| Date Initiated by Firm | November 18, 2024 |
| Date Posted | January 07, 2025 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0815-2025 |
| Recall Event ID |
95855 |
| 510(K)Number | K172702 |
| Product Classification |
Vaporizer, anesthesia, non-heated - Product Code CAD
|
| Product | Tec 820 ISO Anesthesia Vaporizer, Reference Number 1177-9820-ISO |
| Code Information |
GTIN: 00840682124744;
Serial Numbers:
SQAD01037
SQAD01038
SQAD01041
SQAC01464
SQAC01467
SQAC00988
SQAC00987
SQAD01184
SQAD01219
SQAD01221
SQAD01222
SQAC01106
SQAD00048
SQAD00057
SQAD00059
SQAD00116
SQAD00124
SQAD00127
SQAD00129
SQAD00134
SQAD00136
SQAD00137
SQAD00139
SQAD00144
SQAD00576
SQAD00577
SQAD00581
SQAC00844
SQAC00851
SQAC00852
SQAC00854
SQAC00855
SQAC00856
SQAC00857
SQAC00858
SQAC00930
SQAC00842
SQAC00895
SQAC00517
SQAC01192
SQAD00860
SQAC01203
SQAC00989
SQAC00990
SQAC01186
SQAD00142
SQAD00486
SQAD00498
SQAD00831
SQAD00853
SQAD00855
SQAD00856
SQAD00857
SQAD00862
SQAD00863
SQAD00871
SQAD00873
SQAD00881
SQAD00883
SQAD00887
SQAD01109
SQAD01117
SQAD01168
SQAD00704
SQAD00736
SQAD00720
SQAD00747
SQAD00754
SQAD00761
SQAD00784
SQAD01031
SQAD01033
SQAD00523
SQAD00527
SQAD00583
SQAC01468
SQAD00140
SQAD00143
SQAD00808
SQAC01367
SQAC01389
SQAC01397
SQAC01190
SQAD00870
SQAD00859
SQAC01424
SQAC00881
SQAC01457
SQAC01476
SQAC01494
SQAC01445
SQAC01373
SQAC01435
SQAC01447
SQAC01361
SQAD00202
SQAC00941
SQAD00438
SQAD00439
SQAD00441
SQAC00811
SQAC01451
SQAD00024
SQAC00953
SQAD00200
SQAD00201
SQAD00199
SQAD00412
SQAD00413
SQAD00414
SQAD00416
SQAD00417
SQAD00419
SQAD00420
SQAD00421
SQAD00422
SQAD00423
SQAD00424
SQAD00425
SQAD00427
SQAD00428
SQAD00429
SQAD00430
SQAD00431
SQAD00433
SQAD00434
SQAD00435
SQAD00198
SQAC01507
SQAD00017
SQAD00031
SQAD00189
SQAD00220
SQAC01503
SQAC01508
SQAC01453
SQAC01461
SQAD00203
SQAC01220
SQAC01224
SQAC01215
SQAC01232
SQAC01233
SQAC01237
SQAC01458
SQAC01281
SQAC00814
SQAD00844
SQAC01056
SQAC00938
SQAD00008
SQAD00016
SQAD00437
SQAD00832
SQAC01217
SQAC01223
SQAC01228
SQAC00927
SQAC01142
SQAC01143
SQAC01239
SQAC01462
SQAC00875
SQAC01032
SQAC01033
SQAC01046
SQAC01050
SQAC01052
SQAC01054
SQAC01055
SQAC01058
SQAC01060
SQAC01061
SQAC01063
SQAC01064
SQAC01066
SQAC01070
SQAC01071
SQAC01072
SQAC01214
SQAC01226
SQAC01227
SQAC01229
SQAC01238
SQAC01245
SQAC01248
SQAC01249
SQAC01259
SQAC01276
SQAC01277
SQAC01279
SQAC01282
SQAC01285
SQAC01466
SQAC01500
SQAD00375
SQAD00377
SQAD00379
SQAD00380
SQAD00382
SQAD00383
SQAD00386
SQAD00387
SQAD00389
SQAC01470
SQAC01474
SQAD00032
SQAD00228
SQAD00234
SQAD00404
SQAD00426
SQAD00432
SQAD00436
SQAD00447
SQAD00450
SQAD00451
SQAD00452
SQAD00455
SQAD00456
SQAD00457
SQAD00458
SQAD00459
SQAD00460
SQAD00461
SQAD00462
SQAD00463
SQAD00464
SQAD00466
SQAD00467
SQAD00468
SQAD00469
SQAD00471
SQAD00740
SQAD00813
SQAD00828
SQAD00829
SQAD00830
SQAD00833
SQAD00834
SQAD00835
SQAD00836
SQAD00837
SQAD00838
SQAD00839
SQAD00840
SQAD00842
