• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Medline

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Medlinesee related information
Date Initiated by FirmOctober 09, 2024
Date PostedDecember 05, 2024
Recall Status1 Open3, Classified
Recall NumberZ-0649-2025
Recall Event ID 95621
Product Classification Orthopedic tray - Product Code OJH
ProductMEDLINE convenience kits labeled as: 1) ORTHO LAMINECTOMY-LF, REF CDS780112J; 2) NEURO CRANIOTOMY CDS #36-RF, REF CDS780119V; 3) BASIC BACK CDS, REF CDS780147N; 4) INSTRUMENTED BACK CDS, REF CDS780148Q; 5) INSTRUMENTED BACK CDS, REF CDS780148R; 6) TOTAL HIP, REF CDS920027W; 7) TOTAL HIP, REF CDS920027X; 8) TOTAL HIP, REF CDS920027Y; 9) TOTAL KNEE CDS-LF, REF CDS940047AF; 10) TOTAL KNEE CDS-LF, REF CDS940047AG; 11) TOTAL HIP CDS-LF, REF CDS940048AF; 12) TOTAL HIP CDS-LF, REF CDS940048AG; 13) TOTAL HIP CDS-LF, REF CDS940048AI; 14) TOTAL KNEE, REF CDS940072W; 15) TOTAL KNEE, REF CDS940072Y; 16) TOTAL KNEE REPLACEMENT CDS, REF CDS940823AB; 17) TOTAL KNEE CDS, REF CDS940842Y; 18) TOTAL HIP CDS, REF CDS940843AA; 19) FRACTURE TABLE-LF, REF CDS940857K; 20) FRACTURE TABLE-LF, REF CDS940857L; 21) EXTREMITY-LF, REF CDS940860P; 22) EXTREMITY-LF, REF CDS940860Q; 23) KNEE/SHOULDER, REF CDS940911Q; 24) TOTAL HIP CDS, REF CDS940963Y; 25) SHOULDER CDS, REF CDS941065L; 26) TOTAL JOINT-LF, REF CDS941167M; 27) TOTAL JOINT-LF, REF CDS941167N; 28) LOWER EXTREMITY CDS-LF, REF CDS980666I; 29) LOWER EXTREMITY CDS-LF, REF CDS980666J; 30) TOTAL HIP-LF, REF CDS980788O; 31) TOTAL HIP-LF, REF CDS980788P; 32) EXTREMITY KIT, REF CDS981314R; 33) EXTREMITY KIT, REF CDS981314S; 34) ORIF HIP KIT, REF CDS981318T; 35) ORIF HIP KIT, REF CDS981318U; 36) TOTAL HIP KIT, REF CDS981320V; 37) TOTAL HIP KIT, REF CDS981320W; 38) TOTAL KNEE KIT, REF CDS981323AA; 39) TOTAL KNEE KIT, REF CDS981323Y; 40) BI LATERAL KNEE KIT, REF CDS981324T; 41) UPPER EXTREMITY CDS-LF, REF CDS981564G; 42) UPPER EXTREMITY CDS-LF, REF CDS981564I; 43) LAMI, REF CDS981886X; 44) CRANIOTOMY CDS, REF CDS981888W; 45) CRANIOTOMY CDS, REF CDS981888X; 46) TOTAL KNEE CDS, REF CDS981926D; 47) TOTAL KNEE PROCEDURE-LF, REF CDS982082F; 48) BHMC SYSTEM HIP KIT, REF CDS982396N; 49) BAPTIST NORTH TOTAL KNEE CDS, REF CDS982404M; 50) BAP NORTH SHOULDER CDS, REF CDS982406M; 51) BAP NORTH UNIV EXTREMITY CDS, REF CDS982408K; 52) TOTAL KNEE CDS, REF CDS982823Q; 53) LOWER EXTREMITY CDS-LF, REF CDS982927M; 54) LUMBAR LAMINECTOMY CDS-LF, REF CDS982928O; 55) TOTAL HIP REPLACEMENT CDS-LF, REF CDS982929K; 56) TOTAL KNEE REPLACEMENT CDS-LF, REF CDS982931L; 57) UPPER EXTREMITY-LF, REF CDS982932K; 58) LOWER EXTREMITY CDS, REF CDS982943Q; 59) LOWER EXTREMITY CDS, REF CDS982943R; 60) LOWER EXTREMITY CDS, REF CDS982943S; 61) TOTAL KNEE, REF CDS982986T; 62) SPECIALIZED HIP CDS, REF CDS983030V; 63) EXTREMITY CDS, REF CDS983031Q; 64) ZALE SPINE CDS, REF CDS983182J; 65) ZALE SPINE CDS, REF CDS983182K; 66) ZALE CRANIOTOMY CDS, REF CDS983188K; 67) DR GILL'S CDS, REF CDS983189G; 68) ZALE TOTAL KNEE CDS, REF CDS983194I; 69) ZALE TOTAL KNEE CDS, REF CDS983194J; 70) ZALE TOTAL KNEE CDS, REF CDS983194K; 71) ZALE TOTAL KNEE CDS, REF CDS983194L; 72) ZALE TOTAL HIP CDS, REF CDS983195I; 73) ZALE TOTAL HIP CDS, REF CDS983195J; 74) ZALE TOTAL HIP CDS, REF CDS983195K; 75) ZALE EXTREMITY CDS, REF CDS983199K; 76) ZALE TOTAL SHOULDER CDS, REF CDS983203J; 77) ZALE TOTAL SHOULDER CDS, REF CDS983203K; 78) ZALE TOTAL SHOULDER CDS, REF CDS983203L; 79) ORTHO EXTREMITY CDS, REF CDS983408G; 80) ORTHO-CV CDS, REF CDS983415G; 81) PODIATRY CDS, REF CDS983416D; 82) PODIATRY CDS, REF CDS983416G; 83) ORTHO EXTREMITY, REF CDS983468O; 84) LUMBAR LAMINECTOMY NEURO CDS, REF CDS983476J; 85) SITTING CERVICAL CDS, REF CDS983480G; 86) ORTHO SPINE, REF CDS983482L; 87) ARTHROSCOPY SHOULDER, REF CDS983485L; 88) EXTREMITY-LF, REF CDS983487K; 89) ACL-LF, REF CDS983490M; 90) PODIATRY CDS, REF CDS983610L; 91) TOTAL JOINT-LF, REF CDS983863F; 92) UPPER EXTREMITY CDS, REF CDS983907K; 93) UPPER EXTREMITY CDS, REF CDS983907L; 94) LOWER EXTREMITY CDS, REF CDS983909O; 95) LOWER EXTREMITY CDS, REF CDS983909P; 96) GENERAL ORTHO CDS, REF CDS983915L; 97) GENERAL ORTHO CDS, REF CDS983915M; 98) MAJOR ORTHO, REF CDS984253L; 99) LOWER EXTREMTIY, REF CDS984254K; 100) LOWER EXTREMTIY, REF CDS98425
Code Information 86OJF 1) REF CDS780112J, UDI/DI 10195327009663 (each) 40195327009664 (case), Lot Numbers: 22FBF455, 23CBH389, 23FBO500; 2) REF CDS780119V, UDI/DI 10195327114916 (each) 40195327114917 (case), Lot Numbers: 22KBN442, 22HBO015, 23ABF941, 23DBE182, 23BBQ326; 3) REF CDS780147N, UDI/DI 10195327268770 (each) 40195327268771 (case), Lot Numbers: 23DBT165, 22JBU233, 23BBJ852; 4) REF CDS780148Q, UDI/DI 10193489202939 (each) 40193489202930 (case), Lot Numbers: 22JBL294; 5) REF CDS780148R, UDI/DI 10195327268787 (each) 40195327268788 (case), Lot Numbers: 23CBE921, 22LBL862, 23EBA185, 23EBU780; 6) REF CDS920027W, UDI/DI 10193489946956 (each) 40193489946957 (case), Lot Numbers: 22GBL731; 7) REF CDS920027X, UDI/DI 10195327241742 (each) 40195327241743 (case), Lot Numbers: 22JBF953; 8) REF CDS920027Y, UDI/DI 10195327298531 (each) 40195327298532 (case), Lot Numbers: 23CBI608, 23CBF425; 9) REF CDS940047AF, UDI/DI 10195327243890 (each) 40195327243891 (case), Lot Numbers: 23DDC114, 23FDA378, 23IDA521, 23HDB417; 10) REF CDS940047AG, UDI/DI 10195327516710 (each) 40195327516711 (case), Lot Numbers: 23KDA158; 11) REF CDS940048AF, UDI/DI 10195327150259 (each) 40195327150250 (case), Lot Numbers: 22EDB749; 12) REF CDS940048AG, UDI/DI 10195327241896 (each) 40195327241897 (case), Lot Numbers: 22IDA258, 23CDB146, 23IDA944, 23HDA534; 13) REF CDS940048AI, UDI/DI 10195327516703 (each) 40195327516704 (case), Lot Numbers: 23KDA886; 14) REF CDS940072W, UDI/DI 10193489288544 (each) 40193489288545 (case), Lot Numbers: 22DBN494; 15) REF CDS940072Y, UDI/DI 10195327300845 (each) 40195327300846 (case), Lot Numbers: 23EBG408, 23EBE750; 16) REF CDS940823AB, UDI/DI 10193489883282 (each) 40193489883283 (case), Lot Numbers: 22FBP685, 22GBM982, 22FBR756, 23CBS044, 23ABJ584, 22KBN648, 22JBL399, 23EBG438; 17) REF CDS940842Y, UDI/DI 10193489845129 (each) 40193489845120 (case), Lot Numbers: 23EBV624, 23DBT379; 18) REF CDS940843AA, UDI/DI 10195327151409 (each) 40195327151400 (case), Lot Numbers: 22EBD486, 22KBW061, 23DBH163, 23CBV755, 23CBC381, 23EBN012; 19) REF CDS940857K, UDI/DI 10195327247553 (each) 40195327247554 (case), Lot Numbers: 22JBD506, 23ABC389, 23FBM896, 23EBM731; 20) REF CDS940857L, UDI/DI 10195327533311 (each) 40195327533312 (case), Lot Numbers: 23KBF380; 21) REF CDS940860P, UDI/DI 10195327009403 (each) 40195327009404 (case), Lot Numbers: 22FBS102, 22GBM011, 23CBE976, 23CBU516, 23JBH990; 22) REF CDS940860Q, UDI/DI 10195327533304 (each) 40195327533305 (case), Lot Numbers: 23KBX215; 23) REF CDS940911Q, UDI/DI 10195327349073 (each) 40195327349074 (case), Lot Numbers: 23DBD900, 23GBJ502, 23IBB454, 23CBV672, 23GBU420, 23FBR727, 23JBM716, 23JBN033, 23IBV466; 24) REF CDS940963Y, UDI/DI 10193489883671 (each) 40193489883672 (case), Lot Numbers: 22EBN044, 22EBN226, 22GBX355, 22GBR212, 23CBO762, 23BBG535, 22KBE790, 25) REF CDS941065L, UDI/DI 10193489445350 (each) 40193489445351 (case), Lot Numbers: 22KBD585, 23BBN941, 23EBL605; 26) REF CDS941167M, UDI/DI 10195327010454 (each) 40195327010455 (case), Lot Numbers: 22EBM271, 22EBL380, 23ABF515, 23ABD332, 22KBF510, 22IBT122, 23CBU256, 23JBY809, 23JBN695, 23IBC605; 27) REF CDS941167N, UDI/DI 10195327535735 (each) 40195327535736 (case), Lot Numbers: 23KBV806; 28) REF CDS980666I, UDI/DI 10193489485233 (each) 40193489485234 (case), Lot Numbers: 22FME365, 22FMC206, 22JMA036, 23GMB747, 23HMH273, 23KME164, 29) REF CDS980666J, UDI/DI 10195327568566 (each) 40195327568567 (case), Lot Numbers: 24AMF689, 24CME853, 24BMJ472, 24AMJ370, 24DMD132; 30) REF CDS980788O, UDI/DI 10193489233285 (each) 40193489233286 (case), Lot Numbers: 22JLA593, 22LLA050, 23ELA147, 23BLA703; 31) REF CDS980788P, UDI/DI 10195327384722 (each) 40195327384723 (case), Lot Numbers: 24ALA706, 23HLA176, 23GLA079, 23FLA091, 23ELA727, 23LLA549, 23LLB042, 24CLA125, 24DLA497; 32) REF CDS981314R, UDI/DI 10195327080457 (each) 40195327080458 (case), Lot Numbers: 22GMF503, 22KMC777, 22FMF044, 22FMF045, 23HME905, 23GMC775; 33) REF CDS981314S, UDI/DI 10195327539269 (each) 40195327539260 (case), Lot Numbers: 24AMD136; 34) REF CDS981318T, UDI/DI 10195327080174 (each) 40195327080175 (case), Lot Numbers: 23AMD897, 23JMC381; 35) REF CDS981318U, UDI/DI 10195327539245 (each) 40195327539246 (case), Lot Numbers: 23LMA922; 36) REF CDS981320V, UDI/DI 10195327080105 (each) 40195327080106 (case), Lot Numbers: 22GMC259, 22FMI088, 23EMD881, 23EMB534, 23AMC198, 23AME227, 23GME223, 23HMB373, 23KME353; 37) REF CDS981320W, UDI/DI 10195327539306 (each) 40195327539307 (case), Lot Numbers: 23LMG443; 38) REF CDS981323AA, UDI/DI 10195327539313 (each) 40195327539314 (case), Lot Numbers: 24AMC457, 23LMD523; 39) REF CDS981323Y, UDI/DI 10195327080112 (each) 40195327080113 (case), Lot Numbers: 23DME426, 23EMI771, 23JME028, 23JMG482; 40) REF CDS981324T, UDI/DI 10195327080426 (each) 40195327080427 (case), Lot Numbers: 23CMC158, 23CME830; 41) REF CDS981564G, UDI/DI 10193489485226 (each) 40193489485227 (case), Lot Numbers: 22HMA210, 22KMH608, 23AME249, 22LMF194, 23GMB938, 23FME185, 23IMH127, 23IMA748, 23KMI894; 42) REF CDS981564I, UDI/DI 10195327568672 (each) 40195327568673 (case), Lot Numbers: 24BMJ063, 24EMC702, 24DMB943; 43) REF CDS981886X, UDI/DI 10195327145002 (each) 40195327145003 (case), Lot Numbers: 22EBO274, 22JBE914, 22LBF066, 23ABJ993, 23EBS630, 23EBV032, 23JBA693, 23LBB198, 24BBE467, 24CBA388, 24DBK454; 44) REF CDS981888W, UDI/DI 10195327055806 (each) 40195327055807 (case), Lot Numbers: 22DBK058; 45) REF CDS981888X, UDI/DI 10195327144760 (each) 40195327144761 (case), Lot Numbers: 22JBX645, 23DBA432, 23HBT873, 23JBR359, 24DBI356, 24CBD698, 23JBG798, 23JBG925; 46) REF CDS981926D, UDI/DI 10195327141172 (each) 40195327141173 (case), Lot Numbers: 22HBF223, 22GBW207, 22KBO460, 23CBC519, 23BBA923, 23HBP475, 23HBN990, 23EBU765, 23KBF411, 24BBB303, 24EBC197, 24CBK503, 24CBU581; 47) REF CDS982082F, UDI/DI 10193489748765 (each) 40193489748766 (case), Lot Numbers: 22EDB581, 22LDB741, 23CDC093, 23FDB555, 24ADB287, 24DDB001; 48) REF CDS982396N, UDI/DI 10193489985245 (each) 40193489985246 (case), Lot Numbers: 23EBJ391, 23ABH250, 23BBS724, 23GBG962, 23LBF687, 23KBF527, 24BBP516, 24ABE656, 24EBP535, 24EBA091; 49) REF CDS982404M, UDI/DI 10195327010362 (each) 40195327010363 (case), Lot Numbers: 22LBL557, 22JBY563, 23GBG866, 23EBH327, 23JBR515, 23IBG369, 23KBM706, 24ABA228, 24CBA071, 24CBA243, 24DBD789, 24EBF079; 50) REF CDS982406M, UDI/DI 10195327010379 (each) 40195327010370 (case), Lot Numbers: 23ABQ862, 23ABP846, 23EBR980, 23HBX924, 23IBR566, 23IBQ656, 23KBS511, 24BBS092, 24ABV912, 24DBI838; 51) REF CDS982408K, UDI/DI 10195327029159 (each) 40195327029150 (case), Lot Numbers: 22GBU424, 22FBI473, 22FBG543, 23ABC232, 22LBJ951, 22KBU540, 22KBA291, 23HBH304, 23DBD824, 23IBR319, 23EBS632, 24ABV379, 23LBK663, 23KBL500, 24CBB696, 24DBI834, 24EBO787; 52) REF CDS982823Q, UDI/DI 10195327238339 (each) 40195327238330 (case), Lot Numbers: 23ABT038, 22KBT004, 23BBK666; 53) REF CDS982927M, UDI/DI 10193489879155 (each) 40193489879156 (case), Lot Numbers: 22FBX307, 22GBU507, 22JBQ096; 54) REF CDS982928O, UDI/DI 10193489879162 (each) 40193489879163 (case), Lot Numbers: 22GBF591, 22JBI126; 55) REF CDS982929K, UDI/DI 10193489879179 (each) 40193489879170 (case), Lot Numbers: 22GBT440, 22JBH490; 56) REF CDS982931L, UDI/DI 10193489879377 (each) 40193489879378 (case), Lot Numbers: 22GBU520, 22FBG546, 22FBG547, 22GBF485, 22JBH489; 57) REF CDS982932K, UDI/DI 10193489879384 (each) 40193489879385 (case), Lot Numbers: 22GBD664, 22FBU390, 22FBG515; 58) REF CDS982943Q, UDI/DI 10193489363258 (each) 40193489363259 (case), Lot Numbers: 22FBN424; 59) REF CDS982943R, UDI/DI 10195327185268 (each) 40195327185269 (case), Lot Numbers: 23DBR765, 23EBQ503; 60) REF CDS982943S, UDI/DI 10195327432393 (each) 40195327432394 (case), Lot Numbers: 23GBJ647; 61) REF CDS982986T, UDI/DI 10195327175283 (each) 40195327175284 (case), Lot Numbers: 23DBA273, 23ABN818, 23ABJ008, 23FBF632, 23EBG702; 62) REF CDS983030V, UDI/DI 10195327151416 (each) 40195327151417 (case), Lot Numbers: 22GBT329, 22HBL176, 22KBS029; 63) REF CDS983031Q, UDI/DI 10193489848359 (each) 40193489848350 (case), Lot Numbers: 22HBS390, 22JBQ016, 23ABC508, 23EBK971; 64) REF CDS983182J, UDI/DI 10195327147266 (each) 40195327147267 (case), Lot Numbers: 22HBO659, 22GBK869, 22GBK872, 22IBE783; 65) REF CDS983182K, UDI/DI 10195327297503 (each) 40195327297504 (case), Lot Numbers: 23ABB680, 23ABB679, 23DBL294, 23DBL131, 23KBM508, 23JBO877; 66) REF CDS983188K, UDI/DI 10195327105600 (each) 40195327105601 (case), Lot Numbers: 22LBP537, 22JBE497, 22GBZ339, 22GBK865, 23DBF027, 23BBJ482, 22LBS194, 23HBJ188, 23DBG097, 23FBA995, 23IBE032; 67) REF CDS983189G, UDI/DI 10195327164485 (each) 40195327164486 (case), Lot Numbers: 23ABR014, 23JBJ120, 23JBF561, 23HBF458, 23LBQ693; 68) REF CDS983194I, UDI/DI 10193489368017 (each) 40193489368018 (case), Lot Numbers: 22EBG718; 69) REF CDS983194J, UDI/DI 10195327164614 (each) 40195327164615 (case), Lot Numbers: 22FBI875, 22GBO380; 70) REF CDS983194K, UDI/DI 10195327260828 (each) 40195327260829 (case), Lot Numbers: 22LBM727; 71) REF CDS983194L, UDI/DI 10195327379490 (each) 40195327379491 (case), Lot Numbers: 23GBF702; 72) REF CDS983195I, UDI/DI 10193489368024 (each) 40193489368025 (case), Lot Numbers: 22EBI460; 73) REF CDS983195J, UDI/DI 10195327164577 (each) 40195327164578 (case), Lot Numbers: 22GBZ464, 22FBV949, 22GBO379; 74) REF CDS983195K, UDI/DI 10195327297466 (each) 40195327297467 (case), Lot Numbers: 22LBL260; 75) REF CDS983199K, UDI/DI 10195327490140 (each) 40195327490141 (case), Lot Numbers: 23KBF439; 76) REF CDS983203J, UDI/DI 10195327164584 (each) 40195327164585 (case), Lot Numbers: 22GBZ347, 22HBA033, 22GBV925; 77) REF CDS983203K, UDI/DI 10195327260835 (each) 40195327260836 (case), Lot Numbers: 23BBG462, 23DBF658; 78) REF CDS983203L, UDI/DI 10195327379575 (each) 40195327379576 (case), Lot Numbers: 23JBO016, 23JBF453; 79) REF CDS983408G, UDI/DI 10195327150266 (each) 40195327150267 (case), Lot Numbers: 23EDA279, 23CDC647, 23JDC508, 23FDB193; 80) REF CDS983415G, UDI/DI 10195327150181 (each) 40195327150182 (case), Lot Numbers: 22KDC167, 23JDB387, 23JDA841, 23IDA942; 81) REF CDS983416D, UDI/DI 10195327150198 (each) 40195327150199 (case), Lot Numbers: 22GDA312; 82) REF CDS983416G, UDI/DI 10195327279783 (each) 40195327279784 (case), Lot Numbers: 23CDA511; 83) REF CDS983468O, UDI/DI 10195327560959 (each) 40195327560950 (case), Lot Numbers: 24DDA929, 24ADA428, 24EDA794; 84) REF CDS983476J, UDI/DI 10195327497262 (each) 40195327497263 (case), Lot Numbers: 24BDA356, 24ADA401, 23JDA222, 23KDC060, 24DDA670; 85) REF CDS983480G, UDI/DI 10195327497842 (each) 40195327497843 (case), Lot Numbers: 24ADA524, 24CDB716; 86) REF CDS983482L, UDI/DI 10195327497491 (each) 40195327497492 (case), Lot Numbers: 24DBD083, 24BBG190, 24DBD506, 23LBO753; 87) REF CDS983485L, UDI/DI 10195327493073 (each) 40195327493074 (case), Lot Numbers: 23JDB449, 23IDB676, 24ADC157, 23LDB088, 23LDA537, 24CDA450; 88) REF CDS983487K, UDI/DI 10195327493080 (each) 40195327493081 (case), Lot Numbers: 24DDC105, 24ADA398; 89) REF CDS983490M, UDI/DI 10195327493110 (each) 40195327493111 (case), Lot Numbers: 23KDA049, 23IDB775, 24BDA104; 90) REF CDS983610L, UDI/DI 10195327210151 (each) 40195327210152 (case), Lot Numbers: 23HBC818, 23DBE011, 22JBP692, 22KBL228, 23ABS787, 23ABK600; 91) REF CDS983863F, UDI/DI 10195327457600 (each) 40195327457601 (case), Lot Numbers: 23JDC333; 92) REF CDS983907K, UDI/DI 10193489402766 (each) 40193489402767 (case), Lot Numbers: 22GBG415, 22JBQ095; 93) REF CDS983907L, UDI/DI 10195327238346 (each) 40195327238347 (case), Lot Numbers: 23DBN421, 23FBA959, 23FBA960, 23ABJ737, 23ABN875, 23ABR460, 23GBV529, 23FBU350; 94) REF CDS983909O, UDI/DI 10193489402780 (each) 40193489402781 (case), Lot Numbers: 22GBG991, 22GBG385; 95) REF CDS983909P, UDI/DI 10195327238384 (each) 40195327238385 (case), Lot Numbers: 22JBB183, 23BBS793, 22LBL297, 23FBG201, 23EBD620, 23DBD683, 23CBT465, 23GBF419; 96) REF CDS983915L, UDI/DI 10193489402704 (each) 40193489402705 (case), Lot Numbers: 22DBL676, 22FBU344; 97) REF CDS983915M, UDI/DI 10195327238353 (each) 40195327238354 (case), Lot Numbers: 22KBI139, 23BBG644, 22LBT311, 23ABT196, 23DBR773; 98) REF CDS984253L, UDI/DI 10193489288513 (each) 40193489288514 (case), Lot Numbers: 22GBE919, 22DBE741, 22OBM146, 22JBR832, 23CBI600, 23DBD705; 99) REF CDS984254K, UDI/DI 10193489289091 (each) 40193489289092 (case), Lot Numbers: 22GBO260, 22KBP500; 100) REF CDS984254L, UDI/DI 10195327298159 (each) 40195327298150 (case), Lot Numbers: 23CBR791, 23EBJ573, 23ABK592, 23DBD812; 101) REF CDS984279N, UDI/DI 10195327438784 (each) 40195327438785 (case), Lot Numbers: 23GLA089; 102) REF CDS984281K, UDI/DI 10195327291525 (each) 40195327291526 (case), Lot Numbers: 23DMH792; 103) REF CDS984346Q, UDI/DI 10195327404383 (each) 40195327404384 (case), Lot Numbers: 23FBI845; 104) REF CDS984459N, UDI/DI 10195327029142 (each) 40195327029143 (case), Lot Numbers: 22IBJ488, 22GBV039, 23JBF444, 23CBE952, 23KBI538, 24ABN206, 23ABQ886, 23ABI080, 23ABI356, 23GBW460; 105) REF CDS984462M, UDI/DI 10195327029135 (each) 40195327029136 (case), Lot Numbers: 23HBA114, 23FBM912, 23DBB265, 23ABS777, 23ABH406, 22HBQ525, 23FBC858, 22KBU534, 22KBU490, 23HBQ503, 24ABB913, 24GBJ266, 24BBU602, 23EBI540, 24DBI839, 23HBG580; 106) REF CDS984586P, UDI/DI 10195327103378 (each) 40195327103379 (case), Lot Numbers: 22JBN244, 22GBN110, 22JBR373, 23ABL092; 107) REF CDS984597F, UDI/DI 10193489486001 (each) 40193489486002 (case), Lot Numbers: 22KMG311, 22KMD511, 23CMF681, 22LMH126, 23EMD601, 23CMA025, 23DMH793, 23FMI884, 23JMG377, 23KMJ524; 108) REF CDS984597G, UDI/DI 10195327569655 (each) 40195327569656 (case), Lot Numbers: 24DMF470, 24BMJ471, 24BMC871, 24AMH721, 24FMC508, 24DMF900; 109) REF CDS984598G, UDI/DI 