• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Tec 850 SEV

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Tec 850 SEVsee related information
Date Initiated by FirmNovember 18, 2024
Date PostedJanuary 07, 2025
Recall Status1 Open3, Classified
Recall NumberZ-0818-2025
Recall Event ID 95855
510(K)NumberK172702 
Product Classification Vaporizer, anesthesia, non-heated - Product Code CAD
ProductTec 850 SEV Anesthesia Vaporizer, Reference Number 1177-9850-SEV
Code Information GTIN: 00840682124737; Serial Numbers: SQBC03462 SQBD02949 SQBD00338 SQBC03366 SQBC03815 SQBC05612 SQBC05620 SQBC05627 SQBC05635 SQBC05637 SQBC05641 SQBC05642 SQBC03558 SQBD03089 SQBD04439 SQBD04457 SQBD04459 SQBD04468 SQBD04702 SQBC04948 SQBC05702 SQBD02556 SQBC03896 SQBC04666 SQBC03225 SQBD02978 SQBD01384 SQBC02115 SQBC02023 SQBD04606 SQBD01484 SQBD04811 SQBC03549 SQBC03554 SQBC03559 SQBC03888 SQBC03970 SQBC04507 SQBD00211 SQBC02666 SQBC02959 SQBD00220 SQBD00402 SQBC01921 SQBC02087 SQBC02330 SQBC03003 SQBC01843 SQBC01855 SQBC04513 SQBD00326 SQBC05544 SQBC01450 SQBC02011 SQBC04529 SQBD01491 SQBD04711 SQBC01581 SQBC01747 SQBC02108 SQBC02162 SQBC02165 SQBC02183 SQBC02595 SQBC03023 SQBC05440 SQBD00444 SQBD00581 SQBC02204 SQBC02837 SQBD02999 SQBC01275 SQBC01279 SQBC01280 SQBC01283 SQBC01286 SQBC03723 SQBC02951 SQBD03058 SQBC03476 SQBD01380 SQBC03932 SQBC03969 SQBD04515 SQBD01481 SQBD01489 SQBD01315 SQBC01923 SQBD04013 SQBD04032 SQBD04034 SQBD04292 SQBD04307 SQBC02190 SQBC01912 SQBC05459 SQBD02862 SQBD02869 SQBD02886 SQBC02729 SQBC02747 SQBC01321 SQBC01322 SQBC01333 SQBD00159 SQBD02160 SQBC04115 SQBD02228 SQBD02276 SQBC02688 SQBC02753 SQBC02902 SQBD04016 SQBC01702 SQBC01811 SQBC01857 SQBC01864 SQBC01867 SQBC01869 SQBC01871 SQBC01872 SQBC01873 SQBC03257 SQBD04646 SQBD04058 SQBD04065 SQBC02014 SQBC02020 SQBD00130 SQBD04615 SQBC05293 SQBD00470 SQBD00474 SQBD00522 SQBC05037 SQBC02025 SQBD02635 SQBD02736 SQBD02738 SQBD02751 SQBD02754 SQBC04123 SQBD01505 SQBD01510 SQBD04630 SQBC05782 SQBC02814 SQBD00160 SQBD00161 SQBD00172 SQBC03066 SQBD04342 SQBD04052 SQBD04123 SQBD04242 SQBD04250 SQBD04287 SQBD04304 SQBD04325 SQBD04344 SQBD04346 SQBD04271 SQBD03156 SQBD02176 SQBC05369 SQBD03652 SQBD01267 SQBC04784 SQBC05013 SQBC05427 SQBC03055 SQBC03059 SQBC03090 SQBC03117 SQBC03119 SQBC03124 SQBC03127 SQBC03128 SQBC03129 SQBC03424 SQBC03515 SQBC03555 SQBC03560 SQBC03567 SQBC03576 SQBC05768 SQBC05773 SQBC05775 SQBC05776 SQBC05777 SQBD00239 SQBD04355 SQBD05023 SQBD04317 SQBD02837 SQBD02838 SQBD01312 SQBD01314 SQBD02337 SQBD02402 SQBD00692 SQBD00695 SQBD00698 SQBD00700 SQBD00705 SQBC05456 SQBC03004 SQBD04562 SQBD04565 SQBD04566 SQBD04567 SQBD04571 SQBD04573 SQBD00519 SQBD05014 SQBC02742 SQBC05578 SQBC05579 SQBC05580 SQBC05582 SQBC01276 SQBD02836 SQBD04588 SQBD03900 SQBD03932 SQBC01438 SQBC01441 SQBC01717 SQBC01973 SQBC03755 SQBC03968 SQBC03971 SQBD04462 SQBD03008 SQBD03011 SQBC02706 SQBC01988 SQBC03778 SQBC03784 SQBC01543 SQBC01579 SQBC01580 SQBC01587 SQBC02901 SQBC02015 SQBC04196 SQBC01287 SQBD04335 SQBC02363 SQBC02179 SQBC01434 SQBD01374 SQBC01951 SQBC02979 SQBD00593 SQBD00598 SQBD00601 SQBD00602 SQBD00604 SQBD00607 SQBC05383 SQBD03629 SQBD03631 SQBC03189 SQBC03226 SQBC04009 SQBD01371 SQBD01372 SQBD03580 SQBD03838 SQBC03675 SQBC03691 SQBD05016 SQBD02350 SQBD02162 SQBD00712 SQBC04000 SQBC04012 SQBC04024 SQBC04176 SQBC04177 SQBC04178 SQBC04187 SQBC04193 SQBC05463 SQBD01340 SQBC04546 SQBC04971 SQBC01898 SQBC03255 SQBC05634 SQBC02266 SQBD01492 SQBC02772 SQBC02785 SQBC02787 SQBC02788 SQBC02800 SQBC02817 SQBC02017 SQBD04227 SQBD04336 SQBD03195 SQBD04735 SQBC04887 SQBC04923 SQBC04931 SQBC04939 SQBC04941 SQBC02803 SQBC04454 SQBD03205 SQBD00610 SQBD00613 SQBD00614 SQBD00615 SQBD00616 SQBD00618 SQBD00619 SQBD00620 SQBD00621 SQBD00623 SQBD00624 SQBC04534 SQBC04553 SQBD02676 SQBD02872 SQBD02940 SQBD02971 SQBD02973 SQBC01995 SQBC04514 SQBD00486 SQBD00502 SQBC02002 SQBD00658 SQBD00660 SQBD00663 SQBD03066 SQBD03309 SQBD03485 SQBD03491 SQBD03536 SQBD03630 SQBD03633 SQBD03639 SQBD03906 SQBD04121 SQBD04157 SQBD04212 SQBD04583 SQBD04674 SQBD00126 SQBD00134 SQBD00139 SQBD00153 SQBD00154 SQBD00162 SQBD02282 SQBC04462 SQBC05632 SQBD03907 SQBD00441 SQBD01541 SQBD01556 SQBD01558 SQBD01584 SQBD01585 SQBD01586 SQBD01587 SQBD01588 SQBD01589 SQBD01590 SQBD01592 SQBC02239 SQBC02354 SQBC02364 SQBC02371 SQBC02373 SQBC02672 SQBC02674 SQBD02107 SQBD00377 SQBD00622 SQBD00627 SQBD00629 SQBD00632 SQBD01323 SQBC01037 SQBC01049 SQBC01332 SQBC01340 SQBC01345 SQBD04328 SQBC05319 SQBC05371 SQBC05372 SQBC01252 SQBC01261 SQBD00494 SQBD00495 SQBD00496 SQBD00498 SQBD00499 SQBD00501 SQBD01313 SQBD01322 SQBD03651 SQBD03803 SQBC03788 SQBC04878 SQBC03184 SQBD02268 SQBD02270 SQBC01994 SQBC01997 SQBC02163 SQBC02171 SQBC02173 SQBC02184 SQBC02185 SQBD00260 SQBC02022 SQBC02815 SQBC02816 SQBC02818 SQBC01695 SQBC02360 SQBD04904 SQBC04569 SQBD02543 SQBD02588 SQBD04694 SQBD04696 SQBC02101 SQBC02664 SQBC02669 SQBC01710 SQBC03009 SQBD01547 SQBD00456 SQBD01539 SQBD03061 SQBC01996 SQBC03444 SQBD02983 SQBD03000 SQBD01432 SQBD01456 SQBC02712 SQBC03781 SQBC02084 SQBC02315 SQBC02930 SQBC03094 SQBC01985 SQBC04041 SQBC04042 SQBC04091 SQBC04092 SQBD03482 SQBD03483 SQBD03493 SQBD03500 SQBD03649 SQBC01919 SQBC01950 SQBC03381 SQBD03927 SQBD03933 SQBD03937 SQBD03939 SQBD03943 SQBD03945 SQBD03948 SQBD03949 SQBD03950 SQBD03952 SQBC02824 SQBD01383 SQBD01485 SQBD01487 SQBD01490 SQBC01398 SQBD01294 SQBD01297 SQBD01309 SQBD02156 SQBD02159 SQBD02163 SQBD02164 SQBD03445 SQBD04322 SQBC02026 SQBC01808 SQBD00710 SQBC02966 SQBD03976 SQBD01562 SQBD01686 SQBD01702 SQBD01725 SQBD01732 SQBD01734 SQBD05010 SQBD05011 SQBD05017 SQBD05019 SQBD02317 SQBC04552 SQBC04556 SQBD02645 SQBD02667 SQBD02690 SQBC01779 SQBC01949 SQBC01970 SQBC03406 SQBC01936 SQBC03075 SQBC03678 SQBC03816 SQBC03872 SQBD03641 SQBC03258 SQBC05528 SQBC04218 SQBC04228 SQBC04254 SQBC04257 virtual1061324 virtual2061324 SQBC03901 SQBC04096 SQBC04353 SQBC04377 SQBC04500 SQBD00664 SQBD03046 SQBD03047 SQBD03048 SQBD03049 SQBD03050 SQBD03051 SQBD03062 SQBD03064 SQBD03072 SQBD00588 SQBD02553 SQBC02763 SQBC01562 SQBC02019 SQBC02180 SQBC02182 SQBC02186 SQBC01312 SQBC01323 SQBC01326 SQBC01327 SQBC01342 SQBC03557 SQBC03934 SQBC04159 SQBC04167 SQBC04171 SQBC02732 SQBC02758 SQBC02760 SQBD02985 SQBC04154 SQBC04174 SQBC04175 SQBD00131 SQBD00133 SQBD00138 SQBD00140 SQBC02259 SQBD00230 SQBD00037 SQBD00038 SQBD03880 SQBD03562 SQBD02158 SQBD03538 SQBD04698 SQBC03764 SQBC04516 SQBD01555 SQBD01718 SQBD01723 SQBD01724 SQBD01726 SQBD01727 SQBD01728 SQBD01729 SQBD01730 SQBD01731 SQBD01736 SQBD01737 SQBD01738 SQBD01739 SQBD01746 SQBD01747 SQBD01749 SQBD01750 SQBD04514 SQBD04521 SQBD04525 SQBD04526 SQBD04529 SQBC05614 SQBC05623 SQBD00662 SQBD02382 SQBD03240 SQBD03426 SQBD03455 SQBD03518 SQBD03646 SQBD03892 SQBD04155 SQBD01138 SQBD02285 SQBC04555 SQBC05550 SQBD00026 SQBD00041 SQBD02769 SQBD02548 SQBD02589 SQBD02591 SQBD02595 SQBD02412 SQBD02871 SQBC02773 SQBC02795 SQBC01291 SQBC03242 SQBD00672 SQBC03254 SQBC04969 SQBC04972 SQBD01079 SQBC01583 SQBC05394 SQBC04497 SQBC05783 SQBC02103 SQBC02119 SQBC03264 SQBC03259 SQBC03261 SQBC03278 SQBC03283 SQBD00223 SQBD00454 SQBC05668 SQBD01467 SQBD01710 SQBD01712 SQBD01717 SQBD01761 SQBD01777 SQBD01798 SQBD02816 SQBC02120 SQBC01547 SQBD00155 SQBC05745 SQBD04943 SQBD01285 SQBC03245 SQBC03247 SQBC03271 SQBC03272 SQBC03273 SQBC03274 SQBC03556 SQBC03573 SQBC03574 SQBC05484 SQBC01199 SQBC03874 SQBC03917 SQBD02708 SQBC03536 SQBC04779 SQBD01488 SQBC03548 SQBC03562 SQBC01948 SQBC01198 SQBC01195 SQBD04542 SQBC05097 SQBC01188 SQBC03082 SQBC05707 SQBC05005 SQBC05450 SQBC05491 SQBC05703 SQBD00090 SQBD02288 SQBC03569 SQBD04647 SQBC03211 SQBC01722 SQBC01723 SQBC01851 SQBC01853 SQBC02312 SQBC02374 SQBC02585 SQBC02586 SQBC02691 SQBC02879 SQBC02880 SQBC03582 SQBD01714 SQBD01740 SQBD01744 SQBD01748 SQBC05457 SQBC05460 SQBC05461 SQBC05462 SQBC05464 SQBC05465 SQBC05466 SQBC05467 SQBC05470 SQBD01379 SQBD01417 SQBC03925 SQBC04774 SQBD02397 SQBD02406 SQBD02540 SQBD02593 SQBD02646 SQBC01316 SQBC01325 SQBC01632 SQBC01800 SQBC01803 SQBC02210 SQBC01194 SQBC01197 SQBD03865 SQBC02809 SQBC02822 SQBD04465 SQBD04469 SQBC02644 SQBC03081 SQBC04104 SQBC01444 SQBD02627 SQBC05445 SQBD04275 SQBD04297 SQBD04303 SQBC01061 SQBC01095 SQBC01629 SQBC05609 SQBC05611 SQBC05613 SQBC05615 SQBC05617 SQBC05618 SQBC05621 SQBC05622 SQBC05624 SQBD01229 SQBD02103 SQBD02544 SQBD03290 SQBD03352 SQBD03384 SQBD03422 SQBD03451 SQBD03466 SQBD03469 SQBD03514 SQBD03558 SQBD03561 SQBD03589 SQBD03674 SQBD03682 SQBD03743 SQBD03744 SQBD03750 SQBD03753 SQBD03758 SQBD03759 SQBD03763 SQBD03767 SQBD03768 SQBD03774 SQBD03775 SQBD03787 SQBD03792 SQBD03793 SQBD03794 SQBD03795 SQBD03796 SQBD03799 SQBD03809 SQBD03819 SQBD03820 SQBD03822 SQBD04120 SQBD04129 SQBD04142 SQBD04144 SQBD04194 SQBD04213 SQBD04215 SQBD04216 SQBD04221 SQBD04229 SQBD04230 SQBD04231 SQBD04232 SQBD04233 SQBD04241 SQBD04243 SQBD04244 SQBD04245 SQBD04247 SQBD04248 SQBD04643 SQBD04712 SQBD04720 SQBD04722 SQBD04723 SQBD04724 SQBD04726 SQBD04792 SQBD04795 SQBD04799 SQBD04820 SQBD04821 SQBD04822 SQBD04823 SQBD04826 SQBD04832 SQBD04836 SQBD04837 SQBD04838 SQBD04841 SQBC01622 SQBC01654 SQBC01660 SQBC02719 SQBC02730 SQBC02734 SQBC02735 SQBC02736 SQBC02746 SQBC02748 SQBD03038 SQBD03386 SQBD03388 SQBD03389 SQBD03472 SQBC02263 SQBD00638 SQBD04672 SQBC01796 SQBC01850 SQBC01852 SQBC02969 SQBC04730 SQBC04895 SQBC05417 SQBC05455 SQBC02244 SQBC02021 SQBD02832 SQBC02584 SQBC01439 SQBC01446 SQBC01457 SQBC01633 SQBC02379 SQBC02869 SQBC04293 SQBC04311 SQBC04819 SQBC04822 SQBD00212 SQBD00218 SQBD00556 SQBD01704 SQBD01884 SQBD01885 SQBD01918 SQBD01926 SQBD01929 SQBD01930 SQBD01932 SQBD01933 SQBD01934 SQBD01935 SQBD01936 SQBD01937 SQBD01938 SQBD01939 SQBD01940 SQBD01942 SQBD01943 SQBD03235 SQBD03556 SQBD03574 SQBC02358 SQBC02582 SQBC05329 SQBD00206 SQBD00213 SQBC01648 SQBC01658 SQBC01662 SQBD01320 SQBD01336 SQBD01338 SQBD01349 SQBD01367 SQBD01369 SQBD01370 SQBC05258 SQBC05368 SQBC02206 SQBC02988 SQBC03008 SQBC03026 SQBC03029 SQBC03034 SQBC01360 SQBD00221 SQBD00227 SQBD00229 SQBC04106 SQBD00737 SQBD00741 SQBD00742 SQBD00743 SQBC01135 SQBC01608 SQBC01635 SQBC01644 SQBC03005 SQBC03037 SQBC03232 SQBC03277 SQBC03279 SQBC03284 SQBC03286 SQBC03293 SQBC03295 SQBC03298 SQBC03299 SQBC03300 SQBC03304 SQBC03305 SQBC03309 SQBD03057 SQBD03060 SQBD03094 SQBD03211 SQBD03236 SQBD03299 SQBD03343 SQBD03369 SQBD03450 SQBD03457 SQBD03463 SQBD03464 SQBD03481 SQBD03484 SQBD03494 SQBD03519 SQBD03573 SQBD03576 SQBD03577 SQBD03578 SQBC01468 SQBC02996 SQBC03041 SQBC04390 SQBD02917 SQBD02930 SQBD02953 SQBD04503 SQBC02009 SQBC02739 SQBC02752 SQBC02791 SQBC02856 SQBD02473 SQBD02912 SQBD02951 SQBD02956 SQBD02960 SQBD03130 SQBD03171 SQBD03655 SQBD03676 SQBD03685 SQBD03780 SQBD04130 SQBD04334 SQBD04348 