SQAD00843
SQAD00846
SQAD00847
SQAD00889
SQAD01016
SQAD01018
SQAD01030
SQAD01039
SQAD01043
SQAD01044
SQAD01045
SQAD01046
SQAD01047
SQAD01048
SQAD01051
SQAD01053
SQAD01054
SQAD01055
SQAD01060
SQAD01061
SQAD01062
SQAD01063
SQAD01064
SQAD01110
SQAD01111
SQAD01112
SQAD01113
SQAD01114
SQAD01115
SQAD01116
SQAD01118
SQAD01119
SQAD01121
SQAD01122
SQAD01123
SQAD01124
SQAD01125
SQAD01127
SQAD01130
SQAD01132
SQAD01134
SQAD01135
SQAD01136
SQAD01137
SQAD01138
SQAD01139
SQAD01141
SQAD01193
SQAD01197
SQAD01198
SQAD01199
SQAD01200
SQAD01203
SQAD01076
SQAD01080
SQAD01081
SQAD01082
SQAD01083
SQAD01089
SQAD01090
SQAD01091
SQAD01092
SQAD01093
SQAD01094
SQAD01095
SQAD01096
SQAD01097
SQAD01098
SQAD01099
SQAD01104
SQAD01106
SQAC01283
SQAC01049
SQAC01062
SQAC01073
SQAC01274
SQAC01234
SQAC01509
SQAD00002
SQAC00937
SQAC00940
SQAD00009
SQAC00939
SQAC01144
SQAD00376
SQAC00932
SQAD00010
SQAC00821
SQAC00822
SQAC00879
SQAC01497
SQAD00014
SQAD00021
SQAD00209
SQAD00210
SQAD00212
SQAD00207
SQAD00211
SQAC00929
SQAD00841
SQAD00018
SQAD00359
SQAC01504
SQAC00928
SQAD00374
SQAD00381
SQAD00384
SQAD00388
SQAD00390
SQAD00391
SQAD00405
SQAD00407
SQAD00408
SQAD00410
SQAD00411
SQAD00023
SQAC00758
SQAC00764
SQAD00011
SQAC01284
SQAD00443
SQAC01138
SQAC01278
SQAC00926
SQAD00197
SQAD00217
SQAC00933
SQAC00935
SQAC00946
SQAC00959
SQAC00960
SQAC01133
SQAD00445
SQAD00448
SQAD00465
SQAC01247
SQAC01246
SQAD00191
SQAD00446
SQAC01134
SQAC01449
SQAC01450
SQAD00208
SQAD00013
SQAD00845
SQAD01079
SQAD01102
SQAC01455
SQAC01456
SQAD00444
SQAD00449
SQAD00453
SQAD00454
SQAD00215
SQAC01140
SQAC01141
SQAC01341
SQAC01358
SQAC01359
SQAC01360
SQAD00204
SQAC01338
SQAD00205
SQAC01505
SQAC01506
SQAC01459
SQAC01501
SQAC00931
SQAD00218
SQAD00219
SQAD00226
SQAD00015
SQAC00936
SQAC00942
SQAC00945
SQAC00954
SQAC00962
SQAC00934
SQAD00440
SQAC00943
SQAC01448
SQAD00213
SQAD00216
SQAC00951
SQAC00819
SQAC01502
SQAD00221
SQAC01446
SQAC01490
SQAD00223
SQAC00947
SQAC00961
SQAD00442
SQAD00206
SQAC01469
SQAC00513
SQAC00641
SQAC00769
SQAC00859
SQAC00862
SQAC00884
SQAC00888
SQAC00889
SQAD00263
SQAD00001
SQAD00003
SQAD00022
SQAC00509
SQAC00511
SQAC00529
SQAC00530
SQAC00644
SQAC00770
SQAC00771
SQAC00778
SQAC01015
SQAC01043
SQAC01112
SQAC01115
SQAC01148
SQAC01178
SQAC01188
SQAC01191
SQAC01210
SQAC01288
SQAC01334
SQAC01356
SQAC01377
SQAC01499
SQAD00237
SQAD00240
SQAD00246
SQAD00247
SQAD00478
SQAD00479
SQAD00485
SQAD00706
SQAD00713
SQAD01078
SQAD01084
SQAD00481
SQAD00558
SQAD00582
SQAD00592
SQAD00594
SQAD00597
SQAD00932
SQAD01023
SQAD01058
SQAD01066
SQAD01142
SQAD01175
SQAD01223
SQAC00672
SQAC00674
SQAC00675
SQAC00676
SQAC00678
SQAC00684
SQAC00691
SQAC00696
SQAC00741
SQAC00765
SQAC00843
SQAC00847
SQAC00848
SQAC00849
SQAC00850
SQAC00993
SQAC00994
SQAC00995
SQAC00997
SQAC00998
SQAC01075
SQAC01076
SQAC01078
SQAC01079
SQAC01080
SQAC01089
SQAC01102
SQAC01103
SQAC01104
SQAC01105
SQAC01107
SQAC01108
SQAC01110
SQAC01117
SQAC01118
SQAC01119
