLA827, 23HLA259; 57) PAIN1937, UDI/DI 10195327362737 (each) 40195327362738 (case), Lot Numbers: 23ELB122; 58) PAIN1940, UDI/DI 10195327365332 (each) 40195327365333 (case), Lot Numbers: 23ELB080; 59) PAIN1958, UDI/DI 10195327399986 (each) 40195327399987 (case), Lot Numbers: 23GBV335, 23IBL615; 60) PAIN1967, UDI/DI 10195327417482 (each) 40195327417483 (case), Lot Numbers: 23IDA109, 23KDB373; 61) PAIN2014, UDI/DI 10195327510763 (each) 40195327510764 (case), Lot Numbers: 23JBS142; 62) PAIN9025, UDI/DI 10193489777697 (each) 40193489777698 (case), Lot Numbers: 22FLB008, 22LLA741; 63) PAIN9026, UDI/DI 10193489777703 (each) 40193489777704 (case), Lot Numbers: 23ALA479, 23ALA534, 23ALA753, 23BLA647, 23BLB091, 23CLA295; 64) PAIN9028, UDI/DI 10193489295689 (each) 40193489295680 (case), Lot Numbers: 22GLA287, 22ILA124, 23CLA410; 65) PAIN9030, UDI/DI 10193489295641 (each) 40193489295642 (case), Lot Numbers: 22FLB009, 22HLB347, 22ILA078, 23CLA407; 66) PAIN9031, UDI/DI 10195327146429 (each) 40195327146420 (case), Lot Numbers: 22GLA057, 23ALA831, 23ELA887, 23GLA871; 67) PAIN9034, UDI/DI 10195327287757 (each) 40195327287758 (case), Lot Numbers: 23ABB439, 23ABG209, 23BBS341, 23CBO460, 23EBM262, 23EBR039, 23FBI376, 23FBI377; 68) PAIN9035, UDI/DI 10195327288006 (each) 40195327288007 (case), Lot Numbers: 23BBP889; 69) PAIN9036, UDI/DI 10195327288013 (each) 40195327288014 (case), Lot Numbers: 23BBN205, 23GBW665, 23HBQ304, 23HBU314, 23JBO472; 70) REPICATH19, UDI/DI 10193489630848 (each) 40193489630849 (case), Lot Numbers: 22ABM967, 22GBG329, 22HBO372, 23CBI450, 23GBI168, 23GBI169, 23GBI170, 23HBI971; 71) SAMPA0105, UDI/DI 10195327026240 (each) 40195327026241 (case), Lot Numbers: 21JLA580; 72) SAMPA0106, UDI/DI 10195327028169 (each) 40195327028160 (case), Lot Numbers: 21LBY057; 73) SAMPA0110, UDI/DI 10195327097004 (each) 40195327097005 (case), Lot Numbers: 22BLA107, 22CLA719, 22CLA850; 74) SAMPA0111, UDI/DI 10195327204693 (each) 40195327204694 (case), Lot Numbers: 22LLA665
FEI Number 1417592
Recalling Firm/
Manufacturer
MEDLINE INDUSTRIES, LP - Northfield
3 Lakes Dr
Northfield IL 60093-2753
For Additional Information ContactHaley Barclay
866-359-1704
Manufacturer Reason
for Recall
The connectors may not allow catheters to be inserted as intended if the connector is not in the fully open position.
FDA Determined
Cause 2
Under Investigation by firm
ActionAn IMMEDIATE ACTION REQUIRED RECALL notification letter dated 10/16/24 was sent to customers. REQUIRED ACTION: 1.Immediately check your stock for the affected item number and the affected lot numbers which can be located within the recall portal. Quarantine all affected product immediately. 2.Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the response form. 3.Upon receipt of your submitted response form, your account will receive over-labels to place unaffected inventory, with instructions for your staff to remove the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. The login for completing the response form is: Website link: https://recalls.medline.comRecall Reference #: R-24-216-FGRecall Code: 4.If you are a distributor, or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. If you have any questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.
DistributionUS Nationwide. Bermuda, Canada, Panama, UAE, US Virgin Islands
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
-
-