10193489486131 (each) 40193489486132 (case), Lot Numbers: 22EMH742, 22JMA043, 22FMG243, 22HME734, 23CMB375, 23CMB376, 23GMF053, 23FMC757, 23KMJ523, 23IMH761, 23IMC032; 110) REF CDS984598I, UDI/DI 10195327569662 (each) 40195327569663 (case), Lot Numbers: 24AMD497, 24EMF652; 111) REF CDS984639D, UDI/DI 10193489699722 (each) 40193489699723 (case), Lot Numbers: 22EBL309, 22FBN472, 23DBD903, 23ABR103, 23CBS010; 112) REF CDS984641D, UDI/DI 10193489699739 (each) 40193489699730 (case), Lot Numbers: 22FBT396, 22FBD663, 22KBR850, 23DBN261; 113) REF CDS984641F, UDI/DI 10195327435813 (each) 40195327435814 (case), Lot Numbers: 23GBD768, 23LBO224; 114) REF CDS984694D, UDI/DI 10195327461621 (each) 40195327461622 (case), Lot Numbers: 23HMH236, 23HMF875, 23IMG658, 23KME902, 23JMG082, 23JMB329; 115) REF CDS984896I, UDI/DI 10193489983920 (each) 40193489983921 (case), Lot Numbers: 23KBD732, 23KBO191, 24BBF369; 116) REF CDS984896J, UDI/DI 10195327621032 (each) 40195327621033 (case), Lot Numbers: 24EBK430; 117) REF CDS984898J, UDI/DI 10193489981261 (each) 40193489981262 (case), Lot Numbers: 22GBZ344; 118) REF CDS984898K, UDI/DI 10195327265687 (each) 40195327265688 (case), Lot Numbers: 23KBD086, 23ABE795, 24ABH472, 24ABD440, 23HBU517; 119) REF CDS984898L, UDI/DI 10195327621049 (each) 40195327621040 (case), Lot Numbers: 24CBD710; 120) REF CDS985234P, UDI/DI 10195327128401 (each) 40195327128402 (case), Lot Numbers: 23ABS175; 121) REF CDS985235N, UDI/DI 10195327103385 (each) 40195327103386 (case), Lot Numbers: 22GBI200; 122) REF CDS985236O, UDI/DI 10195327103392 (each) 40195327103393 (case), Lot Numbers: 22IBI035, 22FBI859, 22FBK151, 22FBI858, 23ABM166; 123) REF CDS985237L, UDI/DI 10195327173920 (each) 40195327173921 (case), Lot Numbers: 22KBP663, 23ABC074; 124) REF CDS985279K, UDI/DI 10195327497460 (each) 40195327497461 (case), Lot Numbers: 23LDA420, 24CDA763; 125) REF CDS985319I, UDI/DI 10195327091545 (each) 40195327091546 (case), Lot Numbers: 22EBJ454, 22GBF348, 22GBF350, 22EBK321, 22EBI070, 23ABB768, 22JBL431, 22JBH315, 23CBJ711; 126) REF CDS985322O, UDI/DI 10195327091484 (each) 40195327091485 (case), Lot Numbers: 22HBB447, 22FBT914, 23EBE522, 23ABN848, 22LBQ609, 22LBF840, 22JBJ445, 23CBP708; 127) REF CDS985345G, UDI/DI 10195327243586 (each) 40195327243587 (case), Lot Numbers: 22JDA453, 22LDA845, 23CDC650; 128) REF CDS985345I, UDI/DI 10195327428754 (each) 40195327428755 (case), Lot Numbers: 23HDB317, 23JDC127, 23LDA893, 23KDA304, 24BDA211, 24EDA295; 129) REF CDS985346I, UDI/DI 10195327243579 (each) 40195327243570 (case), Lot Numbers: 22LDB527, 23BDB709, 23CDB801, 23FDB094; 130) REF CDS985346K, UDI/DI 10195327452230 (each) 40195327452231 (case), Lot Numbers: 23HDB810, 23KDB011, 24CDC028; 131) REF CDS985347F, UDI/DI 10195327243616 (each) 40195327243617 (case), Lot Numbers: 22KDC138, 22JDB124, 23IDB172, 23KDA093, 23LDB766, 23KDB661, 23CDC499, 23DDB392, 23EDB196, 24EDA310; 132) REF CDS985428L, UDI/DI 10195327470722 (each) 40195327470723 (case), Lot Numbers: 23HBU628, 23HBQ472; 133) REF CDS985428M, UDI/DI 10195327527259 (each) 40195327527250 (case), Lot Numbers: 23KBE466, 24ABB045, 24BBK956, 24ABV395, 24ABS126, 24DBO092; 134) REF CDS985432N, UDI/DI 10195327336127 (each) 40195327336128 (case), Lot Numbers: 23BBE587; 135) REF CDS985432O, UDI/DI 10195327382025 (each) 40195327382026 (case), Lot Numbers: 23EBP900, 23EBE298; 136) REF CDS985432Q, UDI/DI 10195327527198 (each) 40195327527199 (case), Lot Numbers: 23KBS506, 24GBJ137, 24CBN817, 24CBB250, 24DBD900; 137) REF CDS985433M, UDI/DI 10195327336134 (each) 40195327336135 (case), Lot Numbers: 23BBL088, 23EBE899, 23EBC985; 138) REF CDS985433N, UDI/DI 10195327398255 (each) 40195327398256 (case), Lot Numbers: 23FBU308, 23FBO225; 139) REF CDS985433P, UDI/DI 10195327527204 (each) 40195327527205 (case), Lot Numbers: 24DBP406; 140) REF CDS985443J, UDI/DI 10195327228101 (each) 40195327228102 (case), Lot Numbers: 22JBO938, 22JBL358, 22JBJ683, 22IBT393, 22IBJ799, 22JBU319, 23ABM747; 141) REF CDSOPM834W, UDI/DI 10193489748758 (each) 40193489748759 (case), Lot Numbers: 22FDB552, 22JDB343, 23ADB631, 23CDB828, 23JDB638, 24ADA064; 142) REF CDSOPP508AF, UDI/DI 10193489401783 (each) 40193489401784 (case), Lot Numbers: 22DBT540, 22FBN433; 143) REF CDSOPP508AG, UDI/DI 10195327145149 (each) 40195327145140 (case), Lot Numbers: 22HBO323; 144) REF DYNJ00284L, UDI/DI 10193489962970 (each) 40193489962971 (case), Lot Numbers: 22JMD793, 22HMC141, 22GMB675, 23BMB762; 145) REF DYNJ00284M, UDI/DI 10195327412036 (each) 40195327412037 (case), Lot Numbers: 23FMB532; 146) REF DYNJ0101051L, UDI/DI 10195327339067 (each) 40195327339068 (case), Lot Numbers: 23DDA067, 23IDA268, 23IDC014; 147) REF DYNJ0101067N, UDI/DI 10195327104146 (each) 40195327104147 (case), Lot Numbers: 22EMG580, 22GMD049, 23BMA253, 23DMD611, 23AMG496, 23IME763, 23JMB621; 148) REF DYNJ0101282F, UDI/DI 10195327288679 (each) 40195327288670 (case), Lot Numbers: 23ADB994, 23DDB459; 149) REF DYNJ0101282G, UDI/DI 10195327390921 (each) 40195327390922 (case), Lot Numbers: 23IDA957, 23HDA132; 150) REF DYNJ0101283F, UDI/DI 10195327288662 (each) 40195327288663 (case), Lot Numbers: 23DDB252; 151) REF DYNJ0101283G, UDI/DI 10195327391393 (each) 40195327391394 (case), Lot Numbers: 23GDA746, 23JDA056; 152) REF DYNJ0101287J, UDI/DI 10195327286484 (each) 40195327286485 (case), Lot Numbers: 23EDA927; 153) REF DYNJ0101292F, UDI/DI 10193489434040 (each) 40193489434041 (case), Lot Numbers: 22EDB214, 22EDB043; 154) REF DYNJ0101292G, UDI/DI 10195327286255 (each) 40195327286256 (case), Lot Numbers: 23JDA057, 23HDB829, 23DDA637; 155) REF DYNJ0101293F, UDI/DI 10195327286514 (each) 40195327286515 (case), Lot Numbers: 23BDB632, 23EDA019, 23JDA719; 156) REF DYNJ0101300G, UDI/DI 10195327288686 (each) 40195327288687 (case), Lot Numbers: 23BDB321, 23KDC177, 23HDC663, 24BDB589; 157) REF DYNJ0101312K, UDI/DI 10195327333102 (each) 40195327333103 (case), Lot Numbers: 23HBP291, 23HBM321, 23GBJ788; 158) REF DYNJ0101328K, UDI/DI 10195327295943 (each) 40195327295944 (case), Lot Numbers: 23AMA479; 159) REF DYNJ0101328L, UDI/DI 10195327332907 (each) 40195327332908 (case), Lot Numbers: 23JBW902; 160) REF DYNJ0101349M, UDI/DI 10195327295950 (each) 40195327295951 (case), Lot Numbers: 22LMD920; 161) REF DYNJ0101349N, UDI/DI 10195327332914 (each) 40195327332915 (case), Lot Numbers: 23IBT537, 23HMB680; 162) REF DYNJ0105943T, UDI/DI 10195327120672 (each) 40195327120673 (case), Lot Numbers: 22DMD765; 163) REF DYNJ0106852G, UDI/DI 10889942854172 (each) 40889942854173 (case), Lot Numbers: 22KBO320, 23LBH415, 23GBF928; 164) REF DYNJ0111383L, UDI/DI 10193489315714 (each) 40193489315715 (case), Lot Numbers: 23AMC168, 22EMC657, 23DME952, 23FMI326, 23HMD592; 165) REF DYNJ0122428I, UDI/DI 10889942854196 (each) 40889942854197 (case), Lot Numbers: 24DBR554; 166) REF DYNJ0160354C, UDI/DI 10195327320072 (each) 40195327320073 (case), Lot Numbers: 23BDA901; 167) REF DYNJ0190795O, UDI/DI 10193489442618 (each) 40193489442619 (case), Lot Numbers: 23BMI341, 23GMH800, 23JMJ190, 23IMH749, 23FMF312, 23LMH163, 24FME100; 168) REF DYNJ0190827M, UDI/DI 10193489427677 (each) 40193489427678 (case), Lot Numbers: 22EBJ731, 22GBR015, 23ABT829, 23ABS484, 23ABO254, 23JBV636, 23JBB839, 23IBE648, 23LBS788, 24DBA509; 169) REF DYNJ01927K, UDI/DI 10195327491567 (each) 40195327491568 (case), Lot Numbers: 23JBQ414; 170) REF DYNJ02464K, UDI/DI 10195327291129 (each) 40195327291120 (case), Lot Numbers: 22LMC143; 171) REF DYNJ02465J, UDI/DI 10195327291136 (each) 40195327291137 (case), Lot Numbers: 23AMA990, 23KMI177, 23IMF419; 172) REF DYNJ02466K, UDI/DI 10195327291037 (each) 40195327291038 (case), Lot Numbers: 23EMJ733, 23HMD136; 173) REF DYNJ02468K, UDI/DI 10195327291044 (each) 40195327291045 (case), Lot Numbers: 23AMH491, 23HME303; 174) REF DYNJ0282079T, UDI/DI 10195327088590 (each) 40195327088591 (case), Lot Numbers: 22GBT971, 22EBS078, 22KBL646, 23BBD836, 23EBV356, 23HBR885, 23FBR451, 23JBL864; 175) REF DYNJ0340302I, UDI/DI 10195327104191 (each) 40195327104192 (case), Lot Numbers: 22CDC058, 22EDB821, 23EDB166, 23FDA825; 176) REF DYNJ0340302J, UDI/DI 10195327449971 (each) 40195327449972 (case), Lot Numbers: 23GDA731; 177) REF DYNJ0341231M, UDI/DI 10193489429060 (each) 40193489429061 (case), Lot Numbers: 22EMA406, 22GMH120, 22IMA199, 23EMK175, 23BMI342, 23FMF374, 23JMJ187, 23IMH944, 23KME561, 23LMG321, 24EMB392, 24BME816, 24CMB075, 24CMI977; 178) REF DYNJ0366324O, UDI/DI 10195327510688 (each) 40195327510689 (case), Lot Numbers: 23JMH693; 179) REF DYNJ0368519U, UDI/DI 10193489772326 (each) 40193489772327 (case), Lot Numbers: 22FBM508, 22OBC596, 22DBT321; 180) REF DYNJ0368519V, UDI/DI 10195327214302 (each) 40195327214303 (case), Lot Numbers: 22HBK670, 22JBB941; 181) REF DYNJ0368519W, UDI/DI 10195327268190 (each) 40195327268191 (case), Lot Numbers: 22LBG638, 22LBG641, 23DBB518, 23EBE443; 182) REF DYNJ0368519X, UDI/DI 10195327418038 (each) 40195327418039 (case), Lot Numbers: 23FBJ313; 183) REF DYNJ0378245N, UDI/DI 10195327500672 (each) 40195327500673 (case), Lot Numbers: 23IBP545, 23JBG896; 184) REF DYNJ0378266O, UDI/DI 10195327500450 (each) 40195327500451 (case), Lot Numbers: 23JBO727; 185) REF DYNJ0378293N, UDI/DI 10195327500689 (each) 40195327500680 (case), Lot Numbers: 23IBN337, 23JBV569; 186) REF DYNJ0421428O, UDI/DI 10195327089344 (each) 40195327089345 (case), Lot Numbers: 22DMI806; 187) REF DYNJ0421449X, UDI/DI 10195327085025 (each) 40195327085026 (case), Lot Numbers: 22GME045, 22FMB786, 22EMB558, 22GMH968, 23AME130, 23CMA591, 23FMJ595; 188) REF DYNJ0421476T, UDI/DI 10195327083588 (each) 40195327083589 (case), Lot Numbers: 22GMH371, 22EMB325; 189) REF DYNJ0421476U, UDI/DI 10195327249687 (each) 40195327249688 (case), Lot Numbers: 23GMD427, 23BME541, 22LMD501, 23DMB982; 190) REF DYNJ0451527N, UDI/DI 10195327046040 (each) 40195327046041 (case), Lot Numbers: 22FBJ134, 22JBA082; 191) REF DYNJ0451527O, UDI/DI 10195327306083 (each) 40195327306084 (case), Lot Numbers: 23GBL779; 192) REF DYNJ0451527P, UDI/DI 10195327504243 (each) 40195327504244 (case), Lot Numbers: 23IBI357; 193) REF DYNJ0475125K, UDI/DI 10195327498313 (each) 40195327498314 (case), Lot Numbers: 23IBQ973; 194) REF DYNJ0484186O, UDI/DI 10195327505103 (each) 40195327505104 (case), Lot Numbers: 23JBC704; 195) REF DYNJ0487967R, UDI/DI 10889942854387 (each) 40889942854388 (case), Lot Numbers: 23LBA367, 23DBI922, 23ABN926, 23FBT419, 24CBU495; 196) REF DYNJ0488567O, UDI/DI 10195327133986 (each) 40195327133987 (case), Lot Numbers: 22EBQ674; 197) REF DYNJ0488567P, UDI/DI 10195327212926 (each) 40195327212927 (case), Lot Numbers: 23BBH294, 23CBG444, 23EBS558, 23HBP870, 23JBC892, 23KBF853; 198) REF DYNJ0488588P, UDI/DI 10195327133979 (each) 40195327133970 (case), Lot Numbers: 22EBQ413, 22EBP051, 22EBL705, 22EBC278, 22DBR608; 199) REF DYNJ05137F, UDI/DI 10193489298567 (each) 40193489298568 (case), Lot Numbers: 22GDA791, 22FDB899, 22HDA197; 200) REF DYNJ05137G, UDI/DI 10195327224042 (each) 40195327224043 (case), Lot Numbers: 22IDC103, 22IDC163, 22KDB600, 23CDB185, 23ADB946, 23HDB483, 23FDC237; 201) REF DYNJ0519273N, UDI/DI 10195327083472 (each) 40195327083473 (case), Lot Numbers: 22GMH141, 22EMI446, 22LMI802, 22JMG306, 23CMD008; 202) REF DYNJ0530895AO, UDI/DI 10193489782790 (each) 40193489782791 (case), Lot Numbers: 22CMA594, 22DME722, 22GME323, 22IMF976, 23BMD594, 23BMD312; 203) REF DYNJ0532846S, UDI/DI 10195327356781 (each) 40195327356782 (case), Lot Numbers: 23DMA332; 204) REF DYNJ0532852Q, UDI/DI 10195327356705 (each) 40195327356706 (case), Lot Numbers: 23DMA333, 23FMI190; 205) REF DYNJ0536892U, UDI/DI 10195327206239 (each) 40195327206230 (case), Lot Numbers: 23BMC358, 22HMB123, 22KME303, 23CMF965, 23EMC256; 206) REF DYNJ0536903AL, UDI/DI 10193489782783 (each) 40193489782784 (case), Lot Numbers: 22DBI585, 22DBM797, 23CBK347, 22DBH512, 22KBM157, 23DBP819, 23FBE647; 207) REF DYNJ0537284AA, UDI/DI 10195327268084 (each) 40195327268085 (case), Lot Numbers: 23ABJ706, 23DBA241; 208) REF DYNJ0537284AB, UDI/DI 10195327418151 (each) 40195327418152 (case), Lot Numbers: 23EBX477; 209) REF DYNJ0537284X, UDI/DI 10193489681062 (each) 40193489681063 (case), Lot Numbers: 22DBG361, 22DBH399, 22GBD136; 210) REF DYNJ0537284Y, UDI/DI 10195327214111 (each) 40195327214112 (case), Lot Numbers: 22JBB218, 22JBB808; 211) REF DYNJ0587115Q, UDI/DI 10889942796847 (each) 40889942796848 (case), Lot Numbers: 22CMG299, 22FMF669, 22KMH033, 23BMC363, 23BMC362; 212) REF DYNJ0618354V, UDI/DI 10889942860760 (each) 40889942860761 (case), Lot Numbers: 23BMD622; 213) REF DYNJ0646272S, UDI/DI 10889942860920 (each) 40889942860921 (case), Lot Numbers: 22JMG686, 22LMG294; 214) REF DYNJ0774643J, UDI/DI 10195327045982 (each) 40195327045983 (case), Lot Numbers: 22HBJ605, 22EBF304, 22EBF853, 23HBK421; 215) REF DYNJ0826358T, UDI/DI 10193489384314 (each) 40193489384315 (case), Lot Numbers: 23CBV754; 216) REF DYNJ0826379O, UDI/DI 10193489300932 (each) 40193489300933 (case), Lot Numbers: 22EMB415; 217) REF DYNJ0826390X, UDI/DI 10193489265668 (each) 40193489265669 (case), Lot Numbers: 23BMB094, 23BMD322, 22GMC365, 22EMB564, 22CMI377, 23IMH180, 23EMK004; 218) REF DYNJ0826417T, UDI/DI 10193489265675 (each) 40193489265676 (case), Lot Numbers: 22FMH746, 23BMC016, 22JMF901, 23CMD855, 23FMB448, 23IME089, 23FMI373, 24AMB092; 219) REF DYNJ0826438T, UDI/DI 10193489265651 (each) 40193489265652 (case), Lot Numbers: 22JMF791, 22DMH603, 22FMH747, 22LMF278, 23DMB109, 23AMG343, 23KMD955, 23IMH965; 220) REF DYNJ0826492Q, UDI/DI 10193489300949 (each) 40193489300940 (case), Lot Numbers: 22DMF131, 23DMH876, 22FMH748, 23HMG639, 23IMH966, 23JMH219; 221) REF DYNJ0842910G, UDI/DI 10195327286637 (each) 40195327286638 (case), Lot Numbers: 23EDB905, 23FDA509; 222) REF DYNJ0843009F, UDI/DI 10195327288709 (each) 40195327288700 (case), Lot Numbers: 23JDA475; 223) REF DYNJ0843185G, UDI/DI 10195327288631 (each) 40195327288632 (case), Lot Numbers: 23GDA827; 224) REF DYNJ0843201I, UDI/DI 10195327286545 (each) 40195327286546 (case), Lot Numbers: 23DDA017, 23DDA929, 23BDB798, 23JDA011, 23GDA968; 225) REF DYNJ0843339M, UDI/DI 10193489964615 (each) 40193489964616 (case), Lot Numbers: 22EDB876, 22KDB701, 22DDA738; 226) REF DYNJ0843339N, UDI/DI 10195327286552 (each) 40195327286553 (case), Lot Numbers: 23ADA251, 23JDA990, 23IDB329, 23LDA449; 227) REF DYNJ0843648L, UDI/DI 10195327342982 (each) 40195327342983 (case), Lot Numbers: 23EDB923; 228) REF DYNJ0843648M, UDI/DI 10195327491512 (each) 40195327491513 (case), Lot Numbers: 23IDA183; 229) REF DYNJ0843840I, UDI/DI 10195327156534 (each) 40195327156535 (case), Lot Numbers: 22EDA885, 22FDA471, 22GDB104, 23ADB822; 230) REF DYNJ0843840J, UDI/DI 10195327343262 (each) 40195327343263 (case), Lot Numbers: 23FDC326, 23DDA006; 231) REF DYNJ0843840K, UDI/DI 10195327491536 (each) 40195327491537 (case), Lot Numbers: 23IDC068; 232) REF DYNJ0867555K, UDI/DI 10195327026004 (each) 40195327026005 (case), Lot Numbers: 22EMD876; 233) REF DYNJ0885263F, UDI/DI 10195327219635 (each) 40195327219636 (case), Lot Numbers: 23BMH237, 23BMB097, 22IMC316, 23FMG984; 234) REF DYNJ0949139J, UDI/DI 10195327500443 (each) 40195327500444 (case), Lot Numbers: 23IBR863; 235) REF DYNJ0952375N, UDI/DI 10195327156565 (each) 40195327156566 (case), Lot Numbers: 22EDB511, 22GDA482; 236) REF DYNJ0952375O, UDI/DI 10195327343071 (each) 40195327343072 (case), Lot Numbers: 23EDC229, 23GDC003; 237) REF DYNJ09693N, UDI/DI 10195327349196 (each) 40195327349197 (case), Lot Numbers: 23EBQ825, 23CBA158, 23GBK653, 23IBQ939, 23KBG590; 238) REF DYNJ16154G, UDI/DI 10193489486124 (each) 40193489486125 (case), Lot Numbers: 22KMB030, 23IMH968, 23AMG897, 23AMG917, 23FME243; 239) REF DYNJ17330M, UDI/DI 10195327227104 (each) 40195327227105 (case), Lot Numbers: 23FDA066, 23HDB271; 240) REF DYNJ17332M, UDI/DI 10195327227128 (each) 40195327227129 (case), Lot Numbers: 23FDC214, 23HDA042, 23JDA194; 241) REF DYNJ17451P, UDI/DI 10195327491598 (each) 40195327491599 (case), Lot Numbers: 23IBT526, 23KBH396; 242) REF DYNJ21826R, UDI/DI 10195327104412 (each) 40195327104413 (case), Lot Numbers: 22JBB065, 22JBG214, 23BBE709, 23GBA464; 243) REF DYNJ24631O, UDI/DI 10195327059385 (each) 40195327059386 (case), Lot Numbers: 23BBC196; 244) REF DYNJ24636I, UDI/DI 10193489982145 (each) 40193489982146 (case), Lot Numbers: 22FBF313, 23CBT957, 23HBP876, 23KBO981, 23JBJ725, 24DBV407; 245) REF DYNJ24638I, UDI/DI 10193489989953 (each) 40193489989954 (case), Lot Numbers: 23ABJ283, 22JBQ318, 23HBI011, 23DBB295, 23FBL170, 23JBJ640, 24DBV525, 24DBI498; 246) REF DYNJ24639J, UDI/DI 10193489989960 (each) 40193489989961 (case), Lot Numbers: 23EBI295, 23DBB296, 23JBJ726, 24BBK339; 247) REF DYNJ24715L, UDI/DI 10195327165772 (each) 40195327165773 (case), Lot Numbers: 22HMC683, 23GMI336, 24BMA798, 24CMI024, 23LMD460; 248) REF DYNJ26425L, UDI/DI 10195327156985 (each) 40195327156986 (case), Lot Numbers: 23ADA252, 23CDA966, 24BDA884, 23IDB380, 23JDB892, 24BDA961; 249) REF DYNJ28262K, UDI/DI 10195327157012 (each) 40195327157013 (case), Lot Numbers: 22GMD387, 23FMA009, 23AMA810, 23HMC742, 24CMI248; 250) REF DYNJ30224M, UDI/DI 10195327155032 (each) 40195327155033 (case), Lot Numbers: 22EBK924, 22JBH673, 22KBI794, 22LBN714, 23EBU149; 251) REF DYNJ30227K, UDI/DI 10195327155018 (each) 40195327155019 (case), Lot Numbers: 23BBL575, 22LBN755, 22KBH659, 23HBN310; 252) REF DYNJ30802B, UDI/DI 10193489917932 (each) 40193489917933 (case), Lot Numbers: 23DBD087; 253) REF DYNJ31296I, UDI/DI 10195327337681 (each) 40195327337682 (case), Lot Numbers: 23GBP372, 23KBJ324, 23HBW677, 23JBI597, 24BBS906; 254) REF DYNJ31297I, UDI/DI 10195327011932 (each) 40195327011933 (case), Lot Numbers: 22LBN785; 255) REF DYNJ31297J, UDI/DI 10195327337674 (each) 40195327337675 (case), Lot Numbers: 23HBT707, 23JBS439, 24DBH633; 256) REF DYNJ31382D, UDI/DI 10195327141189 (each) 40195327141180 (case), Lot Numbers: 22GMF907, 22FMF594, 23EMF680, 23KBE540, 23JBL300, 23FMG919, 24CBF002; 257) REF DYNJ32629K, UDI/DI 10195327527037 (each) 40195327527038 (case), Lot Numbers: 23KMF693; 258) REF DYNJ33594G, UDI/DI 10195327469573 (each) 40195327469574 (case), Lot Numbers: 