| | Class 2 Device Recall Medline |  |
| Date Initiated by Firm | October 09, 2024 |
| Date Posted | December 05, 2024 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0633-2025 |
| Recall Event ID |
95621 |
| Product Classification |
Plastic surgery and accessories kit - Product Code FTN
|
| Product | MEDLINE convenience kits labeled as:
1) MAMMOPLASTY-LF, REF CDS983478L;
2) MAMMOPLASTY-LF, REF CDS983478M;
3) PLASTICS FREE FLAP, REF CDS983493M;
4) PLASTIC TORSO-LF, REF CDS983496L;
5) PLASTIC TORSO-LF, REF CDS983496M;
6) PLASTIC HEAD AND NECK-LF, REF CDS983498N;
7) ACH PLASTICS BREAST-LF, REF CDS983639J;
8) ACH PLASTICS BREAST-LF, REF CDS983639K;
9) BREAST ABD, REF CDS984267J;
10) DAY SURG PLASTIC CDS, REF CDS984993N;
11) DAY SURG PLASTIC CDS, REF CDS984993O;
12) DAY SURG PLASTIC CDS, REF CDS984993P;
13) PLASTIC PACK, REF DYNJ00264L;
14) LKLND BRST CHST PK RFID-LF, REF DYNJ0101536L;
15) OCULOPLASTY PACK-LF, REF DYNJ0252232U;
16) PLASTICS PACK, REF DYNJ02615G;
17) BREAST PBM-LF, REF DYNJ0318423M;
18) FLAP RECONSTRUCTION PACK-LF, REF DYNJ0378388P;
19) BREAST PACK-LF, REF DYNJ0378394R;
20) BREAST PACK-LF, REF DYNJ0378394S;
21) BREAST PACK-LF, REF DYNJ0378394T;
22) BREAST ABDOMINOPLASTY MZ PK-LF, REF DYNJ0428947K;
23) BREAST ABDOMINOPLASTY MZ PK-LF, REF DYNJ0428947L;
24) ASC MAJOR PLASTIC PACK-LF, REF DYNJ0854517M;
25) ASC ENT PLASTIC PACK-LF, REF DYNJ0854650J;
26) BREAST PACK-LF, REF DYNJ0956628M;
27) BREAST PACK-LF, REF DYNJ0956628N;
28) BREAST PACK-LF, REF DYNJ0956628O;
29) AUGMENTATION #87-RF, REF DYNJ27319U;
30) AUGMENTATION #87-RF, REF DYNJ27319V;
31) PLASTICS PACK-LF, REF DYNJ36107J;
32) PLASTIC MINOR PACK, REF DYNJ37620F;
33) CHEST/BREAST PACK, REF DYNJ39118C;
34) PLASTIC BREAST PACK, REF DYNJ40620C;
35) PLASTIC BREAST PACK, REF DYNJ40620D;
36) BREAST RECONSTRUCTION-HOULE, REF DYNJ44852M;
37) BREAST MINOR PK RFID, REF DYNJ52409J;
38) BREAST MAJOR PACK, REF DYNJ52435J;
39) BREAST MAJOR PK RFID, REF DYNJ52435K;
40) BREAST PACK, REF DYNJ57527C;
41) TRANSGENDER MASTECTOMY 2, REF DYNJ58257G;
42) BREAST PACK SSC, REF DYNJ59733A;
43) BREAST PACK, REF DYNJ60511B;
44) BREAST PACK, REF DYNJ60511F;
45) PLASTIC PACK, REF DYNJ61755B;
46) TYMPANOPLASTY, REF DYNJ63332D;
47) BREAST PACK, REF DYNJ63369D;
48) BREAST SURGERY FOAKS, REF DYNJ63577C;
49) MAJOR PLASTIC FOAKS, REF DYNJ63579C;
50) MAJOR PLASTIC PACK FAIRFAX, REF DYNJ63612C;
51) MAJOR BREAST WOMENS PACK, REF DYNJ63647C;
52) PLASTIC BASIN PACK, REF DYNJ63883B;
53) PLASTICS WOODBURN, REF DYNJ64001B;
54) BREAST PK, REF DYNJ64144B;
55) BREAST PK, REF DYNJ64144C;
56) PACK,PLASTIC, REF DYNJ65477B;
57) PACK,PLASTIC BASIC ASC, REF DYNJ65485A;
58) OSC BREAST PACK, REF DYNJ65808A;
59) CHEST/BREAST PACK, REF DYNJ66495A;
60) PLASTICS PACK, REF DYNJ66610B;
61) BREAST PACK, REF DYNJ66739B;