SQBD04350 SQBD04357 SQBD04363 SQBD04365 SQBD04366 SQBD04369 SQBD04371 SQBD04372 SQBD04373 SQBD04375 SQBD04376 SQBD04377 SQBD04378 SQBD04380 SQBD04381 SQBD04383 SQBD04386 SQBD04388 SQBD04389 SQBD04390 SQBD04391 SQBD04392 SQBD04393 SQBD04395 SQBD04399 SQBD04401 SQBD04403 SQBD04407 SQBD04408 SQBD04410 SQBD04411 SQBD04412 SQBD04413 SQBD04416 SQBD04418 SQBD04420 SQBD04421 SQBD04423 SQBD04424 SQBD04425 SQBD04426 SQBC04476 SQBD03065 SQBD03842 SQBD03430 SQBC04318 SQBC04321 SQBC04322 SQBC04325 SQBC04330 SQBC04383 SQBC04435 SQBC04525 SQBC04716 SQBC04722 SQBC04745 SQBC04747 SQBC04750 SQBC04752 SQBC04785 SQBC04787 SQBC04797 SQBC04801 SQBC04802 SQBC04804 SQBD03477 SQBC02823 SQBD01517 SQBC03250 SQBC03297 SQBC03814 SQBD03866 SQBC05360 SQBC03632 SQBC03728 SQBC03736 SQBC03738 SQBC03961 SQBC04256 SQBC04267 SQBC04268 SQBC04284 SQBC04290 SQBC04291 SQBC04327 SQBC04391 SQBC04398 SQBC04617 SQBC02894 SQBC03223 SQBD02988 SQBD03006 SQBD03007 SQBC01148 SQBC01149 SQBC02686 SQBC02741 SQBC02786 SQBC02789 SQBC02792 SQBC02857 SQBC02858 SQBC03048 SQBC05420 SQBC05422 SQBC05425 SQBC05735 SQBD01425 SQBD01428 SQBD01528 SQBD03546 SQBD04531 SQBC00623 SQBC00657 SQBC01400 SQBC01413 SQBC01414 SQBC01416 SQBC01418 SQBC01419 SQBC01421 SQBC01422 SQBC01423 SQBC01428 SQBC01430 SQBC01431 SQBC01435 SQBC01436 SQBC01437 SQBC04409 SQBC05784 SQBC05787 SQBC05788 SQBD01548 SQBD01975 SQBD01979 SQBD01985 SQBD01987 SQBD02117 SQBD02120 SQBD02121 SQBD02127 SQBD02129 SQBD02130 SQBD02132 SQBD02134 SQBD02139 SQBD02142 SQBD02870 SQBD02874 SQBD03302 SQBD03365 SQBD03376 SQBD03613 SQBD03626 SQBD03844 SQBD03868 SQBD03877 SQBD03895 SQBD03897 SQBD03908 SQBD03910 SQBD03911 SQBD03917 SQBD04014 SQBD04026 SQBD04062 SQBD04066 SQBD04191 SQBD04196 SQBD04197 SQBD04319 SQBD04324 SQBD04569 SQBD04570 SQBD04574 SQBD04576 SQBD04578 SQBD04625 SQBD04686 SQBD04687 SQBD04775 SQBD04791 SQBD04928 SQBD04940 SQBD04941 SQBD04942 SQBD04945 SQBD04961 SQBD04968 SQBD04969 SQBC03868 SQBC04109 SQBC04110 SQBC04758 SQBC04759 SQBD00101 SQBD00107 SQBD00291 SQBD02466 SQBD02518 SQBD02554 SQBD00451 SQBD00483 SQBD02133 SQBD02141 SQBD04604 SQBC05604 SQBD02669 SQBD02773 SQBD02805 SQBD02812 SQBD02813 SQBD02819 SQBD02822 SQBD02825 SQBD02834 SQBD02845 SQBD02846 SQBD02848 SQBD04298 SQBC04874 SQBC04881 SQBD02534 SQBD02537 SQBB05615 SQBC01224 SQBC03804 SQBC03811 SQBC03813 SQBC04015 SQBC04031 SQBC04032 SQBC04247 SQBC04273 SQBC04274 SQBC04277 SQBC04278 SQBC01768 SQBC01769 SQBC01770 SQBC01771 SQBC01772 SQBC01773 SQBC01774 SQBC01775 SQBC01776 SQBC01777 SQBC01780 SQBC01781 SQBC01783 SQBC01784 SQBC01785 SQBC01788 SQBC01789 SQBC01790 SQBC01792 SQBC01793 SQBC01794 SQBC01795 SQBC01797 SQBC01798 SQBC03627 SQBC03720 SQBC03863 SQBC03871 SQBC03924 SQBC03954 SQBC03955 SQBC03965 SQBC03967 SQBC03973 SQBC03976 SQBC03977 SQBC03978 SQBC03979 SQBC03980 SQBC03981 SQBC03982 SQBC03983 SQBC03988 SQBC03989 SQBC03990 SQBC03991 SQBB05655 SQBC02352 SQBC02591 SQBC03542 SQBC04129 SQBC04370 SQBC04378 SQBC04403 SQBC04437 SQBC04439 SQBC04440 SQBC04442 SQBC04447 SQBC04465 SQBC04494 SQBC04528 SQBC04568 SQBC04582 SQBC04591 SQBC04805 SQBC04815 SQBC04816 SQBC04817 SQBC04860 SQBC04873 SQBC04875 SQBC04882 SQBC05334 SQBC05403 SQBC05411 SQBC05442 SQBC05458 SQBC05483 SQBC05562 SQBC05571 SQBC05574 SQBC05576 SQBC05581 SQBC05583 SQBC05585 SQBC05588 SQBC05590 SQBC05591 SQBC05600 SQBC05601 SQBC05602 SQBC05605 SQBC05606 SQBC05607 SQBC05625 SQBC05630 SQBC05631 SQBC05633 SQBC05638 SQBC05644 SQBC05646 SQBC05648 SQBC05651 SQBC05652 SQBC05654 SQBC05657 SQBC05658 SQBC05659 SQBC05660 SQBC05661 SQBC05663 SQBC05664 SQBC05665 SQBC05666 SQBC05667 SQBC05669 SQBC05671 SQBC05673 SQBC05674 SQBC05675 SQBC05676 SQBC05677 SQBC05678 SQBC05679 SQBC05680 SQBC05681 SQBC05683 SQBC05684 SQBC05685 SQBC05686 SQBC05687 SQBC05689 SQBC05692 SQBC05694 SQBC05695 SQBC05696 SQBC05697 SQBC05698 SQBC05700 SQBC05701 SQBD01327 SQBD01328 SQBD01342 SQBD01344 SQBD01346 SQBD01347 SQBD01358 SQBD01359 SQBD01360 SQBD01364 SQBD01875 SQBD01992 SQBD02037 SQBD02039 SQBD02040 SQBD02047 SQBD02049 SQBD02050 SQBD02052 SQBD02055 SQBD02059 SQBD02081 SQBD02082 SQBD02100 SQBD02102 SQBD02104 SQBD02106 SQBD02108 SQBD02116 SQBD02122 SQBD02124 SQBD02128 SQBD02131 SQBD02135 SQBD02137 SQBD02140 SQBD02519 SQBD02546 SQBD02547 SQBD02558 SQBD02580 SQBD02581 SQBD02582 SQBD02583 SQBD02586 SQBD02587 SQBD02590 SQBD03334 SQBD03392 SQBD03421 SQBD03428 SQBD03438 SQBD03439 SQBD03446 SQBD03447 SQBD03454 SQBD03474 SQBD03476 SQBD03533 SQBD03554 SQBD03559 SQBD03560 SQBD03567 SQBD03568 SQBD03569 SQBD03570 SQBD03571 SQBD03572 SQBD03581 SQBD03582 SQBD03592 SQBD03935 SQBD03956 SQBD03992 SQBD03995 SQBD03996 SQBD02457 SQBD02647 SQBD02700 SQBD02783 SQBD02851 SQBD02898 SQBD02947 SQBD02964 SQBD02965 SQBD02966 SQBD02967 SQBD02968 SQBD02969 SQBD02970 SQBD02972 SQBD01348 SQBD03600 SQBD03665 SQBD03841 SQBC01573 SQBC01582 SQBC02609 SQBC02627 SQBC03218 SQBC04760 SQBD00666 SQBD00670 SQBD00671 SQBD02640 SQBD02979 SQBD01849 SQBD02258 SQBD03012 SQBC03248 SQBC03301 SQBC03303 SQBC03359 SQBC03360 SQBC03361 SQBC03365 SQBC03369 SQBC03371 SQBC03372 SQBC03379 SQBC03380 SQBC03402 SQBC03414 SQBC03428 SQBC03430 SQBC03431 SQBC03432 SQBD02657 SQBD02716 SQBD02722 SQBD02725 SQBD02726 SQBD02727 SQBD02728 SQBD02730 SQBD02731 SQBD02732 SQBD02733 SQBD02734 SQBD02735 SQBD02737 SQBD02739 SQBD02740 SQBD02758 SQBD02770 SQBD02776 SQBD02778 SQBD02781 SQBD02785 SQBD02787 SQBD02791 SQBD02794 SQBC01402 SQBC02229 SQBC05710 SQBC01560 SQBC03579 SQBC03588 SQBD01250 SQBD04511 SQBC01611 SQBD03739 SQBD02982 SQBD04005 SQBD02741 SQBC03102 SQBD01319 SQBD01321 SQBD01362 SQBD01363 SQBD01366 SQBD01375 SQBD01361 SQBD04075 SQBD04193 SQBD04203 SQBC01526 SQBC01541 SQBC01578 SQBC01791 SQBC02790 SQBC02897 SQBC02899 SQBC04181 SQBC04183 SQBC04191 SQBC04825 SQBC04879 SQBC04880 SQBD00028 SQBD00029 SQBD00036 SQBD01660 SQBD01671 SQBD01676 SQBD01691 SQBD01174 SQBD01133 SQBD00447 SQBD00418 SQBD00421 SQBD00471 SQBD00475 SQBD01692 SQBD00867 SQBD01443 SQBD01644 SQBD01678 SQBD01685 SQBD00443 SQBC00970 SQBC01384 SQBC01451 SQBC01452 SQBC01458 SQBC01460 SQBC01481 SQBC01486 SQBC01488 SQBC01500 SQBC01501 SQBC01874 SQBC02168 SQBC02284 SQBC02842 SQBC02852 SQBC02853 SQBC02855 SQBC02859 SQBC02866 SQBC02886 SQBC02889 SQBC02898 SQBC02945 SQBC02946 SQBC02955 SQBC02962 SQBC02967 SQBC02976 SQBC03080 SQBC03337 SQBC03374 SQBC03441 SQBC03443 SQBC03446 SQBC03447 SQBC03455 SQBC03456 SQBC03524 SQBC03610 SQBC03635 SQBC03642 SQBC03646 SQBC03740 SQBC03744 SQBC03746 SQBC03749 SQBC03754 SQBC03759 SQBC03760 SQBC03762 SQBC03768 SQBC03770 SQBC03772 SQBC03779 SQBC03780 SQBC03786 SQBC03929 SQBC03937 SQBC03958 SQBC03962 SQBC03964 SQBC03993 SQBC03994 SQBC04226 SQBC04250 SQBC04265 SQBC04270 SQBC04355 SQBC04369 SQBC04389 SQBC04393 SQBC04405 SQBC04436 SQBC04438 SQBC04446 SQBC04479 SQBC04496 SQBC04521 SQBC04541 SQBC04586 SQBC04590 SQBC04624 SQBC04625 SQBC04628 SQBC04636 SQBC04639 SQBC04660 SQBC04661 SQBC04664 SQBC04764 SQBC04768 SQBC04811 SQBC04890 SQBC04894 SQBC04904 SQBC04907 SQBC04916 SQBC04917 SQBC04926 SQBC05062 SQBC05081 SQBC05107 SQBC05118 SQBC05237 SQBC05238 SQBC05241 SQBC05245 SQBC05253 SQBC05257 SQBC05261 SQBC05263 SQBC05264 SQBC05266 SQBC05267 SQBC05269 SQBC05273 SQBC01461 SQBC01489 SQBC01503 SQBC01512 SQBC02270 SQBC02280 SQBC02285 SQBC02296 SQBC02331 SQBC02366 SQBC02375 SQBC02821 SQBC02861 SQBC02862 SQBC02863 SQBC02896 SQBC02903 SQBC02919 SQBC02939 SQBC02953 SQBC03285 SQBC03308 SQBC03319 SQBC03326 SQBC03330 SQBC03339 SQBC03347 SQBC03351 SQBC03353 SQBC03589 SQBC03743 SQBC03945 SQBC03972 SQBC04269 SQBC04566 SQBC04646 SQBD00753 SQBD00773 SQBD00804 SQBD01192 SQBD01193 SQBD01234 SQBD01243 SQBD03551 SQBD04070 SQBD04097 SQBD02333 SQBD00438 SQBD00345 SQBD00367 SQBD00370 SQBD00414 SQBD00415 SQBD01217 SQBD01664 SQBD01673 SQBD01694 SQBD01757 SQBD01760 SQBD01763 SQBD01764 SQBD01768 SQBD01771 SQBD01772 SQBD01775 SQBD01778 SQBD01790 SQBD01795 SQBD00339 SQBD00340 SQBD00449 SQBD00480 SQBD01153 SQBD01167 SQBD01173 SQBD01183 SQBD01189 SQBD01569 SQBD01769 SQBD01773 SQBD01791 SQBD01793 SQBD01794 SQBD03080 SQBD03101 SQBD02113 SQBD03664 SQBD03701 SQBD03709 SQBD03797 SQBD03801 SQBD03864 SQBD01600 SQBD02379 SQBD02391 SQBD02184 SQBD02197 SQBD01440 SQBD01675 SQBC01380 SQBC01385 SQBC01699 SQBC01704 SQBC01708 SQBC03092 SQBC03115 SQBC03125 SQBC03540 SQBC03545 SQBC03609 SQBC03697 SQBC03933 SQBC03995 SQBC04654 SQBC04919 SQBD00481 SQBD00487 SQBD00537 SQBD00543 SQBD00551 SQBD00561 SQBD00580 SQBD00594 SQBD00600 SQBD00733 SQBD01024 SQBD01066 SQBD01067 SQBD01068 SQBD01073 SQBD01074 SQBD01096 SQBD01098 SQBD01099 SQBD01100 SQBD01104 SQBD01109 SQBD01161 SQBD01221 SQBD01225 SQBD01241 SQBD01251 SQBD01255 SQBD01256 SQBD01265 SQBD01271 SQBD01272 SQBD01594 SQBD01683 SQBD01832 SQBD01842 SQBD02280 SQBD02306 SQBD02338 SQBD02366 SQBD02368 SQBD02372 SQBD02374 SQBD02377 SQBD02398 SQBD02400 SQBD02407 SQBD02410 SQBD02411 SQBD02415 SQBD02420 SQBD04059 SQBD04099 SQBD04485 SQBD00545 SQBD00546 SQBD00797 SQBD01091 SQBC01643 SQBC01727 SQBC01231 SQBD00583 SQBD00779 SQBD00894 SQBD01479 SQBD01654 SQBD02361 SQBD02918 SQBD03063 SQBD03105 SQBD03109 SQBD03112 SQBD03114 SQBD03118 SQBD03123 SQBD03136 SQBD03425 SQBD03456 SQBD03499 SQBD03609 SQBD03690 SQBD03694 SQBD03713 SQBD03764 SQBD03804 SQBD03815 SQBD03852 SQBD03863 SQBD03871 SQBD03929 SQBD03930 SQBD03938 SQBD03947 SQBD04109 SQBD04111 SQBD04114 SQBD04125 SQBD04128 SQBD04200 SQBD04283 SQBD04356 SQBD04385 SQBD04422 SQBD04428 SQBD04450 SQBD04467 SQBD04470 SQBD04471 SQBD04473 SQBD04474 SQBD04479 SQBD04480 SQBD04482 SQBD04483 SQBD04484 SQBD04486 SQBD04505 SQBD04506 SQBD04513 SQBD04522 SQBD04523 SQBD04532 SQBD04535 SQBD04541 SQBD04545 SQBD04551 SQBD04553 SQBD04560 SQBD04568 SQBD04582 SQBD04584 SQBD04585 SQBD04593 SQBD04594 SQBD04595 SQBD04598 SQBD04599 SQBD04600 SQBD04602 SQBD04607 SQBD04609 SQBD04611 SQBD04612 SQBD04616 SQBD04622 SQBD04632 SQBD04634 SQBD04635 SQBD04636 SQBD04658 SQBD04683 SQBD04703 SQBD04718 SQBD04790 SQBD04798 SQBD04885 SQBD04890 SQBD04891 SQBD04900 SQBD04918 SQBD04962 SQBD04973 SQBD04974 SQBD04975 SQBD04976 SQBD04980 SQBD04981 SQBD04982 SQBD04984 SQBD04987 SQBD04988 SQBD04990 SQBD04992 SQBD05009 SQBD05020 SQBD05024 SQBD05027 SQBD05028 SQBD05029 SQBD05030 SQBD05031 SQBD05032 SQBD05033 SQBD05034 SQBD05035 SQBD05036 SQBD05039 SQBD05045 SQBD05053 SQBD05054 SQBD05055 SQBD05056 SQBD05060 SQBD05064 SQBD05065 SQBD05072 SQBD05073 SQBD05074 SQBD05075 SQBD05076 SQBD05077 SQBD05078 SQBD05081 SQBD05083 SQBD05084 SQBD05085 SQBD05087 SQBD05092 SQBD05093 SQBD05096 SQBD05099 SQBD05101 SQBD05102 SQBD05103 SQBD05104 SQBD05105 SQBD05106 SQBD05107 SQBD05111 SQBD05118 SQBD05121 SQBD05133 SQBD05136 SQBD05138 SQBD05141 SQBD05145 SQBD05146 SQBD05148 SQBD05151 SQBD00337 SQBD00355 SQBD00388 SQBD00393 SQBD00395 SQBD00396 SQBD00423 SQBD00424 SQBD03104 SQBD03318 SQBD03675 SQBD03704 SQBD03781 SQBD03788 SQBD04054 SQBD04061 SQBD04063 SQBD04069 