23IDC076; 259) REF DYNJ33671G, UDI/DI 10195327469603 (each) 40195327469604 (case), Lot Numbers: 23JDB275; 260) REF DYNJ34420L, UDI/DI 10193489969290 (each) 40193489969291 (case), Lot Numbers: 22FBP867, 22CBX861, 22DBE974; 261) REF DYNJ34420M, UDI/DI 10195327213787 (each) 40195327213788 (case), Lot Numbers: 23ABJ177, 22KBS965, 22IBJ525, 22GBT442, 23CBC266, 23DBC094; 262) REF DYNJ34420N, UDI/DI 10195327384388 (each) 40195327384389 (case), Lot Numbers: 23EBE625; 263) REF DYNJ34428N, UDI/DI 10193489969306 (each) 40193489969307 (case), Lot Numbers: 22KMA354, 22FME407, 22DMG673, 22CMG297, 23CMD488, 23AMI797; 264) REF DYNJ34428O, UDI/DI 10195327384333 (each) 40195327384334 (case), Lot Numbers: 23FMB966; 265) REF DYNJ35432G, UDI/DI 10195327469610 (each) 40195327469611 (case), Lot Numbers: 23GDB898; 266) REF DYNJ35546I, UDI/DI 10195327100346 (each) 40195327100347 (case), Lot Numbers: 22GMD907, 22FMF899, 23IMG957, 23IMB627, 23JMB935; 267) REF DYNJ35835F, UDI/DI 10193489970593 (each) 40193489970594 (case), Lot Numbers: 22LMC292, 22IMA391, 23AMD112, 23DME157, 23BMI107; 268) REF DYNJ35835G, UDI/DI 10195327412005 (each) 40195327412006 (case), Lot Numbers: 23FMA390; 269) REF DYNJ37617I, UDI/DI 10195327497484 (each) 40195327497485 (case), Lot Numbers: 24ADA010, 23JDC007, 23IDB240; 270) REF DYNJ38234F, UDI/DI 10195327227920 (each) 40195327227921 (case), Lot Numbers: 22KBS839, 22KBI198, 22JBF056; 271) REF DYNJ39434J, UDI/DI 10195327360139 (each) 40195327360130 (case), Lot Numbers: 23GMH694, 23JMB538, 23CMG472 23FMG021; 272) REF DYNJ39434K, UDI/DI 10195327526658 (each) 40195327526659 (case), Lot Numbers: 24EMB145, 24AME049; 273) REF DYNJ39437J, UDI/DI 10195327360221 (each) 40195327360222 (case), Lot Numbers: 23CMG203, 23GMG246, 23EMK016, 23HMD212, 23IMF367; 274) REF DYNJ39437K, UDI/DI 10195327526603 (each) 40195327526604 (case), Lot Numbers: 23KMI284, 24BMD691, 24AMD988; 275) REF DYNJ39438I, UDI/DI 10195327360153 (each) 40195327360154 (case), Lot Numbers: 23JMB446, 23CME727, 23EMA386, 23HMC487, 23IMD476, 23EMK017; 276) REF DYNJ39438J, UDI/DI 10195327526627 (each) 40195327526628 (case), Lot Numbers: 24EME404, 23KMG648, 24BMD113, 24CMA721; 277) REF DYNJ40330J, UDI/DI 10195327109141 (each) 40195327109142 (case), Lot Numbers: 22FBQ740, 22FBB791, 22EBB074, 22HBG874, 23CBV711; 278) REF DYNJ40605D, UDI/DI 10193489699746 (each) 40193489699747 (case), Lot Numbers: 22JBG481, 23CBH487, 23EBT505; 279) REF DYNJ40605F, UDI/DI 10195327435837 (each) 40195327435838 (case), Lot Numbers: 23JBR578; 280) REF DYNJ40606D, UDI/DI 10193489699647 (each) 40193489699648 (case), Lot Numbers: 22DBS696, 22FBD687, 23CBU826, 23EBS919; 281) REF DYNJ40611F, UDI/DI 10195327431624 (each) 40195327431625 (case), Lot Numbers: 23IBU427; 282) REF DYNJ40982D, UDI/DI 10889942564002 (each) 40889942564003 (case), Lot Numbers: 22KBP840, 22OBJ515, 23ABM257, 22IBC523, 22FBM880, 22JBO605, 23CBC833, 23EBE098; 283) REF DYNJ40995B, UDI/DI 10889942431236 (each) 40889942431237 (case), Lot Numbers: 22DBQ804, 22DBP068, 22FBC847, 22KBS967, 22LBN471, 23DBM132, 23CBX018; 284) REF DYNJ41030K, UDI/DI 10195327156763 (each) 40195327156764 (case), Lot Numbers: 22IMH451, 22HME079, 22IMG229, 22KMF345, 22GMA588, 22JMH238, 23EMA396, 23CMG600, 23HME100, 23KMD613, 23JMB121, 23IMH192, 24FMI644, 24BME267, 23FME845, 24DMF361; 285) REF DYNJ41031J, UDI/DI 10195327156770 (each) 40195327156771 (case), Lot Numbers: 22IMG230, 22IMA227, 22JMA832, 22KMF346, 22GMI022, 23CME629, 22JMG797, 23IMF440, 23FMF122, 23EMI560, 23IMH981, 24AME178A, 24BMA779, 24AME178, 24BMA020, 24BMI693; 286) REF DYNJ41416G, UDI/DI 10195327220976 (each) 40195327220977 (case), Lot Numbers: 22LBF253, 22LBD615, 23CBF856, 23BBB377, 23DBR926; 287) REF DYNJ42734B, UDI/DI 10193489877311 (each) 40193489877312 (case), Lot Numbers: 22KMA648, 22IME750; 288) REF DYNJ42793D, UDI/DI 10193489427547 (each) 40193489427548 (case), Lot Numbers: 22DMH374, 22CMF261, 23BMD338, 22IMI373, 23AMF364, 23EMB857; 289) REF DYNJ43142D, UDI/DI 10195327431617 (each) 40195327431618 (case), Lot Numbers: 23IBM984, 23FBL003; 290) REF DYNJ43208G, UDI/DI 10193489953053 (each) 40193489953054 (case), Lot Numbers: 22EME372, 22KMH677, 23BMA133; 291) REF DYNJ43215I, UDI/DI 10193489953084 (each) 40193489953085 (case), Lot Numbers: 22EMH454, 22GMB007, 23AMD774; 292) REF DYNJ43216J, UDI/DI 10193489953091 (each) 40193489953092 (case), Lot Numbers: 22FMF377, 22GMG333, 22GMG494, 23AMJ532; 293) REF DYNJ43218F, UDI/DI 10193489953107 (each) 40193489953108 (case), Lot Numbers: 23AMH390, 22LME703, 22IMD857, 23FMA260; 294) REF DYNJ43220G, UDI/DI 10193489953114 (each) 40193489953115 (case), Lot Numbers: 22HMA07023CMF759; 295) REF DYNJ43225G, UDI/DI 10193489958874 (each) 40193489958875 (case), Lot Numbers: 22FMH990, 22EMF784, 22JMI242, 23AME669, 22HMG986, 23CMB185, 23FMI388; 296) REF DYNJ43226I, UDI/DI 10193489953039 (each) 40193489953030 (case), Lot Numbers: 22EMA090, 22GME084, 23AMI071, 22JMG344, 23CMB129, 23CMF288, 23EME963; 297) REF DYNJ43768G, UDI/DI 10193489981797 (each) 40193489981798 (case), Lot Numbers: 22FBJ135, 22GBI865; 298) REF DYNJ43768I, UDI/DI 10195327249175 (each) 40195327249176 (case), Lot Numbers: 22IDB663, 23JDB276, 23KDC250, 23HDB344, 24ADB738, 24CDA125; 299) REF DYNJ43780F, UDI/DI 10193489990140 (each) 40193489990141 (case), Lot Numbers: 23ABL238, 23CBH008, 22JBT939, 23KBE894, 23JBJ733, 23DBP912, 23KBV860, 23HBX578, 24BBJ402, 24CBA721, 24DBQ766; 300) REF DYNJ43781I, UDI/DI 10193489989687 (each) 40193489989688 (case), Lot Numbers: 22DDB534, 23GDB334, 23JDA572, 23HDC670, 23CDA576, 23ADB464, 23ADB505, 24BDA892, 23CDC640; 301) REF DYNJ43785I, UDI/DI 10193489989908 (each) 40193489989909 (case), Lot Numbers: 22GBF626, 22JBA976, 22LBK921, 23ABV318, 22JBK138; 302) REF DYNJ43785J, UDI/DI 10195327325732 (each) 40195327325733 (case), Lot Numbers: 23CBH105, 23CBW026, 23IBP045, 23GBH237, 23JBT551, 23HBP233, 24BBG167, 23LBE280, 24ABJ751, 24CBK842, 24DBV432; 303) REF DYNJ43803B, UDI/DI 10195327243067 (each) 40195327243068 (case), Lot Numbers: 22IMF487, 23IMH444, 23LMC448, 23IME778, 23EMJ046; 304) REF DYNJ43852F, UDI/DI 10195327200541 (each) 40195327200542 (case), Lot Numbers: 23CBU837, 22LBP910, 22LBK345, 23FBK894, 23EBQ613, 23IBT917, 23JBG171; 305) REF DYNJ44001C, UDI/DI 10193489326406 (each) 40193489326407 (case), Lot Numbers: 22FBI040, 22KBL649, 23DBM945, 23JBW572A, 23KBE407, 23JBW572, 23LBL113, 24CBJ890; 306) REF DYNJ44050F, UDI/DI 10193489989915 (each) 40193489989916 (case), Lot Numbers: 22IBV694, 22IBC526, 22KBJ970, 23ABG949, 23DBB309, 23DBM689, 23HBS968, 23GBB813, 23KBG864, 23FBS727, 24ABR380, 24BBK330; 307) REF DYNJ44152I, UDI/DI 10195327412616 (each) 40195327412617 (case), Lot Numbers: 23IBU434, 23GBO640, 23FBL402, 23EBN414; 308) REF DYNJ44166I, UDI/DI 10195327412999 (each) 40195327412990 (case), Lot Numbers: 23IBP289, 23HBR348; 309) REF DYNJ44168G, UDI/DI 10195327412975 (each) 40195327412976 (case), Lot Numbers: 23JBG033, 23HBK346, 23KBK728; 310) REF DYNJ44171I, UDI/DI 10195327413057 (each) 40195327413058 (case), Lot Numbers: 23IBW172, 23FBA570; 311) REF DYNJ44411G, UDI/DI 10195327412678 (each) 40195327412679 (case), Lot Numbers: 23KBG483; 312) REF DYNJ44681J, UDI/DI 10195327130886 (each) 40195327130887 (case), Lot Numbers: 22LDA584, 23BDA484; 313) REF DYNJ44685L, UDI/DI 10195327127404 (each) 40195327127405 (case), Lot Numbers: 22EBU372, 23BBD629, 23EBA774; 314) REF DYNJ44687L, UDI/DI 10195327127411 (each) 40195327127412 (case), Lot Numbers: 23BMC549, 23FMB860; 315) REF DYNJ44688I, UDI/DI 10195327127374 (each) 40195327127375 (case), Lot Numbers: 22DBQ406, 22KBH121, 23BBO707, 23CBS467; 316) REF DYNJ44692F, UDI/DI 10193489931099 (each) 40193489931090 (case), Lot Numbers: 22HDB559; 317) REF DYNJ44692G, UDI/DI 10195327277574 (each) 40195327277575 (case), Lot Numbers: 22LDB435, 22KDA921, 23ADA839; 318) REF DYNJ44765G, UDI/DI 10195327041595 (each) 40195327041596 (case), Lot Numbers: 22FBT048, 22LBK333, 23BBT899; 319) REF DYNJ44849F, UDI/DI 10195327127572 (each) 40195327127573 (case), Lot Numbers: 22KDA717, 23CDA837; 320) REF DYNJ44860I, UDI/DI 10195327127343 (each) 40195327127344 (case), Lot Numbers: 22KDB967, 22KDB783, 23DDA157; 321) REF DYNJ44878D, UDI/DI 10195327127701 (each) 40195327127702 (case), Lot Numbers: 22IDA448, 22GDB722, 22FDA133, 22KDA718, 23ADB011, 23CDC301, 23BDA570; 322) REF DYNJ44884F, UDI/DI 10195327127732 (each) 40195327127733 (case), Lot Numbers: 22FDA348, 22KDB140, 22IDA333, 22HDB242, 22GDB723; 323) REF DYNJ44884G, UDI/DI 10195327298425 (each) 40195327298426 (case), Lot Numbers: 22LDC049, 23BDA921, 23DDA265; 324) REF DYNJ44915M, UDI/DI 10195327298555 (each) 40195327298556 (case), Lot Numbers: 23BDB644; 325) REF DYNJ44915O, UDI/DI 10195327406561 (each) 40195327406562 (case), Lot Numbers: 23EDC394, 23HDA181, 23IDC127; 326) REF DYNJ44972N, UDI/DI 10195327381059 (each) 40195327381050 (case), Lot Numbers: 23HBG262, 23GBT305, 23EBP155, 23EBA093, 23JBY309, 23IBO662; 327) REF DYNJ45332G, UDI/DI 10195327242534 (each) 40195327242535 (case), Lot Numbers: 22KMA935, 22IMF489, 23DMB597, 23BMI199, 23CME732, 23EMF684; 328) REF DYNJ45579B, UDI/DI 10195327059170 (each) 40195327059171 (case), Lot Numbers: 22FBD961, 23DBJ428, 23DBB520, 22LBR094, 23BBD390, 23GBF698, 23EBQ227, 23EBQ405, 23JBQ936, 23LBB064, 24BBN755, 23IBA342, 24EBC381; 329) REF DYNJ46360O, UDI/DI 10195327110215 (each) 40195327110216 (case), Lot Numbers: 23AMA827; 330) REF DYNJ46380N, UDI/DI 10195327110222 (each) 40195327110223 (case), Lot Numbers: 22FMI359, 22EMF968, 22LMG147, 23DMC396; 331) REF DYNJ46380O, UDI/DI 10195327441609 (each) 40195327441600 (case), Lot Numbers: 23GMD060, 23HME464, 23HMG224; 332) REF DYNJ46676F, UDI/DI 10193489887617 (each) 40193489887618 (case), Lot Numbers: 23DDA493, 23FDA402, 22FDC031, 22JDC182, 22LDA136, 23CDB470, 23BDB476; 333) REF DYNJ46689J, UDI/DI 10193489887600 (each) 40193489887601 (case), Lot Numbers: 22CDB871, 22DDC332, 22FDA846, 22GDC191; 334) REF DYNJ46689K, UDI/DI 10195327224059 (each) 40195327224050 (case), Lot Numbers: 23CDC023, 23BDA845, 22JDB062, 22LDB236, 23EDB229, 23FDC371; 335) REF DYNJ46698G, UDI/DI 10195327224066 (each) 40195327224067 (case), Lot Numbers: 22IDB648, 22KDA680, 23CDB321, 23BDA774, 22LDC233, 23DDA884; 336) REF DYNJ47012O, UDI/DI 10195327072209 (each) 40195327072200 (case), Lot Numbers: 23CBW027, 23CBF864; 337) REF DYNJ47176L, UDI/DI 10193489909289 (each) 40193489909280 (case), Lot Numbers: 22JBT414, 23JBK941; 338) REF DYNJ47701I, UDI/DI 10195327357566 (each) 40195327357567 (case), Lot Numbers: 23DBK180, 23EBB152; 339) REF DYNJ47752G, UDI/DI 10195327269968 (each) 40195327269969 (case), Lot Numbers: 23EMI287; 340) REF DYNJ48084D, UDI/DI 10195327019419 (each) 40195327019410 (case), Lot Numbers: 22DMB406, 23AMJ103, 22LME957, 22LMI847, 22JMG004, 23AMH881; 341) REF DYNJ48090F, UDI/DI 10195327019686 (each) 40195327019687 (case), Lot Numbers: 23CMC256, 22DMH446, 22CMH255, 22LMG790, 23AMG372, 23BMC408; 342) REF DYNJ48091F, UDI/DI 10195327019693 (each) 40195327019694 (case), Lot Numbers: 22JMH937, 22IMB869, 23BMD482, 23FMA051; 343) REF DYNJ48130J, UDI/DI 10195327209568 (each) 40195327209569 (case), Lot Numbers: 22KBR701; 344) REF DYNJ48130K, UDI/DI 10195327349448 (each) 40195327349449 (case), Lot Numbers: 23CBE710; 345) REF DYNJ48131J, UDI/DI 10193489848069 (each) 40193489848060 (case), Lot Numbers: 22EBR840, 22HBV293; 346) REF DYNJ48259D, UDI/DI 10193489425338 (each) 40193489425339 (case), Lot Numbers: 22JMF417, 22LMD282, 23HMB682, 23EMF966, 23IMG826, 24BBA279; 347) REF DYNJ49315G, UDI/DI 10195327495718 (each) 40195327495719 (case), Lot Numbers: 23IDC243, 23KDC461, 24CDA382; 348) REF DYNJ49667K, UDI/DI 10195327059422 (each) 40195327059423 (case), Lot Numbers: 22FBX139, 22LBL766, 23CBH006, 23BBL426, 23CBU843, 22JBQ877; 349) REF DYNJ50062M, UDI/DI 10195327059361 (each) 40195327059362 (case), Lot Numbers: 22HBC896, 23ABN437; 350) REF DYNJ50072F, UDI/DI 10193489340921 (each) 40193489340922 (case), Lot Numbers: 22KMC281, 23BMH293, 23CMC263, 23KMB691, 23EMB984, 23HMG455, 24AMI546; 351) REF DYNJ50127C, UDI/DI 10889942916993 (each) 40889942916994 (case), Lot Numbers: 22DBD577; 352) REF DYNJ50621K, UDI/DI 10195327002756 (each) 40195327002757 (case), Lot Numbers: 22CME958, 23BMG900, 22EMF160, 23AME835, 22HMB251, 22JMH260, 22JMG614, 22JMI402, 23BMG421, 23HMG764, 23FMG924; 353) REF DYNJ50622J, UDI/DI 10193489922943 (each) 40193489922944 (case), Lot Numbers: 22GMF307, 22CMD580, 23DMI638, 22JMI258, 22IMC803, 22HMH756, 23DMA353, 23HMH918; 354) REF DYNJ50665N, UDI/DI 10195327542733 (each) 40195327542734 (case), Lot Numbers: 23KBE170, 23LBI160, 24ABK459; 355) REF DYNJ50666M, UDI/DI 10195327600983 (each) 40195327600984 (case), Lot Numbers: 24BBN662, 24EBE931; 356) REF DYNJ50884C, UDI/DI 10195327179519 (each) 40195327179510 (case), Lot Numbers: 23AMF855, 23CMC475, 23BMA394, 23EMJ710, 23EME118, 24DMD645; 357) REF DYNJ50994I, UDI/DI 10195327088637 (each) 40195327088638 (case), Lot Numbers: 23ABR305, 22EBG897, 22HBE747, 22GBV082, 22KBD063, 23HBC086, 23JBB233, 23LBE846, 23FBR628; 358) REF DYNJ51149R, UDI/DI 10195327127015 (each) 40195327127016 (case), Lot Numbers: 23CBA582; 359) REF DYNJ51149S, UDI/DI 10195327510374 (each) 40195327510375 (case), Lot Numbers: 23KBF450; 360) REF DYNJ51150O, UDI/DI 10195327127022 (each) 40195327127023 (case), Lot Numbers: 22FBY642, 22FBF024, 23ABL412, 23ABL413, 23GBL833, 23FBC091, 23HBQ398; 361) REF DYNJ51150P, UDI/DI 10195327510404 (each) 40195327510405 (case), Lot Numbers: 23JBE656, 23JBW291; 362) REF DYNJ51320L, UDI/DI 10193489841374 (each) 40193489841375 (case), Lot Numbers: 22FBW403, 22DBT710, 22JBO825, 23KBF216; 363) REF DYNJ51802J, UDI/DI 10195327475079 (each) 40195327475070 (case), Lot Numbers: 23GDC154, 23IDA304; 364) REF DYNJ51803I, UDI/DI 10195327475062 (each) 40195327475063 (case), Lot Numbers: 23GDC144, 23HDA736; 365) REF DYNJ52007C, UDI/DI 10889942796878 (each) 40889942796879 (case), Lot Numbers: 22CMG176, 23EMA071, 22DMF070, 22JME038, 22JME069, 23CMF980, 22LMI347, 22LMI350, 23BMC588; 366) REF DYNJ52331F, UDI/DI 10195327237691 (each) 40195327237692 (case), Lot Numbers: 23EBI999, 22LBE306, 22KBQ085, 22IBP287, 23DBK756, 23FBJ664; 367) REF DYNJ52332G, UDI/DI 10195327237707 (each) 40195327237708 (case), Lot Numbers: 22IBE036, 22HBV289, 22JBQ833, 23CBH602; 368) REF DYNJ52333D, UDI/DI 10195327237714 (each) 40195327237715 (case), Lot Numbers: 22HBV378, 22JBP714, 22JBP717, 22LBD328, 23BBH346, 23ABO549, 23CBA576; 369) REF DYNJ52398N, UDI/DI 10195327339104 (each) 40195327339105 (case), Lot Numbers: 23FDA686, 23JDA047, 23HDA446; 370) REF DYNJ52405I, UDI/DI 10195327222628 (each) 40195327222629 (case), Lot Numbers: 22JDC201; 371) REF DYNJ52405J, UDI/DI 10195327339166 (each) 40195327339167 (case), Lot Numbers: 23GDB229, 23HDB575, 23DDA852, 23FDA445, 23JDA254; 372) REF DYNJ52415L, UDI/DI 10195327339234 (each) 40195327339235 (case), Lot Numbers: 23GDA970, 23EDC509; 373) REF DYNJ52418J, UDI/DI 10195327222697 (each) 40195327222698 (case), Lot Numbers: 22KDA708; 374) REF DYNJ52418K, UDI/DI 10195327339258 (each) 40195327339259 (case), Lot Numbers: 23DDA689, 23IDC320, 23HDB839; 375) REF DYNJ52420J, UDI/DI 10195327222710 (each) 40195327222711 (case), Lot Numbers: 22KDA303; 376) REF DYNJ52420K, UDI/DI 10195327339272 (each) 40195327339273 (case), Lot Numbers: 23CDA077, 23CDC516, 23HDC167, 23JDA243; 377) REF DYNJ52422F, UDI/DI 10195327222734 (each) 40195327222735 (case), Lot Numbers: 22JDC148; 378) REF DYNJ52422G, UDI/DI 10195327339296 (each) 40195327339297 (case), Lot Numbers: 23HDA050; 379) REF DYNJ52423K, UDI/DI 10195327339302 (each) 40195327339303 (case), Lot Numbers: 23GDA262, 23FDA281, 23IDC072; 380) REF DYNJ52427J, UDI/DI 10195327339340 (each) 40195327339341 (case), Lot Numbers: 23DDA942, 23IDA396; 381) REF DYNJ52430K, UDI/DI 10195327339593 (each) 40195327339594 (case), Lot Numbers: 23DDB533, 23GDA993, 23JDA099; 382) REF DYNJ52431K, UDI/DI 10195327222802 (each) 40195327222803 (case), Lot Numbers: 22KDA296; 383) REF DYNJ52431L, UDI/DI 10195327339609 (each) 40195327339600 (case), Lot Numbers: 23BDA248, 23DDA589, 23EDB819, 23GDA329, 24CDA929, 23IDA827, 23JDA949; 384) REF DYNJ52436K, UDI/DI 10195327222826 (each) 40195327222827 (case), Lot Numbers: 22HDA788; 385) REF DYNJ52436L, UDI/DI 10195327339647 (each) 40195327339648 (case), Lot Numbers: 23CDC601, 23GDC227; 386) REF DYNJ52628B, UDI/DI 10889942872060 (each) 40889942872061 (case), Lot Numbers: 22JDA509, 22GDB458, 23EDA896, 23ADA462; 387) REF DYNJ52638C, UDI/DI 10195327088644 (each) 40195327088645 (case), Lot Numbers: 22KBL650, 23HBF875, 23KBK761, 23KBB235, 23EBE717, 23EBV500, 23FBM239; 388) REF DYNJ53006C, UDI/DI 10193489782745 (each) 40193489782746 (case), Lot Numbers: 22CMD559, 22DMF577, 23BMB746, 23AMJ124, 22HME683, 22GMH417, 22IMI055, 22IMI057; 389) REF DYNJ53322G, UDI/DI 10195327360214 (each) 40195327360215 (case), Lot Numbers: 23CMH443, 23IME654, 23CME925, 23EMG576, 23FMG819; 390) REF DYNJ53322I, UDI/DI 10195327526573 (each) 40195327526574 (case), Lot Numbers: 23KMD822, 23JME049, 24CMI637, 24BMI474, 24AME052; 391) REF DYNJ53555F, UDI/DI 10195327410056 (each) 40195327410057 (case), Lot Numbers: 23EBI165, 23FBL406; 392) REF DYNJ53652C, UDI/DI 10195327012137 (each) 40195327012138 (case), Lot Numbers: 22ELA688, 23ALA032, 22KLA300, 22JLA010; 393) REF DYNJ53652D, UDI/DI 10195327337650 (each) 40195327337651 (case), Lot Numbers: 23CLA577, 23BLB064, 23BLA460, 23KLA826, 23DLA361, 23ELA054, 23JLB077, 24BLA326, 24DLA706, 24DLA103; 394) REF DYNJ53682B, UDI/DI 10193489883084 (each) 40193489883085 (case), Lot Numbers: 23EDB818; 395) REF DYNJ53687B, UDI/DI 10193489886511 (each) 40193489886512 (case), Lot Numbers: 22HDA058; 396) REF DYNJ54149D, UDI/DI 10195327107284 (each) 40195327107285 (case), Lot Numbers: 22DMF213; 397) REF DYNJ54149F, UDI/DI 10195327292614 (each) 40195327292615 (case), Lot Numbers: 22LMD587, 23JMA188, 23IMD234, 23KMI574; 398) REF DYNJ54462C, UDI/DI 10195327075040 (each) 40195327075041 (case), Lot Numbers: 22EDB554, 23DDB397, 23JDA699, 23JDA083; 399) REF DYNJ54463F, UDI/DI 10195327155704 (each) 40195327155705 (case), Lot Numbers: 23CME386, 22EMD530, 22DMI914, 