SQAC01290
SQAC01291
SQAC01292
SQAC01293
SQAC01295
SQAC01300
SQAC01301
SQAC01302
SQAC01303
SQAC01304
SQAC01306
SQAC01307
SQAC01308
SQAC01309
SQAC01342
SQAC01345
SQAC01352
SQAC01362
SQAC01368
SQAC01369
SQAD00406
SQAD00569
SQAD00573
SQAD00575
SQAD00579
SQAD00925
SQAD00945
SQAD00950
SQAD00958
SQAD00966
SQAD01014
SQAD01015
SQAD01028
SQAD01029
SQAD01034
SQAD01056
SQAD01057
SQAD01059
SQAD01071
SQAD01074
SQAC00603
SQAC00642
SQAC00654
SQAC00655
SQAC00726
SQAC00739
SQAC00820
SQAC01165
SQAC01253
SQAD00762
SQAD01052
SQAD00475
SQAD00476
SQAD00477
SQAD00474
SQAD00487
SQAD00488
SQAD00492
SQAD00493
SQAD00742
SQAC01157
SQAC01158
SQAC01161
SQAC01173
SQAD00886
SQAC01318
SQAC01320
SQAC01346
SQAC01349
SQAC01404
SQAC01422
SQAC00812
SQAC00984
SQAC00985
SQAC00991
SQAC00992
SQAC00996
SQAC01395
SQAC01400
SQAC01406
SQAC01412
SQAC00633
SQAC00768
SQAC00774
SQAC00780
SQAD01183
SQAC00880
SQAC00882
SQAD00358
SQAD00852
SQAD00854
SQAD00858
SQAD00884
SQAD01213
SQAD00885
SQAD00607
SQAD00608
SQAD00609
SQAD00615
SQAD00682
SQAC00979
SQAD00331
SQAD00334
SQAD00565
SQAD00996
SQAD00999
SQAD01036
SQAD01040
SQAD01042
|
| FEI Number |
2126677
|
Recalling Firm/ Manufacturer |
GE Medical Systems, LLC
3000 N Grandview Blvd
Waukesha
WI
53188-1615
|
Manufacturer Reason for Recall | Due to a manufacturing issue, certain Tec 6 Plus and Tec 800 series vaporizers can deliver less anesthetic agent than the vaporizer dial setting. |
FDA Determined Cause 2 | Process control |
| Action | An URGENT MEDICAL DEVICE CORRECTION notification letter dated 11/18/24 was sent to customers.
Actions to be taken by Customer/User
You can continue to use the vaporizer in accordance with the instructions in the User Reference Manual (URM) and the actions described below:
1. Perform the Output Concentration Check described in Appendix 1 for every impacted vaporizer (see Affected Product Details in this letter).
" If a vaporizer passes the Output Concentration Check, you can continue to use that vaporizer. GE HealthCare recommends use of anesthetic agent monitoring during vaporizer use.
" If a vaporizer fails the Output Concentration Check, you can continue to use that vaporizer with anesthetic agent monitoring until GE HealthCare corrects the vaporizer.
" If a vaporizer fails the Output Concentration Check and anesthetic agent monitoring is not available, remove the vaporizer from service.
2. Please complete and return the attached Medical Device Notification Acknowledgement Response Form to ARC.FMI34139@gehealthcare.com
3. Please ensure all potential staff in your facility are made aware of this safety notification and the recommended actions.
4. Please retain this document for your records.
If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative. |
| Quantity in Commerce | 644 units |
| Distribution | US Nationwide. Global Distribution. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = CAD
|