SQBD04079 SQBD04085 SQBD04089 SQBD04091 SQBD04098 SQBD01449 SQBD01523 SQBD01545 SQBD01570 SQBD01574 SQBD01575 SQBD01576 SQBD01577 SQBD01578 SQBD01579 SQBD01580 SQBD01581 SQBD01583 SQBD01785 SQBD01797 SQBD01801 SQBD01806 SQBD01809 SQBD01810 SQBD01815 SQBD01825 SQBD00547 SQBC01478 SQBD01070 SQBD00775 SQBD00863 SQBD03342 SQBD03354 SQBD03359 SQBD03362 SQBD03379 SQBD04071 SQBD04074 SQBD04077 SQBD04078 SQBD04087 SQBD00346 SQBD00436 SQBD01206 SQBD01184 SQBD01203 SQBD01171 SQBD01180 SQBD01181 SQBD01199 SQBD01216 SQBD01223 SQBD01433 SQBD01639 SQBD01681 SQBD01693 SQBD00358 SQBD00390 SQBD00401 SQBD00432 SQBD00440 SQBD00461 SQBD00464 SQBD00465 SQBD00515 SQBD00579 SQBD01059 SQBD01116 SQBD01122 SQBD01124 SQBD01452 SQBD01641 SQBD01699 SQBD01708 SQBD00343 SQBD01438 SQBD01261 SQBD02069 SQBD02118 SQBD01069 SQBD01196 SQBD01198 SQBD01202 SQBD01207 SQBC05488 SQBD00330 SQBD00419 SQBD01812 SQBD00912 SQBD01112 SQBD02068 SQBD02928 SQBD02939 SQBD02945 SQBD03254 SQBD03304 SQBD03317 SQBD03322 SQBD03323 SQBD03329 SQBD03330 SQBD03335 SQBD03337 SQBD03347 SQBD03356 SQBD03711 SQBD03745 SQBD03783 SQBD03784 SQBD03791 SQBD03800 SQBD03860 SQBD03872 SQBD03873 SQBD04288 SQBD00429 SQBC02140 SQBC02201 SQBC02220 SQBC02225 SQBC02242 SQBC02275 SQBC02277 SQBC02307 SQBC02313 SQBC02314 SQBC02320 SQBC02327 SQBC02340 SQBC02341 SQBC02344 SQBC02948 SQBC03287 SQBC03318 SQBC03332 SQBC03345 SQBC03352 SQBC04677 SQBC04872 SQBC05122 SQBC05143 SQBC05153 SQBC05155 SQBC05161 SQBC05163 SQBD01194 SQBD01201 SQBD02625 SQBD02897 SQBD02911 SQBD02950 SQBC05724 SQBD00410 SQBD00428 SQBD00476 SQBD01210 SQBD02375 SQBD02413 SQBD02449 SQBD00335 SQBD01065 SQBD00450 SQBD00509 SQBD00605 SQBD00833 SQBD00864 SQBD00878 SQBD00886 SQBD03052 SQBD03079 SQBD03102 SQBD03119 SQBD03127 SQBD03131 SQBD00760 SQBD00767 SQBD00875 SQBD00879 SQBD01431 SQBD01666 SQBD00482 SQBD02206 SQBC01106 SQBC01157 SQBC01164 SQBC01243 SQBC01271 SQBC01341 SQBC01479 SQBC01585 SQBC01705 SQBC01846 SQBC01876 SQBC01889 SQBC01901 SQBC01906 SQBC01909 SQBC01914 SQBC01943 SQBC02117 SQBC02188 SQBC02193 SQBC02200 SQBC02228 SQBC02230 SQBC02231 SQBC02305 SQBC02309 SQBC02324 SQBC02361 SQBC02671 SQBC02675 SQBC02677 SQBC02687 SQBC02694 SQBC02696 SQBC02697 SQBC02699 SQBC02700 SQBC02720 SQBC02728 SQBC02738 SQBC02740 SQBC02777 SQBC02876 SQBC02890 SQBC02916 SQBC02949 SQBC02973 SQBC03020 SQBC03022 SQBC03027 SQBC03061 SQBC03086 SQBC03105 SQBC03126 SQBC03136 SQBC03163 SQBC03168 SQBC03260 SQBC03262 SQBC03312 SQBC03315 SQBC03346 SQBC03449 SQBC03452 SQBC03477 SQBC03480 SQBC03492 SQBC03497 SQBC03501 SQBC03506 SQBC03507 SQBC03520 SQBC03525 SQBC03564 SQBC03604 SQBC03605 SQBC03607 SQBC03613 SQBC03624 SQBC03634 SQBC03656 SQBC03671 SQBC03684 SQBC03700 SQBC03702 SQBC03706 SQBC03707 SQBC03709 SQBC03714 SQBC03716 SQBC03722 SQBC03870 SQBC03927 SQBC04121 SQBC04124 SQBC04127 SQBC04140 SQBC04148 SQBC04266 SQBC04307 SQBC04358 SQBC04359 SQBC04366 SQBC04455 SQBC04515 SQBC04619 SQBC04621 SQBC04627 SQBC04632 SQBC04647 SQBC04662 SQBC04762 SQBC04766 SQBC04789 SQBC05075 SQBC05079 SQBC05082 SQBC05088 SQBC05119 SQBC05125 SQBC05148 SQBC05149 SQBC05150 SQBC05152 SQBC05162 SQBC05165 SQBC05168 SQBC05169 SQBC05170 SQBC05172 SQBC05173 SQBC05174 SQBC05175 SQBC05178 SQBC05179 SQBC05180 SQBC05181 SQBC05182 SQBC05186 SQBC05189 SQBC05217 SQBC05218 SQBC05219 SQBC05223 SQBC05224 SQBC05226 SQBC05230 SQBC05246 SQBC05270 SQBC01496 SQBC01497 SQBC02136 SQBC02154 SQBC02159 SQBC02167 SQBC02169 SQBC02195 SQBC02197 SQBC02199 SQBC02215 SQBC02223 SQBC02224 SQBC02227 SQBC02240 SQBC03578 SQBC03630 SQBC03641 SQBC03758 SQBC03777 SQBC03789 SQBC04080 SQBC04192 SQBC04594 SQBC04641 SQBC04643 SQBC04648 SQBC05066 SQBC05069 SQBC05072 SQBC05073 SQBC05084 SQBC05089 SQBC05110 SQBC05115 SQBC00924 SQBC01159 SQBC01235 SQBC01268 SQBC01270 SQBC01375 SQBC01376 SQBC01379 SQBC01462 SQBC01465 SQBC01480 SQBC01485 SQBC01487 SQBC01490 SQBC01510 SQBC01519 SQBC01528 SQBC02261 SQBC02264 SQBC02282 SQBC02290 SQBC02292 SQBC02295 SQBC02297 SQBC02298 SQBC02300 SQBC02301 SQBC02302 SQBC02311 SQBC02316 SQBC02317 SQBC02321 SQBC02329 SQBC02332 SQBC02333 SQBC02342 SQBC02346 SQBC02349 SQBC02351 SQBC02356 SQBC02369 SQBC02622 SQBC02623 SQBC02698 SQBC02733 SQBC02806 SQBC02832 SQBC02836 SQBC02845 SQBC02865 SQBC02895 SQBC02905 SQBC02907 SQBC02910 SQBC02912 SQBC02915 SQBC02944 SQBC02970 SQBC02974 SQBC03113 SQBC03122 SQBC03123 SQBC03145 SQBC03325 SQBC03350 SQBC03387 SQBC03434 SQBC03453 SQBC04633 SQBC04634 SQBC04653 SQBC04678 SQBC04684 SQBC04689 SQBC04690 SQBC04692 SQBC05053 SQBC05141 SQBC05146 SQBC05156 SQBC05159 SQBC05160 SQBC05164 SQBC05171 SQBC02347 SQBC02175 SQBC02194 SQBC02202 SQBC02221 SQBC02237 SQBC02238 SQBC00659 SQBC00804 SQBC00980 SQBC01162 SQBC01320 SQBC01329 SQBC01383 SQBC01448 SQBC01471 SQBC01502 SQBC01509 SQBC01696 SQBC01766 SQBC01860 SQBC02079 SQBC02250 SQBC02276 SQBC02350 SQBC02367 SQBC02368 SQBC02835 SQBC02887 SQBC02900 SQBC02906 SQBC02911 SQBC02917 SQBC03116 SQBC03322 SQBC03335 SQBC03344 SQBC03348 SQBC03517 SQBC03550 SQBC03626 SQBC03668 SQBC03761 SQBC03765 SQBC03766 SQBC03767 SQBC03776 SQBC03975 SQBC04593 SQBC04596 SQBC04631 SQBC04637 SQBC04640 SQBC04644 SQBC04651 SQBC04656 SQBC04674 SQBC04676 SQBC04679 SQBC04680 SQBC04685 SQBC04693 SQBC01021 SQBC01239 SQBC01483 SQBC01493 SQBC01551 SQBC01576 SQBC01577 SQBC01599 SQBC01607 SQBC01609 SQBC02095 SQBC02226 SQBC02245 SQBC02256 SQBC02322 SQBC02682 SQBC02683 SQBC02685 SQBC02693 SQBC02717 SQBC02726 SQBC03100 SQBC03144 SQBC03148 SQBC03208 SQBC03252 SQBC03314 SQBC03338 SQBC03438 SQBC03490 SQBC03491 SQBC03494 SQBC03505 SQBC03514 SQBC03522 SQBC03532 SQBC03537 SQBC03737 SQBC03931 SQBC04396 SQBC04630 SQBC04669 SQBC05260 SQBC05262 SQBC02278 SQBC02279 SQBC02281 SQBC02310 SQBC02336 SQBC02337 SQBC03166 SQBC03585 SQBC03708 SQBC01344 SQBC01350 SQBC02157 SQBC02236 SQBC02908 SQBC02954 SQBC03079 SQBC03512 SQBC03570 SQBC03571 SQBC03591 SQBC03597 SQBC03637 SQBC03639 SQBC03670 SQBC03696 SQBC03698 SQBC03704 SQBC04122 SQBC04638 SQBC04667 SQBC04754 SQBC01649 SQBC02630 SQBC02925 SQBC02960 SQBC02971 SQBC02972 SQBC02975 SQBC02977 SQBC04208 SQBC04241 SQBC04242 SQBC04243 SQBC04297 SQBC05136 SQBC05157 SQBC05166 SQBC05167 SQBC05177 SQBC05183 SQBC05184 SQBC05185 SQBC05215 SQBC05216 SQBC05220 SQBC05239 SQBC05240 SQBC05244 SQBC05247 SQBC05248 SQBC05254 SQBC05255 SQBC05256 SQBC05259 SQBC05265 SQBC05268 SQBC05272 SQBC05278 SQBC03900 SQBC04248 SQBC04249 SQBC04294 SQBC01030 SQBC01137 SQBC01449 SQBC01455 SQBC01463 SQBC01474 SQBC01498 SQBC02203 SQBC02253 SQBC02283 SQBC02288 SQBC02306 SQBC02319 SQBC02323 SQBC02338 SQBC02667 SQBC02676 SQBC02681 SQBC02684 SQBC02695 SQBC02702 SQBC02716 SQBC02724 SQBC03114 SQBC03130 SQBC03137 SQBC03141 SQBC03142 SQBC03435 SQBC03450 SQBC03504 SQBC03553 SQBC03894 SQBC05067 SQBC05080 SQBC05087 SQBC05140 SQBC05144 SQBC05147 SQBC05154 SQBC01491 SQBC01567 SQBC01569 SQBC01591 SQBC01593 SQBC01612 SQBC01614 SQBC01885 SQBC01892 SQBC01895 SQBC01897 SQBC01910 SQBC01916 SQBC02166 SQBC02241 SQBC03109 SQBC03133 SQBC03503 SQBC03509 SQBC03534 SQBC03543 SQBC03575 SQBC03577 SQBC03608 SQBC03703 SQBC03713 SQBC03717 SQBC00508 SQBC01105 SQBC01115 SQBC01172 SQBC01244 SQBC01269 SQBC01381 SQBC01382 SQBC01387 SQBC01470 SQBC01476 SQBC01482 SQBC01494 SQBC01515 SQBC01880 SQBC02867 SQBC03107 SQBC03118 SQBC03120 SQBC03121 SQBC03131 SQBC03132 SQBC03138 SQBC03150 SQBC03169 SQBC03311 SQBC03495 SQBC03530 SQBC03544 SQBC03614 SQBC05076 SQBC04410 SQBC04929 SQBC01120 SQBC02326 SQBC02328 SQBC02335 SQBC02339 SQBC02348 SQBC02701 SQBC02820 SQBC03106 SQBC03134 SQBC03153 SQBC03541 SQBC03919 SQBC04551 SQBC04567 SQBC04724 SQBC04755 SQBC04803 SQBC04888 SQBC04901 SQBC04903 SQBC04905 SQBC04914 SQBC04918 SQBC04928 SQBC01117 SQBC01330 SQBC01378 SQBC01388 SQBC01466 SQBC01469 SQBC01472 SQBC01524 SQBC01531 SQBC01539 SQBC01540 SQBC01544 SQBC01545 SQBC01546 SQBC01549 SQBC01552 SQBC01555 SQBC01556 SQBC01558 SQBC01584 SQBC02164 SQBC02174 SQBC02208 SQBC02303 SQBC02703 SQBC02704 SQBC02705 SQBC02708 SQBC02710 SQBC02711 SQBC02854 SQBC03165 SQBC03206 SQBC03498 SQBC03592 SQBC03611 SQBC03662 SQBC04295 SQBC04301 SQBC04310 SQBC04361 SQBC04362 SQBC04363 SQBC04368 SQBC04374 SQBC04375 SQBC04395 SQBC04401 SQBC04491 SQBC03205 SQBC03207 SQBC03421 SQBC03496 SQBC03623 SQBC02714 SQBC02723 SQBC03091 SQBC03521 SQBC03593 SQBC03602 SQBC01101 SQBC01108 SQBC01109 SQBC01112 SQBC01114 SQBC01136 SQBC01328 SQBC01339 SQBC01347 SQBC01351 SQBC01353 SQBC01354 SQBC02833 SQBC04357 SQBC04673 SQBC04763 SQBC04770 SQBC01004 SQBC01121 SQBC01475 SQBC01707 SQBC01714 SQBC01865 SQBC01922 SQBC01924 SQBC01945 SQBC01947 SQBC02036 SQBC02039 SQBC02044 SQBC02074 SQBC02085 SQBC02096 SQBC02097 SQBC02099 SQBC02114 SQBC02211 SQBC02299 SQBC02304 SQBC02308 SQBC02318 SQBC02325 SQBC02678 SQBC02689 SQBC02692 SQBC02721 SQBC02913 SQBC02914 SQBC02943 SQBC03313 SQBC03317 SQBC03321 SQBC03328 SQBC03331 SQBC03334 SQBC03341 SQBC03342 SQBC03343 SQBC03356 SQBC03502 SQBC03510 SQBC03523 SQBC03533 SQBC03552 SQBC03590 SQBC03622 SQBC03673 SQBC03701 SQBC03808 SQBC03873 SQBC03875 SQBC03877 SQBC03879 SQBC03891 SQBC04691 SQBC04702 SQBC01272 SQBC01467 SQBC01477 SQBC01484 SQBC02235 SQBC03516 SQBC01374 SQBC01453 SQBC01499 SQBC01508 SQBC03222 SQBC03336 SQBC03349 SQBC00456 SQBC01310 SQBC01447 SQBC01454 SQBC01473 SQBC01511 SQBC01565 SQBC01566 SQBC01568 SQBC01570 SQBC01574 SQBC01575 SQBC01586 SQBC01588 SQBC01589 SQBC01590 SQBC01592 SQBC01594 SQBC01595 SQBC01596 SQBC01597 SQBC01598 SQBC01600 SQBC01601 SQBC01602 SQBC01603 SQBC01604 SQBC01606 SQBC01610 SQBC01613 SQBC01615 SQBC01616 SQBC01617 SQBC01618 SQBC01619 SQBC01620 SQBC01621 SQBC01623 SQBC01624 SQBC01626 SQBC01837 SQBC01878 SQBC01879 SQBC01881 SQBC01882 SQBC01886 SQBC01894 SQBC01899 SQBC01903 SQBC01904 SQBC01908 SQBC01911 SQBC01917 SQBC01918 SQBC01925 SQBC01927 SQBC01930 SQBC01934 SQBC01937 SQBC02110 SQBC02246 SQBC02247 SQBC02249 SQBC02252 SQBC02254 SQBC02255 SQBC02257 SQBC02258 SQBC02260 SQBC02262 SQBC02670 SQBC02718 SQBC02725 SQBC02727 SQBC03164 SQBC03212 SQBC03214 SQBC03275 SQBC03412 SQBC03420 SQBC03500 SQBC03531 SQBC03806 SQBC03812 SQBC03928 SQBC03960 SQBC04133 SQBC04137 SQBC04138 SQBC04139 SQBC04194 SQBC04210 SQBC04263 SQBC04264 SQBC04296 SQBC04304 SQBC04309 SQBC04313 SQBC04317 SQBC04351 SQBC04364 SQBC04365 SQBC04367 SQBC04406 SQBC04408 SQBC04443 SQBC04629 SQBC04645 SQBC04658 SQBC04765 SQBC04772 SQBC05052 SQBC05068 SQBC05074 SQBC05077 SQBC05078 SQBC05086 SQBC05103 SQBC05105 SQBC05109 SQBC05117 SQBC05142 SQBC05145 SQBC01251 SQBC01866 SQBC01868 SQBC01875 SQBC01877 SQBC01884 SQBC01888 SQBC01890 SQBC01891 SQBC01893 SQBC01902 SQBC01905 SQBC01913 SQBC01915 SQBC01926 SQBC01928 SQBC01929 SQBC01931 SQBC01932 SQBC01939 SQBC03427 SQBC03440 SQBC03454 SQBC03459 SQBC03606 SQBC03692 SQBC02153 SQBC02161 SQBC02177 SQBC02216 SQBC02219 SQBC02222 SQBC02233 SQBC02243 SQBD02974 SQBD02987 SQBD03028 SQBD03033 SQBD01050 SQBD01058 SQBD00840 SQBD00884 SQBD00938 SQBD00939 