23JMH897, 23HMF144; 400) REF DYNJ54464G, UDI/DI 10195327155711 (each) 40195327155712 (case), Lot Numbers: 22FMB906, 22FMB839, 22EMA876, 23DMD917; 401) REF DYNJ54934B, UDI/DI 10195327344399 (each) 40195327344390 (case), Lot Numbers: 23BDA904; 402) REF DYNJ55049D, UDI/DI 10195327145705 (each) 40195327145706 (case), Lot Numbers: 22JMC226, 22EMD458; 403) REF DYNJ55667D, UDI/DI 10195327281939 (each) 40195327281930 (case), Lot Numbers: 23BBP785, 23ABN999, 23CBL665, 23IBN555, 23HBQ079, 23JBY358, 23JBY654, 23JBD701; 404) REF DYNJ55733G, UDI/DI 10193489454659 (each) 40193489454650 (case), Lot Numbers: 22JBU639, 23GBJ474, 23CBX532, 23BBE511, 22HBY185; 405) REF DYNJ55851A, UDI/DI 10193489985030 (each) 40193489985031 (case), Lot Numbers: 22CMF878, 22JMA855, 22EMI149; 406) REF DYNJ55852B, UDI/DI 10193489985016 (each) 40193489985017 (case), Lot Numbers: 22IMB760; 407) REF DYNJ55852C, UDI/DI 10195327267575 (each) 40195327267576 (case), Lot Numbers: 23BMH431, 23EMI576; 408) REF DYNJ56070D, UDI/DI 10193489773026 (each) 40193489773027 (case), Lot Numbers: 22EMG162, 22JMB598, 23AMA086, 23BMA752, 23KMD985, 23FMF245, 23EMI181, 23LBV444, 23IMH211, 24BBG378, 24DBA267; 409) REF DYNJ56952C, UDI/DI 10195327485917 (each) 40195327485918 (case), Lot Numbers: 23JMH674, 23LME884, 24EMB237; 410) REF DYNJ57113C, UDI/DI 10195327074630 (each) 40195327074631 (case), Lot Numbers: 22HMI210, 23IBU927, 23HBP819; 411) REF DYNJ57546B, UDI/DI 10195327124663 (each) 40195327124664 (case), Lot Numbers: 22FMD947; 412) REF DYNJ57546C, UDI/DI 10195327216580 (each) 40195327216581 (case), Lot Numbers: 22LMB359, 22KME770, 22HMG755, 23CMD289, 23BMC613; 413) REF DYNJ57648D, UDI/DI 10195327027391 (each) 40195327027392 (case), Lot Numbers: 22EMA228, 22FMI369, 23HMG727, 23FMB585, 23AMH213, 22JMA245, 23FMJ099; 414) REF DYNJ57841C, UDI/DI 10195327488581 (each) 40195327488582 (case), Lot Numbers: 23HBR517, 23IBL726; 415) REF DYNJ57883C, UDI/DI 10195327046019 (each) 40195327046010 (case), Lot Numbers: 22DBS083; 416) REF DYNJ57883D, UDI/DI 10195327204143 (each) 40195327204144 (case), Lot Numbers: 22HBE753, 24ABT515, 23BBM998, 23FBL672, 23CBJ367, 23JBW817, 23IBQ129; 417) REF DYNJ57971B, UDI/DI 10195327049041 (each) 40195327049042 (case), Lot Numbers: 22GBS099; 418) REF DYNJ57971C, UDI/DI 10195327225209 (each) 40195327225200 (case), Lot Numbers: 22KBM557, 22KBM558, 23ABF124, 23JBY165, 23LBD844, 23IBB085, 24ABW581, 24DBD421; 419) REF DYNJ57975D, UDI/DI 10195327225124 (each) 40195327225125 (case), Lot Numbers: 23CMH203, 23AMA728, 22KMD412, 22JMB348, 23BMD721, 23FMC375, 23HMG228, 23JMG688, 23IMF776, 23EMC530, 24AMF292; 420) REF DYNJ57975F, UDI/DI 10195327659431 (each) 40195327659432 (case), Lot Numbers: 24EMD224, 24CMH221, 24HMF645; 421) REF DYNJ58053D, UDI/DI 10195327049058 (each) 40195327049059 (case), Lot Numbers: 22GMH676, 22EMH464; 422) REF DYNJ58053F, UDI/DI 10195327225131 (each) 40195327225132 (case), Lot Numbers: 23GMI449, 23EMB074, 23FMH212, 24BMI409, 24CMB013; 423) REF DYNJ58557C, UDI/DI 10193489972276 (each) 40193489972277 (case), Lot Numbers: 22JMF230, 23AME789, 23FMC763, 23FMI232; 424) REF DYNJ58563D, UDI/DI 10193489972320 (each) 40193489972321 (case), Lot Numbers: 22CMA822, 22CMA823, 22KMH220, 23GMA044, 23FMH095, 22LMG325; 425) REF DYNJ58564D, UDI/DI 10193489972337 (each) 40193489972338 (case), Lot Numbers: 22CMC588, 22KMG860, 23AME526, 23JMB659, 23IMD727, 23KMD387; 426) REF DYNJ58631G, UDI/DI 10195327026585 (each) 40195327026586 (case), Lot Numbers: 22EDB868, 22FDB176, 23CDA513, 23HDA582, 23KDB279, 24ADB364; 427) REF DYNJ58633D, UDI/DI 10195327026592 (each) 40195327026593 (case), Lot Numbers: 22FDB011, 22JDB863, 22KDB382, 23CDA345, 23CDC561, 23DDC160, 23JDB393, 23IDB456, 23IDA932, 23KDC183, 24BDA964; 428) REF DYNJ58665F, UDI/DI 10195327529741 (each) 40195327529742 (case), Lot Numbers: 23KMB334; 429) REF DYNJ59011G, UDI/DI 10195327332648 (each) 40195327332649 (case), Lot Numbers: 23IDC046, 23GDB612; 430) REF DYNJ59025D, UDI/DI 10195327519636 (each) 40195327519637 (case), Lot Numbers: 23JBB388; 431) REF DYNJ59080C, UDI/DI 10195327200299 (each) 40195327200290 (case), Lot Numbers: 22JBM549, 23BBI615, 22LBM723, 23DBB343, 23FBK899, 23HBL384, 23HBG733, 23JBN309; 432) REF DYNJ59106C, UDI/DI 10195327519643 (each) 40195327519644 (case), Lot Numbers: 23JMB434; 433) REF DYNJ59581D, UDI/DI 10195327165765 (each) 40195327165766 (case), Lot Numbers: 22JME822, 23DMI271, 23EMH407, 23EME554, 24AMG444, 24DMB376, 24BMD084, 24BMA718; 434) REF DYNJ59819A, UDI/DI 10195327331597 (each) 40195327331598 (case), Lot Numbers: 23CBF431; 435) REF DYNJ60049B, UDI/DI 10193489782776 (each) 40193489782777 (case), Lot Numbers: 22DBH397, 22JBY793, 22LBK547, 23CBX671, 23FBU291; 436) REF DYNJ60243G, UDI/DI 10195327332884 (each) 40195327332885 (case), Lot Numbers: 23BMD070, 23IMC653; 437) REF DYNJ60546B, UDI/DI 10195327361754 (each) 40195327361755 (case), Lot Numbers: 23CBX535; 438) REF DYNJ60550B, UDI/DI 10195327361853 (each) 40195327361854 (case), Lot Numbers: 23CBQ320; 439) REF DYNJ60553B, UDI/DI 10195327361822 (each) 40195327361823 (case), Lot Numbers: 23CBX537; 440) REF DYNJ60626B, UDI/DI 10193489961904 (each) 40193489961905 (case), Lot Numbers: 22FBS949, 23ABR523, 23DBA353, 23IBS559; 441) REF DYNJ60818B, UDI/DI 10195327454661 (each) 40195327454662 (case), Lot Numbers: 23HBJ977, 23GBS227, 23IBV407; 442) REF DYNJ60818C, UDI/DI 10195327521325 (each) 40195327521326 (case), Lot Numbers: 23JBT400, 23KBJ738, 24ABA318; 443) REF DYNJ60824B, UDI/DI 10195327521288 (each) 40195327521289 (case), Lot Numbers: 23JBT425; 444) REF DYNJ60896B, UDI/DI 10195327500429 (each) 40195327500420 (case), Lot Numbers: 23IBS475, 23IBH842; 445) REF DYNJ60906A, UDI/DI 10193489835960 (each) 40193489835961 (case), Lot Numbers: 22EMH358, 22EME827, 22CMB326, 22FMC970, 22HME390, 22KBS938, 22KBM814, 23EBC437, 23DBC688; 446) REF DYNJ60906B, UDI/DI 10195327500351 (each) 40195327500352 (case), Lot Numbers: 23IBS476; 447) REF DYNJ60921C, UDI/DI 10195327470395 (each) 40195327470396 (case), Lot Numbers: 23GBT246, 23KBG585, 23JBF899; 448) REF DYNJ60930C, UDI/DI 10195327470340 (each) 40195327470341 (case), Lot Numbers: 23HBL078, 23JBL127; 449) REF DYNJ61137F, UDI/DI 10195327470449 (each) 40195327470440 (case), Lot Numbers: 23HBL077, 23IBN862, 23JBV920; 450) REF DYNJ61143F, UDI/DI 10195327470425 (each) 40195327470426 (case), Lot Numbers: 23HBQ168; 451) REF DYNJ61146F, UDI/DI 10195327470418 (each) 40195327470419 (case), Lot Numbers: 23HBM016, 23JBC547; 452) REF DYNJ61152D, UDI/DI 10195327470210 (each) 40195327470211 (case), Lot Numbers: 23GBT618, 23JBY125, 23JBE190; 453) REF DYNJ61155B, UDI/DI 10195327080846 (each) 40195327080847 (case), Lot Numbers: 22EBK970, 23CBU121, 23FBC089, 23GBL975; 454) REF DYNJ61155C, UDI/DI 10195327470487 (each) 40195327470488 (case), Lot Numbers: 23HBL129, 23KBA990; 455) REF DYNJ61156C, UDI/DI 10195327470227 (each) 40195327470228 (case), Lot Numbers: 23JBL221; 456) REF DYNJ61306A, UDI/DI 10195327072261 (each) 40195327072262 (case), Lot Numbers: 22EMA261, 23AMI588, 23EMB090, 23JBW540; 457) REF DYNJ61340A, UDI/DI 10193489873801 (each) 40193489873802 (case), Lot Numbers: 22KBO231, 22HBY295, 22HBE528, 22JBO026, 23BBM091; 458) REF DYNJ61414B, UDI/DI 10193489873900 (each) 40193489873901 (case), Lot Numbers: 22DBU513, 22HBX138, 22HBK088, 22GBA319, 22IBR399, 22LBP338, 22LBO290, 23CBE238, 23BBU334, 23DBC271; 459) REF DYNJ61500, UDI/DI 10193489574098 (each) 40193489574099 (case), Lot Numbers: 22KME391, 22JMI283, 22IMH886, 22GMH683; 460) REF DYNJ61582C, UDI/DI 10195327152161 (each) 40195327152162 (case), Lot Numbers: 22GME126; 461) REF DYNJ61651I, UDI/DI 10195327327125 (each) 40195327327126 (case), Lot Numbers: 23BMC068, 23CMD237; 462) REF DYNJ61663A, UDI/DI 10193489929652 (each) 40193489929653 (case), Lot Numbers: 22IBF170, 23CBF409, 22LBN360, 22LBA795; 463) REF DYNJ61844B, UDI/DI 10195327495954 (each) 40195327495955 (case), Lot Numbers: 23IDB452, 23KDA607, 24ADA592, 24BDC034; 464) REF DYNJ61845A, UDI/DI 10195327496050 (each) 40195327496051 (case), Lot Numbers: 23LDA497, 23JDB429, 23IDB697; 465) REF DYNJ61846A, UDI/DI 10195327496487 (each) 40195327496488 (case), Lot Numbers: 23IDB696; 466) REF DYNJ61849D, UDI/DI 10195327495558 (each) 40195327495559 (case), Lot Numbers: 23KDC327, 23JDA309, 24ADA958, 24CDB779, 24DDB411; 467) REF DYNJ61908C, UDI/DI 10195327281915 (each) 40195327281916 (case), Lot Numbers: 23JBT556, 23GBT165, 23GBM755, 23EBX530, 23HBQ082, 23HBL263; 468) REF DYNJ62031B, UDI/DI 10195327058272 (each) 40195327058273 (case), Lot Numbers: 22DBR051, 22JBW801, 23FBT974, 22LBT054, 23BBC855, 22LBS076, 23CBU872, 23IBV506, 23GBN446; 469) REF DYNJ62034C, UDI/DI 10195327627157 (each) 40195327627158 (case), Lot Numbers: 24CBR453; 470) REF DYNJ62035C, UDI/DI 10195327058258 (each) 40195327058259 (case), Lot Numbers: 22FBL674, 23DBB035, 23JBX179, 23GBP806; 471) REF DYNJ62395C, UDI/DI 10195327316617 (each) 40195327316618 (case), Lot Numbers: 23ABN929, 23EBU099, 23KBQ093; 472) REF DYNJ62395D, UDI/DI 10195327611378 (each) 40195327611379 (case), Lot Numbers: 24BBN072, 24DBT910; 473) REF DYNJ62650B, UDI/DI 10195327121846 (each) 40195327121847 (case), Lot Numbers: 22GBF339, 23BBD864, 22HBM930, 22IBK987, 23CBO687, 23EBE814, 23FBE980; 474) REF DYNJ62668C, UDI/DI 10195327546342 (each) 40195327546343 (case), Lot Numbers: 23KMC875; 475) REF DYNJ62777D, UDI/DI 10195327068028 (each) 40195327068029 (case), Lot Numbers: 22EDA852; 476) REF DYNJ63082B, UDI/DI 10195327102135 (each) 40195327102136 (case), Lot Numbers: 22CMD192, 23EMI678, 22IMA803, 22IMA805, 22LMJ344, 23CMI070; 477) REF DYNJ63090A, UDI/DI 10195327358952 (each) 40195327358953 (case), Lot Numbers: 23FBL996, 23JBX772; 478) REF DYNJ63095A, UDI/DI 10195327358990 (each) 40195327358991 (case), Lot Numbers: 23EBP508; 479) REF DYNJ63300C, UDI/DI 10195327041038 (each) 40195327041039 (case), Lot Numbers: 22FBY595, 22FBX540; 480) REF DYNJ63300D, UDI/DI 10195327275716 (each) 40195327275717 (case), Lot Numbers: 23DMD465, 23IMA358, 23IMG385, 24DMH488; 481) REF DYNJ63301B, UDI/DI 10193489896961 (each) 40193489896962 (case), Lot Numbers: 22FBC835, 22DBN079; 22GBF112; 482) REF DYNJ63301C, UDI/DI 10195327273644 (each) 40195327273645 (case), Lot Numbers: 23EMI318, 23IME153; 483) REF DYNJ63302C, UDI/DI 10195327040987 (each) 40195327040988 (case), Lot Numbers: 22EBC472, 22FBD800; 484) REF DYNJ63302D, UDI/DI 10195327275686 (each) 40195327275687 (case), Lot Numbers: 23FMI709, 23IMA802, 23IMA941; 485) REF DYNJ63303C, UDI/DI 10195327040994 (each) 40195327040995 (case), Lot Numbers: 22HMG296, 22HMG297; 486) REF DYNJ63303D, UDI/DI 10195327275723 (each) 40195327275724 (case), Lot Numbers: 23BMD048, 23IMH305, 23HMI184, 23KMB839; 487) REF DYNJ63305B, UDI/DI 10195327041052 (each) 40195327041053 (case), Lot Numbers: 22FME428; 488) REF DYNJ63306B, UDI/DI 10195327270049 (each) 40195327270040 (case), Lot Numbers: 23JMG735, 23HME655, 23GMD024, 23DMC975, 23BMC104; 489) REF DYNJ63310C, UDI/DI 10195327041168 (each) 40195327041169 (case), Lot Numbers: 22FBU999; 490) REF DYNJ63310D, UDI/DI 10195327270636 (each) 40195327270637 (case), Lot Numbers: 23DMB224, 23IMD732, 23JMG130; 491) REF DYNJ63311B, UDI/DI 10193489896978 (each) 40193489896979 (case), Lot Numbers: 22EBG812, 22EBG200; 492) REF DYNJ63311C, UDI/DI 10195327273637 (each) 40195327273638 (case), Lot Numbers: 23FMG395, 23HME647, 23IME154; 493) REF DYNJ63312C, UDI/DI 10195327270667 (each) 40195327270668 (case), Lot Numbers: 23BMA186, 23AMI101; 494) REF DYNJ63312D, UDI/DI 10195327478476 (each) 40195327478477 (case), Lot Numbers: 23IME155; 495) REF DYNJ63316D, UDI/DI 10195327273545 (each) 40195327273546 (case), Lot Numbers: 23IMA324, 22LMH495, 22LMH499; 496) REF DYNJ63318A, UDI/DI 10193489989441 (each) 40193489989442 (case), Lot Numbers: 22GBH439, 22GBB669; 497) REF DYNJ63318B, UDI/DI 10195327270063 (each) 40195327270064 (case), Lot Numbers: 23AMH442, 23GMC956, 23IMA362, 23JMF710; 498) REF DYNJ63319B, UDI/DI 10193489949391 (each) 40193489949392 (case), Lot Numbers: 22GBJ839; 499) REF DYNJ63319C, UDI/DI 10195327273651 (each) 40195327273652 (case), Lot Numbers: 23AMG001; 500) REF DYNJ63324B, UDI/DI 10193489958140 (each) 40193489958141 (case), Lot Numbers: 22EBI790, 22EBI793, 22IMG131; 501) REF DYNJ63324C, UDI/DI 10195327275761 (each) 40195327275762 (case), Lot Numbers: 23DMC973, 23DMA796; 502) REF DYNJ63324D, UDI/DI 10195327402143 (each) 40195327402144 (case), Lot Numbers: 23IMA813, 23JMG201, 23KMI808; 503) REF DYNJ63339D, UDI/DI 10195327273606 (each) 40195327273607 (case), Lot Numbers: 23HBE292; 504) REF DYNJ63349C, UDI/DI 10193489983746 (each) 40193489983747 (case), Lot Numbers: 22EBF338, 22GMA096, 22FMB860; 505) REF DYNJ63373D, UDI/DI 10195327086152 (each) 40195327086153 (case), Lot Numbers: 22FBY598; 506) REF DYNJ63373F, UDI/DI 10195327270780 (each) 40195327270781 (case), Lot Numbers: 22LBI728, 22LBI729, 23CBT768, 23FBG742, 23IBO340; 507) REF DYNJ63436A, UDI/DI 10193489252972 (each) 40193489252973 (case), Lot Numbers: 22GBJ224, 22GBP404; 508) REF DYNJ63436B, UDI/DI 10195327282240 (each) 40195327282241 (case), Lot Numbers: 22KBQ089; 509) REF DYNJ63436C, UDI/DI 10195327319380 (each) 40195327319381 (case), Lot Numbers: 23FBB921, 23HBJ572; 510) REF DYNJ63473C, UDI/DI 10195327510923 (each) 40195327510924 (case), Lot Numbers: 23IMF358, 24BMA722; 511) REF DYNJ63552B, UDI/DI 10193489900378 (each) 40193489900379 (case), Lot Numbers: 22FBF321, 23BBC141, 22LBJ155, 23BBJ095, 23EBV888, 23GBU067, 23KBI857, 23JBO055, 23IBO111; 512) REF DYNJ63557B, UDI/DI 10193489900392 (each) 40193489900393 (case), Lot Numbers: 22DBN108, 23CBM182, 23BBH750, 23HBP781; 513) REF DYNJ63558B, UDI/DI 10193489900415 (each) 40193489900416 (case), Lot Numbers: 23ABU703, 22LBN733, 23DBJ042, 23GBU068, 23IBQ589; 514) REF DYNJ63559B, UDI/DI 10193489900422 (each) 40193489900423 (case), Lot Numbers: 22DBS937, 22FBS409, 22IBG382, 23ABO512, 23IBK484, 23HBP711, 23GBU069, 23KBI346; 515) REF DYNJ63583B, UDI/DI 10195327122898 (each) 40195327122899 (case), Lot Numbers: 23GBL805; 516) REF DYNJ63604B, UDI/DI 10193489900231 (each) 40193489900232 (case), Lot Numbers: 22FBO473, 22KBH234, 22KBH617, 22KBH240, 23CBF429, 23GBP853, 23JBE719; 517) REF DYNJ63607B, UDI/DI 10193489900255 (each) 40193489900256 (case), Lot Numbers: 22FBV007, 23JBP047, 23JBA953, 23IBA206, 23GBQ019, 23EBQ718, 23CBS745; 518) REF DYNJ63609C, UDI/DI 10195327350284 (each) 40195327350285 (case), Lot Numbers: 23JBQ193, 23HBI039; 519) REF DYNJ63614B, UDI/DI 10195327350543 (each) 40195327350544 (case), Lot Numbers: 23FBE255, 23GBT158, 23CBS818, 23IBA158; 520) REF DYNJ63616B, UDI/DI 10195327350369 (each) 40195327350360 (case), Lot Numbers: 23EBT358, 23DBE423, 23GBO792, 23JBR539; 521) REF DYNJ63620B, UDI/DI 10193489900279 (each) 40193489900270 (case), Lot Numbers: 22DBS615, 22DBS700, 22DBS701; 522) REF DYNJ63639B, UDI/DI 10193489900323 (each) 40193489900324 (case), Lot Numbers: 22FBK829, 22LBP339; 523) REF DYNJ63641B, UDI/DI 10195327122805 (each) 40195327122806 (case), Lot Numbers: 22JBJ076, 22FBV009, 23EBW585, 23IBN177, 23JBC776; 524) REF DYNJ63646B, UDI/DI 10195327350352 (each) 40195327350353 (case), Lot Numbers: 23CBT773, 23JBR021; 525) REF DYNJ63653C, UDI/DI 10195327350468 (each) 40195327350469 (case), Lot Numbers: 23DBB425, 23FBF856, 23IBA351, 23GBS392, 23JBV605; 526) REF DYNJ63654A, UDI/DI 10195327350512 (each) 40195327350513 (case), Lot Numbers: 23CBC038, 23FBD023, 23GBJ801, 24ABW300, 23JBC777; 527) REF DYNJ63831C, UDI/DI 10195327041007 (each) 40195327041008 (case), Lot Numbers: 22FBB477, 22GBD802, 22FBD777; 528) REF DYNJ63831D, UDI/DI 10195327275754 (each) 40195327275755 (case), Lot Numbers: 23BMD043, 23KMF707; 529) REF DYNJ63886A, UDI/DI 10195327221836 (each) 40195327221837 (case), Lot Numbers: 23GDB254, 23KDB405, 23FDB375, 24CDB332; 530) REF DYNJ63939A, UDI/DI 10193489270006 (each) 40193489270007 (case), Lot Numbers: 22EDA758; 531) REF DYNJ63939B, UDI/DI 10195327221850 (each) 40195327221851 (case), Lot Numbers: 22IDC124, 23GDB082, 23JDA267, 23KDB911, 23KDA705, 23CDB169, 24BDA519, 23DDB707, 24CDB509; 532) REF DYNJ63942A, UDI/DI 10193489270013 (each) 40193489270014 (case), Lot Numbers: 22FDA747; 533) REF DYNJ63942B, UDI/DI 10195327221867 (each) 40195327221868 (case), Lot Numbers: 23BDA783; 534) REF DYNJ63942C, UDI/DI 10195327347192 (each) 40195327347193 (case), Lot Numbers: 23JDA396, 23CDB849, 23HDC482, 23GDA461, 23KDB369, 24ADB754; 535) REF DYNJ63957B, UDI/DI 10195327364939 (each) 40195327364930 (case), Lot Numbers: 23DDC317; 536) REF DYNJ63957C, UDI/DI 10195327428860 (each) 40195327428861 (case), Lot Numbers: 23GDB209, 23IDB634, 23KDB938, 23FDB374; 537) REF DYNJ63957D, UDI/DI 10195327580285 (each) 40195327580286 (case), Lot Numbers: 23LDB149, 24ADA221, 24ADB760; 538) REF DYNJ64004D, UDI/DI 10195327118679 (each) 40195327118670 (case), Lot Numbers: 22EDB550, 22HDA301, 23FDA090, 23GDB263, 23JDA746; 539) REF DYNJ64005D, UDI/DI 10195327118662 (each) 40195327118663 (case), Lot Numbers: 22DDA783, 22HDB620, 23CDB599, 23DDC345, 23ADC123, 22FDA748, 23HDB656; 540) REF DYNJ64008C, UDI/DI 10195327118587 (each) 40195327118588 (case), Lot Numbers: 22GDC267, 22EDB024, 23BDA201, 23CDA348, 23IDC065; 541) REF DYNJ64011D, UDI/DI 10195327118594 (each) 40195327118595 (case), Lot Numbers: 23BDA197; 542) REF DYNJ64184C, UDI/DI 10195327278793 (each) 40195327278794 (case), Lot Numbers: 23ABB562, 23FBL254, 23HBM054; 543) REF DYNJ64233B, UDI/DI 10193489382648 (each) 40193489382649 (case), Lot Numbers: 22EMF960, 22KMF038, 22IMG521, 22IMG699, 23BMG941; 544) REF DYNJ64240A, UDI/DI 10193489382631 (each) 40193489382632 (case), Lot Numbers: 22EMA915, 22GMA838, 22JMB039, 22KME606, 22IMF876, 23DMB656, 23CMG377; 545) REF DYNJ64276C, UDI/DI 10195327091095 (each) 40195327091096 (case), Lot Numbers: 22DBB451, 23EBG970, 22IBF538, 22FBX958, 22FBX960, 23ABL863, 23BBQ435; 546) REF DYNJ64344A, UDI/DI 10195327157425 (each) 40195327157426 (case), Lot Numbers: 22FBB836; 547) REF DYNJ64344B, UDI/DI 10195327190293 (each) 40195327190294 (case), Lot Numbers: 22KBJ534, 22LBM629; 548) REF DYNJ64672A, UDI/DI 10193489496963 (each) 40193489496964 (case), Lot Numbers: 