62) MAJOR PLASTICS PACK, REF DYNJ67968A;
63) BREAST PACK, REF DYNJ68464A;
64) FLAP PACK, REF DYNJ80179A;
65) PLASTICS PACK, REF DYNJ81797A;
66) PLASTICS PACK, REF DYNJ81811A;
67) MINOR PLASTICS PACK, REF DYNJ82196;
68) MINOR PLASTICS PACK, REF DYNJ82196A;
69) DELNOR MAJOR PLASTICS PACK, REF DYNJ82248;
70) DELNOR SIMPLE PLASTIC PACK, REF DYNJ82250;
71) DELNOR SIMPLE PLASTIC PACK, REF DYNJ82250A;
72) DELNOR AUGMENTATION PACK, REF DYNJ82265;
73) DELNOR AUGMENTATION PACK, REF DYNJ82265A;
74) AUGMENTATION PACK, REF DYNJ82279;
75) AUGMENTATION PACK, REF DYNJ82279A;
76) PLASTIC SURGERY PACK, REF DYNJ82352;
77) PLASTIC SURGERY PACK, REF DYNJ82352A;
78) FREE FLAP 2 PACK, REF DYNJ82659B;
79) FREE FLAP 1 PACK, REF DYNJ82660A;
80) FREE FLAP 1 PACK, REF DYNJ82660B;
81) FACE PACK, REF DYNJ83532;
82) BREAST RECONSTRUCTION PACK-LF, REF DYNJ83612A;
83) PLASTIC PACK, REF DYNJ83856;
84) BREAST PACK, REF DYNJ84332;
85) MAJOR PLASTIC PACK, REF DYNJ84827;
86) PLASTICS PACK, REF DYNJ84966;
87) PLASTIC PACK, REF DYNJ85702;
88) FACIAL PLASTICS FLAP PACK, REF DYNJ86031;
89) BREAST-CHEST, REF DYNJ900294K;
90) GENERAL PLASTIC-LF, REF DYNJ901738O;
91) TRAM RECONSTRUCTION-LF, REF DYNJ901753L;
92) PLASTIC UNIVERSAL, REF DYNJ901792J;
93) MID HUDSON BREAST, REF DYNJ903614C;
94) MID HUDSON BREAST, REF DYNJ903614D;
95) MID HUDSON BREAST, REF DYNJ903614F;
96) MID HUDSON PLASTICS, REF DYNJ903616F;
97) MID HUDSON PLASTICS, REF DYNJ903616G;
98) MID |
| Code Information |
1) REF CDS983478L, UDI/DI 10195327497286 (each) 40195327497287 (case), Lot Numbers: 23JDA999, 23KDB756, 24ADA957, 24DDA553;
2) REF CDS983478M, UDI/DI 10198459000065 (each) 40198459000066 (case), Lot Numbers: 24EDB504;
3) REF CDS983493M, UDI/DI 10195327497415 (each) 40195327497416 (case), Lot Numbers: 23LDA144, 24ADA328;
4) REF CDS983496L, UDI/DI 10195327497408 (each) 40195327497409 (case), Lot Numbers: 23KDC085, 24ADB601;
5) REF CDS983496M, UDI/DI 10198459000201 (each) 40198459000202 (case), Lot Numbers: 24EDA817;
6) REF CDS983498N, UDI/DI 10198459000188 (each) 40198459000189 (case), Lot Numbers: 24EDA816;
7) REF CDS983639J, UDI/DI 10195327228149 (each) 40195327228140 (case), Lot Numbers: 22IBF012, 22IBF004, 22LBE960, 22KBK687;
8) REF CDS983639K, UDI/DI 10195327351533 (each) 40195327351534 (case), Lot Numbers: 23EBA182, 23EBN996, 23CBV828, 23FBP157, 23FBH151;
9) REF CDS984267J, UDI/DI 10193489288872 (each) 40193489288873 (case), Lot Numbers: 22GBR052, 22FBG539, 22KBU660, 23ABL613, 23DBD827;
10) REF CDS984993N, UDI/DI 10195327031114 (each) 40195327031115 (case), Lot Numbers: 22EBV339;
11) REF CDS984993O, UDI/DI 10195327175764 (each) 40195327175765 (case), Lot Numbers: 22GBF349, 22JBJ441, 23BBV017, 23EBN595, 23IBU110, 23GBE740, 23GBE766, 23IBC614, 23LBC779, 23KBH710;
12) REF CDS984993P, UDI/DI 10195327588007 (each) 40195327588008 (case), Lot Numbers: 24ABH482, 24DBD860;