SQBD01047 SQBD02098 SQBD02214 SQBD02217 SQBD02218 SQBD02223 SQBD02233 SQBD02240 SQBD02251 SQBD02254 SQBD02256 SQBD02257 SQBD02259 SQBD02261 SQBD01200 SQBD00599 SQBD01228 SQBC01653 SQBD00713 SQBD00877 SQBD00930 SQBD00942 SQBD01044 SQBD01045 SQBD01046 SQBD01048 SQBD01049 SQBD01063 SQBD01188 SQBD01227 SQBD01231 SQBD01242 SQBD01253 SQBD01260 SQBD01264 SQBD01270 SQBD01840 SQBD00541 SQBD04080 SQBD04086 SQBD04102 SQBD04103 SQBC02155 SQBC02234 SQBD00597 SQBD01086 SQBD01232 SQBD01238 SQBD00540 SQBD00560 SQBD00595 SQBD01072 SQBD01083 SQBD01087 SQBD01095 SQBD01101 SQBD01254 SQBD01259 SQBD01262 SQBD01268 SQBD01682 SQBD01899 SQBD01900 SQBD01230 SQBD01258 SQBD02057 SQBD02061 SQBD02071 SQBD02074 SQBD02083 SQBD02086 SQBD02087 SQBD02088 SQBD02091 SQBD02093 SQBD02094 SQBD02095 SQBD02096 SQBD02097 SQBD02099 SQBD02101 SQBD02109 SQBD02110 SQBD02111 SQBD02112 SQBD02115 SQBD02119 SQBD03962 SQBD04064 SQBD01190 SQBD02290 SQBD02339 SQBD02376 SQBD02386 SQBD02394 SQBD02437 SQBD01078 SQBD01084 SQBD04220 SQBC04259 SQBD00405 SQBD00880 SQBD01051 SQBD01060 SQBD01064 SQBD00319 SQBD00835 SQBD00872 SQBD00874 SQBD00883 SQBD00885 SQBD01028 SQBD01057 SQBD02044 SQBD02170 SQBD02213 SQBD02221 SQBD02222 SQBD02224 SQBD02229 SQBD02232 SQBD02241 SQBD02262 SQBD02275 SQBD02383 SQBD02389 SQBD03999 SQBD04285 SQBD00908 SQBD01213 SQBD01439 SQBD01766 SQBD01774 SQBD01792 SQBD02250 SQBD01495 SQBD01780 SQBD02203 SQBD00348 SQBD00349 SQBD00711 SQBD00755 SQBD00757 SQBD00758 SQBD00766 SQBD00772 SQBD00781 SQBD00510 SQBD00539 SQBD00730 SQBD00731 SQBD00732 SQBD00736 SQBD01434 SQBD01445 SQBD01451 SQBD01606 SQBD01642 SQBD01657 SQBD01661 SQBD01680 SQBD01684 SQBD01689 SQBD01697 SQBD01705 SQBD01713 SQBD01715 SQBD02180 SQBD02225 SQBD02318 SQBD02326 SQBD02699 SQBD02760 SQBD02890 SQBD02893 SQBD02896 SQBD02902 SQBD02906 SQBD02910 SQBD02914 SQBD02923 SQBD02944 SQBD02952 SQBD02954 SQBD02963 SQBD03393 SQBD04009 SQBD04057 SQBD04088 SQBD04094 SQBD04101 SQBD02349 SQBD02384 SQBD02385 SQBD02387 SQBD02390 SQBD02396 SQBD02405 SQBD00342 SQBD00383 SQBD00389 SQBD00404 SQBD00750 SQBD00770 SQBD00771 SQBD00780 SQBD00791 SQBD00794 SQBD00796 SQBD00798 SQBD00799 SQBD00803 SQBD00806 SQBD01634 SQBD01205 SQBD01687 SQBD01745 SQBD01850 SQBD01855 SQBD01877 SQBD01889 SQBD01902 SQBD01915 SQBD01922 SQBD01944 SQBD01960 SQBD01966 SQBD01455 SQBD00317 SQBD00384 SQBD00394 SQBD00403 SQBD01996 SQBD02000 SQBD02003 SQBD02005 SQBD00382 SQBD00385 SQBD00412 SQBD00434 SQBD01209 SQBD01304 SQBD01412 SQBD01593 SQBD01823 SQBD01831 SQBD01844 SQBD01969 SQBD02006 SQBD02008 SQBD01459 SQBD01701 SQBD01707 SQBD01052 SQBD01080 SQBD01092 SQBD01129 SQBD01152 SQBD01156 SQBD01236 SQBD01247 SQBD01437 SQBD01444 SQBC04997 SQBC05639 SQBC05662 SQBC05709 SQBC05725 SQBC05736 SQBC05737 SQBC05739 SQBC05741 SQBC05742 SQBC05743 SQBD00458 SQBD00459 SQBD00462 SQBD00468 SQBD00479 SQBD01155 SQBD01211 SQBD00718 SQBD00720 SQBD00723 SQBD00740 SQBD00746 SQBD00756 SQBD02340 SQBD02422 SQBD02433 SQBD02436 SQBD02439 SQBD02460 SQBD01226 SQBD02441 SQBD02359 SQBD02370 SQBD00350 SQBD00460 SQBD00542 SQBD00722 SQBD00845 SQBD00847 SQBD00865 SQBD00881 SQBD03108 SQBD03122 SQBD03328 SQBD00956 SQBD00303 SQBD00333 SQBD00334 SQBD00336 SQBD00386 SQBD00387 SQBD01643 SQBD01651 SQBD02381 SQBD02408 SQBD02414 SQBD01102 SQBD03442 SQBD03452 SQBD03091 SQBD03092 SQBD03100 SQBD03106 SQBD03111 SQBD03115 SQBD03117 SQBD03120 SQBD03121 SQBD03125 SQBD03126 SQBD03316 SQBD03331 SQBD03332 SQBD03333 SQBD03339 SQBD03340 SQBD03395 SQBD02895 SQBD00896 SQBD00928 SQBD01082 SQBD01088 SQBD01103 SQBD01105 SQBD01106 SQBD01108 SQBD01111 SQBD01113 SQBD01114 SQBD01115 SQBD01117 SQBD01118 SQBD01119 SQBD01123 SQBD01134 SQBD01197 SQBD01224 SQBD02322 SQBD02399 SQBD02401 SQBD02403 SQBD02409 SQBD02417 SQBD02404 SQBD02899 SQBD02948 SQBD01139 SQBD01142 SQBD01144 SQBD01146 SQBD01158 SQBD01017 SQBD02755 SQBD02903 SQBD00898 SQBD01055 SQBD01056 SQBD01249 SQBD01275 SQBD01829 SQBD01845 SQBD00351 SQBD00466 SQBD00397 SQBD01154 SQBD01093 SQBD01820 SQBD01821 SQBD01822 SQBD00832 SQBD01107 SQBD01120 SQBD01141 SQBD01147 SQBD01149 SQBD01159 SQBD01163 SQBD01164 SQBD01165 SQBD01175 SQBD01182 SQBD00400 SQBD00398 SQBD00416 SQBD02367 SQBD02369 SQBD00489 SQBD01450 SQBD01669 SQBD01690 SQBD00721 SQBD00725 SQBD02303 SQBD02486 SQBD02894 SQBD02961 SQBD02307 SQBD02909 SQBD03346 SQBD04106 SQBD04141 SQBD04150 SQBD04134 SQBD00517 SQBD00765 SQBD00768 SQBD00774 SQBD00792 SQBD00805 SQBD03095 SQBD03116 SQBD03140 SQBD03802 SQBD03890 SQBD02297 SQBD02336 SQBD01652 SQBD01786 SQBD02085 SQBD02189 SQBD02191 SQBD00442 SQBD00488 SQBD00552 SQBD00591 SQBD00353 SQBD00420 SQBD00431 SQBD00453 SQBD01214 SQBD01446 SQBD01706 SQBD01852 SQBD01857 SQBD01858 SQBD01859 SQBD01861 SQBD01862 SQBD01878 SQBD01880 SQBD01882 SQBD01898 SQBD01901 SQBD01903 SQBD01921 SQBD01948 SQBD01954 SQBD01955 SQBD01956 SQBD01957 SQBD01963 SQBD01965 SQBD01971 SQBD01978 SQBD01995 SQBD02320 SQBD02325 SQBD02327 SQBD02335 SQBD02393 SQBD02418 SQBD02419 SQBD02421 SQBD02424 SQBD02428 SQBD02429 SQBD02432 SQBD02434 SQBD02435 SQBD02442 SQBD02444 SQBD02445 SQBD02448 SQBD00426 SQBD00435 SQBD00553 SQBD00584 SQBD00585 SQBD00589 SQBD00626 SQBD01014 SQBD01026 SQBD01030 SQBD01031 SQBD01032 SQBD01033 SQBD01034 SQBD01036 SQBD01037 SQBD01039 SQBD01040 SQBD01042 SQBD01043 SQBD01187 SQBD01195 SQBD01892 SQBD01923 SQBD02092 SQBD02105 SQBD02114 SQBD02187 SQBD02192 SQBD02205 SQBD01235 SQBD03034 SQBD03336 SQBD03366 SQBD03473 SQBD03475 SQBD03492 SQBD03529 SQBD03604 SQBD03615 SQBD03661 SQBD03760 SQBD03770 SQBD03777 SQBD03790 SQBD03818 SQBD03891 SQBD03915 SQBD03954 SQBD03964 SQBD04081 SQBD04082 SQBD04093 SQBD04104 SQBD04107 SQBD04108 SQBD04110 SQBD04112 SQBD04113 SQBD04115 SQBD04116 SQBD04118 SQBD04119 SQBD04122 SQBD04124 SQBD04126 SQBD04127 SQBD04131 SQBD04132 SQBD04135 SQBD04138 SQBD04143 SQBD04145 SQBD04147 SQBD04148 SQBD04149 SQBD04151 SQBD04270 SQBD04280 SQBD04284 SQBD04293 SQBD04294 SQBD04402 SQBD04404 SQBD04417 SQBD04419 SQBD04436 SQBD04437 SQBD04442 SQBD04443 SQBD04444 SQBD04445 SQBD04447 SQBD04448 SQBD04451 SQBD04454 SQBD00409 SQBD00484 SQBD00538 SQBD00761 SQBD01257 SQBD01516 SQBD01568 SQBD00446 SQBD01827 SQBD01640 SQBD01649 SQBD01709 SQBD01716 SQBD01722 SQBD00531 SQBD00734 SQBD00759 SQBD01075 SQBD01076 SQBD01094 SQBD01160 SQBD01162 SQBD01166 SQBD01168 SQBD01169 SQBD01170 SQBD01172 SQBD01176 SQBD01177 SQBD01178 SQBD01191 SQBD02863 SQBD02920 SQBD02921 SQBD02922 SQBD02926 SQBD02931 SQBD02934 SQBD02935 SQBD02938 SQBD02959 SQBD00763 SQBD00890 SQBD00923 SQBD00926 SQBD00927 SQBD01053 SQBD01061 SQBD01062 SQBD01665 SQBD00467 SQBD00352 SQBD01741 SQBD01765 SQBD01782 SQBD03320 SQBD03341 SQBD03358 SQBD03394 SQBD01779 SQBD02363 SQBD02365 SQBD02371 SQBD02373 SQBD02380 SQBD02388 SQBD02231 SQBD02255 SQBD00448 SQBD00437 SQBD00473 SQBD01843 SQBD02252 SQBD00439 SQBD02362 SQBD02378 SQBD00427 SQBD01071 SQBD01077 SQBD01081 SQBD01085 SQBD01097 SQBD00776 SQBD01022 SQBD01121 SQBD01131 SQBD01132 SQBD01135 SQBD01137 SQBD01441 SQBD01442 SQBD01667 SQBD01668 SQBD01670 SQBD01672 SQBD01674 SQBD01677 SQBD01695 SQBD01696 SQBD01698 SQBD01700 SQBD02843 SQBD02905 SQBD02908 SQBD02924 SQBD02927 SQBD02932 SQBD02933 SQBD02937 SQBD02943 SQBD03321 SQBD03355 SQBD01035 SQBD01126 SQBD01128 SQBD01136 SQBD01140 SQBD01148 SQBD01150 SQBD00368 SQBD04090 SQBD04092 SQBD00323 SQBD00363 SQBD00378 SQBD00379 SQBD00422 SQBD00425 SQBD00452 SQBD00477 SQBD00681 SQBD00735 SQBD00747 SQBD00860 SQBD00866 SQBD01759 SQBD01837 SQBD01841 SQBD03778 SQBD03971 SQBD04004 SQBD04272 SQBD04278 SQBD04282 SQBD04290 SQBD04291 SQBD04295 SQBD02416 SQBD02431 SQBD01125 SQBD01127 SQBD01130 SQBD01143 SQBD01145 SQBD01151 SQBD01157 SQBD00455 SQBD00399 SQBD00411 SQBD01186 SQBD01252 SQBD01233 SQBD02062 SQBD02084 SQBD02181 SQBD02185 SQBD02186 SQBD02190 SQBD02196 SQBD02199 SQBD02207 SQBD03989 SQBD04267 SQBD04274 SQBD04277 SQBD04279 SQBD04281 SQBD01789 SQBD01835 SQBD01839 SQBD01851 SQBD01854 SQBD01856 SQBD01865 SQBD01897 SQBD01911 SQBD01968 SQBD00371 SQBD01219 SQBD02219 SQBD02260 SQBD02919 SQBD02056 SQBC05399 SQBC05400 SQBC05404 SQBC05426 SQBC05429 SQBD01215 SQBD01471 SQBD01497 SQBD01615 SQBD01628 SQBD01645 SQBD01647 SQBD01648 SQBD01662 SQBD01663 SQBD01679 SQBD01688 SQBD01863 SQBD01864 SQBD01866 SQBD01888 SQBD01890 SQBD01896 SQBD01904 SQBD01905 SQBD01907 SQBD01908 SQBD01909 SQBD01910 SQBD01912 SQBD01913 SQBD01914 SQBD01916 SQBD01919 SQBD01920 SQBD01928 SQBD01946 SQBD01947 SQBD01949 SQBD01951 SQBD01958 SQBD01962 SQBD01964 SQBD01967 SQBD01970 SQBD01976 SQBD01982 SQBD01986 SQBD01988 SQBD01989 SQBD01990 SQBD01991 SQBD01993 SQBD01997 SQBD01998 SQBD01999 SQBD02002 SQBD02021 SQBD02033 SQBD02042 SQBD02046 SQBD02048 SQBD02051 SQBC00608 SQBC00787 SQBC01442 SQBC03021 SQBC03182 SQBC03216 SQBC03219 SQBC03224 SQBC03843 SQBC05436 SQBD00525 SQBC00493 SQBC01352 SQBC01663 SQBD00572 SQBD03919 SQBD03921 SQBC02936 SQBC03280 SQBD00565 SQBD00567 SQBD00570 SQBD00574 SQBD00576 SQBD00578 SQBD00745 SQBD00748 SQBD00749 SQBD00751 SQBD00752 SQBD00762 SQBD00788 SQBD00818 SQBD00828 SQBD00831 SQBD00834 SQBD00837 SQBD00848 SQBD00857 SQBD00858 SQBD00868 SQBD00869 SQBD00870 SQBD00871 SQBD00889 SQBD00903 SQBD00904 SQBD00905 SQBD00906 SQBD00907 SQBD00909 SQBD00911 SQBD00913 SQBD00914 SQBD00915 SQBD00916 SQBD00917 SQBD00919 SQBD00920 SQBD00921 SQBD00922 SQBD00924 SQBD00925 SQBD00929 SQBD00932 SQBD00933 SQBD01018 SQBD02301 SQBD02989 SQBC04973 SQBC04975 SQBC04976 SQBC04977 SQBC04981 SQBC04982 SQBC04989 SQBC04998 SQBC05015 SQBC05016 SQBC05017 SQBC05018 SQBC05019 SQBC05020 SQBC05021 SQBC05029 SQBC05031 SQBC05036 SQBC05038 SQBC05039 SQBC05042 SQBC05060 SQBC05061 SQBC05064 SQBC05065 SQBC02192 SQBC03167 SQBC03178 SQBC03185 SQBD00661 SQBD01337 SQBD02680 SQBD02683 SQBD02686 SQBD02688 SQBD02709 SQBD04756 SQBD04758 SQBD04760 SQBD03293 SQBD03294 SQBD03460 SQBD03715 SQBD03716 SQBD03719 SQBD03721 SQBD03723 SQBD03726 SQBD03727 SQBC02831 SQBC02860 SQBC02870 SQBC02874 SQBC02875 SQBC03920 SQBC04005 SQBC04006 SQBC04010 SQBC04014 SQBC04017 SQBC04065 SQBC04066 SQBC04067 SQBC04076 SQBC04081 SQBC04082 SQBC04087 SQBC04184 SQBC04186 SQBC04203 SQBC04204 SQBC04205 SQBC04206 SQBC04209 SQBC05326 SQBC05328 SQBC05331 SQBC05332 SQBC05333 SQBC05336 SQBC05337 SQBC05338 SQBC05340 SQBC05341 SQBD01368 SQBD01373 SQBD01376 SQBD01377 SQBD01378 SQBD02495 SQBD02496 SQBD02497 SQBD02498 SQBD02499 SQBD02500 SQBD02501 SQBD02502 SQBD02504 SQBD02505 SQBD02506 SQBD02507 SQBD02510 SQBD02511 SQBD02512 SQBD02513 SQBD02516 SQBD02517 SQBD02526 SQBD02527 SQBD02529 SQBD02607 SQBD02608 SQBD02609 SQBD02614 SQBD02615 SQBD02616 SQBD02617 SQBD02618 SQBD02619 SQBD02639 SQBD02641 SQBD02642 SQBD02644 SQBD02648 SQBD02650 SQBD02651 SQBD02652 SQBD02654 SQBD02655 SQBD02656 SQBD02659 SQBD02660 SQBD02663 SQBD02665 SQBD02666 SQBD02668 SQBD02670 SQBD02671 SQBD02672 SQBD02674 SQBD02675 SQBC01959 SQBC01967 SQBC01971 SQBC01984 SQBC01991 SQBC01992 SQBC01955 SQBC01993 SQBC01957 