22GMA040; 549) REF DYNJ64755C, UDI/DI 10195327665968 (each) 40195327665969 (case), Lot Numbers: 24CDB511; 550) REF DYNJ64796C, UDI/DI 10195327217945 (each) 40195327217946 (case), Lot Numbers: 23DDC094, 23GDB171, 23EDC564; 551) REF DYNJ64907A, UDI/DI 10195327053734 (each) 40195327053735 (case), Lot Numbers: 22EBI609, 22EBP072, 22JBM618, 22LBP896, 22KBD336, 22JBS427; 552) REF DYNJ64907B, UDI/DI 10195327489632 (each) 40195327489633 (case), Lot Numbers: 23IBH849; 553) REF DYNJ64955A, UDI/DI 10195327110949 (each) 40195327110940 (case), Lot Numbers: 22JMA953, 22HMA793, 22FMG704, 23EMB887, 23BMD644, 23BMD059, 23EMG035, 23IMH097, 23JMJ933; 554) REF DYNJ64981F, UDI/DI 10195327528133 (each) 40195327528134 (case), Lot Numbers: 23KMA711; 555) REF DYNJ65065A, UDI/DI 10193489317244 (each) 40193489317245 (case), Lot Numbers: 23AMG500, 22FME986, 23BMC082, 23HMC959, 24DMJ424, 23KMA049, 23LMH308, 24BMJ166, 24AMG861; 556) REF DYNJ65066B, UDI/DI 10193489317183 (each) 40193489317184 (case), Lot Numbers: 23AMG583, 22KME040, 22GMH242, 22GMA626, 22DMF180, 23DMA376, 23BMB169, 24BMA894, 23LMF624, 23KME638, 23FMD960, 23EMI651, 24AMG151, 24FMI648; 557) REF DYNJ65069A, UDI/DI 10193489317237 (each) 40193489317238 (case), Lot Numbers: 22LMD079, 23HMH939, 23KMJ410, 23KME112, 23BME898, 23AMG551, 23BMA302, 23AMB608, 24CMB410; 558) REF DYNJ65240A, UDI/DI 10193489416978 (each) 40193489416979 (case), Lot Numbers: 22DBT346; 559) REF DYNJ65240B, UDI/DI 10195327163778 (each) 40195327163779 (case), Lot Numbers: 22FBX940, 22LBL503, 23DBH093, 23ABT638, 23FBT717; 560) REF DYNJ65246, UDI/DI 10193489257595 (each) 40193489257596 (case), Lot Numbers: 23AMG386, 23FMB173, 23FMI413, 23JMJ638; 561) REF DYNJ65386C, UDI/DI 10195327410087 (each) 40195327410088 (case), Lot Numbers: 23FMC511; 562) REF DYNJ65484A, UDI/DI 10195327303501 (each) 40195327303502 (case), Lot Numbers: 23HMD135, 23CMF388; 563) REF DYNJ65825A, UDI/DI 10195327187729 (each) 40195327187720 (case), Lot Numbers: 23HMB027, 23IMD579; 564) REF DYNJ65855B, UDI/DI 10195327232146 (each) 40195327232147 (case), Lot Numbers: 22HBW485, 22LBD735, 23BBQ528; 565) REF DYNJ66236C, UDI/DI 10193489897272 (each) 40193489897273 (case), Lot Numbers: 22IMA620; 566) REF DYNJ66236D, UDI/DI 10195327275730 (each) 40195327275731 (case), Lot Numbers: 23AMH457, 23GMI240, 23HMI160; 567) REF DYNJ66237C, UDI/DI 10193489897289 (each) 40193489897280 (case), Lot Numbers: 22FBU863; 568) REF DYNJ66237D, UDI/DI 10195327275747 (each) 40195327275748 (case), Lot Numbers: 23FMI227, 23DMD426, 23EMB393; 569) REF DYNJ66357C, UDI/DI 10195327227975 (each) 40195327227976 (case), Lot Numbers: 22IBG338, 22HBX757, 22JBV661, 22JBW838, 23CBI755, 23DBB811; 570) REF DYNJ66649D, UDI/DI 10195327374570 (each) 40195327374571 (case), Lot Numbers: 23DBP833; 571) REF DYNJ66730A, UDI/DI 10195327504465 (each) 40195327504466 (case), Lot Numbers: 23LBD988, 24ABN196, 24BBC802, 24BBC859; 572) REF DYNJ66738A, UDI/DI 10195327504946 (each) 40195327504947 (case), Lot Numbers: 23KBN307, 24ABE488, 24BBF455, 24CBI276, 24EBL622; 573) REF DYNJ66968A, UDI/DI 10195327267117 (each) 40195327267118 (case), Lot Numbers: 23IBJ698; 574) REF DYNJ66969A, UDI/DI 10195327267520 (each) 40195327267521 (case), Lot Numbers: 22LBD280, 23FBO304, 23HBL009, 23IBQ506; 575) REF DYNJ66976A, UDI/DI 10193489879490 (each) 40193489879491 (case), Lot Numbers: 22JBG282, 23KBD050, 22LBD165; 576) REF DYNJ66981A, UDI/DI 10193489879537 (each) 40193489879538 (case), Lot Numbers: 23CBH243, 22DBK542, 22KBL344, 22IBR511, 23HBD638, 23KBC458, 23JBM179, 23KBQ338, 23KBQ341, 23ABK662, 23FBO864, 23DBJ273; 577) REF DYNJ67195C, UDI/DI 10195327299705 (each) 40195327299706 (case), Lot Numbers: 23BBI892; 578) REF DYNJ67196B, UDI/DI 10195327299712 (each) 40195327299713 (case), Lot Numbers: 23CBW673, 22LBO927; 579) REF DYNJ67212A, UDI/DI 10195327150549 (each) 40195327150540 (case), Lot Numbers: 23JBI789, 23KBU521, 22LBD041, 22LBD045, 23BBC539, 23DBM715, 23GBA079; 580) REF DYNJ67212B, UDI/DI 10195327577940 (each) 40195327577941 (case), Lot Numbers: 24ABB293; 581) REF DYNJ67212C, UDI/DI 10195327621735 (each) 40195327621736 (case), Lot Numbers: 24BBU528; 582) REF DYNJ67212D, UDI/DI 10195327668358 (each) 40195327668359 (case), Lot Numbers: 24DBQ160; 583) REF DYNJ67213A, UDI/DI 10195327150556 (each) 40195327150557 (case), Lot Numbers: 22EBP199, 23BBB397, 23BBB201, 23CBA507; 584) REF DYNJ67264A, UDI/DI 10195327039530 (each) 40195327039531 (case), Lot Numbers: 22EBM706, 22JBX213, 22JBG223, 23IBC366, 23HBY717, 23HBB411, 23JBF392, 23KBS927, 23DBK428, 23GBI625; 585) REF DYNJ67264B, UDI/DI 10195327578527 (each) 40195327578528 (case), Lot Numbers: 24ABU946; 586) REF DYNJ67264C, UDI/DI 10195327670153 (each) 40195327670154 (case), Lot Numbers: 24GBW785, 24DBO685; 587) REF DYNJ67265B, UDI/DI 10195327150921 (each) 40195327150922 (case), Lot Numbers: 22GBF576, 22EBM303, 22JBI919, 23JBC924, 23GBQ718, 23KBC712, 23LBC682, 22LBD113, 23ABJ042; 588) REF DYNJ67520G, UDI/DI 10195327334833 (each) 40195327334834 (case), Lot Numbers: 23BBC223, 23ABQ621, 23JBQ768, 23JBF278, 23GBH254, 23KBR697, 24ABF738, 23IBF489, 24CBO857; 589) REF DYNJ67521C, UDI/DI 10195327397913 (each) 40195327397914 (case), Lot Numbers: 23GDC211, 23KDA461, 24BDA015, 23IDA321, 23DDB944; 590) REF DYNJ67522D, UDI/DI 10195327332556 (each) 40195327332557 (case), Lot Numbers: 23KBG850, 23LBJ239, 23LBK732, 23ABN233, 23GBB678; 591) REF DYNJ67574I, UDI/DI 10195327406578 (each) 40195327406579 (case), Lot Numbers: 23FBM267; 592) REF DYNJ67596A, UDI/DI 10193489841060 (each) 40193489841061 (case), Lot Numbers: 22FBU865; 593) REF DYNJ67596B, UDI/DI 10195327217532 (each) 40195327217533 (case), Lot Numbers: 22JBL753, 23CBC289, 23IBR961, 23FBM685, 23KBB954; 594) REF DYNJ67677A, UDI/DI 10195327504779 (each) 40195327504770 (case), Lot Numbers: 23KBM911; 595) REF DYNJ67801B, UDI/DI 10195327228026 (each) 40195327228027 (case), Lot Numbers: 22IBR468, 22IBA999, 22HBX760, 22LBJ174, 23CBB332, 23BBH671, 23DBC354; 596) REF DYNJ67802B, UDI/DI 10195327227951 (each) 40195327227952 (case), Lot Numbers: 22IBB139, 22IBA731, 23ABJ302; 597) REF DYNJ67886B, UDI/DI 10193489980523 (each) 40193489980524 (case), Lot Numbers: 22ILA134, 22HLA375; 598) REF DYNJ67886C, UDI/DI 10195327265359 (each) 40195327265350 (case), Lot Numbers: 22JLB007, 22KLA316, 23CLA643, 23DLA767; 599) REF DYNJ67887B, UDI/DI 10193489970616 (each) 40193489970617 (case), Lot Numbers: 22IMF688; 600) REF DYNJ67959, UDI/DI 10193489459425 (each) 40193489459426 (case), Lot Numbers: 22DBT576, 22DBF011, 22FBO580, 22JBI285, 22IBO190, 23CBA566, 23ABA923; 601) REF DYNJ67983A, UDI/DI 10195327104429 (each) 40195327104420 (case), Lot Numbers: 22EBF444, 22EBF445; 602) REF DYNJ68204A, UDI/DI 10195327026356 (each) 40195327026357 (case), Lot Numbers: 22EMA922; 603) REF DYNJ68204B, UDI/DI 10195327352684 (each) 40195327352685 (case), Lot Numbers: 23EMB404, 23HMF322, 23FMD314, 23EMI706; 604) REF DYNJ68204C, UDI/DI 10195327512095 (each) 40195327512096 (case), Lot Numbers: 23JMG139, 23IMG918, 24BMA772, 24AMC705, 24AMA748, 24AMH321, 24CMC554; 605) REF DYNJ68206A, UDI/DI 10195327026226 (each) 40195327026227 (case), Lot Numbers: 22ELA483, 22KLA629, 22JLB227, 23ALB033; 606) REF DYNJ68206B, UDI/DI 10195327352615 (each) 40195327352616 (case), Lot Numbers: 23HLB045, 23GLA095, 23DLA972, 23CLA562; 607) REF DYNJ68206C, UDI/DI 10195327512118 (each) 40195327512119 (case), Lot Numbers: 23KLA363, 24ALB145; 608) REF DYNJ68553A, UDI/DI 10193489994889 (each) 40193489994880 (case), Lot Numbers: 23CBL534, 22IBD248, 23HBG150, 23BBC646, 23ABI373, 23DBR643, 23GBC957; 609) REF DYNJ69170A, UDI/DI 10195327156497 (each) 40195327156498 (case), Lot Numbers: 22EMG183, 22JMD107; 610) REF DYNJ69170B, UDI/DI 10195327342975 (each) 40195327342976 (case), Lot Numbers: 23GMC991, 23FMB245; 611) REF DYNJ69170C, UDI/DI 10195327491352 (each) 40195327491353 (case), Lot Numbers: 23LMH595, 23JMI162; 612) REF DYNJ69207A, UDI/DI 10195327187712 (each) 40195327187713 (case), Lot Numbers: 23DMC276, 23FMH470; 613) REF DYNJ69207B, UDI/DI 10195327516239 (each) 40195327516230 (case), Lot Numbers: 23IMH529; 614) REF DYNJ69487B, UDI/DI 10195327227159 (each) 40195327227150 (case), Lot Numbers: 23GDB343, 23CDA273, 23BDB852, 23JDA197; 615) REF DYNJ69577C, UDI/DI 10195327137205 (each) 40195327137206 (case), Lot Numbers: 22EBH426, 22FBF033; 616) REF DYNJ69577F, UDI/DI 10195327212896 (each) 40195327212897 (case), Lot Numbers: 22JBX725; 617) REF DYNJ69577G, UDI/DI 10195327302337 (each) 40195327302338 (case), Lot Numbers: 23KBG592, 24ABI922, 23LBL002, 23DBI943, 24ABU087, 23EBW816, 23IBJ299, 23JBL134; 618) REF DYNJ69578B, UDI/DI 10195327130435 (each) 40195327130436 (case), Lot Numbers: 22EME040; 619) REF DYNJ69578D, UDI/DI 10195327302320 (each) 40195327302321 (case), Lot Numbers: 23LMH283, 23KMJ024, 24AMG093, 23KMA437, 23FMA685, 23AMJ500, 23AMB592; 620) REF DYNJ80110A, UDI/DI 10195327098964 (each) 40195327098965 (case), Lot Numbers: 22IBH937, 23CME933, 23KME620; 621) REF DYNJ80130, UDI/DI 10195327010577 (each) 40195327010578 (case), Lot Numbers: 22IMF536; 622) REF DYNJ80140A, UDI/DI 10195327148058 (each) 40195327148059 (case), Lot Numbers: 22IMA277; 623) REF DYNJ80245A, UDI/DI 10195327195076 (each) 40195327195077 (case), Lot Numbers: 22FBW650; 624) REF DYNJ80284, UDI/DI 10195327020750 (each) 40195327020751 (case), Lot Numbers: 22GMA635, 22GMA293, 22DMH685, 22LMG202, 23DMB252, 24DMH294, 23IMA947; 625) REF DYNJ80298, UDI/DI 10195327021979 (each) 40195327021970 (case), Lot Numbers: 22GME140, 22IMB430; 626) REF DYNJ80424A, UDI/DI 10195327234119 (each) 40195327234110 (case), Lot Numbers: 22IBL761, 23BBE371; 627) REF DYNJ80435A, UDI/DI 10195327234102 (each) 40195327234103 (case), Lot Numbers: 22IBL761, 23BBE371, 23DBH684; 628) REF DYNJ80450A, UDI/DI 10195327168919 (each) 40195327168910 (case), Lot Numbers: 22FLB145, 22FLA590, 22GLA921; 629) REF DYNJ80450B, UDI/DI 10195327328542 (each) 40195327328543 (case), Lot Numbers: 23ALB121, 23CLA158, 23HLB046; 630) REF DYNJ80451, UDI/DI 10195327038304 (each) 40195327038305 (case), Lot Numbers: 22ILA015, 22JLA682; 631) REF DYNJ80503A, UDI/DI 10195327225483 (each) 40195327225484 (case), Lot Numbers: 22IMF538, 22KMB095, 23EMI342, 23EMB067, 23FMH680; 632) REF DYNJ80505, UDI/DI 10195327049195 (each) 40195327049196 (case), Lot Numbers: 22GMH017; 633) REF DYNJ80696, UDI/DI 10195327066482 (each) 40195327066483 (case), Lot Numbers: 22EBE965; 634) REF DYNJ80696A, UDI/DI 10195327228415 (each) 40195327228416 (case), Lot Numbers: 22JBC276; 635) REF DYNJ80696B, UDI/DI 10195327274115 (each) 40195327274116 (case), Lot Numbers: 23BBD025, 22LBR313; 636) REF DYNJ80697A, UDI/DI 10195327228521 (each) 40195327228522 (case), Lot Numbers: 22IBL693; 637) REF DYNJ80697B, UDI/DI 10195327274108 (each) 40195327274109 (case), Lot Numbers: 23ABR781, 22JBS491, 23BBO609; 638) REF DYNJ81176, UDI/DI 10195327089443 (each) 40195327089444 (case), Lot Numbers: 22EBM753, 22JBS368, 23ABG787, 23ABG437, 23HBB483, 23CBV612, 23JBN909, 23LBF379, 24GBV429, 24EBE649; 639) REF DYNJ81405A, UDI/DI 10195327470159 (each) 40195327470150 (case), Lot Numbers: 23IBN860; 640) REF DYNJ81598, UDI/DI 10195327124014 (each) 40195327124015 (case), Lot Numbers: 22HBO968, 22EBJ445, 22IBU035, 22HBX019; 641) REF DYNJ81598A, UDI/DI 10195327314934 (each) 40195327314935 (case), Lot Numbers: 22LBP952, 22LBN846, 23DBB923; 642) REF DYNJ81608, UDI/DI 10195327123925 (each) 40195327123926 (case), Lot Numbers: 22IBF053, 22EBJ747, 22JBS356; 643) REF DYNJ81608A, UDI/DI 10195327314927 (each) 40195327314928 (case), Lot Numbers: 23ABA190, 22LBN901, 23DBP300; 644) REF DYNJ81611, UDI/DI 10195327123963 (each) 40195327123964 (case), Lot Numbers: 22JBD143; 645) REF DYNJ81612, UDI/DI 10195327123987 (each) 40195327123988 (case), Lot Numbers: 22EBJ448, 22IBR520, 22HBY228; 646) REF DYNJ81612A, UDI/DI 10195327315061 (each) 40195327315062 (case), Lot Numbers: 22LBN205, 23CBA405, 23DBM719; 647) REF DYNJ81783A, UDI/DI 10195327515874 (each) 40195327515875 (case), Lot Numbers: 23IBT721; 648) REF DYNJ81783B, UDI/DI 10195327532215 (each) 40195327532216 (case), Lot Numbers: 23KBT832; 649) REF DYNJ81786C, UDI/DI 10195327525514 (each) 40195327525515 (case), Lot Numbers: 23JBI949; 650) REF DYNJ81786D, UDI/DI 10195327532222 (each) 40195327532223 (case), Lot Numbers: 23KBT803, 24BBC051, 24CBR666, 24EBD427; 651) REF DYNJ81812B, UDI/DI 10195327681517 (each) 40195327681518 (case), Lot Numbers: 24DDB407; 652) REF DYNJ81840, UDI/DI 10195327149543 (each) 40195327149544 (case), Lot Numbers: 22FMF269; 653) REF DYNJ82178, UDI/DI 10195327182694 (each) 40195327182695 (case), Lot Numbers: 23GDB510, 23CDA038, 23FDA552, 23IDC200, 23KMD170; 654) REF DYNJ82203, UDI/DI 10195327183691 (each) 40195327183692 (case), Lot Numbers: 23HMI536; 655) REF DYNJ82204, UDI/DI 10195327183714 (each) 40195327183715 (case), Lot Numbers: 23EDB094, 23HDA590, 23FDC215; 656) REF DYNJ82204A, UDI/DI 10195327520069 (each) 40195327520060 (case), Lot Numbers: 23JDA223; 657) REF DYNJ82205, UDI/DI 10195327183707 (each) 40195327183708 (case), Lot Numbers: 23EDB908, 23IDB080, 23CDA306; 658) REF DYNJ82258, UDI/DI 10195327181703 (each) 40195327181704 (case), Lot Numbers: 23FDA696, 23DDA374; 659) REF DYNJ82287, UDI/DI 10195327183974 (each) 40195327183975 (case), Lot Numbers: 23FDA654, 23GDB511; 660) REF DYNJ82293, UDI/DI 10195327184315 (each) 40195327184316 (case), Lot Numbers: 23BDA115; 661) REF DYNJ82293A, UDI/DI 10195327503512 (each) 40195327503513 (case), Lot Numbers: 23IDB110, 23JMH999; 662) REF DYNJ82299, UDI/DI 10195327184322 (each) 40195327184323 (case), Lot Numbers: 23HDB220, 23HDA189; 663) REF DYNJ82299A, UDI/DI 10195327503543 (each) 40195327503544 (case), Lot Numbers: 23JMH016, 23JMB780; 664) REF DYNJ82311, UDI/DI 10195327184346 (each) 40195327184347 (case), Lot Numbers: 23FDC334; 665) REF DYNJ82311A, UDI/DI 10195327503475 (each) 40195327503476 (case), Lot Numbers: 23IDB027; 666) REF DYNJ82313, UDI/DI 10195327184339 (each) 40195327184330 (case), Lot Numbers: 23FDA667, 23HDB650, 23IMA325; 667) REF DYNJ82315, UDI/DI 10195327184353 (each) 40195327184354 (case), Lot Numbers: 23CDB089; 668) REF DYNJ82315A, UDI/DI 10195327503505 (each) 40195327503506 (case), Lot Numbers: 23JMH046; 669) REF DYNJ82317, UDI/DI 10195327185008 (each) 40195327185009 (case), Lot Numbers: 23DDA012, 23GDB650, 23GDA239; 670) REF DYNJ82331, UDI/DI 10195327185046 (each) 40195327185047 (case), Lot Numbers: 23ADB798, 23FDA073; 671) REF DYNJ82334, UDI/DI 10195327184988 (each) 40195327184989 (case), Lot Numbers: 23BDA016, 23FDA729; 672) REF DYNJ82340, UDI/DI 10195327185015 (each) 40195327185016 (case), Lot Numbers: 23HDA579, 23IDA276, 23GDA240; 673) REF DYNJ82341, UDI/DI 10195327185084 (each) 40195327185085 (case), Lot Numbers: 23ADC134, 23IMA893; 674) REF DYNJ82343, UDI/DI 10195327185091 (each) 40195327185092 (case), Lot Numbers: 23EDB409, 23IDB878; 675) REF DYNJ82345A, UDI/DI 10195327316556 (each) 40195327316557 (case), Lot Numbers: 23ABE332, 23GBR353, 23CBV824, 23LBT387, 23GBK085, 23GBD178, 23FBE485, 24ABT128; 676) REF DYNJ82365, UDI/DI 10195327186715 (each) 40195327186716 (case), Lot Numbers: 23HDA097, 23JMB452; 677) REF DYNJ82367A, UDI/DI 10195327316563 (each) 40195327316564 (case), Lot Numbers: 23ABI222, 23HBC300, 23LBD137, 24ABT015, 24ABE893, 23GBK112, 23FBU725; 678) REF DYNJ82669, UDI/DI 10195327202811 (each) 40195327202812 (case), Lot Numbers: 23HMF026, 23GMI179, 23HMI734; 679) REF DYNJ82673, UDI/DI 10195327203047 (each) 40195327203048 (case), Lot Numbers: 23DMI297, 23GMD089, 23IMI418, 23HMC831, 23JMJ228, 23LME450; 680) REF DYNJ82675, UDI/DI 10195327203061 (each) 40195327203062 (case), Lot Numbers: 23GMD792, 23HME628; 681) REF DYNJ82685, UDI/DI 10195327202873 (each) 40195327202874 (case), Lot Numbers: 23FMF892, 23GMD848, 23KME727, 23IMG386, 23HMF087; 682) REF DYNJ82689, UDI/DI 10195327202910 (each) 40195327202911 (case), Lot Numbers: 23JMH128; 683) REF DYNJ82693, UDI/DI 10195327202958 (each) 40195327202959 (case), Lot Numbers: 23GMC986, 23IMA575, 23KMF187; 684) REF DYNJ82950, UDI/DI 10195327218966 (each) 40195327218967 (case), Lot Numbers: 22JBD140, 23BBS169; 685) REF DYNJ82953, UDI/DI 10195327218997 (each) 40195327218998 (case), Lot Numbers: 22JBD999, 23DBO510, 23CBJ924; 686) REF DYNJ83005, UDI/DI 10195327225896 (each) 40195327225897 (case), Lot Numbers: 22JMH340; 687) REF DYNJ83005A, UDI/DI 10195327377083 (each) 40195327377084 (case), Lot Numbers: 23HME353, 23KMD629, 23FME324, 24AMD974, 24DME585, 24CMB225; 688) REF DYNJ83006, UDI/DI 10195327225902 (each) 40195327225903 (case), Lot Numbers: 23AMG948, 22JME292; 689) REF DYNJ83006A, UDI/DI 10195327377090 (each) 40195327377091 (case), Lot Numbers: 23LMA658, 23KMB885, 24FMI201, 24AMD927, 24CMC319; 690) REF DYNJ83010, UDI/DI 10195327226251 (each) 40195327226252 (case), Lot Numbers: 22KMA653, 23CMF803; 691) REF DYNJ83010A, UDI/DI 10195327377106 (each) 40195327377107 (case), Lot Numbers: 23HMA251, 23FME163; 692) REF DYNJ83010B, UDI/DI 10195327500733 (each) 40195327500734 (case), Lot Numbers: 23IMF327, 23JMI712, 23LMA550, 23KME568, 24AMG159, 24CMC390; 693) REF DYNJ83107, UDI/DI 10195327232221 (each) 40195327232222 (case), Lot Numbers: 22LMF005; 694) REF DYNJ83107A, UDI/DI 10195327381950 (each) 40195327381951 (case), Lot Numbers: 23DMF950; 695) REF DYNJ83186A, UDI/DI 10195327282745 (each) 40195327282746 (case), Lot Numbers: 22KBM902; 696) REF DYNJ83186B, UDI/DI 10195327364625 (each) 40195327364626 (case), Lot Numbers: 23DBB562, 24ABI748, 23JBE810, 23KBW468, 23KBA804, 23FBE338, 24ABN430, 24BBG127, 23LBP872; 697) REF DYNJ83234, UDI/DI 10195327247782 (each) 40195327247783 (case), Lot Numbers: 23EBC088, 22KBI755, 24FBO388, 24EBA239, 24CBD652; 698) REF DYNJ83420, UDI/DI 10195327265380 (each) 40195327265381 (case), Lot Numbers: 22LBP312, 22LBN377, 23CBH126, 23EBT725, 23HBO639, 23KBL617, 23JBW081, 24ABK394, 24CBA696, 24DBV489; 699) REF DYNJ83495, UDI/DI 10195327271800 (each) 40195327271801 (case), Lot Numbers: 23ABV177, 23IBK256; 700) REF DYNJ83591, UDI/DI 10195327283704 (each) 40195327283705 (case), Lot Numbers: 23ABV178; 701) REF DYNJ83592, UDI/DI 10195327283803 (each) 40195327283804 (case), Lot