13) REF DYNJ00264L, UDI/DI 10193489942019 (each) 40193489942010 (case), Lot Numbers: 22IMH644, 22EMF495, 23AMB776, 22FMD650, 23CMA570, 22KME942;
14) REF DYNJ0101536L, UDI/DI 10195327333041 (each) 40195327333042 (case), Lot Numbers: 23IMB288;
15) REF DYNJ0252232U, UDI/DI 10193489954111 (each) 40193489954112 (case), Lot Numbers: 22FMI316, 22EMD248, 23EMF255, 23HMB707, 23AME630, 23IMA845, 23DMC649, 23IMG327, 23JMI300;
16) REF DYNJ02615G, UDI/DI 10195327291051 (each) 40195327291052 (case), Lot Numbers: 23JMJ208;
17) REF DYNJ0318423M, UDI/DI 10193489442687 (each) 40193489442688 (case), Lot Numbers: 22LMH207, 22JMG688, 23LMB715, 23EMJ272;
18) REF DYNJ0378388P, UDI/DI 10195327089979 (each) 40195327089970 (case), Lot Numbers: 22DBK643, 22DBI785, 23BBC753, 22LBO807, 23CBT793, 23HBY759, 23HBA704;
19) REF DYNJ0378394R, UDI/DI 10195327089900 (each) 40195327089901 (case), Lot Numbers: 22EBM867;
20) REF DYNJ0378394S, UDI/DI 10195327187958 (each) 40195327187959 (case), Lot Numbers: 22FBP935, 22LBO853, 23DBB583, 23EBE187, 23DBJ565, 23FBG607, 23GBH212;
21) REF DYNJ0378394T, UDI/DI 10195327498306 (each) 40195327498307 (case), Lot Numbers: 23IBJ701, 23JBP685;
22) REF DYNJ0428947K, UDI/DI 10195327055578 (each) 40195327055579 (case), Lot Numbers: 22CLB388, 22EMG126, 22ELA518, 22KME295, 22HME064;
23) REF DYNJ0428947L, UDI/DI 10195327329136 (each) 40195327329137 (case), Lot Numbers: 23EMH835, 23CMC033, 23BMD726;
24) REF DYNJ0854517M, UDI/DI 10193489832419 (each) 40193489832410 (case), Lot Numbers: 22HMC268, 22IME146, 23BMD202, 23EMD923, 23FME798, 23JMF894, 24BMD822, 23GMH631, 24DMD934, 23KMI567, 23IMG202, 24CMC965;
25) REF DYNJ0854650J, UDI/DI 10193489832426 (each) 40193489832427 (case), Lot Numbers: 22KMF904, 22JMA456, 23EMD925, 23BMA380, 23LMI046, 23KMJ020, 24BMB393;
26) REF DYNJ0956628M, UDI/DI 10195327117955 (each) 40195327117956 (case), Lot Numbers: 22FMI331;
27) REF DYNJ0956628N, UDI/DI 10195327216450 (each) 40195327216451 (case), Lot Numbers: 22JMB058, 22LMD532, 23AMG987, 23BMC748;
28) REF DYNJ0956628O, UDI/DI 10195327416133 (each) 40195327416134 (case), Lot Numbers: 23HMH285, 23GMI738, 23EMI827, 23IMG042, 23LMA088, 23KMF814, 24BMG288, 23LMF199, 23LMF199A, 24DMF169;
29) REF DYNJ27319U, UDI/DI 10195327104528 (each) 40195327104529 (case), Lot Numbers: 22DBE822, 22DBT722;
30) REF DYNJ27319V, UDI/DI 10195327208127 (each) 40195327208128 (case), Lot Numbers: 22IBJ843, 22IBJ845, 23DBJ736, 23ABF396, 22GBY108, 23EBW485, 23EBQ342;
31) REF DYNJ36107J, UDI/DI 10195327080273 (each) 40195327080274 (case), Lot Numbers: 22FMH984;
32) REF DYNJ37620F, UDI/DI 10195327497392 (each) 40195327497393 (case), Lot Numbers: 23JDA165, 23LDA934, 24CDA370;
33) REF DYNJ39118C, UDI/DI 10195327509750 (each) 40195327509751 (case), Lot Numbers: 23JBT477;