SQBC03666 SQBC04004 SQBC01631 SQBC01935 SQBC01941 SQBC01952 SQBC01954 SQBC01956 SQBC01958 SQBC01962 SQBC01972 SQBC01978 SQBC01979 SQBC01981 SQBC01982 SQBC01986 SQBC01987 SQBC03539 SQBC03856 SQBC03963 SQBC03987 SQBC03997 SQBC05111 SQBC05137 SQBC05233 SQBC05281 SQBC05563 SQBC05592 SQBC05628 SQBD00084 SQBD00100 SQBD00111 SQBD00125 SQBD00158 SQBD00164 SQBD00165 SQBD00166 SQBD00167 SQBD00168 SQBD00171 SQBD00174 SQBD00176 SQBD00177 SQBD00178 SQBD00179 SQBD00180 SQBD00181 SQBD00182 SQBD00183 SQBD00184 SQBD00185 SQBD00190 SQBD00192 SQBD00193 SQBD00195 SQBD00293 SQBD00294 SQBD00298 SQBD00299 SQBD00300 SQBD00301 SQBD00306 SQBD00314 SQBD00315 SQBD00318 SQBD00324 SQBD00325 SQBD00328 SQBD00329 SQBD00332 SQBD00344 SQBD00347 SQBD01507 SQBD01519 SQBD01591 SQBD01605 SQBD01613 SQBD01614 SQBD01616 SQBD01617 SQBD01618 SQBD01619 SQBD01621 SQBD01622 SQBD01624 SQBD01625 SQBD01627 SQBD01629 SQBD01630 SQBD01631 SQBD01632 SQBD01633 SQBD01635 SQBD01636 SQBD01646 SQBD01655 SQBD03812 SQBD01458 SQBC03998 SQBC03999 SQBC04001 SQBC05330 SQBC05691 SQBD01296 SQBD01302 SQBD01308 SQBD01311 SQBD01483 SQBD03585 SQBD03586 SQBD03588 SQBD03663 SQBD03673 SQBD03990 SQBD03993 SQBD03994 SQBD03997 SQBD04003 SQBC01731 SQBC01961 SQBC01968 SQBC02088 SQBC02090 SQBC02091 SQBC02093 SQBC02098 SQBC02102 SQBC02104 SQBC02105 SQBC02109 SQBC02111 SQBC02112 SQBC02113 SQBC02150 SQBC02156 SQBC02170 SQBC02172 SQBC03151 SQBC03152 SQBC03154 SQBC03156 SQBC03467 SQBC03527 SQBC03616 SQBC03619 SQBC03621 SQBC03628 SQBC03651 SQBC03652 SQBC03653 SQBC03654 SQBC03655 SQBC01933 SQBC01960 SQBC01966 SQBC01969 SQBC01964 SQBC01977 SQBC01965 SQBC03699 SQBC04195 SQBC04207 SQBC04211 SQBC04212 SQBC04216 SQBC04217 SQBC04220 SQBC04227 SQBC04229 SQBC04230 SQBC04233 SQBC01963 SQBC01980 SQBC01983 SQBC01989 SQBD04668 SQBC02123 SQBC03899 SQBC04668 SQBD00173 SQBD00175 SQBD04340 SQBD04345 SQBC01505 SQBC02757 SQBC02761 SQBC02756 SQBC03859 SQBC03864 SQBC01802 SQBD04743 SQBD04751 SQBD01240 SQBC04258 SQBC04260 SQBC04261 SQBC05418 SQBC05385 SQBC00465 SQBC01293 SQBC01299 SQBC01537 SQBC01548 SQBC01550 SQBC01564 SQBC01572 SQBC02016 SQBC03177 SQBC03229 SQBC03235 SQBC03288 SQBC03290 SQBC03291 SQBC03292 SQBC03296 SQBC03310 SQBD00295 SQBD00311 SQBD00312 SQBD00316 SQBD00320 SQBD00354 SQBD00469 SQBD00491 SQBD00518 SQBD00524 SQBD00527 SQBD00528 SQBC02981 SQBC02743 SQBC04161 SQBC01495 SQBC01504 SQBC01506 SQBC01507 SQBC01553 SQBC01557 SQBD00143 SQBD00309 SQBD00310 SQBD00321 SQBD00769 SQBD00789 SQBD01410 SQBD01419 SQBD01950 SQBD01953 SQBD01980 SQBD01981 SQBD03078 SQBD03548 SQBD03563 SQBD03738 SQBD03742 SQBD03746 SQBD03748 SQBD03754 SQBD03755 SQBD03757 SQBC03084 SQBC03382 SQBC03383 SQBC03384 SQBC03385 SQBC03386 SQBC03388 SQBC03390 SQBC03391 SQBC03395 SQBC03397 SQBC03398 SQBC03399 SQBC03400 SQBC03403 SQBC03404 SQBC03405 SQBC03408 SQBC03409 SQBC03415 SQBC03417 SQBD01527 SQBD01535 SQBD01537 SQBD01540 SQBC03046 SQBC04224 SQBC03083 SQBC01824 SQBC04221 SQBC05708 SQBD00199 SQBD02793 SQBD02859 SQBC01713 SQBC01799 SQBC02152 SQBC02768 SQBC03140 SQBC03149 SQBC03251 SQBC03256 SQBC03392 SQBC03411 SQBC03419 SQBC03436 SQBC03437 SQBC03457 SQBC03464 SQBC03465 SQBC03468 SQBC03469 SQBC03470 SQBC03471 SQBC03478 SQBC03483 SQBC03486 SQBC04222 SQBC04232 SQBC04234 SQBC04235 SQBC04300 SQBC04302 SQBC04305 SQBC04306 SQBC04312 SQBC04319 SQBC04320 SQBC04356 SQBC04481 SQBC04538 SQBC05292 SQBC05749 SQBC05754 SQBC05755 SQBC05760 SQBC05762 SQBD00507 SQBD00526 SQBD00530 SQBD00533 SQBD00534 SQBD00536 SQBD00550 SQBD00554 SQBD00555 SQBD00557 SQBD00563 SQBD00566 SQBD00571 SQBD00738 SQBD01179 SQBD01245 SQBD01246 SQBD01269 SQBD01274 SQBD01276 SQBD01277 SQBD01279 SQBD01280 SQBD01281 SQBD01282 SQBD01283 SQBD01287 SQBD01288 SQBD01289 SQBD01290 SQBD01291 SQBD01292 SQBD01298 SQBD01299 SQBD01300 SQBD01454 SQBD02027 SQBD02066 SQBD02076 SQBD02077 SQBD02078 SQBD02143 SQBD02145 SQBD02146 SQBD02148 SQBD02149 SQBD02152 SQBD02153 SQBD02154 SQBD02179 SQBD02193 SQBD02210 SQBD02227 SQBD02230 SQBD02236 SQBD02239 SQBD02800 SQBD02801 SQBD02991 SQBD03297 SQBD03440 SQBD03448 SQBD03471 SQBD03524 SQBD03594 SQBD03854 SQBD03981 SQBD03987 SQBD04165 SQBD04195 SQBD04198 SQBD04236 SQBD04254 SQBD04347 SQBD04354 SQBD04362 SQBD04431 SQBD04449 SQBD04492 SQBD04493 SQBD04497 SQBD04614 SQBD04619 SQBD04739 SQBD04740 SQBD04741 SQBD04757 SQBD04761 SQBD05003 SQBD05015 SQBD05022 SQBD05026 SQBD01504 SQBD01514 SQBD01520 SQBD01553 SQBC02010 SQBC03396 SQBC03547 SQBC03676 SQBC03694 SQBC01679 SQBC01697 SQBC01726 SQBC02588 SQBC02668 SQBC04111 SQBC04141 SQBC04237 SQBC04687 SQBC04700 SQBC04710 SQBC04829 SQBC04838 SQBD02216 SQBD02272 SQBD03360 SQBD03377 SQBD03462 SQBD03465 SQBD03486 SQBD03506 SQBD03507 SQBD03509 SQBD03521 SQBD03526 SQBD03668 SQBD03680 SQBD03691 SQBD03712 SQBD02283 SQBD03002 SQBD03659 SQBC05551 SQBC05554 SQBC05555 SQBC05556 SQBC05559 SQBC05561 SQBC05564 SQBC05565 SQBC05566 SQBC05572 SQBC05798 SQBC05803 SQBC05805 SQBC05810 SQBC05811 SQBD01204 SQBD01212 SQBD01305 SQBD01460 SQBD01461 SQBD01464 SQBD01466 SQBD01473 SQBD01525 SQBD01595 SQBD01656 SQBD01703 SQBD01742 SQBD01743 SQBD01751 SQBD01770 SQBD01776 SQBD01784 SQBD01800 SQBD01804 SQBD01811 SQBD01813 SQBD01814 SQBD01816 SQBD01817 SQBD01818 SQBD01819 SQBD01826 SQBD01828 SQBD01830 SQBC04151 SQBC04995 SQBC05307 SQBC05397 SQBC05398 SQBC05526 SQBD01508 SQBD01551 SQBD04341 SQBD04736 SQBD04755 SQBD04759 SQBD04977 SQBD05000 SQBD05018 SQBC04806 SQBD03468 SQBC02884 SQBD04533 SQBD04685 SQBD04991 SQBD04993 SQBD04994 SQBD04995 SQBD04996 SQBD04998 SQBD04999 SQBD05005 SQBD05007 SQBC02883 SQBD03699 SQBC02891 SQBC02892 SQBD04732 SQBD04794 SQBD04997 SQBD05006 SQBC02769 SQBC02778 SQBC02783 SQBC02794 SQBC02796 SQBC02882 SQBD03846 SQBC02027 SQBC02707 SQBC02709 SQBC02713 SQBD03107 SQBD03374 SQBD03380 SQBD03381 SQBD03408 SQBD03431 SQBD03432 SQBD03435 SQBD03436 SQBD03437 SQBC04272 SQBC04285 SQBC05034 SQBC05035 SQBC05040 SQBC05041 SQBC05043 SQBC05044 SQBC05046 SQBC05047 SQBC05048 SQBC05049 SQBC05050 SQBC05054 SQBC05056 SQBC05057 SQBC05058 SQBC05059 SQBC05063 SQBD00457 SQBD01266 SQBD01550 SQBD01925 SQBD01941 SQBD01961 SQBD02011 SQBD02012 SQBD02014 SQBD02017 SQBD02018 SQBD02020 SQBD02023 SQBD02024 SQBD02025 SQBD02026 SQBD02028 SQBD02029 SQBD02032 SQBD02034 SQBD02038 SQBD02041 SQBD02060 SQBD02126 SQBD02138 SQBD02155 SQBD02165 SQBD02166 SQBD02167 SQBD02168 SQBD02169 SQBD02171 SQBD02173 SQBD02174 SQBD02175 SQBD02177 SQBD02178 SQBD02182 SQBD02183 SQBD02188 SQBD03014 SQBD03144 SQBD03145 SQBD03170 SQBD03179 SQBD03180 SQBD03185 SQBD03191 SQBD03198 SQBD03223 SQBD03239 SQBD03241 SQBD03245 SQBD03510 SQBD03545 SQBD03583 SQBD03584 SQBD03593 SQBD03618 SQBD03620 SQBD03627 SQBD03637 SQBD03644 SQBD03233 SQBD03246 SQBD03410 SQBD03429 SQBD03807 SQBD03839 SQBD03840 SQBD03849 SQBD03853 SQBD03855 SQBC03282 SQBC03373 SQBC03375 SQBC03413 SQBC03458 SQBC03473 SQBC03479 SQBC03482 SQBC03487 SQBC03640 SQBC03660 SQBC03669 SQBC03729 SQBC03731 SQBC03732 SQBC03733 SQBC03735 SQBC03817 SQBC03818 SQBC03819 SQBD01345 SQBD01353 SQBD01354 SQBD01356 SQBD01365 SQBD01407 SQBD01408 SQBD01422 SQBD01423 SQBD01426 SQBC01389 SQBC01390 SQBC01396 SQBC02935 SQBC03281 SQBC04718 SQBC04719 SQBC04721 SQBC04736 SQBC04737 SQBC04824 SQBC04830 SQBC04831 SQBC04832 SQBC04844 SQBD02194 SQBD02242 SQBD02243 SQBD02244 SQBD02247 SQBD02248 SQBD02698 SQBD02803 SQBD02818 SQBD02820 SQBD02830 SQBD02840 SQBD02842 SQBD02844 SQBD02856 SQBD03215 SQBD03234 SQBD03399 SQBD03420 SQBD03434 SQBD03537 SQBD03598 SQBD03599 SQBD03602 SQBD03610 SQBD03611 SQBD03848 SQBD03851 SQBD03856 SQBD03857 SQBD03858 SQBD03859 SQBD03882 SQBD03946 SQBC02596 SQBC05454 SQBD02749 SQBC00644 SQBD03513 SQBD02461 SQBC03044 SQBC05447 SQBC04400 SQBC04461 SQBD02714 SQBC04397 SQBD02748 SQBC01554 SQBC01306 SQBD04049 SQBC05814 SQBC05276 SQBD03903 SQBC02759 SQBC02838 SQBC03007 SQBC04098 SQBC04145 SQBC05279 SQBC05713 SQBC05790 SQBD00825 SQBD01711 SQBD04314 SQBC03824 SQBD02522 SQBC03013 SQBC04387 SQBD03782 SQBD03785 SQBD01563 SQBC01033 SQBD00688 SQBD00214 SQBD00558 SQBD01448 SQBC04283 SQBC01836 SQBC01838 SQBC01839 SQBC01840 SQBC01842 SQBC01844 SQBC01849 SQBD02610 SQBC05468 SQBC01668 SQBC02764 SQBC04093 SQBC04097 SQBC03043 SQBC03047 SQBC03088 SQBD00965 SQBD01596 SQBC01974 SQBC03017 SQBD02605 SQBC01670 SQBC01782 SQBC02793 SQBC02827 SQBC02828 SQBC02839 SQBC02840 SQBC02922 SQBC02923 SQBD01972 SQBD01973 SQBC01242 SQBC04077 SQBC02754 SQBC05381 SQBD04044 SQBD00217 SQBC03953 SQBC04540 SQBC04558 SQBC04562 SQBD03018 SQBC03838 SQBC04488 SQBC04571 SQBC04585 SQBD04027 SQBC01294 SQBC01308 SQBD00650 SQBC03244 SQBC03378 SQBC03050 SQBC03053 SQBD02058 SQBC03869 SQBD02212 SQBD03982 SQBC05792 SQBC05793 SQBC05794 SQBD03936 SQBD03512 SQBD02602 SQBD00811 SQBD00895 SQBC03058 SQBC03060 SQBC03062 SQBC03063 SQBC03064 SQBC02992 SQBD00240 SQBD00824 SQBD00855 SQBD00887 SQBD00529 SQBD02007 SQBD02484 SQBD02485 SQBD02603 SQBD02604 SQBC05231 SQBD03458 SQBD01474 SQBD01783 SQBD03654 SQBC04506 SQBC04149 SQBC03837 SQBD04260 SQBC01241 SQBC01305 SQBD00445 SQBC01166 SQBC01170 SQBC05138 SQBC04490 SQBC05771 SQBC01178 SQBC01168 SQBC03422 SQBC03615 SQBC03625 SQBC03661 SQBC03711 SQBC03846 SQBC04002 SQBC04117 SQBC04118 SQBC04565 SQBC04889 SQBC05127 SQBC05221 SQBC05274 SQBC05481 SQBC05780 SQBC05786 SQBC05789 SQBC05796 SQBD00472 SQBD00478 SQBD00532 SQBD00559 SQBD00812 SQBD00823 SQBD00838 SQBD00859 SQBD00910 SQBD01011 SQBD01463 SQBD01465 SQBD01542 SQBD01567 SQBD01598 SQBD01599 SQBD01781 SQBD02043 SQBD02489 SQBD02523 SQBD02528 SQBD02600 SQBD02601 SQBD02613 SQBD03009 SQBD03017 SQBD03035 SQBD03113 SQBD03129 SQBD03134 SQBD03138 SQBD03168 SQBD03208 SQBD03364 SQBD03516 SQBD03579 SQBD03648 SQBD03658 SQBD03703 SQBD03706 SQBD03751 SQBD03761 SQBD03798 SQBD03816 SQBD03827 SQBD03850 SQBD03861 SQBD03934 SQBD03966 SQBD04000 SQBD04011 SQBD04021 SQBD04022 SQBD04023 SQBD04025 SQBD04029 SQBD04037 SQBD04041 SQBD04046 SQBD04050 SQBD04083 SQBD04178 SQBD04187 SQBD04201 SQBD04209 SQBD04228 SQBD04238 SQBD04246 SQBD04262 SQBD04263 SQBD04264 SQBD04269 SQBD04299 SQBD04300 SQBD04302 SQBD04306 SQBD04309 SQBD04310 SQBD04312 SQBD04316 SQBD04323 SQBD04329 SQBD04333 SQBD04339 SQBD04351 SQBD04387 SQBD04414 SQBD04427 SQBD04455 SQBD04501 SQBD04508 SQBD04518 SQBD04547 SQBD04559 SQBD04563 SQBD04577 SQBD04589 SQBD04596 SQBD04601 SQBD04648 SQBD04649 SQBD04650 SQBD04651 SQBD04656 SQBD04664 SQBD04666 SQBD04673 SQBD04676 SQBD04677 SQBD04678 SQBD04680 SQBD04681 SQBD04682 SQBD04697 SQBD04707 SQBD04769 SQBD04770 SQBD04772 SQBD04773 SQBD04778 SQBD04779 SQBD04780 SQBD04781 SQBD04782 SQBD04783 SQBD04784 SQBD04785 SQBD04786 SQBD04796 SQBD04804 SQBD04805 SQBD04815 SQBD04828 SQBD04830 SQBD04834 SQBD04835 SQBD04889 SQBD04894 SQBD04896 SQBD04897 SQBD04898 SQBD04899 SQBD04901 SQBD04907 SQBD04908 SQBD04909 SQBD04911 SQBD04913 SQBD04915 SQBD04919 SQBD04920 SQBD04926 SQBD04931 SQBD04932 SQBD04933 SQBD04934 SQBD04937 SQBD04938 SQBD04939 SQBD04944 SQBD04946 