| Class 1 Device Recall JMC5A Ni Oxygen Concentrator |  |
Date Initiated by Firm | December 09, 2024 |
Date Posted | January 03, 2025 |
Recall Status1 |
Open3, Classified |
Recall Number | Z-0795-2025 |
Recall Event ID |
95818 |
510(K)Number | K230969 |
Product Classification |
Generator, oxygen, portable - Product Code CAW
|
Product | JMC5A Ni/TruAire-5 Oxygen Concentrator, Model: O2C5L |
Code Information |
UDI-DI: 00092237623152.
Serial Number Ranges: JA2311000001-JA2311000740,
JA2311000741-JA2311001480,
JA2311001481-JA2311002220,
JA2311002221-JA2311002960,
JA2311002961-JA2311003700,
JA2312000001-JA2312000740,
JA2312000741-JA2312001480,
JA2312001481-JA2312002080,
JA2401000001-JA2401000740. |
Recalling Firm/ Manufacturer |
JIANGSU JUMAO X-CARE MEDICAL EQUIPMENT CO LTD No 36 Danyan Road, Lianxing Village Danyang China
|
Manufacturer Reason for Recall | Oxygen concentrator may melt or catch on fire during use. |
FDA Determined Cause 2 | Component change control |
Action | On 11/26/2024, a recall notice was sent a distributor asking them to take the steps below.
On 12/09/2024, recall notices were sent from the distributor to customers who were asked to do the following:
This notice needs to be passed on to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred.
All units from the specified serial number range MUST be immediately discontinued from
use and removed from service and promptly returned.
Complete and return the acknowledgement and receipt form via email to recall@compasshealthbrands.com
If you have any questions, please contact:
1-800-376-7263 x444, Monday-Friday 8:00 am EST - 5:00 pm EST, or email recall@compasshealthbrands.com
On 12/18/2024, additional recall notices were sent adding additional affected devices. |
Quantity in Commerce | 6,520 |
Distribution | US: OH |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
510(K) Database | 510(K)s with Product Code = CAW
|
|
|
|