34) REF DYNJ40620C, UDI/DI 10193489699692 (each) 40193489699693 (case), Lot Numbers: 23CBV601, 23CBH954, 23EBT508;
35) REF DYNJ40620D, UDI/DI 10195327435776 (each) 40195327435777 (case), Lot Numbers: 23JBP045;
36) REF DYNJ44852M, UDI/DI 10195327325220 (each) 40195327325221 (case), Lot Numbers: 23AMC530, 23AMC374, 23CMF343, 23FMH424;
37) REF DYNJ52409J, UDI/DI 10195327339203 (each) 40195327339204 (case), Lot Numbers: 23DDB215, 23GDB232;
38) REF DYNJ52435J, UDI/DI 10195327222819 (each) 40195327222810 (case), Lot Numbers: 22KDA287;
39) REF DYNJ52435K, UDI/DI 10195327339630 (each) 40195327339631 (case), Lot Numbers: 23EDA425, 23GDC225, 23IDA847;
40) REF DYNJ57527C, UDI/DI 10195327243470 (each) 40195327243471 (case), Lot Numbers: 23ABG344, 23BBS780;
41) REF DYNJ58257G, UDI/DI 10193489968644 (each) 40193489968645 (case), Lot Numbers: 22EMA050, 22JME557, 22HMH771, 22LMD593, 23CMC086;
42) REF DYNJ59733A, UDI/DI 10195327206413 (each) 40195327206414 (case), Lot Numbers: 22LBI106;
43) REF DYNJ60511B, UDI/DI 10195327134976 (each) 40195327134977 (case), Lot Numbers: 22EBL029;
44) REF DYNJ60511F, UDI/DI 10195327352332 (each) 40195327352333 (case), Lot Numbers: 23CBB490, 23IBU091, 23IBI131, 23JBM061;
45) REF DYNJ61755B, UDI/DI 10195327149550 (each) 40195327149551 (case), Lot Numbers: 22FMG692, 23BMB157, 22JMD332, 22HMF130, 22KMG291;
46) REF DYNJ63332D, UDI/DI 10195327062675 (each) 40195327062676 (case), Lot Numbers: 22DBL674, 22GBC932;
47) REF DYNJ63369D, UDI/DI 10195327273354 (each) 40195327273355 (case), Lot Numbers: 23BMA188, 23JMG362;
48) REF DYNJ63577C, UDI/DI 10195327122850 (each) 40195327122851 (case), Lot Numbers: 22EBT513, 22GBV398, 23IBQ212, 23JBG821;
49) REF DYNJ63579C, UDI/DI 10195327122867 (each) 40195327122868 (case), Lot Numbers: 23HBP091, 23IBT926;
50) REF DYNJ63612C, UDI/DI 10195327350291 (each) 40195327350292 (case), Lot Numbers: 23FBE254, 23IBN288, 23GBT338;
51) REF DYNJ63647C, UDI/DI 10195327350406 (each) 40195327350407 (case), Lot Numbers: 23HBI038, 23CBU235, 23FBR630;
52) REF DYNJ63883B, UDI/DI 10195327014148 (each) 40195327014149 (case), Lot Numbers: 22DDA800, 22EDB037;
53) REF DYNJ64001B, UDI/DI 10195327178062 (each) 40195327178063 (case), Lot Numbers: 22GBW636, 22FBR610, 23JBG814, 23IBN289;
54) REF DYNJ64144B, UDI/DI 10195327366773 (each) 40195327366774 (case), Lot Numbers: 23FBF133, 23GBQ526;
55) REF DYNJ64144C, UDI/DI 10195327513023 (each) 40195327513024 (case), Lot Numbers: 23JBL674;
56) REF DYNJ65477B, UDI/DI 10195327303440 (each) 40195327303441 (case), Lot Numbers: 23AMG835, 23HMC960, 23IME795;
57) REF DYNJ65485A, UDI/DI 10195327303433 (each) 40195327303434 (case), Lot Numbers: 23HMG237;
58) REF DYNJ65808A, UDI/DI 10193489949803 (each) 40193489949804 (case), Lot Numbers: 22FBD780, 22FBX439;