SQBD04947 SQBD04948 SQBD04949 SQBD04950 SQBD04952 SQBD04953 SQBD04954 SQBD04955 SQBD04956 SQBD04957 SQBD04959 SQBD04960 SQBD04963 SQBD04964 SQBD04965 SQBD04966 SQBD04985 SQBD05044 SQBD05049 SQBD05051 SQBD05066 SQBD05067 SQBD05082 SQBD05086 SQBD05088 SQBD05089 SQBD05090 SQBD05091 SQBD05094 SQBD05095 SQBD05097 SQBD05098 SQBD05100 SQBD05109 SQBD05110 SQBD05112 SQBD05113 SQBD05114 SQBD05115 SQBD05116 SQBD05117 SQBD05120 SQBD05122 SQBD05123 SQBD05124 SQBD05125 SQBD05126 SQBD05127 SQBD05128 SQBD05129 SQBD05130 SQBD05131 SQBD05134 SQBD05135 SQBD05137 SQBD05139 SQBD05140 SQBD05142 SQBD05143 SQBD05144 SQBD05147 SQBD05153 SQBD05154 SQBD05160 SQBD05162 SQBD05169 SQBD05170 SQBD05171 SQBD05173 SQBD05176 SQBD05181 SQBD05183 SQBD05185 SQBD05186 SQBD05195 SQBD05196 SQBD05199 SQBD05202 SQBD05203 SQBD05204 SQBD05206 SQBD05209 SQBD05214 SQBD05216 SQBC03739 SQBC03087 SQBC04372 SQBC04399 SQBC03636 SQBD03504 SQBC03645 SQBC04501 SQBC03096 SQBC03836 SQBC04150 SQBC05235 SQBC05236 SQBC04635 SQBC04650 SQBC01290 SQBD03965 SQBC04392 SQBC04457 SQBC04475 SQBC01830 SQBC01831 SQBC01832 SQBD00391 SQBD00392 SQBD01802 SQBC02918 SQBC02920 SQBD00693 SQBD00694 SQBC04153 SQBC04483 SQBD03283 SQBC03849 SQBC03851 SQBC05083 SQBD02054 SQBC03068 SQBC03074 SQBD00657 SQBD00665 SQBD00233 SQBD00235 SQBD00236 SQBD03953 SQBC05318 SQBD02530 SQBC03715 SQBC01217 SQBD00652 SQBD00653 SQBC01247 SQBC04474 SQBC04499 SQBD00237 SQBD00238 SQBC01182 SQBC01187 SQBC01179 SQBC01181 SQBC03921 SQBD00644 SQBD02477 SQBC03307 SQBC05808 SQBC02807 SQBD02053 SQBC02799 SQBC03595 SQBD01462 SQBC05100 SQBC05102 SQBC01238 SQBC03099 SQBD00189 SQBC04504 SQBC04533 SQBC04535 SQBD03497 SQBD01767 SQBC03650 SQBD00216 SQBD03523 SQBD00655 SQBD00656 SQBC03394 SQBD03766 SQBD03779 SQBD03811 SQBD03969 SQBC03098 SQBD00708 SQBC01246 SQBC01229 SQBC01230 SQBC01232 SQBC01234 SQBC03190 SQBC01189 SQBC01190 SQBC04512 SQBC05282 SQBC02927 SQBC01834 SQBD00817 SQBC03688 SQBC05095 SQBD03110 SQBC03067 SQBC03842 SQBD03502 SQBD00331 SQBD00341 SQBD03503 SQBD03505 SQBD00716 SQBD03287 SQBD04028 SQBD00186 SQBC03644 SQBD03070 SQBD04042 SQBC04893 SQBD01472 SQBD00224 SQBC05433 SQBD03942 SQBC01262 SQBD00685 SQBD02611 SQBC01819 SQBD02015 SQBC01263 SQBD00820 SQBD04043 SQBD03528 SQBD03958 SQBC03418 SQBD00979 SQBD02001 SQBD00696 SQBD00697 SQBD00699 SQBD00701 SQBD00702 SQBD00703 SQBD00704 SQBD00706 SQBD02462 SQBC04492 SQBD00707 SQBC03246 SQBC04478 SQBC04480 SQBD03931 SQBD03961 SQBD00980 SQBD01284 SQBD01945 SQBC03049 SQBD02597 SQBC01701 SQBC01818 SQBC05285 SQBC04564 SQBD03769 SQBD03771 SQBC05287 SQBC05288 SQBC05123 SQBD00205 SQBC04485 SQBC02952 SQBC02995 SQBC03000 SQBC03006 SQBD01601 SQBD03152 SQBD03157 SQBC02813 SQBC02826 SQBC03820 SQBC03638 SQBD04265 SQBC02928 SQBD00228 SQBD00232 SQBD03488 SQBD02598 SQBD02599 SQBD00888 SQBD00893 SQBC02834 SQBC03834 SQBC05809 SQBD03970 SQBC01214 SQBC03844 SQBD03862 SQBD03870 SQBD03874 SQBD03881 SQBD03885 SQBC03897 SQBC04172 SQBD02208 SQBD03887 SQBC01820 SQBD01602 SQBD03128 SQBD04048 SQBD00364 SQBC04394 SQBD02209 SQBD02211 SQBD02612 SQBD01029 SQBD02004 SQBC04078 SQBC04079 SQBD03972 SQBD03984 SQBC05813 SQBD02508 SQBC03033 SQBC05384 SQBC00401 SQBC04898 SQBD02465 SQBD02467 SQBD02475 SQBD01755 SQBD01756 SQBC04595 SQBD04311 SQBD04313 SQBC04510 SQBD00813 SQBD03206 SQBD03479 SQBC02929 SQBC02940 SQBC03289 SQBD01787 SQBD02201 SQBD02202 SQBD01807 SQBC03887 SQBD00977 SQBC04885 SQBC05797 SQBD01561 SQBC04921 SQBD00369 SQBD03786 SQBC05132 SQBC05252 SQBC04511 SQBC04523 SQBC04526 SQBD00648 SQBC04459 SQBC01259 SQBC01260 SQBC02942 SQBC01821 SQBC02941 SQBD00827 SQBD00978 SQBC04532 SQBC04665 SQBD03625 SQBC01161 SQBD03963 SQBC01236 SQBD00361 SQBD00234 SQBC05769 SQBC05799 SQBC03727 SQBD03905 SQBD03923 SQBC02690 SQBC05229 SQBD03904 SQBC03658 SQBC03685 SQBC03831 SQBC04189 SQBD03960 SQBC05112 SQBD00897 SQBC05139 SQBD02531 SQBC03056 SQBC03057 SQBC04103 SQBC04182 SQBC01240 SQBD03153 SQBC04060 SQBD03470 SQBD00372 SQBD00373 SQBC01827 SQBC04489 SQBD00649 SQBC05134 SQBD03498 SQBD03773 SQBD01808 SQBD03133 SQBD00406 SQBD00901 SQBC04681 SQBC04855 SQBC04871 SQBC04876 SQBD02525 SQBD03039 SQBD03284 SQBC03036 SQBC03719 SQBC03845 SQBC04519 SQBC04899 SQBC04900 SQBC04902 SQBC04906 SQBC05135 SQBC05151 SQBC05232 SQBC05290 SQBC05396 SQBD00359 SQBD00362 SQBD00633 SQBD00777 SQBD00829 SQBD00830 SQBD01027 SQBD02009 SQBD02010 SQBD03071 SQBD03490 SQBC04498 SQBC04503 SQBC04530 SQBC03841 SQBD00222 SQBD00646 SQBC04466 SQBC04468 SQBC04450 SQBD00225 SQBC04074 SQBC04549 SQBD02524 SQBC02196 SQBD03762 SQBD02487 SQBD00821 SQBC03724 SQBC03725 SQBC03726 SQBC03659 SQBD03926 SQBD04055 SQBC05099 SQBC03710 SQBC04470 SQBC04471 SQBC05480 SQBD01597 SQBD02481 SQBD02483 SQBC01249 SQBD01005 SQBD01006 SQBD00376 SQBD02514 SQBD00814 SQBD02532 SQBC03032 SQBD01758 SQBD01436 SQBD01788 SQBD03823 SQBD03896 SQBD03898 SQBC03038 SQBC03883 SQBC03902 SQBD04002 SQBD04007 SQBC04062 SQBD00642 SQBD00207 SQBC05128 SQBC05130 SQBD03888 SQBD03889 SQBD03899 SQBD03902 SQBD03979 SQBD04039 SQBD04188 SQBD04053 SQBC04920 SQBC01264 SQBC01273 SQBC04659 SQBC02749 SQBC04219 SQBC04373 SQBD00374 SQBD00375 SQBC04063 SQBD04047 SQBC04897 SQBD00654 SQBD00365 SQBC05234 SQBC04536 SQBD00815 SQBC03847 SQBC05392 SQBD00643 SQBD00380 SQBD01008 SQBC01835 SQBC02921 SQBC05299 SQBC05469 SQBC05479 SQBC01258 SQBC01203 SQBC05815 SQBC01248 SQBD00846 SQBC04119 SQBD00636 SQBC01833 SQBC04812 SQBD00612 SQBD00717 SQBD00724 SQBC02812 SQBC03853 SQBC03855 SQBC03867 SQBD00381 SQBD03814 SQBC05271 SQBD03517 SQBD02016 SQBC03835 SQBC05228 SQBC04486 SQBC01829 SQBD02019 SQBD03975 SQBC04563 SQBD00645 SQBD02458 SQBD02459 SQBC03093 SQBC03095 SQBC05795 SQBD03165 SQBD04301 SQBD00241 SQBD00242 SQBD00826 SQBC05121 SQBC05225 SQBC03821 SQBC03221 SQBC03649 SQBC04560 SQBC04545 SQBC04544 SQBC02750 SQBD03772 SQBD03789 SQBD04261 SQBD04266 SQBC03030 SQBC03944 SQBC04896 SQBD00851 SQBD03515 SQBC04892 SQBD02509 SQBC05093 SQBC03949 SQBC04262 SQBC04542 SQBC04547 SQBD02596 SQBC03664 SQBC03730 SQBC03956 SQBC05387 SQBC05120 SQBD01796 SQBD01799 SQBD01803 SQBD01805 SQBC04473 SQBC04477 SQBC03795 SQBC03054 SQBC03052 SQBD00647 SQBC04508 SQBC05133 SQBD01470 SQBC04456 SQBC04411 SQBD00617 SQBD02552 SQBD02492 SQBC03674 SQBC04114 SQBC04120 SQBC04143 SQBC04168 SQBC04169 SQBC04839 SQBD00497 SQBD00500 SQBD01020 SQBD01021 SQBD01526 SQBD01530 SQBD01536 SQBD01543 SQBD01544 SQBD01874 SQBD01876 SQBD01891 SQBD02073 SQBD02075 SQBD02079 SQBC02040 SQBC03170 SQBC03171 SQBC04495 SQBC01990 SQBC03393 SQBC04469 SQBC05094 SQBC05098 SQBD04432 SQBD03427 SQBD04906 SQBC00984 SQBC00998 SQBC00999 SQBC01000 SQBC01001 SQBC01009 SQBC01019 SQBC01031 SQBC01040 SQBC01041 SQBC01042 SQBC01043 SQBC01193 SQBC01257 SQBC01307 SQBC01366 SQBC01368 SQBC01369 SQBC01370 SQBC01371 SQBC01464 SQBC01733 SQBC02715 SQBC02737 SQBC02801 SQBC02872 SQBC02937 SQBC02947 SQBC02978 SQBC02980 SQBC02983 SQBC02984 SQBC02985 SQBC02986 SQBC02987 SQBC02989 SQBC02990 SQBC02991 SQBC02997 SQBC03663 SQBC03695 SQBC03757 SQBC04003 SQBC04016 SQBC04025 SQBC04026 SQBC04028 SQBC04030 SQBC04033 SQBC04034 SQBC04035 SQBC04038 SQBC04039 SQBC04040 SQBC04061 SQBC04068 SQBC04069 SQBC04070 SQBC04075 SQBD01959 SQBC03854 SQBD03765 SQBC05283 SQBC05294 SQBC01456 SQBC01459 SQBD01019 SQBD01023 SQBD01025 SQBC01748 SQBC01348 SQBC01665 SQBC02994 SQBC03078 SQBC04064 SQBC04156 SQBC04190 SQBC04539 SQBC04570 SQBC04583 SQBC04751 SQBC04865 SQBC04868 SQBC04938 SQBC04951 SQBC04952 SQBC04953 SQBC04955 SQBC04956 SQBC04957 SQBC04959 SQBC04960 SQBC04961 SQBC04962 SQBC04963 SQBC04964 SQBC04965 SQBC04966 SQBC04967 SQBC04968 SQBC04974 SQBC04979 SQBC04980 SQBC05071 SQBC05113 SQBC05116 SQBC05124 SQBC05289 SQBC05324 SQBC05640 SQBC05647 SQBC05711 SQBD00032 SQBD00050 SQBD00092 SQBD00095 SQBD00127 SQBD00135 SQBD00188 SQBD00201 SQBD00204 SQBD00215 SQBD00253 SQBD00255 SQBD00269 SQBD00275 SQBD00279 SQBD00283 SQBD00285 SQBD00286 SQBD00287 SQBD00289 SQBD00290 SQBD00292 SQBD00296 SQBD00297 SQBD00302 SQBD00304 SQBD00305 SQBD00307 SQBD00308 SQBD00322 SQBD00586 SQBD00590 SQBD01041 SQBD01435 SQBD02045 SQBD02195 SQBD02198 SQBD02234 SQBD02235 SQBD02237 SQBD02245 SQBD02249 SQBD02274 SQBD02278 SQBD02281 SQBD02284 SQBD02286 SQBD02287 SQBD02289 SQBD02291 SQBD02292 SQBD02293 SQBD02294 SQBD02295 SQBD02296 SQBD02298 SQBD02299 SQBD02300 SQBD02302 SQBD02304 SQBD02305 SQBD02308 SQBD02309 SQBD02310 SQBD02311 SQBD02312 SQBD02319 SQBD02321 SQBD02323 SQBD02324 SQBC05401 SQBD02975 SQBD02994 SQBC01561 SQBC02001 SQBC02003 SQBC02013 SQBC02024 SQBC02029 SQBC02031 SQBC02032 SQBC02035 SQBC02178 SQBC02181 SQBC02771 SQBC03368 SQBD00611 SQBD02125 SQBD02476 SQBD02804 SQBD02841 SQBD02873 SQBD02875 SQBD02887 SQBD02888 SQBD03459 SQBD03643 SQBD03914 SQBD03916 SQBD03920 SQBD03924 SQBC01637 SQBC04328 SQBC04331 SQBC04333 SQBC04336 SQBC04340 SQBC04341 SQBC04342 SQBC04343 SQBC04345 SQBC04346 SQBC04347 SQBC04348 SQBC04350 SQBC04723 SQBC05412 SQBD00564 SQBD01513 SQBD01604 SQBD03525 SQBD03606 SQBD04276 SQBD04475 SQBD01529 SQBD01638 SQBD01735 SQBD01974 SQBD01983 SQBD01984 SQBD02063 SQBD02067 SQBD02070 SQBD02080 SQBD00640 SQBD00641 SQBD03053 SQBD03237 SQBD03530 SQBD03531 SQBD03535 SQBD03544 SQBD03656 SQBD03662 SQBC02615 SQBC02647 SQBD03391 SQBD04360 SQBC04072 SQBC04147 SQBC04160 SQBD02981 SQBC04199 SQBC04236 SQBC04245 SQBC05346 SQBC05348 SQBC05352 SQBD03005 SQBD03027 SQBD03183 SQBD04289 SQBD04368 SQBD04627 SQBD04708 SQBD04747 SQBB05948 SQBB05960 SQBB06409 SQBC02847 SQBC03561 SQBC03665 SQBC03679 SQBC03681 SQBC03783 SQBC03785 SQBC03787 SQBC03796 SQBC03798 SQBC03861 SQBC04037 SQBC04073 SQBC04095 SQBC04252 SQBC04271 SQBC04275 SQBC04279 SQBC04282 SQBC04286 SQBC04288 SQBC04292 SQBC04298 SQBC04299 SQBC04413 SQBC04445 SQBC04448 SQBC04467 SQBC04517 SQBC04524 SQBC04714 SQBC04720 SQBC05365 SQBC05424 SQBD00360 SQBD02520 SQBD02535 SQBD02536 SQBD02538 SQBD02541 SQBD02555 SQBD02557 SQBD03305 SQBD03461 SQBD03496 SQBD03501 SQBD03511 SQBD03527 SQBD03534 SQBD03550 SQBD03666 SQBD03670 SQBD03684 SQBD03692 SQBD03700 SQBD03702 SQBD03707 SQBD03708 SQBD03717 SQBD03718 SQBD03720 SQBD03722 SQBD03724 SQBD03728 SQBD03729 SQBD03733 SQBD03734 SQBD03735 SQBD03741 SQBD03747 SQBD03749 SQBD03752 SQBD03808 SQBD03023 SQBD03409 SQBC04652 SQBC04602 SQBC04616 SQBC04657 SQBD03444 SQBD03597 SQBC04561 SQBD03169 SQBD03400 SQBD03417 SQBC04412 SQBC03513 SQBC04407 SQBD03371 SQBD03378 SQBD03413 SQBD03415 SQBD03418 SQBD03419 SQBD03424 SQBD03423 SQBD03174 SQBD03192 SQBD03194 SQBD03199 SQBD03201 SQBD03202 SQBD03204 SQBD03074 SQBD03075 SQBD03076 SQBD03084 SQBD03085 SQBD03087 SQBD03097 SQBD03142 SQBD03177 SQBD03188 SQBD03189 SQBD03193 SQBD03200 SQBD03367 SQBD03373 SQBC03535 SQBC04557 SQBD03146 SQBD03147 SQBD03160 SQBD03073 SQBD03082 SQBD03083 SQBD03175 SQBD03186 SQBD03149 SQBD03151 SQBD03184 SQBD03196 SQBD03143 SQBD03148 SQBD03154 SQBD03368 SQBD03383 SQBC01397 SQBC01403 SQBC01404 SQBC03065 SQBC03070 SQBC03071 SQBD03406 SQBD03414 SQBC03756 SQBC03930 