Numbers: 23BMG540; 702) REF DYNJ83628, UDI/DI 10195327286897 (each) 40195327286898 (case), Lot Numbers: 23HMA558, 23FMB991, 23EMD373, 23IMC902, 23KMJ347, 24BMB018; 703) REF DYNJ83628A, UDI/DI 10195327396817 (each) 40195327396818 (case), Lot Numbers: 23CBC376; 704) REF DYNJ83660, UDI/DI 10195327292430 (each) 40195327292431 (case), Lot Numbers: 23CBS225; 705) REF DYNJ83664, UDI/DI 10195327292553 (each) 40195327292554 (case), Lot Numbers: 23BBP135; 706) REF DYNJ83760, UDI/DI 10195327301743 (each) 40195327301744 (case), Lot Numbers: 23BBP136, 23BBP535; 707) REF DYNJ83761, UDI/DI 10195327301750 (each) 40195327301751 (case), Lot Numbers: 23CBV738; 708) REF DYNJ84057A, UDI/DI 10195327358020 (each) 40195327358021 (case), Lot Numbers: 23CBV739; 709) REF DYNJ84060A, UDI/DI 10195327357993 (each) 40195327357994 (case), Lot Numbers: 23HMA973, 23IMD249; 710) REF DYNJ84185, UDI/DI 10195327340575 (each) 40195327340576 (case), Lot Numbers: 23HME837, 23IMF690; 711) REF DYNJ84186, UDI/DI 10195327340568 (each) 40195327340569 (case), Lot Numbers: 23FBN190; 712) REF DYNJ84288, UDI/DI 10195327348182 (each) 40195327348183 (case), Lot Numbers: 23KME091; 713) REF DYNJ84967, UDI/DI 10195327441807 (each) 40195327441808 (case), Lot Numbers: 24DBB343; 714) REF DYNJ85243, UDI/DI 10195327482756 (each) 40195327482757 (case), Lot Numbers: 22GBM717, 22EBA964; 715) REF DYNJ85467A, UDI/DI 10195327637941 (each) 40195327637942 (case), Lot Numbers: 22JBS880, 22JBU383, 22KBQ468, 23CBN666, 23HBC933, 23EBA463, 24ABJ696, 24BBT055, 23KBO627; 716) REF DYNJ900023R, UDI/DI 10193489841381 (each) 40193489841382 (case), Lot Numbers: 23CBJ714, 23GBA115; 717) REF DYNJ900023S, UDI/DI 10195327213640 (each) 40195327213641 (case), Lot Numbers: 22GBV188, 22FBF101, 22EBO298, 22EBB375, 23ABF776, 22LBM184, 23EBT067; 718) REF DYNJ900268G, UDI/DI 10195327348168 (each) 40195327348169 (case), Lot Numbers: 22GBR251, 22GBR253, 23GBC525, 23ABM532, 23DBN404, 23EBG707, 23EBS012; 719) REF DYNJ900283K, UDI/DI 10193489915976 (each) 40193489915977 (case), Lot Numbers: 22FBF090, 22JBP203; 720) REF DYNJ900288G, UDI/DI 10193489915105 (each) 40193489915106 (case), Lot Numbers: 23BBU941, 23EBI554, 23GBL756; 721) REF DYNJ900302G, UDI/DI 10193489915488 (each) 40193489915489 (case), Lot Numbers: 22FBN690, 22FBN691, 22IBM297, 22GBH609, 22KBD779, 23ABS791, 23ABQ840, 23ABK496, 22LBC403, 23ABM639, 23ABK504, 23FBG452; 722) REF DYNJ900303M, UDI/DI 10195327175269 (each) 40195327175260 (case), Lot Numbers: 23EBQ520, 23BBS235, 23LBV758, 23KBL260, 24ABR230, 24EBF033, 24EBP422; 723) REF DYNJ900304N, UDI/DI 10195327175276 (each) 40195327175277 (case), Lot Numbers: 22FBN690, 22FBN691, 22IBM297, 22GBH609, 22KBD779, 23ABS791, 23ABQ840, 23ABK496, 22LBC403, 23ABM639, 23ABK504, 23FBG452; 724) REF DYNJ900414K, UDI/DI 10195327029432 (each) 40195327029433 (case), Lot Numbers: 23EBQ520, 23BBS235, 23LBV758, 23KBL260, 24ABR230, 24EBF033, 24EBP422; 725) REF DYNJ900419P, UDI/DI 10195327043773 (each) 40195327043774 (case), Lot Numbers: 22FBY504, 22KBG527, 23ABB690, 23CBE025, 23CBD656, 23JBB780, 23GBG691, 23HBT096, 24ABE589, 23LBF707, 24BBQ926, 24BBB082, 24EBA349, 24DBD222, 24GBL525; 726) REF DYNJ900565I, UDI/DI 10193489595369 (each) 40193489595360 (case), Lot Numbers: 23ABC040, 22KBJ801, 22LBL182, 23BBL703, 23DBH249; 727) REF DYNJ900570I, UDI/DI 10193489848670 (each) 40193489848671 (case), Lot Numbers: 23ALA606, 23CLA120, 24DLA275, 23ILA275, 23GLA212, 23LLA791, 23HLA377, 23GLA531, 24FLA281, 24ELA550; 728) REF DYNJ900606F, UDI/DI 10193489463750 (each) 40193489463751 (case), Lot Numbers: 22FDA713, 22FDA433, 22EDB172; 729) REF DYNJ900606G, UDI/DI 10195327224097 (each) 40195327224098 (case), Lot Numbers: 22IDB441, 22KDA284, 23BDB270, 23ADB286, 23DDA543; 730) REF DYNJ900624F, UDI/DI 10193489463743 (each) 40193489463744 (case), Lot Numbers: 22EDB592, 22FDB481, 22FDC080; 731) REF DYNJ900624G, UDI/DI 10195327224134 (each) 40195327224135 (case), Lot Numbers: 22HDB968, 22KDA295, 23EDA423, 23BDC197, 23ADA511; 732) REF DYNJ900651C, UDI/DI 10193489436716 (each) 40193489436717 (case), Lot Numbers: 22JBL369, 22IBV324, 22LBG981, 23EBK655, 23CBO965; 733) REF DYNJ900658B, UDI/DI 10193489436709 (each) 40193489436700 (case), Lot Numbers: 22IBU638, 22DBN187, 22DBN529, 23ABP621, 23EBJ341, 23BBF724; 734) REF DYNJ900713D, UDI/DI 10193489951318 (each) 40193489951319 (case), Lot Numbers: 22GBS111, 22FBU842, 23DBJ848, 23CBJ961, 23ABC194, 22FBS164, 22LBS197; 735) REF DYNJ900718C, UDI/DI 10193489951578 (each) 40193489951579 (case), Lot Numbers: 22EBM260, 22HBF175, 23CBS226, 23CBS227, 22KBF981, 23ABM644, 23DBO409; 736) REF DYNJ900719B, UDI/DI 10193489951585 (each) 40193489951586 (case), Lot Numbers: 23ABQ858, 23ABD345; 737) REF DYNJ900720D, UDI/DI 10193489951592 (each) 40193489951593 (case), Lot Numbers: 22FBO545, 22JBF689, 22IBN802, 22GBO891, 23ABP512, 22JBY062, 22LBC649, 23DBD541, 23EBJ422; 738) REF DYNJ900722C, UDI/DI 10193489951615 (each) 40193489951616 (case), Lot Numbers: 22GBS256, 23DBK565, 23CBV883, 23CBS228, 23CBS229, 23CBJ715, 22LBD478, 22KBF311, 23ABP483; 739) REF DYNJ900745D, UDI/DI 10193489526059 (each) 40193489526050 (case), Lot Numbers: 23FBK849, 23ABR066, 22KBR848, 23EBG717, 23DBD756, 24DBL753, 24ABY468, 24ABS951; 740) REF DYNJ900747G, UDI/DI 10889942365067 (each) 40889942365068 (case), Lot Numbers: 22JBK492, 23DBD481, 23BBI448, 23ABM357, 23ABD436, 22LBF812, 22KBS011, 23FBV621, 23IBB489, 23EBL072, 24GBX528, 23LBH747, 24ABX872, 24DBM828; 741) REF DYNJ900748K, UDI/DI 10193489655247 (each) 40193489655248 (case), Lot Numbers: 22JBP204, 22KBS040, 23ABK742, 23EBF972, 23BBI402, 23FBA664, 23EBV896, 23IBD644, 24ABJ166, 23KBH147, 23KBC019, 23GBK063, 24CBU901, 24CBM212, 24EBM861; 742) REF DYNJ901064F, UDI/DI 10193489944396 (each) 40193489944397 (case), Lot Numbers: 22HMH812, 22HMF271, 22HME031; 743) REF DYNJ901064I, UDI/DI 10195327232597 (each) 40195327232598 (case), Lot Numbers: 22KMC858, 22JMC315, 23CME948, 23HMD823, 23GMB741, 23FMI200, 23FMC572, 23BMF414, 23BMF424, 23DMF243; 744) REF DYNJ901069K, UDI/DI 10195327229948 (each) 40195327229949 (case), Lot Numbers: 23AMF005, 23AMA681, 22LMI497, 22LMF749, 22LMB479, 23CMA535, 23BMI246, 23DMB761; 745) REF DYNJ901074G, UDI/DI 10193489944624 (each) 40193489944625 (case), Lot Numbers: 22HMF139; 746) REF DYNJ901074J, UDI/DI 10195327230036 (each) 40195327230037 (case), Lot Numbers: 23DMI015; 747) REF DYNJ901111C, UDI/DI 10195327127503 (each) 40195327127504 (case), Lot Numbers: 22KBI772, 22GBR320, 23CBR809, 23ABM465, 24BBO660, 23JBT819, 23EBL425; 748) REF DYNJ901748I, UDI/DI 10195327495725 (each) 40195327495726 (case), Lot Numbers: 23LDB126, 23JDA078, 24ADB909, 24DDB346; 749) REF DYNJ901751Q, UDI/DI 10195327496067 (each) 40195327496068 (case), Lot Numbers: 24ADB894, 24CDB979; 750) REF DYNJ901752Q, UDI/DI 10195327496173 (each) 40195327496174 (case), Lot Numbers: 24BDB918, 24ADA253, 23JDA606, 23LDA958, 23LDA660, 23JDB757, 24DDA210, 24JDA090, 24FDC008, 24EDA615; 751) REF DYNJ901755Q, UDI/DI 10195327582418 (each) 40195327582419 (case), Lot Numbers: 24AMD405; 752) REF DYNJ901764O, UDI/DI 10195327582401 (each) 40195327582402 (case), Lot Numbers: 24AMF214; 753) REF DYNJ901808I, UDI/DI 10195327496043 (each) 40195327496044 (case), Lot Numbers: 23KDA545, 23JDB826, 24ADA149, 24CDA340, 24DDB799; 754) REF DYNJ901816I, UDI/DI 10195327495701 (each) 40195327495702 (case), Lot Numbers: 24DDB344, 23LDB492; 755) REF DYNJ902058C, UDI/DI 10193489359459 (each) 40193489359450 (case), Lot Numbers: 23ABK598, 22JBV808, 23CBC488; 756) REF DYNJ902058D, UDI/DI 10195327422479 (each) 40195327422470 (case), Lot Numbers: 23KBR583, 23KBG149; 757) REF DYNJ902058F, UDI/DI 10195327572389 (each) 40195327572380 (case), Lot Numbers: 24CBA332, 24BBK015, 24CBS468; 758) REF DYNJ902061G, UDI/DI 10195327086756 (each) 40195327086757 (case), Lot Numbers: 22HBD345, 22JBR422, 22KBC742, 23EBF045, 23EBE314, 23BBN957, 23BBB178; 759) REF DYNJ902061I, UDI/DI 10195327422349 (each) 40195327422340 (case), Lot Numbers: 23KBB124, 23JBL348; 760) REF DYNJ902061J, UDI/DI 10195327572402 (each) 40195327572403 (case), Lot Numbers: 24DBI718, 24CBK603, 24CBK416; 761) REF DYNJ902133K, UDI/DI 10195327497248 (each) 40195327497249 (case), Lot Numbers: 24ADB559, 23LDA150, 23JDB901; 762) REF DYNJ902282I, UDI/DI 10193489983890 (each) 40193489983891 (case), Lot Numbers: 22FDA613, 22DDC497, 22IDB317, 22HDA055; 763) REF DYNJ902282J, UDI/DI 10195327313319 (each) 40195327313310 (case), Lot Numbers: 23HDA883, 23FDB844, 23EDB115, 23CDB987, 23ADB634, 22LDB600, 23IDB843, 23KDB385; 764) REF DYNJ902282K, UDI/DI 10195327585174 (each) 40195327585175 (case), Lot Numbers: 24EDA308; 765) REF DYNJ902313G, UDI/DI 10195327493943 (each) 40195327493944 (case), Lot Numbers: 23KDA051, 23IDB388, 24BDB273, 23LDB168, 24CDB486; 766) REF DYNJ902365L, UDI/DI 10195327144753 (each) 40195327144754 (case), Lot Numbers: 23ABD120, 23IBH853, 23GBJ208, 23IBP814, 23KBA934; 767) REF DYNJ902395D, UDI/DI 10195327457556 (each) 40195327457557 (case), Lot Numbers: 23JDB263, 23IDB602; 768) REF DYNJ902396G, UDI/DI 10195327457570 (each) 40195327457571 (case), Lot Numbers: 23HDB038; 769) REF DYNJ902699P, UDI/DI 10193489841015 (each) 40193489841016 (case), Lot Numbers: 22EBO528; 770) REF DYNJ902699Q, UDI/DI 10195327213657 (each) 40195327213658 (case), Lot Numbers: 23HBF074; 771) REF DYNJ902763G, UDI/DI 10195327461713 (each) 40195327461714 (case), Lot Numbers: 23KBD941, 23KBE318, 24BBG303, 24ABP224; 772) REF DYNJ902776F, UDI/DI 10195327462116 (each) 40195327462117 (case), Lot Numbers: 24ABW975, 24ABH021; 773) REF DYNJ902782G, UDI/DI 10195327461690 (each) 40195327461691 (case), Lot Numbers: 23JBB093, 24ABP601, 24BBH057; 774) REF DYNJ902974N, UDI/DI 10195327147501 (each) 40195327147502 (case), Lot Numbers: 22IDA129; 775) REF DYNJ902977N, UDI/DI 10195327206673 (each) 40195327206674 (case), Lot Numbers: 22LDB526, 22KDA429; 776) REF DYNJ902978L, UDI/DI 10195327147525 (each) 40195327147526 (case), Lot Numbers: 23CDA033, 23GDB554; 777) REF DYNJ902979O, UDI/DI 10195327144340 (each) 40195327144341 (case), Lot Numbers: 22LDA303; 778) REF DYNJ903053F, UDI/DI 10195327461614 (each) 40195327461615 (case), Lot Numbers: 23IMA506, 23IMH666; 779) REF DYNJ903053G, UDI/DI 10195327555115 (each) 40195327555116 (case), Lot Numbers: 24AMA676, 24AMF097; 780) REF DYNJ903053I, UDI/DI 10195327612986 (each) 40195327612987 (case), Lot Numbers: 24CMG800; 781) REF DYNJ903230F, UDI/DI 10195327461607 (each) 40195327461608 (case), Lot Numbers: 23IMA463, 23JMA077; 782) REF DYNJ903230G, UDI/DI 10195327555122 (each) 40195327555123 (case), Lot Numbers: 24BMD681, 24AMG423; 783) REF DYNJ903534D, UDI/DI 10195327461638 (each) 40195327461639 (case), Lot Numbers: 23IMH904, 23HMG998, 23IMF613; 784) REF DYNJ903561C, UDI/DI 10195327348212 (each) 40195327348213 (case), Lot Numbers: 23HBN852; 785) REF DYNJ903606C, UDI/DI 10193489959413 (each) 40193489959414 (case), Lot Numbers: 22GBT643; 786) REF DYNJ903606D, UDI/DI 10195327237592 (each) 40195327237593 (case), Lot Numbers: 22LBH046, 22KBP516, 22KBJ512, 22JBP030; 787) REF DYNJ903606F, UDI/DI 10195327368425 (each) 40195327368426 (case), Lot Numbers: 23EBH942, 23HBM044, 23GBN295, 24ABX857, 23JBU718, 24BBQ150, 23FBS614; 788) REF DYNJ903606G, UDI/DI 10195327658205 (each) 40195327658206 (case), Lot Numbers: 24DBK732; 789) REF DYNJ903607G, UDI/DI 10195327373238 (each) 40195327373239 (case), Lot Numbers: 23IBW024, 24DBN828, 23LBD696; 790) REF DYNJ903608C, UDI/DI 10195327237615 (each) 40195327237616 (case), Lot Numbers: 23ABE687, 23ABC300; 791) REF DYNJ903608D, UDI/DI 10195327368401 (each) 40195327368402 (case), Lot Numbers: 23JBN509, 24ABT610, 24BBD902; 792) REF DYNJ903610C, UDI/DI 10195327237479 (each) 40195327237470 (case), Lot Numbers: 22KBC036, 22KBA216, 22LBS792, 23BBB781; 793) REF DYNJ903610F, UDI/DI 10195327364441 (each) 40195327364442 (case), Lot Numbers: 23DBR784; 794) REF DYNJ903610G, UDI/DI 10195327400927 (each) 40195327400928 (case), Lot Numbers: 24CBG839, 23JBE453, 23JBD275, 23HBU311, 795) REF DYNJ903611B, UDI/DI 10888277761667 (each) 40888277761668 (case), Lot Numbers: 22GBP110; 796) REF DYNJ903611C, UDI/DI 10195327237486 (each) 40195327237487 (case), Lot Numbers: 23BBO734, 22KBL171, 22JBO970, 23ABA728; 797) REF DYNJ903611D, UDI/DI 10195327344733 (each) 40195327344734 (case), Lot Numbers: 23CBD751; 798) REF DYNJ903611G, UDI/DI 10195327400842 (each) 40195327400843 (case), Lot Numbers: 23KBJ499, 24CBF351, 23IBK710, 23IBK768, 24ABS861; 799) REF DYNJ903611I, UDI/DI 10195327648237 (each) 40195327648238 (case), Lot Numbers: 24DBR209, 24DBR261; 800) REF DYNJ903613G, UDI/DI 10195327400835 (each) 40195327400836 (case), Lot Numbers: 23HBT724, 24BBQ921, 24EBI139, 24EBI392, 24BBC917; 801) REF DYNJ903620F, UDI/DI 10195327323639 (each) 40195327323630 (case), Lot Numbers: 23BBQ238; 802) REF DYNJ903620G, UDI/DI 10195327364410 (each) 40195327364411 (case), Lot Numbers: 24ABS999, 23IBQ541; 803) REF DYNJ903621F, UDI/DI 10195327237585 (each) 40195327237586 (case), Lot Numbers: 23BBN895, 22JBR475; 804) REF DYNJ903621G, UDI/DI 10195327373092 (each) 40195327373093 (case), Lot Numbers: 23EBH910, 23IBO529, 23HBL684, 24ABR045; 805) REF DYNJ903621I, UDI/DI 10195327648244 (each) 40195327648245 (case), Lot Numbers: 24DBC507; 806) REF DYNJ903763K, UDI/DI 10195327217655 (each) 40195327217656 (case), Lot Numbers: 22KBP183; 807) REF DYNJ903763L, UDI/DI 10195327334406 (each) 40195327334407 (case), Lot Numbers: 23BBS389; 808) REF DYNJ903763N, UDI/DI 10195327467463 (each) 40195327467464 (case), Lot Numbers: 23IBF008, 23HBH072, 23KBW125, 23JBH997, 24BBI400, 24ABO355, 24DBF357; 809) REF DYNJ903763O, UDI/DI 10195327675837 (each) 40195327675838 (case), Lot Numbers: 24DBQ302, 24DBQ917; 810) REF DYNJ903764J, UDI/DI 10195327217518 (each) 40195327217519 (case), Lot Numbers: 23ABL498, 23BBB953; 811) REF DYNJ903766I, UDI/DI 10195327217686 (each) 40195327217687 (case), Lot Numbers: 23CBG888, 23BBG549, 23ABN709, 23ABN187, 23EBQ535, 24BBD230, 24CBM973; 812) REF DYNJ903766J, UDI/DI 10195327675943 (each) 40195327675944 (case), Lot Numbers: 24DBQ916; 813) REF DYNJ903800I, UDI/DI 10193489274479 (each) 40193489274470 (case), Lot Numbers: 22LBF915, 23CBI550, 22LBB553, 22LBB549, 23CBA567, 23HBT725, 24ABB989, 23KBT454, 23JBT594, 23GBF923, 24BBG975, 24ABT474, 24CBA746, 24BBK582; 814) REF DYNJ903800J, UDI/DI 10195327640682 (each) 40195327640683 (case), Lot Numbers: 24DBE535; 815) REF DYNJ903801D, UDI/DI 10193489274974 (each) 40193489274975 (case), Lot Numbers: 22FBG658, 22HBS607, 23HBN657, 23BBH118, 23CBE017, 23IBG494, 23JBQ332, 24BBK583, 23LBN311; 816) REF DYNJ903801F, UDI/DI 10195327640675 (each) 40195327640676 (case), Lot Numbers: 24DBC481, 24EBE301; 817) REF DYNJ903802D, UDI/DI 10193489275582 (each) 40193489275583 (case), Lot Numbers: 23DBE028, 22JBW593, 23ABB749, 23FBV467, 23EBS008, 24ABL803, 23LBF498; 818) REF DYNJ903802F, UDI/DI 10195327640668 (each) 40195327640669 (case), Lot Numbers: 24CBP664; 819) REF DYNJ903803C, UDI/DI 10193489274981 (each) 40193489274982 (case), Lot Numbers: 22GBH541, 22GBU423, 22LBS208, 22KBE702, 23ABQ937, 22JBW280; 820) REF DYNJ903863C, UDI/DI 10193489232974 (each) 40193489232975 (case), Lot Numbers: 22IMF361, 22IMC641, 22DMA800; 821) REF DYNJ903863D, UDI/DI 10195327236632 (each) 40195327236633 (case), Lot Numbers: 22LMD818, 23CMF085, 23CMF358, 23CMB373, 23GMC259, 23HMA303; 822) REF DYNJ903870D, UDI/DI 10193489415971 (each) 40193489415972 (case), Lot Numbers: 22IBB541, 22EBL327, 22HBL344, 22HBM583; 823) REF DYNJ904084F, UDI/DI 10193489282849 (each) 40193489282840 (case), Lot Numbers: 22HDB932, 22JDB404, 23ADA882, 22KDB333, 22LDA036; 824) REF DYNJ904109F, UDI/DI 10193489841404 (each) 40193489841405 (case), Lot Numbers: 23GBP508, 23EBC803, 23DBG140, 23KBG241; 825) REF DYNJ904113L, UDI/DI 10193489841367 (each) 40193489841368 (case), Lot Numbers: 22GBI998, 22GBI999; 826) REF DYNJ904113N, UDI/DI 10195327272043 (each) 40195327272044 (case), Lot Numbers: 23FBM599, 23FBF556, 22LBL190; 827) REF DYNJ904205G, UDI/DI 10195327450816 (each) 40195327450817 (case), Lot Numbers: 23HDC122; 828) REF DYNJ904206G, UDI/DI 10195327450823 (each) 40195327450824 (case), Lot Numbers: 23HDC360, 23IDA118; 829) REF DYNJ904209I, UDI/DI 10195327450830 (each) 40195327450831 (case), Lot Numbers: 23HDC084, 23KDA682, 23JDA024; 830) REF DYNJ904210I, UDI/DI 10195327450847 (each) 40195327450848 (case), Lot Numbers: 23GDC141, 23IDA584; 831) REF DYNJ904399F, UDI/DI 10195327545536 (each) 40195327545537 (case), Lot Numbers: 24AMD530; 832) REF DYNJ904475K, UDI/DI 10195327186302 (each) 40195327186303 (case), Lot Numbers: 23ABL097; 833) REF DYNJ904480K, UDI/DI 10195327186456 (each) 40195327186457 (case), Lot Numbers: 22KBC062, 23ABG264, 22LBT100; 834) REF DYNJ904481J, UDI/DI 10195327186364 (each) 40195327186365 (case), Lot Numbers: 22GBG217, 22HBO530, 22HBD285; 835) REF DYNJ904482F, UDI/DI 10195327165888 (each) 40195327165889 (case), Lot Numbers: 22GBA664; 836) REF DYNJ904482G, UDI/DI 10195327186500 (each) 40195327186501 (case), Lot Numbers: 22FBU445, 22GBZ137; 837) REF DYNJ904487J, UDI/DI 10195327186395 (each) 40195327186396 (case), Lot Numbers: 22KBS411; 838) REF DYNJ904489I, UDI/DI 10195327166212 (each) 40195327166213 (case), Lot Numbers: 22FBO367; 839) REF DYNJ904497J, UDI/DI 10195327186227 (each) 40195327186228 (case), Lot Numbers: 22FBY372, 22HBD286; 840) REF DYNJ904498K, UDI/DI 10195327186234 (each) 40195327186235 (case), Lot Numbers: 23ABH742, 22GBT218; 841) REF DYNJ904589F, UDI/DI 10195327497361 (each) 40195327497362 (case), Lot Numbers: 24ADB251; 842) REF DYNJ904648F, UDI/DI 10193489931228 (each) 40193489931229 (case), Lot Numbers: 22EBJ527, 22FBL680, 22JBB373; 843) REF DYNJ904648G, UDI/DI 10195327277833 (each) 40195327277834 (case), Lot Numbers: 22KBP499, 23ABE040, 23CBJ375; 844) REF DYNJ904698F, UDI/DI 10195327169282 (each) 40195327169283 (case), Lot Numbers: 23CDC646, 22IDB991, 23EDA530; 845) REF DYNJ904702G, UDI/DI 10195327169312 (each) 40195327169313 (case), Lot Numbers: 22GDB187, 22HDA865, 23ADA273, 22LDA441, 23DDC167, 23CDB936, 23BDB141; 846) REF DYNJ904703F, UDI/DI 10195327169268 (each) 40195327169269 (case), Lot Numbers: 