59) REF DYNJ66495A, UDI/DI 10195327247584 (each) 40195327247585 (case), Lot Numbers: 22JMB371, 23BMH635;
60) REF DYNJ66610B, UDI/DI 10195327429140 (each) 40195327429141 (case), Lot Numbers: 23GBF354, 23KBC530, 23HBV806, 24ABW336, 24BBN977, 24DBF700, 24CBN383;
61) REF DYNJ66739B, UDI/DI 10195327504441 (each) 40195327504442 (case), Lot Numbers: 23KBM905, 23LBR954;
62) REF DYNJ67968A, UDI/DI 10193489962949 (each) 40193489962940 (case), Lot Numbers: 22IMF531, 22GMG777, 23BMC966, 23DMA746;
63) REF DYNJ68464A, UDI/DI 10195327261665 (each) 40195327261666 (case), Lot Numbers: 23ABL973, 23ABK919, 23ABH279, 22LBJ691, 23BBE426, 23FBH839;
64) REF DYNJ80179A, UDI/DI 10195327194949 (each) 40195327194940 (case), Lot Numbers: 22LBJ892, 23BBU702, 22IBJ389, 23FBC812;
65) REF DYNJ81797A, UDI/DI 10195327515973 (each) 40195327515974 (case), Lot Numbers: 23KBD075, 23IBT729, 23LBM384, 24BBC801, 24BBG132, 24CBK447, 24CBC579;
66) REF DYNJ81811A, UDI/DI 10195327681562 (each) 40195327681563 (case), Lot Numbers: 24DBK129;
67) REF DYNJ82196, UDI/DI 10195327183509 (each) 40195327183500 (case), Lot Numbers: 23DDB164, 23GDB788, 23HDB846;
68) REF DYNJ82196A, UDI/DI 10195327519827 (each) 40195327519828 (case), Lot Numbers: 23LMG620, 23JDA365;
69) REF DYNJ82248, UDI/DI 10195327181741 (each) 40195327181742 (case), Lot Numbers: 23GDA030, 23GDA857;
70) REF DYNJ82250, UDI/DI 10195327181758 (each) 40195327181759 (case), Lot Numbers: 23FDA957, 23HDB240;
71) REF DYNJ82250A, UDI/DI 10195327528584 (each) 40195327528585 (case), Lot Numbers: 23KME039;
72) REF DYNJ82265, UDI/DI 10195327182274 (each) 40195327182275 (case), Lot Numbers: 23FDA050;
73) REF DYNJ82265A, UDI/DI 10195327528393 (each) 40195327528394 (case), Lot Numbers: 23JMJ376, 23KMH770;
74) REF DYNJ82279, UDI/DI 10195327183950 (each) 40195327183951 (case), Lot Numbers: 23GDB647, 23IMF720;
75) REF DYNJ82279A, UDI/DI 10195327545161 (each) 40195327545162 (case), Lot Numbers: 23KMF508;
76) REF DYNJ82352, UDI/DI 10195327186630 (each) 40195327186631 (case), Lot Numbers: 23DDA656, 23IDA294, 23FDA737;
77) REF DYNJ82352A, UDI/DI 10195327545772 (each) 40195327545773 (case), Lot Numbers: 23KMD874;
78) REF DYNJ82659B, UDI/DI 10195327557799 (each) 40195327557790 (case), Lot Numbers: 23LBH530, 24BBO870;
79) REF DYNJ82660A, UDI/DI 10195327489687 (each) 40195327489688 (case), Lot Numbers: 23JBD568;
80) REF DYNJ82660B, UDI/DI 10195327557706 (each) 40195327557707 (case), Lot Numbers: 23KBL438, 24ABN390, 24BBI263;
81) REF DYNJ83532, UDI/DI 10195327276416 (each) 40195327276417 (case), Lot Numbers: 23FBL704, 23HBL044, 23JBI232, 23KBI654;
82) REF DYNJ83612A, UDI/DI 10195327433703 (each) 40195327433704 (case), Lot Numbers: 23KMI546, 23IMG920, 23GMI169;
83) REF DYNJ83856, UDI/DI 10195327318253 (each) 40195327318254 (case), Lot Numbers: 23FBO433, 23FBO778, 23CBM467;