SQBC04698 SQBD03307 SQBD03385 SQBD03255 SQBD03449 SQBC04767 SQBC04769 SQBC04771 SQBC04775 SQBC04777 SQBC04626 SQBC04649 SQBC04699 SQBC04790 SQBC04622 SQBC04675 SQBC04704 SQBC04744 SQBC04749 SQBC04185 SQBC04780 SQBD03124 SQBD03139 SQBD03150 SQBD03405 SQBD03416 SQBC04382 SQBC04493 SQBD03411 SQBD03090 SQBD03099 SQBD03173 SQBD03181 SQBD03190 SQBD03197 SQBC04580 SQBC04670 SQBD03135 SQBD03158 SQBD03159 SQBD03161 SQBD03372 SQBD03375 SQBD02995 SQBD03054 SQBD03068 SQBD03096 SQBD03212 SQBD03081 SQBD03086 SQBD03088 SQBD03093 SQBD03103 SQBD03182 SQBC04743 SQBC04748 SQBC04773 SQBC04776 SQBD02865 SQBD02976 SQBD02990 SQBD02993 SQBD02996 SQBD02998 SQBD03003 SQBD03015 SQBD03098 SQBD04394 SQBD04398 SQBD04400 SQBD04438 SQBD04519 SQBD04590 SQBD02977 SQBD03010 SQBD03155 SQBD03216 SQBD03218 SQBC04712 SQBC04725 SQBC04728 SQBC04733 SQBD03396 SQBD03397 SQBD03402 SQBC03598 SQBD03162 SQBD03207 SQBD03166 SQBD03209 SQBD03214 SQBD03004 SQBC04709 SQBC03805 SQBC04696 SQBC04701 SQBD03059 SQBD03176 SQBD03178 SQBD03203 SQBC05714 SQBC05720 SQBD00504 SQBD00784 SQBD00785 SQBD00787 SQBD00790 SQBD00836 SQBD00839 SQBD01564 SQBD01565 SQBD01566 SQBD01571 SQBD01572 SQBD03453 SQBC04688 SQBC04695 SQBC04344 SQBC04451 SQBD03401 SQBD03403 SQBD03412 SQBC04201 SQBC04255 SQBC04381 SQBC04663 SQBC03518 SQBC03565 SQBC03566 SQBC04574 SQBD03132 SQBD03163 SQBD03167 SQBC04686 SQBC04703 SQBC04708 SQBC02005 SQBC02008 SQBC01804 SQBC01805 SQBC01806 SQBC03076 SQBC03407 SQBC03423 SQBC03519 SQBC03690 SQBC03782 SQBC03862 SQBC03882 SQBC03905 SQBC03912 SQBC03914 SQBC03915 SQBC03916 SQBC03936 SQBC03938 SQBC03941 SQBC04840 SQBD01538 SQBD01870 SQBD01886 SQBD02013 SQBD00066 SQBD00068 SQBD00070 SQBD00072 SQBD00078 SQBD00081 SQBC02774 SQBC02781 SQBC04810 SQBC05599 SQBC05603 SQBC05608 SQBC05626 SQBC05589 SQBD04618 SQBD04970 SQBC01940 SQBC01944 SQBC02094 SQBC02999 SQBC03001 SQBC03002 SQBC03014 SQBC03016 SQBC01433 SQBC01887 SQBC01942 SQBC01946 SQBC04083 SQBC04200 SQBC04213 SQBC04214 SQBD03894 SQBC05380 SQBC05382 SQBC05386 SQBC03876 SQBC04202 SQBC03833 SQBC04036 SQBC01440 SQBC01639 SQBC02850 SQBC03024 SQBC05366 SQBC01206 SQBC02121 SQBC03045 SQBC04008 SQBC01215 SQBC03865 SQBC04089 SQBC04113 SQBC02012 SQBC04112 SQBC04781 SQBC04792 SQBD00152 SQBD00157 SQBD00170 SQBC05730 SQBC05740 SQBC05744 SQBC05753 SQBD01879 SQBD02392 SQBD02450 SQBD01333 SQBC00886 SQBC02148 SQBC03040 SQBC03042 SQBD00463 SQBD00492 SQBD00506 SQBD00511 SQBD00514 SQBD00568 SQBD00569 SQBD00573 SQBD00575 SQBD00577 SQBD00609 SQBD00690 SQBD00754 SQBD00816 SQBD00843 SQBD00844 SQBD00849 SQBD00850 SQBD00852 SQBD00853 SQBD00854 SQBD00856 SQBD01385 SQBD01386 SQBD01387 SQBD01388 SQBD01389 SQBD01390 SQBD01391 SQBD01392 SQBD01393 SQBD01394 SQBD01395 SQBD01396 SQBD01397 SQBD01398 SQBD01399 SQBD01401 SQBD01402 SQBD01403 SQBD01404 SQBD01405 SQBD01406 SQBD01411 SQBD01413 SQBD01415 SQBD01557 SQBD01559 SQBD01573 SQBD01752 SQBD01753 SQBD01917 SQBD02064 SQBD02065 SQBD02172 SQBD02267 SQBD04540 SQBC01492 SQBC01703 SQBC02362 SQBC02849 SQBC02864 SQBC02909 SQBC02924 SQBC02931 SQBC02933 SQBD03172 SQBC01138 SQBC01274 SQBC01536 SQBC01559 SQBC01563 SQBC01953 SQBC01976 SQBC03370 SQBC03376 SQBD00098 SQBD00132 SQBD00187 SQBD00197 SQBD00203 SQBD02035 SQBC00622 SQBC01445 SQBC04472 SQBC05377 SQBC05477 SQBC05778 SQBD00035 SQBD02144 SQBD03553 SQBC03270 SQBC03355 SQBC03362 SQBC03363 SQBC03367 SQBC03923 SQBC04126 SQBC04131 SQBC04132 SQBD00592 SQBD02226 SQBD02246 SQBD02253 SQBD02271 SQBD02277 SQBD00094 SQBD01318 SQBD01352 SQBD01355 SQBD04456 SQBD04461 SQBD04463 SQBD04464 SQBD04472 SQBD04476 SQBD04488 SQBD04499 SQBD04642 SQBD04644 SQBC01650 SQBC01655 SQBC02045 SQBC02048 SQBC02050 SQBC02053 SQBC02055 SQBC02100 SQBC02118 SQBC02124 SQBC02265 SQBC02267 SQBC02268 SQBC02271 SQBC02272 SQBC02273 SQBC03253 SQBC03538 SQBC03880 SQBC03886 SQBC03890 SQBC03892 SQBC03893 SQBC03898 SQBC03908 SQBC03910 SQBC04215 SQBC05114 SQBC05286 SQBC05370 SQBD03552 SQBD03557 SQBD03564 SQBD03565 SQBD03614 SQBD03616 SQBD03619 SQBD03621 SQBD03622 SQBD03623 SQBC03884 SQBC04683 SQBC04877 SQBD00669 SQBD00673 SQBD00691 SQBD00739 SQBD03370 SQBD03508 SQBD04018 SQBD04591 SQBC05395 SQBC05406 SQBC05407 SQBC05408 SQBC05410 SQBC05413 SQBC05414 SQBC05415 SQBC05416 SQBC05419 SQBC05423 SQBC05430 SQBC05431 SQBC05432 SQBC05435 SQBC05437 SQBC05438 SQBC05439 SQBC05441 SQBC05443 SQBD03671 SQBD03672 SQBD04162 SQBD04168 SQBD00882 SQBD00899 SQBC00619 SQBC01386 SQBC01391 SQBC01392 SQBC01401 SQBC01443 SQBC01744 SQBC01786 SQBC01920 SQBC02018 SQBC02054 SQBC02056 SQBC02057 SQBC02059 SQBC02060 SQBC02065 SQBC02068 SQBC02069 SQBC02070 SQBC02072 SQBC02187 SQBC02207 SQBC02209 SQBC02212 SQBC02213 SQBC02214 SQBC02217 SQBC02251 SQBC02355 SQBC02590 SQBC02594 SQBC02602 SQBC02607 SQBC02637 SQBC02640 SQBC02745 SQBC02751 SQBC02765 SQBC02770 SQBC02775 SQBC02776 SQBC02782 SQBC02784 SQBC02797 SQBC02798 SQBC02802 SQBC02804 SQBC02805 SQBC02843 SQBC02846 SQBC02877 SQBC02878 SQBC02881 SQBC02885 SQBC02957 SQBC02993 SQBC03089 SQBC03175 SQBC03236 SQBC03239 SQBC03240 SQBC03263 SQBC03276 SQBC03599 SQBC03705 SQBC03747 SQBC03763 SQBC03791 SQBC03848 SQBC03858 SQBC03984 SQBC03985 SQBC03986 SQBC04251 SQBC04414 SQBC04444 SQBC04453 SQBC04460 SQBC04464 SQBC04543 SQBC04548 SQBC04778 SQBC05291 SQBC05295 SQBC05296 SQBC05297 SQBC05300 SQBC05301 SQBC05302 SQBC05304 SQBC05305 SQBC05306 SQBC05308 SQBC05309 SQBC05310 SQBC05312 SQBC05313 SQBC05314 SQBC05316 SQBC05317 SQBC05322 SQBC05339 SQBD00631 SQBD00634 SQBD00678 SQBD00679 SQBD00680 SQBD00682 SQBD00686 SQBD00687 SQBD00764 SQBD00778 SQBD00782 SQBD00783 SQBD00786 SQBD00800 SQBD00802 SQBD00807 SQBD00808 SQBD00809 SQBD00810 SQBD00822 SQBD00842 SQBD00891 SQBD00892 SQBD00900 SQBD00902 SQBD01010 SQBD01185 SQBD01824 SQBD02452 SQBD02629 SQBD02632 SQBD02633 SQBD02636 SQBD02638 SQBD02692 SQBD02694 SQBD02696 SQBD02697 SQBD02719 SQBD02720 SQBD02729 SQBD02756 SQBD02757 SQBD02761 SQBD02766 SQBD02767 SQBD02771 SQBD02772 SQBD02774 SQBD02775 SQBD02790 SQBD02824 SQBD02849 SQBD02853 SQBD02858 SQBD02861 SQBD02868 SQBD02997 SQBD04067 SQBD04352 SQBD04353 SQBD04358 SQBD04530 SQBD02817 SQBD02827 SQBD04315 SQBD04343 SQBD04364 SQBD04367 SQBD04370 SQBD04930 SQBD04972 SQBC04146 SQBC03750 SQBC03745 SQBC04315 SQBC04332 SQBC04337 SQBD00231 SQBC05421 SQBC03210 SQBC03389 SQBC03475 SQBC05655 SQBC05732 SQBC05733 SQBC05715 SQBD04638 SQBC04116 SQBC04682 SQBC05712 SQBC05716 SQBC05719 SQBC05722 SQBC05723 SQBD02521 SQBD04544 SQBD04550 SQBD04592 SQBD04603 SQBD04629 SQBD03879 SQBD03883 SQBD01416 SQBC05672 SQBC02868 SQBC04303 SQBC05393 SQBC01826 SQBC02286 SQBC02291 SQBD04320 SQBD04326 SQBC04809 SQBC05444 SQBC05448 SQBC05453 SQBD00562 SQBC02287 SQBC02289 SQBC04537 SQBC05746 SQBC05747 SQBC05748 SQBC05750 SQBC05757 SQBC05758 SQBC05759 SQBC05761 SQBC04142 SQBC05451 SQBC05452 SQBC04240 SQBC04244 SQBC04795 SQBC05210 SQBC05475 SQBC05688 SQBC05704 SQBC05705 SQBC05706 SQBC05717 SQBC05726 SQBC05727 SQBC05728 SQBD01409 SQBD02136 SQBD02150 SQBD02488 SQBD02623 SQBD02798 SQBD02857 SQBD02884 SQBD04308 SQBD04318 SQBD04330 SQBD04639 SQBD04640 SQBD04729 SQBD04754 SQBC02294 SQBD02806 SQBC04505 SQBC04717 SQBC04740 SQBC04788 SQBC05350 SQBC03797 SQBC03807 SQBC04922 SQBC04924 SQBC04936 SQBC04943 SQBC04949 SQBC04958 SQBC04988 SQBC04990 SQBC04992 SQBC04993 SQBD01906 SQBD02425 SQBD02847 SQBD02878 SQBD02879 SQBD02883 SQBD02885 SQBC04833 SQBC05327 SQBD01457 SQBD01480 SQBD01502 SQBD01515 SQBC02617 SQBC02766 SQBD04641 SQBD04655 SQBC05345 SQBC05353 SQBC05354 SQBD03608 SQBD04575 SQBD04621 SQBD04689 SQBC03771 SQBD03867 SQBD03875 SQBD03878 SQBD04695 SQBC03792 SQBC03800 SQBC03801 SQBC03802 SQBC03803 SQBC03809 SQBC03825 SQBC03826 SQBC03828 SQBC03829 SQBC01265 SQBC01292 SQBC01298 SQBC01998 SQBC01999 SQBC02000 SQBC02004 SQBC02006 SQBC02007 SQBC03103 SQBC03172 SQBC03183 SQBC03187 SQBC03238 SQBC03241 SQBC03266 SQBC03267 SQBC03268 SQBC03269 SQBC03617 SQBD00169 SQBD00603 SQBD00606 SQBD03016 SQBC02248 SQBC02871 SQBD02826 SQBC02851 SQBC02873 SQBC04672 SQBC04954 SQBC05001 SQBC05476 SQBC05478 SQBC05482 SQBC05499 SQBC05500 SQBC05514 SQBC03775 SQBC03799 SQBD04613 SQBC05055 SQBC05503 SQBC05511 SQBC05515 SQBC05517 SQBD03847 SQBD01881 SQBC05512 SQBC04158 SQBD02833 SQBD02839 SQBD01603 SQBC01828 SQBC05367 SQBC03035 SQBC04198 SQBC05108 SQBC04155 SQBC05504 SQBC05505 SQBC04694 SQBC02844 SQBC02848 SQBC04179 SQBD04338 SQBD04359 SQBC02051 SQBC04239 SQBC03633 SQBC03204 SQBC01334 SQBC02149 SQBC05208 SQBC02293 SQBC03173 SQBC03198 SQBC03201 SQBC02038 SQBC02049 SQBC02147 SQBD01038 SQBD02515 SQBD02713 SQBD02891 SQBD02941 SQBD02946 SQBD02955 SQBD02962 SQBC03682 SQBC04385 SQBC01778 SQBC01787 SQBC01801 SQBC01807 SQBC01809 SQBC01810 SQBC01812 SQBC01816 SQBC01817 SQBC01823 SQBC02107 SQBC02648 SQBC00843 SQBC00848 SQBD00121 SQBD00128 SQBC02359 SQBC01815 SQBC01725 SQBC02129 SQBC02131 SQBC02132 SQBC04671 SQBC01814 SQBD01834 SQBC05629 SQBC02377 SQBC02378 SQBC05213 SQBC04007 SQBC03939 SQBC03996 SQBC04013 SQBC04022 SQBC02372 SQBC05214 SQBC02052 SQBC05212 SQBC05211 SQBC01628 SQBC02376 SQBC04706 SQBD04202 SQBD04204 SQBD04208 SQBC03721 SQBC05222 SQBC02144 SQBD02549 SQBD02679 SQBD02710 SQBD02712 SQBD02743 SQBD02744 SQBC02122 SQBC02134 SQBC02142 SQBC02126 SQBC02130 SQBC02133 SQBC02143 SQBC02145 SQBC02146 SQBC04402 SQBC05101 SQBD02455 SQBD02542 SQBD02562 SQBD02563 SQBD02564 SQBD02565 SQBD02566 SQBD02567 SQBD02568 SQBD02569 SQBD02570 SQBD02571 SQBD02572 SQBD02573 SQBD02574 SQBD02575 SQBD02576 SQBD02579 SQBD02584 SQBD02721 SQBD02724 SQBD02752 SQBD02788 SQBD02789 SQBD02802 SQBD02807 SQBD02808 SQBD02809 SQBD02815 SQBD03213 SQBD02577 SQBD00967 SQBC02139 SQBC02151 SQBD02031 SQBC02125 SQBC02138 SQBC02033 SQBC02127 SQBC02028 SQBC02043 SQBC05801 SQBC02034 SQBC02932 SQBC02934 SQBC02938 SQBC04125 SQBC04380 SQBC04388 SQBC04531 SQBC04738 SQBC04782 SQBC04783 SQBC05298 SQBD00676 SQBD02630 SQBD02723 SQBD02828 SQBD02831 SQBD02852 SQBD02860 SQBD03398 SQBD03607 SQBD03831 SQBD00861 SQBD00873 SQBD00918 SQBD01498 SQBD03443 SQBD03845 SQBD05013 SQBD04626 SQBC01542 SQBC01281 SQBC01659 SQBC01667 SQBC01677 SQBC01678 SQBC01728 SQBC03072 SQBC03073 SQBC03097 SQBC03108 SQBC03110 SQBC03112 SQBC03508 SQBC04983 SQBD00630 SQBD03467 SQBD03021 SQBD03022 SQBD03024 SQBD03026 SQBD03030 SQBD03031 SQBD03032 SQBD03037 SQBD03044 SQBD03045 SQBD03067 SQBD03918 SQBC04316 SQBC04323 SQBC04326 SQBC04335 SQBC04338 SQBD01013 SQBD01836 SQBC05718 SQBC05729 SQBC05731 SQBC05734 SQBD01301 SQBD01339 SQBD01341 SQBD01350 SQBD01351 SQBD03549 SQBD03628 SQBD03669 SQBD03678 SQBD03689 SQBD03693 SQBD03695 SQBD03696 SQBD03697 SQBD03714 SQBC00514 SQBC00997 SQBC01296 SQBC01336 SQBC01513 SQBC01514 SQBC01517 SQBC01520 SQBC01521 SQBC01522 SQBC01523 SQBC01525 SQBC01527 SQBC01529 SQBC01530 SQBC01532 SQBC01533 SQBC01534 SQBC01535 SQBC01538 SQBC01841 SQBC01847 SQBC01848 SQBC04731 SQBC04826 SQBC04827 SQBC04841 SQBC04842 SQBC04843 SQBC04847 SQBC04848 SQBC05391 SQBD00493 SQBD00505 