22GDB058, 22FDA909, 22IDA556, 22JDB907, 23DDB953, 23CDB132, 23CDA593, 23BDA147; 847) REF DYNJ904704D, UDI/DI 10195327126247 (each) 40195327126248 (case), Lot Numbers: 22GDB156, 22LDA718, 23DDA759; 848) REF DYNJ904765B, UDI/DI 10195327010485 (each) 40195327010486 (case), Lot Numbers: 22GBQ785, 22GBK512, 23CBC561, 23JBO143, 23IBT377; 849) REF DYNJ904765C, UDI/DI 10195327535742 (each) 40195327535743 (case), Lot Numbers: 23KBV823; 850) REF DYNJ904860C, UDI/DI 10195327124793 (each) 40195327124794 (case), Lot Numbers: 22HMD644; 851) REF DYNJ904860D, UDI/DI 10195327222536 (each) 40195327222537 (case), Lot Numbers: 22JMC904, 22LMA059, 22KMG895, 22KMF847, 23HMH067, 23FMI126, 23CMF389, 23CMF009, 23CMD418, 23IMD317, 24AMA086, 23KMI863, 23KME314, 24AMJ020; 852) REF DYNJ904906G, UDI/DI 10195327065164 (each) 40195327065165 (case), Lot Numbers: 22FBN681, 22FBN682, 22FBS643, 22IBA018; 853) REF DYNJ904906I, UDI/DI 10195327247805 (each) 40195327247806 (case), Lot Numbers: 22KBT768, 23DBD885, 23FBQ980, 23FBP020, 23HBX774; 854) REF DYNJ904993L, UDI/DI 10195327186272 (each) 40195327186273 (case), Lot Numbers: 22IBO970; 855) REF DYNJ904995G, UDI/DI 10195327165680 (each) 40195327165681 (case), Lot Numbers: 22FDA708; 856) REF DYNJ904996I, UDI/DI 10195327190286 (each) 40195327190287 (case), Lot Numbers: 23ABI076; 857) REF DYNJ905075F, UDI/DI 10195327045302 (each) 40195327045303 (case), Lot Numbers: 22FBE945, 22GBE231, 22KBM325, 22JBH247; 858) REF DYNJ905075G, UDI/DI 10195327233839 (each) 40195327233830 (case), Lot Numbers: 22KBI003, 22KBI004, 22LBH034, 23DBD746, 23DBA929, 23FBG302; 859) REF DYNJ905078F, UDI/DI 10193489437768 (each) 40193489437769 (case), Lot Numbers: 22NBF134, 22GBB090, 22KBH368, 22JBY286; 860) REF DYNJ905080G, UDI/DI 10195327046552 (each) 40195327046553 (case), Lot Numbers: 23ABA878, 23EBS459; 861) REF DYNJ905112B, UDI/DI 10193489887624 (each) 40193489887625 (case), Lot Numbers: 22EDB438, 22FDB733; 862) REF DYNJ905112C, UDI/DI 10195327224158 (each) 40195327224159 (case), Lot Numbers: 22JDA094, 23ADA503, 22KDA336, 23BDB721; 863) REF DYNJ905196I, UDI/DI 10193489466744 (each) 40193489466745 (case), Lot Numbers: 22HME748, 22GMF184, 22FMA081; 864) REF DYNJ905196J, UDI/DI 10195327230821 (each) 40195327230822 (case), Lot Numbers: 22JMD693, 23AMI844, 23AME521, 22JMF861, 23DMB374, 22LMB547, 23CMD438; 865) REF DYNJ905334C, UDI/DI 10195327638016 (each) 40195327638017 (case), Lot Numbers: 24EBG630, 24CBO936, 24HBO616, 24GBH746; 866) REF DYNJ905443C, UDI/DI 10193489883688 (each) 40193489883689 (case), Lot Numbers: 22JBB831, 22IBN307, 22GBY782, 22GBK088, 22FBP816, 22DBO430, 23BBA885, 22KBV220, 23EBE319, 23CBO783; 867) REF DYNJ905464G, UDI/DI 10193489841022 (each) 40193489841023 (case), Lot Numbers: 22GBM716, 22FBV950; 868) REF DYNJ905464I, UDI/DI 10195327217549 (each) 40195327217540 (case), Lot Numbers: 23DBK508, 23CBP055, 23GBU897, 23EBN913, 23IBW052, 23IBQ658, 23HBJ896, 23JBU586; 869) REF DYNJ905469C, UDI/DI 10195327200282 (each) 40195327200283 (case), Lot Numbers: 22KBR546, 22KBR551, 23CBV484, 23EBB489, 23JBI044, 23KBR699, 23KBR556, 24BBT857; 870) REF DYNJ905471B, UDI/DI 10195327200510 (each) 40195327200511 (case), Lot Numbers: 23DBI548, 22KBR568, 22KBR569, 23HBY031, 24BBK748, 23LBN256, 23JBP241, 23JBF443; 871) REF DYNJ905604F, UDI/DI 10195327066550 (each) 40195327066551 (case), Lot Numbers: 22HBK220, 22GBW114, 22GBW115, 22KBV697; 872) REF DYNJ905604G, UDI/DI 10195327329662 (each) 40195327329663 (case), Lot Numbers: 23EBF189, 23CBP671; 873) REF DYNJ905605F, UDI/DI 10195327066567 (each) 40195327066568 (case), Lot Numbers: 22EBG716, 22EBI074, 22KBO825, 23BBA688; 874) REF DYNJ905605G, UDI/DI 10195327329679 (each) 40195327329670 (case), Lot Numbers: 23FBF791; 875) REF DYNJ905617F, UDI/DI 10195327675851 (each) 40195327675852 (case), Lot Numbers: 24EBG331; 876) REF DYNJ905813D, UDI/DI 10195327094188 (each) 40195327094189 (case), Lot Numbers: 23CDC412, 23ADC095, 23IDB681, 23HDB420, 23LDA261, 23KDA402, 24ADB902; 877) REF DYNJ905814F, UDI/DI 10195327129606 (each) 40195327129607 (case), Lot Numbers: 22JDB123, 22GDA019, 22KDC152, 23DDB941, 23BDB252, 23GDA260, 23IDB391, 23HDB799, 24BDB864, 24ADA055; 878) REF DYNJ905974A, UDI/DI 10195327037727 (each) 40195327037728 (case), Lot Numbers: 22DBU738; 879) REF DYNJ905974B, UDI/DI 10195327311315 (each) 40195327311316 (case), Lot Numbers: 23GBO163, 24BBE649, 23KBG150, 23IBL206; 880) REF DYNJ905978A, UDI/DI 10195327030629 (each) 40195327030620 (case), Lot Numbers: 23ABS782, 23ABQ824, 22GBO887, 23ABM782, 23FBR258, 23EBN920; 881) REF DYNJ905992D, UDI/DI 10195327451301 (each) 40195327451302 (case), Lot Numbers: 23GBS854, 23HBX882; 882) REF DYNJ905992F, UDI/DI 10195327527211 (each) 40195327527212 (case), Lot Numbers: 24BBB309, 24ABV739, 24DBN864; 883) REF DYNJ906053D, UDI/DI 10193489984576 (each) 40193489984577 (case), Lot Numbers: 22DDC498; 884) REF DYNJ906053G, UDI/DI 10195327293062 (each) 40195327293063 (case), Lot Numbers: 22KDB741; 885) REF DYNJ906053I, UDI/DI 10195327313333 (each) 40195327313334 (case), Lot Numbers: 23CDA692, 23EDB002, 23ADA201; 886) REF DYNJ906053J, UDI/DI 10195327518233 (each) 40195327518234 (case), Lot Numbers: 23KDC354, 24ADA141, 24BDA491; 887) REF DYNJ906161A, UDI/DI 10193489233278 (each) 40193489233279 (case), Lot Numbers: 22JLA637, 23FLA345, 22LLA948, 22LLA588, 23CLB249, 23BLA914, 23ILA274, 23GLA749, 23LLA802, 23KLA293, 23KLA415, 23JLA817, 23LLA905, 24ALA873, 24CLA142, 24ELA833; 888) REF DYNJ906163B, UDI/DI 10193489819410 (each) 40193489819411 (case), Lot Numbers: 22FBV613, 22HBS633, 22KBM272, 23BBE573, 23LBA471, 23FBN758, 23EBL829, 23KBK823, 23GBE754, 24CBA354, 24BBF086; 889) REF DYNJ906166A, UDI/DI 10195327036607 (each) 40195327036608 (case), Lot Numbers: 22KBT481, 23EBT073, 23BBQ295, 23BBG458, 23DBM963, 23IBK707, 24CBP673, 24ABQ967, 24EBC230; 890) REF DYNJ906168C, UDI/DI 10195327120658 (each) 40195327120659 (case), Lot Numbers: 23FBD132, 23BBP552, 23IBM223; 891) REF DYNJ906223B, UDI/DI 10195327344405 (each) 40195327344406 (case), Lot Numbers: 23CBX460, 23CBI971; 892) REF DYNJ906261C, UDI/DI 10195327344436 (each) 40195327344437 (case), Lot Numbers: 23DBE791; 893) REF DYNJ906262C, UDI/DI 10195327344368 (each) 40195327344369 (case), Lot Numbers: 23FBR888; 894) REF DYNJ906317G, UDI/DI 10195327370909 (each) 40195327370900 (case), Lot Numbers: 23JBT885, 23IBP592, 23GBK296, 23GBA476; 895) REF DYNJ906409A, UDI/DI 10193489874853 (each) 40193489874854 (case), Lot Numbers: 22KBN513, 22FBS708; 896) REF DYNJ906409C, UDI/DI 10195327319809 (each) 40195327319800 (case), Lot Numbers: 23KBC002; 897) REF DYNJ906448C, UDI/DI 10195327167578 (each) 40195327167579 (case), Lot Numbers: 23BBR525, 22GBD044, 22GBD045, 23ABQ954, 22LBL552; 898) REF DYNJ906448F, UDI/DI 10195327436421 (each) 40195327436422 (case), Lot Numbers: 23JBH738, 23KBC876; 899) REF DYNJ906450C, UDI/DI 10193489900439 (each) 40193489900430 (case), Lot Numbers: 22EBB644, 22EBC235; 900) REF DYNJ906450D, UDI/DI 10195327170660 (each) 40195327170661 (case), Lot Numbers: 23BBO821; 901) REF DYNJ906450F, UDI/DI 10195327348656 (each) 40195327348657 (case), Lot Numbers: 23CBP017, 23HBM407, 23KBA715; 902) REF DYNJ906451C, UDI/DI 10195327167448 (each) 40195327167449 (case), Lot Numbers: 22LBQ816; 903) REF DYNJ906451D, UDI/DI 10195327348649 (each) 40195327348640 (case), Lot Numbers: 23FBV262, 23JBA209; 904) REF DYNJ906457C, UDI/DI 10195327167561 (each) 40195327167562 (case), Lot Numbers: 22KBV019, 23ABR027, 23CBW474; 905) REF DYNJ906457D, UDI/DI 10195327436650 (each) 40195327436651 (case), Lot Numbers: 23IBD447; 906) REF DYNJ906458C, UDI/DI 10193489900200 (each) 40193489900201 (case), Lot Numbers: 22EBF057, 22FBQ883; 907) REF DYNJ906458D, UDI/DI 10195327167462 (each) 40195327167463 (case), Lot Numbers: 22GBN080, 22FBV564, 23ABQ884, 22KBW787, 23BBQ294; 908) REF DYNJ906458F, UDI/DI 10195327355883 (each) 40195327355884 (case), Lot Numbers: 23GBU813, 23KBW142, 23KBW056, 23IBF302, 24BBP102, 24GBR743, 24CBO485; 909) REF DYNJ906469D, UDI/DI 10195327041731 (each) 40195327041732 (case), Lot Numbers: 22DDA747, 22FDA670; 910) REF DYNJ906469F, UDI/DI 10195327210939 (each) 40195327210930 (case), Lot Numbers: 22HDC339, 23IDC208; 911) REF DYNJ906479C, UDI/DI 10195327024307 (each) 40195327024308 (case), Lot Numbers: 22HDB405, 22GDA236; 912) REF DYNJ906479D, UDI/DI 10195327202682 (each) 40195327202683 (case), Lot Numbers: 23DDB574, 23FDB867, 23FDB142, 23HDC215, 23JDB994, 23JDB288; 913) REF DYNJ906485D, UDI/DI 10195327025250 (each) 40195327025251 (case), Lot Numbers: 22GDB445; 914) REF DYNJ906485F, UDI/DI 10195327202767 (each) 40195327202768 (case), Lot Numbers: 23EDA580, 23CDB072, 23JDB048; 915) REF DYNJ906495F, UDI/DI 10195327210977 (each) 40195327210978 (case), Lot Numbers: 23GDA783, 23FDB527; 916) REF DYNJ906496F, UDI/DI 10195327370954 (each) 40195327370955 (case), Lot Numbers: 23GLA582, 23FLA898, 23JLA675, 23ILA204, 23ILA505, 23ILA840; 917) REF DYNJ906539F, UDI/DI 10195327427412 (each) 40195327427413 (case), Lot Numbers: 23GBE724; 918) REF DYNJ906542A, UDI/DI 10193489270020 (each) 40193489270021 (case), Lot Numbers: 22HDB047; 919) REF DYNJ906542B, UDI/DI 10195327221874 (each) 40195327221875 (case), Lot Numbers: 23FDA462, 24GDA418; 920) REF DYNJ906602C, UDI/DI 10193489466768 (each) 40193489466769 (case), Lot Numbers: 22GME256, 22FMD475; 921) REF DYNJ906602D, UDI/DI 10195327230814 (each) 40195327230815 (case), Lot Numbers: 23DME802, 23BMG010, 23BMB203, 22LMB267, 23GMG878; 922) REF DYNJ906614B, UDI/DI 10193489345322 (each) 40193489345323 (case), Lot Numbers: 22FBS243, 23ABD346, 23ABQ836, 23LBH022, 23IBV173, 23IBV010, 23GBC569, 23GBD902, 23KBP690, 23KBP510; 923) REF DYNJ906651F, UDI/DI 10193489992342 (each) 40193489992343 (case), Lot Numbers: 22LDB126, 22KDC114; 924) REF DYNJ906658F, UDI/DI 10195327000899 (each) 40195327000890 (case), Lot Numbers: 22JDA648, 22KDA427, 22LDA989; 925) REF DYNJ906660G, UDI/DI 10195327070038 (each) 40195327070039 (case), Lot Numbers: 22IDB435, 22GDB475, 22JDB999, 22KDC189, 23CDC432; 926) REF DYNJ906664G, UDI/DI 10195327070175 (each) 40195327070176 (case), Lot Numbers: 22LDB485, 23BDA224, 23GDA475, 23FDA867; 927) REF DYNJ906665G, UDI/DI 10195327092481 (each) 40195327092482 (case), Lot Numbers: 22EDB767, 22EDA992, 22CDC258, 22FDA817, 22LDA149, 22IDA259, 22GDB555, 23ADB845; 928) REF DYNJ906666J, UDI/DI 10195327129590 (each) 40195327129591 (case), Lot Numbers: 22GDA384, 22FDA952, 22FDA056, 22HDA321, 23ABR423, 23ABB003, 22KBC065, 23CBA140, 23ABR422, 22GDB274; 929) REF DYNJ906723A, UDI/DI 10195327059866 (each) 40195327059867 (case), Lot Numbers: 22DDC232, 22FDA614; 930) REF DYNJ906890B, UDI/DI 10195327303587 (each) 40195327303588 (case), Lot Numbers: 23FBO221, 23KBT243, 23KBE431, 23CBP900; 931) REF DYNJ906891C, UDI/DI 10195327303570 (each) 40195327303571 (case), Lot Numbers: 23CBP901, 24DBS776, 24CBN224; 932) REF DYNJ906892A, UDI/DI 10195327303426 (each) 40195327303427 (case), Lot Numbers: 23DBQ690, 23HBN943, 23KBG736, 23GBV420, 23KBB202; 933) REF DYNJ906893D, UDI/DI 10195327303372 (each) 40195327303373 (case), Lot Numbers: 23IMC078; 934) REF DYNJ906894A, UDI/DI 10195327303365 (each) 40195327303366 (case), Lot Numbers: 23HLB075, 23ELA244, 23CLA420, 23BLA206; 935) REF DYNJ906911A, UDI/DI 10195327303594 (each) 40195327303595 (case), Lot Numbers: 23CBH442, 23ABC522, 23FBV559, 23FBC212, 23HBS918; 936) REF DYNJ906950B, UDI/DI 10195327381011 (each) 40195327381012 (case), Lot Numbers: 23HBB790, 23EBK134, 23KBR567, 23KBI260, 23LBB806; 937) REF DYNJ906950C, UDI/DI 10195327566807 (each) 40195327566808 (case), Lot Numbers: 24ABW981, 24DBD940, 24ABL051; 938) REF DYNJ906955D, UDI/DI 10195327381141 (each) 40195327381142 (case), Lot Numbers: 23HBB476, 23KBQ769, 23KBP472, 23LBB805, 23FBC881, 23IBP953; 939) REF DYNJ906956C, UDI/DI 10195327381158 (each) 40195327381159 (case), Lot Numbers: 23HBC932, 23EBG857, 23EBC449, 23KBP686, 23KBK837, 23LBP373, 23FBI132; 940) REF DYNJ906956D, UDI/DI 10195327567125 (each) 40195327567126 (case), Lot Numbers: 24ABH035, 24BBC929, 24DBP220; 941) REF DYNJ906957B, UDI/DI 10195327381165 (each) 40195327381166 (case), Lot Numbers: 23FBI110, 23KBO184, 23KBB084, 23EBM815; 942) REF DYNJ906957C, UDI/DI 10195327567132 (each) 40195327567133 (case), Lot Numbers: 24CBM172, 24BBJ355, 24ABH098, 24DBN816; 943) REF DYNJ906974F, UDI/DI 10195327502881 (each) 40195327502882 (case), Lot Numbers: 23KBB168, 23JBU563, 23LBS348; 944) REF DYNJ906974I, UDI/DI 10195327621414 (each) 40195327621415 (case), Lot Numbers: 24GBH723, 24CBM366, 24DBG769; 945) REF DYNJ906975D, UDI/DI 10195327502874 (each) 40195327502875 (case), Lot Numbers: 23LLA276, 23KLA951; 946) REF DYNJ907022C, UDI/DI 10193489846577 (each) 40193489846578 (case), Lot Numbers: 22GBM076, 22HBS352, 22EBA303, 23BBL462, 23ABI384, 23EBB693, 23DBH227; 947) REF DYNJ907038D, UDI/DI 10195327228163 (each) 40195327228164 (case), Lot Numbers: 22IBG109, 22IBG110, 23ABL541, 23ABB862, 22KBI775, 22JBX889; 948) REF DYNJ907038F, UDI/DI 10195327372682 (each) 40195327372683 (case), Lot Numbers: 23EBJ311, 23CBX587, 23CBQ477, 23FBO835; 949) REF DYNJ907222B, UDI/DI 10195327324278 (each) 40195327324279 (case), Lot Numbers: 23EMD890, 23EMB435; 950) REF DYNJ907227C, UDI/DI 10195327145378 (each) 40195327145379 (case), Lot Numbers: 22FBO226, 22FBI627; 951) REF DYNJ907227D, UDI/DI 10195327247430 (each) 40195327247431 (case), Lot Numbers: 22JBR773, 23FBF615, 23ABC457, 23KBF400, 23KBB212, 23LBO474, 24FMI517; 952) REF DYNJ907249B, UDI/DI 10195327138585 (each) 40195327138586 (case), Lot Numbers: 22KME146, 22JMI139, 23CMG071, 23BMD847, 23BMG030, 23BMF499, 23GMC765, 23CMC866; 953) REF DYNJ907249C, UDI/DI 10195327486037 (each) 40195327486038 (case), Lot Numbers: 23LMH533, 23KMD710, 24AMH722, 24CMD208, 24DMB386, 24EMG375; 954) REF DYNJ907252A, UDI/DI 10193489872484 (each) 40193489872485 (case), Lot Numbers: 22GBS302, 22DBH380, 22DBJ406, 22JBR774, 22LBC149, 23DBM247, 23EBP777, 23KBG206, 23HBF251, 23LBG693; 955) REF DYNJ907253A, UDI/DI 10193489879506 (each) 40193489879507 (case), Lot Numbers: 22GBV712, 22GBR355, 22GBK981, 22DBG367, 22DBF377, 22LBP264, 23DBL869, 23EBQ837, 23EBR132, 23GBG859, 23LBM201, 23LBG921, 23KBF383; 956) REF DYNJ907330F, UDI/DI 10195327345471 (each) 40195327345472 (case), Lot Numbers: 23JBL940, 23IBQ659, 23KBT044; 957) REF DYNJ907331I, UDI/DI 10195327345518 (each) 40195327345519 (case), Lot Numbers: 23CBH324, 23FBF070, 23JBH942; 958) REF DYNJ907394B, UDI/DI 10195327359225 (each) 40195327359226 (case), Lot Numbers: 23DMH233, 23EMD892; 959) REF DYNJ907398A, UDI/DI 10195327324384 (each) 40195327324385 (case), Lot Numbers: 23EME758; 960) REF DYNJ907403A, UDI/DI 10195327359218 (each) 40195327359219 (case), Lot Numbers: 23GMH562, 23EMC989; 961) REF DYNJ907412A, UDI/DI 10195327335731 (each) 40195327335732 (case), Lot Numbers: 23DMF107, 23EME405; 962) REF DYNJ907419A, UDI/DI 10195327358754 (each) 40195327358755 (case), Lot Numbers: 23FMD834, 23FMA440, 23DME759; 963) REF DYNJ907420A, UDI/DI 10195327358761 (each) 40195327358762 (case), Lot Numbers: 23DMF108, 23FME339, 23EMK083, 23HMA290; 964) REF DYNJ907487F, UDI/DI 10195327515560 (each) 40195327515561 (case), Lot Numbers: 23KBW901; 965) REF DYNJ907516G, UDI/DI 10195327345532 (each) 40195327345533 (case), Lot Numbers: 23DBD216, 23JBE155; 966) REF DYNJ907516I, UDI/DI 10198459059469 (each) 40198459059460 (case), Lot Numbers: 24HBU187, 24GBR656; 967) REF DYNJ907526F, UDI/DI 10195327345501 (each) 40195327345502 (case), Lot Numbers: 23JBA742; 968) REF DYNJ907527G, UDI/DI 10195327345327 (each) 40195327345328 (case), Lot Numbers: 23DBN688, 23IBA261; 969) REF DYNJ907528F, UDI/DI 10195327345334 (each) 40195327345335 (case), Lot Numbers: 23EBA831, 23GBI252, 23HBX983, 23JBS791; 970) REF DYNJ907581A, UDI/DI 10195327045296 (each) 40195327045297 (case), Lot Numbers: 22GBD237, 22IBU694, 23CBM613; 971) REF DYNJ907588I, UDI/DI 10195327345341 (each) 40195327345342 (case), Lot Numbers: 23CBJ928, 23DBP536, 23FBN592, 23FBN449, 23IBB003, 23KBG198; 972) REF DYNJ907788C, UDI/DI 10195327141059 (each) 40195327141050 (case), Lot Numbers: 22HBA124; 973) REF DYNJ907788F, UDI/DI 10195327273057 (each) 40195327273058 (case), Lot Numbers: 23CBV889, 23DBD546; 974) REF DYNJ907790B, UDI/DI 10193489999273 (each) 40193489999274 (case), Lot Numbers: 22GBA671; 975) REF DYNJ907790C, UDI/DI 10195327185305 (each) 40195327185306 (case), Lot Numbers: 22HBS804, 23EBT254, 23CBX589; 976) REF DYNJ907790D, UDI/DI 10195327431655 (each) 40195327431656 (case), Lot Numbers: 23GBH340; 977) REF DYNJ908005, UDI/DI 10193489960372 (each) 40193489960373 (case), Lot Numbers: 22GBX610, 22HBS815, 23ABM610, 23DBE071; 978) REF DYNJ908009A, UDI/DI 10195327113674 (each) 40195327113675 (case), Lot Numbers: 22FMD128, 22FMG824; 979) REF DYNJ908009B, UDI/DI 10195327257347 (each) 40195327257348 (case), Lot Numbers: 23FMH537, 23DMC858, 23DMB913, 23JMD035; 980) REF DYNJ908063A, UDI/DI 10195327125554 (each) 40195327125555 (case), Lot Numbers: 23DBD691, 23EBA988, 23EBN881, 23EBC773; 981) REF DYNJ908067A, UDI/DI 10195327125721 (each) 40195327125722 (case), Lot Numbers: 22HBH011, 22FBW369, 22FBW370, 22FBO644, 22EBO471, 23DBE108; 982) REF DYNJ908070A, UDI/DI 10195327125752 (each) 40195327125753 (case), Lot Numbers: 22EBM394, 22FBG080, 22EBU045, 22EBN035, 23ABM724; 983) REF DYNJ908071A, UDI/DI 10195327125769 (each) 40195327125760 (case), Lot Numbers: 22EBJ761, 22EBJ762, 23DLA327, 23ELA964, 23HLA995, 23FLA953; 984) REF DYNJ908073, UDI/DI 10193489967869 (each) 40193489967860 (case), Lot Numbers: 22KBE178, 22KBP162, 23CBM743, 23ABJ520, 23GBH156; 985) REF DYNJ908173C, UDI/DI 10195327345457 (each) 40195327345458 (case), Lot Numbers: 23CBC564, 23FBV454, 23FBN722, 23HBY069, 23JBG923; 986) REF DYNJ908256, UDI/DI 10195327016159 (each) 40195327016150 (case), Lot Numbers: 22KBG059, 22KBF265, 23ABM691; 987) REF DYNJ908290A, UDI/DI 10195327216610 (each) 40195327216611 (case), Lot Numbers: 22JMB449; 988) REF DYNJ908292, UDI/DI 10195327029289 (each) 40195327029280 (case), Lot Numbers: 23BBS142, 23ABP750, 22JBU899, 23FBF303, 23EBF157, 23EBC430, 24ABG432, 23KBJ003, 23JBV879, 24BBJ154, 23LBK688, 24CBU510; 989) REF