84) REF DYNJ84332, UDI/DI 10195327353117 (each) 40195327353118 (case), Lot Numbers: 23DMB713;
85) REF DYNJ84827, UDI/DI 10195327425272 (each) 40195327425273 (case), Lot Numbers: 23HBE654, 23EBX040, 23JBI656;
86) REF DYNJ84966, UDI/DI 10195327441821 (each) 40195327441822 (case), Lot Numbers: 23JBJ781, 23KBB702, 23JBW571, 23KBG577;
87) REF DYNJ85702, UDI/DI 10195327536213 (each) 40195327536214 (case), Lot Numbers: 24ABP053, 24DBR982;
88) REF DYNJ86031, UDI/DI 10195327581633 (each) 40195327581634 (case), Lot Numbers: 24BLB039;
89) REF DYNJ900294K, UDI/DI 10193489915150 (each) 40193489915151 (case), Lot Numbers: 22KBP158, 23ABM718, 22JBS854, 23EBF969, 23EBR997;
90) REF DYNJ901738O, UDI/DI 10195327495930 (each) 40195327495931 (case), Lot Numbers: 24ADA162, 23KDA408;
91) REF DYNJ901753L, UDI/DI 10195327496180 (each) 40195327496181 (case), Lot Numbers: 24BDA185, 23LDB476;
92) REF DYNJ901792J, UDI/DI 10195327546564 (each) 40195327546565 (case), Lot Numbers: 24AMJ347, 23LMD856;
93) REF DYNJ903614C, UDI/DI 10193489280234 (each) 40193489280235 (case), Lot Numbers: 22GBY708;
94) REF DYNJ903614D, UDI/DI 10195327237516 (each) 40195327237517 (case), Lot Numbers: 23ABM705;
95) REF DYNJ903614F, UDI/DI 10195327368371 (each) 40195327368372 (case), Lot Numbers: 24CBH093, 23HBD137, 24DBR891;
96) REF DYNJ903616F, UDI/DI 10195327237530 (each) 40195327237531 (case), Lot Numbers: 23BBB789;
97) REF DYNJ903616G, UDI/DI 10195327372262 (each) 40195327372263 (case), Lot Numbers: 23IBG231, 23IBG829, 23HBN599, 23EBI668, 23FBB347, 24BBG027, 23GBS726;
98) REF DYNJ903616I, UDI/DI 10195327648374 (each) 40195327648375 (case), Lot Numbers: 24DBL776;
99) REF DYNJ903616J, UDI/DI 10198459007279 (each) 40198459007270 (case), Lot Numbers: 24IBL099, 24HBL800, 24GBK425;
100) REF DYNJ906471F, UDI/DI 10195327210946 (each) 40195327210947 (case), Lot Numbers: 22IDB704, 22LDB605, 23IDB173;
101) REF DYNJ907141A, UDI/DI 10193489999938 (each) 40193489999939 (case), Lot Numbers: 22FBM161, 22FBG627, 22GBY527, 22JBR768, 23ABQ820, 23ABE732, 23EBU779, 23EBA967, 23CBW388;
102) REF DYNJ907141B, UDI/DI 10195327450434 (each) 40195327450435 (case), Lot Numbers: 23HBA138;
103) REF DYNJ907410A, UDI/DI 10195327335823 (each) 40195327335824 (case), Lot Numbers: 23EMD893;
104) REF DYNJ907410B, UDI/DI 10195327462222 (each) 40195327462223 (case), Lot Numbers: 23HMF127, 23IMF385;
105) REF DYNJ907791B, UDI/DI 10195327185237 (each) 40195327185238 (case), Lot Numbers: 22JBN762, 22GBK931, 22KBE789, 22KBB846, 23CBN749, 23CBJ391, 23CBE960, 23DBM205;
106) REF DYNJ907791D, UDI/DI 10195327439859 (each) 40195327439850 (case), Lot Numbers: 23GBB518;
107) REF DYNJ908058A, UDI/DI 10195327125424 (each) 40195327125425 (case), Lot Numbers: 22GBM150, 22FBR755, 22FBM688, 22LBC294, 22KBF561, 22JBE835, 23ABM655, 23EBA419;