SQBD01295 SQBD01330 SQBD01331 SQBD01332 SQBD01334 SQBD01335 SQBD01381 SQBD01382 SQBD01400 SQBD01420 SQBD01421 SQBD01424 SQBD01427 SQBD01429 SQBD01430 SQBD01532 SQBC05250 SQBD01521 SQBD01524 SQBC01303 SQBC01417 SQBC01429 SQBC02762 SQBC02767 SQBC05325 SQBC05320 SQBC05321 SQBC05323 SQBC01691 SQBC02663 SQBC02665 SQBC02673 SQBC03186 SQBD04605 SQBC03157 SQBD01007 SQBD02980 SQBD03029 SQBD04762 SQBD03069 SQBD03137 SQBD03540 SQBD01511 SQBD03603 SQBD03605 SQBD05001 SQBC01627 SQBC01630 SQBC01634 SQBC01638 SQBC01640 SQBC01641 SQBC01642 SQBC01645 SQBC01646 SQBC01647 SQBD02637 SQBD02795 SQBC04487 SQBD02606 SQBD02780 SQBD02797 SQBD03909 SQBD03940 SQBC03631 SQBC03546 SQBD03042 SQBD03591 SQBD03595 SQBD03596 SQBD03632 SQBD03636 SQBD03642 SQBC02779 SQBC05311 SQBC05375 SQBC05379 SQBC05495 SQBC05496 SQBC05498 SQBC05502 SQBC05508 SQBC05509 SQBC05510 SQBC05513 SQBC05518 SQBC05519 SQBC05520 SQBC05521 SQBC05522 SQBC05524 SQBC05525 SQBC05527 SQBC05529 SQBC05531 SQBC05532 SQBC05533 SQBC05535 SQBC05536 SQBC05537 SQBC05538 SQBC05539 SQBC05540 SQBC05541 SQBC05542 SQBC05545 SQBC05546 SQBC05547 SQBC05549 SQBC05552 SQBC05553 SQBC05557 SQBC05558 SQBC05560 SQBD02867 SQBC03718 SQBC03940 SQBC03943 SQBD01220 SQBD01222 SQBD01263 SQBD01273 SQBD01475 SQBD01476 SQBD01477 SQBD01478 SQBD01486 SQBD01493 SQBD01496 SQBD01499 SQBD01500 SQBD01531 SQBD01533 SQBD01609 SQBD01620 SQBD01623 SQBD01637 SQBD01833 SQBD01838 SQBD01867 SQBD01869 SQBD01872 SQBD01952 SQBD02151 SQBD02881 SQBD02984 SQBD02986 SQBD03013 SQBD03036 SQBD03040 SQBD03055 SQBD03056 SQBD03164 SQBD03210 SQBD03217 SQBD03219 SQBD03221 SQBD03225 SQBD03226 SQBD03227 SQBD03228 SQBD03229 SQBD03230 SQBD03231 SQBD03232 SQBD03238 SQBD03242 SQBD03243 SQBD03244 SQBD03247 SQBD03248 SQBD03249 SQBD03250 SQBD03251 SQBD03252 SQBD03253 SQBD03256 SQBD03257 SQBD03259 SQBD03279 SQBD03281 SQBD03288 SQBD03289 SQBD03291 SQBD03292 SQBD03295 SQBD03296 SQBD03300 SQBD03301 SQBD03303 SQBD03306 SQBD03308 SQBD03311 SQBD03315 SQBD03325 SQBD03326 SQBD03327 SQBD03344 SQBD03345 SQBD03348 SQBD03349 SQBD03350 SQBD03351 SQBD03433 SQBD03487 SQBD03489 SQBD03539 SQBD03541 SQBD04721 SQBD04745 SQBC02345 SQBC02353 SQBC02601 SQBC02612 SQBC02620 SQBC02634 SQBC02635 SQBC02636 SQBC02650 SQBC02652 SQBC02654 SQBD00485 SQBD00490 SQBD00503 SQBD00508 SQBD00535 SQBD00548 SQBD00549 SQBD00674 SQBD00675 SQBD00677 SQBD00684 SQBD00795 SQBD00801 SQBD00819 SQBD01506 SQBD01721 SQBD01754 SQBD02592 SQBC01277 SQBC01625 SQBD03901 SQBD02792 SQBD02799 SQBD02811 SQBC05643 SQBC05699 SQBC05807 SQBD04211 SQBD00248 SQBD00249 SQBD00259 SQBC05770 SQBC05772 SQBC04020 SQBC04021 SQBC02958 SQBC05085 SQBC05376 SQBD00001 SQBD00004 SQBD00005 SQBD00007 SQBD00008 SQBD00016 SQBD00022 SQBD00031 SQBD00033 SQBD00034 SQBD00039 SQBD00040 SQBD00042 SQBD00043 SQBD00045 SQBD00046 SQBD00047 SQBD00048 SQBD00049 SQBD00051 SQBD00052 SQBD00053 SQBD00054 SQBD00055 SQBD00062 SQBD00063 SQBD00064 SQBD00065 SQBD00091 SQBD00099 SQBD00108 SQBD00109 SQBD00110 SQBD00112 SQBD00113 SQBD00114 SQBD00115 SQBD00117 SQBD00118 SQBD00119 SQBD00120 SQBD00122 SQBD00123 SQBD00124 SQBD00129 SQBD00136 SQBD00137 SQBD00141 SQBD00142 SQBD00144 SQBD00145 SQBD00146 SQBD00147 SQBD00148 SQBD00149 SQBD00150 SQBD00151 SQBD00243 SQBD00244 SQBD00245 SQBD00246 SQBD00247 SQBD00251 SQBD00254 SQBD00256 SQBD00257 SQBD00261 SQBD00262 SQBD00263 SQBD00264 SQBD00265 SQBD00266 SQBD00267 SQBD00268 SQBD00270 SQBD00271 SQBD00272 SQBD00273 SQBD00274 SQBD00276 SQBD00277 SQBD00278 SQBD00280 SQBD00281 SQBD00282 SQBD00284 SQBD00288 SQBD00945 SQBD00946 SQBD00947 SQBD00948 SQBD00949 SQBD00950 SQBD00951 SQBD00952 SQBD00953 SQBD00954 SQBD00955 SQBD00961 SQBD00962 SQBD00963 SQBD00964 SQBD00966 SQBD00969 SQBD00970 SQBD00971 SQBD00972 SQBD00973 SQBD00974 SQBD00975 SQBD00976 SQBD00981 SQBD00982 SQBD00983 SQBD01004 SQBD01009 SQBD01012 SQBD02594 SQBC02679 SQBC03493 SQBC03511 SQBC03528 SQBC03596 SQBC03620 SQBC03822 SQBC03926 SQBC04018 SQBC04019 SQBC04023 SQBC04029 SQBC04047 SQBC04739 SQBC04927 SQBC04932 SQBC04945 SQBC05275 SQBC05682 SQBC05693 SQBC05774 SQBC05781 SQBD00116 SQBD00163 SQBD00521 SQBD00523 SQBD02279 SQBD03077 SQBD03480 SQBD03520 SQBD03532 SQBD03653 SQBD03657 SQBD03679 SQBD03705 SQBD03737 SQBD03912 SQBD04024 SQBD04031 SQBD04035 SQBD04038 SQBD04040 SQBD04045 SQBD04056 SQBD04060 SQBD04105 SQBD04117 SQBD04133 SQBD04136 SQBD04139 SQBD04146 SQBD04152 SQBD04153 SQBD04156 SQBD04159 SQBD04160 SQBD04164 SQBD04167 SQBD04169 SQBD04170 SQBD04171 SQBD04173 SQBD04175 SQBD04176 SQBD04177 SQBD04179 SQBD04181 SQBD04182 SQBD04183 SQBD04184 SQBD04185 SQBD04186 SQBD04189 SQBD04190 SQBD04192 SQBD04217 SQBD04218 SQBD04219 SQBD04222 SQBD04225 SQBD04226 SQBD04234 SQBD04237 SQBD04239 SQBD04240 SQBD04256 SQBD04257 SQBD04258 SQBD04259 SQBD04268 SQBD04581 SQBD00582 SQBC03587 SQBC03881 SQBD03020 SQBD03647 SQBD04154 SQBD04206 SQBD04223 SQBD04224 SQBD01317 SQBD02265 SQBC04441 SQBD00943 SQBD01469 SQBD01719 SQBC01255 SQBC01288 SQBC01335 SQBC03445 SQBC03618 SQBC03687 SQBC03689 SQBC05765 SQBC05766 SQBD01650 SQBC04223 SQBC04352 SQBC04360 SQBC04572 SQBC04573 SQBC02030 SQBC02037 SQBC02041 SQBC02042 SQBC02046 SQBC02047 SQBC02058 SQBC02061 SQBC02062 SQBC02063 SQBC02064 SQBC02066 SQBC02071 SQBC02073 SQBC02075 SQBC02076 SQBC02077 SQBC02078 SQBC02080 SQBC02082 SQBC02086 SQBC02089 SQBC03085 SQBC03181 SQBC03192 SQBC03220 SQBC03230 SQBC03231 SQBC03233 SQBC03234 SQBC03677 SQBC03693 SQBC03774 SQBC03903 SQBC03904 SQBC03906 SQBC03907 SQBC03942 SQBC03946 SQBC03952 SQBC03957 SQBC03966 SQBD02314 SQBD02315 SQBD02328 SQBD02329 SQBD02330 SQBD02331 SQBD02332 SQBD02334 SQBD02342 SQBD02343 SQBD02344 SQBD00587 SQBD00625 SQBD00841 SQBD00862 SQBD00934 SQBD00936 SQBD00937 SQBD00940 SQBD00941 SQBD00944 SQBD00957 SQBD00958 SQBD00959 SQBD00960 SQBD00968 SQBD04337 SQBD04349 SQBC01359 SQBC01364 SQBC01406 SQBC01407 SQBC01410 SQBC01411 SQBC01412 SQBC01415 SQBC01420 SQBC01424 SQBC01425 SQBC01426 SQBC01427 SQBC04099 SQBC04102 SQBC04793 SQBC04794 SQBC04796 SQBC04821 SQBC04835 SQBC04837 SQBD03041 SQBD03806 SQBD03830 SQBD03832 SQBD03833 SQBD03834 SQBD03835 SQBD03836 SQBD03837 SQBD04430 SQBC02176 SQBC02613 SQBC02628 SQBC03147 SQBD02705 SQBD02747 SQBC02926 SQBC03010 SQBC03025 SQBC03031 SQBC03011 SQBC03012 SQBD02238 SQBD02263 SQBD02264 SQBD02269 SQBD02273 SQBC02081 SQBC02141 SQBC02160 SQBC02629 SQBC03188 SQBC03474 SQBC04502 SQBC04509 SQBD00191 SQBD00194 SQBD00196 SQBD00198 SQBD00200 SQBD00202 SQBD00208 SQBD00209 SQBD00210 SQBD00226 SQBD00250 SQBD00252 SQBD00258 SQBD01015 SQBD01016 SQBD01239 SQBD02072 SQBD02089 SQBD02090 SQBD03617 SQBD03634 SQBD03667 SQBD03826 SQBD03928 SQBD04033 SQBD04036 SQBD04072 SQBD04158 SQBD04161 SQBD04163 SQBD04172 SQBD04174 SQBD04207 SQBD04210 SQBD04441 SQBD04446 SQBD04452 SQBD04453 SQBD04460 SQBD04466 SQBD04477 SQBD04481 SQBD04487 SQBD04489 SQBD04490 SQBD04494 SQBD04495 SQBD04496 SQBD05163 SQBD02545 SQBD02643 SQBD01414 SQBD04429 SQBD04738 SQBD04753 SQBD03686 SQBD03776 SQBD04382 SQBD04434 SQBD04435 SQBD04440 SQBC01432 SQBD04433 SQBC01571 SQBC02137 SQBC02205 SQBC01304 SQBC05434 SQBC05446 SQBC05449 SQBD04763 SQBD04764 SQBD04771 SQBD04808 SQBD04817 SQBD04921 SQBD04927 SQBD04951 SQBD04967 SQBD05041 SQBD05046 SQBD05047 SQBD05059 SQBD05061 SQBD05062 SQBD05063 SQBD05068 SQBD05069 SQBD05070 SQBD05071 SQBC04742 SQBC05284 SQBC05373 SQBC05374 SQBC04324 SQBC04339 SQBC04818 SQBC03243 SQBD00020 SQBC02755 SQBC02810 SQBC02811 SQBC02825 SQBC02829 SQBC02830 SQBC05355 SQBC04170 SQBC05351 SQBD03710 SQBD04030 SQBD04458 SQBC04482 SQBD00313 SQBD00727 SQBD00744 SQBD01303 SQBD01306 SQBD01307 SQBD01310 SQBD01316 SQBD01324 SQBD01325 SQBD01326 SQBD01329 SQBC01349 SQBC02592 SQBC02611 SQBC02614 SQBC02616 SQBC02618 SQBC02621 SQBC02638 SQBC02655 SQBC02956 SQBC02998 SQBC03015 SQBC03051 SQBC03069 SQBC03203 SQBC03213 SQBC03215 SQBC03227 SQBC03354 SQBC03357 SQBC03358 SQBC03461 SQBC03463 SQBC03466 SQBC03572 SQBC03748 SQBC03769 SQBC03866 SQBC03885 SQBC03889 SQBC03895 SQBC03922 SQBC03959 SQBC04011 SQBC04053 SQBC04101 SQBC04105 SQBC04107 SQBC04144 SQBC04162 SQBC04163 SQBC04165 SQBC04166 SQBC04238 SQBC04253 SQBC04354 SQBC04384 SQBC04463 SQBC04484 SQBC04613 SQBC04697 SQBC04707 SQBC04741 SQBC04746 SQBC04820 SQBC04828 SQBC04856 SQBC04857 SQBC04858 SQBC04859 SQBC04861 SQBC04862 SQBC04864 SQBC04866 SQBC04867 SQBC04870 SQBC04883 SQBC04933 SQBC04935 SQBC04942 SQBC04944 SQBC04946 SQBC04947 SQBC05187 SQBC05188 SQBC05190 SQBC05192 SQBC05193 SQBC05194 SQBC05195 SQBC05196 SQBC05197 SQBC05199 SQBC05200 SQBC05201 SQBC05202 SQBC05203 SQBC05205 SQBC05206 SQBC05207 SQBC05209 SQBC05473 SQBC05610 SQBC05763 SQBC05764 SQBC05767 SQBC05779 SQBC05791 SQBC05800 SQBC05802 SQBC05804 SQBC05812 SQBD00714 SQBD01494 SQBD01534 SQBD01607 SQBD01608 SQBD01610 SQBD01611 SQBD01612 SQBD01860 SQBD01868 SQBD01871 SQBD01873 SQBD01883 SQBD01887 SQBD01893 SQBD01894 SQBD01895 SQBD02539 SQBD02718 SQBD02763 SQBD03285 SQBD03407 SQBD03495 SQBD03612 SQBD03681 SQBD03683 SQBD03740 SQBD03813 SQBD03824 SQBD03884 SQBD03941 SQBD03955 SQBD03959 SQBD03967 SQBD03977 SQBD03980 SQBD03986 SQBD03988 SQBD04010 SQBD04012 SQBD04017 SQBD04019 SQBD04020 SQBD04361
FEI Number 2126677
Recalling Firm/
Manufacturer
GE Medical Systems, LLC
3000 N Grandview Blvd
Waukesha WI 53188-1615
Manufacturer Reason
for Recall
Due to a manufacturing issue, certain Tec 6 Plus and Tec 800 series vaporizers can deliver less anesthetic agent than the vaporizer dial setting.
FDA Determined
Cause 2
Process control
ActionAn URGENT MEDICAL DEVICE CORRECTION notification letter dated 11/18/24 was sent to customers. Actions to be taken by Customer/User You can continue to use the vaporizer in accordance with the instructions in the User Reference Manual (URM) and the actions described below: 1. Perform the Output Concentration Check described in Appendix 1 for every impacted vaporizer (see Affected Product Details in this letter). " If a vaporizer passes the Output Concentration Check, you can continue to use that vaporizer. GE HealthCare recommends use of anesthetic agent monitoring during vaporizer use. " If a vaporizer fails the Output Concentration Check, you can continue to use that vaporizer with anesthetic agent monitoring until GE HealthCare corrects the vaporizer. " If a vaporizer fails the Output Concentration Check and anesthetic agent monitoring is not available, remove the vaporizer from service. 2. Please complete and return the attached Medical Device Notification Acknowledgement Response Form to ARC.FMI34139@gehealthcare.com 3. Please ensure all potential staff in your facility are made aware of this safety notification and the recommended actions. 4. Please retain this document for your records. If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
Quantity in Commerce8319 units
DistributionUS Nationwide. Global Distribution.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = CAD
-
-