DYNJ908295, UDI/DI 10195327029579 (each) 40195327029570 (case), Lot Numbers: 22FBP832, 22JBQ931, 23EBL980, 23ABH904, 23CBI554, 23BBO814, 23KBT462, 23KBC620, 23LBU359, 24BBC925, 24CBF715, 24DBN038; 990) REF DYNJ908296, UDI/DI 10195327033392 (each) 40195327033393 (case), Lot Numbers: 22FBS143, 22FBS144; 991) REF DYNJ908299, UDI/DI 10195327030285 (each) 40195327030286 (case), Lot Numbers: 23ABQ876, 23DBE301, 23DBD644, 23ABM209, 23ABF344, 23BBR968, 23EBO972; 992) REF DYNJ908331B, UDI/DI 10195327449636 (each) 40195327449637 (case), Lot Numbers: 23HBG005; 993) REF DYNJ908332B, UDI/DI 10195327474416 (each) 40195327474417 (case), Lot Numbers: 23JBM281; 994) REF DYNJ908345, UDI/DI 10195327048365 (each) 40195327048366 (case), Lot Numbers: 22LME372; 995) REF DYNJ908346, UDI/DI 10195327048372 (each) 40195327048373 (case), Lot Numbers: 22GMB091, 22FMF893, 22HMF816; 996) REF DYNJ908348, UDI/DI 10195327048501 (each) 40195327048502 (case), Lot Numbers: 22GMD710, 22HMF303, 22FMA020, 22LMH351, 22LMF421, 22JMF319, 22KMB101; 997) REF DYNJ908363, UDI/DI 10195327058319 (each) 40195327058310 (case), Lot Numbers: 22DDB949; 998) REF DYNJ908363A, UDI/DI 10195327163082 (each) 40195327163083 (case), Lot Numbers: 22FDA839, 22GDB474; 999) REF DYNJ908363D, UDI/DI 10195327369453 (each) 40195327369454 (case), Lot Numbers: 23HDA539, 23GDA374, 23KDA772; 1000) REF DYNJ908363F, UDI/DI 10195327576332 (each) 40195327576333 (case), Lot Numbers: 24ADB827, 24BDA921, 24EDA369, 24CDB360, 24FDA234; 1001) REF DYNJ908374A, UDI/DI 10195327175696 (each) 40195327175697 (case), Lot Numbers: 22FDB256, 22FDB975, 22HDB892, 22HDB095; 1002) REF DYNJ908374D, UDI/DI 10195327369460 (each) 40195327369461 (case), Lot Numbers: 23JDA117, 23FDA796, 23IDA784; 1003) REF DYNJ908519A, UDI/DI 10195327470401 (each) 40195327470402 (case), Lot Numbers: 23KBS441; 1004) REF DYNJ908520, UDI/DI 10195327082178 (each) 40195327082179 (case), Lot Numbers: 22IBV995, 22EBG946, 22LBF448, 23BBF957; 1005) REF DYNJ908520A, UDI/DI 10195327342302 (each) 40195327342303 (case), Lot Numbers: 23CBO976, 23BBP231, 23GBG150; 1006) REF DYNJ908520B, UDI/DI 10195327470470 (each) 40195327470471 (case), Lot Numbers: 23JBY177, 23LBB383, 23HBT110; 1007) REF DYNJ908521, UDI/DI 10195327082161 (each) 40195327082162 (case), Lot Numbers: 22FBU636, 22FBK410, 22IBH122, 23BBK665; 1008) REF DYNJ908521A, UDI/DI 10195327342296 (each) 40195327342297 (case), Lot Numbers: 23DBB093, 23FBB333; 1009) REF DYNJ908521B, UDI/DI 10195327470371 (each) 40195327470372 (case), Lot Numbers: 23IBC633; 1010) REF DYNJ908587, UDI/DI 10195327108304 (each) 40195327108305 (case), Lot Numbers: 22GDB889; 1011) REF DYNJ908587A, UDI/DI 10195327298975 (each) 40195327298976 (case), Lot Numbers: 23ADB763, 23ADA639; 1012) REF DYNJ908587B, UDI/DI 10195327369477 (each) 40195327369478 (case), Lot Numbers: 23GDA286, 23HDB750; 1013) REF DYNJ908587C, UDI/DI 10195327576219 (each) 40195327576210 (case), Lot Numbers: 24BDA532, 24DDB345; 1014) REF DYNJ908638, UDI/DI 10195327122287 (each) 40195327122288 (case), Lot Numbers: 22EBR654, 22EBL550, 22KBW220, 22KBP678, 23BBN934, 23ABQ841, 23ABO551, 23DBE144, 23DBB106, 23CBP300; 1015) REF DYNJ908642A, UDI/DI 10195327168407 (each) 40195327168408 (case), Lot Numbers: 22FBR725, 22JBL299; 1016) REF DYNJ908642B, UDI/DI 10195327243432 (each) 40195327243433 (case), Lot Numbers: 22KBS080, 22KBA822, 22JBE869; 1017) REF DYNJ908642C, UDI/DI 10195327315009 (each) 40195327315000 (case), Lot Numbers: 23ABP725, 23CBQ351, 23DBH519; 1018) REF DYNJ908644A, UDI/DI 10195327168650 (each) 40195327168651 (case), Lot Numbers: 22FBP241; 1019) REF DYNJ908644B, UDI/DI 10195327243449 (each) 40195327243440 (case), Lot Numbers: 22IBL137, 22KBS065; 1020) REF DYNJ908644C, UDI/DI 10195327315016 (each) 40195327315017 (case), Lot Numbers: 23BBL706, 23CBT963; 1021) REF DYNJ908647A, UDI/DI 10195327168438 (each) 40195327168439 (case), Lot Numbers: 22IBE709, 22JBH372, 23ABM544, 22JBR824, 23EBG073; 1022) REF DYNJ908674, UDI/DI 10195327136581 (each) 40195327136582 (case), Lot Numbers: 22GDA786; 1023) REF DYNJ908674A, UDI/DI 10195327251741 (each) 40195327251742 (case), Lot Numbers: 22LDA348; 1024) REF DYNJ908674B, UDI/DI 10195327334000 (each) 40195327334001 (case), Lot Numbers: 23DDC352; 1025) REF DYNJ908674C, UDI/DI 10195327428952 (each) 40195327428953 (case), Lot Numbers: 23JDA143, 23IDA761, 23HDC130, 23GDB834, 23GDA042; 1026) REF DYNJ908676B, UDI/DI 10195327434861 (each) 40195327434862 (case), Lot Numbers: 23HDB965, 23JDC407; 1027) REF DYNJ908681, UDI/DI 10195327139681 (each) 40195327139682 (case), Lot Numbers: 22GDB816, 22LDA354, 23DDA403, 23JDC215, 23KDB420, 23IDA575, 24ADC245; 1028) REF DYNJ908682B, UDI/DI 10195327334031 (each) 40195327334032 (case), Lot Numbers: 23JDC268; 1029) REF DYNJ908683, UDI/DI 10195327139698 (each) 40195327139699 (case), Lot Numbers: 22HDC171; 1030) REF DYNJ908683A, UDI/DI 10195327251772 (each) 40195327251773 (case), Lot Numbers: 22KDC147; 1031) REF DYNJ908683B, UDI/DI 10195327396435 (each) 40195327396436 (case), Lot Numbers: 23IDB009; 1032) REF DYNJ908780, UDI/DI 10195327183356 (each) 40195327183357 (case), Lot Numbers: 23CDA809, 23FDA816, 23HDB677, 23DDA731; 1033) REF DYNJ908782, UDI/DI 10195327183325 (each) 40195327183326 (case), Lot Numbers: 23FDA586, 23IDB890; 1034) REF DYNJ908786, UDI/DI 10195327183332 (each) 40195327183333 (case), Lot Numbers: 23CDB081, 23JME836, 23IDA376; 1035) REF DYNJ908798, UDI/DI 10195327183349 (each) 40195327183340 (case), Lot Numbers: 23CDB381, 23FDA714, 23HDC011; 1036) REF DYNJ908799, UDI/DI 10195327183370 (each) 40195327183371 (case), Lot Numbers: 23DDB167, 23FDA791, 23IDC106, 23HDA802; 1037) REF DYNJ908800, UDI/DI 10195327183387 (each) 40195327183388 (case), Lot Numbers: 23FDB403, 23HDC467; 1038) REF DYNJ908800A, UDI/DI 10195327520816 (each) 40195327520817 (case), Lot Numbers: 23KBO196, 23KBD414, 23KBW418; 1039) REF DYNJ908801, UDI/DI 10195327183394 (each) 40195327183395 (case), Lot Numbers: 22IDA928, 23IDA359, 23GDB160; 1040) REF DYNJ908801A, UDI/DI 10195327520809 (each) 40195327520800 (case), Lot Numbers: 23LBN623, 23KBE475, 23JBY024; 1041) REF DYNJ908802, UDI/DI 10195327183363 (each) 40195327183364 (case), Lot Numbers: 23DDB298, 23FDA792, 23HDB652; 1042) REF DYNJ908802A, UDI/DI 10195327520793 (each) 40195327520794 (case), Lot Numbers: 23LBT966, 23KBT242; 1043) REF DYNJ908811A, UDI/DI 10195327525422 (each) 40195327525423 (case), Lot Numbers: 23KMI866; 1044) REF DYNJ908816, UDI/DI 10195327186531 (each) 40195327186532 (case), Lot Numbers: 22KDB737; 1045) REF DYNJ908817, UDI/DI 10195327186548 (each) 40195327186549 (case), Lot Numbers: 23JMB238, 23DDC016, 23JMI125, 23IDA754; 1046) REF DYNJ908819, UDI/DI 10195327185626 (each) 40195327185627 (case), Lot Numbers: 23EDA729, 23DDB170, 23IDA168, 23LDA179, 23JDA966; 1047) REF DYNJ908828, UDI/DI 10195327185725 (each) 40195327185726 (case), Lot Numbers: 23FDA575, 23KMD570, 23IDB082; 1048) REF DYNJ908831, UDI/DI 10195327185640 (each) 40195327185641 (case), Lot Numbers: 23CDB151, 23IDA920; 1049) REF DYNJ908832, UDI/DI 10195327185657 (each) 40195327185658 (case), Lot Numbers: 23EDA173, 23FDA820, 23IMH088, 23IDB313, 23JMI448; 1050) REF DYNJ908865C, UDI/DI 10195327481988 (each) 40195327481989 (case), Lot Numbers: 23ILA619, 23JLA397; 1051) REF DYNJ908979, UDI/DI 10195327238995 (each) 40195327238996 (case), Lot Numbers: 22KBB139; 1052) REF DYNJ908998, UDI/DI 10195327237783 (each) 40195327237784 (case), Lot Numbers: 22KDB214; 1053) REF DYNJ909000A, UDI/DI 10195327428327 (each) 40195327428328 (case), Lot Numbers: 23HBQ467, 23GBK254, 23JBH758; 1054) REF DYNJ909002, UDI/DI 10195327238148 (each) 40195327238149 (case), Lot Numbers: 23EBI750, 23EBT346, 23FBG449; 1055) REF DYNJ909002A, UDI/DI 10195327428334 (each) 40195327428335 (case), Lot Numbers: 23HBQ478, 23KBK334; 1056) REF DYNJ909004, UDI/DI 10195327238971 (each) 40195327238972 (case), Lot Numbers: 22KDB988; 1057) REF DYNJ909004A, UDI/DI 10195327444129 (each) 40195327444120 (case), Lot Numbers: 23GDB166; 1058) REF DYNJ909005, UDI/DI 10195327238988 (each) 40195327238989 (case), Lot Numbers: 22KDB989, 23CDC590, 23GDB399, 23FDB461; 1059) REF DYNJ909019, UDI/DI 10195327247799 (each) 40195327247790 (case), Lot Numbers: 22LBP962; 1060) REF DYNJ909019A, UDI/DI 10195327596842 (each) 40195327596843 (case), Lot Numbers: 24CBK591; 1061) REF DYNJ909143B, UDI/DI 10195327498436 (each) 40195327498437 (case), Lot Numbers: 23JMD034; 1062) REF DYNJ909161, UDI/DI 10195327281854 (each) 40195327281855 (case), Lot Numbers: 23ABL819, 23EBD646; 1063) REF DYNJ909162, UDI/DI 10195327281861 (each) 40195327281862 (case), Lot Numbers: 23ABL818; 1064) REF DYNJ909380, UDI/DI 10195327323134 (each) 40195327323135 (case), Lot Numbers: 23FDB613; 1065) REF DYNJ909400, UDI/DI 10195327336103 (each) 40195327336104 (case), Lot Numbers: 23BBR478, 23FBS605, 23FBK753; 1066) REF DYNJ909405, UDI/DI 10195327338435 (each) 40195327338436 (case), Lot Numbers: 23CMB928; 1067) REF DYNJ909412, UDI/DI 10195327339814 (each) 40195327339815 (case), Lot Numbers: 23EMA940, 23KMI102, 23KMB366, 23HMC618, 23JMC097, 23JMB263, 23IMB861, 23HMI892, 23JMI583, 23JMI722, 23JMC920, 24BMF068, 24AMA070; 1068) REF DYNJ909413, UDI/DI 10195327339845 (each) 40195327339846 (case), Lot Numbers: 23EME763; 1069) REF DYNJ909414, UDI/DI 10195327339944 (each) 40195327339945 (case), Lot Numbers: 23DMB749. 23HMB359; 1070) REF DYNJ909415, UDI/DI 10195327340063 (each) 40195327340064 (case), Lot Numbers: 23FMG135, 23EMG876, 23IMG783, 23IMA715, 23JMH846, 23JMB339, 24AME242; 1071) REF DYNJ909422, UDI/DI 10195327347741 (each) 40195327347742 (case), Lot Numbers: 23DBD577; 1072) REF DYNJ909423, UDI/DI 10195327347727 (each) 40195327347728 (case), Lot Numbers: 23DBF657; 1073) REF DYNJ909424, UDI/DI 10195327347765 (each) 40195327347766 (case), Lot Numbers: 23DBF804; 1074) REF DYNJ909462, UDI/DI 10195327361563 (each) 40195327361564 (case), Lot Numbers: 23DBM644, 23GBH101, 23FBB537, 23JBD213; 1075) REF DYNJ909463, UDI/DI 10195327361570 (each) 40195327361571 (case), Lot Numbers: 23GBT173, 23EBN778, 23EBM890, 23DBR999; 1076) REF DYNJ909464, UDI/DI 10195327361594 (each) 40195327361595 (case), Lot Numbers: 23DBN690; 1077) REF DYNJ909494, UDI/DI 10195327374907 (each) 40195327374908 (case), Lot Numbers: 23HBE924; 1078) REF DYNJ909494A, UDI/DI 10195327528836 (each) 40195327528837 (case), Lot Numbers: 23KBN713, 23KBN908; 1079) REF DYNJ909500, UDI/DI 10195327384951 (each) 40195327384952 (case), Lot Numbers: 23KBN713; 1080) REF DYNJ909502, UDI/DI 10195327384968 (each) 40195327384969 (case), Lot Numbers: 23KBN908; 1081) REF DYNJ909505, UDI/DI 10195327386030 (each) 40195327386031 (case), Lot Numbers: 23FMD557; 1082) REF DYNJ909510, UDI/DI 10195327387235 (each) 40195327387236 (case), Lot Numbers: 23GLA881, 23GLA007; 1083) REF DYNJ909579, UDI/DI 10195327409715 (each) 40195327409716 (case), Lot Numbers: 23HBM873, 23JBR241; 1084) REF DYNJ909582, UDI/DI 10195327413590 (each) 40195327413591 (case), Lot Numbers: 23JBT141, 23FBP595; 1085) REF DYNJ909590A, UDI/DI 10195327501105 (each) 40195327501106 (case), Lot Numbers: 23JBQ279, 23IBS510; 1086) REF DYNJ909593, UDI/DI 10195327413644 (each) 40195327413645 (case), Lot Numbers: 23FBT925; 1087) REF DYNJ909595, UDI/DI 10195327413583 (each) 40195327413584 (case), Lot Numbers: 23JBT858, 23JBL547; 1088) REF DYNJ909625, UDI/DI 10195327415143 (each) 40195327415144 (case), Lot Numbers: 23IBL744; 1089) REF DYNJ909657, UDI/DI 10195327431532 (each) 40195327431533 (case), Lot Numbers: 23KBF228, 23FBQ974, 24BBJ335, 24ABR146, 24CBA927, 24EBD307; 1090) REF DYNJ909658, UDI/DI 10195327431549 (each) 40195327431540 (case), Lot Numbers: 24ABO335; 1091) REF DYNJ909687, UDI/DI 10195327442293 (each) 40195327442294 (case), Lot Numbers: 23IBE284; 1092) REF DYNJ909795, UDI/DI 10195327480363 (each) 40195327480364 (case), Lot Numbers: 24ABK091; 1093) REF DYNJ909796A, UDI/DI 10195327639266 (each) 40195327639267 (case), Lot Numbers: 24JBB147, 24HBD833, 24GBL455, 24DBK383; 1094) REF DYNJ909797, UDI/DI 10195327481513 (each) 40195327481514 (case), Lot Numbers: 23LLA172; 1095) REF DYNJ909930B, UDI/DI 10195327638023 (each) 40195327638024 (case), Lot Numbers: 24DBJ984, 24GBD808; 1096) REF DYNJ909970, UDI/DI 10195327522179 (each) 40195327522170 (case), Lot Numbers: 24ALA572, 24ALA502; 1097) REF DYNJ910010, UDI/DI 10195327542443 (each) 40195327542444 (case), Lot Numbers: 24ADC084; 1098) REF DYNJ910068, UDI/DI 10195327567163 (each) 40195327567164 (case), Lot Numbers: 24BBN397; 1099) REF DYNJ910071, UDI/DI 10195327567194 (each) 40195327567195 (case), Lot Numbers: 24BBN398; 1100) REF DYNJ910085, UDI/DI 10195327569105 (each) 40195327569106 (case), Lot Numbers: 23LBO017, 24CBM480, 24BBR419, 24BBD693; 1101) REF DYNJ910132, UDI/DI 10195327597214 (each) 40195327597215 (case), Lot Numbers: 24DBI622; 1102) REF DYNJ910418, UDI/DI 10195327681531 (each) 40195327681532 (case), Lot Numbers: 24EBH428; 1103) REF DYNJ910419, UDI/DI 10195327681555 (each) 40195327681556 (case), Lot Numbers: 24EBM474; 1104) REF DYNJC9270K, UDI/DI 10195327109172 (each) 40195327109173 (case), Lot Numbers: 22HDA634, 23CDA802, 23GDC010; 1105) REF DYNJC9270L, UDI/DI 10195327476236 (each) 40195327476237 (case), Lot Numbers: 23HDB757; 1106) REF DYNJC9271N, UDI/DI 10195327109189 (each) 40195327109180 (case), Lot Numbers: 23ABD952, 22LBF230, 23BBU052, 23ABR018, 23HBP102; 1107) REF DYNJC9271O, UDI/DI 10195327476007 (each) 40195327476008 (case), Lot Numbers: 23KBG170; 1108) REF DYNJC9272O, UDI/DI 10195327109097 (each) 40195327109098 (case), Lot Numbers: 22FDB204, 23FDA025, 23HDB305, 23HDA828; 1109) REF DYNJC9272P, UDI/DI 10195327476229 (each) 40195327476220 (case), Lot Numbers: 23IDC228; 1110) REF DYNJC9309K, UDI/DI 10195327109103 (each) 40195327109104 (case), Lot Numbers: 22GBN804, 23ABC988, 23DBM210; 1111) REF DYNJC9320I, UDI/DI 10193489989939 (each) 40193489989930 (case), Lot Numbers: 23HDA687; 1112) REF DYNJC9323K, UDI/DI 10195327027490 (each) 40195327027491 (case), Lot Numbers: 23FBG142, 23LBT898, 24DBC169, 24EBC455, 24DBR463; 1113) REF DYNJC9324G, UDI/DI 10193489989946 (each) 40193489989947 (case), Lot Numbers: 22IBE723, 22FBP755, 22JBQ158, 22HBD446, 22GBX113, 22HBH857; 1114) REF DYNJV0277J, UDI/DI 10193489972146 (each) 40193489972147 (case), Lot Numbers: 22GMA861, 22JMG291, 23AMF743, 22KMD625, 23EMH516; 1115) REF DYNJV0277K, UDI/DI 10195327441074 (each) 40195327441075 (case), Lot Numbers: 23FMI731; 1116) REF DYNJV0279I, UDI/DI 10193489971989 (each) 40193489971980 (case), Lot Numbers: 22CMB321; 1117) REF DYNJV0279J, UDI/DI 10195327161798 (each) 40195327161799 (case), Lot Numbers: 22EME514, 23BMA404, 22JME083, 22JME084, 23DMC089, 23EMJ121; 1118) REF DYNJVA900018B, UDI/DI 10193489379518 (each) 40193489379519 (case), Lot Numbers: 23BLB028; 1119) REF DYNJVB1276C, UDI/DI 10195327295158 (each) 40195327295159 (case), Lot Numbers: 23ALA976, 23GLA480, 23GLA099, 23DLB133, 23CLA044, 22LLA333, 24CLA095, 24ALA960, 23KLB076, 23HLB285, 24DLA755, 24DLA325; 1120) REF DYNJVB1277D, UDI/DI 10195327482107 (each) 40195327482108 (case), Lot Numbers: 23HLA529; 1121) REF DYNJVB91006, UDI/DI 10193489801552 (each) 40193489801553 (case), Lot Numbers: 22ILB119, 22KLA992, 23DLA379, 23ALB020, 23ALA419, 23GLA235, 23ELB243; 1122) REF DYNJVB91038B, UDI/DI 10195327171759 (each) 40195327171750 (case), Lot Numbers: 23KDA679, 23HDB476, 23HDA877, 24BDB627, 23BDB258, 23BDA097, 24BDB591, 23CDC501, 23EDB105; 1123) REF DYNJVB91047C, UDI/DI 10195327171797 (each) 40195327171798 (case), Lot Numbers: 22JDA678; 1124) REF DYNJVB91047D, UDI/DI 10195327312299 (each) 40195327312290 (case), Lot Numbers: 23GDB483, 23EDB468, 23IDA442; 1125) REF DYNJVB91048D, UDI/DI 10195327171704 (each) 40195327171705 (case), Lot Numbers: 22JDA345, 22GDA981, 22KDB953; 1126) REF DYNJVB91048F, UDI/DI 10195327312305 (each) 40195327312306 (case), Lot Numbers: 23BDA154, 23EDB601, 23KDA631, 23JDB496, 23IDC212, 24ADA251, 24ADC222, 24DDA319, 24CDC026, 24DDC083; 1127) REF DYNJVB91059, UDI/DI 10193489801187 (each) 40193489801188 (case), Lot Numbers: 22HLA783, 22ELB050, 22FLA771; 1128) REF DYNJVB91059B, UDI/DI 10195327550158 (each) 40195327550159 (case), Lot Numbers: 24ALA957, 24CLA534, 24DLA595; 1129) REF DYNJVB91128B, UDI/DI 10193489986600 (each) 40193489986601 (case), Lot Numbers: 22ILA838; 1130) REF DYNJVB91128C, UDI/DI 10195327295165 (each) 40195327295166 (case), Lot Numbers: 23CLA774, 22LLA614; 1131) REF DYNJVB91128D, UDI/DI 10195327367916 (each) 40195327367917 (case), Lot Numbers: 23ILA486, 23HLA869, 23HLA337, 23GLA693, 23FLA076, 23DLA842, 24CLA151, 23JLA241, 24DLB010, 24CLA896; 1132) REF PHS102077N, UDI/DI 10195327353759 (each) 40195327353750 (case), Lot Numbers: 23DDA851, 23FDA520, 23EDB088, 23IDC335, 23HDA278
FEI Number 1417592
Recalling Firm/
Manufacturer
MEDLINE INDUSTRIES, LP - Northfield
3 Lakes Dr
Northfield IL 60093-2753
For Additional Information ContactHaley Barclay
866-359-1704
Manufacturer Reason
for Recall
Medline is recalling medical convenience manufactured using certain lots of SafeAir Smoke Evacuation Pencil which were subsequently recalled by Stryker.
FDA Determined
Cause 2
Under Investigation by firm
ActionMedline issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 10/11/2024 via mail and email. The notice explained the problem with the Stryker component, potential risk, requested the kits be quarantined pending additional labeling and o respond through their portal: https://recalls.medline.com. Once the response is received, Medline will provide additional labeling to be applied to the affected kits which states the following: "Attention: R-24-219, This kit contains a recalled item, STRYKER SafeAir Smoke Evacuation Pencil, VPN: 0703-047-000, 0703-047-001, 0703-047-002, 0703-046-000, 0703-046-001, 0703-046-002. DO NOT USE THIS COMPONENT. The rest of the kit may be used. Please remove this item only and replace with product from supply. *Do not cover or remove this sticker*". Those further distributing the product were directed to notify their customers.
Quantity in Commerce252162 units
DistributionUS Nationwide distribution.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
-
-