108) REF DYNJ908258C, UDI/DI 10195327243876 (each) 40195327243877 (case), Lot Numbers: 23FDA366, 23HDA578, 23JDB900;
109) REF DYNJ908691A, UDI/DI 10195327515911 (each) 40195327515912 (case), Lot Numbers: 23LBQ414, 24DBA370;
110) REF DYNJ908729, UDI/DI 10195327159221 (each) 40195327159222 (case), Lot Numbers: 22GBV793, 23EBI552, 23HBS915, 23HBS528, 23JBV713, 23LBH123, 24ABW912, 24ABF765;
111) REF DYNJ908729A, UDI/DI 10195327641375 (each) 40195327641376 (case), Lot Numbers: 24CBP680, 24DBS149;
112) REF DYNJ908790, UDI/DI 10195327183295 (each) 40195327183296 (case), Lot Numbers: 23CDB747, 23CDA590, 23GDA891, 23IDB773;
113) REF DYNJ908790A, UDI/DI 10195327520410 (each) 40195327520411 (case), Lot Numbers: 24AME679;
114) REF DYNJ908839, UDI/DI 10195327184438 (each) 40195327184439 (case), Lot Numbers: 23ABT334, 23ABL506;
115) REF DYNJ909203, UDI/DI 10195327304065 (each) 40195327304066 (case), Lot Numbers: 23BBG452;
116) REF DYNJ909410, UDI/DI 10195327338473 (each) 40195327338474 (case), Lot Numbers: 23FMI603;
117) REF DYNJ909786, UDI/DI 10195327475802 (each) 40195327475803 (case), Lot Numbers: 23IBK726;
118) REF DYNJ9426113M, UDI/DI 10195327057114 (each) 40195327057115 (case), Lot Numbers: 22FLA762, 23CMF426;
119) REF DYNJVB91043, UDI/DI 10193489284454 (each) 40193489284455 (case), Lot Numbers: 22GDA314, 22IDA697 |
| FEI Number |
1417592
|
Recalling Firm/ Manufacturer |
MEDLINE INDUSTRIES, LP - Northfield 3 Lakes Dr Northfield IL 60093-2753
|
| For Additional Information Contact | Haley Barclay 866-359-1704 |
Manufacturer Reason for Recall | Medline is recalling medical convenience manufactured using certain lots of SafeAir Smoke Evacuation Pencil which were subsequently recalled by Stryker. |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | Medline issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 10/11/2024 via mail and email. The notice explained the problem with the Stryker component, potential risk, requested the kits be quarantined pending additional labeling and o respond through their portal: https://recalls.medline.com. Once the response is received, Medline will provide additional labeling to be applied to the affected kits which states the following: "Attention: R-24-219, This kit contains a recalled item, STRYKER SafeAir Smoke Evacuation Pencil, VPN: 0703-047-000, 0703-047-001, 0703-047-002, 0703-046-000, 0703-046-001, 0703-046-002. DO NOT USE THIS COMPONENT. The rest of the kit may be used. Please remove this item only and replace with product from supply. *Do not cover or remove this sticker*".
Those further distributing the product were directed to notify their customers.
|
| Quantity in Commerce | 25877 units |
| Distribution | US Nationwide distribution. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|